Survey							
                            
		                
		                * Your assessment is very important for improving the workof artificial intelligence, which forms the content of this project
* Your assessment is very important for improving the workof artificial intelligence, which forms the content of this project
International Atomic Energy Agency Medical exposure in radiology: Optimization of protection !! CONTINUATION OF FILE VIII.3 Equipment ! Module VIII.3 - Part 1: Design considerations for the equipment International Atomic Energy Agency Topic 4: Computed Tomography The CT equipment • The CT scanner • Hounsfield units • Difference with projection radiology • Latest generations of scanners Module VIII.3-Part 1. Design considerations equipment Continued International Atomic Energy Agency 3 CT: “Computed” and “Tomography” CT is a tomographic imaging technique, generating cross-sectional image in the axial plane CT techniques use high kV with heavy filtration A fan beam is passed through the patient Transmitted radiation is measured by array detectors CT images or “sections” are maps of µ They are derived by mathematical analysis of multiple projections (use of computer): filtered back projection Module VIII.3-Part 1. Design considerations equipment Continued International Atomic Energy Agency 4 Image and “Hounsfield” number • • • • • Pixel size depends on FOV and matrix • • Matrix size 512x512 or 1024x1024 12 bits (4096 gray levels) Voxel is: • volume element=pixel area x slice thickness Relative attenuation coefficient µ is expressed in Hounsfield units or CT numbers HUt = 1000 x (µt - µwater) / µwater By definition HUwater = 0, HUair = -1000 Attention: µ will depend on kV…so will HU Module VIII.3-Part 1. Design considerations equipment Continued International Atomic Energy Agency 5 Image display: windowing CT is a DIGITAL IMAGE Window settings determine HOW tissue attenuation values are displayed Module VIII.3-Part 1. Design considerations equipment Continued International Atomic Energy Agency 6 Window settings: optimized for mediastinum Module VIII.3-Part 1. Design considerations equipment Continued International Atomic Energy Agency 7 Window settings: optimized for lung Module VIII.3-Part 1. Design considerations equipment Continued International Atomic Energy Agency 8 CT scanner • • • • Computed Tomography (CT) was introduced into clinical practice in 1972 and revolutionized X Ray imaging by providing high quality images which reproduced transverse cross sections of the body. Tissues are therefore not superimposed on the image as they are in conventional projections The technique offered in particular improved low contrast resolution for better visualization of soft tissue, but with relatively high absorbed radiation dose Dose distribution different due to rotation Module VIII.3-Part 1. Design considerations equipment Continued International Atomic Energy Agency 9 Dose distributions in projection RL and CT Head exam Conventional: dose decreases from entrance to exit, ratio 1/100 …1/1000 Rotational scanning gives dose more concentrated at center of rotation Also dose distribution in slices, but contribution outside imaged slice Module VIII.3-Part 1. Design considerations equipment Continued International Atomic Energy Agency 10 3rd & 4th generation scanners 3rd generation: rotating fan beam and array of detectors (rotate-rotate) Module VIII.3-Part 1. Design considerations equipment Continued 4th generation: rotating source and fixed ring of detectors (rotate-stationary) International Atomic Energy Agency 11 A look inside a rotate/rotate CT Detector Array and Collimator Module VIII.3-Part 1. Design considerations equipment Continued X Ray Tube International Atomic Energy Agency 12 3rd & 4th generation scanners • Slice (section) acquisition: 1-2 seconds • Cables limit rotation to 1 revolution • Modern scanners: slip ring technology • Beam is highly filtered (-> 10mmAl HVL) • Heat loading on tube very high • Collimation • defines section thickness • reduces scatter • @ 1mm slice also additional collimation at detector Module VIII.3-Part 1. Design considerations equipment Continued International Atomic Energy Agency 13 Helical or spiral CT • Slip ring technology allows continuous rotation • The patient can then be moved continuously • through the beam, making the examination much faster Continuous data acquisition and table feed X Ray beam Direction of patient movement Module VIII.3-Part 1. Design considerations equipment Continued International Atomic Energy Agency 14 New possibilities for procedures • The new helical scanning CT units allow a range of new features, such as : • CT fluoroscopy, where the patient is stationary, but • • the tube continues to rotate multislice CT, where multiple slices can be collected