Download CT Scan Parameters and Radiation Dose

Survey
yes no Was this document useful for you?
   Thank you for your participation!

* Your assessment is very important for improving the workof artificial intelligence, which forms the content of this project

Document related concepts

Medical imaging wikipedia , lookup

X-ray wikipedia , lookup

Radiation therapy wikipedia , lookup

Positron emission tomography wikipedia , lookup

Center for Radiological Research wikipedia , lookup

Neutron capture therapy of cancer wikipedia , lookup

Industrial radiography wikipedia , lookup

Radiosurgery wikipedia , lookup

Nuclear medicine wikipedia , lookup

Backscatter X-ray wikipedia , lookup

Radiation burn wikipedia , lookup

Sievert wikipedia , lookup

Image-guided radiation therapy wikipedia , lookup

Fluoroscopy wikipedia , lookup

Transcript
CT Scan Parameters and Radiation
Dose: Practical Advice for
Radiologists
Siva P. Raman, MD, Mahadevappa Mahesh, MS, PhD,
Robert V. Blasko, BS, RT(R)(CT), Elliot K. Fishman, MD
Although there has been increasing recognition of the importance of reducing radiation dose when performing
multidetector CT examinations, the increasing complexity of CT scanner technology, as well as confusion
about the importance of many different CT scan parameters, has served as an impediment to radiologists
seeking to create lower dose protocols. The authors seek to guide radiologists through the manipulation of 8
fundamental CT scan parameters that can be altered or optimized to reduce patient radiation dose, including
detector configuration, tube current, tube potential, reconstruction algorithm, patient positioning, scan range,
reconstructed slice thickness, and pitch. Although there is always an inherent trade-off between image quality
or noise and patient radiation dose, in many cases, a reasoned manipulation of these 8 parameters can allow the
safer imaging of patients (with lower dose) while preserving diagnostic image quality.
Key Words: Radiation dose, automated tube current modulation, low kVp, pitch, iterative reconstruction
J Am Coll Radiol 2013;10:840-846. Copyright © 2013 American College of Radiology
INTRODUCTION
Over the past several years, there has been increasing
recognition of the importance of reducing radiation dose
in radiologic studies, particularly with regard to multidetector CT. The growth in the volume of multidetector
CT studies performed in the United States, the increasing use of CT in susceptible populations (including children), and growing concerns on the part of the general
public with regard to radiation exposure have provided
an impetus for performing these studies with the least
possible radiation dose [1]. It is now believed that as
many as 0.4% of all malignancies in the United States
can be traced back to radiation from CT studies performed between 1991 and 1996 and that radiation from
CT studies currently being performed may ultimately
account for 1.5% to 2% of all cancers in the future [2].
Unfortunately, despite this growing awareness among
radiologists, the introduction of each new generation of
CT scanners has resulted in scanning protocols that are
increasingly complex and difficult to manipulate for a
radiologist whose primary expertise lies in clinical imaging rather than medical physics. When facing increasing
criticism regarding CT dose, the best way to tackle the
issue is to understand all the factors and parameters that
Department of Radiology, Johns Hopkins University, Baltimore, Maryland.
Corresponding author and reprints: Siva P. Raman, MD, Johns Hopkins
University, Department of Radiology, JHOC 3251, 601 N Caroline Street,
Baltimore, MD 21287; e-mail: [email protected].
can affect radiation dose and image quality and examine
how these can be altered to reduce dose.
In this review, we seek to provide radiologists with a
simple approach to the manipulation of 8 CT parameters
that can be changed by the radiologist, while designing or
altering scan protocols, to reduce patient radiation dose:
(1) detector configuration, (2) tube current, (3) tube
potential (kVp), (4) reconstruction algorithm, (5) patient positioning, (6) scan range, (7) reconstructed slice
thickness, and (8) pitch.
DETECTOR CONFIGURATION
Detector configuration is a term encompassing the number of data channels being used in the z axis and the
“effective detector thickness” of each data channel. For
example, a detector configuration of 64 ⫻ 0.5 mm would
suggest the use of 64 data channels in the z-axis, each of
which has an effective thickness of 0.5 mm. Notably, the
effective detector thickness represents the smallest possible reconstructed slice thickness: if a study is acquired
with an effective detector thickness of 0.5 mm, images
cannot be reconstructed to any smaller interval (such as
