Download FTIR Instrumentation

Survey
yes no Was this document useful for you?
   Thank you for your participation!

* Your assessment is very important for improving the workof artificial intelligence, which forms the content of this project

Document related concepts

Ellipsometry wikipedia , lookup

Optical tweezers wikipedia , lookup

Super-resolution microscopy wikipedia , lookup

Holography wikipedia , lookup

Photoacoustic effect wikipedia , lookup

Rutherford backscattering spectrometry wikipedia , lookup

Spectral density wikipedia , lookup

Confocal microscopy wikipedia , lookup

Diffraction grating wikipedia , lookup

Night vision device wikipedia , lookup

Phase-contrast X-ray imaging wikipedia , lookup

Light wikipedia , lookup

Spectrum analyzer wikipedia , lookup

Thomas Young (scientist) wikipedia , lookup

Anti-reflective coating wikipedia , lookup

Chemical imaging wikipedia , lookup

Harold Hopkins (physicist) wikipedia , lookup

Reflecting telescope wikipedia , lookup

Microscopy wikipedia , lookup

Very Large Telescope wikipedia , lookup

Retroreflector wikipedia , lookup

Ultrafast laser spectroscopy wikipedia , lookup

Magnetic circular dichroism wikipedia , lookup

Nonlinear optics wikipedia , lookup

Gamma spectroscopy wikipedia , lookup

AFM-IR wikipedia , lookup

Optical coherence tomography wikipedia , lookup

Opto-isolator wikipedia , lookup

Vibrational analysis with scanning probe microscopy wikipedia , lookup

X-ray fluorescence wikipedia , lookup

Astronomical spectroscopy wikipedia , lookup

Ultraviolet–visible spectroscopy wikipedia , lookup

Transcript
FTIR_Instrumentation_short.pdf, RL 2008
FTIR Instrumentation
Adopted from the FTIR lab instruction by H.-N. Hsieh, New Jersey Institute of
Technology: http://www-ec.njit.edu/~hsieh/ene669/FTIR.html
1. IR Instrumentation
Two types of instrumentation are used to obtain infrared spectra:
•
•
Dispersive spectrophotometers, which use a monochromator to produce an
infrared spectrum one resolution element at a time.
Michelson interferometers, which use a moving mirror to create an
interference pattern, or interferogram, from which all resolution elements are
determined simultaneously.
2. FTIR Spectroscopy History
Chemical infrared spectroscopy emerged as a science in the 1880s. In the 1890s, A.
A. Michelson, to further his studies of the speed of light, invented the interferometer.
In the early 1940s, chemical infrared spectroscopy was still an immature scientific
field. But with commercial development of the optical null dispersive
spectrophotometer, later that decade, chemical infrared spectroscopy came into
widespread use. Dispersive instruments proved the tremendous value of infrared
analysis, and soon became the mainstay of organic characterization laboratories.
In 1949 astrophysicist Peter Fellgett used an interferometer to measure light from
celestial bodies and produced the first Fourier transform infrared spectrum. But for
many years, only a few advanced research groups with access to large, expensive
computers and with personnel able to wait up to 12 hours to transform an
interferogram into a spectrum used Fourier transform infrared (FTIR) spectroscopy.
For two decades more, FTIR spectrometers were limited to studying problems not
solvable with dispersive techniques.
In the late 1960s when microcomputers able to do the Fourier transform became
available, commercial FTIR spectrometers appeared. The 1966 development of the
Cooley-Tukey algorithm, which quickly does a Fourier transform (the Fast Fourier
Transform or FFT), was also instrumental in the commercialization of FTIR
spectrometers. However, the first FTIR spectrometers were large and expensive, and
were found primarily in a few well-to-do research labs.
Gradually, technology reduced the cost, increased the availability, and enhanced the
capabilities of FTIR spectroscopy systems. The performance-to-price ratio provided
by FTIR spectrometers today was unthinkable only a decade ago.
1
FTIR_Instrumentation_short.pdf, RL 2008
3. Theory of Interferometer Operation
The optical system in an FTIR spectrometer is very simple: the interferometer
requires two mirrors, an infrared light source, an infrared detector, and a beamsplitter.
fixed mirror
beam splitter
collimator
moving
mirror
detector
source
focussing
mirror
Figure 1. The Michelson interferometer
The beamsplitter is the heart of the interferometer. Essentially a half-silvered mirror,
the beamsplitter reflects about half of an incident light beam while simultaneously
transmitting the remaining half. One half of this split light beam travels to the