simultaneously 3-dimensional CT and CT endoscopy • All these new technologies and applications: • Constant increase in number of examinations • High contribution to collective effective dose Module VIII.3-Part 1. Design considerations equipment Continued International Atomic Energy Agency 15 CT contribution to number of exams and to collective effective dose (FRG 1990-92) % freq X Ray exams FRG 1990-02 Misc 30% CT 4% % coll effect dose X-ray FRG 1990-92 Misc 16% Spine 10% Chest 18% CT 35% Urography 9% Angio/DSA 10% Extremities 20% Angio/DSA 1% GI 15% Dental 17% Chest 5% Spine 10% FRG: Federal Republic of Germany Module VIII.3-Part 1. Design considerations equipment Continued International Atomic Energy Agency 16 Factors influencing dose in CT Module VIII.3-Part 1. Design considerations equipment Continued International Atomic Energy Agency 17 Equipment related factors • Wave form of generator: ~nil, low ripple • Range of tube current settings, steps • Beam filtration • Beam shaper • Focus axis distance: • shorter geometry (60 cm instead of 80) gives increase of wCTDI, but allows less mAs • Slice collimation • For ~1mm slices: dose increase • Scan field and scan angle Module VIII.3-Part 1. Design considerations equipment Continued International Atomic Energy Agency 18 Dose and spiral CT • With identical protocol settings (Slice thickness, table feed, mAs product per scan or rotation, scan length) dose in spiral CT would be same as in sequential scanning • For interpolation purpose: ~1 additional rotation => slight increase in dose (<10%) • If pitch is taken >1: less loss of information, but dose is reduced • BUT often INCREASE: • Scanning over longer distances • Multi-phase studies, same body scanned repeatedly Module VIII.3-Part 1. Design considerations equipment Continued International Atomic Energy Agency 19 Summary on this topic CT • We learned about the principle of CT scanner, the different generations and the factors influencing this high dose examination Module VIII.3-Part 1. Design considerations equipment Continued International Atomic Energy Agency 20 International Atomic Energy Agency Topic 5: X Ray beam characteristics Beam description, factors affecting beam quality and effect on imaging process. • • • • Contents of this topic Factors affecting X Ray beam and image • • • • • • Spectra Beam quality Tube current Influence of mAs Filtration Wave-form, ripple Geometric factors • • • Inverse square law Anode angle Heel effect Unwanted radiation • • Stray radiation, scatter, leakage Fighting against scatter Enhancing contrast • • Lowering kV Contrast agents Module VIII.3-Part 1. Design considerations equipment Continued International Atomic Energy Agency 22 Factors affecting the X Ray beam •Tube current •kVp value •Ripple •Filtration Module VIII.3-Part 1. Design considerations equipment Continued International Atomic Energy Agency 23 Photon spectra of X Rays kVp ≤ charact. line en. Module VIII.3-Part 1. Design considerations equipment Continued kVp > charact. line en. International Atomic Energy Agency 24 Beam “Quality” • “Measures” penetrating power • Subjectively describes the shape of continuous • spectrum Function of: • kVp • filtration • high voltage supply characteristics • Diagnostic radiology beams are POLYCHROMATIC • ± Quantification: effective energy • With moderate filtration: Eeff ≤ 2/3 kVp • With heavy filtration: Eeff ≤ 1/2 kVp Module VIII.3-Part 1. Design considerations equipment Continued International Atomic Energy Agency 25 Changing the tube current Change of QUANTITY NO change of quality Effective kV not changed Module VIII.3-Part 1. Design considerations equipment Continued International Atomic Energy Agency 26 Changing current x time product (mAs) 70 kV- 50 mAs 70 kV- 80 mAs Module VIII.3-Part 1. Design considerations equipment Continued 70 kV- 25 mAs International Atomic Energy Agency 27 Effect of changing kVp Change in QUANTITY & Change in QUALITY - spectrum shifts to higher energy - characteristic lines appear Module VIII.3-Part 1. Design considerations equipment Continued International Atomic Energy Agency 28 Importance of correct choice of kV Right image is made with correct kV; left image with too high kV, leading to loss of contrast. Module VIII.3-Part 1. Design considerations equipment Continued International Atomic Energy Agency 29 Remember generator waveforms Single phase single pulse kV ripple (%) 100% Single phase 2-pulse 13% Three phase 6-pulse 4% Three phase 12-pulse Line voltage 0.01 s 0.02 s Module VIII.3-Part 1. Design considerations equipment Continued International Atomic Energy Agency 30 Effect of waveform or ripple • During 1/2 mains cycle kV fluctuates between • • • • min and max At any moment spectrum defined by kV The more ripple, the more low energy photons Less ripple, beam is harder We do not need