Transcript
By
Fatimah Abu-Dayah
/2008F.ABUDAYAH
1
Clinical objectives
By the end of this lecture you will be able
to:
Define pacemaker
Differentiate types of pacemaker
List function of pacemaker
Assist and monitor pt under going pacing
Identifying pt’s educational needs
2
/2008F.ABUDAYAH
Out line









Introduction
Definition of cardiac pacing
Clinical Indication
Pacemaker design
Pacemaker function
Types of pacing
Nursing diagnosis
Nursing intervention
Pt’s education
/2008F.ABUDAYAH
3
Normal conductive system of the
heart
/2008F.ABUDAYAH
4
Definition of cardiac pacing
It is an electric device that
delivers direct electrical
stimulation to stimulate the
myocardium to depolarize
,initiating a mechanical
contraction.
/2008F.ABUDAYAH
5
Clinical Indication
1.
2.



3.
/2008F.ABUDAYAH
Symptomatic bradycardia
Symptomatic heart block
2nd degree heart block
3rd or complete heart block
Bifasicular or transfasicular
bundle branch blocks.
Prophylaxis
6
Pacemaker Design
1.
2.
/2008F.ABUDAYAH
Pulse generator
leads
7
Pacemaker Design
Pulse generator
 In permanent pacemaker is
encapsulated in a metal can ,to
protect the generator from
electromagnetic interference
/2008F.ABUDAYAH
8
Pacemaker Design
Pulse generator
 Temporary pacing system
generator is externally
contained in a small box
/2008F.ABUDAYAH
9
Pacemaker Design
Pulse generator
 Transcutanus external pacing
system house the generator in a
piece of equipment similar to
portable ECG monitor.
/2008F.ABUDAYAH
10
Pacemaker Design
Pacemaker lead
1. Single chamber (unipolar)
pacemaker
 Lead placed in atrium or ventricle
 Produce large spic on the ECG
 Sensing and pacing in the chamber
where the lead is located
 More likely to be affected by
electromechanically interference
/2008F.ABUDAYAH
11
Single chamber (unipolar
/2008F.ABUDAYAH
12
/2008F.ABUDAYAH
13
Pacemaker Design
2.




Dual-chamber (bipolar) pacemaker
One Lead located in the atrium and
one in the ventricle
Sensing and pacing in both
chambers mimicking the normal
heart function
Produce in visible spic in the ECG
Less affected by electromechanical
interference.
/2008F.ABUDAYAH
14
Dual-chamber (bipolar) pacemaker
/2008F.ABUDAYAH
15
/2008F.ABUDAYAH
16
Pacemaker function
1.
2.
3.
/2008F.ABUDAYAH
Pacing function
Sensing function
Capture function
17
Pacing function
Atrial pacing:
stimulation of RT atrium produce spic
on ECG preceding P wave
/2008F.ABUDAYAH
18
Pacing function
Ventricle pacing :
stimulation of RT or LT ventricle
produce a spic on ECG preceding
QRS complex.
/2008F.ABUDAYAH
19
Pacing function
AVpacing:
direct stimulation of RT atrium
and either ventricles mimic
normal heart conduction
/2008F.ABUDAYAH
20
Sensing function
Sensing :
Ability of the cardiac pace maker
to see intrinsic cardiac activity
when it occurs.
/2008F.ABUDAYAH
21
Sensing function
Demand:
 pacing stimulation delivered only if
the heart rate falls below the preset
limit.
Fixed:
 no ability to sense. constantly
delivers the preset stimulus at preset
rate.
Triggered:
delivers stimuli in response to
(sensing )cardiac event.
/2008F.ABUDAYAH
22
Capture function
Capture:
Ability of the pacemaker to
generate a response from the
heart (contraction) after
electrical stimulation.
/2008F.ABUDAYAH
23
Capture function
1.
2.
/2008F.ABUDAYAH
Electrical capture :
indicated by P or QRS
following and corresponding to
a pacemaker spike.
Mechanical capture:
palpable pulse corresponding
to the electrical event.
24
Pacing types



Permanent
Temporary
biventricular
/2008F.ABUDAYAH
25
Types of pacing
1.



/2008F.ABUDAYAH
Permanent pacemaker
Used to treat chronic heart
condition
Surgically placed
transvenuosly under local
anesthesia
Pulse generator placed in a
pocket subcutaneously ,can be
adjusted externally
26
Permanent pacemaker
/2008F.ABUDAYAH
27
Types of pacing
2.



Temporary pacemaker
Placed during emergencies
Indicated for pts’ high
degree heart block or
unstable bradycardia
Can be placed transvenosly,
epicardially,transcutanusly
or transthorasicly
/2008F.ABUDAYAH
28
Types of pacing
3.


Biventricular pacemaker
Used in sever heart failure
Utilize three leads in right
atrium, right ventricle and
left ventricle to coordinate
ventricular coordination
and improve cardiac
out put
/2008F.ABUDAYAH
29
Equipments




Transvenous pacing
catheter
EKG machine
Pacemaker generator
with battery and cable
Emergency crash cart




/2008F.ABUDAYAH
Lidocaine
Defibrillator
(2) 5cc syringe with
22 and 25 gauge
needles
External Pacer
Sterile gown, gloves,
mask
30
INSERTION SITES




Left Subclavian (most reliable)
Internal jugular (lower
incidence of pneumothorax)
Femoral vein
Brachial vein
/2008F.ABUDAYAH
31
INSERTION PROCEDURE



