Download Flow Characteristics of the Medtronic CoreValve: Difficulties

Survey
yes no Was this document useful for you?
   Thank you for your participation!

* Your assessment is very important for improving the workof artificial intelligence, which forms the content of this project

Document related concepts

Cardiac contractility modulation wikipedia , lookup

Turner syndrome wikipedia , lookup

Marfan syndrome wikipedia , lookup

Echocardiography wikipedia , lookup

Hypertrophic cardiomyopathy wikipedia , lookup

Lutembacher's syndrome wikipedia , lookup

Jatene procedure wikipedia , lookup

Pericardial heart valves wikipedia , lookup

Quantium Medical Cardiac Output wikipedia , lookup

Mitral insufficiency wikipedia , lookup

Aortic stenosis wikipedia , lookup

Transcript
132
Original articles
|
March 2014 - Issue 3
Flow Characteristics of the Medtronic
CoreValve: Difficulties Estimating Aortic Valve
Cross-Sectional Area Following Transcatheter
Aortic Valve Implantation
Alison Duncan MRCP, PhD; Eric Lim FRCS; Sarah Barker MSc; Carlo Di Mario FRCP; Simon
Davies FRCP; Neil Moat FRCS
Abstract
Background: Echocardiographic evaluation after transcatheter aortic valve implantation (TAVI) includes estimation of effective
orifice area (EOA). EOA calculation depends on sub-valvular stroke volume (SV), which depends on sub-valvular diameter and
velocity time integral (VTI). The Medtronic CoreValve area changes throughout its length. We aimed to (i) compare SV at two
sites of flow acceleration: ‘pre-stent’ and ‘in-stent, pre-valve’, (ii) assess effects of possible differences in sub-valvular SV on
EOA, and (iii) assess agreement of measurement of EOA calculation after CoreValve TAVI.
Methods: We studied 43 patients after CoreValve implantation. All had transthoracic echocardiography 5-7 days after TAVI.
Sub-valvular SV was measured ‘pre-stent’ and ‘in-stent, pre-valve’. Measurement agreement was assessed by root mean
square (RMS) differences and Bland-Altman analyses.
Results: SV was consistently higher ‘in-stent, pre-valve’ compared with ‘pre-stent’ (62±20ml vs. 53±19ml, p<0.001), so that
EOA was correspondingly larger using ‘in-stent, pre-valve’ measurements (1.7±0.5cm2 vs. 1.4±0.5cm2, p<0.001). Betweenobserver RMS difference for calculation of EOA was higher ‘in-stent, pre-valve’ compared to ‘pre-stent’ (0.53 cm2 vs.
0.23cm2, difference from zero 0.17, p=0.002). Though sub-valvular diameter measurements were variable, VTI variability was
additionally higher ‘in-stent, pre-valve’ compared to ‘pre-stent’ (0.42cm vs. 0.6cm, difference from zero -1.74, p=0.11).
Conclusion: Calculation of EOA after CoreValve TAVI is highly dependent on sub-valvular sample position. EOA may be
underestimated using ‘pre-stent’ SV, and overestimated using ‘in-stent, pre-valve’ SV. Limitations in SV reproducibility
suggests EOA should be used in conjunction with other indices of valve function in serial assessment of CoreValve function
following TAVI.
Key words:
Transcatheter aortic valve implantation, Effective aortic valve area, Stroke volume, Pre-stent, In-stent pre-valve,
Agreement
Introduction
Aortic stenosis is the most frequent type of valvular
heart disease in Europe and North America, and Doppler
echocardiography is the preferred technique for assessing
the severity of native aortic valve stenosis and prosthetic
aortic valve function after surgical aortic valve replacement
(AVR)1. Trans-catheter aortic valve implantation (TAVI) is a
less-invasive alternative for patients with severe symptomatic
aortic stenosis who are at high-risk for surgical AVR2,3. Survival
after TAVI is higher compared to medical therapy alone4, and is
comparable with surgical AVR at 1-year5. The need for accurate
follow-up assessment of TAVI devices is therefore evident.
Echocardiography is the primary imaging modality for the
assessment of prosthetic valve function, including after TAVI6.
The American Society of Echocardiography and The European
Association of Echocardiography7-10 advocate an integrated
approach, using both flow-dependent (mean gradient) and
flow-independent (effective orifice area: EOA) assessment. If
there is discordance between these measurements, the Doppler
velocity index (DVI) is calculated.
