Download Imaging Essentials Before VAD Placement

Survey
yes no Was this document useful for you?
   Thank you for your participation!

* Your assessment is very important for improving the workof artificial intelligence, which forms the content of this project

Document related concepts

Remote ischemic conditioning wikipedia , lookup

Electrocardiography wikipedia , lookup

Coronary artery disease wikipedia , lookup

Management of acute coronary syndrome wikipedia , lookup

Artificial heart valve wikipedia , lookup

Heart failure wikipedia , lookup

Cardiac surgery wikipedia , lookup

Cardiac contractility modulation wikipedia , lookup

Myocardial infarction wikipedia , lookup

Aortic stenosis wikipedia , lookup

Mitral insufficiency wikipedia , lookup

Echocardiography wikipedia , lookup

Hypertrophic cardiomyopathy wikipedia , lookup

Lutembacher's syndrome wikipedia , lookup

Dextro-Transposition of the great arteries wikipedia , lookup

Quantium Medical Cardiac Output wikipedia , lookup

Atrial septal defect wikipedia , lookup

Ventricular fibrillation wikipedia , lookup

Arrhythmogenic right ventricular dysplasia wikipedia , lookup

Transcript
Imaging Essentials before Ventricular Assist Device (VAD) Placement: What Does
the Surgeon Need to Know?
Michele L. Sumler, M.D.
Assistant Professor of Anesthesiology
Emory School of Medicine
Division of Cardiothoracic Anesthesiology
At the conclusion of this lecture, the participant should be able to:
1. Identify the essential components of an intraoperative TEE examination in a patient who presents for
VAD placement.
2. Review the importance of detecting intra-cardiac shunting in patients undergoing VAD placement.
3. Discuss the clinical significance of valvular regurgitation in patients undergoing VAD placement.
Left ventricular assist devices (LVADs), have emerged as a relevant option
for improving quality of life and survival in patients with end-stage heart failure. The most common
indications include bridge to transplant (BTT), destination therapy (DT) and bridge to recovery (BTR).
Technological advancements have led to the use of continuous flow devices which are improved
compared to previous pulsatile models as far as efficiency, size, implantability, extended support, and
overall patient outcomes are concerned. LVAD implantation improves exercise tolerance and end-organ
dysfunction and can improve hemodynamics. Echocardiography is an important imaging modality used
not only in the diagnosis of heart failure, but in the intraoperative implantation and management of
LVADs. Echocardiographers must develop a systematic approach to echocardiographic assessment of
LVAD implantation.
Pre-Implantation Exam
Prior to beginning the procedure, certain conditions must be addressed that would hinder
proper device function. Since the pump simply pulls blood from whatever chamber where the inflow
cannula is placed, any obstruction to flow into that chamber must be prevented. In the case of an LVAD,
mitral stenosis must be ruled out. Because of the vacuum-like effect, conditions that would cause the
pump to pull-in unwanted blood must also be addressed. Intracardiac shunts, for example, are
accentuated and can cause mixing of deoxygenated blood and hypoxemia. Similarly, aortic insufficiency
can cause the LVAD to pull blood from the ascending aorta, compromising forward flow. Analogous
situations can occur with RVADs on the right side of the heart. In addition to conditions compromising
pump flow, the pre-implantation TEE should identify any thrombus, particularly at the site of inflow
cannula implantation. Similarly, atherosclerotic plaques at the site of outflow cannula placement should
be avoided.
Right ventricular failure is a significant concern when an LVAD is being implanted, and may
complicate management in up to 50% of cases. For this reason, the pre-implantation TEE should
thoroughly evaluate RV function, as well as the severity of tricuspid regurgitation. Several scoring
systems have been developed in order to predict the need for an RVAD following LVAD implantation.
Among them, the “CRITT” score has been shown useful and easy for the intraoperative
echocardiographer to apply.(5)
Cardiac Shunts
Assessment for cardiac shunting prior to VAD insertion is performed to evaluate the risk of
significant right to left shunting and to determine the risk of paradoxical embolization. These shunts may
be at two levels, atrial or ventricular. A ventricular shunt occurs if there is a ventricular septal defect and
these may occur following an acute myocardial infarction. It is important to detect these prior to VAD
insertion, as they would require closure at the time of VAD insertion. A more frequent cardiac shunt is at
the atrial level, due to either a patent foramen ovale (PFO) or atrial septal defect (ASD). Both these
defects have the risk of intracardiac shunting which may cause hypoxia or paradoxical emboli. A PFO or
ASD can be detected and assessed using transesophageal echocardiography, with direct visualization of
the defect on 2D imaging. Color Doppler imaging aids in determining the direction of flow. However, an
atrial shunt is a function of the pressure differential between the right and left atria and this can be a
dynamic parameter. A PFO with right to left shunting may not become apparent until after LVAD
