Download General Screen Film Radiography and Its Limitations

Survey
yes no Was this document useful for you?
   Thank you for your participation!

* Your assessment is very important for improving the workof artificial intelligence, which forms the content of this project

Document related concepts

Proton therapy wikipedia , lookup

Medical imaging wikipedia , lookup

Radiosurgery wikipedia , lookup

Radiographer wikipedia , lookup

Image-guided radiation therapy wikipedia , lookup

Backscatter X-ray wikipedia , lookup

Industrial radiography wikipedia , lookup

X-ray wikipedia , lookup

Fluoroscopy wikipedia , lookup

Transcript
Chapter 2
General Screen Film Radiography and Its Limitations
X-rays have been in use for over a century since the report of their discovery by
Röentgen (1896). In the health and medical areas, x-rays have been used for both
diagnosis and treatment of patients’ conditions. In diagnostic medical imaging
(radiography), x-rays are used in a wide variety of examinations. Examples of
medical imaging examinations can be found in Table 2.1.
Table 2.1
Types of medical imaging examinations using x-rays as the
energy source (Ballinger, 1991; Fauber, 2000; Gunn, 2002)
general radiography – single projection images
fluoroscopy – single projection imaging to display function or anatomical
motion
angiography – a rapid series of single projection images to capture
motion of blood flow
computed tomography (CT) – multiple projection imaging to display a
cross-section of the anatomy
bone mineral densitometry (BMD) – analysis of the bone strength.
X-rays are electromagnetic (EM) radiation with characteristics of short wavelengths,
very high frequencies and very high energy. High energy EM radiation exhibits
characteristics of both waves and discrete bundles of energy called photons. X-ray
photon energy is usually measured in kilo-electron volts (keV) and diagnostic x-ray
energies are typically 10 to 150 keV. Such high energy EM radiation is ionising.
Ionising radiation has potential harmful effects when it irradiates human tissue
(Bushberg et al, 2002; Curry et al, 1990; Graham & Cloke, 2003).
In diagnostic medical imaging, x-ray production and its attenuation in matter is well
described by Bushberg et al, (2002), Bushong (2001), Curry et al (1990), Dowsett,
Kenny & Johnston (1998), Graham & Cloke, (2003), Thompson et al (1994) and
5
Webb (1988). An important consideration for this project is that a heterogeneous
intensity x-ray beam results from differential attenuation within the patient’s
anatomy.
The exit intensities of the x-ray photons will depend on the properties of the body
within the irradiated area and the characteristics of the x-ray beam. The mechanisms
that result in reduction of x-ray intensities are the types of interaction with matter,
such as the photoelectric effect and Compton scatter. Exit intensity of the x-ray beam
can be considered as a function of the probability that an x-ray photon will have an
attenuating event. Properties of the body that affect attenuation are effective atomic
number of the anatomical material, electron density of the anatomical material, and
thickness of the anatomical area or the distance the x-ray photons travel through
anatomy. As the effective atomic number of the tissue increases, the probability of an
interaction of the x-ray photon with the anatomy increases; as electron density of the
anatomy increases, the probability of an interaction of the x-ray photon with the
anatomy increases; and as the x-ray photon’s path distance through the body
increases, the probability of an interaction of the x-ray photon with the anatomy
increases. The degree of attenuation is also affected by the energy of the photons of
the entrance x-ray beam. As x-ray photon energy increases, the probability of an
interaction of the x-ray photons with the anatomy decreases (Bushberg et al, 2002;
Curry et al, 1990; Graham & Cloke, 2003).