0.25 mm). Both the number of channels used and the
effective detector thickness can be varied depending on
how many channels in a detector array are used, which
channels in a detector array are used, and the manner in
which different channels are combined [3]. Different scanners from different manufacturers have different included
840
1546-1440/13/$36.00
●
© 2013 American College of Radiology
http://dx.doi.org/10.1016/j.jacr.2013.05.032
Raman et al/CT Scan Parameters and Radiation Dose 841
detector channels, and the detector configurations available
on different equipment can vary widely [4].
The product of the detector configuration (the number of data channels multiplied by the effective detector
row thickness) is equivalent to the beam collimation. The
physical manifestation of the beam collimation is the
x-ray beam size as defined at the gantry isocenter [5]. In
the example given previously, with a detector configuration of 64 ⫻ 0.5 mm, the beam collimation would be 32
mm. This has a critical impact on patient dose because
the x-ray beam is actually slightly wider than the beam
collimation as a result of a penumbra effect (also referred
to as “overbeaming”), resulting in a small amount of
“wasted” radiation dose with each rotation [6]. As a result, if a smaller beam collimation (4 ⫻ 0.5 mm) is used,
the relative wasted dose is relatively high compared with
the primary beam size. On the other hand, with a larger
64 ⫻ 0.5 mm collimation, the wasted dose is relatively
low compared with the primary beam size. However, it
should be noted that with the newest CT scanners, the
penumbra effect is much less apparent compared with
earlier generations of spiral CT technology because the
extent of the overbeaming or penumbra effect is less
pronounced (relative to the beam width). Therefore, the
need to obtain a smaller effective detector thickness to
improve spatial resolution in the z-axis must be balanced
against the increased radiation dose associated with such
a detector configuration, as well as longer scan times
[3,5].
The detector configuration should be determined on
the basis of the type of study performed, the necessary
slice thickness for multiplanar reformations (MPRs), and
the need for 3-D images. For routine acquisitions, without the need for thin-section MPR images or 3-D imaging, an extremely thin effective detector thickness is
unnecessary. In such cases, if thick-section 5-mm images
in the axial, coronal, and sagittal planes are sufficient for
radiologist review, the effective detector thickness can be
widened to 1.25 or 1.5 mm (rather than 0.75 or 0.5
mm), without a marked impact on image quality (either
for the axial images or MPR images). However, if thinsection MPRs or high-resolution 3-D images (with increased spatial resolution in the z-axis) are required, a
thin effective detector thickness (0.5-0.75 mm) may be
necessary, although there is an associated dose penalty.
most applications. This computer software, which has
various names depending on the specific vendor (ie,
CARE Dose 4D on Siemens, Dose-Right on Phillips,
Auto mA/Smart mA on GE, and SUREExposure 3D on
Toshiba), automatically increases the mAs in those parts
of the body with the greatest attenuation and decreases
the mAs in those parts of the body with lower attenuation. As a result, the software usually increases the mAs in
the shoulder and hips (which have relatively more attenuation) and decreases the mAs in the abdomen and thorax (which have relatively less attenuation) [1,7]. In most
cases, automated tube current modulation reduces patient dose and minimizes photon starvation artifacts.
Automated tube current modulation works on different principles depending on the specific vendor, although a combination of 4 different strategies is generally
used: Patient size modulation varies the mAs on the basis
of a global evaluation of the overall size of the patient as
seen on the scout radiograph. Z-axis modulation changes
the mAs constantly along the z-axis of the patient depending on the patient attenuation at each point, as
determined using the scout image. Angular (x, y) modulation changes the mAs as the x-ray tube rotates 360°
around the patient to account for different attenuations
in different projections of the x-ray beam [1]. Combined
x, y, and z modulation adjusts the mAs in all 3 axes on the
basis of the patient’s attenuation [8]. The user must define a minimum acceptable image quality, the definition
of which will vary depending on the specific vendor and
system. For example, GE users must specify a noise index
value that is maintained as a constant on every image in a
series, while Siemens users must specify an image quality
reference mAs, the mAs that would be used for an average-sized patient [1].
