* Your assessment is very important for improving the workof artificial intelligence, which forms the content of this project
Download V. Images and Results with Si
Survey
Document related concepts
Transcript
Characterization of a Dual Energy Angiographic System with CCD+FOS and Si-strips Detectors G. Baldazzi Member, IEEE, D. Bollini, A.E. Cabal Rodriguez, M. Gambaccini, M. Gombia, P. Grybos J. Lopez Gaitan, G. Pancaldi, F. Prino, L. Ramello, L. Roma, P. L. Rossi, A. Sarnelli, K. Swientek, A. Taibi, A. Tuffanelli, M. Zuffa We have investigated a non-conventional angiographic imaging methodology called Dual Energy Angiography (DEA) based on a quasi-monochromatic X-rays source. An experimental DEA apparatus was developed. The Bragg monochromator, mounted on a standard W-anode X-ray tube, generates two thin parallel beams with energies peaked before (at EL = 31.1 keV) and after (at EH = 35.3 keV) the iodine K-edge (EK = 33.2 keV). Images result as the difference between the logarithms of the transmitted intensities of the two beams. A dedicated dynamic phantom, simulating different tissues absorption and containing calibrated vessels, was scanned under the monochromator slits by a PC-controlled scanning system. Two types of detectors were used and tested: a linear CCD coupled with FOS and a Si-strips detector with energy window discriminator. In this work resulting imaging capabilities of the DEA experimental system were compared with those of a commercial SIEMENS ANGIOSTAR digital subtraction angiographic apparatus. Abstract-- I. INTRODUCTION Digital Subtraction Angiographic technique (DSA) is one of the most invasive and risky radiodiagnostic procedures [1]. In fact, the long X-ray fluoroscopy time – needed to verify the insertion and site of catheters and wire guides, the injection of the iodine contrast medium (with a concentration of 320370 G. Baldazzi, D. Bollini and M. Gombia are with the University of Bologna, Dep.t of Physics, viale Berti Pichat 6/2, 40127 Bologna, Italy (telephone: +39051-2095152, e-mail: [email protected]). G. Baldazzi, D. Bollini, G. Pancaldi and M. Zuffa are with the INFN Section of Bologna, viale Berti Pichat 6/2, 40127 Bologna, Italy. A.E. Cabal Rodriguez is with CEADEN, Havana, Cuba. M. Gambaccini and A. Sarnelli are with the University of Ferrara, Dep.t of Physics, via Paradiso 12, 44100 Ferrara, Italy. P. Grybos and K. Swientek are with the AGH University of Science and Technology, Dep.t of Nuclear Electronics, Krakow, Poland. J. Lopez Gaitan is with the Universidad de los Andes, Bogota, Colombia. F. Prino and L. Ramello are with The University of Piemonte Orientale Dep.t of Science and Advanced Technologies and INFN, via Bellini 25/G, 15100 Alessandria, Italy. L. Roma is with the University Hospital S. Orsola, via Massarenti 9, 40138 Bologna, Italy. P. L. Rossi is with the University of Bologna Health Physics Service, via Massarenti 9, 40138 Bologna, Italy. mgI/ml) and the sequences of radiological frames could represent a risk for patient [2]. In particular, visualization of small vessels, as in pediatric patients, needs higher iodine concentration (370 mgI/ml) and longer fluoroscopy time (20 minutes or more), that could determine very high dose values (also Entrance Skin Dose values of 1 Gy or more) if added to the necessity of repeating X-rays exposure for multi-projection diagnostic images. Because traditional DSA – also known as “temporal DSA” – is based on logarithmic subtraction of images acquired before and after contrast medium injection, only the part of the X-rays spectrum with energy surrounding the iodine K-edge value (E=33.17 keV) produces the angiographic image. So, large part of the polychromatic spectrum contributes only to patient dose or image penumbra [3]. Different imaging systems based on added filtration of the primary beam were proposed [4] but difficulties involving the