1. Check that patient has a patent IV,
and that the defibrillator, emergency cart
and appropriate medications are
available.
obtain consent (time permitting).
Obtain vital signs and ECG rhythm strip
prior to insertion. Connect to 12 lead EKG
and continuously monitor before, during
and after
/2008F.ABUDAYAH
32
INSERTION PROCEDURE



Anesthetize the area locally.
Prepare the external temporary
generator:
Portable Chest X-ray is required
to confirm placement.
/2008F.ABUDAYAH
33
Applying transcutaneous pacing





Anterior/posterior:
Anterior/anterior:
Module on stand by. minimal out
put
Connect pacing to external
module
Increase milliamp until a pacing
spike and corresponding QRS
are seen.plpate pulse
/2008F.ABUDAYAH
34
Complication






Movement and dislocation of the lead
Injury
Bleeding and hematoma
Ventricular ectopy or VT from wall
stimulation
Infection
Cardiac tamponad
/2008F.ABUDAYAH
35
Nursing diagnosis





Decreased cardiac output related to
potential pacemaker mal function
Risk for injury related to
peumothorax
Impaired physical mobility related to
restriction of movement.
Acute pain related to surgical incision
or external pacing stimuli.
Disturbed body image related to
pacemaker implementation.
/2008F.ABUDAYAH
36
Nursing intervention
1.






/2008F.ABUDAYAH
Maintain adequate cardiac output
Record information after insertion
pacemaker model ,mode, program
setting,pt’s rhythm
Attach ECG for continues monitoring
Analyze rhythm strips as per
protocol
Monitor vital signs
Monitor urine output
Observe for dysrhythmia
37
Nursing intervention
Avoid injury
 Obtain chest x-ray to
check lead wire
position
2.



Monitor for sign and symptom of
hemothorax
Monitor for sign and symptom of
pneumothorax
Evaluate evidence for bleeding
/2008F.ABUDAYAH
38
Nursing intervention
3.




/2008F.ABUDAYAH
Monitor for evidence of lead
migration and perforation of
heart
Observe for muscle twitching
and hiccups
Evaluate chest pain
Auscultate foe friction rub
Observe for signs of cardiac
tamponade
39
Nursing intervention
4.


/2008F.ABUDAYAH
Provide electrically safe
environment
Protect exposed parts of
electrode leads with rubber
Wear rubber gloves when
touching a temporary pacing
lead
40
Nursing intervention
Be aware of hazards in the facility
that can interfere pacemaker and
cause failure
 Avoid use of electrical razor
 Avoid direct placement of defibrillator
paddles over the generator, should be
placed 4-5 inches away.
 Pt’s with permanent pacemaker
should never exposed to MRI because
it may alter and erase the program
memory.
 Caution must be used if pt will receive
/2008F.ABUDAYAH
41
radiation therapy.
5.
Nursing intervention
6.


/2008F.ABUDAYAH
Prevent accidental pacemaker
malfunctions
Use external plastic covering
over external generator all
times
Secure temporary pace maker
over pt’s chest or wrist never
hang it over iv pole
42
Nursing intervention



Place a sign over pt's bed
alerting personnel to the
presence of pacemaker.
Evaluate transecutanuse
pacing every 2 hr
Monitor for electrolyte
imbalances, hypoxia and
myocardial infarction.
/2008F.ABUDAYAH
43
Nursing intervention
7.





/2008F.ABUDAYAH
Preventing infection
Take temp every 4hrs
Observe for sign and symptoms of
infection
Clean incision site with sterile
technique
Monitor vein which pacing placed in
for phlipaitis
Administer antibiotic as ordered.
44
Nursing intervention
8.
9.
10.
11.



Relieving anxiety
Reliving pain.
Maintaining a positive body image
Minimizing the effect of immobility
Rest for 24-48 hrs post pacing
insertion
Deep breathing exercise
Restrict movement of affected
extremity
/2008F.ABUDAYAH
45
Patient education
Anatomy and physiology of the heart
2. Pacemaker function
3. Activity
Specific instruction include
 Not to lift items over 1.4kg or
perform difficult arm maneuver.
 Avoid excessive stretching or
bending excessive.
 Avoid contact sport,tennis,gulfing
until advised by doctor.
 Sexual activity can be resumed
/2008F.ABUDAYAH
46
when desired
1.
Patient education
4.




Pacemaker failure
Teach pt to check own pulse
at least weekly for 1 min
Report slowing on the pulse
less or greater than the setting rate
Report sign and symptom as
palpitation ,fatigue ,dizziness
,prolonged hiccups
Wear identification bracelet and carry
a pacemaker identification cared.
/2008F.ABUDAYAH
47
Patient education
5.



/2008F.ABUDAYAH
Electromagnetic interference
Caution pt that EMI could interfere
with pacemaker function.
Explain that high energy radar, TV
and radio transmetters,MRI,large
motors may affect the pacemaker
function.
Teach pt to move 4-6 m away from
source and check pulse. it should
return to normal.
48
Patient education



Most pacemaker equipped with
internal filters to prevent
interaction with cell phone.
Tell pt that antitheft devices and
airport security alarms may affect
pacemaker and trigger security
alarm.
Household and kitchen appliance
will not affect pacemaker.
/2008F.ABUDAYAH
49
Patient education
6.




Care of pacemaker site.
Wear loose-fitting
clothes around pacemaker
Watch sign and symptom
of infection
Keep incision site clean
and dry. not to scrub site
Advise well balanced diet.
/2008F.ABUDAYAH
50
References
Sandra M. Nettina
MSN, APRN, BC, ANP
Manual of Nursing Practice
Eighth Edition
Braunner&SuDDARTH’S
Textbook of medical surgical
nursing 10th edition
/2008F.ABUDAYAH
51