Download Pressure-Volume Theory of Operation

Survey
yes no Was this document useful for you?
   Thank you for your participation!

* Your assessment is very important for improving the workof artificial intelligence, which forms the content of this project

Document related concepts

Molecular scale electronics wikipedia , lookup

Decibel wikipedia , lookup

Ohm's law wikipedia , lookup

Transcript
Pressure-Volume
Theory of Operation
Pressure-Volume Technology
Conductance Theory of Operation
Deriving ventricular volume from a Conductance Catheter is
based on a very simple electrical principle: Ohm’s Law:
Voltage (V) = Current (I) X Resistance (R)
V = IR
Conductance (G) rather than resistance is the parameter
of interest. Since conductance is the inverse of resistance,
Ohm’s Law can be rewritten as:
Voltage = Current/Conductance
V = I/G
Conductance Catheters are comprised of both excitation
electrodes and recording electrodes. The excitation
electrodes (most distal and proximal electrodes on the
Catheter) generate an electrical field inside the heart from
the aortic valve to the apex. This field is generated as a
result of an alternating current being applied (at a constant
magnitude) between these 2 outermost electrodes. The
inner recording electrodes measure voltage change which is
proportional to a change in resistance.
The
electrical Scisense
field cannot
be restricted
to just the blood
Transonic
Technical
Note
volume and must pass through some of the cardiac muscle.
This means that the measured conductance value (Gx) is
actually a combination of blood conductance (Gb) and muscle
or parallel conductance (Gp).
Conductance excitation electrodes create an
electric field while
sensing electrodes
measure
A. Henton
and F. Konecny
the voltage change which allows for the
March, 2013
calculation of resistance and conductance.
mathematical
reviewaof
SV correction
InA1981,
Dr. Baan et. al. proposed
relationship
between factors employed
time-varying measurements of total conductance (Gx)
equation and Wei’s equation:
to time-varying changes in ventricular volume (Vol). This
volume formula takes into account the distance between the
recording electrodes (L), blood resistivity (ρ), and the parallel
conductance (Gp). It also takes into account the non-uniform
nature
of the
electrical field with the field correction factor,
Baan’s
equation:
Where:
alpha (α).
ρ = blood resistivity
L = measuring electrodes distance
in Baan’s
BAAN’S EQUATION
𝑆𝑆𝑆𝑆
α = SV
correction factor = 𝑐𝑐𝑐𝑐𝑐𝑐𝑐𝑐𝑐𝑐𝑐𝑐𝑐𝑐𝑐𝑐𝑐𝑐𝑐𝑐𝑐𝑐 A. Henton and F. Konecny
Transonic Scisense Technical
Note
1
𝑆𝑆𝑆𝑆𝑟𝑟𝑟𝑟𝑟𝑟𝑟𝑟𝑟𝑟𝑟𝑟𝑟𝑟𝑟𝑟𝑟𝑟
March, 2013
𝑉𝑉𝑉𝑉𝑉𝑉 = 𝜌𝜌𝜌𝜌2 (𝐺𝐺𝑥𝑥 −G.𝑝𝑝 )
Gx = measured TOTAL conductance
𝛼𝛼
Conductance uses a circuit model where both
conductance
Gp = parallel/muscle
ρ = Blood resistivity
(Gb) equations
and cardiac
(Gm) are conductive
Therefore, we can compare SV correction
process between Baan’sblood
and Wei’s
bymuscle
the following:
measured saline
together
as a single conductance
(assumed to be removed byand
hypertonic
injection)
L = Measuring electrode distance
value (G𝐺𝐺
and phase components are ignored.
x)𝑏𝑏
𝑆𝑆𝑆𝑆𝑐𝑐𝑐𝑐𝑐𝑐𝑐𝑐𝑐𝑐𝑐𝑐𝑐𝑐𝑐𝑐𝑐𝑐𝑐𝑐𝑐𝑐
)
VS
α = Baan’s SV correction factor = (
𝑆𝑆𝑆𝑆𝑟𝑟𝑟𝑟𝑟𝑟𝑟𝑟𝑟𝑟𝑟𝑟𝑟𝑟𝑟𝑟𝑟𝑟
𝛾𝛾
Gx = Measured total conductance
(1 −
)
Note: for the purpose
ofwethis
document
weBaan’s
will ‘alpha’
not use
a hypertonic
saline
processfactor
to calculate
However,
must
recognize that
correction
is a constant,
whilebolus
Wei’s correction
is
Gp = Baan’s parallel/muscle
conductance
(assumed
dynamic
and
defined
by
the
following:
gamma,
a
non-linear
constant,
and
Gb,
the
unique
value
of
blood
the average
Vp to satisfy
Baan’s equation.
