* Your assessment is very important for improving the workof artificial intelligence, which forms the content of this project
Download www.hik-consulting.pl
Coronary artery disease wikipedia , lookup
Heart failure wikipedia , lookup
Mitral insufficiency wikipedia , lookup
Antihypertensive drug wikipedia , lookup
Electrocardiography wikipedia , lookup
Quantium Medical Cardiac Output wikipedia , lookup
Artificial heart valve wikipedia , lookup
Jatene procedure wikipedia , lookup
Myocardial infarction wikipedia , lookup
Congenital heart defect wikipedia , lookup
Lutembacher's syndrome wikipedia , lookup
Heart arrhythmia wikipedia , lookup
Dextro-Transposition of the great arteries wikipedia , lookup
Phonocardiograph Trainer ST2356 -c hik w. ww Operating Manual on Ver 1.1 pl g. ltin su An ISO 9001 : 2000 company 94-101, Electronic Complex Pardeshipura, Indore- 452010, India Tel : 91-731- 2570301/02, 4211100 Fax: 91- 731- 2555643 email : [email protected] Website : www.scientech.bz Toll free : 1800-103-5050 ST2356 pl g. ltin su on -c hik w. ww Scientech Technologies Pvt. Ltd. 2 ST2356 Phonocardiogram Trainer ST2356 Table of Contents Introduction 4 2. Features 5 3. Technical Specifications 6 4. Panel Controls and Indicators 8 5. Explanation of the Block Diagram 9 6. Human Cardio- Vascular System 10 7. Heart Sounds 13 8. Phonocardiography 17 9. Glossary of Human Cardio- Vascular System Terms 20 10. Operating Instructions 21 11. Real Time Phonocardiogram Display Software 12. Experiments w. ww 1. 22 hik Experiment 1 To observe Phonocardiogram waveforms (PCG) of Subject (Human body) • Experiment 2 To hear the Phonocardiogram signals (PC G) of Subject (Human body) 23 24 Warranty 14. List of Accessories Scientech Technologies Pvt. Ltd. pl 13. g. ltin su on -c • 25 25 3 ST2356 Introduction Phonocardiograph Trainer ST2356 is equipment used to record the sound generated due to mechanical activity of heart i.e. pumping operation of heart and lungs- cardiac sounds, murmurs and respiratory sounds. It provides high fidelity recording of heart sounds, respiratory sounds and murmurs. Heart sound is a physiological phenomenon which provides diagnostic information in a frequency band from 20 Hz to 1000 Hz. The electronic sensor is placed on the subject's chest. PCG unit has a low noise and distortion- free amplifier and the Filter selector and Gain control is adjusted depending on the user's requirement of pitch and intensity. The heart sounds can be heard clearly on the headphone. From the output sockets provided, the PCG unit can be connected to an Oscilloscope or Phonocardiogram data acquisition software, to obtain the Phonocardiogram of subject (human body). pl g. ltin su on -c hik w. ww Scientech Technologies Pvt. Ltd. 4 ST2356 Features • Provides Amplified Phonocardiograph Output • User Selectable Filter Section • On board Variable Gain Control Facility • Separate Test-Points to observe Waveforms after each Block • Buffer stage for Oscilloscope Display • With Real time PCG Acquisition Software • Self contained, Easy to Operate • Specially Designed for Educational Purpose. pl g. ltin su on -c hik w. ww Scientech Technologies Pvt. Ltd. 5 ST2356 Technical Specification Frequency response : 1Hz – 10 KHz CMRR : Better than 80 db Filter ranges : 25Hz - 100 Hz 50Hz - 100 Hz 100Hz - 750 Hz 250Hz - 1.2 KHz Gain adjustment : Variable Audio amplifier for Phonocardiogram output with Headphone out : 230V, ±10%, 50Hz Dimensions (mm) : W325, H90, D255 Weight : 3 ½ Kg approximately pl g. ltin su on -c hik w. ww Power Supply Scientech Technologies Pvt. Ltd. 6 ST2356 pl g. ltin su on -c hik w. ww Figure 1 Scientech Technologies Pvt. Ltd. 7 ST2356 Panel Controls and Indicators Power ‘On/Off’ Rocker switch for supplying power to the instrument. Filter Selection Switch : Control used to select a particular from the various ranges of Band Pass Filters to Pass suitable frequency bands of interest. Filters : 25Hz - 100 Hz : 50Hz-100 Hz : 100Hz - 750 Hz : 250Hz - 1.2 KHz Gain Adjust : Control used to adjust the gain of Main amplifier. ww w. Audio ‘On/Off’ : Toggle switch used to ‘On/Off’ the Audio section. (Power amplifier input + Headphone output) -c hik To Headphone : Headphone jack used to connect headphone to hear the sound. pl g. ltin su on PC Interface : Audio jack to connect the audio cable provided with accessories to user's PC Line-In / Mic- In Port. Scientech Technologies Pvt. Ltd. 8 ST2356 Explanation of Block Diagram 1. Phonocardiogram sensor : The basic transducer for the phonocardiogram is a microphone having the necessary frequency response, generally ranging from below 5 Hz to above 1000 Hz. These microphones sense the external noise along with the heart sounds and murmurs during recording. External noises, such as air conditioners, typewriters, machinery, street noises and voices from radios and televisions, etc, create the vibrations within the same frequency range of the heart sounds, resulting into artifacts. Therefore specially designed noise free rooms are used for PCG recording. 