Download Power Handling Capacity of the 100 Series Probes

Survey
yes no Was this document useful for you?
   Thank you for your participation!

* Your assessment is very important for improving the workof artificial intelligence, which forms the content of this project

Document related concepts

Decibel wikipedia , lookup

Current source wikipedia , lookup

Skin effect wikipedia , lookup

Opto-isolator wikipedia , lookup

Electronic engineering wikipedia , lookup

History of electric power transmission wikipedia , lookup

Resistive opto-isolator wikipedia , lookup

Wireless power transfer wikipedia , lookup

Three-phase electric power wikipedia , lookup

Spectral density wikipedia , lookup

Pulse-width modulation wikipedia , lookup

Electromagnetic compatibility wikipedia , lookup

Electrification wikipedia , lookup

Electric power system wikipedia , lookup

Utility frequency wikipedia , lookup

Switched-mode power supply wikipedia , lookup

Audio power wikipedia , lookup

Earthing system wikipedia , lookup

Power electronics wikipedia , lookup

Buck converter wikipedia , lookup

Power engineering wikipedia , lookup

Mathematics of radio engineering wikipedia , lookup

Islanding wikipedia , lookup

Heterodyne wikipedia , lookup

Ohm's law wikipedia , lookup

Mains electricity wikipedia , lookup

Metadyne wikipedia , lookup

Rectiverter wikipedia , lookup

Zobel network wikipedia , lookup

Nominal impedance wikipedia , lookup

Alternating current wikipedia , lookup

Transcript
Power Handling Capacity of the 100 Series
Probes
Introduction
EMC probes like the Beehive Electronics 100 series probes (figure 1) can be used in two
ways. They can be used as receiving antennas, to measure an electromagnetic field.
Alternatively, when driven by a signal generator, they can be used as transmitters. When
used this way, they can be used to test the susceptibility of circuits to electromagnetic
fields.
Figure 1
Probe Current Handling Capacity
The power handling capacity of the probes is limited by the current-handling capacity of
its traces and components. At low frequencies, this is determined by the probe’s DC
resistance. As frequency increases, dielectric losses and conductor skin depth reduce the
current handling capacity of the probes. The maximum current for the probes is shown in
figure 2.
1 Beehive Electronics · Tel: (707) 824-9206 · [email protected]
8555 Lawrence Lane, Sebastopol CA 95472
rev. 1.1
Max Current, 100 Series Probes
1.2
1
Current (A, rms)
0.8
0.6
0.4
0.2
0
1.00E+03
1.00E+04
1.00E+05
1.00E+06
1.00E+07
1.00E+08
1.00E+09
1.00E+10
Freq (Hz)
Figure 2
Probe Power Handling Capacity
The impedance of the 100 series probes is a complex function of frequency. The
magnetic field probes, models 100 A, B, and C, approximate a short at DC. The 100D
electric field probe is an open circuit at DC. As the frequency increases, the impedance
of the probes varies in a complex fashion. This varying impedance is taken into account
in determining the sensitivity of the probes; and does not cause any problems in normal
operation. For determining the maximum power handling capacity of the probe,
however, the impedance is important.
For the purposes of this application note, we calculate the power handling in the most
conservative way: by assuming that the probe is a short to ground. A 50 ohm source will
deliver the most current when it is loaded by a short circuit. The graph in figure 3 shows
the power setting of a 50 ohm signal generator that will deliver the current shown in
figure 2 above.
2 Beehive Electronics · Tel: (707) 824-9206 · [email protected]
8555 Lawrence Lane, Sebastopol CA 95472
rev. 1.1
Max Power Setting for 50 Ohm Source Driving 100 Series Probes
50
45
Power (dBm)
40
35
30
25
20
1.00E+03
1.00E+04
1.00E+05
1.00E+06
1.00E+07
1.00E+08
1.00E+09
1.00E+10
Freq (Hz)
Figure 3
If a specific application requires more that the power shown in figure 3, a more accurate
calculation can be done by measuring the complex impedance of the probe at the
frequency of interest. With the current rating of the probe from figure 2 and the
measured impedance, it is possible to calculate the power setting of a signal generator
that will deliver that current into the probe’s measured impedance.
These ratings were determined by analyzing the case where the probe is used as a
transmitter, i.e. when it is driven by a signal generator through the connector. They also
apply when it is used as a receiver as well, i.e. when the probe is connected to a spectrum
analyzer.
Low Frequency Considerations for the 100D E-Field Probe
At low frequencies, the impedance of the 100D E field probe is very high. As a result, if
the currents in figure 2 were applied to the probe, very high voltages would result. For
this reason, it is necessary to specify a maximum voltage rating for the 100D as well.
This is determined by the voltage rating of the SMB connector, and is 335V rms.
Summary
The graphs above show the current- and power-handling capacity of the 100 series EMC
probes. For more information, please contact Beehive Electronics at (707) 824-9206 or
[email protected]
3 Beehive Electronics · Tel: (707) 824-9206 · [email protected]