Health Policy Advisory Committee on Technology Technology Brief Update Implantable carotid sinus baroreflex device for the treatment of drug-resistant hypertension July 2014 © State of Queensland (Queensland Department of Health) 2014 This work is licensed under a Creative Commons Attribution Non-Commercial No Derivatives 3.0 Australia licence. In essence, you are free to copy and communicate the work in its current form for non-commercial purposes, as long as you attribute the authors and abide by the licence terms. You may not alter or adapt the work in any way. To view a copy of this licence, visit http://creativecommons.org/licenses/by-nc-nd/3.0/au/deed.en. For further information, contact the HealthPACT Secretariat at: HealthPACT Secretariat c/o Clinical Access and Redesign Unit, Health Service and Clinical Innovation Division Department of Health, Queensland Level 2, 15 Butterfield St HERSTON QLD 4029 Postal Address: GPO Box 48, Brisbane QLD 4001 Email: [email protected] Telephone: +61 7 332 89180 For permissions beyond the scope of this licence contact: Intellectual Property Officer, Department of Health, GPO Box 48, Brisbane QLD 4001, email [email protected], phone (07) 3328 9824. Electronic copies can be obtained from: http://www.health.qld.gov.au/healthpact DISCLAIMER: This Brief is published with the intention of providing information of interest. It is based on information available at the time of research and cannot be expected to cover any developments arising from subsequent improvements to health technologies. This Brief is based on a limited literature search and is not a definitive statement on the safety, effectiveness or costeffectiveness of the health technology covered. The State of Queensland acting through Queensland Health (“Queensland Health”) does not guarantee the accuracy, currency or completeness of the information in this Brief. Information may contain or summarise the views of others, and not necessarily reflect the views of Queensland Health. This Brief is not intended to be used as medical advice and it is not intended to be used to diagnose, treat, cure or prevent any disease, nor should it be used for therapeutic purposes or as a substitute for a health professional's advice. It must not be relied upon without verification from authoritative sources. Queensland Health does not accept any liability, including for any injury, loss or damage, incurred by use of or reliance on the information. This Brief was commissioned by Queensland Health, in its role as the Secretariat of the Health Policy Advisory Committee on Technology (HealthPACT). The production of this Brief was overseen by HealthPACT. HealthPACT comprises representatives from health departments in all States and Territories, the Australian and New Zealand governments and MSAC. It is a sub-committee of the Australian Health Ministers’ Advisory Council (AHMAC), reporting to AHMAC’s Hospitals Principal Committee (HPC). AHMAC supports HealthPACT through funding. This brief was prepared by Jonathan Henry Jacobsen from ASERNIP-S. TECHNOLOGY BRIEF UPDATE 2014 Technology, Company and Licensing Register ID WP066 Technology name Rheos® Baroreflex Hypertension Therapy™ System Patient indication Patients with drug resistant hypertension Stage of development in Australia Yet to emerge Established Experimental Established but changed indication or modification of technique Should be taken out of use Investigational Nearly established Australian Therapeutic Goods Administration approval Yes ARTG number (s) No Not applicable International utilisation Country Level of Use Trials underway or completed Canada Czech Republic Germany Latvia Netherlands United States of America Poland Switzerland Limited use Rheos® Baroreflex Hypertension Therapy™ System Update: July 2014 Widely diffused 1 2014 Evidence and Policy 2014 Safety and effectiveness Two case series (level IV Interventional evidence) evaluating implantable baroreflex stimulation for drug-resistant hypertension were included in this update. However, the included studies used different iterations of the baroreflex stimulation devices. Bakris