* Your assessment is very important for improving the workof artificial intelligence, which forms the content of this project
Download Analytical Chemistry CHEM 21122
Survey
Document related concepts
Transcript
2/9/2005 Course content & Exam Sampling and chemometrics (6 h) Analytical spectroscopy: (6 h) Analytical electrochemistry: (6 h) Analytical separation : (6 h) Analytical instrumentation : (6h) Prof. K. A. S. Pathiratne(06 L) Dr . Russel De Silva (12 L) Dr. Sri Skandaraja (12 L) Analytical Spectroscopy End of semester exam : two hours, five Questions. Expect in-class exams + take home assignments too. Spectroscopy Latin: “spectrum” — ghost or spirit Greek: “σκοπειν” — to look carefully We do not look directly at the matter (atoms, molecules) but at its “ghost” We observe the interaction of electromagnetic radiation with different properties of the matter. Each type of spectroscopy gives a different picture of the matter → the spectrum. Spectroscopy is the study of the interaction between matter and radiated energy. The spectrum is the variation of the intensity of the radiation as a function of the frequency or wavelength Spectroscopy is any procedure that uses the interaction of Electromagnetic Radiation (EMR) with matter to identify and/or to estimate an analyte. Quantitative Analysis Band spectra molecules ions atoms Mixtures solid liquid gas solutions Qualitative Analysis line spectra 1 2/9/2005 Electromagnetic radiation (E-M radiation or EMR) is a form of energy that exhibits wave-like behavior as it travels through space. (Dual nature of light) EMR has both electric and magnetic field components, which oscillate in phase perpendicular to each other and perpendicular to the direction of energy propagation. Properties of light waves Wavelength (): Distance from one wave peak to the next. Units: m, cm, m, nm Frequency (): Number of peaks that pass a given point per second. Units: Cycles/second or s-1 or Hertz (Hz) Wave number : Number of waves per cm. Units: cm-1 Electromagnetic radiation consists of discrete packets of energy, which we call photons. Photons are the particles of light or the quanta of light and each photon carries the energy, E (Joule). E hυ where h is the Planck’s constant (=6.626x10-34 J s) E = hυ = h c = hc υ λ Where c is the speed of light in vaccum ( = 2.998 x 108 m/s) The greater the energy, the higher the frequency and wave number and the shorter the wavelength Electromagnetic Spectrum 2 2/9/2005 3 2/9/2005 Transmission and Absorption Transmittance and Absorbance Transmittance and Absorbance For a T = 0.7, There are two quantities that relate the change in the intensity of EMR before (I0) , and after, (I) interaction with matter. 1. Transmittance, T, is simply defined as “the fraction of light that reaches a detector after passing through a sample” T I I 0≤T≤1 The percent transmittance, %T, is simply 100 T I %T x 100 0 < %T < 100 I 2. Absorbance, defined as: I 0<A<2 A = log ( ) A= log T I %T = 70% then, A =-log T = 0.155 When no light is absorbed , P=P0 , A=0, and %T = 100% If 90% of the light is absorbed, 10% is transmitted, thus T = I/I0 = 10/100 = 0.1 and A = -log T = - log 0.1 = 1 Beer–Lambert law the Beer–Lambert law, also known as Beer's law or the Lambert– Beer law or the Beer–Lambert–Bouguer law (named after August Beer, Johann Heinrich Lambert, and Pierre Bouguer) relates the absorption of light to the properties of the material through which the light is travelling. 4 2/9/2005 Beer-Lambert Law Example Absorbance of light of a particular wavelength is directly proportional to: Concentration, c, mol/L (M) of absorbing species in the sample (A c) Path length of light, (b or l), in cm through the sample (A b ) Find the absorbance and transmittance of a 0.00240 M solution of a substance with molar absorptivity of 313 M-1 cm-1 at of 520 nm when measured in a cell of 2.00 cm pathlength. Thus, A = bc This relation is commonly known as Beer-Lambert law (M-1 cm-1 )is called the molar absorptivity or molar absorption coefficient (or Extinction coefficient) “Absorbance of 1 M solution measured in a cell of 1 cm pathlength” , is characteristic for each substance at a particular wavelength, . A = bc = (313 M-1 cm-1)(2.00 cm)(0.00240 M) = 1.50 - log T = A log T = -A T = 10logT = 10-A = 10-1.50 = 0.0316 Absorbance & Concentration Molar absorptivity ( Beer’s Law A c) Increasing Concentration ε = 2212 (mol L-1)-1 cm-1 ε = 0.02 (mol L-1)-1 cm-1 Increasing absorbance KMnO4 Lambert’s law (A b ) 5 2/9/2005 Transmittance and Concentration Absorbance & concentration T=10-A =10- bc A = bc = mx Slope = b 1 Transmittance, T Absorbance, A y Transmittance decreases exponentially as concentration increases 0.5 Solution with T = 50% 0 Concentration 12.5% 25% P0 = 100% 32 50% Concentration not zero Deviation from Beer’s law At high concentrations, the proximity of molecules to each others alters their absorptivity. Scattering and changes in refractive index will also contribute. the error by possible stray light is significant . At low concentrations, error due to sensitivity of the instrument is significant . If the absorbing molecule participates in a concentration- dependent chemical equilibrium: HA A- + H + The monochromator generally fails to choose only one λ, instead it projects a very narrow band on the sample (recall that Beer’s law is applied at one certain λ). The range 0.1 - 1.0 absorbance(79.4 - 10%T) is considered ok for analytical purposes. However the recommended best range to work is 0.2 -0.8. (63.1 – 15.8%T) Absorption spectrum Shows the variation of Absorbance with wave length. We usually select the wavelength of max intensity for the analysis of any compound. 6 2/9/2005 Instrumentation Instrumentation (spectrophotometers) 1. Sources of light Continuous source, that can emit radiation of all wavelengths within the spectral region for which they are to be used. Tungsten Lamp (320-2500 nm) used for visible and near IR spectrophotometers. Deuterium Lamp (200-400 nm) used for UV spectrophotometers. Line sources, emit one or very narrow bands of radiation whose wavelengths are known exactly. Light emitting diodes (LED) used for fluorimeteric measurements. Laser and photodiode, that can be used in different types of spectrophotometers according to selected wavelength. Hallow cathod lamps used in AAS. A single beam spectrophotometer • The spectrophotometer is composed of a polychromatic light source separated into narrow band of wavelength by a wavelength selector (Monochromator), passed through the sample compartment and the transmitted intensity, P, after the sample is measured by a detector . • Single beam, means single light path through sample to detector. 2. Wavelength selector – Filters • Filters permit certain bands of wavelength (bandwidth of ~ 50 nm) to pass through. 2. Wavelength selector – Filters • The simplest kind of filter is colored glass, in which the coloring species absorbs a broad portion of the spectrum (complementary color) and transmits other portions (its color). • Disadvantage: • They are not very good wavelength selectors and can’t be used in instruments utilized in research. • This is because they allow the passage of a broad bandwidth which gives a chance for deviations from Beer’s law and allows interference with impurities.39 Wavelength selectors (monochromator) A monochromator disperses light into its component wavelengths and select a narrow band of wavelengths to pass to the sample or detector. It can be in the form of prism or gratings. Types of gratings Echellette-type grating Reflection followed by either constructive or destructive interferences Concave type grating Collimating/focusing is not necessary so its cost effective with more energy throughput Prisms are not preferable at long wave lengths as they don’t offer enough resolution and the resolution power varies with temperature (glass is affected by heat) thus they are now replaced by different types of gratings. Holographic grating Two lasers with different angles – fringes 6000/mm for 50 cm possible 7 2/9/2005 3- Sample compartment (cells) Importance of Gratings Spectral purity Free of unwanted, scattered or stray radiation. Reflections of optical components. Mechanical imperfections Scattering by dust particles. For Visible and UV spectroscopy, a liquid sample is usually contained in a cell called a cuvet. Glass is suitable for visible but not for UV spectroscopy because it absorbs UV radiation. Quartz can be used in UV as well as in visible spectroscopy Glass 400-3000 nm (vis-near IR) Opaque Face Dispersion Ability to separate small wavelength differences. 