2015 5th International Conference on Biomedical Engineering and Technology (ICBET 2015)
IPCBEE vol.81 (2015) © (2015) IACSIT Press, Singapore
DOI: 10.7763/IPCBEE. 2015. V81. 1
Measurement of Skin Surface pH with a Non-invasive Dry pH Sensor
Takuto Nagashima 1, Takashi Komeda 1, Shin-ichiroh Yamamoto 1, Tatsuhiko Yajima 2, and
Takehito Kemuriyama 3
1
2
Shibaura Institute of Technology, 307 Fukasaku, Minuma-ku, Saitama 337-8570, Japan
Saitama Institute of Technology, 1690 Fusaiji, Okabe-cho, Oosato-gun, Saitama 369-0293, Japan
3
National Defense Medical College, 3-2 Namiki, Tokorozawa-shi, Saitama 359-0042, Japan
Abstract. Recent reports suggest that skin barrier function and atopic dermatitis are related to skin surface
pH. We are developing a portable multifunctional skin measurement system for evaluating data such as skin
pH and water content. The system uses a non-invasive method for measuring the skin’s surface pH and does
not require addition of water. Previously, we reported a flat, dry pH sensor based on the Nernstian response
that was fabricated by ion plating. When the sensor was used on the skin, the voltage was constantly elevated
because of water evaporation from the skin. Herein, we measured the skin surface pH with the dry pH sensor
and a glass electrode to compare non-invasive measurements and measurements with added water. The
sensor was tested by using a standard pH solution and a correlation between pH and voltage was found. All
experiments were carried out on the same region of the left forearm of healthy male subjects at constant room
temperature and humidity (23 °C, 30% RH). Skin surface pH was measured for 2 min with the dry electrode
and with a glass electrode, and then the skin water content was measured. The increase in voltage caused by
water evaporation from the skin was subtracted from the dry pH sensor results. The skin surface pH was
different between the non-invasive dry sensor and the conventional sensor. The response of the pH sensor
voltage was affected by skin water evaporation and by water-soluble substances in the stratum corneum.
Keywords: Skin surface pH, flat dry-pH-Sensor, measurement, pH
1. Introduction
Human skin consists of three main types of tissues: the subcutaneous tissue, the dermis, and the
epidermis, which includes the outer part of the skin called the stratum corneum (Fig. 1). Skin surface pH is
measured on the stratum corneum, which contains various acids, such as amino acids, lactic acid, and fatty
acids, as well as numerous other compounds that form the acid mantle. The acid mantle provides a defense
mechanism against pathogenic microorganisms and plays an important role in skin barrier homeostasis [1],
[2].
Measurements of the skin surface pH and water content of the stratum corneum are useful in many noninvasive techniques used to assess atopic dermatitis and the physical properties of the skin barrier objectively
[3], [4]. In addition, common loss-of-function genetic variants of the epidermal barrier protein, filaggrin, are
a major predisposing factor for atopic dermatitis [5], suggesting that lack of filaggrin is involved in the
mechanism of atopic dermatitis. Recent research suggests that skin barrier function is related to the skin
surface pH and this plays a role in atopic dermatitis [6], [7]. Therefore, the skin surface pH may reflect the
skin’s barrier properties and it might be an important index for evaluating skin condition and health. We
intend to develop a portable multifunctional skin measurement system for evaluating data such as skin
surface pH, water content, and temperature. The system uses a non-invasive method for measuring the skin’s
surface pH and does not require addition of water. Previously, we reported a dry pH sensor consisting of a

Corresponding author. Tel.: +81 48 688 9477; fax: +81 48 688 9477.
E-mail address: [email protected].
1
platinum wire electrode and an Ag/AgCl wire electrode, based on the Nernstian response [8]. The voltage
was constantly elevated during dry pH sensor measurements. We developed another flat, dry pH sensor
based on the Nernstian response that was fabricated by ion plating. This paper describes the effect on the
skin water content of measuring the skin surface pH with the dry pH sensor. The relationship between the pH
and voltage was determined by using a standard pH buffer solution.
Skin surface pH
Amino
acids
Sweat
Lipid
Evaporation
Epidermis
Lactic
acid
Stratum
corneum
Dermis
Hypodermis
Fig. 1: Skin structure, skin surface pH and water evaporation
2. Material and Methods
2.1. Flat, Dry pH Sensor
A photograph of the sensor is shown in Fig. 2 (A). The electrodes are connected through copper tape,
silver paste, and leads. A schematic cross-sectional view of the sensor film layers is shown in Fig. 2 (B).
Base layers of titanium, platinum, and silver were prepared by ion plating. The titanium layer was used for
improving adhesion of the platinum and titanium layers to the substrate. A silver chloride layer was prepared
by anodic polarization. The reference electrode consisted of silver, silver chloride, and potassium chloride.