et al1 evaluated the safety and efficacy of the Rheos® Baroreflex Hypertension Therapy™ System (CVRx®, Inc., Minneapolis, MN, USA), and Hoppe et al2 evaluated the Barostim neo™ (CVRx®, Inc.), a second generation device that supersedes the Rheos® System. In contrast with its predecessor, the Barostim neo™ system has a new unilateral 1 mm electrode and a smaller, more advanced stimulator. The case series study by Bakris et al1 enrolled participants from the Rheos Pivotal Trial3, a multicentre, double-blind, randomised, placebo-controlled crossover trial of the Rheos® System, to assess the long-term safety and efficacy of baroreflex activation therapy. The key enrolment criterion for the original Rheos Pivotal Trial was drug-resistant hypertension, defined as systolic blood pressure of at least 160 mmHg and a diastolic blood pressure of at least 80 mmHg after at least one month of maximally tolerated therapy with at least three antihypertensive medications, including a diuretic. The primary efficacy outcomes for Bakris et al1 were the mean change in systolic blood pressure and the mean number of blood pressure medications taken per patient following device implantation. The primary safety outcome was the number of deaths. Similarly, the primary outcomes in Hoppe et al 2 were the mean change in systolic blood pressure and the mean number of adverse events six months after the intervention. A summary of the studies is provided in Table 1. Table 1 Characteristics of included studies evaluating implantable baroreflex stimulation for hypertension Study Bakris et al 2012 Hoppe et al 2012 Design (evidence level) Case study (IV) Case study (IV) Prospective Prospective Device Rheos® System Barostim neo™ Number of patients 276 30 Length of follow-up 22 to 53 months 6 months Conflict of interest The first, second, fifth and last authors are paid consultants or advisors for CRVx®, Inc. The first six authors and the last author are paid consultants for CRVx®, Inc. The fourth author is an employee of CRVx®, Inc. The seventh and eighth authors are employees of CRVx®, Inc. Rheos® Baroreflex Hypertension Therapy™ System Update: July 2014 2 Bakris et al 2012 The purpose of this study was to assess long-term blood pressure control in patients with drug-resistant hypertension who participated in the original Rheos Pivotal Trial 3 (see November 2011 technology update for study details).1 Of the original Rheos Pivotal Trial cohort (n=322), 276 patients either consented to or remained eligible to consent to the long-term follow-up. Twenty-one patients withdrew during the initial trial3 and a further 25 did not consent to further participation (two patients died, five had the device removed and 18 withdrew). Consenting patients were divided into three groups: clinically significant responders (n=244), intermediate responders (n=32) and withdrawn (n=46). The inclusion and exclusion criteria for each group are listed in Table 2. Patients were followed up for an average of 28 months (standard deviation [SD] 9; range 22 to 53 months) after the Rheos Pivotal Trial. Table 2 Bakris et al1 study group categories Study group Inclusion/exclusion criteria Clinically significant responders Systolic blood pressure <140 mmHg or <130 mmHg in patients with diabetes or renal disease A sustained reduction in systolic blood pressure of 20 mmHg following device activation A sustained reduction in systolic blood pressure of 20 mmHg following device deactivation Received a replacement device following completion of the Rheos Pivotal Trial Did not receive a replacement device or have a sustained decrease in systolic blood pressure Intermediate responders Withdrawn Patients who withdrew from the Rheos Pivotal Trial Safety Two deaths occurred among the patients who were eligible but were not yet enrolled in the long-term study (one unclassified death occurred during sleep and one as a result of ischaemic encephalopathy). Two deaths occurred due to cardiopulmonary arrest during the long-term follow-up. None of the deaths were attributed to the device. Efficacy The results for the primary efficacy endpoints are summarised