X-ray beams are comprised of photons of many different energies, and as such are
polyenergetic. A characteristic of an x-ray beam that affects the rate of attenuation,
and hence the exit intensity of the beam, is its effective energy. Effective energy of
the x-ray beam can be considered as the weighted mean of the x-ray photon energies
within the beam (Bushberg et al, 2002; Carlsson & Carlsson, 1984). A common
means of measurement of the effective energy of the x-ray beam is the measurement
of the half value thickness (HVT) of the material that is irradiated. The HVT is the
thickness of the material irradiated that reduces the intensity of the exit x-ray beam to
half of its entrance intensity. Common materials used for the measurement of
effective energy of an x-ray beam are aluminium (Al) and copper (Cu). HVT, at
diagnostic x-ray energies, is typically stated in millimetres of Al or Cu (Bushberg et
al, 2002; Carlsson & Carlsson, 1984).
6
Attenuation of polyenergetic x-ray beams can then be considered as a function of
their effective energy or HVT. As the effective energy of x-ray beams increases, the
probability of interactions of the x-ray photons decreases. A higher effective energy
x-ray beam will result in a higher exit intensity than when the effective energy of the
entrance beam is lower.
Subject contrast is the differences in exit intensities of the x-ray beam, within the
irradiated field, that result from different rates of attenuation within the irradiated
body. Large subject contrast results from large differences in attenuation rates. Large
subject contrast will typically result from a low effective energy beam as well as
large differences in attenuation (Bushong, 2001; Fauber, 2000; Gunn, 2002).
Subject contrast produces differences in optical densities that can be visualised
within the resulting image. Optical density differences are known as radiographic or
image contrast. The degree of radiographic contrast depends on the subject contrast
and also depends on factors within the image receptor (Fauber, 2000; Gunn, 2002;
Thompson et al, 1994).
2.1
Filtration and Shaping of the X-ray Beam
Polyenergetic x-ray beams used in medical imaging have a maximum photon energy,
which corresponds to the peak kilovoltage (kVp) applied across the x-ray tube. The
kVp is determined by the radiographer, depending upon the desired characteristics of
the radiographic image. A polyenergetic x-ray beam produced within the x-ray tube
anode has a low number of high energy photons and there are an increasing number
of photons at each energy level as the photon energy decreases from the maximum
energy. This is depicted graphically as the dashed line in Figure 2.1 (Curry et al,
1990). In Figure 2.1, the maximum photon energy is 150 keV which results from
setting 150 kVp across the x-ray tube.
7
Figure 2.1
Typical energy spectrum of a 150kVp x-ray beam (Curry et al, 1990,
p.33)
Filtration is the removal of x-ray photons from the beam by attenuation when the
beam is passed through a medium (Bushberg et al, 2002; Curry et al, 1990;
Thompson et al, 1994). There are two types of general filtration: inherent and added
filtration.
X-ray photons are produced by interaction of an accelerated electron with a target
atom. The atoms producing the x-ray photons are often at depth within the target. As
the photons travel within the target itself, attenuation can occur. As the photons
travel further through the x-ray tube, attenuation can again occur within the glass of
the x-ray tube, within the cooling oil that surrounds the glass x-ray tube and within
other parts of the x-ray tube housing and collimator. This is the process of inherent
filtration. Lower energy photons have a high probability of attenuation. Photons with
energies below 15 keV are fully attenuated by the x-ray tube’s inherent filtration
(Bushberg et al, 2002; Curry et al, 1990; Graham & Cloke, 2003; Thompson et al,
1994). The solid line in Figure 2.1 depicts a typical shape of the x-ray spectrum
following inherent filtration of the Bremsstrahlung and characteristic radiation.
8
Added filtration is the addition of other material in the path of the x-ray beam prior to
the x-ray beam entering the patient’s anatomy. The prime purpose of added filtration
is the further removal of low energy photons from the x-ray beam. The effect of
added filtration on the shape of the x-ray spectrum can be seen in Figure 2.2 (Curry
et al, 1990). Low energy photons, if allowed to enter the patient’s body, would have
a high probability of being fully attenuated by the anatomy and hence increasing the
absorbed dose to the patient without contributing to the image (Bushberg et al, 2002;
Curry et al, 1990; Thompson et al, 1994).