these low energy photons: • Absorbed in patient or scattered • Dose to patient increases • Do not reach the image receptor • HF or 12-pulse will reduce dose to patient • And give less noisy images (less scatter) Module VIII.3-Part 1. Design considerations equipment Continued International Atomic Energy Agency 31 Ripple increases image noise The left image is made with an old (high ripple) generator. The increased noise led to loss of detail compared to modern equipment Module VIII.3-Part 1. Design considerations equipment Continued International Atomic Energy Agency 32 Filtration Change in QUANTITY & Change in QUALITY spectrum shifts to higher energy 1- Spectrum out of anode 2- After window tube housing (INHERENT filtration) 3- After ADDITIONAL filtration Module VIII.3-Part 1. Design considerations equipment Continued International Atomic Energy Agency 33 Inherent filtration •Tube housing window 1 •Inherent filtration: –Glass envelope –Cooling oil –Beryllium window (thin) •“Radioluscent” •You can see the filament: -->1 •Filtration expressed in equivalent Al thickness •Typically: 1mm Al •Housing: lead shield ! Module VIII.3-Part 1. Design considerations equipment Continued International Atomic Energy Agency 34 Additional filtration Additional filtration: •preferably removes lower energy photons • increases effective energy Eeff • lowers intensity • reduces patient dose ! •but increases tube loading Module VIII.3-Part 1. Design considerations equipment Continued International Atomic Energy Agency 35 K-edge filtration Relative intensity Metals with K-edges in diagnostic range Ex: Erbium (Mo, Rh,Sn) Max absorption above K-edge Allow modification spectrum: mammo K-edge Tin (Sn) allows lightweight shielding in protective aprons kV Module VIII.3-Part 1. Design considerations equipment Continued International Atomic Energy Agency 36 Geometrical properties Inverse square law Anode angle Heel effect Module VIII.3-Part 1. Design considerations equipment Continued International Atomic Energy Agency 37 Inverse square law: dose reduction Fluency of beam constant Dose (or doserate) per cm2 proportional to 1/x2 Dose = 1 Dose = 1/4 Dose = 1/9 Module VIII.3-Part 1. Design considerations equipment Continued International Atomic Energy Agency 38 Inverse square law Focus 20cm Front 20cm Patient Image Receptor Reduces geometric unsharpness Reduces patient dose • • 100cm Same dose D at image receptor Increase at entrance point: • • Patient Image Receptor Left: (40/20)2 => 4 D Right (120/100)2 => 1.44 D Module VIII.3-Part 1. Design considerations equipment Continued Front International Atomic Energy Agency 39 20cm Anode angle and resolution Geometrically: size of the effective focal spot directly related to the sine of the angle of the anode. A the angle of the anode is decreased, the effective focal spot is also decreased. Penumbra and geometrical unsharpness decreases Module VIII.3-Part 1. Design considerations equipment Continued International Atomic Energy Agency 40 Heel effect Plot of intensity shows variation along anode to cathode axis Heel effect lower by large FID: not the full cone of radiation (C to B) is used Module VIII.3-Part 1. Design considerations equipment Continued International Atomic Energy Agency 41 Unwanted radiation Module VIII.3-Part 1. Design considerations equipment Continued International Atomic Energy Agency 42 Leakage radiation •Radiation transmitted thtrough tube housing • Image taken on film • Cathode side • Need for Radiation protection •At installation •In use •After maintainance Leakage Module VIII.3-Part 1. Design considerations equipment Continued International Atomic Energy Agency 43 Stray radiation • Is the sum of the leakage and scattered radiation • Scattered radiation has been deviated after leaving the tube • Scatter occurs from all material exposed to the radiation: patient, table, filters,room walls Module VIII.3-Part 1. Design considerations equipment Continued International Atomic Energy Agency 44 Scattered Radiation • Effect on image quality • increasing of noise • loss of contrast • Effect on patient dose • increasing of superficial and depth dose • Possible reduction through : • • • •  use of grid  limitation of the field to the useful portion  limitation of the irradiated volume (e.g.:breast compression in mammography)  use of air gap Module VIII.3-Part 1. Design considerations equipment Continued International Atomic Energy Agency 45 Anti-scatter grid Source of X Rays Patient Scattered X Rays Lead Film and cassette Useful X Rays Module VIII.3-Part 1. Design considerations equipment Continued International Atomic Energy Agency 46 Anti-scatter grid • Radiation emerging from the patient • primary beam : contributes to the image • scattered radiation does reach the detector and • contributes to background without