Calculation of EOA is performed by echo-Doppler, and is based
on the use of the continuity equation. This requires calculation
of sub-valvular stroke volume (SV), obtained as the product
of velocity time integral (VTI) and cross-sectional area (CSA)
measured simultaneously at a pre-defined region below the
aortic valve. However, the design of transcatheter devices
complicates conventional calculation of EOA due to difficulties
pre-specifying the exact sub-valvular region of interest. The two
TAVI devices most commonly used in Europe and the USA are
the Edwards SAPIEN XT THV device (Edwards Lifesciences,
Irvine, California), and the Medtronic CoreValve prosthesis
(Medtronic, Minneapolis, Minnesota). There are important shape
differences between a fully deployed SAPIEN device and a
fully deployed CoreValve. The SAPIEN valve is cylindrical when
fully expanded and the neo-aortic leaflets are relatively close
to the ventricular end of the device (Figure 1A). In contrast,
the CoreValve is hour-glass shaped (minimum dimension at
neo-aortic valve leaflet level), and is a longer device, so that
the neo-aortic valve leaflets are relatively further away from
the ventricular end of the device (Figure 1B). These differences
in neo-aortic valve height are confirmed in representational
2-dimensional echo pictures shown in Figures 2A and 2B.
The unique shape of different TAVI devices has specific
influence on sub-valvular flow acceleration across the implanted
valve. In the SAPIEN device, flow acceleration occurs ‘prestent’ and ‘in-stent, pre-valve’11-13. Previous studies thus
recommend using sub-valvular SV values ‘pre-stent’ to avoid
‘in-stent’ sub-valvular flow acceleration. However, these studies
have assessed only the haemodynamics of the SAPIEN device.
March 2014 - Issue 3
|
133
Original articles
Figure 1: Comparing distance from base of device to base of aortic valve leaflets in Edwards SAPIEN vs. Medtronic CoreValve in vitro.
A: The distance from the ventricular aspect of the device to the base of the neo-aortic leaflets in the SAPIEN valves is short (7.7mm,
8.7mm, and 11.6mm in 23mm, 26mm, and 29mm SAPIEN devices respectively). Source: Edwards Lifesciences, Irvine, California.
The effect of the site of sub-valvular stroke volume calculation
on EOA following CoreValve implantation has received less
attention.
The aims of our study were therefore (i) to compare calculated
sub-valvular values of SV at two pre-defined sample volume
positions; ‘pre-stent’ and ‘in-stent, pre-valve’ in patients
undergoing purely CoreValve TAVI, and (ii) to assess the effect
of any differences in sub-valvular SV on calculated EOA. We
also sought to evaluate the agreement of measurements of subvalvular SV at both ‘pre-stent’ and ‘in-stent, pre-valve’ levels,
and the effect of sub-valvular variability on EOA in a group of
patients undergoing CoreValve TAVI.
Methods
Patients
Forty-three consecutive patients with symptomatic severe aortic
stenosis who underwent TAVI with a CoreValve at The Royal
Brompton Hospital, London, UK were studied. All patients
had severe calcific aortic stenosis (aortic valve area ≤1cm2 or
BSA-indexed AVA <0.6 cm2/m2, and mean aortic valve pressure
gradient >40mmHg). All underwent CoreValve implantation
using transoesophageal and fluoroscopic guidance. The study
complied with the Declaration of Helsinki, and was approved by
the Royal Brompton Research Committee. All patients provided
signed informed consent before the procedure.
Echocardiographic Assessment
All patients had a transthoracic echocardiogram performed
by the same echocardiographer between 5 and 7 days after
the TAVI procedure. Echocardiograms were obtained using a
Philips ie33 echo machine and a S5 multifrequency transducer
(Andover, Massachusetts, USA). With the subject at rest in
the semi-lateral position, standard 2-dimensional parasternal
long-axis views were used to determine end-diastolic and
end-systolic left ventricular (LV) cavity size (LVEDD and LVESD
respectively). The same views were also used to obtain a
zoomed full screen of the LV outflow tract, native aortic valve
(before TAVI), transcatheter device, neo-aortic valve (after TAVI)
and aortic root. Apical 5- and 3-chamber views were used to
obtain a zoomed full screen of the transcatheter device. The