The platinum surface functioned as the working electrode through the Nernstian response. p-Xylene, Nafion,
and potassium chloride were prepared by chemical vacuum deposition. p-Xylene functioned as a selective
membrane for protons. The poly(4-methyl-1-pentene) membranes were prepared by vacuum casting.
(i)
(ii)
TPX
p-XY
Ti
Pt
Naf+KCl
AgCl
Ag
Substrate
(A)
(B)
2
Fig. 2: (A) Flat, dry pH sensor. Substrate dimensions: 25 mm × 1.1 mm, electrode area: 5 × 10 mm. (B) Schematic
cross-sectional view of sensor film layers. (i) Working electrode and (ii) reference electrode. TPX: poly(4-methyl-1pentene); p-XY: p-xylene; Naf: Nafion
2.2. Measurements
The measurements were performed by using a DC microvolt ammeter (PM-18U, TOA Co. Ltd., Japan)
with a personal computer equipped with an analog/digital interface and a C language measurement program.
Zero offset calibration was carried out before measurements using two standard Ag/AgCl electrodes (RE-1B,
BAS Inc., Japan) with 0.01 mol/L NaOH solution. The sensor was used to measure five pH buffer solutions
(pH 1.68, 4.01, 6.86, 9.18, 10.02) on a sampling sheet (Y011, Horiba Ltd., Japan). Sampling time was about
0.1 s and the total measurement time was 120 s for the pH voltage response. The pH voltage response was
treated as a moving average.
Figure 3 (A) shows the pH sensor voltage response as a function of time. The low and high pH solution
responses exhibit some drift. However, the intermediate pH voltage responses were sufficiently stable. Fig. 3
(B) shows the relationship between the pH voltage responses over 30–60 s and the average response is
2
plotted as a linear approximation. We focused on the average values because the intermediate pH voltage
responses were stable and the skin measurement took 120 s.
(A)
(B)
250
250
pH 1.68
200
Voltage[mV]
Voltage[mV]
150
pH 4.01
pH 6.86
100
50
0
pH 9.18
-50
pH 10.02
30
60
Time[s]
90
150
100
50
0
-50
-100
0
y = -33.3 x + 273.1
R²= 0.95
200
-100
120
0
2
4 pH
6
8
10
Fig. 3: (A) Sensor voltage response as a function of time for various pH standard buffer solutions. (B) Relationship
between pH and voltage responses (averaged over 30-60s)
All measurements were carried out on the same region of the left forearm of two normal male subjects
(A and B) at constant temperature and humidity (23 °C, 30% RH). The subjects were asked to rest for about
15 min with the forearm exposed and with an earth connection on their wrist to reduce noise.
Non-invasive skin surface pH measurements with the dry pH sensor were performed for 2 min without
added water and were repeated 5 times. The responses showed some instability; therefore, they were
analyzed with a linear approximation for 3-8 s, 30-120 s and 90-120 s, with R2 > 0.9 and standard deviation
of < 2.5 mV. One response measurement was selected from each subject.
After addition of water, skin surface pH was measured with a skin pH meter (PH900, Courage +
Khazaka Electronic GmbH, Germany). The glass probe had a flat surface membrane with a reference contact,
and it was calibrated using buffer solutions with pH values of 4.01 and 6.86. For every measurement, the
electrode was rinsed ethanol and ionized water. The average was taken of three measurements performed
five times and the maximum and minimum results were removed.
Skin water content was measured by a Corneometer (CM825, Courage + Khazaka Electronic GmbH,
Germany) in arbitrary units. Skin water evaporation was measured by keeping the probe in contact with the
skin for 120 s to monitor water content. The probe is coated with a hydrophobic material, so that water
evaporating from the skin is collected on the surface of the sensor.
3. Results and Discussion
The skin surface pH determined with added water was 4.6 for subject A and 4.7 for subject B. Fig. 4
shows the skin surface pH voltage response measured with the dry pH sensor. The response was elevated in
both subjects. Our previous results suggest that the elevation was caused by the evaporation of the skin’s
water [9]. Therefore, the elevated response was subtracted using the slope of the linear approximation for
90–120 s.