1 cm Cuvettes & other sample holders Short pathlength 44 5. Detectors 2. Photomultiplier tube It is a very sensitive device in which electrons emitted from the photosensitive cathode strike a second surface called dynode which is positive with respect to the original cathode. Electrons are thus accelerated and can knock out more than one electrons from the dynode. h If the above process is repeated several times, so more than 106 electrons are finally collected for each photon striking the first cathode. The electrons flow through vacuum to an anode to produce current which is propor-tional to radiation intensity. Plastic containers (VIS) Examples of Detectors (cont.) 4-Examples of Detectors 1. Vacuum Phototube Phototube emits electrons from a photosensitive,-vely charged cathode when struck by visible or UV radiation. NaCl 200-15,000 nm (UV-far IR) Transparent Face Light gathering 1 cm pathlength cuvet Long pathlength Silica/quartz 200-3000 nm (UV-near IR) eanode -V light amplifier dynodes electrons photochathode anode e- Photosensitive cathode 47 voltage divider network 48 high voltage 8 2/9/2005 Double Beam spectrophotometers Photodiode Array-multiplex Good for quick (fraction of a second) scanning of a full spectrum Uses semiconductor material: Remember: n-type silicon has a conduction electron – P or As doped p-type silicon has a ‘hole’ or electron vacancy – Al or B doped A diode is a pn junction: under forward bias, current flows from n-Si to p-Si under reverse bias, no current flows In double beam arrangement, the light alternately passes through the sample and reference (blank), directed by rotating half-sector mirror (chopper) into and out of the light path. When light passes through the sample, the detector measures the Boundary is called a depletion layer or region radiant power P. When the chopper diverts the beam through the blank solution, the detector measures P0. The beam is chopped several times per second and the electronic circuit automatically compares P and P0 to calculate absorbance and Transmittance. Double Beam spectrophotometers (cont.) Double Beam spectrophotometers (cont.) Advantages of double beam over single beam instruments: The absorption in the sample is automatically corrected for the absorption occurring in the blank, since the readout of the instrument is the difference between the sample beam and the blank beam. Automatic correction for changes of the source intensity and changes in the detector response with time or wavelength because the two beams are compared and measured at the same time. Automatic scanning and continuous recording of spectrum (absorbance versus wavelength). Diode array Spectrophotometers It’s a double beam spectrophotometer having multiple diodes (300-500) placed side by side and insulated from one another in the form of a multi-layer sandwich. The common use of a diode array is to monitor light that has passed through a liquid sensor cell as in a multiwavelength liquid chromatography detector. It needs, due to its high scanning rate, a data acquisition unit and should be connected to computer and this allows scanning of full spectrum at the same time with high speed. Single beam spectrophotometer is inconvenient because: The sample and blank must be placed alternately in the light path. For measurements at multiple wavelengths, the blank must be run at each wavelength. Time consumption and possible errors due to stray light or voltage changes during measurements. Diode array Spectrophotometers (cont.) Advantages of Diode array over other double beam spectrophotometers Fast spectral acquisition (in a fraction of a second). No wavelength selection needed, all wavelength are recorded simultaneously while in others only one narrow band reaches the detector at any time. No source intensity lost because we do not use slits as in conventional spectrophotometers. 9 2/9/2005 Spectrophotometric Applications Spectronic D20 Selection of wavelength Selection of wavelength Crystal Violet Absorption Spectrum the wavelength should be If the solution contains more than one absorbing species, 1.4 1.2 Absorbance chosen to give the highest 1 the wavelength should be chosen, whenever possible, in a 0.8 possible sensitivity to be sure 0.6 0.2 concentration can be measured region at which the other species does not absorb radiation max 0.4 that the lowest sample 0 200 250 300 350 with fair accuracy. 400 450 500 wavelength, nm 550 600 650 700 750 measured with good accuracy determination. 