Download Series 100 Manual (TH-W-APK-00-00-0612-05-A ).indd

Survey
yes no Was this document useful for you?
   Thank you for your participation!

* Your assessment is very important for improving the workof artificial intelligence, which forms the content of this project

Document related concepts

Resistive opto-isolator wikipedia , lookup

Cellular repeater wikipedia , lookup

Interlaced video wikipedia , lookup

Immunity-aware programming wikipedia , lookup

Tektronix analog oscilloscopes wikipedia , lookup

Oscilloscope wikipedia , lookup

Radio transmitter design wikipedia , lookup

Time-to-digital converter wikipedia , lookup

Switched-mode power supply wikipedia , lookup

HD-MAC wikipedia , lookup

Valve RF amplifier wikipedia , lookup

Analog television wikipedia , lookup

Peak programme meter wikipedia , lookup

Analog-to-digital converter wikipedia , lookup

Multimeter wikipedia , lookup

Opto-isolator wikipedia , lookup

Oscilloscope history wikipedia , lookup

Rectiverter wikipedia , lookup

High-frequency direction finding wikipedia , lookup

Transcript
TH-W-APK-00-00-0612-05-A
Series 100 Manual
Series 100 ACT-PAK Manual
ENCLOSURES
Environmental Operating Conditions
Standard
Ambient Temperature: 0º F to 120º F
Relative Humidity: Non-condensating
A free standing or panel mountable package that meets
NEMA1 Requirements.
PHYSICAL SIZE / WEIGHT
NEMA4X
A wall mountable fiberglass enclosure.
FRONT PANEL
Input Pulse Indicator
A red LED that blinks on and off to display each pulse
received from meter.
NEMA1
NEMA4X
Length
6.35”
5.9”
Height
3.0”
5.9”
Width
5.3”
3.8”
Weight
2.4 Lbs
2.5 Lbs
OUTPUTS AVAILABLE
Power Indicator
Standard Proportional DC Output: 4-20 mA
A red LED located below the pulse indicator, indicating the
instruments power supply is functional.
All instruments are supplied with a 4-20 mA DC output
proportion to the flow rate through the metering device. This
signal has the capability of driving external equipment with
an impedance from 0 to 500 ohms without recalibration.
Totalizer / Flow Rate Indicator
Displays both total consumed and flow rate on an
eightdigit transmissive Red LED backlight 0.46” display.
A single display indicates flow rate designated by an R.
The accumulating totalizer is indicated by a blank and the
resettable totalizer annuniciator is a b.The display can be
toggled manually with the front panel SELECT switch or
programmed to automatically to display flow rate and total
consumed on the two available totalizers. The display may
be left in either mode. The number of active digits depends
on the parameters of the input signal.
SPECIFICATIONS
Input Signals
General Requirements
Input Frequency (full Scale) – 1 to 1000 Hz
Maximum loop resistance – 40 ohms
Duty cycle – 50/50 ± 20%
Contact Closure Inputs
Type – SPST ( contact bounce less than 5 ms, interrupting
current – 50 ma max)
Solid State Inputs
Optional Outputs
Option A: Dual 4-20 mA Output Signals: Two (2) 4-20 mA
signals proportional to the flow through the metering device.
With the standard 4-20 mA output, this option provides for
a total of three (3) 4-20 mA signals. If isolation between
devices is necessary an analog isolator is recommended.
May require an enclosure upgrade.
Option Bb: Totalized Contact Output A SPDT contact
closure rated 5 amps, resistive 30 VDC or 250 VAC. The
value of the pulse and the pulse duration may be changed
in the field. Closures smaller than the meter sweep can be
accurately programmed.
Option C1: Keying Contact: A SPST mercury-wetted contact
closure (bounce free, rated 1 amp 24 VDC or 0.1 amps at 117
VAC, all resistive.
Option I: Input Compensator: An input signal multiplier used
when totalizer quantities that are smaller than standard are
specified. Usually not required with the option Bb.
Option K: Slave 4-20 mA Output: A second 4-20 mA signal.
Calibrated the same as the existing 4-20 mA output. If
isolation between devices is necessary an analog isolator
is recommended.
Electrical
Option L: Flow Alarm: A programmable SPDT contact
closure that is activated when the flow rate exceeds or falls
below a preset limit. The switching point may be programmed
