Download Flow Cytometry and Sorting, Part 1

Survey
yes no Was this document useful for you?
   Thank you for your participation!

* Your assessment is very important for improving the workof artificial intelligence, which forms the content of this project

Document related concepts

Cytoplasmic streaming wikipedia , lookup

Confocal microscopy wikipedia , lookup

Amitosis wikipedia , lookup

Organ-on-a-chip wikipedia , lookup

Circular dichroism wikipedia , lookup

Transcript
Flow Cytometry
Principles & practice of “Fluorescence
Spectroscopy in Biological Diagnosis
& Research”
Dr.Hekmatimoghaddam
Assistant professor of pathology
Definitions

Flow Cytometry
– Measuring properties of cells in flow

Flow Sorting
– Sorting (separating) cells based on properties
measured in flow
– Also called Fluorescence-Activated Cell
Sorting (FACS)
Basics of Flow Cytometry
•Cells in suspension
Fluidics
•flow in single-file through
•an illuminated volume where they
Optics
•scatter light and emit fluorescence
•that is collected, filtered and
Electronics
•converted to digital values
•that are stored on a computer
Fluidics
Need to have cells in suspension flow in
single file through an illuminated volume
 In most instruments, accomplished by
injecting sample into a sheath fluid as it
passes through a small (50-300 µm)
orifice

Flow Cell
Injector
Tip
Fluorescence
signals
Focused laser
beam
Sheath
fluid
Fluidics
When conditions are right, sample fluid
flows in a central core that does not mix
with the sheath fluid
 This is termed Laminar flow

Fluidics

The introduction of a large volume into a
small volume in such a way that it
becomes “focused” along an axis is called
Hydrodynamic Focusing
Fluidics - Differential Pressure
System
Use air (or other gas) to pressurize sample
and sheath containers
 Use pressure regulators to control
pressure on each container separately

Fluidics - Differential Pressure
System
Sheath pressure will set the sheath
volume flow rate (assuming sample flow
is negligible)
 Difference in pressure between sample
and sheath will control sample volume
flow rate
 Control is not absolute - changes in
friction cause changes in sample volume
flow rate

Fluidics - Differential Pressure
System
C. Göttlinger, B. Mechtold, and A. Radbruch
Fluidics - Particle Orientation
and Deformation
“a: Native human
erythrocytes near the
margin of the core stream
of a short tube (orifice).
The cells are uniformly
oriented and elongated by
the hydrodynamic forces
of the inlet flow.
b: In the turbulent flow
near the tube wall, the
cells are deformed and
disoriented in a very
individual way. v>3 m/s.”
V. Kachel, et al. - MLM Chapt. 3
Fluidics - Flow Chambers
Flow
through
cuvette
(sense in
quartz)
H.B. Steen - MLM Chapt. 2
Flow Cell
Injector
Tip
Fluorescence
signals
Focused laser
beam
Sheath
fluid
Optics
Need to have a light source focused on
the same point where cells have been
focused (the illumination volume)
 Two types of light sources

– Lasers
– Arc-lamps
Optics - Light Sources

Lasers
– can provide a single wavelength of light (a
laser line) or (more rarely) a mixture of
wavelengths
– can provide from milliwatts to watts of light
– can be inexpensive, air-cooled units or
expensive, water-cooled units
– provide coherent light
Optics - Light Sources

Arc-lamps
– provide mixture of wavelengths that must be
filtered to select desired wavelengths
– provide milliwatts of light
– inexpensive, air-cooled units
– provide incoherent light
Optics - Forward Scatter
Channel
When a laser light source is used, the
amount of light scattered in the forward
direction (along the same axis that the
laser light is traveling) is detected in the
forward scatter channel
 The intensity of forward scatter is
proportional to the size, shape and optical
homogeneity of cells (or other particles)

Forward Angle Light Scatter
Laser
FALS Sensor
Optics - Side Scatter Channel
When a laser light source is used, the
amount of light scattered to the side
(perpendicular to the axis that the laser
light is traveling) is detected in the side
or 90o scatter channel
 The intensity of side scatter is proportional
to the internal structure and granularity of
cells (or other particles)

90 Degree Light Scatter
Laser
FALS Sensor
90LS Sensor
Optics - Light Scatter

Forward scatter tends to be more sensitive
to surface properties of particles (e.g., cell
ruffling) than side scatter
– can be used to distinguish live from dead cells

