3.The signal might be interpreted by the device as
a programming signal, leading to inappropriate
reprogramming.
4. A continuous train of electrical impulses
(electrocautery), conducted down the lead can
induce ventricular or atrial fibrillation.
5.High levels of current can pass through the
device, down the lead, and into the endocardium,
causing thermal burns at the lead–tissue interface.
This can raise the pacing threshold.