presence of aortic regurgitation (AR) was assessed by colour
and continuous wave Doppler in zoomed parasternal long
and short axis windows, 5- and 3-chamber views, and from
supra-sternal imaging of proximal descending aorta. Severity
was graded semi-quantitatively as none, trivial, mild, moderate
or severe
according
to ASE
guidelines7.
Mitral
regurgitation
was also
assessed
according
to ASE
guidelines7.
Aortic Annulus
Lowest
distance aortic
Pulsed wave
valve from
Doppler
ventricular end
was used
device 12mm.
to obtain
sub-valvular
B: In contrast, the distance from the ventricular
VTI at two
separate and aspect of the device to the base of the
neo-aortic leaflets in the CoreValve (purple
distinct sites
semicircle and arrow) is longer, and is always
(see below).
12mm, irrespective of which CoreValve size
In general,
is used (Source: Medtronic, Minneapolis,
pulsed-wave
spectra were Minnesota).
optimised to
eliminate spectral broadening and to display modal velocities,
and the nominal pulsed-wave sample size was kept constant
at 4mm. Pulsed-wave Doppler recordings (both ‘pre-stent’ and
‘in-stent, pre-valve’) were obtained by a single observer in both
the 5- and 3-chamber views over 3 consecutive cardiac cycles.
Peak velocity and the velocity time integral (VTI) were traced.
For all pulsed-wave Doppler recordings, (both ‘pre-stent’
and ‘in-stent, pre-valve’) the average result from the 6 traces
obtained from 2 views was reported.
Peak and mean trans-prosthetic pressure gradients were
recorded using continuous wave Doppler from the apical
5-chamber, apical 3-chamber, and right parasternal windows,
over 3 consecutive cardiac cycles in each window. Highest
values were taken. LV end-diastolic (LVEDV) and end-systolic
(LVESV) volumes were measured in the apical 4-chamber and
2-chamber windows. Since all patients had either none or trace
mitral regurgitation, biplane stroke volume was calculated
using volumetric methods by subtracting LVESV from LVEDV.
LV ejection fraction was calculated using Simpson’s biplane
method. All recordings were made at a sweep speed of 100
mm/s, with a superimposed ECG (lead II).
134
Figure 2: Comparing distance from base of device to base of aortic
valve leaflets in Edwards SAPIEN vs. Medtronic CoreValve in vivo.
Original articles
|
March 2014 - Issue 3
Figure 3: Measurement of sub-valvular velocity at two different sites
in the same patient
A: ‘Pre-stent’ sub-valvular diameter, measured in mid-systole,
just underneath the apical margin of the valve stent (in this
example, diameter was 1.73cm).
Figure 2A: In a 26mm Edwards SAPIEN prosthesis, the neoaortic valve measured 0.9cm from the ventricular end of the
device in a transthoracic 5-chamber view.
B: ‘Ín-stent, pre-valve’ s ub-valvular diameter, measured in midsystole, just below the hinge-points of the neo-aortic valve (in
this example, diameter was 1.45cm).
Figure 2B: In a 26mm Medtronic CoreValve prosthesis by
comparison, the neo-aortic valve measured 1.2cm from
the ventricular end of the device in a similar transthoracic
5-chamber view.
Site-Specific Measurement of Sub-Valvular SV
Sub-valvular diameter: Using a zoomed 2D parasternal long
axis full-screen display, ‘pre-stent’ diameter was measured
(in mid-systole, using inner-edge to inner-edge methodology,
perpendicular to the TAVI device long axis) just underneath
the apical margin of the valve stent12-14. After that, ‘in-stent,
pre-valve’ diameter was measured (again using a zoomed
parasternal mid-systolic frame, using inner-edge to inner-edge
methodology perpendicular to the TAVI device long axis) just
below the hinge points of the visible prosthetic leaflets15. Crosssectional area at the ‘pre-stent’ and ‘in-stent, pre-valve’ levels
were then calculated as π(d/2)2, assuming a circular crosssection.
Sub-valvular VTI: Using a zoomed 2D full-screen display of the
trans-catheter valve in both an apical 5- and 3-chamber views,
‘pre-stent´ VTI was recorded in mid-systole with the pulsedwave Doppler sample volume just underneath the apical margin
C: ‘Pre-stent’ sub-valvular VTI, measured with sample volume
just underneath the apical margin of the valve stent (in this
example, VTI was 16cm).
D: ‘Ín-stent, pre-valve’ sub-valvular VTI, measured just below