2. Pre-amplifier and Band Pass Filter selective network : A Pre-amplifier with similar response of microphones characteristics is required, which may offer a selective Low Pass Filter to allow the high frequency cutoff to be adjusted for noise and other considerations. ww Various ranges of selective band pass filters allow passing suitable frequency bands of interest, so that particular heart sound frequencies can be recorded. : 25Hz - 100 Hz : 50Hz - 100 Hz : 100Hz - 750 Hz : 250Hz - 1.2 KHz ltin su Main Amplifier : on -c 3 hik w. Filters 4. Audio Section : pl g. The amplifier section amplifies the signal to the desired level. The filtered signal has amplitude of some mill volts hence it is amplified using main amplifier with. Being the Phonocardiogram signals are in audible frequency range, an audio amplifier and speaker are incorporated in Phonocardiograph. By listening to the sounds produced during mechanical action of heart activity, it is also possible to diagnose various cardiovascular disorders. Audio section consists of Audio Power amplifiers with Head phone jack. Headphone jack is used to hear the phonocardiogram signals on headphone. Scientech Technologies Pvt. Ltd. 9 ST2356 Human Cardio- Vascular System The Anatomy of the human heart The heart is basically a hollow muscular pump, which pushes the blood through out the body via the blood vessels. It is located between the lungs and slightly to the left of center. The heart is an involuntary muscle that has approximately seventy to ninety contractions per minute during a restful state. It begins to pump early in the life of a fetus and will continue unceasingly until death. Heart Wall The heart wall is divided into three layers : • Pericardium • Myocardium • Endocardium Chambers : hik w. ww The heart is djvided by a partition or septum into two halves. The halves are in turn divided into chambers. The upper two chambers of the heart are called atria and the lower two chambers are called ventricles. Valves allow blood to flow in one direction between the chambers of the heart. The heart has four distinct chambers. -c Right atrium is the thin-walled area that receives the venous or "used" blood returning to the body by the veins. 2. Right ventricle is the "pump" area of the heart's right side. The atrium dumps the blood into the ventricle where it is then pumped out the pulmonary arteries and to the lungs. 3. Left atrium receives the oxygenated blood returning from the lungs. 4. Left ventricle has the thickest walls of all. It is from this chamber the blood is pumped out of the heart, into the aorta and out to the rest of the body. pl g. ltin su on 1. Heart valves : 1. Tricuspid valve is the one located at the entrance of the right ventricle. It prevents the blood from washing back into the right atrium. 2. Pulmonary semilunar valve is located between the right ventricle and the pulmonary artery. 3. Mitral valve is made of very heavy cusps and is located at the entrance of the left ventricle. This is a powerful valve that closes as the left ventricle begins each of its contractions to ensure the oxygenated blood doesn't re-enter the left atrium. 4. Aortic valve is located, as its name would imply, between the left ventricle's exit and the aorta itself. Scientech Technologies Pvt. Ltd. 10 ST2356 Even though the heart is split up into two distinct halves, these two must work together to function properly. ww Figure 2 w. Heart as a pump : Human heart pl g. ltin su on -c hik Blood from the body that enters the right side of the heart contains carbon dioxide, a gaseous waste the cells produce in creating energy. Blood enters the right atrium through the superior vena cava and inferior vena cava. The atrium fills with blood and then contracts, squeezing the blood through the tricuspid valve into the right ventricle. After the ventricle is filled, pressure forces the tricuspid valve to close and the pulmonic valve, leading to the pulmonary artery, to open. The ventricle contracts and the blood gush through the pulmonary artery and into the lungs. In the lungs, carbon dioxide is removed from the blood and oxygen is added. The oxygenated blood then flows through the pulmonary veins to the left side of the