Download Use of Light-Emitting Diode Fluorescence Microscopy to Detect Acid

Survey
yes no Was this document useful for you?
   Thank you for your participation!

* Your assessment is very important for improving the workof artificial intelligence, which forms the content of this project

Document related concepts

Surround optical-fiber immunoassay wikipedia , lookup

History of tuberculosis wikipedia , lookup

Tuberculosis wikipedia , lookup

Mycobacterium tuberculosis wikipedia , lookup

Transcript
MAJOR ARTICLE
Use of Light-Emitting Diode Fluorescence
Microscopy to Detect Acid-Fast Bacilli in Sputum
Ben J. Marais,1,2,3 Wendy Brittle,1 Katrien Painczyk,1 Anneke C. Hesseling,1 Nulda Beyers,1 Elizabeth Wasserman,4
Dick van Soolingen,6 and Rob M. Warren5
1
Desmond Tutu TB Centre, 2Department of Paediatrics and Child Health, 3Ukwanda Centre for Rural Health, 4Division of Medical Microbiology,
National Health Laboratory Service, and 5Department of Science and Technology, National Research Foundation Centre of Excellence in
Biomedical Tuberculosis Research, Medical Research Council Centre for Molecular and Cellular Biology, Stellenbosch University, Tygerberg,
Cape Town, South Africa; and 6National Institute of Public Health and the Environment, Bilthoven, The Netherlands
Background. Fluorescence microscopy offers well-described benefits, compared with conventional light microscopy, for the evaluation of sputum smear samples for tuberculosis. However, its use in resource-limited settings
has been limited by the high cost of the excitatory light source. We evaluated the diagnostic performance of
fluorescence microscopy, using novel light-emitting diode (LED) technology as an alternative to the conventional
mercury vapor lamp (MVP).
Methods. Routinely collected sputum specimens from persons suspected to have tuberculosis who attended
community clinics were stained with auramine O and were evaluated using 2 different excitatory light sources
(MVP and LED); these specimens were then Ziehl-Neelsen stained and reexamined using light microscopy. Two
microscopists independently evaluated all smears. Bacterial culture provided the gold standard.
Results. Of the 221 sputum specimens evaluated, 36 (16.3%) were positive for Mycobacterium tuberculosis by
culture. Sensitivity and specificity documented for the different modalities were 84.7% and 98.9%, respectively,
for the LED assessment; 73.6% and 99.8%, respectively, for the MVP assessment; and 61.1% and 98.9%, respectively,
for light microscopy. k values for interreader variation were 0.87 for the LED assessment, 0.79 for the MVP
assessment, and 0.77 for light microscopy. The mean time to read a negative smear was 1.4 min with fluorescence
microscopy and 3.6 min with light microscopy, reflecting a time savings of 61% with fluorescence microscopy.
Conclusion. LED fluorescence microscopy provides a reliable alternative to conventional methods and has
many favorable attributes that facilitate improved, decentralized, diagnostic services.
Sputum smear microscopy is the only diagnostic test
available in most resource-limited settings for the evaluation of patients with symptoms suggestive of pulmonary tuberculosis (TB). Since the initial description
of the auramine O fluorescence microscopy technique
by Hagemann [1] in 1937, numerous reports have confirmed the superior diagnostic performance of fluorescence microscopy, compared with Ziehl-Neelsen (ZN)
staining and light microscopy [2–7]. In a systematic
review of 18 studies, Steingart et al. [7] reported that
Received 31 January 2008; accepted 24 March 2008; electronically published
4 June 2008.
Reprints or correspondence: Dr. Ben J. Marais, Dept. of Paediatrics and Child
Health, Faculty of Health Sciences, Stellenbosch University, PO Box 19063,
Tygerberg 7505, South Africa ([email protected]).
Clinical Infectious Diseases 2008; 47:203–7
2008 by the Infectious Diseases Society of America. All rights reserved.
1058-4838/2008/4702-0010$15.00
DOI: 10.1086/589248
fluorescence microscopy of auramine-stained smears
provides similar specificity and increased sensitivity
(mean improvement of 10%), compared with light microscopy of ZN-stained smears.