10-5 M at max, while at other wavelength (1), it may not be detected at all. 57 max 1 sample Absorbance concentration (10-5 M) can be 10-3 M 10-4 M 5x10-5 M Absorbance For example, the lowest sample or its absorbance is minimum. This will eliminate or at least minimize the interference by the second species, in the 10-2 M Reagent (blank) wavelength wavelength m EXAMPLE It is preferable to choose the wavelength at which the absorbance will not significantly change if the wavelength is slightly changed, i.e., A / is minimum. At a wavelength corresponding to broad horizontal band on the spectrum (band A), the radiation is mainly absorbed to the same extent (A / zero). Absorbance Broad horizontal bands Band B A shoulder 575 nm Acidic form 379 nm Basic form Band A A 445 nm Absorbance Selection of wavelength isobestic point 503 nm 440 nm Wavelength Choose the proper for detection of a weak acid (HA following cases: H+ + A-) in the 1- Determination of the acidic form in presence of the basic form: 445 nm 2- Determination of the basic form in absence of the acid form: 379 or 575nm 3- Determination of the basic form in presence of the acid form: 575 nm m wavelength 4- Determination of both species: 503 nm (isobestic point) 10 2/9/2005 Spectrophotometric Applications 1. Direct methods Isosbestic point An isobestic point is a specific wavelength at which two chemical species have the same molar absorptivity(ε) or -more generally- are linearly related. The word derives from two Greek words: isos: equal, and sbestos: extinguishable. a. Single-point calibration method Standard with measured absorbance Astd = b Cstd Unknown with measured absorbance Aunk = b Cunk When an isosbestic plot is constructed by the superposition of the absorption spectra of two species (whether by using molar absorptivity for the representation, or by using absorbance and keeping the same molar concentration for both species), the isosbestic point corresponds to a wavelength at which these spectra cross each other. Isosbestic point is very important in removing errors due to compounds that can undergo chemical reactions as weak acids or bases. ( This decreases deviation from beer’s law) Ratio the two equations Aunk/ Astd = b Cunk / b Cstd Solve for Cunk Aunk/ Astd = Cunk /Cstd Note that, the standard solution of the species under test should be prepared and measured under the same conditions of the analyte (in the same cell and at the same ). Calibration curve b. By direct extrapolation From a previously plotted calibration curve for the same compound. Series of standard solutions of the analyte is prepared by successive dilutions. Matrix effect, solvent effects and interference In chemical analysis, matrix refers to the components of a sample other than the analyte. Solvent, buffers, reagents, other ions & molecules Blank solution. • Solvent blank • Reagent blank(a blank solution that contains the reagents used in analysis) • Method blank (a blank solution that has been handled similar to a sample, and to which the same reagents have been added, that had contact to the same type of vessels and that was treated by a similar procedure. Standard additions method Ideally the use of the Beer’s law plot of A vs conc. should be sufficiently accurate for the evaluation of data. However, there are frequently what are known as matrix effects where substances other than the analyte affect the absorbance reading of the solution at the wavelength chosen for the analysis. The method of standard additions is used to counteract the matrix effect which we can not overcome or account for. Standard additions method usually involves adding one or more increments of a standard solution to sample aliquots of the same size (spiking). interference 11 2/9/2005 Standard additions method (cont.) c. Analysis of a mixture Example A 2.00 mL urine specimen was treated with reagent to generate a color with phosphate, following which the sample was diluted to 100 mL. To a second 2.00 mL sample was added exactly 5.00 mL of a phosphate solution containing 0.0300 mg phosphate /mL, which was treated in the same way as the original sample. The absorbance of the first solution was 0.428 and that of the second 0.538. Calculate the concentration in milligrams phosphate per liter of the specimen. Cx Detector does not distinguish among absorbance readings of individual analytes. Absorbance of analytes are additive For our example: A λ mixture ε λA b [A] ε Bλ b [B] and more generically: A ε xb [X] ε y b [Y] ε z b [Z] 0.428 x 0.0300 (mg PO 43- /ml) x 5.00ml 0.292 mg PO 43- /ml sample (0.538 0.428) x 2.00 ml sample A1= c1b A2= (c1-c2)b Determination of X , X , Y and Y c. Analysis of a mixture (cont.) Prepare a standard solutions of pure X. Measure the absorbance of these solutions at ’ and ’’. Plot a calibration graph between absorbance readings and A mixture contains two analytes, X and Y Measure absorbance of the mixture at max for X Measure absorbance of the mixture at max for Y A concentrations of the standards at both s. From the slopes of the two graphs, the values of X and X can be determined. The values of Y and can be determined in a similar way. A Xb[ X] Yb[ Y] Y 0.6 A X b[ X] Y b[ Y] Y By knowing the values of molar absorptivities at the two wavelengths, the above two equations can be solved simultaneously for the two unknowns [X] and [Y]. 