information, and contrast is lowered scattered radiation also contributes to unnecessary dose • the grid (between patient and film) eliminates most of scattered radiation on image receptor Module VIII.3-Part 1. Design considerations equipment Continued International Atomic Energy Agency 47 Anti-scatter grid BUT: to keep same dose on image receptor increase dose Grid NOT recommended for extremities or pediatric radiology Module VIII.3-Part 1. Design considerations equipment Continued International Atomic Energy Agency 48 Scatter reduction by reducing field size Module VIII.3-Part 1. Design considerations equipment Continued International Atomic Energy Agency 49 Scatter removal by air gap Module VIII.3-Part 1. Design considerations equipment Continued International Atomic Energy Agency 50 Scatter removal by narrow beam geometry Module VIII.3-Part 1. Design considerations equipment Continued International Atomic Energy Agency 51 Contrast agent • Tissue characteristics and photon energy • determine photoelectric absorption Basic idea to enhance contrast: lower kV • cf. mammography • But lower kV, lower penetration • How to improve contrast? • Introducing contrast agents: Ba, Iodine • High atomic number : photoelectric ~ Z3 • K-edge falls within diagnostic imaging kV Module VIII.3-Part 1. Design considerations equipment Continued International Atomic Energy Agency 52 Summary on what we learned about the X Ray beam (1) • Which factors influence the X Ray spectrum? • tube potential • kVp value • W, Mo, Rh etc. • inherent + additional • wave shape of tube potential • anode track material • X Ray beam filtration • How to enhance contrast • Lower kV • Use of contrast agents: Barium, Iodine Module VIII.3-Part 1. Design considerations equipment Continued International Atomic Energy Agency 53 Summary on what we learned about the X Ray beam (2) • We became familiar with geometrical properties • Inverse square law • Anode angle • Heel effect • We learned about some unwanted effects • • • • Leakage radiation Stray radiation Scattered radiation How to “remove” unwanted scatter radiation • • • • Reduce field size Grid Air gap Narrow beam geometry Module VIII.3-Part 1. Design considerations equipment Continued International Atomic Energy Agency 54 International Atomic Energy Agency Topic 6: Special applications Dental equipment, Pediatric Equipment Summary of special applications • Pediatric radiology • Specificity • Requirements on equipment, rooms and accessories • Positioning and immobilization • Considerations for the use of the equipment • Protective shielding • Dental radiology • Low dose high frequency • The equipment • Radioprotection and patient dose Module VIII.3-Part 1. Design considerations equipment Continued International Atomic Energy Agency 56 Equipment for pediatric room Module VIII.3-Part 1. Design considerations equipment Continued International Atomic Energy Agency 57 Specificity of pediatric radiology • Longer life expectance • Risk of late detrimental radiation effect greater • Estimated 2-3 x greater than @ 30-40y • or 5-7 x greater than after 50y Justification and optimization even more important • Smaller size require adapted radiographic • • techniques and exposure factors Positioning and lack of co-operation requires short exposures Functional differences (e.g. higher heart rate, faster respiration, inability to stop breathing on command, increased intestinal gas etc.) Module VIII.3-Part 1. Design considerations equipment Continued International Atomic Energy Agency 58 General, equipment and installation considerations (1) • Short exposure times can improve image • • quality. So the generator should have enough power to allow short exposure times (3 milliseconds) and the timer should allow short exposure times. The use of mobile X Ray units in pediatrics could raise special problems (low power:motion blurring). The generator should be of high frequency to improve the accuracy and reproducibility of exposures. Module VIII.3-Part 1. Design considerations equipment Continued International Atomic Energy Agency 59 General, equipment and installation considerations (2) • • • • Automatic exposure control (AEC) devices should be used with caution in pediatrics (they should be adapted specifically to pediatrics).. Careful manual selection of exposure factors usually results in lower doses X Ray rooms for pediatrics should be designed for improving the child’s cooperation (control panel with easy patient visibility and audio communication, etc.). Fast film-screen combinations have advantages (reduction of dose) Module VIII.3-Part 1. Design considerations equipment Continued International Atomic Energy Agency 60 General, equipment and installation considerations (3) • Low-absorbing materials in cassettes, • • tabletops, etc. specially important in