TECHNICAL SPECIFICATION FOR PURCHASE OF EQUIPMENT IN AYUB
TEACHING HOSPITAL..
ANESTHESIA MACHINE WITH VENTILATOR
QUANTITY: 08Nos

Anesthesia 3 gas unit (O2, N2O, Air)

Digital Display of Pipeline & cylinder pressure on screen.

Central Gas Driven Unit with Central Gas hoses

Pin index cylinder yokes for Oxygen & Nitrous Oxide (One for each)

Connections for large Oxygen and Nitrous Cylinders (One for each)

Power outlet with socket outlet
Anesthesia Ventilator (STANDARD)

The ventilator shall be capable of ventilating Adult & Pediatric patient down to 2 kg
body weight.

Screen Size 6“Color or more.

Auto-clavable breathing system.

Heated and dehumidified breathing system.
Modes of Ventilation

Volume Preset time cycle ventilator (IPPV mode)

Pressure Controlled Mode.

Manual / Spontaneous.
Control Ranges:

Frequency: 4 to 60 bpm or better

PEEP: 0 to 20 cmH2O

Trigger sensitivity: 2 to 15 I/min or better
Continuous monitoring of O2 concentration, Frequency, Tidal Volume, Minute Volume,
Pressure

Breathing mode selection (Standby / Volume / Spontaneous & pressure)

Fresh Gas Decoupling System

Built in oxygen monitoring

Automatic Flow sensor & O2 Calibration

Battery Back up More than 30 minutes.

Power Supply 220VAC, 50Hz

Supplied with Adult and Pediatric Anesthesia breathing circuit
Optional :

Isoflorane & Sevoflorane both vaporizer will be supplied with each unit.
SUCTION MACHINE HEAVY DUTY

Heavy duty Mobile Suction Unit with twin jars of capacity up to 4 or 5 liters each

Auto-clavable.

10X bacterial filter

Footswitch

Complete aspiration set

220V/50Hz. I Phase

Trolley with lockable wheels.

Nozzles of different sizes or shapes.
CAUTERY MACHINE

QUANTITY: 10 Nos
QUANTITY 08 Nos.
Provide PURE CUT, BLEND (1), BLEND (2),BLEND(3), COAG1(spray), COAG2(contact)and
AUTO BIPOLAR, FOOTSWITCH BIPOLAR, Digital 400Watt.

Percentage output digital display shows on front panel.

It is controlled by Microprocessor which is responsible for controlling and monitoring of all
the functions and features in unit.

Output values (CUT, COAG, BIPOLAR) are remained in the memory after switching off.
These values will be come up to the display at turning on the power.

Automatic bipolar output without using footswitch or with hand.

This can be hand or foot operated.

A bipolar output power profile that provides good resolution to permit small changes in power
levels for delicate work yet has sufficient maximum output power capability to satisfy the
requirements of the more demanding Procedures.

Spray Coagulation provides the Trans Urethral Resection (T. U. R) and Endoscopic
procedures in urology. ( 400D)

Accessibility to parts and ease of maintenance that is unsurpassed.

Versatility, Reliability.

Designed for safety / Designed for Ease of use.
Adaptive REM Monitoring/PCS ( Patient Control System):
After the patient return electrode is applied, and the initial impedance is between 5 ohms and 135
ohms, the “adaptive” REM system measures the contact impedance and uses it as a reference value
or above 135 ohms, the REM alarm sounds and the Mono-Polar output is disabled.
Technical Specifications:
Output Power
 PURE CUT
400W AT 300-500Ω
 BLEND 1
270W AT 300-500Ω
 BLEND 2
200W AT 300-500Ω
 BLEND 3
160W AT 300-500Ω
 SPRAY COAG
70W AT 300-500Ω
 BIPLAR COAG
70W AT 100Ω
Operating Frequency: 400KHZ + 10%
 Input Voltage: 110/220V, 50/60 Hz
 Enclosure Leakage Current
0.1 µA
 Patient Leakage Current
7 µA
 Earth Leakage Current
03.358 mA
 RF Leakage Current
43 mA
 Protection Class / Type: Class I / Type CF
STANDARD ACCESSORIES:
No:
Standard Accessories
Qty
1
Footswitch 2 pedal with cable
01
2
Footswitch 1 pedal with cable
01
3
Reusable Foot switching pencil with cable
01
4
Disposable 2 push button monopolar pencil with cable and
01
electrode
5
Bipolar forceps and Cable asset
01
6
Reusable patient plate with cable for adult
01
7
Disposable patient plate for adult
01
8
Set of electrodes (6Pieces) for reusable and disposable handle
01
9
Power cable
01
OPERATION TABLE
QUANTITY 08 Nos.

Multi Sectional Electro-Hydraulic operated operation table

Sturdy enough to hold a weight of Minimum 180kg in the normal position

Minimum 3-sections or more excluding head rest

The leg section will be of split led type.

Fixed column with removable with removable top

Table top should be of radiolucent material

The mattress covers with washable, antistatic and of fire resistant material

The movements shall be controlled through handset controller

Side rails for accessories attachment

Trendelenberg and reverse trendelenberg position and lateral left right position
between Flex and reflex

Raising/ lowering of the head section

Raising/ lowering of the leg section.
STANDARD ACCESSORIES FOR TABLE: (All Stainless steel)

Shoulder Supports Set of 2

Universal Lateral Supports

Lateral Supports 2 Nos.

General Purpose Head Support

Anesthetic Screen Frame

Knee Crutches 2 Nos.

Raised Arm Board 1 ,

Arm Board

Kidney Elevator

Cassette Holder

Knee Rest

Douche Tray Hanging Type

Douche Tray Sliding Type
TRACTION TABLE
QUANTITY 02 Nos.

Multi Sectional Electro-Hydraulic operated operation table

Full length radiolucent top with integral X-ray cassette tunnel

Slim line column offset relative to tabletop and base.

Trendelenberg and reverse trendelenberg position and lateral left right position
between Flex and reflex

Raising/ lowering of the head section

Raising/ lowering of the leg section.

Longitunanal displacement .

600mm wide top plus optional width extenders

Interchangeable general and orthopedic surgery trunk sections

Auto self-locking, interchangeable head, leg and infill sections

Covers are made of impact resistant acrylic capped ABS

Battery status indicators for primary and back-up batteries on handset and table base

Standby control panel on table column

Tabletop break offset from column

A choice of general and orthopedic trunks allow the table to be easily configured for
general, orthopedic and specialist procedures

Offset column break to improves the imaging flexibility

Extended height range for optimum clinician comfort in use

200kg or more patient weight

Orthopedic docking cart facilitates rapid and secure table set-up

interchangeable head and leg sections offer

Corded handset for simple, intuitive control

prevent fluid and dirt ingress
TABLE MAIN FEATURES INCLUDES THE FOLLOWINGS:

Modular orthopedic

Easy adjustment

Easy configuration

Override control

range of traction and orthopedic accessories easily secure to the operating table

Dynamic Hip Screw (DHS)

Tibia and lateral femoral nailing

orthopedic system - kits and spares

To provides excellent patient support and allows positioning without the need for manual
handling.

Country of origin must be USA, Japan, UK & Europe.