Figure 2.2
X-ray spectra at 90kVp with added filtration (filtered) and without
added filtration (unfiltered) (Curry et al, 1990, p.89)
Added filtration will remove more lower energy photons from the beam than higher
energy photons. Added filtration assists in lowering the patient absorbed dose.
Another effect of placing added filtration in the path of the x-ray beam is that of
shifting the effective energy of the beam. In Figure 2.2 it can be seen that the
addition of filtration to the beam has increased the effective energy of the beam.
Added filtration can be any material that is placed in the path of the x-ray beam. It is
usually measured in equivalent thickness of aluminium. When the prime purpose is
9
dose reduction, added filtration material is of uniform thickness across the entire
x-ray field. Minimum amounts of added filtration are usually regulated by statutory
authorities. In Australia, these standards are regulated by various state Acts and
Regulations such as the New South Wales Radiation Control Act (NSW) (1990), the
Radiation Control Regulation (NSW) (2003) and the Radiation Guidelines 6 (2004).
Aluminium and copper are the most common added filtration material (Bushberg et
al, 2002; Curry et al, 1990; Thompson et al, 1994). Typical thicknesses of added
filtration material in diagnostic radiography are 2 – 3 mm of Al or 0.1 mm of Cu
(Carlsson & Carlsson, 1996). The use of erbium filter material has been reported by
Chakera et al (1982), Shrimpton et al (1988) and Cranage et al (1992). Other authors
(Koedooder & Venema, 1986; MacDonald-Jankowski & Lawinski, 1992; Regano &
Sutton, 1992; Sanborg et al, 1993; Tapiovaara et al, 1999; Villegran et al, 1978)
have reported on the use of other filtration material in medical x-ray use such as
samarium (Sm), gadolinium (Gd), holmium (Ho), ytterbium (Yb), tungsten (W),
yttrium (Y), niobium (Nb) and other metals and materials.
Different filter materials attenuate the x-ray beam in different manners and hence
produce different dose rates to the patient. The prime purpose of many authors
(Chakera et al, 1982; Kohn et al, 1988; MacDonald-Jankowski & Lawinski, 1992;
Regano & Sutton, 1992; Sanborg et al, 1993; Shrimpton et al, 1988) has been to
document the improved dose reduction when particular filter materials are placed in
the x-ray beam.
The choice of filtration material can have effects other than dose reduction in
medical imaging. Kohn et al (1988) used five radiologists to compare image quality
when different filter materials were used in the x-ray beam. The radiologists found
no difference in image quality of skull images when Al, Cu and Y filter material
were used. The radiologists preferred images of the hands when Al was the filter
material used. Image contrast was reduced when Cu filter material was used in
paediatric barium examinations (Hansson et al, 1997). In low kVp dental
radiographic examinations reported by Shibuya et al (2000), filter materials of Al,
Nb, Gd and a composite material of Cu, Al and tin were used. In this case, the
composite material improved diagnostic performance over the other filter materials.
10
Williamson et al (1994) reported on the use of K-edge filtration materials, where the
binding energy of the K-edge was within the diagnostic medical imaging range of
energies, between 17 and 68keV. One of the conclusions drawn by Williamson et al
(1994) was that reduced image contrast occurred when K-edge filters were used
compared to when Al filters were used.
2.2
Film/Screen Radiography
X-ray intensities that exit the body are recorded for viewing and storage. In general
radiography, the x-ray image may be recorded using a variety of means. The type of
general radiography is named after the recording media used. The types of general
radiography are listed in Table 2.2.
Table 2.2
Types of general radiography recording media
(Bushberg et al, 2002; Bushong, 2001; Gunn, 2002)
film/screen (F/S) radiography using film and intensifying screens
digital radiography (DR) such as:
computed radiography (CR) using photostimulable phosphors
flat panel systems such as:
direct radiography using photoconductor material with thin film
transistors
indirect radiography using intensifying screens and detectors.
