* Your assessment is very important for improving the workof artificial intelligence, which forms the content of this project
Download Spectrophotometer
Survey
Document related concepts
Transcript
Spectrophotometer, Installation Maintenance and Repair Prof. Moustafa Moustafa Mohamed Spectrophotometer • The word spectrophotometer is derived from the Latin word spectrum, which means image, and the Greek word phos or photos, which means light. • Generally, light from a lamp with special characteristics is guided through a device, which selects and separates a determined wave length and makes it pass through a sample. • The light intensity leaving the sample is captured and compared with that which passed through the sample. Transmittance, which depends on factors such as the substance concentration is calculated from this intensity ratio. PURPOSE OF THE EQUIPMENT • The spectrophotometer is used in the laboratory for determining the concentration of a substance in a solution allowing a qualitative or quantitative analysis of the sample. Conventional spectrophotometer • Upon passing of in mediums, light undergoes a series of phenomena. • Among these are featured reflection, refraction, diffraction, absorption, diffusion, polarization and other phenomena measured by various instruments and devices. The table below shows the wavelength ranges used for carrying out spectrophotometry tests. • The diagram in Figure 27 shows that the incidental radiation [Io] can undergo a series of transformations. It can be reflected [Ir], transmitted [It], diffused [Id], absorbed and directly emitted as fluorescence [If ]. The phenomena on which spectrophotometry is based are mainly absorption and transmission. In order to understand how, it is necessary to take Beer Lambert’s law into account. Beer Lambert’s Law. • it identifies the relationship between the concentration of the sample and the intensity of light transmitted through it. • The transmittance [T] is the fraction of the incidental light of determined wavelength passing through the sample. Where: • It = intensity of the transmitted radiation • Io = intensity of the incidental radiation • The percentage of transmittance [%T] can be expressed by the following equation: • The concentration of light absorbing molecules in a sample is proportional to the absorbance [A] of that sample. It is expressed mathematically as: • Where: – A = Absorbance measured – ε = Molecule absorbance coefficient [litres/moles/cm] – l = Distance of the trajectory traversed (path length) by the light in the sample – c = Sample concentration [moles/litres] • Absorbance [A] is related to transmittance [T] through thefollowing equation: Absorbance phenomenon The graphs presented next demonstrate how absorbance [A]and transmittance [T] vary as a function of the concentration [C] according to Beer Lambert’s law. Transmittance graph Absorbance graph • In conclusion it can be inferred that by increasing the concentration of a substance, the transmittance is decreased and, upon increasing the concentration of the substance, absorbance is increased. • The linearity of Beer Lambert’s law is affected if the following conditions occur. – Displacement of the sample’s chemical balance as a function of the concentration. – Deviations in the absorbance coefficients, greater concentrations than 0.01 M due to electrostatic interaction between nearby molecules. – Changes in the refraction index at high concentrations of the analyte. – Diffusion of light due to particles in the sample. – Fluorescence or phosphorescence of the sample. – Non-monochromatic radiation. SPECTROPHOTOMETER COMPONENTS • The most important components of a spectrophotometer are the following. – – – – – The light source The monochromator The sample carrier The detector system The reading system Light source • Depending on the type of spectrophotometer, the light source can be – a tungsten lamp for visible light or – a deuterium arc lamp for ultraviolet light. – Some manufacturers have designed spectrophotometers with long lasting xenon intermittent lamps emitting light in the visible and ultraviolet ranges. • The lamp(s) come factory-assembled on a base that ensures a fixed position, to maintain optical adjustment and focus when operating or when replacing the bulb. • The typical radiating energy emitted from a tungsten lamp is between 2600 and 3000°K (Kelvin degrees). Monochromator • The monochomator is a set of elements used to disperse white light into waves of different wavelengths, one of which is used in the sample reading. • In general, it has an entry crevice or groove which limits the light radiation produced by the source and confines it to a determined area; • a set of mirrors for transmitting light through the optic system; an element for separating the light radiation wavelengths (which may be a prism or a diffraction (or transmission) grating); and an exit opening for selecting the wavelength required to illuminate the sample. Diffraction