Instead we will employ the admittance correction process
to be removed
by hypertonic
saline injection)
measured throughout the cardiac cycle. Therefore, it is our opinion that to compare these
reliant on phase conductance
shift measurement.
By doing this we can better compare resulting volume
two correction processes it is best to use a sample data set and follow the process of each equation.
measurements derived by each Baan’s and Wei’s equations and understand the effect of each unique SV
RPV-3-tn Rev. A 4/13
The following guideline shows how to compare the SV correction processes using LabScribe2 software.
correction factor. Therefore, after satisfying Vp by admittance we can understand Baan’s equation as:
Pressure-Volume
Pressure-Volume Technology Theory of Operation Cont.
Conductance Theory of Operation Cont.
However, Baan assumed alpha to be a constant with a value of one for a uniform current field distribution
(in reality electrical field strength decreases non-linearly with distance). Alpha can be calculated from the SV
conductance ratio (see equation on page 1) or by cuvette calibration. Both of these methods give a single
constant value for alpha. Parallel or muscle conductance (Gp) is often determined by hypertonic saline injection
which temporarily changes blood conductance but not myocardial conductance; allowing for the parallel
conductance value to be determined from the graph of changing conductance. This produces a single constant
value for parallel conductance.
IMPACT OF PHYSIOLOGY ON CONDUCTANCE MEASUREMENTS
At systole there is relatively little blood in the ventricle which
means that a larger portion of the electrical field passes through
the myocardium. Thus, myocardial resistance contributes more to
the total measured conductance than blood at this time. However,
because the hypertonic saline bolus method provides an average
measurement of muscle contribution, it is typical for the derived
volume to be overestimated at systole.
At diastole there is a large quantity of blood in the ventricle and the
heart walls have expanded. This means that most of the electrical
field is passing though blood with a very small contribution from the
myocardium. Thus, the measured conductance value is almost entirely At Systole
blood conductance. However, the same value of parallel conductance
is still subtracted from the total conductance which leads to an under
estimation of blood volume.
The electrical field strength decreases in a non-linear manner with
increasing field size. This means measurements of blood conductance
further from the Catheter do not have the same strength as those
nearer to the Catheter. Without correction this leads to an under
estimation of total volume. The larger the volume which is being
measured, the greater the under estimation: volume measurements
at diastole are thus more prone to under estimation than those at
systole. Alpha attempts to correct some of this error but fails to
address the non-linearity of the electric field or the varying strength
the under estimation has at different phases of the heart cycle.
At Diastole
Pressure-Volume
Transonic Scisense Technical Note
A. Henton and F. Konecny
March, 2013
Pressure-Volume Technology Theory of Operation Cont.
A mathematical review of SV correction factors employed in Baan’s
Admittance Theory of Operation
equation and Wei’s equation:
Admittance technique is an extension of the Conductance
method which measures both resistive and capacitive properties
of blood and muscle. In the electric field blood is purely resistive,
but muscle has both capacitive and resistive properties. This
allows
separation
of the muscle component of conductance from
Baan’s
equation:
Where:
Transonic
Technical
Note
A. Henton and F. Konecny
that
of blood,Scisense
using electric
field theory.