In addition to increased sensitivity, fluorescence microscopy also allows more-rapid screening of sputum
smear specimens. From an operational perspective, this
is highly advantageous, particularly when high numbers
of samples are screened per day, because the majority
of laboratory time is spent confirming negative smear
results. According to the International Union Against
Tuberculosis and Lung Disease technical guidelines for
sputum microscopy, at least 5 min of screening is required to correctly identify a negative smear result when
conventional light microscopy is used [8]. However,
under routine field conditions, the time spent per slide
is often far less than the minimum required. An operational study from Cameroon demonstrated a median
sputum microscopy examination time of only 2 min
LED Fluorescence Microscopy to Detect AFB • CID 2008:47 (15 July) • 203
[9]. Almost 50% of the cases detected through a thorough 10min evaluation were missed during routine investigation [9],
which demonstrates the negative impact that conventional light
microscopy may have on early case detection and diagnostic
delay. A comparative study reported that a mean time of 1 min
to examine a sputum smear with fluorescence microscopy
achieved higher sensitivity and equivalent specificity than did
conventional light microscopy with an examination time of 4
min [10]. Despite the clear operational advantages of fluorescence microcopy, conventional light microscopy remains the
most widely used diagnostic test in resource-limited settings.
The main reason that fluorescence microscopy is not used
more widely is the limited lifespan (typically 200–300 h) and
the high cost of the short-arc mercury vapor lamp (MVP),
which has traditionally been used as the excitatory light source.
Repeated on-and-off switching, as may occur with unreliable
local power supply, shortens the lifespan even further [11]. In
addition, MVPs are energy inefficient and require an extensive
power supply; they may also fail catastrophically and release
toxic mercury into the environment [11]. Fluorescence microscopes provided by donor agencies often fall into disuse because
of high maintenance costs [12]. Light-emitting diode (LED)
technology provides a cheap and reliable light source with a
usable lifespan of 150,000 h; repeated on-and-off switching
does not reduce its usable lifespan, and it does not pose a
potential toxicity risk [11]. Initial studies indicated that LED
fluorescence microscopy, with use of a royal blue LED light,
offers a valid alternative to the MVP [11, 12], but data regarding
its diagnostic use remain limited. Interreader variability and
implementation considerations have not been evaluated under
routine conditions.
Our study evaluated the diagnostic performance of auramine
O fluorescence microscopy with use of an LED excitatory light
source, compared with the traditional MVP. In addition, both
fluorescence modalities were compared with conventional light
microscopy of ZN-stained smears. Mycobacterial culture provided the gold standard.
affinity for the mycolic acid contained in the cell walls of mycobacteria. With auramine O staining, mycobacteria appear as
bright yellow fluorescent rods when viewed under an excitatory
light source. Auramine O is excited by blue light (wavelength,
450–480 nm) and emits in the green-yellow range (wavelength,
500–600 nm). The royal blue LED light source used (350 mÅ,
1 W, Royal Blue Luxeon emitter; Philips Lumileds Lighting
Company) has an absorption maximum of 450 nm, which
matches the absorption maximum of auramine O.
Figure 1 provides an overview of the specimen collection and
evaluation procedures. Research slides were prepared in 2
batches of 10–20 specimens per week. Slides were stained with
auramine O and were read by 2 independent microscopists;
the slides were first read with the MVP, and then read with the
LED light source. An eye piece with ⫻10 magnification and an
objective with ⫻40 magnification (total magnification, ⫻400)
were used. The microscopists were blinded to all previous results. Slides were randomly reassorted after each evaluation,
and separate data capture forms were used for each modality
to eliminate the possibility of the first reading influencing the
second. After completion of fluorescence microscopy, the same
slide was ZN stained using standard methodology [8], and the
stained slide was evaluated using a conventional light microscope. As recommended by the International Union Against
Tuberculosis and Lung Disease technical guide, 100 fields were
covered with the battlement method, with use of an eye piece
METHODS
We conducted a cross-sectional, laboratory-based study. Consecutive study specimens were collected from adult patients
suspected to have pulmonary TB who routinely attended 2
primary health care clinics in Cape Town, South Africa. The
overall TB incidence recorded at the study setting and its immediate surroundings was 845 cases per 100,000 population in
2004 [13].
Smears were prepared from routine sputum specimens collected during a 3-month study period (April–June 2007). Standard procedures were performed for decontamination and concentration of specimens, preparation of slides, and
fluorochrome staining with auramine O [14], which has an
204 • CID 2008:47 (15 July) • Marais et al.
Figure 1. Overview of specimen collection and evaluation procedures.
An asterisk (*) indicates that the result of the routine diagnostic smear
was positive, and the remaining sample was inoculated into 7H9 liquid
broth medium with use of the Mycobacterial Growth Indicator Tube (MGIT)
system. A pound sign (#) indicates that the result of the routine diagnostic
smear was negative, and the remaining sample was inoculated onto
Löewenstein-Jensen (LJ) solid medium. TB, tuberculosis.
with ⫻10 magnification and an oil immersion objective with
⫻50 magnification (total magnification, ⫻500) [8]. Performing
a ZN stain after initial auramine O staining should not adversely
affect the outcome and is accepted as standard practice (S. H.