reduction of the Signal to Noise Ratio (SNR) represent a limit to the clinical application of these techniques. Recently, the application of the DEA technique with high intensity highly monochromatic synchrotron radiation [5],[6],[7],[8],[9],[10],[11], generated angiographic images with reduction of contrast medium concentration and patient dose, but cannot represent a practical solution for clinical applications. To overcome these fundamental limitations, we have developed and tested a new imaging system [12] based on the Bragg diffraction of the primary polychromatic X-ray beam on a highly oriented pyrolitic graphite mosaic crystal. The experimental DEA apparatus utilize a custom Bragg monochromator fitted on a W-anode X-ray tube and produces two thin (1504 mm2) and quasi-monochromatic parallel beams. The detector system was constituted by two HAMAMATSU S7199 linear image sensors CCD coupled with Fiber Optic plate with CsI(Tl) Scintillator (FOS) mounted on a micrometric motion system. Phantoms were scanned between the monochromator and the detector and an adapted version of the logarithmic subtraction, that we have called hybrid subtraction, provide the enhanced image of the vessels morphology. A second type of detector was used. It was a Si strips linear detector [13], [14]. This detector has spectroscopic capabilities and permits to work in an energetic window to reduce scattering noise. On the other hand, the difference between the intensities of the two beams, the energetic shift (that is the difference in energy measured along a transversal section of each slit) and intrinsic properties of the detectors could represent limitations of the DEA system. For this reasons, the aim of this work is to characterize and evaluate the performance of our DEA apparatus, comparing the image results with a professional DSA angiograph used in clinical applications. II. MATERIALS AND METHODS A. X-ray source: the Monochromator The polychromatic beam generated by a conventional X-ray tube impinge on a 002-pyrolytic graphite mosaic crystal (28601 mm3 with certified mosaic spread 0.26) with an incidence angle 3.2. This is the Bragg angle [15] needed to select diffracted beams with peaks energies lower and higher of the iodine K-edge by means of two slit collimators [16]. The choice of the mosaic graphite crystal was made to gain in diffracted photonic intensities [17], to have spectra with sufficiently great fluence and a FWHM of about 3 keV. Fig. 1. Spectrometric analysis of the beams along coronal plane (perpendicular to the slits): spectrua was acquired using a nitrogen cooled HPGe detector with a collimator of 0.21 mm2 translated under the slits. In order to define the energetic properties of the beams as well as the photonic fluxes, a spectrometric analysis was performed. Figure 1 shows the shape of the quasimonochromatic energetic beams. Their properties are summarized in Table I. According to Zachariasen theory for the Bragg diffraction on graphite mosaic crystal [18], the photonic flux of the Low Energy Beam (LEB EL = 31.1 keV) is roughly 3 times more intense of the High Energy Beam (HEB EH = 35.3 keV). Table I: Beams Properties @ 70 kV, 25 mA anode voltage and current, 25 mm slit-detector distance. LEB HEB Mean Energy (Peak Centroid in keV) 31.1 35.3 FWHM (keV) 2.3 2.3 Mean Photon Flux (photons×mm-2×mA-1s-1) 1.9E+04 5.2E+03 B. The CCD detector The detector, produced by HAMAMATSU, is made with two CCD chips mounted side to side closely (with only 130200 m of dead space) and coupled with CsI(Tl) FOS. The CCD pixel size is 4848 m2 and the overall detector size is 15361282 square pixels with an active area of 147.4566.144 mm2 (plus the dead space between the chips). CsI(Tl) scintillator is grown as needle-shaped crystals, each having a diameter of about 2 m to increase internal conversion efficiency between X-rays and light photons The CCDs