0.4 X 0.3 ’’ 0.2 0.1 X 0 0 2 4 6 8 10 12 Concentration of standard soltion of X 69 Example Example The concentration of Fe3+ and Cu2+ in a mixture can be determined 4-, which following their reaction with hexacayano-ruthenate (II) Ru(CN)6 forma a purple blue complex with Fe3+ (max= 550 nm), and a pale green complex with Cu2+ (max=396 nm). The molar absorbitivities (M-1cm-1) for the metal complexes at the two wave lengths are summarized below. When a sample ’ 0.5 Absorbance X () () Fe3+ 550 9970 396 84 Cu2+ 34 856 containing Fe3+ and Cu2+ is analyzed in a cell with pathlength of 1.00 cm, the absorbance at 550 nm is 0.183 and at 396 nm is 0.109. What are the molar concentrations of Fe3+ and Cu2+ in the Substituting known values into equation given in slide 22 gives: A550 = 0.183 = 9970CFe + 34CCu A396= 0.109 = 84CFe + 856CCu To determine the CFe and CCu, we solve the 1st equation for CCu CCu = 0.183-9970CFe/34 And substitute the result into the 2nd equation, 0.109 = 84CFe + 856 (0.183-9970CFe/34) =4.607-(2.51x105)CFe Thus Concentration of Fe3+ = 1.80x10-5 M and by substituting back in equation of mixture’s absorbance at 396 , concentration of Cu2+ = 1.26x10-4M. sample? 71 12 2/9/2005 2- Spectrophotometric titrations 2- Spectrophotometric titrations radiation in the UV-Vis. regions, absorbance measurements at fixed wavelength can be used to locate the endpoint of the titration: At + T P The absorbance of the analyte solution is measured after each addition of the titrant and the end point is located from the plot of the absorbance readings and the volume of the titrant added. Let’s consider the analysis of hydrogen peroxide with potassium permanganate in an acidic solution. The potassium permanganate or MnO4- is the only colored substance in the reaction. How would the absorbance change as titrant was added? 5H2O2 + 2MnO4- + 6H+ 5O2 (g) + (Analyte) Absorbance If the analyte (At), the titrant, or the reaction product absorbs (Titrant) purple 2Mn2+ + 8H2O (colorless products) Equivalence point MnO4- reacting, color disappears MnO4accumulates Volume of titrant (mL KMnO4) To determine the concentation of the analyte 2 (MV)H2O2 = 5 (MV)MnO4- Stoichiometry of complex ions Examples Determination of metal ligand ratio Solutions of cation and ligand with identical formal concentrations are mixed in varying volume ratios; but V T = const. A plot of A vs volume ratio (volume ratio = mole fraction) gives maximum absorbance when there is a stoichiometric amount of the two. 76 Example (cont.) Determination metal ligand ratio (cont.) Example To determine the formula of a the complex between Fe2+ and 0-phenathroline, a series of solutions was prepared in which the total concentration of metal and ligand was held constant at 3.15x10-4M. The absorbance of each solution was measured at a wavelength of 510nm. Using the data given in table determine the formula of the complex XL Absorbance 0.0 0.000 0.1 0.116 0.2 0.231 0.3 0.347 0.4 0.462 0.5 0.578 0.6 0.693 0.7 0.809 0.8 0.693 0.9 0.347 1.0 0.000 Determination metal ligand ratio (cont.) Thus, formula of the metal-ligand complex is Fe(o-phenthroline)32+ 77 78 13 2/9/2005 Absorption spectra of organic molecules Determination of acid-base dissociation constant Let’s consider, as a simple example, an acid–base reaction of * or * molecular orbital (anti-bonding) the general form HIn + H2O H3O+ + In– where HIn and In– are the conjugate weak acid and weak base forms of a visual acid–base indicator. The equilibrium constant for this reaction is s or p atomic orbital, Atom X The following formulae can be used for calculating Ka and pKa of weak acids and bases: s or p atomic orbital, Atom Y non-bonding orbital or molecular orbital (bonding) 79 Vacuum UV or Far UV (λ<190 nm ) Absorption spectra of organic molecules (cont.) (1) (2) (3) (4) Electrons in molecule can be classified into four different types: Closed shell electrons that are not involved in bonding. These have very high excitation energies and do not contribute to absorption in UV-visible region. Covalent single bond electrons (, or sigma electrons). These are firmly held and their excitation required energies corresponding to vacuum UV region, thus they are not widely used for analytical purpose because of experimental difficulties to work in this region. Paired non bonding outer-shell electrons( n electrons) such as those on N, S, O and halogens. These are less tightly held than electrons