ρ = blood resistivity
March, 2013
The capacitive property of muscle causes a time (phase) delay
L = measuring electrodes distance
in measured signal (see graph at bottom right). By tracking this𝑆𝑆𝑆𝑆
α = SV
correction
factor = 𝑐𝑐𝑐𝑐𝑐𝑐𝑐𝑐𝑐𝑐𝑐𝑐𝑐𝑐𝑐𝑐𝑐𝑐𝑐𝑐𝑐𝑐
delay known
phase angle in real time
and
mathematically
1 as the
𝑆𝑆𝑆𝑆𝑟𝑟𝑟𝑟𝑟𝑟𝑟𝑟𝑟𝑟𝑟𝑟𝑟𝑟𝑟𝑟𝑟𝑟
2
relating
to Correction:
the resistance
tissue, the ADV500
Wei’s=itSV
𝑉𝑉𝑉𝑉𝑉𝑉
𝜌𝜌𝜌𝜌
(𝐺𝐺𝑥𝑥 −G.of𝑝𝑝 )the myocardial
Gx = measured
TOTAL conductance
𝛼𝛼 continuous, non-invasive tracking
System allows
of muscle/
=
parallel/muscle
conductance
GSV
pheartbeat. The phase
parallel
conductance
(Gm) throughout
the
To demonstrate
the calculations
for Wei’s
correction factor, we will use two sample points in the
(assumed
be
removed
by hypertonic saline injection)
angle reports heart tissueTransonic
intrusion into
the
field to
as
the
heart Note
Scisense
Technical
Henton and F. K
cardiac cycle (End-SystoleTransonic
and End-Diastole
as determined
by dP/dt
inflection points). This is required as A.
Scisense
Technical
A. Henton and F. K
contracts and expands, and
as expected,
this measurement
is Note
Transonic
Scisense
Technical
Note
A. Henton and
F. K
Marc
Wei’s SVatcorrection
is dynamic
and
therefore
unique
to any
stage of the cardiac cycle. The same
Marc
greatest
systole and
lowest at
diastole.
This
provides
a great
Marc
advantage
overabove
classical
conductance
volumetry
which treats
variables used
in Baan’s
SV correction
will apply:
Note: for
the purposeasofa this
document
wethan
will not
use a hypertonic saline bolus process to calculate
parallel
conductance
constant,
rather
a dynamic
Admittance uses a circuit model where blood
Wei’s
SV
Correction:
variable
which
changes
throughout
the
cardiac
cycle.
L = 4.5
mm Vp to satisfy Baan’s
the
average
Instead we will employ the admittance correction process
Wei’sequation.
SV Correction:
Wei’s SV Correction:
is conductive (Gb) and cardiac muscle is both
The
ADV500
system
an equation
developed
by Dr.
m) and capacitive (Cm).
reliant
on phase
shiftemploys
measurement.
By doing
this we can
better compareconductive
resulting(Gvolume
To
demonstrate
the
calculations
SV
correction
factor,
we
will
use
two
sample points
in
th
ρ = 1.2 Ohm·m
Chia-Ling
Wei to convert conductance
to volume
instead for
of Wei’s
To
demonstrate
the
calculations
for
Wei’s
SV
correction
factor,
we
will
use
two
points
in
th
measurements derived byTo
each
Baan’s andthe
Wei’s
equationsfor
and
understand
the effect
of each
unique
SV sample
demonstrate
calculations
Wei’s
SV
correction
factor,
we
will
use
two
sample
points
in
th
cardiac
cycle
(End-Systole
and
End-Diastole
as
determined
by
dP/dt
inflection
points).
This
is
requ
the traditional Baan’s equation.
In
Baan’s
equation
the
Field
cardiac
cycle
(End-Systole
and
End-Diastole
as
determined
by
dP/dt
inflection
points).
This
is
requ
correction
factor.
Therefore,
after
satisfying
Vp
by
admittance
we
can
understand
Baan’s
equation
as:
cardiac
cycle
(End-Systole
and
End-Diastole
as
determined
by
dP/dt
inflection
points).
This
is
requ
Correction
alpha
(α),
is assumed
to beisconstant
despite
SVreference =Factor,
32uL (as
derived
from
hi-resolution
echocardiography)
Wei’s
SV
correction
dynamic
and
therefore
unique
to
any
stage
of
the
cardiac
cycle.
The
same
Wei’s
SV
correction
is
dynamic
and
therefore
unique
to
any
stage
of
the
cardiac
cycle.
The
same
the non-linear nature of the
electrical
field. However,
Wei’s
Wei’s
SV
correction
is
dynamic
and
therefore
unique
to
any
stage
of
the
cardiac
cycle.