were mounted inside a vacuum proof case of Cu with the dedicated electronics. C. The Si-strips detector The detector consists of two one-dimensional silicon array (to cover contemporarily both beams) with photon counting capabilities. A double energy threshold is implemented to exclude scattered radiation. Any array has 384 silicon microstrips of 100 m pitch and 300 m Si-wafer thickness. Each array is equipped with 6 RX64 ASICs. The ASIC includes a charge preamplifier, a shaper, a discriminator and a 20-bit counter for each of its 64 channels [14]. D. Dynamic phantom The test object used to evaluate imaging properties of the systems is shown in Figure 2. It consists of 5 different epoxy blocks with catheters of different diameters: (2.000.05) mm, (1.800.05) mm, (1.600.05) mm, (1.000.05) mm, (0.800.05) mm and (0.400.05) mm. This phantom has a particular design that can confuse the Automatic Exposure Control (AEC) always active in the Siemens angiograph. So, the epoxy blocks are surrounded by a lead layer that masks the X-rays excess. On the same support, a bar pattern (for modulation transfer function analysis) and inserts with high and low contrast were embedded in order to obtain a multi-purpose object to test the physical and imaging properties of every angiographic system. Catheters are filled automatically by a pump: this feature permits to analyze the vessels perfusion as well as to acquire mask image – in DSA – and subtracting images without phantom movements to eliminate image artifacts. III. DEA IMAGE PROCESSING A. DEA imaging Two images (LEB and HEB images) were recorded scanning the dynamic phantom along the orthogonal direction of the beams. Images are composed of N frames, each one with 25 rows of 3072 pixels. The image matrices contain (N25)3072 numerical data with 12 bits of dynamical range. SI (i, j ) ln HEB final (i, j ) ln LEB final (i, j ) , where i 1,...,25 N j 1,...,3072. (1) This allows to reduce background noise in subtracted image and to enhance iodine contrast. IV. IMAGES AND RESULTS WITH CCD DETECTOR Water Bone Phantom 1 Phantom 2 Phantom 3 Phantom 4 Phantom 5 Lin. Abs. Coeff. (1/cm) 2.5 Energy range of interest for D.E.A. 2.0 1.5 1.0 In order to evaluate and compare the image properties, we have collected images of the dynamic phantom with our D.E.A. system and with a commercial D.S.A. apparatus, filling catheters with different iodine concentration (starting from 185 mgI/ml – i.e. ½ of typical iodine concentration injected in pediatric patient during an angiographic procedure - to 20 mgI/ml). 0.5 0.0 20 30 40 50 60 70 80 Energy (keV) Fig. 3.a: Images of the dynamic phantom filled with 185 mgI/ml. On the left, DEA image compared with the DSA image on the right. The Phantom 1 is on the top, the Phantom 5 on the bottom. Fig. 2: Every block of the dynamic phantom was characterized in terms of linear absorption coefficient (1/cm) with a X-ray monochromatic source. The first block on the left is the phantom 1 and so on. Images are weighted by means of a flat field process, to correct the different efficiency and dark current of each CCD pixels and are normalized for X-ray tube fluctuations by reading the counts of a custom miniaturized solid-state exposure meter fitted inside the monochromator. B. Hybrid subtraction Since the energy spread between the two quasimonochromatic beams is of 4 keV, it is not possible to apply the classic logarithmic subtraction algorithm between LEB and HEB images because of the energy dependence of the absorption coefficients. If applied to DEA, the classic DSA algorithm produce an incomplete background contrast cancellation, reducing visibility of vessels filled by contrast medium. For this reason, we have implemented a hybrid algorithm that weight HEB and LEB images by using two parameters (and ) correlated to the differences of between tissues: Fig. 3.b: The same as 3.a but with 