Material that fluoresces under irradiation from x-rays was first noted by Röentgen
(1896). Intensifying screens for use with film as a joint recording medium were first
designed by Thomas Edison (Thompson et al, 1994). The purpose of the intensifying
screen is to convert the x-ray energy to light. The use of intensifying screens
increases the efficiency of exposure of the film over exposure directly by the x-ray
beam. The use of intensifying screens decreases the absorbed dose received by the
patient compared to x-rays directly exposing the film. Films are typically exposed by
95 – 99% light and 1 – 5% x-ray photons when intensifying screens are used. A
measure of the screen’s efficiency is its intensifying factor (IF). IF is a comparison of
exposures with and without the use of intensifying screens to achieve the same
11
optical density on the film (Bushong, 2001; Curry et al, 1990; Fauber, 2000;
Thompson et al, 1994 ).
Many different fluorescent materials have been used in intensifying screens.
Different materials fluoresce at different wavelengths of light. Examples of some
fluorescent materials and their spectral emission used in intensifying screens are:•
calcium tungstate (CaWO4) – blue light (peak response ≈ 420nm);
•
gadolinium oxysulphide: terbium activated (Gd2O2S:Tb) – green light (peak
response ≈ 550nm);
•
lanthanum oxybromide: thulium activated (LaOBr:Tm) – blue light (peak
response ≈ 380 to 420nm).
(Bushberg et al, 2002; Curry et al, 1990; Graham, 2003)
X-ray films are designed to be sensitive to the specific spectral emission of a given
intensifying screen. Consequently, films and screens are matched for optimised
efficiency of conversion of x-ray photons to optical densities on the film. The F/S
combinations can exhibit different characteristics of conversion efficiency (speed),
spatial resolution or detail visualised in the image. It is generally accepted that high
speed F/S combinations will result in lower spatial resolution. The use of a higher
spatial resolution and lower speed F/S combination will result in a high absorbed
dose to the patient (Bushong, 2001; Fauber, 2000; Gunn, 2002; Thompson et al,
1994).
Optical density (OD) is a logarithm of the ratio of the amount of incident light to the
transmitted light through the x-ray film when viewing the image. OD measurements
on x-ray film range between the base fog of the film, typically an OD of 0.2, to the
maximum density of the film, Dmax, at a typical OD of 3.5 to 4. When the number of
x-ray photons reaching the F/S is high, the OD is high. When there has been no x-ray
exposure to the F/S, the OD measured is that of the base fog of the film. High speed
F/S combinations require fewer x-ray photons to achieve the same OD as lower
speed F/S combinations (Fauber, 2000; Gunn, 2002).
12
The OD at a specific point on the film is dependant upon several factors. These
factors are the initial number of photons produced in the x-ray tube, loosely referred
to as exposure; the amount of attenuation of the x-ray beam as it travels through the
body; the F/S speed; and the film processing (Fauber, 2000; Gunn, 2002; Thompson
et al, 1994).
Another characteristic of F/S combinations is F/S latitude. Latitude is the range of
relative exposures required to produce a range of OD. The range of OD is usually
between values of 0.25 and 2.0. Figure 2.3 (Curry et al, 1990, p.159) shows plots of
OD resulting from various exposures reaching the F/S. If a small range of exposures
produces these ODs, the latitude is said to be narrow (Figure 2.3 a.). If a larger range
of exposure produces the same ODs, the latitude is said to be wide or broad (Figure
2.3 b.).
Figure 2.3
Characteristic curves showing: a. Narrow latitude; b. Wide latitude
(Curry et al, 1990, p.159)
The dynamic range of an x-ray film is equivalent to the film’s latitude. The range of
OD on an x-ray film, typically 0.25 to 3.0, is equivalent to a dynamic range of 1:100
(Dowsett et al, 1998).
The effective energy of the x-ray beam can also affect film latitude. An increase of
effective energy of the beam, either through increasing the kVp or through the
13
addition of filtration to the beam, will increase the latitude (Bushberg et al, 2002;
Curry et al, 1990; Fauber, 2000; Gunn, 2002).