In almost all cases, automated tube current modulation
should be enabled and used by all users, although some
significant caveats and exceptions must be kept in mind:
●
●
TUBE CURRENT
Increases in tube current or the product of tube current
and scan time (mAs) result in improved image quality,
decreased image noise, and increased patient dose. In
general, the relationship between tube current and patient dose is essentially linear, with increases in mAs
resulting in a comparable percentage increase in patient
dose [3]. Although tube current can be manually controlled, most operators use automated tube current modulation (also known as automated exposure control) for
●
Small or pediatric patients, when incorrectly positioned
or improperly centered within the CT gantry, can result
in excessively noisy images with artificially low mAs. Not
only must technologists be properly taught to correctly
position patients, but it may also be helpful to modify the
noise index or increase the reference mAs for smaller
patients (while still capping the maximum mAs to avoid
inadvertently high radiation doses) [8,9].
Many institutions do not use automated tube current
modulation when performing scans of the head, particularly in pediatric patients, because of the difficulty in
correctly placing the head in the isocenter of the gantry. A
small deviation of the head from the isocenter can result
in a significant increase in image noise.
Patients’ arms should be out of the field of view for
both scout-image and axial-image acquisition. Any
discrepancy in the position of the arms between the
scout and axial images can adversely affect the algorithm’s function and theoretically increase dose. If
842 Journal of the American College of Radiology/ Vol. 10 No. 11 November 2013
lips), which are relevant to any discussion of tube current.
These two essentially synonymous terms, defined as tube
current ⫻ rotation time/pitch, are used to convey the
concept that as the pitch increases on these two vendors’ scanners (with resultant increases in noise), the
scanner software will automatically increase the tube
current proportionally to compensate for the increased pitch. As long as the effective mAs or mAs per
slice remains constant, the noise level (and dose) will
remain unchanged [11].
Kvp
Fig 1. CT colonoscopic prone image acquired at a fixed
low 50-mAs protocol.
●
●
●
●
neck and chest CT studies are to be performed at the
same time, either two separate acquisitions should be
performed (with the arms out of the field of view for
each), or a single acquisition with the arms up by the
neck should be used [8].
Only some vendors’ software can correctly account for
metallic orthopedic hardware. Users should check
with their vendors as to whether their specific software
is able take this into account. If not, the modulation
software could theoretically increase patient dose [8].
In such cases, automatic tube current modulation
should be turned off [10].
The same noise levels may not be necessary for every
scan. Each type of study should be individually appraised, and the appropriate noise level should be determined. Moreover, for extremely low dose studies
(such as screening chest CT), it may be easier to use a
low-fixed-mAs protocol (Fig. 1) [8].
In extremely large or obese patients, there may be a
conflict between the specified noise index (or reference
mAs) and the maximum allowed mAs because the
maximum mAs may not be sufficient to provide diagnostic images. In such cases, the maximum mAs may
need to be increased or a higher kVp used [9]. Moreover, extreme care should be taken when using tube
current modulation with obese patients, particularly
on the latest generation of scanners that can produce
extremely high tube currents. In these cases, the tube
current must be appropriately limited or “capped,” or
else patient doses could increase markedly [10].
Pediatric modulation settings should not be used for
adults, and vice versa, because some scanner software
packages use separate modulation curves for children
and adults. Such use may increase doses for children or
result in inadequate doses for adults.
In addition, users should be aware of the terms effective
mAs (used by Siemens) and mAs per slice (used by Phil-
Reducing kVp can be an effective means of reducing the
radiation dose imparted during an examination. As a
general rule of thumb, the radiation dose changes with
the square of kVp, and a reduction in kVp from 120 to
100 reduces radiation dose by 33%, while a further reduction to 80 kVp can reduce dose by 65% [12,13]. As a
further advantage, particularly in angiographic studies, in
which vascular contrast is critical, reductions in kVp result
in an increased attenuation of iodine (even with a constant