in the field. Contact ratings are the same as option Bb.
117 volts ± 10%, 60 Hz, 1 Phase, 3 wire circuit
Power Consumption: 10 volt amps (watts), maximum.
Option P: Panel Mounting: NEMA1 instruments are supplied
with the hardware necessary for panel mounting.
An open collector design conforming to switch specifications
listed above.
Communication Ports: Two (2) serial communication ports
RS485 COMM1, and RS232 COMM2, are available. Only
one port can be used at a time. External programming by the
user is required.
TH-W-APK-00-00-0612-05-A
Page 2 of 8
FLOW RATE INDICATOR / TOTALIZER
Display
An 8-digit transmissive Red (Green also available) LED,
0.46” high (For outdoor or high ambient light conditions,
reflective readouts may be ordered).
Memory
TERMINAL
NUMBER
FUNCTION
IC , EROFI
(Belden 8760)
HSPU, OMNI, HET
(Belden 8770)
15
Input Signal
White
White
14
Shield
Bare (Shield)
Bare (Shield)
13
Ground
Black
Black
12
+12VDC
No
Connection
Red
Non-volatile E2PROM memory retains all programming
parameters and totalized data when power is removed.
AC POWER INPUTS
Front Panel Buttons
SELECT: Toggles display between totalizers and flow rate
indicator, also advances menu selections in the program
mode.
TERMINAL NUMBER
3
Earth Ground
2
117 VAC
1
117 VAC
RESET: Sets resettable totalizer(s) to zero and changes
display in the programming mode.
OPTION C1
Front Panel Displays
TERMINAL NUMBER
COUNT DISPLAY: 8-digit display flashes “Cnt OVEr” to
indicate that the totalizer has turned over.
RATE DISPLAY: 6-digit flashes “OLOLOL” when the
calibrated flow rate is exceeded.
DISPLAY DESIGNATOR: (on top left side of display)
R – Designates that the flow rate is active on the display
b – Designates the resettable totalizer is displayed.
_ – (Blank) Indicates the accumulating totalizer is displayed.
FUNCTION
FUNCTION
10
Common
11
Normally Open
PROPORTIONAL OUTPUT
TERMINAL NUMBER
FUNCTION
4
Negative
5
Positive
NOTE: Terminal #5 is common to both option K and the 4-20
ma
Outputs
OPTION BB OR OPTION L
Proportional Output – 4-20 mA, 500 ohms maximum
impedance.
TERMINAL NUMBER
Option A – Dual Analog Outputs - A dual 4-20 mA output
signal in addition to the standard 4-20 mA signal supplied.
The option A signals may be calibrated to different flow rate
parameters than the standard 4-20 mA signal. These signals
operate through a maximum impedance of 500 ohms.
Option B, Bb,L – SPDT, Rated 5 amps resistive, 24 VDC or 3
amps, resistive, 117 VAC. Pulse duration for Bb and L can be
set in the field between 0.01 to 99.99 seconds, or set to latch.
Option B duration is set to approximately 40 ms.
Options C1,D – SPST, mercury-wetted bounce free rated 1
amp, 24 VDC or 0.1 amp at 117 VAC. All resistive loads.
FUNCTION
7-16
Normally Open
8-17
Common
9-18
Normaly Closed
OPTION K
TERMINAL NUMBER
FUNCTION
5
Positive
6
Negative
For connections to options A, B C2 or D see option board
insert.
Error Indicators
Electrical Connections
Cnt OVEr — indicates that totalizer A has rolled over.
Complete all connections to the metering device and
instrument before applying power. If it becomes necessary
to disconnect any of these connections, first remove the AC
power to the instrument.
bCntOVEr—indicates the totalizer b has rolled over.
OLOLOL—indicates that the flow rate has exceeded the
calibrated flow.
Totalizer Over—To reset the totalizers back to zero (0) locate
the three (3) pins connector in the corner of the PC board
opposite terminals 12 through 15 (see connection diagram).
TH-W-APK-00-00-0612-05-A
Page 3 of 8
Remove the power from the instrument. Move the jumper on
the left and middle pins to the right hand and middle pins.
Reapply the power momentarily. Turn the power off and
move the jumper back to the left and middle pins. Reapply
the power; the totalizers should read zero (0).
Flow Rate Over—There are a number of discrepancies that
can cause the OLOLOL indication, listed below.
•
•
•
Instrument is calibrated for a different meter.
Meter flow rate exceeds instrument calibration rate.
Meter maximum flow capacity is exceeded.
To eliminate a flow rate calibration problem check that the
meter and instrument calibration agree. If recalibration is