heart. Oxygenated blood from the lungs enters and fills the left atrium. The atrium then contracts, which squeezes the blood through the mitral valve into the left ventricle. After blood fills the ventricle, the mitral valve closes and the aortic valve opens. Blood pours into the aorta and flows through arteries to the body tissues. Scientech Technologies Pvt. Ltd. 11 ST2356 ww -c hik w. Figure 3 Both sides of the heart pump blood at the same time. As the right ventricle contracts and sends blood to the lungs, the left ventricle contracts and squeezes blood out to the body. The heart's cycle of activity has two periods, systole and diastole. Systole occurs when the ventricles contract, and diastole when they relax. One complete contraction and relaxation of the heart muscle makes up one heartbeat. su The Heart's Conduction System : on The heart-rate is a rate at which the heart beats per minute. There are four basic components to the heart’s conduction system Inter-nodal fibre bundles 3. Atrioventricular node (A V node) 4. Atrioventricular bundle pl 2. g. Sinoatrial node (SA node) ltin 1. At the right top corner of the heart there is a special group of excitable cells, called natural pacemaker or sinoatrial node. This natural pacemaker generates electrical impulses, spontaneously. At the lower part of the right atrium there is another mass of specialized group of cells called atrioventriculer node. From the atrioventriculer node a bundle of conducting fibers called bundle of his, passes down to interventriculer septum. A natural pacemaker generates electrical impulses at regular rate. To initiate the heartbeat the action potentials generated by the natural pacemaker or S.A. node gets propagated in all directions along the surface of both atria and atrioventriculer node. This spreads through out the right and left atrium, their wall tissues and results into contraction of atria. Now the waveform reaches to the A.V. node through special nerve fibers which provide the delay in propagation so as to have proper timing Scientech Technologies Pvt. Ltd. 12 ST2356 between the pumping action of atrium and ventricles .During the delay time the atria completes their contraction forcing blood in to ventricles in order to complete their filling. At this point A. V. node initiates an impulse, which gets propagated into the ventricles throughout bundles of his then into left and right bundle branch and further into purkinje fibres causing contraction of both the ventricles and forcing blood in to lungs and the whole body. During the contraction of ventricles the atria complete their filling and to initiate the next heart beat a pacemaker generates another electrical impulse. With the natural pacemaker providing the impulse, the rate of contraction of the heart is maintains and controlled. Normally this action occurs for 60-100 times in a minute, when additional blood is required, the flow must be increased. This is achieved by generating the impulse at faster rate by natural pacemaker. ltin su on -c hik w. ww Heart Conduction System pl g. Figure 4 Heart Sounds For centuries the medical profession has been aided in its diagnosis of certain types of heart disorders by the sounds and vibrations associated with the beating of the heart and the pumping of blood. Blood flow and heart movements in the cardiovascular system create minute vibrations, which radiate to the heart surface. The technique of listening to sounds produced by the organs and vessels of the body is called auscultation, and it is still in common use today. During his training the physician learns to recognize sounds or changes in sounds that he can associate with various types of disorders. In spite of its widespread use, however, auscultation is rather subjective, and the amount of information that can be obtained by listening to the sounds of the heart depends largely on the skill, experience, and hearing ability of the physician. Scientech Technologies Pvt. Ltd. 13 ST2356 The heart sounds heard by the physician through his stethoscope actually occur at the time of closure of major valves in the heart. This timing could easily lead to the false assumption that the sounds which are heard are primarily caused by the snapping together of the vanes of these valves. In reality, this snapping action produces almost no sound; because of the cushioning effect of the blood. The principle cause of heart sounds seems to be vibrations set up in the blood inside the heart by the sudden closure of the valves. These vibrations, together with eddy currents induced in the blood as it is forced through the closing valves, produce