Download 3200 - transmille.net

Survey
yes no Was this document useful for you?
   Thank you for your participation!

* Your assessment is very important for improving the workof artificial intelligence, which forms the content of this project

Document related concepts

Mercury-arc valve wikipedia , lookup

Power engineering wikipedia , lookup

Variable-frequency drive wikipedia , lookup

Rectifier wikipedia , lookup

Islanding wikipedia , lookup

Electrical ballast wikipedia , lookup

Three-phase electric power wikipedia , lookup

History of electric power transmission wikipedia , lookup

Opto-isolator wikipedia , lookup

Electromagnetic compatibility wikipedia , lookup

Resistive opto-isolator wikipedia , lookup

Electrical substation wikipedia , lookup

Buck converter wikipedia , lookup

Surge protector wikipedia , lookup

Current source wikipedia , lookup

Switched-mode power supply wikipedia , lookup

Ground loop (electricity) wikipedia , lookup

Ohm's law wikipedia , lookup

Automatic test equipment wikipedia , lookup

Ground (electricity) wikipedia , lookup

Multimeter wikipedia , lookup

Voltage optimisation wikipedia , lookup

Alternating current wikipedia , lookup

Stray voltage wikipedia , lookup

Electrical wiring in the United Kingdom wikipedia , lookup

Mains electricity wikipedia , lookup

Earthing system wikipedia , lookup

Portable appliance testing wikipedia , lookup

Transcript
3200 Electrical Test Equipment
Calibrator
The Total Solution to Electrical
Test Equipment Calibration
3200 Electrical Test Equipment Calibrator –
Five Calibrators In One
Calibrates…





INSULATION TESTERS
RCD TESTERS
LOOP TESTERS
PORTABLE APPLIANCE TESTERS
HIPOT / BREAKDOWN TESTERS
3200 Electrical Test Equipment Calibrator –
Insulation Tester Calibration
 Insulation Resistance from 100kohms to 10Gohms at 1kV
 Measurement of Insulation Test Voltage at 1mA load
 AC Voltage Calibration to 400 Volts
 Continuity Resistance from 0 to 1kohms
 Continuity Resistance Current to 300mA Measured into 1ohm
Calibrating Insulation testers with the 3200 is simple. All the functions
required are in one instrument, making calibration fast and simple.
Use with ProCal software to automate and save even more time.
3200 Electrical Test Equipment Calibrator –
RCD Tester Calibration
Residual circuit breaker testers are designed to test and measure
the break time of AC mains (RCB) trips at set test currents. Fault
current is allowed to pass between live phase and earth.
The ‘fault’ current is started by pressing a button on the tester,
usually an LCD counter on the tester measures the time in
milliseconds until the mains supply is disconnected.
The 3200 has been specially designed to make
the calibration of RCD testers safe & straight
forward, which can otherwise be quite difficult
and even dangerous.
To calibrate an RCD tester, it is necessary to measure the current taken from the
mains supply by the tester on each of its ranges and the time period measured shown
on the testers display to the disconnection of supply.
The 3200 can calibrate both functions. Intelligent firmware measures the correct fault
current ignoring pre-test currents from testers such as the AVO CM500. The 3200 also
provides accurate timed disconnection of the supply to verify the testers timing
accuracy.
Some RCB testers have additional functions such as which half cycle of the mains the
fault current started, fault current duration etc. which can also be tested by the 3200.
3200 Electrical Test Equipment Calibrator –
Loop Tester Calibration
 Automatic Measurement and compensation of
supply loop impedance
 Automatic detection of faulty Tester preventing damage to 3200
 8 loop values from 0.05ohm to 1kohm.
To calibrate a Loop Tester simple connect to the 3200, select the loop
function and run auto loop. Select the required loop value on the 3200
and press test. The 3200 automatically checks for a faulty loop tester
before applying the full mains supply.
3200 Electrical Test Equipment Calibrator –
PAT Calibration
 Earth Bond Resistance
 Earth Bond Current Measurement to 30 Amps
 Insulation resistance from 1M to 10Gohms
 Leakage Current Measurement
 Class 1 & 2 Flash Voltage Measurement
There are several different measurement functions to be
calibrated on a PAT tester. The functions required are all in
the 3200 ensuring fast and efficient calibration of these
instruments.
3200 Electrical Test Equipment Calibrator –
HIPOT / Breakdown Tester Calibration
 3kV & 12kV AC/DC Measurement Ranges
 200uA • 2mA • 20mA Current Ranges
 200uA : 5M •10M •20M Loads
 2mA : 500k •1M •2M Loads
 20mA : 50k •100k •200k Loads
 Current ranges : 1kV maximum
 Power On & Overload Indicators
 Requires 3200 Calibrator
VOLTAGE MEASUREMENT
Connects between ground and appropriate voltage terminal
Select voltage range (3kV / 12kV AC/DC) and read back the
voltage measured from the display of the 3200.
CURRENT LOAD
Three electronically selectable ranges, nine current
loads Select current range using the 3200 ‘softkeys’
Connect breakdown tester between ground and the
appropriate resistor socket.