dose of contrast) as a result of the photoelectric effect and
approaching the k-edge of iodine, potentially improving
contrast-to-noise ratios. Nevertheless, unlike reductions in
mAs, which have a linear and relatively predictable effect on
image noise and contrast-to-noise ratios, decreases in kVp
can result in nonlinear, exponential increases in image noise,
often necessitating a concomitant increase in mAs to preserve image quality. Regardless, several studies evaluating
the use of low kVp in a number of different protocols (angiography, venous phase studies, cardiac and coronary studies, etc) have shown that such reductions can markedly
reduce radiation doses, while still preserving acceptable image quality [12,14,15].
It is now widely accepted that low-kVp protocols
(100, 80, and even 70 kVp) are most useful in thin,
nonobese patients, particularly in vascular or angiographic studies, in which improvements in the attenuation or conspicuity of iodine can be helpful in terms of
image contrast-to-noise ratios (Fig. 2) [16]. Most scanners offer no tools to correctly select an appropriate kVp,
and the decision to decrease kVp must be made at the
discretion of the radiologist, taking into account the type
of study being performed (ie, arterial or venous phase
acquisition), the portions of the body being imaged, and
a global assessment of patient body habitus. Although
some sites have attempted to use algorithms that base the
decision on a patient’s body mass index, this may often be
unsatisfactory because it is the distribution of the patient’s body fat that has the ultimate impact on image
quality. Data regarding the correct kVp selection in any
given patient are still evolving, and each practice must
determine its comfort level with low-kVp techniques by
slowly instituting such changes over time. Although the
Society of Cardiovascular Computed Tomography has
published recommendations regarding kVp levels in patients undergoing cardiac CT, including a suggested kVp
Raman et al/CT Scan Parameters and Radiation Dose 843
Fig 2. Pediatric chest CT study acquired at 80 kVp. A
low–tube voltage technique may be most useful in thin or
pediatric patients.
of 100 in patients weighing ⱕ90 kg, no similar guidelines have been published for conventional body and
neurologic imaging applications [17].
Notably, the increase in image noise associated with
lower kVp techniques must be balanced by an increase in
the tube current. The reference mAs (on Siemens scanners) or noise index (on GE scanners) must be appropriately altered to increase tube current, but the tube current
should not be raised too much because the dose savings
from a lower kVp would be canceled out. Instead, the
tube current must be mildly increased, preserving image
quality (with perhaps minimal degradation), while still
ensuring some dose savings [16].
RECONSTRUCTION ALGORITHM
Traditionally, CT reconstruction algorithms have used
filtered back projection (FBP). The mathematical algorithms underlying FBP assume a “nonexact” relationship
between the projection data acquired at the CT scanner
and the data displayed in the final image. As a result, FBP
removes very little noise from the final image, is limited
in the use of data with lower contrast-to-noise ratios, and
generally requires studies to be performed with larger
radiation doses to be of diagnostic quality [1,18].
As a result, each of the major CT vendors now offers its
own proprietary iterative reconstruction algorithms as a replacement for FBP. Iterative reconstruction, now increasingly
feasible because of better algorithms and improvements in
hardware, involves the use of computationally intensive iterative “correction loops,” which repeatedly compare the reconstructed data with an “ideal” image, until an acceptable final
image is created [1]. Accordingly, these iterative techniques
allow the reconstruction of images with improved image
quality, even in cases in which the imparted radiation dose
and contrast-to-noise ratios are low. A number of studies
using each of the different vendors’ software packages have
concluded that iterative reconstruction is a viable option in
terms of both reducing radiation dose and improving image
quality [19-22]. For example, a study by Kaza et al [23]
found that iterative reconstruction techniques could lower
radiation dose by up to 30% in CT enterographic studies,
while still maintaining acceptable image quality.
Overall, when iterative reconstruction is available on a
group’s CT scanner, there is little downside to its implementation over more traditional FBP techniques in
terms of either diagnostic quality or radiation dose.
However, regardless of which vendor’s iterative reconstruction algorithm is used, there is no simple way to
determine the extent to which dose can subsequently be