95 mgI/ml. For each image and for each block of the dynamic phantom, an average contrast profile of the vessels was calculated in order to define the sensitivity of the systems. In particular, for every block and for every iodine concentration was selected a ROI as great as the block analyzed (i.e., to minimize statistical noise fluctuations), was measured 20 transversal profiles inside the ROI and was evaluated the average contrast profile as the mean of these 20 profiles. Profile (Grey Level) 180 160 140 120 100 80 60 40 0 200 400 600 800 1000 1200 Distance (pixel) Fig. 3.c: The same as 3.a but with 46 mgI/ml. Fig. 4.b: Example of average contrast profiles evaluated on images. These are DEA profile for the phantom 4. Different curves represent profiles obtained with a different iodine concentration (from 185 mgI/ml to 20 mgI/ml). The “overshoots” on the catheter edges are due to the convolution with a nonoptimized filter. Same graphs were calculated for every block obtaining similar shapes. 100% 90% 80% MTF 70% 60% 50% 40% DEA 30% 20% DSA 10% Fig. 3.d: The same as 3.a but with 20 mgI/ml. 0% 0 1 200 3 4 5 180 Fig. 5: Modulation transfer Function (MTF) of the DSA angiograph in comparison with the one of the experimental DEA apparatus. 160 140 120 V. IMAGES AND RESULTS WITH SI-STRIPS DETECTOR 100 80 60 40 0 50 100 150 200 250 300 350 Pixel # Fig. 4.a: Example of average contrast profiles evaluated on images. These are DSA profile for the phantom 4. Different curves represent profiles obtained with a different iodine concentration (from 185 mgI/ml to 20 mgI/ml). Same graphs were calculated for every block obtaining similar shapes. Because of the large time required to scan an image with the thin detectors (300 m) we have acquired only a reduced number of catheters. Figure 6 shows subtracted images of the only first tree vessels filled with growing concentrations of iodine. Relative intensity 300 0.0 -0.2 300 um pixel Profile (Grey Level) 2 Frequency (lp/mm) -0.4 200 -0.6 -0.8 Pixels -1.0 100 0 100 200 300 0 0 20 40 60 80 100 um pixel 100 120 Fig. 6.a: The Phantom 1 filled with 185 mgI/ml. On the left the intensity profile. [3] 300 100 um pixel [4] 200 [5] 100 [6] 0 0 20 40 60 80 300 um pixel 100 120 140 [7] Fig. 6.b: The Phantom 1 filled with 95 mgI/ml. 0 20 40 60 80 100 120 140 [8] 300 300 100 um pixel [9] 200 200 100 100 [10] 0 0 0 20 40 60 80 300 um pixel 100 120 [11] 140 Fig. 6.c: The Phantom 1 filled with 20 mgI/ml. [12] VI. DISCUSSION AND CONCLUSIONS Fig. 4 shows the average profiles for the block labeled phantom 4 when iodine concentration decreases. The DEA system can visualize catheters much smaller than those seen by the conventional angiograph. Smaller catheter is never seen by DSA. But this may be related to its less spatial resolution (Fig. 5). The more important thing is the better capability of DSA to distinguish catheters filled with very small concentration of iodine. Besides, noise in CCD images is much higher then in DSA ones. This is due to various elements: a) the lower photon fluence and consequently the worst statistics, b) the mechanical movements required for DEA, b) the X-rays energy shift in the coronal direction, d) the post-processing algorithm implemented on DSA are not yet well optimized. In Si-strips images, the low fluence feels even more but images are, in complex, less noisy for respect to the CCD detector. We believe that a more accurate filtering process and the use of powerful RX tube may overcome the major problems of this new technique. VII. REFERENCES [1] [2] UNSCEAR , Report of the United Nations Scientific Committee on the Effects of Atomic Radiation to the General Assembly, Annex D – Medical Radiation Exposures (2000) E. Vano, J. Goicolea, C. Galvan, L. Gonzalez, L. Meiggs, JI Ten, C. Macay, Skin radiation injuries in patients following repeated coronary angioplasty procedures, Br.J.Radiol. 