Download display

Survey
yes no Was this document useful for you?
   Thank you for your participation!

* Your assessment is very important for improving the workof artificial intelligence, which forms the content of this project

Document related concepts

Loudspeaker wikipedia , lookup

Stray voltage wikipedia , lookup

Dynamic range compression wikipedia , lookup

Ground (electricity) wikipedia , lookup

Three-phase electric power wikipedia , lookup

Pulse-width modulation wikipedia , lookup

Ground loop (electricity) wikipedia , lookup

Public address system wikipedia , lookup

Rectifier wikipedia , lookup

Negative feedback wikipedia , lookup

Islanding wikipedia , lookup

Power inverter wikipedia , lookup

Alternating current wikipedia , lookup

Scattering parameters wikipedia , lookup

Bode plot wikipedia , lookup

Schmitt trigger wikipedia , lookup

Voltage optimisation wikipedia , lookup

Two-port network wikipedia , lookup

Heterodyne wikipedia , lookup

Power electronics wikipedia , lookup

Buck converter wikipedia , lookup

Tube sound wikipedia , lookup

Tektronix analog oscilloscopes wikipedia , lookup

Oscilloscope types wikipedia , lookup

Switched-mode power supply wikipedia , lookup

Resistive opto-isolator wikipedia , lookup

Mains electricity wikipedia , lookup

Regenerative circuit wikipedia , lookup

Audio power wikipedia , lookup

Metadyne wikipedia , lookup

Oscilloscope history wikipedia , lookup

Amplifier wikipedia , lookup

Wien bridge oscillator wikipedia , lookup

Opto-isolator wikipedia , lookup

Transcript
IFCN standards1 for digital recording of
clinical EEG Verification2 3
Square-Wave Calibration Test
Summary: Square-wave signals must be recorded at the beginning, using a series of 100 µV square waves of each
recording, each 1–2 s long.
1.
Set up function generator to output a square wave with a 1mVp-p amplitude and a frequency of 1 Hz. A
voltage divider circuit must be used to step down the voltage to 100 µV. The schematic is shown below.
CHANNEL A
Signal Generator
Voltage Divider
Electrodes
R1
R3
10
V1
1mVdc
5k
R2
91
R4
5k
0
CHANNEL B
Electrodes
Right Leg Driver
R8
R5
5k
5k
0
0
R9
5k
Figure 1 - Square Wave Calibration Test Setup
2.
3.
Apply the square wave signal to the electrodes of one channel. Ensure that the right leg driver is connected
to ground and the unused channel inputs are also grounded.
Use an oscilloscope to ensure that the square wave is obtained at the output of the analog board with the
required gain.
1
Standards Document: https://edge.rit.edu/content/P08050/public/References/IFCN_Standards_106_259.pdf
2
Omitting noise tests because low noise components have been chosen.
3
Omitting 60 Hz notch filter test as it will be implemented in software.
Anti-Aliasing High-Filter Test
Summary: Prior to sampling at 256 samples/sec., an anti-aliasing high filter at 70 Hz must be used, with a roll-off of
at least 12 dB/octave.
1.
2.
3.
Set up function generator to output sine wave with a 1mVp-p amplitude and a frequency of 1.0 kHz. The
gain of the amplifier may need to be adjusted via the potentiometer to avoid clipping.
Apply the sine wave signal to the electrodes of one channel. Ensure that the right leg driver is connected to
ground and the unused channel inputs are also grounded.
Using an oscilloscope verify that the gain is approximately 0dB as measured at the output. See figure
below for expected frequency response of amplifier.
80
Amplifier Total Response (Gain vs. Freq.), Differential
40
0
SEL>>
-40
DB(V(C27:2))
180d
0 dB Gain at 1 kHz
-60 dB/dec anti-aliasing filter
-0d
Amplifier Total Response (Phase vs. Freq.), Differential
-193d
-387d
1.0mHz
3.0mHz
P(V(C27:2))
10mHz
30mHz
100mHz
300mHz
1.0Hz
3.0Hz
10Hz
30Hz
100Hz
300Hz
Frequency
Figure 2 – Gain (TOP) and Phase (BOTTOM) vs. Frequency of Amplifier
Low-Filter Test
Summary: Whenever possible the low filter should be set to 0.16 Hz or less for recording.
1.
Set up function generator to output sine wave with a 1mVp-p amplitude and a frequency of 2.0 mHz. The