reduced. In general, once iterative reconstruction is instituted, dose should gradually be decreased over time (via
changes in the acceptable noise index or reference mAs
on the system’s automatic exposure control software)
until reaching a point at which the radiologist no longer
feels comfortable with the image quality. Importantly,
the images acquired with iterative reconstruction can
have a smoothed, “artificial” appearance compared to
FBP images, necessitating some acclimatization on the
part of the interpreting radiologist.
PATIENT POSITIONING
In a study by Li et al [24] in 2006, 95% of patients in the
series were found to be improperly centered within the
scanner gantry by the CT technologist. Although this
may seem to be of only minimal importance, improper
positioning can have a significant impact on both image
noise and patient surface dose. Habibzadeh et al [25]
found improper centering of phantoms in the scanner
gantry to be associated with an average surface dose increase of 23% and an image noise increase of 7%. Abnormal positioning in both the horizontal and vertical axes
can have a similar impact, as a study by Kaasalainen et al
[26] using child-sized phantoms found an increase in the
breast surface dose of 16% and in the thyroid gland of
24% for phantoms improperly positioned in the vertical
axis. Just as important, small variations in patient positioning can have a marked effect on image noise, as
vertical displacement of a patient by just 6 cm can result
in an increase in noise of 22%. Needless to say, this can
have a dramatic secondary effect on radiation dose if
automatic exposure control is used because radiation
doses can be markedly increased by the software to compensate for this increased noise [26-28]. Although patient positioning must be carefully monitored in all
patients, extra attention should be paid to this issue in
pediatric patients and patients with smaller body habitus
because small errors in positioning such patients are more
likely to have a significant impact on both noise and
dose. Moreover, although correct patient centering is
844 Journal of the American College of Radiology/ Vol. 10 No. 11 November 2013
an important factor regardless of the type of scanner
being used, it is particularly important when using dualsource CT scanners. Given the scanner geometry of dualsource mode (with the second source typically having a
smaller field of view), the dose penalty from centering the
patient away from the scanner isocenter can be greater
than with conventional single-source acquisitions.
The primary cause for this association between improper patient centering and increased patient radiation
dose is the use of bowtie filters in modern CT scanners.
These filters normally compensate for patient attenuation during tube rotation, resulting in increased x-ray
intensity to the thickest parts of the patient (ie, the center
of the patient) and decreased intensity to the thinnest
parts of a patient (ie, the patient surface). The bowtie
filter functions under the assumption that the patient is
correctly centered within the gantry [26]. In optimal
circumstances, the bowtie filter can reduce patient dose
by up to 50% [24]. However, when the patient is improperly positioned, the mathematical assumptions underlying this filter break down, and doses to the surface of
the body increase.
In general, incorrect positioning in the horizontal axis
should not be a problem in the vast majority of patients.
Virtually every recent scanner has a laser guidance system
that should allow the patient to be correctly placed at the
isocenter of the scanner, and all technologists should be
trained to correctly use this guidance mechanism. However, correct positioning in the vertical axis can be more
tricky: in patients who are markedly scoliotic or unable to
lie completely flat in the gantry, portions of the patient,
by necessity, will inevitably be positioned outside of the
isocenter with vertical displacement. In addition, technologists must be very careful when placing pillows under the patient, as excessive elevation of the patient’s head
with pillows may cause portions of the patient’s upper
body and head to be above the scanner’s isocenter. Unfortunately, an error in positioning may not be immediately apparent to either the radiologist or the technologist
because there may not be a dramatic change in the volume CT dose index: increased surface dose to portions of
the body above the scanner’s isocenter may be balanced
by decreased surface dose to portions of the body below
the isocenter, and the volume CT dose index may seem
unchanged, although portions of the body have received
markedly increased skin dose. As a result, this error must
be prevented through active training of technologists,