in Table 3. Overall, there was a sustained decrease in mean systolic and diastolic blood pressure in all three groups, compared with baseline values (p<0.001). Importantly, 55 per cent of the clinically significant responders maintained systolic blood pressures of less than 140 or less than 130 mmHg for patients with diabetes or kidney disease. By 12 months after device implantation, the mean number of medications used per patient had dropped significantly compared with baseline (p<0.001) in the clinically significant responder group, but it had returned to pre- Rheos® Baroreflex Hypertension Therapy™ System Update: July 2014 3 implant levels by the last follow-up visit. The number of medications used per patient decreased by an average of 0.6, making the clinical significance of this result unclear. Table 3 Efficacy outcomes for patients with a Rheos® System implant (Bakris et al 2012) Clinically significant responder (n=244) Pre-implant 12-month follow-up Last visit† Intermediate responder (n=32) Withdrawn (n=46) Pre-implant 12-month follow-up Last visit† Pre-implant 12-month follow-up Last visit† Mean blood pressure (SD) (mmHg) Systolic 177 (22) 139 (27)* 142 (29)* 177 (27) 155 (27)* 157 (22)* 187 (22) 161 (33)* 152 (33)* 85 (16)* 86 (18)* 105 (18) 95 (21)* 95 (19)* 108 (14) 92 (18)* 91 (23)* 79 (12) 75 (13) 75 (14) 74 (15) 73 (15) 74 (14) 5.9. (2.2) 5.2 (1.6) 5.0 (2.0) 4.8 (2.0) Diastolic 102 (15) Mean heart rate (SD) (beats/minute) 73 (15) 71 (14) 72 (14) Mean number of hypertensive medications used per patient 5.3 (1.9) 4.7 (2.1)* 5.0 (2.0) 6.2 (1.9) 5.4 (1.9) SD: standard deviation, *p<0.001 compared with pre-implantation, †Last visit between 22 and 53 months after implantation Hoppe et al 2012 Thirty adults with drug-resistant hypertension (defined as a systolic blood pressure ≥140 mmHg despite taking three antihypertensive medications, including a diuretic) were enrolled into the case series study (level IV Intervention evidence).2 Patients were recruited prospectively from seven centres within Canada and the European Union. Patients diagnosed with secondary hypertension, unstable angina, syncope, known or suspected baroreflex failure or autonomic neuropathy, previous myocardial infarction or cerebral vascular accident three months prior to the predicted implant date were excluded from the trial. All patients received a Barostim neo™ device, which was activated two weeks after implantation. Six-month follow-up data are currently available. Baseline patient demographics are outlined in Table 4. In general, patients were middle aged, overweight and under intensive treatment for hypertension. Safety No deaths were reported. Four adverse events occurred throughout the study period: a device pocket haematoma, a self-inflicted wound complication, intermittent pain lateral to the device and intermittent pain near the device. All adverse events were attributable to the procedure. However, each patient recovered with no recurrent side effects. Rheos® Baroreflex Hypertension Therapy™ System Update: July 2014 4 Table 4 Baseline characteristics of patients (n=30) included in Hoppe et al 2012 Variable Mean (SD) or n (%) Women 16 (53%) Age (years) 57 (12) Weight (kg) 85 (16) Body mass index (kg/m2) 30 (4.2) Blood pressure (mmHg) Systolic 172 (20) Diastolic 100 (14) Heart rate (beats/min) 75 (12) Number of medications used 6 (3) History of sleep apnoea 2 (7%) History of chronic kidney disease 3 (10%) History of diabetes 7 (23%) History of heart failure 1 (3%) History of renal nerve ablation 6 (20%) Efficacy The implant procedure time was approximately 107 minutes (SD 28), of which 44 minutes (SD 28) was spent mapping the procedure. There was a short learning curve associated with the procedure. By the third implant procedure, the overall procedure time was reduced to 97 minutes (SD 29) (p<0.05). Of the 30 patients, 90 per cent (n=27) received the implant on the right side, with most neck incisions being longer than 5 cm. Twenty-nine patients completed the six-month follow-up and achieved the primary efficacy objective of a reduction in office-cuff systolic blood pressure (Table 5). Interestingly, implanting the device but not activating it also reduced blood pressure, but the statistical significance of this effect was not reported. Additional endpoints included the mean change in heart rate (5 beats/minute [SD 3]) and mean number of medications used (0.3 ± 0.3), but these values were not statistically different from baseline (p>0.05 for both variables). A subgroup analysis of patients with prior renal nerve ablation (another invasive treatment for drug-resistant hypertension) showed that they also achieved a reduction in both blood pressure and heart rate. Systolic and diastolic blood pressure decreased by a mean of 22.3 (SD 9.8) and 11.3 (SD 8.1) mmHg, respectively, while the heart rate dropped by a mean of 5 ± 8 beats per minute. However, the statistical significance of these findings was not reported. Rheos® Baroreflex Hypertension Therapy™ System Update: July 2014 5 Table 5 Change in office-cuff systolic blood pressure in Hoppe et al 2012 (n=30) Mean Systolic blood pressure (SD) (mmHg) Mean Diastolic blood pressure (SD) (mmHg) Statistical significance Pre-implant 171.7 (20.2) 99.5 (13.9) NR Post-implant before activation 160.6 (29.3) 94.6 (19.00) NR Mean change from baseline at 3 months after activation 26.1 (3.3) NR P<0.001 Mean change from baseline at 6 months after activation 26.0 (4.4) NR P<0.001 NR, not reported 2014 Economic evaluation Borisenko et al 2013 The purpose of the economic evaluation was to compare the cost-effectiveness of Rheos® Baroreflex Hypertension Therapy™ System as a second-line treatment for drug-resistant hypertension with optimal medical therapy in a German population.4 The cost-effectiveness of the Rheos® System was determined using a decision analytic model that combined a decision tree with a Markov model. Data from the Rheos Pivotal Trial3 and the European Device Based Therapy in Hypertension Trial (DEBuT-HT) trial5 were used for the efficacy inputs of the model. Using a 50 year old patient with a high risk of end-organ damage as the patient population, the Rheos® System was found to reduce the lifetime incidence of myocardial infarction, stroke, heart failure and end-stage renal disease by 19, 35, 12 and 23 per cent compared with optimal medical management. Additionally, the Rheos® System increased survival from 15.9 to 17.6 years and quality-adjusted life-years from 13.9 to 16.1, with an incremental cost of AUD 26,012.14, compared with continued optimal medical management. The incremental cost-effectiveness ratio (difference in average costs compared with the difference in average quality-adjusted life-years between the study cohorts) determined that the Rheos® System was cost effective (base case analysis was AUD 12,007.38, which was lower than the standard incremental cost-effectiveness ratio threshold of EUR 35,000 [AUD 53,900] per quality-adjusted life-year used in European studies). Rheos® Baroreflex Hypertension Therapy™ System Update: July 2014 6 2014 Ongoing research Table 6 Current clinical trials evaluating the Barostim neo™ system Trial Identifier/ Location Trial status N Details Diagnosis Interventions Estimated completion date NCT01471834 Ongoing, but not recruiting 40 Case series Drug-resistant hypertension Barostim neo™ system July 2015 Recruiting 310 Drug-resistant hypertension Barostim neo™ system July 2015 Canada, Netherlands NCT01679132 Multicentre USA RCT Multicentre Optimal medical management NCT01471834: primary outcome is change in systolic blood pressure six months after the intervention; secondary outcomes were not listed. NCT01679132: primary outcomes are the number of adverse events 30 days after the intervention and the change in office-cuff systolic blood pressure from baseline to six months; secondary outcomes include the mean change in office-cuff and 24-hour ambulatory systolic blood pressure from baseline to 12 months after the intervention and the incidence of all hypertensive emergencies within six months of the intervention. 