the hinge points of the visible prosthetic leaflets (in this example,
VTI was 25cm)
March 2014 - Issue 3
|
135
Original articles
Figure 4: Between-observer agreement of calculated Effective Orifice Area
A
B
C
D
There was close agreement between two independent sonographers for calculating effective aortic orifice area (EOA) when subvalvular measurements were taken ‘pre-stent’ (A) compared to ‘in-stent, pre-valve’ (B). Bland-Altman plots also demonstrated less
variability in EOA calculations when sub-valvular measurements were taken ‘pre-stent’ (C) compared to ‘in-stent, pre-valve’ (D).
of the valve stent12-14. Subsequently, ‘in-stent, pre-valve’ VTI
was measured in mid-systole with the pulsed-wave Doppler
sample volume approximately 1cm below the neo-aortic valve9.
At both sub-valvular sites, the average result from the apical 5and 3-chamber views was recorded.
Doppler velocity index (DVI)
The Doppler velocity index (DVI) was calculated ‘pre-stent’
(using the ratio of the maximum ‘pre-stent’ sub-valvular velocity
to the peak trans-prosthetic velocity) and ‘in-stent, pre-valve’
(using the ratio of the maximum ‘in-stent, pre-valve’ sub-valvular
velocity to the peak trans-prosthetic velocity). DVI was also
obtained for VTI (rather than maximum velocity) using similar
methods.
Data analysis
According to previously reported data from the Placement of
Aortic Transcatheter Valve (PARTNER) study4, the standard
deviation for valve area for the SAPIEN valve is approximately
0.5 cm2. In a power calculation, a sample size of at least 20
patients was required in order for the present study to have
80% power to detect a mean difference in EOA >60% using
2 different sub-valvular sampling positions in the CoreValve
prosthesis at a 5% significance level. Statistical analysis
was performed using R Core Team 2013, R Version 3.0.2 (R
Foundation for Statistical Computing, Vienna, Austria. URL
http://www.R-project.org/.) Normally distributed continuous
variables were expressed as mean±standard deviation.
Continuous variables between groups were compared using
paired Student t-tests. A significance level of p<0.05 was
employed throughout the study.
Agreement
Fifteen patients were selected randomly to assess agreement
of calculated EOA at both ‘pre-valve’ and ‘in-stent, prevalve’ levels. Between-observer agreement of calculated
EOA was determined by having a second and experienced
echocardiographer perform a separate echocardiogram on
each patient (immediately after the initial full echocardiographic
study to minimise patient variability). During the two studies,
both sonographers independently performed their own
measurements of sub-valvular diameter and VTI, at ‘pre-valve’
and ‘in-stent, pre-valve’ levels. Peak aortic velocity and VTI was
measured by only the first sonographer as part of a complete
echo examination. Reproducibility was expressed as the root
136
Original articles
|
March 2014 - Issue 3
Figure 5: Between-observer Agreement of Calculated Stroke Volume depending on sample site.
A
B
C
D
E
F
G
H
Agreement was generally good for
diameter measurement at ‘pre-stent’ level
(A), though there was some variability
on Bland-Altman plots (B), particularly
when diameter measurements were
larger. Agreement was better for VTI
measurement at ‘pre-stent’ level (C), with
minimal variation in the mean difference
from zero (D).
Agreement was less good for diameter
measurement at ‘in-stent, pre-valve’ level
(E) compared to ‘pre-stent’, with the
second sonographer generally producing
higher measurements (F).
Agreement was worse for ‘in-stent, pre-valve’ VTI measurement (G) compared to ‘pre-stent’ level, with wide confidence intervals on a BlandAltman plot (H).
mean square (RMS) difference (± standard deviation of the
difference of the means) between duplicate values. BlandAltman plots were created to assess variability and bias16,
determined for both diameter and VTI measurements, at ‘prevalve’ and ‘in-stent, pre-valve’ levels. Finally, the overall effect
of agreement of these two measurements, at both sub-valvular
sites, on calculated EOA was determined.
Results
Baseline Characteristics
The 43 consecutive patients studied had a mean age of
79±8years, 21 of whom were male. Before TAVI, all patients