The
same
variables
used
above
in
Baan’s
SV
correction
will
apply:
variables
used
above
in
Baan’s
SV
correction
will
apply:
equation
corrects
for
the
nonhomogeneous
nature
of
the
1
Gb-ED = 855 uS =20.855 mS
variables used above in Baan’s SV correction will apply:
Catheter’s
by assuming a non-linear
𝑉𝑉𝑉𝑉𝑉𝑉 = electrical
𝜌𝜌𝜌𝜌 (𝐺𝐺field
𝑏𝑏 ) distribution
LL == 4.5
mm
𝛼𝛼
relationship
between
conductance
and
4.5
mm
Gb-ES = 465 uS = 0.465 mSL = 4.5 mm volume, gamma (γ), thus
improving accuracy over a wider volume range.
ρ = 1.2 Ohm·m
1.2
To measure blood volumeρ
time values are needed for
ρin== real
1.2 Ohm·m
Ohm·m
myocardial conductivity and permittivity
(for σ/ε ratio/heart
2 −4𝑎𝑎𝑎𝑎
SV
== 32uL
(as
derived from
hi-resolution
echocardiography)
reference
−𝑏𝑏
±√𝑏𝑏
𝐺𝐺𝑏𝑏 Equation:
SV
32uL
(as
from
hi-resolution
type),
blood
resistivity
(
ρ),
and
reference
stroke
volume
(SV).
reference
Wei’s
(as derived
derived
from
hi-resolution echocardiography)
echocardiography)
�1 − 𝛾𝛾 � , 𝑤𝑤ℎ𝑒𝑒𝑒𝑒𝑒𝑒 γ = SVreference = 32uL Where:
Default values for ‘Heart Type’2𝑎𝑎
(S/E ratio) ρare
provided
and most
=
blood
resistivity
== 855
uS
== 0.855
mS
commonly used. However,Gb-ED
researchers
can
study
this value
using
Gb-ED
mS
measuring
electrodes
distance
Gb-ED
= 855
855 uS
uS
=L =0.855
0.855
mS
a tetrapolar surface probe provided with the ADV500. Stroke
BLOOD conductance
G0.465
Gb-ES
=
465
uS
=
mS
b = measured
volume can be 2measured via
other
technologies.
Gb-ES
= 465
uS
0.465
mS
uS ==−
0.465
mSusing2 (0.855
𝑎𝑎 = 𝑆𝑆𝑆𝑆 − 𝜌𝜌𝐿𝐿 (𝐺𝐺𝑏𝑏−𝐸𝐸𝐸𝐸 −Gb-ES
𝐺𝐺𝑏𝑏−𝐸𝐸𝐸𝐸=)465
 32
(1.2)(4.5)
− 0.465)
=involving
22.5230
(calculated
admittance
process,The
removal
of a “delay” compared
output voltage
shows
parallel/muscle
conductance
using
phase
measurement)
to
the
input
voltage
signal
used to generate
ρ
=
Blood
resistivity
WEI’S EQUATIONS
𝑏𝑏 = −𝑆𝑆𝑆𝑆 · (𝐺𝐺𝑏𝑏−𝐸𝐸𝐸𝐸 + 𝐺𝐺𝑏𝑏−𝐸𝐸𝐸𝐸 ) −32(0.855
0.465)
=2 −4𝑎𝑎𝑎𝑎
- 42.2400
2 −4𝑎𝑎𝑎𝑎
−𝑏𝑏+±√𝑏𝑏
−𝑏𝑏
±√𝑏𝑏
L = Measuring
electrode
distance
2 −4𝑎𝑎𝑎𝑎
1
±√𝑏𝑏
2 �1 − 𝐺𝐺𝐺𝐺𝑏𝑏𝑏𝑏 � , 𝑤𝑤ℎ𝑒𝑒𝑒𝑒𝑒𝑒
2
γ = γ = −𝑏𝑏
−4𝑎𝑎𝑎𝑎
𝑉𝑉𝑉𝑉𝑉𝑉 =
) 𝐺𝐺𝛾𝛾𝑏𝑏��G,, 𝑤𝑤ℎ𝑒𝑒𝑒𝑒𝑒𝑒
2𝑎𝑎±√𝑏𝑏
�1𝑏𝑏−
γγ == −𝑏𝑏
𝐺𝐺𝑏𝑏 𝜌𝜌𝜌𝜌 (𝐺𝐺
2𝑎𝑎
= Measured
blood
𝛾𝛾 b𝑤𝑤ℎ𝑒𝑒𝑒𝑒𝑒𝑒
𝑐𝑐 = 𝑆𝑆𝑆𝑆 · (1−
𝐺𝐺𝑏𝑏−𝐸𝐸𝐸𝐸
·) 𝐺𝐺𝑏𝑏−𝐸𝐸𝐸𝐸 �1
 −32(0.855
∙
0.465)
=
12.7224
2𝑎𝑎
𝛾𝛾
2𝑎𝑎
𝛾𝛾
conductance
where,
Field Correction Factor:
SV = Stroke volume
the electric field. The signal delay, caused by
myocardial capacitance, is measured in terms of
degrees and is referred to as “Phase angle θ.” The
admittance magnitude (conductance) is impacted
by both the blood and muscle.