Siddiqi, personal communication). All evaluations were performed within a 2-day period.
The number of acid-fast bacilli observed was quantified according to Centers for Disease Control and Prevention guidelines (table 1). No adjustment was made for the slight difference
in magnification between fluorescence (⫻400) and light (⫻500)
microscopy. Specimens were routinely cultured for research
purposes; because of the cost considerations, specimens for
which the result of the routine diagnostic smear was found to
be positive were cultured in liquid (7H9 broth) medium with
use of the automated Mycobacterial Growth Indicator Tube
system (BD Diagnostic Systems), and smear specimens for
which the results of the routine diagnostic smear were negative
were cultured on solid Löewenstein-Jensen medium.
Anonymous, unlinked data were entered into an Excel
spreadsheet (Microsoft). Descriptive and comparative data
analyses were performed using Stata, version 9.2 (StataCorp);
95% CIs for the k statistics were calculated using the bootstrap
method. Sensitivity, specificity, and interreader agreement were
compared among all 3 modalities, with culture as the gold
standard. Ethics approval was granted by the Committee for
Human Research, Stellenbosch University (protocol N07/06/
136).
Table 1. Sputum smear grading, according to the number of
acid-fast bacilli (AFB) visualized, for light and fluorescence
microcopy.
CDC guidelinea
Study definition
Negative
Doubtful positive
Negative
Scanty
Per 100 fields
Per 10 fields
1+
2+
Positive
Positive
Per single field
3+
4+
Positive
Positive
No. of AFB seen
0
1–2 per whole smear
1–9
19 per single field
NOTE. CDC, Centers for Disease Control and Prevention.
a
From Kent and Kubica [15].
copists read each slide only once; therefore, intrareader variability could not be calculated. With fluorescence microscopy,
the mean time to read a negative smear result was 1.4 min; no
difference was noted between the LED and the MVP. With
conventional light microscopy, the mean time spent to read a
negative smear result was 3.6 min, demonstrating time savings
of 61% with fluorescence microscopy.
DISCUSSION
RESULTS
Of the 221 sputum specimens evaluated, 36 (16%) were confirmed by culture to be positive for Mycobacterium tuberculosis.
Table 2 shows the yield of sputum smear–positive specimens
identified with each of the diagnostic modalities evaluated. The
yield achieved with fluorescence microscopy (mean yield for
the LED and MVP combined) exceeded the yield achieved with
light microscopy (30 [14%] of 221 specimens vs. 24 [11%] of
221 specimens; OR, 1.29; 95% CI, 0.70–2.37), but the difference
was not statistically significant. The yields achieved with fluorescent microscopy using different excitatory light sources
(MVP and LED) were comparable (33 [15%] of 221 specimens
vs. 27 [12%] of 221 specimens; OR, 1.26; 95% CI, 0.71–2.26).
The sensitivity and specificity achieved with the various modalities were 84.7% and 98.9%, respectively, for the LED assessment; 73.6% and 99.8%, respectively, for the MVP assessment; and 61.1% and 98.9%, respectively, for light microscopy
of ZN-stained smears.
Table 3 shows the number of positive smear results identified
by each microscopist, in addition to discrepant results and k
values for interreader variability. Interreader agreement was
highest with the use of LED fluorescence microscopy (k, 0.87),
although the 95% CIs of all 3 modalities overlapped. Micros-
No statistically significant differences were noted, but our findings support previous studies that demonstrated the superior
diagnostic performance of fluorescence microscopy, compared
with conventional light microscopy [2–7]. The excellent sensitivity and specificity achieved and the short evaluation time
required for fluorescence microscopy indicate the potential
benefit of making the modality more widely available, especially
in resource-limited settings with a high burden of TB. The time
saving achieved with fluorescence microscopy did not result
from a reduction in the number of fields screened (100 fields
were screened with both modalities) but can be ascribed to
quicker scanning of each field because of increased visibility of
the mycobacteria; the decreased magnification used during
fluorescence microscopy, compared with light microscopy
(⫻400 vs. ⫻500), may also have contributed, particularly toward the sensitivity differences noted. Additional advantages of
fluorescence microscopy include the simplicity of the fluorochrome staining method, compared with ZN methods [7].
Most importantly, our study confirmed previous observations that the LED provides a reliable alternative light source
for fluorescence microscopy [11, 12]; in fact, although differences were not statistically significant, the highest sensitivity
and lowest interreader variability were achieved with LED fluorescence microscopy. This is an important finding, because the
availability of a robust and cheap excitatory light source makes
fluorescence microscopy a more feasible option in resourcelimited settings. In addition, the LED light source is highly
energy efficient, compared with the traditional MVP [12 , 16].
