Download Signal Conversion

Survey
yes no Was this document useful for you?
   Thank you for your participation!

* Your assessment is very important for improving the work of artificial intelligence, which forms the content of this project

Document related concepts

Operational amplifier wikipedia , lookup

Amplifier wikipedia , lookup

Regenerative circuit wikipedia , lookup

Battle of the Beams wikipedia , lookup

Switched-mode power supply wikipedia , lookup

Broadcast television systems wikipedia , lookup

Tektronix analog oscilloscopes wikipedia , lookup

Resistive opto-isolator wikipedia , lookup

Oscilloscope wikipedia , lookup

Power electronics wikipedia , lookup

Signal Corps (United States Army) wikipedia , lookup

Rectiverter wikipedia , lookup

Dynamic range compression wikipedia , lookup

ATSC tuner wikipedia , lookup

Phase-locked loop wikipedia , lookup

Mixing console wikipedia , lookup

Time-to-digital converter wikipedia , lookup

Telecommunication wikipedia , lookup

Radio transmitter design wikipedia , lookup

Integrating ADC wikipedia , lookup

Cellular repeater wikipedia , lookup

Oscilloscope types wikipedia , lookup

Valve RF amplifier wikipedia , lookup

Coupon-eligible converter box wikipedia , lookup

Index of electronics articles wikipedia , lookup

Analog television wikipedia , lookup

HD-MAC wikipedia , lookup

Oscilloscope history wikipedia , lookup

High-frequency direction finding wikipedia , lookup

Television standards conversion wikipedia , lookup

Opto-isolator wikipedia , lookup

Analog-to-digital converter wikipedia , lookup

Transcript
Signal Conversion
73
3.4.5 Analog multiplexer
When several signals need to be converted, it is necessary to either provide an A/D
converter for each signal or use an analog multiplexer to direct the various signals
to a single converter. For most biomedical signals, the required conversion rates
are low enough that multiplexing the signals is the appropriate choice.
Common analog multiplexers utilize either JFET or CMOS transistors. Figure
3.18 shows a simple CMOS analog switch circuit. A number of these switches are
connected to a single Vout to make a multiplexer. The switches should operate in
a break-before-make fashion to ensure that two input lines are not shorted together.
Other attributes to be considered include on-resistance, leakage currents, crosstalk,
and settling time.
ø
p-type
Vin
Vout
n-type
ø
Figure 3.18 A simple CMOS analog switch. The basic functional block of a CMOS analog
multiplexer.
3.4.6 Amplifiers
The biomedical signal produced from the sensor or electrode is typically quite
small. As a result, the first step before the A/D conversion process is often amplification. Analog amplification circuits can also provide filtering. Analog filters are
often used to bandlimit the signal prior to sampling to reduce the sampling rate required to satisfy the sampling theorem and to eliminate noise.
General
For common biomedical signals such as the ECG and the EEG, a simple instrumentation amplifier is used. It provides high input impedance and high CMRR
(Common Mode Rejection Ratio). Section 2.4 discusses the instrumentation amplifier and analog filtering in more detail.
Micropower amplifiers