who must be instructed not only to correctly position the
patients but also to turn off automatic tube current modulation in any case in which there is difficulty in properly
positioning the patient at the isocenter of the scanner.
SCAN RANGE
For many CT applications, significant reductions in scan
range may be neither possible nor desirable. Nevertheless, the scan range should be reduced to the needed
Fig 3. CT coronary artery study with a minimized scan
range encompassing only the necessary aspect of the
thorax in the z-axis.
minimum for any examination, particularly when imaging structures such as the heart, for which an increased
scan range is unnecessary (Fig. 3). Although cardiac studies are certainly the most obvious applications in which a
limited scan length can be used, a small scan range may
be possible in many traditional body imaging applications as well: Patel et al [29] found that using a small scan
range (from the top of the aortic arch to the bottom of the
heart) in CT pulmonary angiographic studies can allow
diagnosis of pulmonary embolism without any loss of
sensitivity but with a reduction in radiation dose of 48%.
Unfortunately, even if the scan range is kept to the
absolute minimum, helical CT requires the acquisition
of raw data on either end of the scan length, outside of the
intended scan range, to acquire sufficient projection data.
This can particularly be a problem with higher pitches
and can result in additional dose as a result of wasted dose
at the margins of the scan range. However, this effect can
be mitigated with the use of dynamic z-collimation. Unlike a fixed collimator, which is fully open throughout the
scan range, the dynamic z-collimator opens only that
portion of the collimator entering the scan range (while
at the beginning of the scan) and closes the proximal edge
of the collimator while leaving the scan range, thus minimizing dose outside the intended scan range and resulting in a more rectangular dose profile. Dose savings with
the use of dynamic z-collimation can be incremental,
although the exact extent of savings will depend on the
CT pitch [30,31].
RECONSTRUCTED SLICE THICKNESS
As the reconstructed slice thickness decreases, the number of photons within each voxel also decreases, resulting
in increased image noise. To maintain constant noise
levels within an image with a smaller slice thickness, the
radiation dose must be consequently increased [3]. With
the introduction of automatic tube current modulation,
the interaction between the user-specified level of acceptable noise, reconstructed slice thickness, and radiation
dose can be quite complex and will undoubtedly vary
depending on the CT scanner and the specific automatic
tube current modulation software being used [32]. In
Raman et al/CT Scan Parameters and Radiation Dose 845
general, the larger the reconstructed slice thickness used,
the lower the patient dose. It is worth repeating, however,
that the reconstructed slice thickness cannot be smaller
than the acquired slice thickness.
PITCH
Pitch in the multidetector, spiral CT era is defined as
table travel per rotation divided by beam collimation.
Pitch ⬍1 suggests overlap between adjacent acquisitions, pitch ⬎1 implies gaps between adjacent acquisitions, and pitch of 1 suggests that acquisitions are
contiguous, with neither overlap nor gaps [7]. A smaller
pitch, with increased overlap of anatomy and increased
sampling at each location, results in an increased radiation dose. Alternatively, a larger pitch implies gaps in the
anatomy and hence lower radiation dose. As a result, if all
other parameters are unchanged, increasing pitch reduces
radiation dose in a linear fashion [7,33]. Low-pitch technique is associated with less image noise, fewer artifacts,
and improved signal-to-noise and contrast-to-noise ratios [34]. For routine body CT protocols, pitch ⬎1 is
generally acceptable with no compromise to image quality. However, using higher pitches (pitch ⬎1.5) can result in interpolation artifacts and image noise that is
typically unacceptable. For most cardiac CT applications, pitch values ⬍0.5 are routinely used, resulting in
higher doses in such studies. These lower pitch values are
used to overcome motion artifacts and are also stipulated
by the reconstruction algorithms in use. As mentioned
earlier, one major caveat to this discussion of the normal
relationship between pitch and radiation dose is the fact
that certain scanners use the concept of effective mAs or
mAs per slice [33]. In these cases, the effect of pitch on
dose is negated by proportional increases in tube current
to maintain similar image noise.
Although higher pitch values have traditionally been
thought to be associated with poor image quality, this is