interferometer's moving mirror while the other half travels to the interferometer's
stationary mirror. The two mirrors reflect both beams back to the beamsplitter where
each of the two beams is again half reflected and half transmitted. Two output beams
result: one travels to the detector as the other travels to the source.
When the two beams return to the beamsplitter, an interference pattern, or
interferogram, is generated. This interference pattern varies with the displacement of
the moving mirror, that is, with the difference in pathlength in the two arms of the
interferometer. The interference pattern, detected by the infrared detector as variations
in the infrared energy level, is what ultimately yields spectral information.
4. More Interferometer Theory
An example of how light is affected in an interferometer further illustrates
interferometer operation. Consider the interference pattern generated when
2
FTIR_Instrumentation_short.pdf, RL 2008
monochromatic light, such as that produced by a helium-neon (HeNe) laser, is the
light source.
An idealized beamsplitter:
- neither generates nor absorbs radiation.
- transmits half of all incident light.
- reflects half of all incident light.
All light energy that enters the interferometer from the source either returns back to
the source or travels to the infrared detector. As the moving mirror scans back and
forth, more or less than half of the total light energy that originally entered the
interferometer reaches the detector. All light that does not reach the detector reaches
the source instead.
For example, if both interferometer mirrors are equidistant from the beamsplitter, the
two light beams travel equal distances. As each beam returns to the beamsplitter, each
is again split and then recombined with half of the light from the other interferometer
arm. The two component beams that make up the output beam passed to the detector
are completely in phase with each other and, therefore, interfere constructively. The
two component beams that make up the output beam passed to the source are
completely out of phase with each other and, therefore, interfere destructively. In this
case, all of the original light energy goes to the detector and none goes to the source.
By contrast, if the moving mirror is displaced one-quarter wavelength of the HeNe
laser radiation (to produce optical retardation of one-half wavelength), light returning
from the interferometer's moving mirror reaches the beamsplitter one-half wavelength
out of phase with respect to the light returning from the fixed mirror. This results in
complete destructive interference in the detected output beam so no light energy
reaches the detector, and all light returns to the source.
Figure 2. Constructive (left) and destructive (right) interference.
As the moving mirror scans back and forth, the detector observes light and dark bands
corresponding to partial and total constructive and destructive interference in the
detected output beam.
In this monochromatic light example, the interference signal the detector receives
produces a cosine interferogram. This wave goes from full intensity at the point where
both mirrors are equidistant from the beamsplitter to zero intensity when the moving
mirror has traveled one-quarter wavelength. Full intensity is again achieved when the
moving mirror has traveled one-half or multiples of one-half wavelength. Similarly,
zero intensity occurs at each multiple of one-quarter wavelength of mirror travel.
3
FTIR_Instrumentation_short.pdf, RL 2008
An infrared source is polychromatic, emitting light over a broad range of frequencies.
Each frequency still yields a unique cosine signal: the resulting interferogram
represents the sum of all cosine waves generated by each individual infrared
frequency. However, only at the point where both mirrors are equal distances from the
beamsplitter does complete constructive interference occur--simultaneous
constructive interference of all wavelengths. At any other point, only some
wavelengths interfere constructively while others do not.
The Fourier transform resolves the frequency and intensity of each cosine wave in the
interferogram. In other words, this algorithm converts the measured .intensity-versusmirror-displacement signal (the interferogram) into a plot of intensity versus
frequency (a spectrum).
Figure 3. Spectrum with three
frequencies with different amplitudes
(top), interferograms for the three
separate frequencies (middle), and
Interferograms for the three mixed
frequencies.
Δx
Δx
Δx = 0
5. Advantage
5.1 Multiplex Advantage (Fellgett's Advantage)
An interferometer does not separate light into individual frequencies before