vibrations in the walls of the heart chambers and in the adjoining blood vessels. With each heartbeat the normal heart produces two distinct sounds that are audible through the stethoscope, often described as lub-dub. A graphic representation of the heart sounds is known as PCG or Phonocardiogram. There are four basic sounds that occur during the sequence of one complete cardiac cycle. The relationship of these sounds is shown in figure 5 su on -c hik w. ww pl g. ltin Figure 5 The first heart sound (Lub Part) is a low pitch sound. It has a frequency in the range of 30 to 45 Hz. This heart sound occurs at the termination of atrial contraction and at the onset of ventricular contraction. The initial movement of blood from the ventricles, closure of the atrioventricular valves and the resulting abrupt cessation of blood movement into the atria all contribute to the first heart sound. The mitral and tricuspid valve closures are generally attributed with a large contribution to the first heart sound. This heart sound occurs approximately at the time of the ‘QRS’ complex of the ECG complex. The second sound (Dub Part) is high pitch sound. It has a frequency between 50 to 70 Hz. It is caused by the closure of the semi lunar valves that is closure of aortic and pulmonary valves, which release the blood for systemic and pulmonary circulation. These valves close at the end of systole, just before the atrioventricular valves reopen. The second heart sound occurs about the time of the end of the ‘T’ wave of the ECG complex. It is louder than first heart sound. The third heart sound has a very low frequency, normally below 30 Hz. It is sometimes heard, especially in young adults. This sound occurs from 0.1 to 0.2 seconds after the second heart sound. It is due to the rush of blood from the atria into the ventricles, which causes turbulence and some vibration of the ventricular walls. Scientech Technologies Pvt. Ltd. 14 ST2356 This sound actually appears before the atrial contraction. This means, Ih (inrush of blood to the ventricles creates this passive sound. It is generated by a blood pushed only with the venous pressure at the inlets of the atria and pulled by the force of gravity into the ventricles. Actually about seventy percent of blood flow into the ventricles occurs before atrial contraction. The fourth heart sound is called atrial heart sound, which is not audible but may be visible on graphic recording. This heart sound occurs, when the atria actually do contract, squeezing the reminder of the blood into the ventricles. The inaudibility of this heart sound is a result of low amplitude and low frequency of the vibrations. Figure 5 shows the time relationships between the first, second, and third heart sounds with respect to the electrocardiogram, and the various pressure waveforms. Opening and closing times of valves are also shown. This figure should also be compared with figure 6. pl g. ltin su on -c hik w. ww Figure 6 Scientech Technologies Pvt. Ltd. 15 ST2356 Murmurs : If the functioning of the heart is abnormal, then additional sounds called murmurs are heard between the normal heart sounds. The murmurs originate in the cardiovascular system as a result of turbulent blood flow. This turbulence normally produces higher. pitch sounds, which make murmurs distinguishable from the basic heart sounds. The typical conditions in the cardiovascular system, which create blood flow turbulence, include local obstructions, shunts, abrupt changes in diameter, and valve insufficiency or incompetence. The turbulence during the normal functioning of the cardiovascular system does not produce audible sound. Murmurs on the other hand often produce much higher pitched sound that has frequencies in the range of 100 to 600Hz. w. ww Murmurs are generally caused either by improper opening of valves (which requires the blood to be forced through a small aperture) or by regurgitation, which results when the valves do not close completely and allow some backward flow of blood. In either case, the sound is due to high-velocity blood flow through a small opening. Another cause of murmurs can be a small opening in the septum, which separates the left and right sides of the heart. In this case, pressure differences between the two sides of the heart force blood through the opening, usually from the left ventricle into the right ventricle, bypassing the systemic circulation. -c hik Normal heart sounds are quite short in duration, approximately one tenth of a second for each, while murmurs usually extend between the normal sounds. Figure 7 shows a record of normal heart sounds and