Survey
* Your assessment is very important for improving the work of artificial intelligence, which forms the content of this project
* Your assessment is very important for improving the work of artificial intelligence, which forms the content of this project
Prototype Feasibility Assessment Senior Design PO8023 1/18/2008 In order to accurately assess the feasibility of the selected design concepts for team PO8023, Air Muscle Artificial Limb, a prototype was assembled using the parts and procedures listed herein. Design and Build: I. Design & Construction Based on initial concept generation exercises it was determined that it would be beneficial for the team to construct a prototype of the finger design by using a dimension fused deposition modeler (FDM). This rapid prototyping machine used the existing computer aided design (CAD) model to create the finger by laying down ABS plastic in layers until the final design was complete, as shown in figure 1. Figure 1: Prototype finger made from ABS Plastic A custom made Lego palm and forearm, figure 2, was constructed as closely to the final design scale as possible to allow simulation of the prototype finger design with air muscles and verify all motion clearances. Figure 2: Prototype Lego palm and forearm The fingers were wired to air muscles using Berkley Steelon Nylon Coated wires and extension ligaments were fitted using Polypropylene Covered Elastic Cord as shown in figure 3. Figure 3: Assembled finger with tendon cabling and extension ligaments All completed parts were then assembled into the final prototype as shown in figure 4. Figure 4: Final assembled prototype for demonstration II. Parts List Description Quantity Units Air Muscle Artificial Limb Assembly Lego Forearm & Palm Assembly 1 1 Ea. Ea. Index Finger Assembly Common Finger Tip (Distal Phalanx) IF_Section 2 (Middle Phalanx) IF_Section 1_Upper (Proximal Phalanx) IF_Section 1_Lower (Metacarpals) 1 1 1 1 1 Ea. Ea. Ea. Ea. Ea. Middle Finger Assembly Common Finger Tip (Distal Phalanx) MF_Section 2 (Middle Phalanx) MF_Section 1_Upper (Proximal Phalanx) MF_Section 1_Lower (Metacarpals) 1 1 1 1 1 Ea. Ea. Ea. Ea. Ea. Ring Finger Assembly Common Finger Tip (Distal Phalanx) RF_Section 2 (Middle Phalanx) RF_Section 1_Upper (Proximal Phalanx) RF_Section 1_Lower (Metacarpals) 1 1 1 1 1 Ea. Ea. Ea. Ea. Ea. Polypropylene Covered Elastic Cord 3/32" Diameter (50 ft.) Double Pinch Zinc-Pltd Stl Hose & Tube Clamp 1/8" to 5/32" Clamp Diameter Range Berkley Steelon Nylon Coated Wire Berkley Wire Leader Connector Sleeves SPRO Ball Bearing Swivel with Interlock Snap 25 25 30 24 9 Ft. Ea. Ft. Ea. Ea. Air Muscle Assembly 3 Ea. III. Prototype Issues One of the main problems with the prototype was that the tendon cabling could not be run through the palm the same way that it would be on the final design. Despite this limitation of the prototype the team used alternate routing of the cables to achieve motion of the fingers. Another drawback of the prototype was that it was difficult to service, especially during initial assembly and when adding air muscles. One of the main reasons for this is that there was no cable guide to help position the tendon cabling running through the palm to the fingers. The final issue that was solved during the prototype assembly process was that the diameter of the holes for the extension ligaments had to be increased in order to attach the elastic cord. IV. Influences on Final Design Based on the building and testing conducted with the prototype Air Muscle Artificial Limb, the team learned that serviceability must be a main focus on the final design. Increased serviceability in the final design will allow for adjustments to be made as necessary through the final assembly process. The team also decided to use both rubber bands and elastic cord for the return ligaments in order to help return the finger to its rest position. The stiffer elastic cord will begin returning the finger to its initial position while the shorter rubber band will finish the fingers extension. V. Conclusions The prototype was successful