LED Fluorescence Microscopy to Detect AFB • CID 2008:47 (15 July) • 205
Table 2. Mean yield of sputum smear–positive specimens achieved with the use of 3
different microscopic modalities for 221 adults suspected to have pulmonary tuberculosis.
Fluorescence microscopy of
auramine O–stained specimens,
no. (%) of specimens
Mercury vapor lamp
Light-emitting diode
Light microscopy of
Ziehl-Neelsen–stained specimens,
no. (%) of specimens
Negative
Scanty
187 (85)
7 (3)
183 (83)
5 (2)
191 (86)
6 (3)
Positive
27 (12)
33 (15)
24 (11)
Result
NOTE. Data represent the mean yield for the 2 microscopists combined.
The low energy requirements of LED fluorescence microscopy
provides prospect for the development of battery-operated fluorescence microscopes that can be used in areas with unreliable
or absent power supply, as is frequently the case in remote and/
or resource-limited settings.
Although the idea was not formally evaluated during the
study, microscopists remarked that, with the LED light source,
mycobacteria remained easily visible despite the absence of a
darkened environment. Microscopists are usually advised to
perform fluorescence microscopy in a darkened environment,
which is a major constraint in resource-limited settings where
a darkened location may not be readily available. The observation that a darkened environment is not necessary requires
further evaluation, but if substantiated, it would greatly enhance
the practical feasibility of using LED fluorescence microscopy
to provide decentralized diagnostic services. Such services have
been demonstrated to improve case detection and reduce diagnostic delay, especially in rural and remote areas [17].
A previous study commented that more “scanty” smears were
observed with the MVP, compared with the LED; this was
potentially attributed to photobleaching of the auramine O
stain after MVP exposure [11]. It is unlikely that photobleaching influenced our results; batches of auramine O–stained
smears were read on the same day, and there were no differences
between the numbers of scanty smears detected with the 2
modalities (7 scanty smears with the MVP and 5 scanty smears
with the LED). In fact, an increased proportion of positive
smear specimens were detected with the LED assessment, compared with the MVP assessment (33 [15%] specimens vs. 27
[12%] specimens), and the LED assessment was performed after
MVP exposure. The improved yield was not statistically significant, but we can conclude that LED fluorescence microscopy
is no less sensitive than microscopy with the traditional MVP.
Sputum culture is widely regarded to be the most sensitive
test (i.e., the gold standard) for the detection of pulmonary
TB, but its routine use in resource-limited settings is hampered
by excessive cost, slow turnaround times, and the need for
adequate laboratory infrastructure. In practice, improvements
in direct sputum sample evaluation that result from improved
sensitivity and/or improved access to decentralized diagnostic
services remain highly relevant. There is a definite need for
improved access to sputum culture to establish drug susceptibility in areas with high rates of drug-resistant TB and in
HIV-infected patients who have negative sputum smear results.
However, the need for rapid smear results and effective treatment of the most infectious TB cases remains paramount; in
addition, fluorescence microscopy seems to be more efficient
than light microscopy for detection of TB in HIV-infected patients with paucibacillary disease [2].
Our study was limited by small numbers and failed to demonstrate statistically significant differences. However, we believe
that the study results are sufficient to demonstrate that LED
fluorescence microscopy offers a highly feasible alternative—
with at least similar diagnostic performance—to conventional
Table 3. Diagnostic yield and interreader variation for 2 microscopists evaluating 3 different sputum microscopic modalities for the identification of acid-fast bacilli in sputum
samples from 221 adults suspected to have pulmonary tuberculosis.
No. of sputum smear specimens determined to be positive
for Mycobacterium tuberculosis
Diagnostic modality
Microscopist 1
Microscopist 2
No. of
discrepant
results
k value (95%
CI)
Light microscopy
23
25
10
0.77 (0.62–0.89)
Fluorescence microscopy
Mercury vapor lamp
25
29
10
0.79 (0.64–0.90)
Light-emitting diode
33
32
7
0.87 (0.76–0.96)
206 • CID 2008:47 (15 July) • Marais et al.
methods. Although the inconsistent use of culture media (because of financial constraints) in our study may be criticized,
this would not have introduced bias, because a single culture
medium was used for each specimen. Therefore, all 3 modalities
were consistently evaluated against the identical gold standard
test. Performance of the ZN stain on the same slide that was
initially stained with auramine O could have affected the sensitivity of the ZN stain, but this is considered to be unlikely
and is accepted as standard practice. Despite this potential reservation, the decision was made to evaluate the various modalities with use of the very same slide to ensure optimal comparison, because different slides prepared from the same
specimen may be highly variable.