2.3
Radiographic Contrast
Radiographic contrast is a measure of the difference between the OD of one region in
the image and the OD of another region. Radiographic contrast can be measured
objectively through the use of densitometers, which measure OD at a point on the
x-ray film. Measurement of two ODs at different points on a film will provide a
measurement of radiographic contrast. Radiographic contrast can also be assessed
subjectively. When an x-ray image comprises mainly areas of high OD (black) and
low OD (white), it is referred to as exhibiting high radiographic contrast. When the
image has a broad range of OD, typified by the highest OD not being black and
lowest OD not being white, it is referred to as exhibiting low radiographic contrast or
wide latitude. Radiographic contrast is directly proportional to the angle, γ, of the
straight line regions of the characteristic curve of the F/S (Bushberg et al, 2002;
Fauber, 2000; Gunn, 2002). Comparison of angles of the straight line of the F/S
characteristic curves can be seen in Figure 2.3. Figure 2.3a has a high γ and high
radiographic contrast, whereas Figure 2.3b has a low γ and low radiographic contrast.
Figure 2.4 shows a comparison of high and low radiographic contrast images. The
image in Figure 2.4a has an appearance of strong blacks and whites. It exhibits high
radiographic contrast. The bones of the feet are better visualised in this image than in
Figure 2.4b. Figure 2.4b exhibits low radiographic contrast. This image displays the
soft tissue regions such as muscles better than Figure 2.4a (Bushong, 2001; Fauber,
2000; Gunn, 2002).
14
Figure 2.4
Radiographic images with:
a. High radiographic contrast
b. Low radiographic contrast
Radiographic contrast within the image results from the variable attenuation of the
x-ray beam and from the latitude. High radiographic contrast occurs when there are
large differences in attenuation within the x-ray field and/or there is a narrow F/S
latitude or dynamic range and/or there is a low effective x-ray beam energy. Low
radiographic contrast or wide latitude occurs when there are small differences in
attenuation within the x-ray field and/or there is a wide F/S latitude or dynamic range
and/or there is a high effective x-ray beam energy. A low or high effective x-ray
beam energy is set by radiographers as either a low or a high kVp setting. The
amount of added filtration is usually constant within the x-ray tube and as such is not
usually considered to have an effect on radiographic contrast (Bushong, 2001;
Fauber, 2000; Gunn, 2002).
It is a general principle in radiography to maximise radiographic contrast for most
radiographic examinations. Bones are better visualised with high radiographic
15
contrast images. Soft tissue examinations and chest radiographic examinations are
exceptions to this general rule. High latitude films and/or high kVp techniques are
often used in the radiographic examination of the chest. Soft tissue such as muscle
has inherent low attenuation differences and hence has low subject contrast within
the tissue. Radiographic contrast is still maximised for soft tissue radiographic
examinations. Low radiographic contrast images will result from imaging soft tissue
anatomical regions (Bushong, 2001; Fauber, 2000; Gunn, 2002).
Radiographic contrast within the image is optimised so as to display the desired
anatomical area under examination. Optimisation is effected through the
radiographer’s selection of appropriate exposure factors such as the kVp and F/S
combination.
2.4
Limitations of Film/Screen Radiography
The dynamic range of a medical imaging system is the system’s ability to record the
signal and represent the anatomical detail. In general radiography, the signal is the
radiation that exits the body. The dynamic range will depend upon attenuation
characteristics of the body. In some instances the F/S recording devices may have a
narrower dynamic range than that of the exit radiation (Dowsett et al, 1998).
In F/S general radiography, radiographers must choose the desired contrast
appearance of the radiographic image. Radiographers may produce a radiographic
image with high contrast or a radiographic image with low contrast. Radiographers
can not produce a single image that demonstrates both high and low radiographic
contrast at the same time. Two radiographic exposures on two films must be made to
achieve an examination that has both high and low radiographic contrast or wide
latitude.