74(887), pp. 1023-1031, (2001) [13] [14] [15] [16] [17] [18] G. Baldazzi, I. Corazza, P.L. Rossi, G. Testoni, T. Bernardi, R. Zannoli, In vivo effectiveness of gadolinium filter in paediatric cardiac angiography in terms of image quality and radiation exposure, Phys. Medica XVIII(3), pp. 109-114, (2002) M.S. Van Lysel, Optimization of beam parameters for dual-energy digital substraction angiography, Med. Phys. 21(2), pp. 219-226 (1994) A.C. Thompson, E. Rubenstein, H. D. Zeman, R. Hofstadter, J. N. Otis, J. C. Giacomini, H. J. Gordon, G. S. Brown, W. Thomlinson, R. S. Kernoff, Coronary angiography using synchrotron radiation, Rev. Sci. Instrum. 60, pp. 1622-1627, (1989) C. Gary, M. Piestrup, D. Boyers, C. Pincus, R. Pantell, G. Rothbart, Noninvasive digital energy subtraction angiography with a channelingradiation x-ray source, Med. Phys. 20, pp. 1527-1535 (1993) N. Nariyama, S. Tanaka, Y. Nakane, H. Nakashima, H. Hirayama, S. Ban, Y. Namito, K. Hyodo, T. Takeda, Comparison of in-phantom dose distributions for coronary angiography using an x-ray machine and synchrotron radiation, Med. Phys. 28(1), pp. 16-21 (2001) H. Elleuame et al., First human transvenous coronary angiography at the European Synchrotron Radiation Facility, Phys. Med. Biol. 45(9), pp. 39-43 F. Esteve et al., Coronary angiography with synchrotron X-ray source on pigs after iodine or gadolinium intravenous injection, Acad. Radiol., 9 Suppl.1, (May 2002) T. Yamashita, S. Kawashima, M. Ozaki, M. Namiki, T. Hirase, N. Inoue, K. Hirata, K. Umetani, K. Sugimura, M. Yokoyama, Mouse coronary Angiography using synchrotron radiation microangiography, Circul. 105, pp. 3-4, (2002) . S. Ohtsuka, Y. Sugishita, T. Takeda, Y. Itai, J. Tada, K. Hyodo, M. Ando, Dynamic Intravenous Coronary Arteriography Using 2D monochromatic Synchrotron Radiation, Br. J. Radiol., 72(1999) , pp. 24-28 G. Baldazzi, D. Bollini, E. Celesti, M. Gambaccini, M. Gombia, C. Labanti, A. Taibi, A. Tuffanelli, First results with a novel X-ray Source for Dual Energy Angiography, IEEE NSS MIC 2001 Conference Record, San Diego (Ca), November 2001. G.Baldazzi, D.Bollini, A.E.Cabal Rodriguez, W.Dabrowski, A.Diaz Garcia, M.Gambaccini, P.Giubellino, M.Gombia, P.Grybos, M.Idzik, A.Marzari-Chiesa, L.M.Montano Zetina, F.Prino, L.Ramello, M.Sitta, K.Swientek, A.Taibi, A.Tuffanelli, R.Wheadon, P.Wiacek, X-ray imaging with a silicon microstrip detector coupled to the RX64 ASIC, Nuclear Instruments and Methods in Physics Research A 509 (2003)315-320. P.Rato Mendes, M.C.Abreu, G.Baldazzi, D.Bollini, A.E.Cabal Rodriguez, W.Dabrowski, A.Diaz Garcia, M.Gambaccini, P.Giubellino, M.Gombia, P.Grybos, M.Idzik, A.Marzari-Chiesa, L.M.Montano, F.Prino, L.Ramello, S.Rodrigues, M.Sitta, P.Sousa, K.Swientek, A.Taibi, A.Tuffanelli, R.Wheadon, P.Wiacek, Silicon strip detectors for twodimensional soft X-ray imaging at normal incidence, Nuclear Instruments and Methods in Physics Research A 509 (2003) 333-339. A. Tuffanelli, A. Taibi, G. Baldazzi, D. Bollini, M. Gombia, L. Ramello, M.Gambaccini, Novel x-ray source for dual-energy subtraction angiography, SPIE Medical Imaging Conference Record 2002, San Diego (Ca), February 2002 A. Sarnelli, A. Taibi, A. Tuffanelli, G. Baldazzi, D. Bollini, A.E. Cabal Rodriguez, M. Gorbia, F. Prino, L. Ramello, E. Tomassi, M. Gambaccini, K-edge digital subtraction imaging based on a dichromatic and compact x-ray source, Phys. Med. Biol. 49, pp. 3291-3305 (2004) A. Tuffanelli, S. Fabbri, A. Sarnelli, A. Taibi, M. Gambaccini, Evaluation of a dichromatic X-ray source for dual-energy imaging in mammography, Nucl. Instr. And Meth. A, 489, pp. 209-218 (2002) M. Ohler, M. Sanchez del Rio, A. Tuffanelli, M. Gambaccini, A. Taibi, A. Fantini, G. Pareschi, X-ray topographic determination of the granular structure in a graphite mosaic crystal: a three-dimensional reconstruction, J. Appl. Cryst., 33, pp.1023-1030 (2000)