In conclusion, our study adds to the body of evidence that
demonstrates that LED fluorescence microscopy provides a
valid alternative to conventional methods. Its low cost, sturdiness, energy efficiency, and perceived efficacy in the absence
of a darkened environment are highly favorable attributes. In
the short term, optimized microscopy offers the most realistic
option for improved case finding in resource-limited settings;
large-scale field trials are required to assess the advantages and
feasibility of replacing conventional ZN light microscopy with
LED fluorescence microscopy as a first-line diagnostic test.
Acknowledgments
We thank the primary health care clinics involved, for assisting with
sputum collection, and Dr. Ivan Toms (City of Cape Town Health Department and the National Health Laboratory Service), for allowing us to
conduct the study.
Potential conflicts of interest. All authors: no conflicts.
References
1. Hagemann PKH. Floureszenzmikroskopische untersuchungen uber virus und andere microben. Zentralbl Bakteriol 1937; 140:184.
2. Prasanthi K, Kumari AR. Efficacy of fluorochrome stain in the diagnosis
of pulmonary tuberculosis co-infected with HIV. Indian J Med Microbiol 2005; 23:179–85.
3. Kivihya-Ndugga LE, van Claaff MR, Githui WA, et al. A comprehensive
comparison of Ziehl-Neelsen and fluorescence microscopy for the diagnosis of tuberculosis in a resource poor urban setting. Int J Tuberc
Lung Dis 2003; 7:1163–71.
4. Singh NP, Parija SC. The value of fluorescence microscopy of auramine
stained sputum smears for the diagnosis of pulmonary tuberculosis.
Southeast Asian J Trop Med Public Health 1998; 29:860–3.
5. Githui W, Kitui F, Juma ES, et al. A comparative study on the reliability
of fluorescence microscopy and Ziehl-Neelsen method in the diagnosis
of pulmonary tuberculosis. East Afr Med J 1993; 70:263–6.
6. Hänscheid T, Ribeiro CM, Shapiro HM, Perlmutter NG. Fluorescence
microscopy for tuberculosis diagnosis. Lancet Infect Dis 2007; 7:236–7.
7. Steingart KR, Henry M, Ng V, et al. Fluorescence versus conventional
sputum smear microscopy for tuberculosis: a systematic review. Lancet
Infect Dis 2006; 6:570–81.
8. International Union Against Tuberculosis and Lung Disease (IUATLD)
technical guide: sputum examination for tuberculosis by direct microscopy in low-income countries. 5th ed. Paris: IUATLD, 2000.
9. Cambanis A, Ramsay A, Wirkom V, Tata E, Cuevas LE. Investing time
in microscopy: an opportunity to optimise smear-based case detection
of tuberculosis. Int J Tuberc Lung Dis 2007; 11:40–5.
10. Bennedsen J, Larson SO. Examination for tubercle bacilli by fluorescence microscopy. Scand J Respir Dis 1966; 47:114–20.
11. Anthony RM, Kolk AH, Kuijper S, Klatser PR. Light emitting diodes
for auramine O fluorescence microscopic screening of Mycobacterium
tuberculosis. Int J Tuberc Lung Dis 2006; 10:1060–2.
12. Hung NV, Sy DH, Anthony RM, Cobelens FGJ, van Soolingen D.
Fluorescence microscopy for tuberculosis diagnosis. Lancet Infect Dis
2007; 7:238–9.
13. Marais BJ, Hesseling AC, Gie RP, Schaaf HS, Beyers N. The burden
of childhood tuberculosis and the accuracy of routine surveillance data
in a high-burden setting. Int J Tuberc Lung Dis 2006; 10:259–63.
14. Siddiqi SH. MGIT procedure manual prepared for the FIND MGIT
demonstration project. Sparks, MD: BD Diagnostic Systems, 2005.
15. Kent PT, Kubica GP. Public health microbiology: a guide for the level
III laboratory. Atlanta, GA: US Department of Health and Human
Services, US Government Printing Office, 1985.
16. Toman K. What are the advantages and disadvantages of fluoresence
microscopy? In: Frieden T, ed. Toman’s tuberculosis: case detection,
treatment and monitoring—questions and answers. 2nd ed. Geneva,
Switzerland: World Health Organization, 2004.
17. El-Sony AI, Mustafa SA, Khamis AH, et al. The effect of decentralization on tuberculosis services in three states of Sudan. Int J Tuberc
Lung Dis 2003; 7:445–50.
LED Fluorescence Microscopy to Detect AFB • CID 2008:47 (15 July) • 207