X-ray photon intensities or exposures that exit the anatomy must fall on the linear
region of the F/S characteristic curve. The linear region of the F/S characteristic
curve is shown in Figure 2.3 and is typically between OD values of 0.5 to 2.0. The
linear region of the F/S characteristic curve is effectively the dynamic range of the
16
F/S system. If x-ray exit intensities are outside this latitude or dynamic range, any
change in exposure will not result in a proportional change of OD that could be
visualised on the film.
Many regions of the body have areas that have both very high and very low
attenuation within a single x-ray field. Examples of such regions of the body are the
chest, where the lungs have low attenuation and the thoracic spine has relatively high
attenuation and the shoulder where peripheral regions have lower attenuation. Such
anatomical areas have a high dynamic range of exit x-ray intensities. When a low
effective energy (low kVp) x-ray beam is used, the difference in attenuation within
these areas is accentuated, and a very high radiographic contrast image will result. If
the exit beams that emerge from two or more regions are above the exposure
required to produce the Dmax of the image these regions will appear black on the
image. In this example, the exit intensities are above the dynamic range of the F/S
system. A linear increase in exposure will not result in a corresponding linear
increase in OD. No radiographic contrast difference between these regions will be
visualised and as such these areas will be indistinguishable. Similarly, if two or more
regions attenuate the beam so that no x-ray photons exit, the regions appear white
and are indistinguishable. This occurs even if the anatomical regions have different
levels of attenuation characteristics. Here the exit intensities are below the dynamic
range of the F/S system.
The usual solution to this problem in radiography is to increase the effective energy
of the beam. Radiographers increase kVp to increase the so-called penetration of the
beam. As kVp is increased there is a corresponding increase of the x-ray beam’s
effective energy. As the effective energy of the x-ray beam is increased there is a
decrease in attenuation of the beam and an increase in intensity of the exit beam. The
dynamic range of the exit radiation is reduced. The advantage of this method is that
latitude is increased and all anatomy is visualised in one image. The γ of the F/S
system is decreased and the linear response region of the characteristic curve is
increased. The disadvantage is that radiographic contrast will decrease across all
anatomy visualised in the image. The general principle of maximising radiographic
contrast has been upheld. An alternative method could be to undertake two
radiographic exposures to optimise radiographic contrast across all anatomy within
17
the field of interest. This approach is not generally undertaken as it requires two
exposures of ionising radiation and an increase in absorbed dose by the patient.
There are many anatomical regions of the body where large attenuation differences
occur or where there is a wide dynamic range within the x-ray field. Some examples
are provided in Table 2.3.
Table 2.3
Examples of regions where wide dynamic ranges exist within the
anatomy (Ballinger, 1991)
Anatomical
Explanation of Wide Dynamic Range
Region
chest
large attenuation differences between the air-filled lungs and the thoracic
spine
thoracic spine
large anatomical thickness differences between the superior and inferior
portions of the thoracic spine
shoulder
large anatomical thickness differences between the edges of the
shoulder and the chest region within the image
facial bone
lateral image required to display both soft tissue and bony anatomy
cervical spine
lateral image required to display both the cervical vertebra and the
cervico-thoracic junction – large attenuating differences between the
cervical vertebra and the cervico-thoracic junction regions
cervical spine
lateral image required to display both soft tissue and bony anatomy
thoraco-lumbar
lateral image required to display both the thoracic and the lumbar spine
spine
hip / neck of
cross-table lateral used – large attenuation differences between the
femur
pelvic region and the neck of femur within the image
femur
large anatomical thickness differences between the superior and inferior
regions of the femur
feet
anatomical thickness differences between the tarsal region and the
metatarsal/phalanges of the foot
hands
lateral image required to display both the metacarpals and the
phalanges
abdomen
horizontal ray required to show air/fluid differences – attenuation
differences are further increased when barium is introduced to outline
the gut
18
2.5
Shaped Tissue Compensation Filters
Filtration material that is non-uniform in thickness may be placed in the x-ray beam
to compensate for the varying degrees of attenuation of the anatomy within the
irradiated area (Bushong, 2001; Curry et al, 1990; Thompson et al, 1994). Curry et al
(1990) describe the use of a wedge shaped filter material. Wedge shaped filters are
occasionally used to obtain radiographic images of more uniform optical density
when a part being examined diminishes greatly in thickness from one side of the
x-ray field to the other. (Curry et al, 1990)
According to Bushong (2001), one of the most difficult tasks facing the radiographer
is to produce an image with a uniform optical density when examining a body part
that varies greatly in thickness or tissue composition. Bushong recommends the use
of a compensating filter when this occurs. A trough shaped filter is recommended for
chest radiographic examinations and a wedge shaped filter for examinations of the
feet.