2 (𝐺𝐺
2
2 (𝐺𝐺 )  32 − )
𝑎𝑎
=
𝑆𝑆𝑆𝑆
−
𝜌𝜌𝐿𝐿
𝐺𝐺
(1.2)(4.5)
2 (𝐺𝐺
2 (0.855 − 0.465) = 22.5230
𝑎𝑎 = 𝑆𝑆𝑆𝑆 −
− 𝜌𝜌𝜌𝜌
−32
𝐺𝐺𝑏𝑏𝑏𝐸𝐸𝐸𝐸
𝑏𝑏−𝐸𝐸𝐸𝐸
𝑏𝑏𝑏𝐸𝐸𝐸𝐸
)) 
𝑎𝑎
=
𝑆𝑆𝑆𝑆
−
𝜌𝜌𝐿𝐿
−
𝐺𝐺
−
(1.2)(4.5)
2 (𝐺𝐺𝑏𝑏−𝐸𝐸𝐸𝐸
2 (0.855 − 0.465) = 22.5230
𝑏𝑏−𝐸𝐸𝐸𝐸
𝑏𝑏−𝐸𝐸𝐸𝐸
𝑎𝑎
=
𝑆𝑆𝑆𝑆
−
𝜌𝜌𝐿𝐿
−
𝐺𝐺

32
−
(1.2)(4.5)
(0.855 − 0.465) = 22.5230
𝑏𝑏−𝐸𝐸𝐸𝐸
𝑏𝑏−𝐸𝐸𝐸𝐸
2
−𝑏𝑏 ±√𝑏𝑏2 −4𝑎𝑎𝑎𝑎
42.24 ±�−42.24 −4(22.5230 ∙ 12.7224)
𝑏𝑏 = −𝑆𝑆𝑆𝑆
· (𝐺𝐺𝑏𝑏𝑏𝐸𝐸𝐸𝐸
𝐺𝐺𝑏𝑏𝑏𝐸𝐸𝐸𝐸 )= 1.4984
γ=
 𝑏𝑏𝑏𝑏 =
and
)+−32(0.855
𝐺𝐺
−𝑆𝑆𝑆𝑆 ·· (𝐺𝐺
+
0.465)
== --0.3770
42.2400
𝑏𝑏−𝐸𝐸𝐸𝐸 +
𝑏𝑏−𝐸𝐸𝐸𝐸
𝐺𝐺
−32(0.855
+
0.465)
42.2400
2𝑎𝑎
𝑏𝑏−𝐸𝐸𝐸𝐸 +
𝑏𝑏−𝐸𝐸𝐸𝐸 )
(𝐺𝐺2(22.5230)
)
+
𝐺𝐺
𝑏𝑏 =
= −𝑆𝑆𝑆𝑆
−𝑆𝑆𝑆𝑆 · (𝐺𝐺
−32(0.855
+
0.465)
=
42.2400
𝑏𝑏−𝐸𝐸𝐸𝐸
𝑏𝑏−𝐸𝐸𝐸𝐸
𝑐𝑐 = 𝑆𝑆𝑆𝑆 · 𝐺𝐺𝑏𝑏𝑏𝐸𝐸𝐸𝐸 · 𝐺𝐺𝑏𝑏𝑏𝐸𝐸𝐸𝐸
𝑐𝑐𝑐𝑐 =
𝑆𝑆𝑆𝑆 ·· 𝐺𝐺

∙ 0.465)
== 12.7224
𝑏𝑏−𝐸𝐸𝐸𝐸 ·· 𝐺𝐺
𝐺𝐺𝑏𝑏−𝐸𝐸𝐸𝐸
 32(0.855
32(0.855
0.465)
12.7224
𝑏𝑏−𝐸𝐸𝐸𝐸
Note: Wei’s equation always
the larger
postive
solution ∙for
γ (1.4984)
𝑐𝑐 =
= 𝑆𝑆𝑆𝑆
𝑆𝑆𝑆𝑆uses
· 𝐺𝐺
𝐺𝐺𝑏𝑏−𝐸𝐸𝐸𝐸
𝑏𝑏−𝐸𝐸𝐸𝐸 · 𝐺𝐺𝑏𝑏−𝐸𝐸𝐸𝐸  32(0.855 ∙ 0.465) = 12.7224
Pressure-Volume
Comparing Conductance vs Admittance
The Scisense ADV500 Pressure-Volume System is capable of being used in either Conductance or Admittance
mode. Both methods have value depending on what the researcher is looking to observe.