gratings have the advantage of eliminating the non-linear dispersion and being insensitive to changes in temperature. Detector system • The detection system can be designed with photocells, phototubes, photodiodes or photomultipliers. This depends on the ranges of wavelength, the sensitivity and the required speed of response. The detection system receives light from the sample and converts it into an electrical signal proportional to the energy received. This electrical signal can be processed and amplifi ed to be interpreted by the reading system. A summary of advantages and disadvantages of devices normally used in detection systems is included in the following table (see opposite). Reading system • The signal which leaves the detector goes through various transformations. It is amplified and transformed until its intensity becomes a proportional transmittance/absorbance percentage. There are analogous reading systems (displaying results on a reading scale) or digital ones (showing results on a screen). • Analogous indicators traditionally bear the name meters. Their exactitude depends among other factors, on the length and the number of divisions of the scale (the more divisions, the more exact it is). Their main disadvantage is that they can be incorrectly read, due to the operators’ fatigue or errors identifying scales when there are several. • Digital indicators usually show results on a screen as illuminated alpha numerals. This makes reading errors less likely. Advantages and disadvantages of common detection devices INSTALLATION REQUIREMENTS • For the correct functioning of a spectrophotometer, the following is required: – An electric supply source that complies with the norms and standards used in the country. In American countries, voltages of 110 V and frequencies of 60 Hz are generally used. Other parts of the World require 220-230V/50-60 Hz. – A clean, dust free, environment. – A stable work table away from equipment that generate vibrations (centrifuges, agitators). SPECTROPHOTOMETER MAINTENANCE • Routine maintenance required vary in complexity, ranging from careful cleaning of components to specialized procedures carried out by a trained specialized technician or engineer with the technical information for different manufacturers’ models and designs. • Following manufacturer’s instructions and careful use will guarantee a prolonged operational life. In recent models, manufacturers have incorporated automatic routines of calibration and verification. • In this document general maintenance recommendations applicable to a wide range of spectrophotometers are presented. It is emphasized that specialized routines can only be performed according to the specific manufacturer’s recommendations for each particular model. General routine maintenance for a spectrophotometer in good condition and the frequency of estimated checks are as follows: Inspection of the instrument’s surroundings • Frequency: Annually – The area in which the spectrophotometer is installed must be inspected visually and tested electrically in order to guarantee the safety of the operator. – The inspection covers the electrical installation and the installation area (physical infrastructure related to the spectrophotometer). Electrical installation • It must be verified and tested for ensuring the following: – There is an electrical outlet or receptacle with a ground pole. – The receptacle is in good condition and is no further than 1.5 m from the spectrophotometer. – The voltage is of an appropriate level and must not vary by more than 5% of the voltage specified on the equipment’s plate. – The receptacle’s polarity is correct. • These tests must be done by an electrical technician or an engineer and results must be recorded to allow follow-up over time. Installation area 1. Check that there is free space around the spectrophotometer for two purposes. - First, for the connecting cables to pass without hindrances and for other components or support equipment (e.g. the voltage stabilizer). Second, to allow adequate ventilation of the equipment when it is in operation. 2. Test the integrity of the counter, its state and cleanliness. 3. Verify that there is no equipment installed that can transmit vibrations in proximity. (E.g. centrifuges). 4. Verify that it is not affected by excessively humid conditions, dust or high temperatures. 5. Avoid installing the equipment where it receives direct solar radiation. 6. Do not install the equipment where there are magnetic fields or intense electromagnetic radiation. 7. Ensure installation area is free from the influence of gases and corrosive substances. The most important aspects are: 1. Check that the work table supporting the spectrophotometer is in good condition. 2. Test the spectrophotometer. Verify that buttons or control switches and mechanical closures are mounted firmly and that their identification labels are clear. 3. Ensure that accessories are clean, not showing cracks and that their functional state is optimal. 