2014 Other issues During the first 12 months of the Rheos Pivotal Trial, adjunct pharmacotherapies could be altered at the discretion of the physician. Therefore, it was not clear whether this influenced patient outcomes. The Barostim Neo System has received CE mark approval in Europe.6 2014 Summary of findings Baroreflex stimulation is a novel therapy for drug-resistant hypertension. The largest data series (n=276) demonstrated a sustained reduction in both systolic and diastolic blood pressure by the study end (22 to 53 months’ follow-up). However, the Rheos Pivotal Trial was a randomised crossover trial, so the long-term efficacy of baroreflex stimulation compared with optimal medical management (the current standard for drug-resistant hypertension) remains undetermined. The other included study was small and had limited follow-up. However, the study reported that baroreflex stimulation had a good safety profile and few adverse events. In addition, there was a sustained decrease in systolic blood pressure by the six-month follow-up. However, whether these surrogate outcomes of reduced blood pressure and lowered heart rate translate to a lowered incidence of cardiovascular or renal diseases remains unknown. Furthermore, the potential conflicts of Rheos® Baroreflex Hypertension Therapy™ System Update: July 2014 7 interest, lack of appropriate comparators and long-term follow-up limits the conclusions that can be made regarding baroreflex stimulation. 2014 HealthPACT assessment Currently there is a lack of clinical and cost-effectiveness evidence to support the use of the baroreflex stimulation to reduce hypertension. It should be noted that the Baroreflex Rheos device has been superseded by the Barostim neo, which is not available in Australia. A large, prospective RCT on the Barostim neo device is due for completion in 2015, the results of which, if favourable, will be identified by horizon scanning activities. Therefore it is recommended that no further research on behalf of HealthPACT is warranted at this time. 2014 Number of studies included All evidence included for assessment in this Technology Brief has been assessed according to the revised NHMRC levels of evidence. A document summarising these levels may be accessed via the HealthPACT web site. Total number of studies 2 Total number of Level IV studies: 2 2014 References 1. Bakris, G. L., Nadim, M. K. et al (2012). 'Baroreflex activation therapy provides durable benefit in patients with resistant hypertension: results of long-term followup in the Rheos Pivotal Trial'. J Am Soc Hypertens, 6 (2), 152-8. 2. Hoppe, U. C., Brandt, M. C. et al (2012). 'Minimally invasive system for baroreflex activation therapy chronically lowers blood pressure with pacemaker-like safety profile: results from the Barostim neo trial'. J Am Soc Hypertens, 6 (4), 270-6. 3. Bisognano, J. D., Bakris, G. et al (2011). 'Baroreflex activation therapy lowers blood pressure in patients with resistant hypertension: results from the double-blind, randomized, placebo-controlled rheos pivotal trial'. J Am Coll Cardiol, 58 (7), 765-73. 4. Borisenko, O., Beige, J. et al (2014). 'Cost-effectiveness of Barostim therapy for the treatment of resistant hypertension in European settings'. J Hypertens, 32 (3), 68192. 5. Scheffers, I. J., Kroon, A. A. et al (2010). 'Novel baroreflex activation therapy in resistant hypertension: results of a European multi-center feasibility study'. J Am Coll Cardiol, 56 (15), 1254-8. 6. CVRx® (2011). CVRx® Receives CE Mark Approval and Introduces New Implantable Device for Hypertension, the Barostim neo™. [Internet]. CVRx®. Available from: http://www.cvrx.com/intl_news/cvrx-receives-ce-mark-approval-and-introducesnew-implantable-device-for-hypertension-the-barostim-neo/ [Accessed 18 Mar 2014]. Rheos® Baroreflex Hypertension Therapy™ System Update: July 2014 8 PRIORITISING SUMMARY (UPDATE 2011) REGISTER ID S000091 NAME OF TECHNOLOGY RHEOS HYPERTENSION THERAPY SYSTEMTM PURPOSE AND TARGET GROUP PATIENTS WITH DRUG RESISTANT HYPERTENSION STAGE OF DEVELOPMENT (IN AUSTRALIA) Yet to emerge Experimental Investigational Nearly established Established Established but changed indication or modification of technique Should be taken out of use AUSTRALIAN THERAPEUTIC GOODS