several-types of murmurs. pl g. ltin su on There is also a difference in frequency range between normal and abnormal heart sounds. The first heart sound is composed primarily, of energy in the 30- to 45-Hz range, with much of the sound below the threshold of audibility. The second heart sound is usually higher in pitch than the rust, with maximum energy in the 50- to 70Hz range. The third heart sound is an extremely weak vibration, with most of its energy at or below 30 Hz. Murmurs, on the other hand, often produces much higher pitched sounds. One particular type of regurgitation, for example, causes a murmur in the 100 to 600-Hz range. Scientech Technologies Pvt. Ltd. 16 ST2356 ww w. Figure 7 Phonocardiography hik su on -c In the early days of auscultation a physician listened to heart sounds by placing his ear on the chest of the patient, directly over the heart. Then someone developed the idea of transmitting heart sounds from the patient's chest to the physician's ear via "a section of cardboard tubing. This was the forerunner of the stethoscope, pl g. ltin The stethoscope (from the Greek word, stethos, meaning chest, and skopein, meaning "to examine")" is simply a device that carries sound energy from the chest of the patient to the ear of the physician via a column of air. There are many forms of stethoscopes, but the familiar configuration has two earpieces connected to a common bell or chest piece. Since the system is strictly acoustical, there is no amplification of sound, except for any that might occur through resonance and other acoustical characteristics. Unfortunately, only a small portion of the energy in heart sounds is in the audible frequency range. Thus, since the dawning of the age of electronics, helps the medical profession by amplifying the heart sounds, with the idea that if the sound level could be increased, a greater portion of the sound spectrum could be heard and greater diagnostic capability might be achieved. In addition, high-fidelity equipment would be able to reproduce the entire frequency range, much of which is missed by the stethoscope. Instruments for graphically recording heart sounds have been more successful. As stated a graphic record of heart sounds is called a phonocardiogram. The instrument for producing this recording is called a phonocardiograph Thus, the sounds and murmurs generated in the cardiovascular system are extremely low in intensity and these sounds and murmurs are commonly recorded using two techniques: Phonocardiography and Pulse wave cardiography. Scientech Technologies Pvt. Ltd. 17 ST2356 To provide adequate heart sound recording, specially designed highly sensitive microphones are used. These microphones sense the external noise along with the heart sounds and murmurs during recording. External noises, such as air conditioners, typewriters, machinery, street noises and voices from radios and televisions, etc, create the vibrations within the same frequency range of the heart sounds, resulting into artifacts. Therefore specially designed noise free rooms are used for PCG recording. The readout of a phonocardiograph is either a high-frequency chart recorder or an oscilloscope. Because most, pen galvanometer recorders have an upper-frequency limitation of around 100 or 200 Hz, photographic or light-galvanometer recorders are required for faithful recording of heart sounds. Although normal heart sounds fall well within the frequency range of pen recorders, the high-frequency murmurs that are often important in diagnosis require the greater response of the photographic device. pl g. ltin su on -c hik w. ww The presence of higher frequencies (murmurs) in the phonocardiogram indicates a possible heart disorder. For this reason, a spectral analysis of heart sounds can provide a useful diagnostic tool for discriminating between normal and abnormal hearts. This type of analysis, however, requires a digital computer with a high-speed analog-to digital conversion capability and some form of Fourier-transform software. A typical spectrum of heart sounds is shown in figure 8. Figure 8 The four basic locations for surface heart sound pickups are shown in figure these areas can slightly change from patient to patient because of heart location, size and orientation. They are described in Table as below: Scientech Technologies Pvt. Ltd. 18 ST2356 Figure 9 Locations Short form Details 1 Aortic area AORT Second intercostals space a right sternal margin 2 Pulmonary 2 Tricuspid area LLSB 4 Mitral area APEX ww S. No PULM Lower left sternal border that is fourth intercostals -c hik w. Second intercostals space a leght