measurement.
This means each point in the interferogram contains information from each
wavelength in the input light. In other words, if 8,000 data points along the
interferogram are collected, each wavelength in the input light is sampled 8,000 times.
By contrast, a dispersive spectrophotometer that measures 8,000 individual points
across a spectrum samples each wavelength only once.
5.2 Throughput Advantage (Jacquinot's Advantage)
The simple optical path of the interferometer (no slits and fewer optical elements)
means more energy gets to the sample than is possible with dispersive
4
FTIR_Instrumentation_short.pdf, RL 2008
spectrophotometers. This means more energy reaches the detector, increasing the
spectrum's potential signal-to-noise ratio.
Together the multiplex and throughput advantages allow an FTIR spectrometer to
obtain a high-quality infrared spectrum in a fraction of the time needed to get the
same spectrum on a dispersive instrument.
Also, to increase the resolution of dispersive instruments, the slits through which light
must pass are narrowed, thereby decreasing energy throughput. In an FTIR
spectrometer, resolution is increased by lengthening the moving mirror stroke length
with no decrease in energy throughput. As wavelength resolution increases, the
advantages of interferometric versus dispersive measurements increase.
5.3 Frequency Precision (Conne's Advantage)
With dispersive instruments, frequency precision and accuracy depend on: 1)
calibration with external standards and 2) the ability of electromechanical
mechanisms to uniformly move gratings and slits during and between scans.
By contrast, the interferometer has an internal frequency standard, generally a heliumneon laser. Also, both mirror movement and detector sampling are clocked by the
interferometer fringes from the laser's monochromatic light. All frequencies in the
output spectrum are calculated from the known frequency of the laser light.
An FTIR spectrometer easily achieves frequency precision and accuracy of better than
0.01 wavenumbers. This means that spectra collected with an FTIR spectrometer can
be quantitatively compared whether they were collected five minutes or five years
apart.
5.4 Constant Spectral Resolution
In an FTIR spectrometer, the resolution of the measured spectrum is the same for all
frequencies, not varied throughout the spectrum as is often true with dispersive
instruments.
6. Resolution
In a dispersive spectrophotometer, resolution depends on slit width and on the
reciprocal of the grating's linear displacement. As the slits are narrowed, the range of
frequencies passed to the detector decreases, resulting in higher resolution. In an FTIR
spectrometer, resolution is increased by increasing optical retardation, that is, by
increasing how far the moving mirror moves.
A relationship between resolution and optical retardation can be empirically derived
by considering a spectrum made up of two sharp bands. Each band is at a distinct
frequency and generates a sinusoidal signal. The total signal at the detector is the sum
of the two signals.
If the two spectral frequencies are relatively far apart, the resulting signals are quite
different and only a short movement of the mirror is needed to measure a complete
5
FTIR_Instrumentation_short.pdf, RL 2008
cycle of the pattern made by the addition of the two waveforms. On the other hand, if
the two frequencies are closer together, the interferometer needs to scan a larger
distance to measure a full cycle of the addition pattern.
The relationship between spectral line spacing and the interferogram is
R = 1/d
where R is the spectral resolution in wavenumbers and d is the optical retardation in
centimeters. Thus, the length of mirror travel (retardation) must be doubled to reduce
instrument resolution by half, or, in other words, to improve instrument resolution by
a factor of two.
Several different criteria are used to measure FTIR spectrometer resolution. These
include the Rayleigh criterion and the full-width-at-half-height (FWHH) method.
Measuring instrument resolution with the FWHH method involves collecting an
absorption spectrum with very sharp peaks, such as that obtained from low-pressure
CO at about 5 torr pressure.
fixed mirror
beam splitter
electronics
moving mirror
source
aperture
sample
detector
Figure 4: Left: Outer view of the Bruker IFS 66v/S FTIR spectrometer. (A) Sample chamber, (B)
Interferometer chamber, (C) Laser compartment, (D) Electronic compartment and vacuum control, (E)
Detector chamber. Right: Schematic internal view. Bold lines mark the optical pathway.
6