Conductance
Admittance
• Measures voltage magnitude
• Measures voltage magnitude and phase angle
• Harder to determine position of Catheter in
ventricle
• Uses Baan’s equation to determine volume
• Parallel conductance (Gp) is assumed constant
• Parallel conductance determined from
hypertonic saline injection after the
experiment
• Field Correction Factor is assumed constant
(alpha)
• Requires empirical reference stroke volume
to derive α or an approximation (typically 1)
can be used
• Volume calculation is done post experiment
with no chance to correct for protocol or
surgical errors.
• Tends to overestimate volume due to constant
nature of α as observed with echocardiography.
• Traditional technique with a solid body of
papers that validate the basic principle of
conductance catheter volumetry.
• Phase angle useful in locating Catheter in ventricle
• Uses Wei’s equation to determine volume
• Muscle conductance (Gm) varies throughout cardiac
cycle
• Parallel conductance determined from phase shift in
real-time (no hypertonic saline injection required)
• Requires sigma/epsilon ratio (conductivity/
permittivity) of heart muscle. Default values
are commonly used, or can be measured using
tetrapolar Calibration Probe.
• Field Correction Factor is non-linear (gamma)
• Requires empirical reference stroke volume to derive
gamma (γ)
• Volume calculation is in real time. Corrections to
experimental protocol or surgery can be made before
experiment is concluded.
• Closer approximation to absolute systolic and diastolic
volume as observed with echocardiography.
• Innovative technology that builds directly on the
foundation of conductance catheter volumetry.
Conductance method
measure pressure and
magnitude in real-time,
creating pressuremagnitude loops. Volume
can only be calculated
post-experiment.
Admittance method
measure pressure,
volume, phase
and magnitude in
real-time, creating
pressure-volume loops.
Transonic Systems Inc. is a global manufacturer of innovative biomedical measurement
equipment. Founded in 1983, Transonic sells “gold standard” transit-time ultrasound
flowmeters and monitors for surgical, hemodialysis, pediatric critical care, perfusion,
interventional radiology and research applications. In addition, Transonic provides
pressure and pressure volume systems, laser Doppler flowmeters and telemetry systems.
www.transonic.com
AMERICAS
EUROPE
ASIA/PACIFIC
JAPAN
Transonic Systems Inc.
34 Dutch Mill Rd
Ithaca, NY 14850
U.S.A.
Tel: +1 607-257-5300
Fax: +1 607-257-7256
[email protected]
Transonic Europe B.V.
Business Park Stein 205
6181 MB Elsloo
The Netherlands
Tel: +31 43-407-7200
Fax: +31 43-407-7201
[email protected]
Transonic Asia Inc.
6F-3 No 5 Hangsiang Rd
Dayuan, Taoyuan County
33747 Taiwan, R.O.C.
Tel: +886 3399-5806
Fax: +886 3399-5805
[email protected]
Transonic Japan Inc.
KS Bldg 201, 735-4 Kita-Akitsu
Tokorozawa Saitama
359-0038 Japan
Tel: +81 04-2946-8541
Fax: +81 04-2946-8542
[email protected]