Side scatter tends to be more sensitive to
inclusions within cells than forward scatter
– can be used to distinguish granulated cells
from non-granulated cells
Fluorescence Detectors
Laser
Freq
FALS Sensor
Fluorescence
Fluorescence detector
(PMT3, PMT4 etc.)
Optics - Filter Properties
Long pass filters transmit wavelengths
above a cut-on wavelength
 Short pass filters transmit wavelengths
below a cut-off wavelength
 Band pass filters transmit wavelengths
in a narrow range around a specified
wavelength

– Band width can be specified
Standard Long Pass
Filters
Light Source
520 nm Long Pass Filter
Transmitted Light
>520 nm
Light
Standard Short Pass Filters
Light Source
575 nm Short Pass Filter
Transmitted Light
<575 nm
Light
Standard Band Pass Filters
630 nm BandPass Filter
White Light Source
Transmitted Light
620 -640 nm Light
Optics - Filter Properties
When a filter is placed at a 45o angle to a
light source, light which would have been
transmitted by that filter is still transmitted
but light that would have been blocked is
reflected (at a 90o angle)
 Used this way, a filter is called a dichroic
filter or dichroic mirror

Dichroic Filter/Mirror
Filter placed at 45o
Light Source
Transmitted Light
Reflected light
Optics - Filter Layout
To simultaneously measure more than one
scatter or fluorescence from each cell, we
typically use multiple channels (multiple
detectors)
 Design of multiple channel layout must
consider

– spectral properties of fluorochromes being
used
– proper order of filters and mirrors
Common
Laser
Lines
350
300 nm
457 488 514
400 nm
500 nm
610 632
600 nm
700 nm
PE-TR Conj.
Texas Red
PI
Ethidium
PE
FITC
cis-Parinaric acid
Example Channel Layout for
PMT
Laser-based Flow
Cytometry
4
Flow cell
PMT
Dichroic
Filters
3
PMT
2
Bandpass
Filters
PMT
1
Laser
Optics - Detectors

Two common detector types
– Photodiode
 used for strong signals when saturation is a
potential problem (e.g., forward scatter detector)
– Photomultiplier tube (PMT)
 more sensitive than photodiode but can be
destroyed by exposure to too much light
Summary of Part 1
•Cells in suspension
Fluidics
•flow in single-file through
•an illuminated volume where they
Optics
•scatter light and emit fluorescence
•that is collected, filtered and
Electronics
•converted to digital values
•that are stored on a computer
Typical Research Cytometer
(Coulter 753) (1980s)
$200-300,000
Detectors
Lasers
Fluidics
Computers
Laser Power Supply
Typical Clinical
Cytometer
Computer System
Detector &
Mechanical Fluidics
$90-120,000
Clinical Applications Of
Flow Cytometric Analysis
FlowCytometric(immunophenotypic)
Classification Of Leukemias
Immunophenotyping
CD4
CD2
0
0
3
0
Count
7
0
1
1
0
1
5
Immunophenotyping
.1
1
10
Log FITC
100
1000
CD4/CD8 Quadstats
100
1
2
45%
1
10
2%
27% 3
4
.1
26%
.1
1
10
100
Log FITC Fluorescence (CD8)
1000
The Cell Cycle
G2
M
G1
S
G0
Quiescent cells
Normal Cell Cycle
G2
M
G0
DNA Analysis
G1
G0G1
s
C
o
u
n
t
G2 M
s
0
200
400
600
4N
2N
DNA content
800
1000
300
225
DNA index 1.21
150
Aneuploid peak
0
75
Counts
DNA Analysis
0
200
400
600
PI Fluorescence
800
1000
112
112
150
150
Reticulocyte Analysis
RMI = 34
Count7 5
Cou7nt5
RMI = 0
R4 R3 R2 R1
0
0
37
37
R4 R3 R2 R1
.1
1
10
100
log Thiazole Orange
1000
.1
1
10
100
1000
log Thiazole Orange
Light Scatter Gating
1000
Side Scatter Projection
Forward Scatter Projection
Neutrophils
800
1000
100
50
600
200
40
20
Monocytes
400
30
15
Lymphocytes
0
8
200
Forward Scatter
Scale
0
200
400
600
800
90 Degree Scatter
1000