no longer true with the latest dual-source CT technology.
A study by Apfaltrer et al [35] using pitch values ⬎3 on a
dual-source scanner for vascular studies suggested that
image quality was more than acceptable, with radiation
dose savings as high as 45% to 50%. The newest dualsource scanners offer high-pitch or FLASH mode acquisitions with pitch values ⬎3, consequently decreasing
scan times precipitously and potentially reducing radiation doses significantly. These high-pitch modes markedly reduce scan times and eliminate many motionrelated artifacts, particularly those previously seen with
cardiac imaging [36]. However, it should be noted that
there are significant caveats that should induce some
caution before using high-pitch or FLASH acquisitions
on every case: Although the data suggest that FLASH
acquisitions offer a significant dose reduction for vascular
studies in the chest, for which retrospective or prospective gating would otherwise be used to prevent motionrelated artifacts, it is unclear whether there is actually any
true dose savings in nonvascular studies in the chest or
abdomen. In particular, in such cases, some systems will
attempt to maintain a constant effective mAs (Siemens)
or mAs per slice (GE) (tube current ⫻ rotation time/
pitch) and will consequently increase the tube current to
compensate for the increased pitch. In addition, in patients who are large, the system may attempt to increase
the tube current but may be precluded from doing so as a
result of a capped or limited mAs, and image noise may
actually rise. Accordingly, in nonvascular studies in
which scan speed is not as important, using a high-pitch
technique may offer little dose savings and potentially
worse image quality. That said, high-pitch techniques on
these scanners still offer significant value in nonvascular
studies in children, in whom scan speed is critical.
CONCLUSIONS
As both the general public and clinicians have become
increasingly aware of the concerns associated with multidetector CT–related radiation dose, it has become critical for radiologists to better understand the CT scanner
technology they work with. In particular, a detailed understanding of a few basic CT scan parameters is essential, and
knowledge of how to manipulate these parameters to produce diagnostic images at lower doses is critical for safe
imaging.
TAKE-HOME POINTS
●
●
●
●
●
●
Several multidetector CT parameters can be changed
to reduce radiation dose, including detector configuration, tube current, kVp, reconstruction algorithm,
patient positioning, scan range, reconstructed slice
thickness, and pitch.
Although all users should generally use automated
tube current modulation, users must be careful about
using this software in obese patients, children, and
patients with metallic hardware.
Low-kVp protocols may be most useful in thin, nonobese patients, particularly in vascular or angiographic
studies.
Iterative reconstruction techniques allow the reconstruction of images with improved image quality, even
in cases in which the imparted radiation dose and
contrast-to-noise ratios are low.
The scan range should be reduced to the needed minimum for any examination.
Incorrect positioning of patients in the scanner gantry can
be associated with significant radiation dose penalties.
REFERENCES
1. Raman SP, Johnson PT, Deshmukh S, et al. CT dose reduction applications: available tools on the latest generation of CT scanners. J Am Coll
Radiol 2013;10:37-41.
2. Brenner DJ, Hall EJ. Computed tomography—an increasing source of
radiation exposure. N Engl J Med 2007;357:2277-84.
846 Journal of the American College of Radiology/ Vol. 10 No. 11 November 2013
3. Saini S. Multi-detector row CT: principles and practice for abdominal
applications. Radiology 2004;233:323-7.
21. Martinsen AC, Saether HK, Hol PK, et al. Iterative reconstruction reduces abdominal CT dose. Eur J Radiol 2012;81:1483-7.
4. Mahesh M. CT physics: the basics of multi-detector physics. Philadelphia: Lippincott Williams & Wilkins; 2009.
22. Leipsic J, Labounty TM, Heilbron B, et al. Estimated radiation dose
reduction using adaptive statistical iterative reconstruction in coronary
CT angiography: the ERASIR study. AJR Am J Roentgenol 2010;195:
655-60.
5. Verdun FR, Gutierrez D, Schnyder P, et al. CT dose optimization when
changing to CT multi-detector row technology. Curr Probl Diagn Radiol
2007;36:176-84.
6. Hamberg LM, Rhea JT, Hunter GJ, et al. Multi-detector row CT: radiation dose characteristics. Radiology 2003;226:762-72.
7. Goldman L. Principles of CT: multislice CT. J Nucl Med Technol 2008;
36:57-68.
8. Singh S, Kalra MK, Thrall JH, et al. Automatic exposure control in CT:
applications and limitations. J Am Coll Radiol 2011;8:446-9.