sternal margin su on Apex area of fifth intercostals space at the midclavicular line pl g. ltin In addition to these four standard locations, other areas of interest may include the base of the neck on the right side. Erb's area i.e. the third left interspace at sternal margin, and points along the right sternal border. For all locations, it is important to remember that the pick-up be placed over the interspaces and not over the ribs or cartilages. In pulse wave. cardiography, air coupled crystal microphones are used. The information from the pulse wave cardiography is very useful for diagnosing vessel compliance characteristics of the heart at the apex. It is used to record carotid pulse, jugular pulse and apex-cardiogram. A funnel shaped pressure cup is connected to the pulse pick up microphone via rubber tube and applied on the recording site with petroleum jelly to form a good seal. It is held firmly in place with the hand for proper recording. Scientech Technologies Pvt. Ltd. 19 ST2356 Glossary of Human Cardio- Vascular System Terms Artery : A muscular blood vessel that carries blood away from the heart Arrhythmia : Pronounced 00 RIHTH mee uh, is an abnormal heart rhythm. Atrium : One of the chambers of the heart that receives blood directly from a vein Circulatory system : The system of the body responsible for internal transport. Composed of the heart, blood vessels, lymphatic vessels, lymph, and the blood Cardiology : Is the branch of medicine that deals with the diagnosis and treatment of disorders of the heart w. ww Diastole : Pronounced dy AS too lee, is the period of heart activity when the ventricles relax. hik Diastolic pressure : The decreased pressure due to the relaxation of the ventricles is called diastolic pressure. on -c Electrocardiograph (ECG) : Is an instrument used to detect heart damage or diagnose heart disorders. ltin su Heart : The muscular organ composed of cardiac muscle that is responsible for pumping blood throughout the body. Phonocardiograph (PCG) : pl g. Normal sinus rhythm: Normal pumping action of heart generates 60 -100 heartbeats per minute. Is an instrument used to detect heart sounds generated and to find damage or diagnose heart disorders. Septum : the wall dividing the two ventricles. Systole : is the period of heart activity when the ventricles contract. Systolic pressure : The increased pressure due to the contraction of the ventricles is called systolic pressure. Ventricle : One of the muscular chambers of the heart that is responsible for pumping blood from the heart into the arteries. Scientech Technologies Pvt. Ltd. 20 ST2356 Operating Instructions 1. Phonocardiogram sensor must be properly placed on the chest of subject (human body) to hear the heart sounds at various points distinctly. 2. During the monitoring patient should be completely in rest position. 3. Connect the phonocardiogram sensor cable to the instrument only after proper placement of sensor on patient's chest positions to reduce the false sounds generated during placing the sensor. Phonocardiogram Display Software : This software is used to display phonocardiogram signals detected by the ST2356 Phonocardiograph Trainer consists of sensor and hardware kit. pl g. ltin su on -c hik w. ww Scientech Technologies Pvt. Ltd. 21 ST2356 Real Time Phonocardiogram Display Software This Software Contains Two Modules (windows) : 1. Display Window 2. Control Panel Window 1. Display Window : Time Domain Window time domain. 2. - Display the phonocardiogram signals on it in Frequency Domain Window- Display the phonocardiogram signals on it in frequency domain. Control Panel Window: User interface for Start/Stop the Display • Allows setting of Time base and amplitude ranges on Display Window. • Allows Printing of Phonocardiogram signals System Requirements : hik w. ww • : Windows 98/2000/Me/Xp RAM : 64 MB and onwards Space Required : 10MB Screen Resolution : 1024 x 768 pixels ltin su About Copyright : on -c 0/8 pl g. This software is protected by copyright law and international treaties .Unauthorized reproduction or distribution of this software or any portion may result in severe civil and criminal penalties and will prosecute to the maximum extent possible under law. Installation Help : Instructions followed by the user for Real Time Phonocardiogram Display Software. • Insert the Real time Phonocardiogram display Software CD into CD drive. It automatically installs the software in to specify installed location (by default c:\programfiles\installed location) of user's PC. • Execute the Real time Phonocardiogram Display Software Application from start menu! Phonocardiogram Displays Software/PCG.exe Note : To acquire the real time phonocardiogram signals on display window of software, connect PC Interface jack provided on the Trainer to PC's Mic-in/Line in port. Then execute the software. Scientech Technologies Pvt. Ltd. 22 ST2356 Experiment 1 Objective : To observe Phonocardiogram waveforms (PCG) of subject (Human body) Equipments Needed : 1. Phonocardiograph Trainer ST2356 2. Phonocardiogram sensor 3. Oscilloscope Real time PCG Software Procedure : Connect the Phonocardiograph Trainer ST2356 to AC mains. 