necessary refer to module #2 in the Instrument Recalibration
Section.
CntA dP – (Numeric selection depends on the factors
determined below.)
CntA ScF – (This factor equals the desired totalizer registration
divided by the number of pulses for that registration. The
number varies with each meter and registration. Example for
a W2000 DRE meter reading in gallons. 500.9663 pulses
equals 1000 gallons. Therefore, the factor is 1/500.9883 or
0.0019961. Only four(4) digits can be entered for the factor.
Therefore, the number entered would be 0.0020, the factor
rounded off. To improve the accuracy the decimal should
be moved two (2) places and the factor becomes 0.19961.
Since the decimal was moved two (2) places, the selection
for CntA dP above is 0.00 and the factor entered into Cnt
ScF is 0.1996.
CntA rSt—to CtLd
Calibrating The Readout
CntA dir—NOr
CAUTION: All instruments are preprogrammed to
operate with the meter specified when ordered and
no additional programming is required. Entering the
program mode may cause erroneous operation. This
should only be done when the operating specifications
have changed.
CntA Ld—000000
To enter the programming mode, depress and hold the SEL
button on the readout for 2 seconds. In the program mode
the display will show Pro and flash to nO. There are five
(5) programming modules available. Press the RST button
to step through the modules. When the correct module is
displayed, press the SEL button to enter that module.
Cntb Bat—nO
Cntb dP – (Usually the same as CntA dP) If the units of
measure are not the same use the procedure listed above
to determine the decimal point location and factor. Possible
examples are one totalizer reading in gallons and the second
in cubic feet, cubic meters or imperial gallons.
Cntb ScF – (Usually the same as CntA ScF, if not use the
procedure listed above.)
RSt P – UP – nO
USEr InP – reset
In the module the function being programmed is shown then
flashes to the value presently entered for the function.
USEr ASn – both A – b
Pressing the RST button stops the display from flashing and
places the unit into the data selection mode.
Module #2 (2 – rAte)
In the selection mode the operator presses the RST button
to scroll through the parameters available for that menu.
RAtE dP – The decimal location depends on the flow rate
value. Typically is set to display the multiplier to use all six
digits on the display. (This setting is determined below.)
When the desired selection is displayed press the SEL
button to store the data and advance to the next menu. When
setting numerical values use the RST button to increment
the flashing digit, and momentarily press the SEL button to
advance to the next digit. To start the least significant digit
flashing momentarily depress the RST button.
To exit the programming mode and save all parameters
step through the module until one of the five main modules
is displayed, then press the SEL button until Pro nO is
displayed. Pressing the SEL button at this point enters the
data and closes the program mode.
THIS PROGRAMMING PROCEDURE ONLY APPLIES 5
TO VERSION 3.0 AND 3.1 READOUTS.
PROGRAM MENUS
Module #1 (1 – input)
InP A-b – dual Cnt
RAtE Enb – yES
RAtE dSP – This number represents the maximum flow
rate to be displayed. In establishing this number, all 6 of the
available digits should be used if possible. (This setting is
determined below.)
RAtE InP – (The Hertz value at the programmed flow rate.
This number should be enhanced as follows: For a W2000
DRE the maximun flow rate is 2500 GPM. At 20.87 hertz.
The 2500 could be used for RAtE dSP above, but to improve
the accuracy it should be set to 25000. The decimal was
moved one place to the right, the hertz value must also be
moved one place to keep proportions the same. Therefore,
RAtE dP should be 0.0, RAte dSP should be 25000.0 and
the hertz set to 208.7.
TH-W-APK-00-00-0612-05-A
Page 4 of 8
LO – Udt – Set to 5.0. this sets the number of seconds
between flow rate display updates. If the flow rate indicator
is varying due to fluctuating meter signals, increasing this
number reduces display fluctuations. May be set between
0.1 and 99.99 seconds.