had severe symptomatic aortic stenosis (peak and mean transaortic pressure gradient were 80±17mmHg and 47±12mmHg
respectively, DVI was 0.22±0.06, and calculated aortic valve
area was 0.5±0.2cm2, Table 1). Sixteen patients received
a 26mm CoreValve, 22 patients a 29mm CoreValve, and 5
patients a 31mm CoreValve. A transfemoral route was used
in 35 patients, a subclavian route in 3 patients, and a direct
aortic approach in 5 patients. Seven patients underwent postimplantation balloon dilatation. Full final valve deployment was
achieved in all patients as determined by both fluoroscopic and
trans-oesophageal echocardiographic imaging. In no patient
was the CoreValve deemed to be low in the LVOT (measured on
fluoroscopy as depth of the ventricular end of the TAVI device
relative native non-coronary cusp of the aortic valve <10mm)15.
Transthoracic echocardiography was repeated 5-7 days
after TAVI. There were significant reductions in peak transaortic gradient (to 16±3mmHg), mean trans-aortic gradient
(to 5±2mmHg, both p<0.001), SV measured by Simpson’s
biplane (from 62± 24ml to 54± 23ml, p=0.04), but there were
no significant change in LV chamber size, chamber volumes, or
LVEF (Table 1).
Doppler velocity index (DVI)
Values for both sub-valvular velocity and VTI were significantly
higher ‘in-stent, pre-valve’ compared with ‘pre-stent´ (Table 2,
Figure 3a and 3b). Consequently, DVI was significantly higher
March 2014 - Issue 3
|
137
Original articles
Table 1: Baseline Echocardiographic Characteristics of the Study Population
Variable
LVEDV (ml)
Before TAVI
After TAVI
P value
125±56
121±45
0.23
LVESV (ml)
63±42
66±43
0.06
Biplane SV (ml)
62±24
54±23
0.04
LVEF (%)
52±16
46±16
0.06
LVEDD (mm)
51±10
50±8
0.08
LVESD (mm)
36±10
36±8
0.81
Peak trans-aortic gradient (mmHg)
80±17
16±3
<0.001
Mean trans-aortic gradient (mmHg)
47±12
5±2
<0.001
Peak trans-aortic VTI (cm)
121±23
39±11
<0.001
0.22±0.06
See table 2
NA
Doppler velocity index (V1/V2)
Data are expressed as mean±SD
LVEDV: left ventricular end-diastolic volume, LVESV: left ventricular end-systolic volume; LVEF: left ventricular ejection fraction, LVEDD:
left ventricular end-diastolic diameter; LVESD: left ventricular end-systolic diameter; ; V1: maximum velocity at sub-valvular level; V2 max:
maximum trans-prosthetic velocity
when ‘in-stent, pre-valve’ values were used compared to ‘prestent’ values (using peak velocity: 0.7±0.2 vs. 0.5±0.1; using
peak VTI: 0.7±0.2 vs. 0.5±0.2 respectively, both p<0.001, Table
2).
Sub-Valvular Stroke Volume
‘Pre-stent’: mean diameter was 2.0±0.1cm (Table 2), which
corresponded to a cross-sectional area (3.1±0.4cm2); VTI was
16.9±5.2cm, so that calculated SV was 53±20ml. This was not
different from SV calculated using Simpson’s biplane method,
(53±20ml vs. 55±23ml, p=0.81).
‘In-stent, pre-valve’: mean diameter was smaller (1.7±0.1cm)
compared to ‘pre-stent’ (Figure 3) so that cross-sectional area
was correspondingly lower (2.3±0.3cm2). However, VTI was
significantly higher compared to values ‘pre-stent’ (26.5±7.4cm,
Figure 3), so that SV was correspondingly greater (62±20ml,
all p<0.001). Calculated SV ‘in-stent, pre-valve’ using Doppler
methods was also significantly greater than SV calculated using
Simpson’s biplane method (62±20ml vs. 55±23ml, p=0.30).
Calculated Aortic Effective Orifice Area
Consistently higher values of SV at ‘in-stent, pre-valve’ level
resulted in larger calculated aortic EOA when ‘in-stent, prevalve’ values were used compared to ‘pre-stent’ values
(1.7±0.5cm2 vs. 1.4±0.5cm2, p<0.001, Table 2).
Agreement
Discussion
The rapid expansion in the number of patients with severe
aortic stenosis receiving a transcatheter aortic implantation
has brought a need for precise echo-Doppler evaluation of
transcatheter valve function. Previous studies in the relatively
short Edwards-SAPIEN valve recommend using the LVOT
diameter and velocity “obtained immediately proximal to the
stent” (i.e. ‘pre-stent’) to calculate sub-valvular SV. However,
site-specific calculation of sub-valvular SV has received little
attention in the other widely commercially available device in
Europe and the US – the Medtronic CoreValve. We report our
experience in patients after receiving the CoreValve prosthesis,
and highlight the influence of the precise site of measurement of