60
Voltage[mV]
A
40
20
B
0
0
30
60
Time[s]
90
120
Fig. 4: Non-invasive skin surface voltage response measurement with the dry pH sensor (subject A, B)
3
Figure 5 shows the skin surface pH measurements obtained with the dry pH sensor. The pH voltage
response was obtained using following equation (where y is the voltage [mV] and x is the pH) according to
the results in Fig. 3 (B).
y = -33.3 x + 273.1
9
pH
8
B
7
A
6
5
4
0
30
60
Time[s]
90
120
Fig. 5: Non-invasive skin surface pH measurement with the dry pH sensor (Subject A, B)
These results show that the dry pH sensor voltage response on the skin was not affected by water
evaporation alone during the measurements. The skin pH at the surface of the skin results from water-soluble
substances in the stratum corneum, including sweat, sebum, carbon dioxide [10], and natural moisturizing
factor. Both subjects had almost the same pH values for the conventional skin pH meter, which required
water addition. However, the skin surface pH measurements with the dry pH sensor were different from the
conventional measurements. This suggests that the non-invasive method could affect other factors that
determine skin pH.
4. Conclusion
A dry pH sensor for non-invasive skin pH measurements was investigated. The pH voltage response was
good for standard pH solutions. We subtracted the effect of water evaporation during the non-invasive
measurements using our sensor, and then converted the results to pH. The skin pH values obtained by the
conventional probe that requires additional water and our dry probe were different. These results indicate
that our non-invasive dry pH sensor may measure different skin condition parameters because the watersoluble substances in the stratum corneum affect pH. We are investigating the differences between our probe
and the conventional probe by using other skin measurement parameters, such as water content and sebum.
5. References
[1] Darlenski, R., Sassning, S., Tsankov, N., Fluhr, J.W. Non-invasive in vivo methods for investigation of the skin
barrier physical properties. European Journal of Pharmaceutics and Biopharmaceutics. 2009, 72 (2): 295-303.
[2] Eberlein-Konig, B., Schafer, T., Huss-Marp, J., Darsow, U., Mohrenschlager, M., Herbert, O., Abeck, D., Kramer,
U., Behrendt, H., Ring, J. Skin surface pH, stratum corneum hydration, trans-epidermal water loss and skin
roughness related to atopic eczema and skin dryness in a population of primary school children. Acta DermatoVenereologica. 2000, 80 (3): 188-191.
[3] Elias, P.M., Steinhoff, M. “Outside-to-Inside” (and Now Back to “Outside”) Pathogenic Mechanisms in Atopic
Dermatitis. J Invest Dermatol. 2008, 128 (5): 1067-1070.
[4] Hatano, Y., Man, M.-Q., Uchida, Y., Crumrine, D., Scharschmidt, T.C., Kim, E.G., Mauro, T.M., Feingold, K.R.,
Elias, P.M., Holleran, W.M., 2009. Maintenance of an Acidic Stratum Corneum Prevents Emergence of Murine
4
Atopic Dermatitis. J Invest Dermatol. 2009, 129 (7): 1824-1835.
[5] Palmer, C.N.A., Irvine, A.D., Terron-Kwiatkowski, A., Zhao, Y., Liao, H., Lee, S.P., Goudie, D.R., Sandilands, A.,
Campbell, L.E., Smith, F.J.D., O'Regan, G.M., Watson, R.M., Cecil, J.E., Bale, S.J., Compton, J.G., DiGiovanna,
J.J., Fleckman, P., Lewis-Jones, S., Arseculeratne, G., Sergeant, A., Munro, C.S., El Houate, B., McElreavey, K.,
Halkjaer, L.B., Bisgaard, H., Mukhopadhyay, S., McLean, W.H.I. Common loss-of-function variants of the
epidermal barrier protein filaggrin are a major predisposing factor for atopic dermatitis. Nat Genet. 2006, 38 (4):
441-446.
[6] Rippke, F., Schreiner, V., Doering, T., Maibach, H.I. Stratum corneum pH in atopic dermatitis: Impact on skin
barrier function and colonization with Staphylococcus aureus. American Journal of Clinical Dermatology. 2004, 5
(4): 217-223.
[7] Rippke, F., Schreiner, V., Schwanitz, H.J. The acidic milieu of the horny layer: New findings on the physiology
and pathophysiology of skin pH. American Journal of Clinical Dermatology. 2002, 3 (4): 261-272.
[8] Yajima, T., Namie, H., Takanari, H. Electrochemical Characteristics of Platinum Surface Protected by Plasma
Polymerization Membrane(1)Nernstian Response to Proton. Journal of Photopolymer Science and Technology.
2002, 15 (2): 279-282.
[9] Nagashima, T., Komeda, T., Koyama, H., Yamamoto, S., Yajima, T. Development of A Dry-pH-Sensor for Skin
Surface pH Measurement. 2012 International Conference on Life Science and Engineering. 2012, IPCBEE 45
(10): 49-53.
[10] Parra, J. L., Paye, M. EEMCO guidance for the in vivo assessment of skin surface pH. Skin Pharmacology and
Applied Skin Physiology. 2003, 16 (3): 188-202.
5