Equipment Specifications for Digital PH Meter
UNSPSC Code:
ECRI Code:
1 Description of Function
Sl
Name
1.1
A pH meter is an electronic instrument used to measure the pH
(acidity or basicity) of a Blood Sample in blood bank.
Technical Bidders
Specs
Deviation
quoted
if any
by bidder
2 Operational Requirements
Sl
Name
2.1
Portable digital system is required.
Technical Bidders
Specs
Deviation
quoted
if any
by bidder
3 Technical Specifications
Sl
Name
3.1
pH range should be 0-14
3.2
Milli volt range should be 0-1999+/- mv
3.3
Resolution 0.01 pH
3.4
Repeatability +/- 0.01pH +/- 1 digit
3.5
Standardization range +/- 2 pH
3.6
Temp. compensation 0-100 degree C
3.7
Display 4 digit LED with automatic polarity
3.8
Polarizing current 10 micro ampere
3.9
Dimensions 240x230x115 mm
3.10
Weight approx 2 kg
Technical Bidders
Specs
Deviation
quoted
if any
by bidder
4 System Configuration Accessories, spares and consumables
Sl
Name
4.1
System as specified-
4.2
Accessories to be supplied –part of glass and reference electrodes,
electrode holder
4.3
All consumables required for installation and standardization of
system to be given free of cost.
Technical Bidders
Specs
Deviation
quoted
if any
by bidder
4.4
System should be quoted with complete accessories and probes to
make it functional as per technical specifications
5 Environmental factors
Sl
Name
5.1
Shall meet IEC-60601-1-2 :2001(Or Equivalent BIS) General
Requirements of Safety for Electromagnetic Compatibility.or should
comply with 89/366/EEC; EMC-directive.
5.2
The unit shall be capable of being stored continuously in ambient
temperature of 0 -50deg C and relative humidity of 15-90%
5.3
Thu unit shall be capable of operating in ambient temperature of 2030 deg C and relative humidity of 80%.
Technical Bidders
Specs
Deviation
quoted
if any
by bidder
6 Power Supply
Sl
Name
6.1
Should work on 220-240V AC as well as power supply. Mains adaptor
to be supplied.
Technical Bidders
Specs
Deviation
quoted
if any
by bidder
7 Standards and Safety
Sl
Name
7.1
Should be FDA , CE,UL or BIS approved product
7.2
Protection level should be certified for IP 67
7.3
Should comply with IEC 61010-1 Safety requirements for electrical
equipment for measurement, control, and laboratory use - Part 1:
General requirements
7.4
Comprehensive warranty for 2 years and 5 years AMC after warranty
7.5
Manufacturer/Supplier should have ISO certification for quality
standards.
Technical Bidders
Specs
Deviation
quoted
if any
by bidder
8 Documentation
Sl
Name
8.1
Certificate of calibration and inspection from factory.
8.2
List of Equipments available for providing calibration and routine
Preventive Maintenance Support. as per manufacturer documentation
in service/technical manual.
8.3
List of important spare parts and accessories with their part number
and costing.
Technical Bidders
Specs
Deviation
quoted
if any
by bidder