HI-Udt – Set to 90.0. Sets the number of seconds the
instrument will wait for a meter pulse before going to zero
(0). May be set between 0.1 and 99.9.
If Count b was selected above set the value to represent the
total number of counts on b that will occur before the relay
actuates.
SP1 Out—Nor
SP1 LIT—Nor
SPt P-UP – Set to SAVE.
SP1 tYPE—Hi-Act
Module #3 (3-dSPLAy)
SP1 Stby– nO
SEL Enb – Set to yES.
SP1 AUtO—ZErO-Str
RSt Enb - This step programs the counter(s) to be reset to
zero (0) with the RST button. Typically set to Count b to reset
the b totalizer with the RST button. To reset the A totalizer set
to Count A, to make both resettable set to both A-b. When
using option Bb the b counter must be set to reset.
SP1 OFF2—nO
d-ScroLL – Typically set to nO, set to yES to have the display
automatically scroll between A total, b total and the flow rate.
d-Color - Specifies the color of the display (Typically set to
rEd, if green is desired set to Grn)
d-LEVEL – Sets display intensity ( Typically set to 5, to
reduce the intensity, reduce the number)
Pro Code – Factory set to 124. Although a lock code is not
necessary, it is an electronic lock to prevent tampering. It
may be changed to any three (3) digit number.
CodE VEr—nO FACt Set - Set to no. Setting to yES causes
the program stored in memory to be erased.
Module #4 (4-SetPt)
This module is only active when options Bb or L are present.
SPt SEL—SP1
SP2 Enb—nO
SPt ASn – Set to rAtE for option L. Contact is then based on
the flow rate. Set to Count b for option Bb. Contact output is
based on quantity.
SPt ACt – If rAtE was selected above set to BOUnd if
contacts are to close when the flow rate equals or exceeds
the set point and close below the set point.
Set to LAtCH to hold the contacts closed once the flow
quals the set point. A manual reset is required to release the
latched contacts.
For quantity applications where Count b was selected above
set to t-OUt .
SPt tOUt - Only active if t-OUt was selected above. SET the
number of seconds that the relay remains closed each time
it is activated.
SPt VAL – If rAtE was selected above,set the actual flow rate
where the contact closure is desired.
SP2 OFF1—nO
Ch-COLOR – Typically set to nO. Set to yES to force the
backlight color to change when the setpoint output is
activated.
To activate set point 2. Set the first step SPt SEL to SP2
and the second step SP2 Enb to yES. Follow the above
procedure starting with SPt ASn.
Module #5— (5—SeriAL)
The serial communications are only accessible when an
optional RS232 or RS485 serial communications module
is ordered in the instrument. A separate data sheet and
programming is included with this option.
Setting Option B Factors
The option B factor is coded in binary and can be changed
in the field to any whole number between 1 and 4095. To
determine the factor the following procedure should be used.
1. Determine the quantity that the contact closure
represents.
2. Determine the value of a pulse from the meter in either
quantity per pulse (cubic feet/pulse) or pulses per
quantity (pulses/cubic foot).
3. Determine factor as follows (assuming desired output is
in 1000 cubic feet increments and meter output is 0.564
cubic feet per meter pulse).
a. Calculate the pulses per cubic foot by determining the
reciprocal of 0.564.
1/0.564= 1.7730496 Pulses/Ft3
NOTE: If meter data is available in pulses per cubic foot, this
step can be eliminated.
b. Multiply the pulses per cubic foot by the quantity to be
indicated. (In this example 1000).
1.7730496 X 1000 = 1773.0496 pulses per 1000 ft3.
c. Since the circuitry is only capable of handling whole
numbers, the factor must be rounded to the nearest
whole number (1773).
TH-W-APK-00-00-0612-05-A
Page 5 of 8
NOTE: When the factor is rounded , a small error is
introduced, in this example .0028%. In applications where
this error is not acceptable the addition of option I or using
option Bb will reduce the error considerably.
d. To set the factor on the option B on the option B board
locate the 12 position dip switch on the board. To start,
place all 12 switches in the off position. Setting the factor
is accomplished by turning the dip switch ON so that the