insertion because the LVAD reduces left heart pressures and hence promotes right to left shunting;
particularly in the setting of pulmonary hypertension, which may be common in this subset of patients.
Valvular Dysfunction
Aortic valvular dysfunction is clinically important in the setting of VAD insertion. Uncorrected
aortic regurgitation has a negative impact on forward flow provided by an LVAD due to regurgitation of
VAD flow back into the left ventricular cavity. It is generally recommended that moderate and greater
levels of severity of aortic regurgitation should be corrected at the time of VAD insertion (13). However,
questions still remain as to which intervention is most appropriate. This choice in part depends upon
whether recovery is expected or not, and whether a continuous or pulsatile device is being considered.
Options include over-sewing the valve, repairing the valve or replacing the valve (13, 14). Aortic stenosis
is usually not of such clinical significance compared to aortic regurgitation. However in continuous flow
devices, depending upon the configuration, aortic stenosis may limit forward flow and may need surgical
replacement at the time of VAD insertion.
Tricuspid regurgitation is a relatively common condition in patients being assessed for
mechanical support. This is typically due to right heart failure secondary to chronic elevation of the
pulmonary pressures due to left heart failure. There are numerous complex factors that can influence
the degree of tricuspid regurgitation post LVAD insertion. The decision to correct tricuspid regurgitation
at the time of VAD insertion is a complex one. If significant tricuspid regurgitation is present and it is
expected that improvement of this would result in improved RV function, then intervention on the
tricuspid valve may be warranted (13).
References
1.
Samuels LE, Casanova-Ghosh E, Droogan C. Cardiogenic shock associated with loco-regional
anesthesia rescued with left ventricular assist device implantation. Journal of cardiothoracic
surgery 2010;5:126.
2.
Rose EA, Gelijns AC, Moskowitz AJ, Heitjan DF, Stevenson LW, Dembitsky W, Long JW, Ascheim
DD, Tierney AR, Levitan RG, Watson JT, Meier P, Ronan NS, Shapiro PA, Lazar RM, Miller LW,
Gupta L, Frazier OH, Desvigne-Nickens P, Oz MC, Poirier VL. Long-term use of a left ventricular
assist device for end-stage heart failure. The New England journal of medicine 2001;345:143543.
3.
Naidu SS. Novel percutaneous cardiac assist devices: the science of and indications for
hemodynamic support. Circulation 2011;123:533-43.
4.
Catena E, Tasca G. Role of echocardiography in the perioperative management of mechanical
circulatory assistance. Best practice & research Clinical anaesthesiology 2012;26:199-216.
5.
Atluri P, Goldstone AB, Fairman AS, MacArthur JW, Shudo Y, Cohen JE, Acker AL, Hiesinger W,
Howard JL, Acker MA, Woo YJ. Predicting right ventricular failure in the modern, continuous
flow left ventricular assist device era. The Annals of thoracic surgery 2013;96:857-63; discussion
63-4.
6.
Ammar KA, Umland MM, Kramer C, Sulemanjee N, Jan MF, Khandheria BK, Seward JB, Paterick
TE. The ABCs of left ventricular assist device echocardiography: a systematic approach.
European heart journal cardiovascular Imaging 2012;13:885-99.
7.
Kormos RL, Teuteberg JJ, Pagani FD, Russell SD, John R, Miller LW, Massey T, Milano CA,
Moazami N, Sundareswaran KS, Farrar DJ. Right ventricular failure in patients with the
HeartMate II continuous-flow left ventricular assist device: incidence, risk factors, and effect on
outcomes. The Journal of thoracic and cardiovascular surgery 2010;139:1316-24.
8.
Pretorius M, Hughes AK, Stahlman MB, Saavedra PJ, Deegan RJ, Greelish JP, Zhao DX. Placement
of the TandemHeart percutaneous left ventricular assist device. Anesthesia and analgesia
2006;103:1412-3.
9.
Patel KM, Sherwani SS, Baudo AM, Salvacion A, Herborn J, Soong W, Kendall MC. Echo rounds:
the use of transesophageal echocardiography for confirmation of appropriate Impella 5.0 device
placement. Anesthesia and analgesia 2012;114:82-5.
10.
Dandel M, Weng Y, Siniawski H, Stepanenko A, Krabatsch T, Potapov E, Lehmkuhl HB, Knosalla C,
Hetzer R. Heart failure reversal by ventricular unloading in patients with chronic
cardiomyopathy: criteria for weaning from ventricular assist devices. European heart journal
2011;32:1148-60.
11.
Scalia GM, McCarthy PM, Savage RM, Smedira NG, Thomas JD. Clinical utility of
echocardiography in the management of implantable ventricular assist devices. Journal of the
American Society of Echocardiography. 2000 Aug;13(8):754-63.
12.
Horton SC, Khodaverdian R, Chatelain P, McIntosh ML, Horne BD, Muhlestein JB, et al. Left
Ventricular Assist Device Malfunction: An Approach to Diagnosis by Echocardiography. Journal
of the American College of Cardiology. 2005;45(9):1435-40.
13.
Rao V, Slater JP, Edwards NM, Naka Y, Oz MC. Surgical management of valvular disease in
patients requiring left ventricular assist device support. Ann Thorac Surg. 2001 May 1,
2001;71(5):1448-53.
14.
Bryant AS, Holman WL, Nanda NC, Vengala S, Blood MS, Pamboukian SV, et al. Native Aortic
Valve Insufficiency in Patients With Left Ventricular Assist Devices. Ann Thorac Surg. 2006
February 1, 2006;81(2):e6-8.