4. Confirm that mechanical adjustment parts (nuts, screws, bolts, etc.) are adjusted and are in good condition. 5. Check that electrical connectors do not have cracks or ruptures, that they are joined correctly to the line. 6. Verify that cables are not showing signs of splicing, that they are not frayed and that they do not have worn-out insulation. 7. Check that cables securing devices and terminals are free of dust, filth or corrosion. These same cables must not be worn out or show signs of deterioration. 8. Check that the grounding system (internal and external) is standardized, of approved type, functional and correctly installed. 9. Ensure that circuit switches or interrupters, the fuse box and indicators are free from dust, filth and corrosion. 10. Check the external electrical components for signs of overheating. General maintenance Cleaning of spills • In case of a leak in the sample holder or carrier, the spill must be cleaned according to the following procedure: 1. Turn off the spectrophotometer and disconnect the cable from the electrical feed. 2. Use a syringe for cleaning the sample holder. Absorb as much liquid that can possibly be extracted. 3. Dry the sample holder with a medicinal cotton bud. 4. Use lens paper or a clean piece of soft textured cloth for cleaning the window of the photocell. 5. Clean the exterior of the instrument with a piece of cloth moistened with distilled water. Cleaning of quartz cuvettes • To maintain quartz cuvettes in good condition: – Wash the cuvettes using a diluted alkaline solution such as NaOH 0.1 M and a diluted acid such as HCl, 0.1 M. – Rinse cuvettes several times with distilled water. – Conduct rigorous and careful cleaning procedures on cuvettes if samples used can deposit films. – Some manufacturers recommend using special detergents for cleaning cuvettes. Battery changes Of low battery indication appears on the instrument’s screen. • Turn off the spectrophotometer. • Disconnect the electrical feed cable. • Open the battery compartment and remove the worn out batteries. • Clean the electrical contact points. • Install new batteries with the same specifications as the originals. • Close the compartment. • Reconnect the equipment. • Adjust the date and time information. Change of bulb/lamp • Common lamp change steps are as follows: – Verify that the bulb is not functioning In modern equipment, a sign will appear on the screen or an error code. In old equipment, the light will simply no longer work. – Turn off the spectrophotometer. – Disconnect the feed cable. – Undo the screws securing the top of the lamp’s compartment. – Undo the screws keeping the lamp’s mechanism fixed. – Undo the screws fastening the electrical connection cable to the lamp – Install a new lamp with the same characteristics as the original. Use gloves to avoid getting fingerprints on the surface of the lamp. – Reconnect the electrical feed cables to the lamp. – Reinstall the screws keeping the lamp in place. – Replace the screws securing the lamp’s compartment’s cover. – Reconnect the spectrophotometer. – Turn the equipment ON and carry out the equipment’s recalibration procedure stipulated by the manufacturer. Preventive Maintenance – Clean the spectrophotometer externally using a piece of fine cloth dampened with distilled water. – Inspect and clean the electrical feed cable. – Verify that the lamp is clean and in good state. – Check the protection fuse. – Put the instrument in the operational configuration. – Activate the “on” switch and allow it to warm up for five (5) minutes. Verify that: • The lights or pilot indicators work. • The reading indicators stay on zero (0). • The light source works. – Carry out an escaping current test in the “on” and “off ” position. – Calibrate the front panel of the spectrophotometer according to the manufacturer’s instructions. – Measure the equipment’s sensitivity. – Conduct a test according to Beer’s law. – Return the spectrophotometer to the initial configuration . GOOD PRACTICES WHEN USING THE SPECTROPHOTOMETER 1. Calibrate the spectrophotometer every time a set of samples is to be analyzed. 2. Keep the cover of the sample holder and compartment closed during the measurement. 3. Avoid reusing disposable cuvettes. 4. Only use quartz cuvettes for carrying out analysis under 310 nm. 5. Avoid the use of plastic cuvettes if using organic solvents. 6. Use high quality boron silicate glassware for preparing standards. 7. Avoid the use of sodium glass (sodium oxide) 7. Carefully clean the glass cuvettes after use. 8. Use high quality reagents. The diluents used (water or solvents) must be free of impurities. 9. Verify that samples or standards did not degas inside the cuvettes. 10. Take into account that not all substances comply with Beer’s law. Carry out linearity tests on the range of concentrations to be used. It is recommended to prepare a group of known high standard solutions and verify the results.