sum of the factors equals the number desired. There is a
unique combination of factors for each desired number.
They are determined in the following manner, select
the largest binary number from the chart that does not
exceed the balance, continue selecting and subtracting
until the balance equals zero (0). At this point, all factors
subtracted must be set into the instrument by placing the
associated switch in the ON position.
3. Apply the formula noted in step #1 to determine the
calculated Range capacitor value. Example: For a
W2000 Turbo Meter at 2500 GPM.
C = 1.54 / 20.874
C = .0738 microfards.
4. A .0738 microfard capacitor is not a standard, that
cannot be procured. To be in the calibration range of
the adjusting pots, the C value selected must be within
10% of the calculated valve. This may require paralling
capacitors to reach the desired value. To achieve
maximum adjusting range, the final value should be a
close to the calculated value as possible.
Example: The calculated value is .0738 microfarads,
therefore, the final capacitor value must be within the 10%
range of .0664 to .0810 microfarads. Although there is a
standard .068 microfarad capacitor available, it is at the
low end of the range. To move the final selection closer
to the middle, the .068 microfarad capacitor should be
paralleled with an .0068 microfarad capacitor increasing
the value to .0748 microfarads. Because you are limited
to available capacitor values and all capacitors have a
tolerance some trail and error may be necessary if a
capacitance meter is not available.
5. Install the correct capacitor values on the PC Board.
The stability of the range capacitor affects stability of the
converter; only polycarbonate or polystrene capacitors
should be used.
The D/A converter is factory set to a precise frequency to
match the water meter or signaling device. If it becomes
necessary to recalibrate the D/A converter, the following
procedure must be followed and high quality test equipment
used. Any component changes must be made with equivalent
components.
Range Capacitor Selection
The range capacitor value is determined by the frequency
output and the desired range for the meter that is driving it.
To determine the Range Capacitor value:
Test Equipment Setup (Using a milliampere meter)
To calibrate the converter you need a signal generator with a
digital readout and a digital VOM that can read milliamperes
in the 20 milliampere range. A voltmeter can be used with a
precision resistor. The signal generator must have a open
collector or a switch output. See option 1. The typical signal
generator has a pulse type output. To use this type of signal
generator a switching type transistor must be connected
between the signal generator and the instrument being
calibrated. See figure #2.
1. Use the formula C range 1.54 / f (C value is in
microfarads.)
2. Determine the full scale frequency value for the signal
source. Exp. A 6” W2000 Turbo Meter outputs 20.874
hertz at 2500 GPM. If the 20 mA signal represents 2500
GPM then use 20.874 hertz for the (f) value in the above
formula. If the 20 mA is to equal a lower flow rate then
scale the frequency value to the lower requirement. For
example assuming a desired maximum flow of 1500
GPM or 60% of the maximum flow range, the frequency
at that flow rate is 60% of the 20.874 hertz or 12.524
hertz.
1. Disconnect the AC power to the instrument.
2. Disconnect any equipment connected to the 4 - 20 mA
output terminals.
3. Connect a milliampere meter directly across the 4-20
mA terminals on the instrument.
TH-W-APK-00-00-0612-05-A
Page 6 of 8
4. Connect the signal generator to the instrument being
calibrated. The plus output of the signal generator is
connected to the Signal Input terminal of the instrument.
Connect the minus or ground terminal of the signal
generator to the Ground terminal of the instrument. The
+12 VDC terminal is not connected. Using a voltmeter
insure that the 12 VDC output is present. (See Figure#2)
5. Apply power to the instrument.
6. Set the signal out of the signal generator to 0 Hertz out.
7. Check the output current by placing a milliampere