sub-valvular SV in calculating EOA.
Table 2: Flow Characteristics of the transcatheter CoreValve
depending on sub-valvular sample volume positioning
Sample site
Variable
‘Pre-stent’
‘In-stent,
pre-valve’
P
Peak sub-valvular
velocity (m/s)
0.9±0.2
1.3±0.4
<0.001
Peak sub-valvular VTI
(cm)
16.9±5.2
26.5±7.4
<0.001
Peak trans-aortic
velocity (m/s)
2.0±0.5
2.0±0.5
NA
Peak trans-aortic VTI
(cm)
39±11
39±11
NA
DVI (V1/V2)
0.5±0.1
0.7±0.2
<0.001
DVI (VTI1/VTI2)
0.5±0.2
0.7±0.2
<0.001
Diameter (cm)
2.0±0.1
1.7±0.1
<0.001
CSA (cm2)
3.1±0.4
2.3±0.3
<0.001
SV (ml)
53±19
62±20
<0.001
EOA (cm2)
1.4±0.5
1.7±0.5
<0.001
Data are expressed as mean±SD
CSA: cross-sectional area; Doppler velocity index (DVI); EOA:
effective valve area; SV: stroke volume; VTI: velocity time integral;
V1: maximum velocity at sub-valvular level; V2 max: maximum transprosthetic velocity; VTI1: maximum VTI at sub-valvular level; VTI2
max: maximum trans-prosthetic VTI
Findings
Our study demonstrates significant reductions in both peak and
mean trans-aortic gradients, and substantial increases in EOA,
one week after CoreValve implantation in consecutive patients
undergoing TAVI for severe aortic stenosis. Our study confirmed
that there are two regions of flow acceleration in the LV outflow
tract after CoreValve: the first at the ventricular aspect of the
device (‘pre-stent’), and the second within the valve (‘in-stent,
pre-valve’). We found significantly higher values for sub-valvular
SV at ‘in-stent, pre-valve’ level, which obviously led to a higher
DVI and a larger EOA at ‘in-stent, pre-valve’ level. Our EOA
estimations therefore are similar with those of by Jilaihawi et al
(15) when 7 values are used for EOA calculation, while our ‘pre-
138
Original articles
|
March 2014 - Issue 3
Table 3: Between-observer agreement
‘Pre-stent’
1st Observer
2nd Observer
RMS
Difference from zero (95% CI)
p
EOA (cm2)
1.4±0.5
1.2±0.4
0.30
0.17 (-0.17 → 0.52)
=0.002
Diameter (cm)
2.1±0.2
2.0±0.5
0.19
0.12 (-0.19 → 0.42)
=0.01
VTI (cm)
15.1±5.5
14.6±5.3
0.56
0.43 (-0.31 → 1.17)
<0.001
‘In-stent, pre-valve’
1st Observer
2nd Observer
RMS
Difference from zero (95% CI)
p
EOA (cm2)
2.3±0.4
1.9±0.5
0.53
0.34 (-0.52 → 1.21)
= 0.01
Diameter (cm)
2.08±0.06
1.82±0.11
0.27
0.25 (0.02 → 0.49)
<0.001
VTI (cm)
25.1±6.4
26.8±6.1
0.42
-1.74 (-9.50 → 6.02)
=0.11
EOA: effective orifice area; RMS: root mean square; VTI: velocity time integral
stent´ values for EOA are consistently lower. Inconsistencies
between ‘pre-stent’ and ‘in-stent, pre-valve’ SV measurements
could not be attributed to the presence of paravalvular AR, as
this was mild in every case and post-TAVI AR severity was not
a patient selection criterion. Values of SV calculated using LV
volumes were similar to those obtained from ‘pre-stent’ subvalvular calculation, and significantly lower compared to those
obtained from ‘in-stent, pre-valve’ sub-valvular calculation,
suggesting ‘pre-stent’ values more likely reflect true subvalvular SV. Thus ‘pre-stent’ measures should be used when
estimating EOA after CoreValve TAVI.
Error in Calculating Sub-Valvular Volume
Our study highlights difficulties in EOA calculation after TAVI
depending on measurement site of sub-valvular SV. For
example, ‘pre-stent’ SV (53±19ml) was significantly lower than
‘in-stent, pre-valve’ SV (62±20ml, p<0.001), despite no loss of
mass between the two sites. A possible explanation may be the
assumption that the LVOT annulus is circular. At the level of the
neo-aortic valve, the device is circular, but within the LVOT, the
device takes on the LVOT shape and is likely to be elliptical.
Using a circular formula for LVOT would underestimate EOA
‘pre-stent’ and be more accurate ‘in-valve, pre-stent’. Flow
acceleration ‘in-stent, pre-valve’ could also account for higher
VTI and thus higher SV.
To investigate potential source(s) of error in obtaining in
sub-valvular SV calculation, we assessed between-observer
agreement for SV and EOA at both ‘pre-stent’ and ‘in-stent, prevalve’ levels. 95% limits of agreement for EOA were wider ‘instent, pre-valve’ compared with ‘pre-stent’, and the largest error
was VTI measurement ‘in-stent, pre-valve’. This is unsurprising,
since placing the sample volume “approximately 1cm below