in demonstrating the motion clearances in the finger design and the feasibility of the connection points for the tendon cabling that will cause finger motion based on air muscle contraction. Air Muscle: I. Design & Construction Air Muscle Build Process Step 1 Step 2 Step 3 Step 4 Step 5 Step 6 Step 7 Step 8 Step 9 Cut rubber tubing of desired diameter to desired length. Thread a thin metal rod through the rubber tubing. Insert the rubber tubing (containing the metal rod) through the mesh. Remove the metal rod – align the edge of the tube to the edge of the mesh. With a heat knife, cut the mesh where the tubing ends. (Now the mesh and tubing should be concentric and of the same length) Place a metal crimping piece on each end of the air muscle. Insert Eye Hook into one end of the rubber tube Insert the appropriate airside connector in the other end of the air muscle. Crimp both ends of the air muscle, creating a seal between the mesh, tubing, and end fitting Air muscles should always be tested for leaks at the pressure they will be operated at before use. Figure 5: Assembled Air Muscle II. Parts List Description Air Muscle Assembly Mesh – 3/8in PET (neon red) Tubing - Super Soft Latex Rubber Tubber 1/8" ID, 3/8" OD, 1/8" Wall Thickness Nckl-Pltd Brass Push-to-Connect Tube Fitting Expanding Tube Stem for 1/8" Stem OD X 1/4" Tube OD Double pinch zinc-pltd stl hose and tube clamp 13/64" to 9/32" PKG=25 per pack Eye Hook #12 Quantity Units 8 8 ft ft 18 Ea. 36 Ea. 18 Ea. III. Prototype Issues Problems with the air muscles were encountered during life cycle testing. The air side connector failed after running for an average of 1500 cycles. During this testing, the connector would be pulled free from the air muscle. This failure mode was the only failure mode that appeared during air muscle testing. Of all the air side connectors tested, the Nickel Plated Push-to-Connect fitting worked the longest. IV. Influences on Final Design Based on the failure mode of the air muscle, ease of serviceability is the key to the success of the final design. As the air muscles approach failure they can be replaced by new muscles to avoid the disassembly of the air side fitting. On the final design the air muscles will be accessible to make serviceability simple. V. Conclusions The air muscle design is suitable for the final limb design. The air muscles will last around 1500 cycles and be available for service. Improving the air side connection method would increase the cycle life of the muscles and reduce the need for replacement. Controls: I. Design & Construction For the initial prototype testing, the valves and relay board were assembled as they would be for the final product. The valves were connected first utilizing the following instructions. The fasteners on the end of coil and DIN connector combination were loosened, rotated 180°, and re-tightened. The gasket was placed over the manifold and aligned with the three-port holes of the individual station. Each valve was aligned to the manifold station and fastened with two mounting screws. The air supply fitting was fastened to the frontface center through the hole on manifold. Plugs were also placed on the remaining two through-holes on front-face. Plugs were placed on the back-face center through-hole and the remaining two through- holes on back-face were left open for exhaust. Figure 6: Valves with Air Supply Fittings The threads of the needle valve were and valve fittings with then wrapped with thread seal tape. The push-to-connect valve fittings were screwed into the designated Ab/Adduction solenoid port holes and needle valves into assigned flexion solenoid port holes. The appropriate end of the push-to-connect fitting was thread into one of the flexion solenoid needle valves, and then the air muscle feed tube was inserted into the push-to-connect end of the fitting ensuring that the tube is fully sealed. The DIN connector was disassembled by unscrewing the connector from the coil and removing the transparent shell, as shown in Figure 7. Figure 7: DIN Connector Terminal 1 was connected