gain of the amplifier may need to be adjusted via the potentiometer to avoid clipping.
2. Apply the sine wave signal to the electrodes of one channel. Ensure that the right leg driver is connected to
ground and the unused channel inputs are also grounded.
3. Using an oscilloscope verify that the gain is approximately 0dB as measured at the output. See figure
below for expected frequency response of amplifier.
1.0KHz
80
Amplifier Total Response (Gain vs. Freq.), Differential
40
0
SEL>>
-40
DB(V(C27:2))
180d
-0d
0 dB Gain at 2 mHz
Amplifier Total Response (Phase vs. Freq.), Differential
-193d
-387d
1.0mHz
3.0mHz
P(V(C27:2))
10mHz
30mHz
100mHz
300mHz
1.0Hz
3.0Hz
10Hz
30Hz
100Hz
300Hz
Frequency
Figure 3 - Gain (TOP) and Phase (BOTTOM) vs. Frequency of Amplifier
Electrode Impedances Test
Summary: Electrode impedances should be kept below 5 kΩ.
1.
2.
3.
4.
5.
6.
Set up a DC Power Supply to output 1 VDC,p-p.
Ground both the inputs to the electrode buffer amp of channel A as shown in figure below.
Ground right leg driver input and inputs of channel B as shown.
Connect 5 kΩ. resistor in series with output of DC power supply as shown, (Note the output impedance of
DC power supply).
Measure voltage across the electrode as shown.
The electrode impedance is calculated as follows:
Equation 1 - Calculation of Electrode Impedance
Relectrode 
Rsource  5k
1  Velectrode
 Velectrode 
1.0KHz
Channel A Connections
Channel B and RLD Connections
Voltage Source Resistance
+ Known Resistance
Electrode Contacts (unknown)
Value Used
4
2
+
3
U1A
-
Voltage Generator
+
5k
R5
TLC277/101/TI
OUT
Velectode
1Vdc
-
V-
Relectrode
5k
1
0
R9
8
V2
R8
Right Leg Driver
5k
V+
Rknown + Rsource
CHANNEL B Electrodes
0
0
0
Electrode Buffer
Amplifier
Figure 4 - Test Setup to Measure Electrode Impedance
Preamplifier Input Impedances
Summary: Preamplifier input impedances must be more than 100 MΩ.
TLC277: 10,000,000 MΩ.
INA114: 100,000 MΩ.
Common Mode Rejection Ratio (CMRR) Test*
Summary: The common mode rejection ratio must be at least 110 dB for each channel measured at amplifier input.
1.
2.
3.
4.
5.
Apply a 0.1mV differential signal to one channel of the amplifier.
Ground the inputs of the other channel and the right leg driver.
Measure the differential voltage gain of the amplifier using an oscilloscope.
Apply a 0.1mVcommon signal to one channel of the amplifier.
Measure the common mode voltage gain of amplifier using an oscilloscope.
6. The CMRR is calculated as:
Equation 2 - CMRR Calculation
CMRR  20  log( DifferentialGain
)
CommonModeGain
*The CMRR for the INA114 is typically 106 dB at a gain of 10.
*The CMRR for the TLC277 is typically 84 dB.
Additional Amplifier Verification Tests
Analog Board Gain / Redundancy Test
Summary: The analog board gain must be 4500V/V (±400V/V). Two analog boards will be made with identical
components and verified to operate similarly.
1.
Apply the same 0.1mV differential signal to one channel of the both the amplifier boards.
2.
3.
4.
5.
For both the analog boards, ground the inputs of the other channel and the right leg driver.
Measure the differential voltage gain of both the amplifier using an oscilloscope.
Ensure the gain value is 4500V/V (±400V/V).
Measure the common mode voltage gain of amplifier using an oscilloscope.
Power Consumption / Analog Board Lifetime Test
Summary: The power consumption of the circuit must be 150mW +/- 10mW. The circuit must also operate
continuously for a time of at least 24 hours.
1.
2.
Use a DC Power Supply to power the analog board by setting the DC output of the supply to 6V. Do not
turn on the output of the power supply yet.
Connect a 100Ω resistor in series with the positive output of the power supply as shown in the figure
below.
Operating Voltage / Battery Voltage
Summary: The circuit must operate on four AA alkaline batteries for a total battery voltage of 5.6V+/- 0.4V.
Simulated EEG Waveform Test
Summary: Using a simulated EEG waveform we will input it to the amplifier and make sure the amplifier is
functioning properly.
Motion Artifact Test
Summary: The amplifier will be exposed to motion artifacts.