Thompson et al (1994) recommend the use of compensating filters when there are
large differences in tissue density within the anatomical region and the goal is to
visualise the entire structure without making additional exposures. They describe the
use of wedge filters and trough filters. Figure 2.5 shows sectional planes through
tissue compensation filters (TCFs) and their relationship to the x-ray beam and the
anatomy.
19
Figure 2.5
Cross-sections through wedge and trough TCFs (Thompson et al,
1994)
Other authors discuss various other uses and shapes of TCFs. Feczko et al (1983)
used various shaped filters in horizontal-beam lateral decubitus radiograph
examinations of barium filled abdomens. They (1983) performed 30 examinations, of
which 28 showed marked quality improvement through the use of TCFs. The shapes
of the TCFs used are shown in Figure 2.6.
Crow, Guinto & Segura (1983) examined the use of TCFs to compensate for thinner
anatomical regions when undertaking arch aortograms. TCFs have been reported to
improve the quality of radiographs of the shoulder (Vezina, 1985). Gray, Hoffman &
Peterson (1983) and Butler et al (1986) have reported on the use of TCF in
radiographic examinations of the scoliotic spine. A single exposure was used to
obtain excellent quality radiographs of the lower leg using a TCF (Petersen & Rohr,
1987). Marugg et al (1990) designed a holder to allow the use of multiple TCFs
during one radiographic examination.
20
Figure 2.6
Cross-sections
through
tissue
compensation
filters
used
for
horizontal-ray barium examinations (Feczko et al, 1983, p.849)
TCFs have proved to be useful devices to assist in overcoming the F/S limitation
where large subject attenuation differences exist within the irradiated field. The
insertion of a varying thickness TCF into the x-ray beam attenuates the beam by
different amounts within the field. There is an assumption that, prior to entering the
anatomy, the x-ray beam is uniform in intensity across the field. Insertion of the TCF
into the entrance beam alters the intensities across the field. The TCF modifies the
intensities within the entrance beam so that the exit beam has a reduced dynamic
range. The result is that the exit beam is more uniform or has a narrower dynamic
range whilst still containing exit intensity differences so that OD differences exist
within the image.
TCF sizes and shapes are predetermined. Radiographers select the most appropriate
size and shape for a particular radiographic examination and patient shape, from a
limited range. Modification to the TCF size and shape can not be made easily.
Goodsitt et al (1998) examined TCF use in mammography. Their object was to
develop a range of TCF shapes to suit individual breast shapes. They concluded that
three or four TCF shapes would be needed to match the contour shape of the
21
compressed
breast
for
all
mammography
examinations.
Mammography
radiographers would then be able to select the most appropriate TCF shape for
individual mammography examinations.
The advantage of using a TCF in F/S radiography where there are anatomical regions
with large attenuation differences within the x-ray field, is that all of the anatomy can
be visualised with only one exposure of ionising radiation to the patient.
Radiographers can select an appropriate kVp to optimise radiographic contrast
without the concern of under- or over-exposing areas of anatomy within the x-ray
field. Radiographic contrast is maximised over the entire image whilst still enabling
all anatomy to be visualised.
