Download Thyroid Stimulating Hormone

Survey
yes no Was this document useful for you?
   Thank you for your participation!

* Your assessment is very important for improving the workof artificial intelligence, which forms the content of this project

Document related concepts

Hypothalamus wikipedia , lookup

Signs and symptoms of Graves' disease wikipedia , lookup

Growth hormone therapy wikipedia , lookup

Hypopituitarism wikipedia , lookup

Hypothyroidism wikipedia , lookup

Hyperthyroidism wikipedia , lookup

Transcript
Thyroid
Stimulating
Hormone
3rd Generation
Ultra Sensitive Assay
Table of Contents
Introduction
2
Thyroid function tests (TFTs) is a collective term
for blood tests used to check the function of the
thyroid. TFTs may be requested if a patient is
thought to suffer from hyperthyroidism (overactive thyroid) or hypothyroidism (underactive
thyroid), or to monitor the effectiveness of either
thyroid-suppression or hormone replacement
therapy. It is also requested routinely in conditions
linked to thyroid disease, such as atrial fibrillation,
depression and anxiety disorder. Thyroid-stimulating hormone (TSH) test is the first line test of
choice for evaluating thyroid function.
Introduction
3
TSH
Materials and Methods
4
Results and Interpretation
5
Conclusion
Notes
vantix® DIAGNOSTICS
Thyroid
Stimulating
Hormone
Third Generation
Ultra Sensitive Assay
Dr. Peter Swarbrick
Dr. Neil Blair
Dr. Ian Beech
Dale Weeks BSc
vantix® DIAGNOSTICS
Abbey Barns Unit 4
Duxford Road
Ickleton, Cambridge CB10 1SX
UK +44 (0) 1799 533160
Published in June 2014
Find out more at vantixdx.com
2 | vantix® DIAGNOSTICS
TSH
Materials and Methods
Thyroid-stimulating hormone (TSH) stimulates the
thyroid gland to synthesize and secrete thyroid
hormone. TSH serum measurements are used to
detect primary hypo- and hyperthyroidism. TSH is
the most sensitive test for thyroid hormone function. TSH is produced in the pituitary gland. The
production of TSH is controlled by Thyrotropin-releasing hormone (TRH), which is produced in the
hypothalamus. TSH levels may be suppressed by
excess free T4 thyroid hormone in the blood. The
capability of a TSH assay to distinguish between
normal and subnormal concentrations is essential
for thyroid testing to ultimately yield information
that is useful to clinicians treating patients with
subnormal TSH/FT4 concentrations.
Seventy five (75) sera samples were supplied by
a large research hospital with known TSH values
ranging from 0.02 mlU/L through 120 mlU/L. The
samples included normal, elevated or suppressed
individual results. TSH assays were performed simultaneously on the vantix® IBB II platform1. Using
the vantix® flow cell chip each sample was aspirated over the biosensors which had anti TSH antibody bound to the surface. In the next conjugate
step, dilutions of second TSH antibody conjugated
to HRP, was run over the sensors for 80 seconds
forming a sandwich immunoassay with TSH. The
final step was to flow vantix® measuring solution
over the sensors, which generates the mV electrical signal in the presence of HRP on the sensor.
The quantified amount of analyte was determined
by taking the mV electrical signal for each sample
and comparing it to standard curve values derived
from known calibrators using the biosensor/assay
system. A log logistic curve fitting was employed.
The performance goal of the vantix® TSH assay
is quantification at a lower value of 0.02 mlU/L
through an upper limit of 75 mlU/L. This is considered 3rd generation or ultrasensitive performance.
Both of these terms relate to the evolution of the
TSH test. Over time, increasingly sensitive and specific TSH tests have been developed and adopted.
Most laboratories now use the 3rd generation/
ultrasensitive TSH test as their “TSH test.” Third
generation assays have far superior sensitivity in
the subnormal TSH range compared to the second-generation assay and about 100 times more
sensitive than the first-generation TSH test.
vantix® DIAGNOSTICS | 3
Results and
Interpretation
Conclusion
The vantix® TSH assay showed a 0.9851 R2 correlation to the results achieved by a large research
hospital’s testing of the same samples.
The vantix® system has demonstrated in clinical
samples strong agreement with the reference
lab method at a large research hospital. The 75
samples were across the dynamic range of a 3rd
generation TSH assay comprised of both diseased
and healthy patients.
TSH Correlations Log Logistic
Clinical Range
0.3 - 3
2
R = 0.9851
3rd Generation Sensitivity
vantix® measurement log (mlU/I)
80.00
70.00
60.00
50.00
40.00
30.00
20.00
10.00
—
4 | vantix® DIAGNOSTICS
—
0.02
0.1
—
0.00
—
—
0.00
1
10
100
Large research hospital’s measurement (mlU/I)
Notes
The IBB II system is a semi-automated liquid
aspirating reader and disposable chip that is the
primary development platform for the vantix®
point of care system
1
How the vantix® System Will Work
vantix® DIAGNOSTICS | 5
www.vantixdx.com