pediatrics radiology. The antiscatter grid in pediatrics gives limited improvement in image quality and increases patient dose given the smaller irradiated volume (and mass) the scattered radiation is less Antiscatter grid should be removable in pediatric equipment, particularly fluoroscopic systems Module VIII.3-Part 1. Design considerations equipment Continued International Atomic Energy Agency 61 General, equipment and installation considerations (4) • Image intensifiers should have high conversion • • • factors for reducing patient dose in fluoroscopic systems. Additional tube filtration may allows dose reductions. Pulsed fluoroscopy is recommended since it allows patient dose reduction as in adults For CT examinations, the use of specific technical radiographic parameters for CT examinations should be promoted (lower mAs than for adults, and lower kV in some cases) Module VIII.3-Part 1. Design considerations equipment Continued International Atomic Energy Agency 62 Dental radiology Module VIII.3-Part 1. Design considerations equipment Continued International Atomic Energy Agency 63 Dental radiology • High frequency (~1/4 of all examinations) but • • • • • low dose technique Wide range of entrance doses:0.5 - 150mGy Dental radiology contributed for 1% of collective dose from medical diagnostics Image quality often very low Many of them in private practice with no medical physicist or RP officer and have no medical physics support available Organs at risk: parathyroid, thyroid, larynx, parotid glands Module VIII.3-Part 1. Design considerations equipment Continued International Atomic Energy Agency 64 Intra-oral X-ray dental equipment Some technical parameters: -Tube 65kV 7.5 mA -Filtration: 2mm Al -Focus skin distance: 20cm -Field size: 6cm diameter Module VIII.3-Part 1. Design considerations equipment Continued International Atomic Energy Agency 65 Panoramic dental X-ray equipment Some technical parameters: -Tube 60-80kVkV -Current 4-10mA Module VIII.3-Part 1. Design considerations equipment Continued International Atomic Energy Agency 66 Image receptors in dental radiology Intraoral Radiology • Small films (2 x 3 or 3 x 4 cm) in light-tight envelopes (no screen) Digital intraoral sensors. Compared with category E film, the radiation dose is reduced by 60%. • • Panoramic Radiology • Film-screen combination • Digital sensors. • Compared with film-screen sensitivity class 200, the radiation dose is reduced by 50-70%. Module VIII.3-Part 1. Design considerations equipment Continued International Atomic Energy Agency 67 What we learned about X Ray equipment • We studied X Ray production (the tube, the • • • generator,…) and the variety of radiological equipment (for plain radiography, fluoroscopy, CT, …) We learned about the basic interactions leading to the image formation Image intensifiers may drastically improve image quality and reduce staff dose and the beam characteristics, including the « undesirable » apects (scatter, leakage) Module VIII.3-Part 1. Design considerations equipment Continued International Atomic Energy Agency 68 What we learned about X Ray equipment (ctnd) • Specific examinations need dedicated • • equipment (dental, pediatric, mammographic) Digital images can enhance contrast and extract regions of interest Equipment used by non-radiologists (surgeons, cardiologists,…) require special technical requirements Module VIII.3-Part 1. Design considerations equipment Continued International Atomic Energy Agency 69 Where to get more information (1) • • • • • The Physics of Diagnostic Imaging, David J. Dowsett, Patrick A. Kenny and R. Eugene Johnston, Chapman & Hall Medical, ISBN 0-412-40170-1 Equipment for diagnostic radiology, E. Forster, MTP Press, 1993 The Essential Physics of Medical Imaging, Williams and Wilkins. Baltimore:1994 Imaging systems in medical diagnostics, Krestel ed., Siemens, 1990 The AAPM/RSNA Physics Tutorial for Residents General Overview of Fluoroscopic Imaging, B. A. Schueler, Radiographics Vol 20, 1115-1126,(2000) Module VIII.3-Part 1. Design considerations equipment Continued International Atomic Energy Agency 70 Where to get more information (2) • • • • • European Guidelines on quality criteria for CT. Report EUR 167262, Office for Official Publications of the European Communities, 1999, Luxemburg Radiation exposure in Computed Tomography; 4th revised Edition, December 2002, H.D.Nagel, CTB Publications, D-21073 Hamburg Rational use of diagnostic imaging in pediatrics. WHO, 1987 European guidelines on quality criteria for diagnostic radiographic images. Report EUR16261 (1996) Radiation protection and quality assurance in dental radiology. Radiation protection 81. European Commission.(1995) CG-89-95-971-EN-C Module VIII.3-Part 1. Design considerations equipment Continued International Atomic Energy Agency 71