meter directly across the 4 - 20 mA output terminals. It
should read 4 milliamperes, if it does not adjust the zero
potentiometer until the meter reads 4.
NOTE: The 4 - 20 mA signal does not respond immediately
when the potentiometer is adjusted. Adjustment should be
in small increments allowing time for the current to stabilize
before further adjusting.
8. Set the signal generator to the full scale hertz value. The
milliampere meter should read 20 milliamperes. If it does
not, adjust the SPAN potentiometer to 20 milliamperes.
Again, adjustment should be in small increments
allowing time for the current to stabilize before further
adjusting.
9. Return to step #6 and readjust the zero setting if
necessary, then again check the full scale value. This
process of checking the zero and full scale values may
have to be repeated 4 or 5 times. The adjustment should
be smaller each time until the final values are achieved.
10. Set the signal generator to half scale (50% of the full
scale hertz). Check the milliampere meter, it should read
12 milliamperes. If it does not check zero and full scale
again. If the 12 milliampere output can not be achieved,
the circuit is not linear and can not be calibrated. It
should be returned for repair.
11. Disconnect the milliampere meter and reconnect the
water meter transmitter. Apply power.
Test Equipment Setup (Using a volt meter)
To calibrate the converter you need a signal generator with
a digital readout and a digital VOM that can read voltage in
the range determined by the resistor selected. The signal
generator must have a open collector or a switch output. The
typical signal generator has a pulse type output. To use this
type of signal generator a switching type transistor must be
connected between the signal generator and the instrument
being calibrated. See figure #1 and #2.
1. Disconnect the AC power to the instrument.
2. Disconnect any equipment connected to the 4 - 20 mA
output terminals.
3. Connect a precision resistor across the 4- 20 mA
terminals on the instrument The resistor value can be
between 1 to 500 ohms. The value selected depends on
the voltage required by the external device being driven,
resistor availability and the range of the voltmeter. For
this procedure we used a 500 ohm resistor . The voltage
out would be 2 (.004 X 500) to 10 (.020 X 500) volts.
4. Connect the signal generator to the instrument being
calibrated. The plus output of the signal generator is
connected to the Signal Input terminal of the instrument.
Connect the minus or ground terminal of the signal
generator to the Ground terminal of the instrument. The
+12 VDC terminal is not connected. Using a voltmeter
insure that the 12 Vdc output is present. (See Figure#2)
5. Apply power to the instrument.
6. Set the signal out of the signal generator to 0 Hertz out.
7. Check the output voltage by placing a volt meter directly
across the resistor. It should read 2 volts, if it does not
adjust the zero potentiometer until the meter reads 2
volts.
NOTE: The voltage does not respond immediately when the
potentiometer is adjusted. Adjustment should be in small
increments allowing time for the current to stabilize before
further adjusting.
8. Set the signal generator to the full scale hertz value.
The volt meter should read 10 volts. If it does not, adjust
the SPAN potentiometer to 10 volts. Again, adjustment
should be in small increments allowing time for the
voltage to stabilize before further adjusting.
9. Return to step #6 and readjust the zero setting if
necessary, then again check the full scale value. This
process of checking the zero and full scale values may
have to be repeated 4 or 5 times. The adjustment should
be smaller each time until the final values are achieved.
10. Set the signal generator to half scale (50% of the full
scale hertz). Check the volt meter, it should read 6 volts.
If it does not check zero and full scale again. If the 6 volt
output can not be achieved, the circuit is not linear and
can not be calibrated. It should be returned for repair.
11. Disconnect the milliampere meter and reconnect the
water meter transmitter. Apply power.
READ BEFORE CHANGING THE PROGRAM
UNLOCK PROCEDURE