Radiographic image optimisation is generally within a radiographer’s autonomous
duties, and the appropriate use of TCFs in the radiographic examination is part of
image optimisation. Selection of the appropriately shaped TCF and precise
placement of the TCF within the beam are judgement issues of the radiographer
performing the radiographic examination.
2.6
Computed Radiography
Computed radiography (CR), as an alternative to screen-film general radiography, is
another means of image capture, storage and display in diagnostic radiography. The
first CR system was announced by the Fuji Photo Film Company in 1981
(Schaetzing et al, 1990). Since then such systems have been manufactured and
marketed by a number of different companies. There are many advantages and
disadvantages between CR and F/S. These are discussed in detail in Chapter 4.
Computed radiography uses photostimulable phosphors (PSPs) to capture a latent
image on the imaging plate. PSPs are barium fluorohalides activated with europium
(BaFX:Eu2+ where the halide X is typically bromine or iodine). X-ray photon energy
ionises the Eu atoms and the free electrons become trapped in so called F-centres.
During reading out of the image, a red laser scans and exposes the imaging plate. The
photon energy of the red laser light is absorbed by the F-centres, allowing the trapped
22
electron to become mobile and reabsorbed by the Eu atoms. During this deexcitation, energy is released in the form of blue-green light. A picture element or
pixel is a quantised value that represents a small area at location (x, y) in a digital
image (Baxes, 1994; Jain, 1989). The intensity of the blue green light is recorded and
converted to a pixel value. The resulting CR image is comprised of multiple rows
and columns of pixels representing the x-ray photon intensities at locations (x, y)
(Bushberg et al, 2002; Bushong, 2001; Dowsett et al, 1998; Weiser, 1997).
One advantage of CR over F/S is the increased dynamic range of CR images (Balter,
1990; MacMahon & Vyborny, 1994; Schaetzing et al, 1990; Siebert, Shelton &
Moore, 1996; Weiser, 1997). Dynamic range in CR is equivalent to F/S exposure
latitude. Figure 2.7 (Bushberg et al, 2002) provides a comparison of the dynamic
ranges of CR and typical F/S exposure latitude. CR is linear to x-ray exposure over a
greater range of exposures than is F/S. CR has a resulting increased exposure latitude
or dynamic range over F/S. A typical dynamic range of the imaging plate used in CR
to capture the image is 1,000:1 (Vuylsteke & Schoeters, 1994).
Radiographic contrast in F/S is limited by the exposure latitude. High radiographic
contrast results when the exposure latitude is narrow. The greater dynamic range of
CR allows x-ray photon exit intensities from high or low attenuating anatomical
regions, which could fall outside the exposure latitude of F/S, to be recorded. In
Figure 2.7, exposures of 0.01 and 0.1 would result in the same OD (white) using an
F/S combination. Using these exposures in CR, differences in CR signal exist and
OD differences can be visualised that would not be visualised using F/S.
Viewers of CR images can take advantage of CR’s greater dynamic range through
manipulation of the displayed brightness and contrast of the image. Look-up tables
(LUTs) are means of controlling the displayed brightness and contrast of digital
images (Artz, 1997; Baxes, 1994; Freedman & Artz, 1997b). The image pixel values
are converted through the LUT to display values visualised on the computer monitor.
CR images can be displayed with a broad or narrow displayed contrast. Narrow
displayed contrast images are similar in appearance to a narrow radiographic contrast
in F/S radiographic imaging. Broad displayed contrast images are similar in
appearance to a wide radiographic contrast in F/S radiographic imaging.
23
Figure 2.7
Dynamic range of CR vs exposure latitude of F/S (Bushberg et al,
2002, p.296)
Where there are anatomical regions with large attenuation differences within the
x-ray field, the advantage of CR over F/S is that all of the anatomy can be visualised
with one exposure of ionising radiation to the patient. The displayed CR image can
then be adjusted to optimise the displayed brightness and contrast of individual
anatomical regions within the image.
24