and can be excited in UV-visible region. Electrons in (pi) orbital, as in double and triple bonds. These are the most readily excited and are responsible for the majority of electronic species in UV-Visible region. UV/VIS 82 Electronic Spectra and Molecular Structure The electronic transition that takes place in the UV-Vis regions of spectrum is due to the absorption of radiation by: Specific types of functional groups or bonds (called CHROMOPHORS) that contains valence electrons with relatively low excitation energy. Absorption characteristics of some common chromophors Chromophore Example C=C Alkenes Excitation λmax (nm) ε π π* 190 15,000 C≡C Alkynes π π* 190 10,000 C=O Ketones n π π* π* 290 195 15 10,000 N=O Nitro-compounds n π π* π* 275 200 17 5,000 C-X X=Br X=I Alkyl halides n n σ* σ* 205 255 200 360 The high value of ε for * indicates that these transition are the most easily occurring and thus the most probable transition occurs between these levels. Note that all the above transitions are observed in the ultraviolet region. That is the reason why most simple organic compounds are colorless 83 14 2/9/2005 Effect of Conjugation (cont.) Effect of Conjugation -electrons are considered to be further delocalized by conjugation as this delocalization leads to decrease in the energy gap between and *. * * * CH2=CH2 Absorption maxima are thus shifted to CH2=CH-CH=CH2 conjugated system longer wavelengths (Bathochromic shift Or red shift). The higher the extent of conjugation, the longer the wavelength of transition. BLUE shift RED shift Also, the molar absorptivity (ε) roughly doubles with each new conjugated double bond (Hyperchromic effect). 85 86 Effect of Conjugation (cont.) Some highly conjugated organic compounds such as carotene and lycopene absorb radiations at longer wavelength enough to be observed in the visible region i.e., they are colored. Effect of solvents The solvents used to obtain UV-VIS spectra can also influence the band shape and wavelength of the measured absorption bands. Non-polar, non-hydrogen bonding solvents approximate the -CAROTENE conditions in the vapor state and often give spectra with more fine structure. Polar and/or hydrogen bonding solvents tend to give more band broadening. Because the ground and excited states often have different charge LYCOPENE 87 distributions the interactions with polar solvents can shift the transitions of n→π* to shorter wavelengths (hypsochromic shift, blue shift); and the π →π* transitions to longer wavelengths (bathochromic shift, red shift) 88 The effects of substitution Effect of solvents (cont.) 89 Auxochrome function group Auxochrome is a functional group that does not absorb in UV region but has the effect of shifting chromophore peaks to longer wavelength (bathochromic effect) as well as increasing their intensity (hyperchromic effect). 90 15 2/9/2005 Analysis of non absorbing species • Many non-absorbing analytes can be determined spectrophotometrically by causing them to react with chromophoric reagents to give products that absorb strongly in the UV-Vis regions (products of high ). • Most of the spectrophotometric clinical tests are based on such kind of interactions . to the colorless solution containing Fe2+ add buffer and excess ferrozine (chromophoric reagent) where a purple complex is formed. Measure the absorbance of at 562 nm (max for the complex) Fe2+ + 3ferrozine colorless [(ferrozine)3Fe]4purple complex emission can be induced or spontaneous Emission intensity I = kP0C Where p0 is the incident radiant power C is concentration of emitting species k is a constant An isolated atom scatters light because the electric field of the incident light wave forces the electrons in the atom to oscillate back and forth about their equilibrium position. By the laws of electromagnetism, when a charge changes its velocity, it emits radiation. Inelastic scattering is important for gases. In liquids, random molecular motions broaden Rayleigh scattering. Light is emitted uniformly in all directions in the plane ⊥ to oscillation, but decreases in amplitude as the viewing angle shifts away from that plane. 