the neo-aortic valve” to obtain VTI is clearly less exact than
placing it “just underneath the apical margin of the valve stent”.
Moreover, flow is not laminar ‘in-stent, pre-valve’, and optimal
velocity signals may originate much closer to the valve leaflets.
Our results suggest that variability in VTI may be the primary
reason for poor agreement between the ‘in-stent, pre-valve’
while diameter measurements accounts for bias measurement
between the two observers.
Clinical Significance
Our study underlines the difficulties of accurate calculation of
valve area, in particular influence of the site of measurement
of sub-valvular SV in estimating EOA following CoreValve
TAVI, and suggests that SV and EOA are larger when ‘in-stent,
pre-valve’ measurements were used. Since EOA is operator-
dependent, absolute cut-off values should not be relied upon
for clinical decision-making and should be considered in
combination with documentation of other measures of valve
function. Our results support recently published VARC-2
recommendations, which suggest that the patient’s own initial
post-implant study should be used as a reference for long-term
follow-up, and that EOA should not be used in isolation in serial
follow-up assessment of valve function14.
Limitations
Our study lacks a ‘gold-standard’ comparison group for
assessment of SV and EOA and thus it is difficult to confirm
which site of SV measurement (‘pre-valve’ or ‘in-stent, prevalve’) is actually correct. We could have compared our EOA
calculations with direct measurement of the internal geometric
orifice area of a similar group of fully-expanded in-vitro
valves, but this would have ignored effect of valve leaflets,
and would not have taken into account leaflet distortion at
implantation. Three-dimensional imaging may have been
helpful in measuring the cross-sectional area of the LVOT17-19. If
CoreValve implantation site were low, ‘pre-stent’ diameter and
LVOT may have been taken in the LV, but this was not the case
in any patient. Finally, we did not take into account pressure
recovery phenomenon that may occur downstream of the valve
in patients with small aortic diameters; however, the smallest
CoreValve used was the 26mm valve, so that this was unlikely
to be a major source of error.
Conclusions
Our study emphasises the importance of sample volume
positioning in calculating sub-valvular stroke volume, and thus
effective aortic valve area, in the functional assessment of a
Medtronic CoreValve transcatheter device. Since aortic valve
area measurements are operator-dependent, and variable, we
suggest that measures of EOA should not be used alone in
the assessment of CoreValve function following TAVI. Rather,
calculation of effective aortic valve area should be used in
conjunction with a description of valve morphology, flow rate,
pressure gradients, Doppler velocity index, ventricular function,
ventricular size and wall thickness, degree of valve calcification,
blood pressure, and functional status. If calculation of EOA is
required, this should be performed using ‘pre-stent’ measures,
noting in particular the variability associated with ‘in-stent,
pre-valve’ VTI measurement. However, since well-established
normal transcatheter valve gradients and EOAs based on preimplant aortic annular dimensions do not exist, the patient’s
own initial post-implant study is best placed as a reference for
serial comparisons.
March 2014 - Issue 3
|
Original articles
Correspondence to:
Dr. A.M. Duncan
The Royal Brompton Hospital
Sydney Street, London, SW3 6NP, United Kingdom.
Telephone: 44 (0) 207 351 8121
Fax: 44 (0) 207 351 8604
E-mail: [email protected]
References
1. Vahanian A, Alfieri O, Andreotti F, Antunes MJ, Barón-Esquivias G,
Baumgartner H, et al. Guidelines on the management of valvular heart
disease (version 2012). The Joint Task Force on the Management of
Valvular Heart Disease of the European Society of Cardiology (ESC) and
the European Association for Cardio-Thoracic Surgery (EACTS). Eur Heart
J. 2012;33(19):2451-96.
2.Grube E, Schuler G, Buellesfeld L, Gerckens U, Linke A, Wenaweser P,
et al. Percutaneous aortic valve replacement for severe aortic stenosis
in high-risk patients using the second- and current third-generation