The following procedures involve programming changes
when making these changes use the Programming Menus.
After making a change press and hold the SEL button for
approximately 2 seconds to enter the change into memory.
After making all changes always return to normal operation
before removing power to permantely enter all changes into
memory.
To enter the program mode, press and hold the SEL button
on the readout until it flashes Pro CodE (If it does not display
PrO nO press the SEL button until it does) and flashes over
to 000 with the least significant digit flashing. Press the RST
TH-W-APK-00-00-0612-05-A
Page 7 of 8
button 4 times to read 004 (if you exceed the desired number
keep pressing the RST button until the number comes back
to 4. When the 4 is displayed, press and hold the SEL button
until the second digit starts flashing, reading 004. Press the
RST button 2 times to read a 2 the readout should read 124.
If it does not, repeat the above process until the proper code
is displayed. When the correct code is displayed, press and
hold the SEL button until the display flashes PrO nO.
Setting the A Totalizer to a Preset Number.
1. Determine the number to be preset into totalizer A.
2. Go to Pro—nO If instrument is locked follow the Unlock
PROGRAMMING OPTION L
1. Go to PrO nO. If instrument is locked follow the Unlock
procedure above to unlock.
2. Press the RST button four (4) times the readout should
read 4 SEtPt.
3. Press the SEL button once. Readout should read SPt
SEL – SP2. If not press the RST button until it does.
4. Press the SEL button 2 times. Readout should read SP2
- ASn - rAtE. If not press the RST button until it does.
5. Press the SEL button the readout should read SP2
Procedure above to unlock.
Act - bOUNd* or LAtCH**. If not press the RST button
until it does. (See * and ** below)
3. Press the RST button once.
6. Press the SEL button one (1) time the readout should
Go to Module #1. Display should read 1– INPUt.
4. Press the SEL button 4 times. Readout reads CntA
rSt - to ZEro press RST to change to CtLd.
5. Press the SEL button 2 times the readout reads
CntA Ld—enter the desired number to be preset.
6. Press and hold the SEL button for 2 seconds to enter
number.
7. Press the SEL button to go to Pro—nO.
8. Press the RST button 3 times. Display should read 3 dSPLAy.
9. Press the SEL button 2 times. The readout reads RSt
Enb - Count b
10. Press the RST button to change to Count A.
11. Press the SEL button to until the readout returns to
normal operation.
12. Press the SEL button to go to Counter A.
13. Press the RST button. Observe that the desired number
was entered into counter A.
14. Go to Pro-nO If instrument is locked follow the
Unlock Procedure above to unlock.
15. Press the RST button once.
16. Go to Module #1. Display should read 1– INPUt.
17. Press the SEL button 4 times. Readout reads CntA
rSt - to CtLd press RST to change to ZEro.
18. Press the SEL button to go to Pro—nO.
19. Press the RST button 3 times. Display should read 3 dSPLAy.
20. Press the SEL button 2 times. The readout reads RSt
Enb - Count A .
21. Press the RST button to change to Count b.
22. Press the SEL button to return to normal operation.
read SP1 - VAL and change over to a number. This is the
flow rate where the set point will activate if acceptable
press the SEL button if not change to the desired
number and press SEL button.
7. If SP2- Act is set to LAtCH and you want to unlatch with
the RST button on the front of the readout.
8. Press the SEL button 5 times. Readout should read
SP2-rSt yES, if not set to yES with the RST button.
In all other cases SP2-rSt should be set to nO.
9. Press the SEL button until the readout returns to normal
operation.
* bOUNd causes the relay to close when the flow rate equals
or exceeds set point. If the flow rate falls below the set point
the relay will open.
** LAtCH causes the relay to close when the flow equals
or exceeds the set point. Once closed it will remain closed
regardless of flow rate until the RST button is pushed.
Change Counter A and/or Counter b Scale Factor
1. If the counter scale factor must be changed, this is
accomplished in Module #1.