to the relay board (Measurement Computing USB SSR 24) and terminal 2 to ground. As shown in the schematic below, terminal 2 was daisy chained to each of the other DIN connectors with the first or last connector terminated to the ground wire of the power supply. Similarly, the power wire was connected to a positive terminal of the relay board and daisy chained to every other positive terminal. Each negative relay terminal was connected directly to one solenoid valve at terminal 2, such that each relay will control one valve. Relay Board + + 1 Power Supply + 2 + 3 + 4 - - - 1 1 1 + 5 - + 6 - + 7 - + 8 - + 9 - VCC_BAR 2 2 1 2 1 2 1 2 2 Solenoid Valve DIN Connectors 2 2 1 2 1 2 1 2 1 2 1 + 10 - 1 Figure 8: Relay Board and Solenoid Valve Connections + 11 - 12 - - The relay board connections are also shown in Figure 9. Figure 9: Relay Board with Connections to Valves The relay board was also connected to the computer via USB. The code that was used in LabVIEW to control a single flexion air muscle consisted of a simple loop that considered the percentage the finger moved due to one valve cycle (a single valve being turned on and off once). The number of cycles needed to reach the target percentage was then determined. Figure 10 provides a simple diagram of the LabVIEW program logic. Active Valve Delay Deactivate Valve Delay *run loop N times (where N is # of cycles needed to reach target Figure 10: LabVIEW Program Logic II. Parts List Description Quantity Units Valve Assembly Clippard 4-Way 3 Position: Double Solenoid, Closed Center, 12 VDC, 1/8" NPT Clippard 4-Way 3 Position: Double Solenoid, Exhaust Center, 12 VDC, 1/8" NPT Clippard 6 Valve Manifold 1/4" Ports Clippard Needle Valves Legris Needle Valve Fittings Mal Straight Legris Manifold Fittings Mal Straight, 1/4" Tube x 1/4" NPT Male Wires Sub-D25 Slider Plug Electrical Tape Shielded Hood D25 Desk Top Power Supply: 2.1 x 5.5 x 10mm, Center + Straight DC Connector 1 3 3 1 9 9 1 1 1 1 1 1 Ea. Ea. Ea. Ea. Ea. Ea. Ea. 100 Ft. Ea. 20 Ft. Ea. Ea. Relay Assembly Opto 22 Relays Relay Board / DAQ USB Cable, MaleA to MaleB Power Supply Sub-D25 Slider Jack Shielded Hood D25 1 16 1 2 1 1 1 Ea. Ea. Ea. Ea. Ea. Ea. Ea. Potentiometer Assembly Poteniometer DAQ, Personal Measurement Device: PMD-1208LS Shielded Hood D25 Sub-D25 Slider Plug Sub-D25 Slider Jack Slide Potentiometer Sensors: 1Kohms Travel = 60mm 1 1 2 1 1 3 Ea. Ea. Ea. Ea. Ea. Ea. Misc. Labview 1 1 Ea. Ea. III. Prototype Issues The primary issues that were apparent during the prototyping analysis were due to the lack of completion in the LabVIEW coding. Although LabVIEW was appropriately interfacing with the valves, the incremental applications of pressure were jerky and the amount was not controllable (the user was not able to fill the air muscle, for instance, 20% of the total inflation capability). Also, the LabVIEW computer interface was not set-up to be user-friendly; only a person that understands the software and code would be able to use it. Furthermore, the potentiometers were not included in the prototype, so there was no feedback during this initial testing. IV. Influences on Final Design Based on the initial prototype testing, it was evident that the valves were correctly assembled, the relay board and DAQ were properly wired, and the interfacing with LabVIEW was feasible. It was also apparent that further work on the LabVIEW code is necessary in order for the pressurized air to be controlled more smoothly and in regulated increments. Ultimately, the computer interface will need to be more user-friendly so that an average costumer would be able to use the program. V. Conclusions From this initial testing, it was clear that the valves, relay board, and DAQ have been correctly compiled and that they will work for this project. The prototype was also successful in showing that LabVIEW is capable of controlling the valves and that the interfacing works correctly. Lastly, it became apparent that there will need to be some sort of method for holding all of the electronic equipment to make transportation more manageable.