self-expanding CoreValve prosthesis: device success and 30-day clinical
outcome. J Am Coll Cardiol 2007;50:69-76.
3. Moat NE, Ludman P, de Belder MA, Bridgewater B, Cunningham AD,
Young CP, et al. Long-term outcomes after transcatheter aortic valve
implantation in high-risk patients with severe aortic stenosis: the U.K.
transcatheter device (United Kingdom Transcatheter Aortic Valve
Implantation) Registry. J Am Coll Cardiol. 2011;58:2130-8.
4.Leon MB, Smith CR, Mack M, Miller DC, Moses JW, Svensson LG, et al.
Transcatheter aortic-valve implantation for aortic stenosis in patients who
cannot undergo surgery. N Engl J Med 2010;363:1597– 607.
5.Smith CR, Leon MB, Mack M, Miller DC, Moses JW, Svensson LG, et al.
Transcatheter versus surgical aortic valve replacement in high risk patients.
N Engl J Med 2011; 364:2187–98.
6. Rosenhek R, Iung B, Tornos P, Antunes MJ, Prendergast BD, Otto CM,
et al. ESC Working Group on Valvular Heart Disease Position Paper:
assessing the risk of interventions in patients with valvular heart disease.
Eur Heart J 2012;33: 822–8.
7. Zogbi WA, Chambers JB, Dumesnil JG, Foster E, Gottdiener JS, Grayburn
PA, et al. Recommendations for evaluation of prosthetic valves with
echocardiography and Doppler ultrasound: a report from the American
Society of Echocardiography’s Guidelines and Standards Committee and
the Task Force on Prosthetic Valves. J Am Soc Echocardiogr 2009;22:975–
1014.
8. Zamorano JL, Badano LP, Bruce C, Chan KL, Gonçalves A, Hahn RT, et
al. EAE/ASE recommendations for the use of echocardiography in new
transcatheter interventions for valvular heart disease. European Journal of
Echocardiography (2011) 12, 557–584.
9. Baumgartner H, Hung J, Bermejo J, Chambers JB, Evangelista A, Griffin
BP, et al. American Society of Echocardiography; European Association of
Echocardiography.et al Echocardiographic assessment of valve stenosis:
EAE/ASE recommendations for clinical practice. J Am Soc Echocardiogr
2009;22:1-23.
10.Pibarot P, Dumesnil JG. Doppler echocardiographic evaluation of
prosthetic valve function. Heart. 2012 Jan;98:69-78.
11.Spethmann S, Dreger H, Schattke S, Baldenhofer G, Saghabalyan D,
Stangl V, et al. Doppler haemodynamics and effective orifice areas of
Edwards SAPIEN and CoreValve transcatheter aortic valves. Eur Heart J
Cardiovasc Imaging. 2012;13(8):690-6.
12.Clavel MA, Rodés-Cabau J, Dumont É, Bagur R, Bergeron S, De
Larochellière R, et al. Validation and characterization of transcatheter aortic
valve effective orifice area measured by Doppler echocardiography. JACC
Cardiovasc Imaging. 2011 Oct;4(10):1053-62.
13.Shames S, Koczo A, Hahn R, Gillam L. Flow characteristics of the SAPIEN
aortic valve: the importance of recognizing in-stent flow acceleration
for the echocardiographic assessment of valve function. J Am Soc
Echocardiogr 2012;25(6);603-9.
14.Kappetein AP, Head SJ, Généreux P, Piazza N, van Mieghem NM,
Blackstone EH, et al. Updated standardized endpoint definitions for
transcatheter aortic valve implantation: the Valve Academic Research
Consortium-2 consensus document. J Am Coll Cardiol. 2012;60(15):143854.
15.Jilaihawi H, Chin D, Spyt T, Jeilan M, Vasa-Nicotera M, BenceJ, et al.
Prosthesis-patient mismatch after transcatheter aortic valve implantation
with the Medtronic-Corevalve bioprosthesis. European Heart Journal
2010;31:857–864.
16.Bland JM, Altman DG. Statistical methods for assessing agreement
between two methods of clinical measurement. Lancet 1986;(i):307-10.
17.Smith LA, Dworakowski R, Bhan A, Delithanasis I, Hancock J, MacCarthy
P, et al. Real-time three-dimensional transoesophageal echocardiography
adds value to transcatheter aortic valve implantation. J Am Soc
Echocardiogr 2013;26(4):359-69.
18.Gonclaves A, Almeria C, Marcos-Alberca P, Feltes G, Hernandez-Antolin
R, Rodriques E, et al. Three-dimensional echocardiography in paravalvular
139
aortic regurgitation after transcatheter aortic valve implantation. J Am Soc
Echocardiogr 2012;25(1):47-55.
19.Hahn RT, Khalique O, Williams MR, Koss E, Paradis JM, Daneault
B, et al. Predicting paravalvular regurgitation folowing transcatheter
valve replacement: utility of a novel method for three-dimensional
echocardiographic measurements of the aortic annulus. J Am Soc
Echocardiogr 2013;26(9):1043-52.