2. Switch on the trainer by Mains switch. 3. Now place the phonocardiogram sensor on the chest of subject near the heart 4. location. (See figure 6). And then connect the cable to the kit. Observe the Phonocardiogram wave forms at preamplifier block output (TP 1) of trainer on oscilloscope. Select appropriate filter from a multiple steps Band Pass Filters employed here allows only a signal of selected bandwidth to pass to next circuit. Four filters are used for these purposes : 25Hz - 100 Hz -c Filters Bandwidth hik w. 5. ww 1. on 50Hz - 100 Hz su 100Hz - 750 Hz 250Hz - 1.2 KHz ltin g. Observe the signals at selected filter block output (TP3/TP5/TP7/TP9) of trainer. After observing the selected filter output, then observe main amplifier block input TP10/Tpll of trainer. And observe main amplifier output TP12 on oscilloscope. (Adjust gain of the amplifier if required). 7. Connect the main amplifier output TPL2 of trainer to the power amplifier (Audio amplifier) input (TP) 7 to observe the EMG signals sound on speaker. Observe the effects on signals & a cracking sound during muscular contraction. pl 6. Note : Phonocardiogram signals generated can also be observed on PC by connecting PC Interface jack provided on the Trainer to PC's Mic-in/Line in port. Then execute the software. Scientech Technologies Pvt. Ltd. 23 ST2356 Experiment 2 Objective : To hear the Phonocardiogram signals (PC G) of subject (Human body) Equipments Needed : 1. Phonocardiograph Trainer ST2356 2. Phonocardiogram sensor 3. 4. Oscilloscope Caddo 802 or equivalent Headphone Procedure : Connect the Phonocardiograph Trainer ST2356 to AC mains. 2. Switch on the trainer by Mains switch. 3. Now place the phonocardiogram sensor on the chest of subject near the heart location. (See figure 6). And then connect the cable to the kit. 4. Connect Headphone to the Headphone jack (To Headphone) provided on the Trainer. 5. Select Audio ‘On/Off’ switch to On position and hear the Phonocardiogram (PCG) signals sound on headphone. 6. Select appropriate filter from a multiple steps Band Pass Filters employed here allows only a signal of selected bandwidth to pass to next circuit. Four filters are used for these purposes : 25Hz - 100 Hz : 50Hz - 100 Hz : 100Hz - 750 Hz : 250Hz - 1.2 KHz Scientech Technologies Pvt. Ltd. pl Observe the effects on signals & sound. g. ltin su Filters Bandwidth on -c hik w. ww 1. 24 ST2356 Warranty 1) We guarantee the product against all manufacturing defects for 24 months from the date of sale by us or through our dealers. Consumables like dry cell etc. are not covered under warranty. 2) The guarantee will become void, if a) The product is not operated as per the instruction given in the operating manual. b) The agreed payment terms and other conditions of sale are not followed. c) The customer resells the instrument to another party. d) Any attempt is made to service and modify the instrument. The non-working of the product is to be communicated to us immediately giving full details of the complaints and defects noticed specifically mentioning the type, serial number of the product and date of purchase etc. 4) The repair work will be carried out, provided the product is dispatched securely packed and insured. The transportation charges shall be borne by the customer. hik w. ww 3) List of Accessories -c Phonocardiogram Sensor................................................................ 1 No Headphone ..................................................................................... 1 No 3. Stereo pin Cable ............................................................................. 1 No 4. Phonojack Connector Cable............................................................ 1 No 5. Phonocardiogram Analysis S/W CD ............................................... 1 No 6. Mains Cord..................................................................................... 1 No. 7. e-Manual ........................................................................................ 1 No. pl g. ltin su on 1. 2. Scientech Technologies Pvt. Ltd. 25