9. Kalra M, Maher M, Toth T, et al. Techniques and applications of automatic tube current modulation for CT. Radiology 2004;233:649-57.
23. Kaza RK, Platt JF, Al-Hawary MM, et al. CT enterography at 80 kVp
with adaptive statistical iterative reconstruction versus at 120 kVp with
standard reconstruction: image quality, diagnostic adequacy, and dose
reduction. AJR Am J Roentgenol 2012;198:1084-92.
24. Li J, Udayasankar UK, Toth TL, et al. Automatic patient centering for
MDCT: effect on radiation dose. AJR Am J Roentgenol 2007;188:547-52.
25. Habibzadeh MA, Ay MR, Asl AR, et al. Impact of miscentering on patient
dose and image noise in x-ray CT imaging: phantom and clinical studies.
Phys Med 2012;28:191-9.
10. Mahesh M, Fishman EK. CT dose reduction strategy: to modulate dose or
not in certain patients? J Am Coll Radiol 2012;9:931-2.
26. Kaasalainen T, Palmu K, Lampinen A, et al. Effect of vertical positioning
on organ dose, image noise and contrast in pediatric chest CT-phantom
study. Pediatr Radiol 2013;43:673-84.
11. Primak AN, McCollough CH, Bruesewitz MR, et al. Relationship between noise, dose, and pitch in cardiac multi-detector row CT. Radiographics 2006;26:1785-94.
27. Toth T, Ge Z, Daly MP. The influence of patient centering on CT dose
and image noise. Med Phys 2007;34:3093-101.
12. Gnannt R, Winklehner A, Eberli D, et al. Automated tube potential
selection for standard chest and abdominal CT in follow-up patients with
testicular cancer: comparison with fixed tube potential. Eur Radiol 2012;
22:1937-45.
13. Paul JF. Individually adapted coronary 64-slice CT angiography based on
precontrast attenuation values, using different kVp and tube current
settings: evaluation of image quality. Int J Cardiovasc Imaging 2011;
27(suppl):53-9.
14. Feuchtner GM, Jodocy D, Klauser A, et al. Radiation dose reduction by
using 100-kV tube voltage in cardiac 64-slice computed tomography: a
comparative study. Eur J Radiol 2010;75:e51-6.
15. Dong F, Davros W, Pozzuto J, et al. Optimization of kilovoltage and tube
current-exposure time product based on abdominal circumference: an
oval phantom study for pediatric abdominal CT. AJR Am J Roentgenol
2012;199:670-6.
16. McCollough CH, Chen GH, Kalender W, et al. Achieving routine submillisievert CT scanning: report from the summit on management of
radiation dose in CT. Radiology 2012;264:567-80.
17. Halliburton SS, Abbara S, Chen MY, et al. SCCT guidelines on radiation
dose and dose-optimization strategies in cardiovascular CT. J Cardiovasc
Comput Tomogr 2011;5:198-224.
18. Fleischmann D, Boas FE. Computed tomography— old ideas and new
technology. Eur Radiol 2011;21:510-7.
28. Gudjonsdottir J, Svensson JR, Campling S, et al. Efficient use of automatic exposure control systems in computed tomography requires correct
patient positioning. Acta Radiol 2009;50:1035-41.
29. Patel H, Coughlin B, LaFrance T, et al, eds.Comparison of full chest CTA
with limited CTA and triple rule-out CTA for PE detection and effective
dose implications. Chicago, Illinois: Radiological Society of North America; 2007.
30. Shirasaka T, Funama Y, Hayashi M, et al. Reduction of the unnecessary
dose from the over-range area with a spiral dynamic z-collimator: comparison of beam pitch and detector coverage with 128-detector row CT.
Radiol Phys Technol 2012;5:53-8.
31. Christner JA, Zavaletta VA, Eusemann CD, et al. Dose reduction in
helical CT: dynamically adjustable z-axis X-ray beam collimation. AJR
Am J Roentgenol 2010;194:W49-55.
32. Kanal K, Stewart B, Kolokythas O, et al. Impact of operator-selected
image noise index and reconstruction slice thickness on patient radiation
dose in 64-MDCT. AJR Am J Roentgenol 2007;189:219-25.
33. Mahesh M, Scatarige J, Cooper J, et al. Dose and pitch relationship for a
particular multislice CT scanner. AJR Am J Roentgenol 2001;177:
1273-5.
34. Hu H, Fox SH. The effect of helical pitch and beam collimation on the
lesion contrast and slice profile in helical CT imaging. Med Phys 1996;
23:1943-54.
19. Desai GS, Uppot RN, Yu EW, et al. Impact of iterative reconstruction on
image quality and radiation dose in multidetector CT of large body size
adults. Eur Radiol 2012;22:1631-40.
35. Apfaltrer P, Hanna EL, Schoepf UJ, et al. Radiation dose and image
quality at high-pitch CT angiography of the aorta: intraindividual and
interindividual comparisons with conventional CT angiography. AJR
Am J Roentgenol 2012;199:1402-9.
20. Korn A, Fenchel M, Bender B, et al. Iterative reconstruction in head CT:
image quality of routine and low-dose protocols in comparison with standard
filtered back-projection. AJNR Am J Neuroradiol 2012;33:218-24.
36. Korn A, Fenchel M, Bender B, et al. High-pitch dual-source CT angiography of supra-aortic arteries: assessment of image quality and radiation
dose. Neuroradiology 2013;55:423-30.