2. Go to PRO nO If instrument is locked follow the
Unlock procedure above to unlock.
3. Press the RST button once the readout should read
1 - INPUt
4. See procedure for Module #1 in the Program Modules
under CntA Scf. In the example the CntA ScF was 0.1996
with two (2) decimals entered under CntA dP. Following
the same logic 50.09663 pulses equals 100 gallons, the
reciprocal is 0.019961. To change to 100 gallon counts
change CntA dP to 0.0 (one (1) decimal ) and leave CntA
ScF at 0.1996. To change to 10 gallon pulses, change
Cnt dP to 0 (no decimals). After completing, press
SEL until the display shows Pro nO. Pressing it again
will return the display to normal operation. The above
procedure may be used for counter b by changing Cntb
Scf and Cntb dP.
TH-W-APK-00-00-0612-05-A
Page 8 of 8
Reducing the value of the counter b allows for smaller
quantities to be detected by option Bb. If the value of
counter b is 1000 gallons the Bb option can only count in
1000 gallon increments. Setting the value to 10 gallons then
any increment of 10 can be programmed. For example a
750 gallon contact closure would not be possible with 1000
gallon increments, but with 10 gallon increments with a set
point of 75 ( SP1 VAL = 75) it can be achieved. Reducing the
value of a count may cause a slight error in the total value.
Programming Option Bb ( Pulse Width and/or Value)
1. Determine the factor to obtain the desired contact
closure output. Example: For a W2000 DRE 500.96639
pulses represent 1000 gallons. To determine the factor,
take the reciprocal of the pulses per 1000 gallons.
(1/500.96639) = .001996)
Totalizer Reset
The electronic totalizer flashes “Cnt OVEr” to indicate that
the totalizer has rolled over. When this occurs, a manual
reset is necessary. With a NEMA4X enclosure, open the
door. Locate the 3 pin connector on the printed circuit board;
it is on the side of the board opposite the transformer above
the mounting hole. There is a 2 place (blue or black) jumper
on the 2 pins on the left side of the connector. Remove the
jumper and place it temporarily on the center and right hand
pins. This must be done with the power on, if preferable, turn
the power to the instrument off before moving the jumper.
After placing the jumper on the center and right pins, turn the
power on; this resets the totalizer to zero (0). After resetting,
move the jumper back to the orginal position.
2. Go to Pro nO. If instrument is locked follow the Unlock
Procedure above to unlock.
3. Press the RST button four (4) times the readout should
read 4 - SEtPt.
4. Press the SEL button once. Readout should read SPt
SEL - SP1. If not read SP! press the RST button until
it does.
5. Press the SEL button 2 times. Readout should read SP1
ASn - count b. If not press the RST button until it does.
6. Press the SEL button the readout should read SP1
Act t-Out. If not press the RST button until it does.
7. Press the SEL button the readout should read SP1 tOUt and change over to a number. This is the pulse
width in seconds if acceptable press SEL if not change
to desired number and press SEL button.
8. Pressing the SEL button above causes the display to
read SP1 VAL and change over to a number. This is the
number of counts on counter b that equal a pulse out.
For example if counter b is reading 10 gallon increments
and SP1 Val reads 100 this will create 1000 gallon
pulses. (10 X 100 = 1000)
9. Press the SEL button until the readout returns to normal
operation.
© All products purchased and services performed are subject to Sensus’ terms of sale, available at either; http://na.sensus.com/TC/TermsConditions.pdf or 1-800-METER-IT. Sensus reserves the right to modify
these terms and conditions in its own discretion without notice to the customer.
This document is for informational purposes only, and SENSUS MAKES NO EXPRESS WARRANTIES IN THIS DOCUMENT. FURTHERMORE, THERE ARE NO IMPLIED WARRANTIES, INCLUDING
WITHOUT LIMITATION, WARRANTIES AS TO FITNESS FOR A PARTICULAR PURPOSE AND MERCHANTABILITY. ANY USE OF THE PRODUCTS THAT IS NOT SPECIFICALLY PERMITTED HEREIN
IS PROHIBITED.
8601 Six Forks Road, Suite 700
Raleigh, NC 27615
1-800-638-3748
www.sensus.com/water