16 2/9/2005 Determination of acid-base dissociation constant (cont.) Determination of acid-base dissociation constant (cont.) Example 1 Example 2 The acidity constant for an acid–base indicator was determined by preparing three solutions, each of which has a total indicator concentration of 5.00x10–5 M. The first solution was made strongly acidic with HCl and has an absorbance of 0.250. The second solution was made strongly basic and has an absorbance of1.40. The pH of the third solution was measured at 2.91, with an absorbance of0.662. What is the value of Ka for the indicator? The Absorbance of a fixed concentration of a pharmaceu-tical compound (phenylephrine) at 292 nm is found to be 1.224 in 0.1M NaOH and 0.02 in 0.1M HCl. Its absorbance in a buffer of pH 8.5 is found to be 0.349. Calculate the pKa value of its acidic phenolic hydroxyl group. The value of Ka is determined as [H3O]+ = -antilog 2.91 = 1.23x10-3M pKa 8.5 log 1.224 0.349 8.5 0.402 8.902 0.349 0.02 99 100 Derivative spectrophotometry Derivative spectrophotometry (cont.) Two closely spaced wavelengths are alternately passed through the sample. To Derivative spectrphotometry can be used to clarify bands in more complex UV spectra. The main effect of derivatization is background correction and better selectivity than normal spectrophotometry for resolving binary mixtures and some ternary mixtures. Derivative spectrophotometry have vast applications in analysis of pharmaceutical compounds that contain more than one active ingredient at the same time. 101 record spectra, one wavelength is set so as to lag behind the other by a fixed amount during the wavelength scanning process. Three points on the derivative curve are of special significance. The point where the curve intercepts the zero line corresponds to the wavelength of the conventional peak's maximum . The negative portion of the derivative curve has a minimum value, and the positive has a maximum value. The horizontal distance between the two is a measure of peak bandwidth This derivatization is done automatically through spectrophotometer. and the vertical distance from the maximum to the minimum relates to the amplitude of the conventional spectral peak and is dependent on chromophore concentration. 102 17 2/9/2005 Derivative spectrophotometry (cont.) Example, UV spectrum of pseudoephedrine, dextromethorphan and triprodline, pesudoephedrine lies on top of a large background because of the other two compounds, which have much stronger chromophoric groups than pesudoephedrine. By special calculations, done through softwares, the amount of pseudoephedrine can be tested by plotting a calibration graph using information obtained from 1st and 2nd derivative absorption spectra. Application of UV spectrophotometry in dissolution testing In recent years, dissolution testing has become increasingly important in the pharmaceutical industry. Dissolution testing provides information on batch homogeneity and conformity, and is routinely used for quality control purposes. In addition, conclusions about the in vivo behavior of the product can be drawn from its in-vitro dissolution behavior. Therefore, the dissolution test has become an indispensable tool in the development of new solid oral dosage forms and the assessment of the physical stability of the formulation. An appropriate in-vitro-dissolution test method should meet the following criteria: - Sufficient discriminatory power Identification of critical manufacturing variables - Low variability of the method - Proof of batch homogeneity 103 Application of UV spectrophotometry in dissolution testing (cont.) Application of UV spectrophotometry in dissolution testing (cont.) Traditionally, dissolution testing has been carried out on solid dosage forms using a standard dissolution test apparatus. Samples of the dissolving active within the dosage form are taken for estimation from the dissolution tester either manually, or using more automated means such as a pump which has been connected to a suitable measurement device, usually a UV-VIS Spectrophotometer. The amount dissolved is usually calculated by comparison to a standard calibration curve. 105 106 In-class exam 1 Take home exam 1 Calculate the absorbance of an organic dye (C= 7 x 10 -4 mol L-1), knowing that the molar absorbtivity is 650 mol-1 L cm-1 and that the length of the optical path of the cell used is 2 x 10 -2 m. Calculate the absorbance of an organic dye (C= 7 x 10 -4 mol L-1), knowing that the molar absorbtivity is 650 mol-1 L cm-1 and that the length of the optical path of the cell used is 2 x 10 -2 m. What would happen to the (i) absorbance & (ii) transmittance, if the cell used was double its present thickness? What would happen to the (i) absorbance & (ii) transmittance, if the cell used was double its present thickness? 18 2/9/2005 In-class exam 2 Calculate the energy of a mole of protons corresponding to a wavelength of 300 nm. the speed of light in vaccum ( = 2.998 x 108 m/s) the Planck’s constant (=6.626x10-34 J s) 19