ADMINISTRATION APPROVAL Yes No Not applicable INTERNATIONAL UTILISATION COUNTRY Czech Republic Germany Latvia Netherlands United States Poland Switzerland Trials Underway or Completed LEVEL OF USE Limited Use Widely Diffused 2011 SAFETY AND EFFECTIVENESS ISSUES Study description Results from the multicentre, double-blind, randomised, placebo-controlled Rheos Pivotal Trial were available for inclusion in this update (Bisognano et al 2011). The European Device Based Therapy in Hypertension Trial (DEBuT-HT) also provided data on the safety and efficacy of the Rheos® Baroreflex system (Scheffers et al 2010). Although numerous supplementary articles have been published which present preliminary data on the aforementioned trial, the publication by Scheffers et al (2010) presents the most detailed and long-term dataset to date. Rheos® Baroreflex Hypertension Therapy™ System Update: 2011 1 The Rheos Pivotal Trial (NCT00442286) was a Phase III clinical trial of 265 patients with resistant hypertension, designed to assess the safety and efficacy of the Rheos system (Bisognano et al 2011). The key enrolment criterion was resistant hypertension, defined as systolic blood pressure (SBP) of ≥160 mm Hg and diastolic blood pressure (DBP) of ≥80 mm Hg following at least one month of maximally tolerated therapy with at least three antihypertensive medications, including a diuretic. Other enrolment criteria included an ambulatory SBP of ≥135 mm Hg for a 24 hour average, and an absence of clinically significant orthostatic blood pressure (BP) changes. A total of 49 centres were used to screen 590 potential patients between March 2007 and November 2009. After screening for eligibility, 264 patients were found to be ineligible. Of the 326 patients who were eligible for implantation, four were not implanted with the device, as they failed to exhibit an acute testing response during surgery. The device, consisting of a pulse generator and leads attached to each carotid sinus, was implanted by either a vascular, cardiothoracic, or neurosurgeon in a total of 322 patients. Of these patients, 265 were randomised in a 2:1 fashion, 181 to group A and 84 to group B. Sample size was calculated to adequately power all primary endpoints, with 148 patients required in group A and 74 in group B. Following implantation, the device was switched off for one month in all patients. Patients in Group A then received baroreflex activation therapy (BAT) for the first six months, while BAT was delayed for six months in Group B patients. Subsequently, patients in both groups received BAT for a further six months. Patients and investigators were blinded to treatment until after the 12-month follow-up visit. The average patient follow-up was 21 ± 8 months, and one patient in group A was lost to follow-up before the 12-month visit (Bisognano et al 2011). An intention to treat analysis was used in this study, with unblinded and withdrawn subjects treated as failures. Both treatment groups had similar baseline characteristics, with no significant differences observed between the groups for age, sex, race, BP, heart rate, body mass index (BMI), comorbidities, and number and type of BP medications. Group A consisted of 116 males (64%) and 65 females (36%), with a mean age of 53.7 ± 10.5 years, mean SBP of 169 ± 26 mm Hg, mean DBP of 101 ± 17 mm Hg, taking a mean of 5.2 ± 1.6 BP medications. Group B consisted of 46 males (55%) and 38 females (45%), with a mean age of 52.4 ± 9.8 years, mean SBP of 168 ± 24 mm Hg, mean DBP of 100 ± 14 mm Hg, taking mean of 5.2 ± 1.8 BP medications (Bisognano et al 2011). The DEBuT-HT trial was a prospective, multicentre, nonrandomised feasibility study, which aimed to assess the safety and efficacy of the Rheos system in resistant hypertensive patients (Scheffers et al 2010). To be eligible for inclusion, patients were required to be over 21 years of age with BP of ≥160/90 mm Hg, despite receiving at least three antihypertensive agents, including a diuretic. All medications were kept constant two months prior, and three months following, implantation. Exclusion criteria included baroreflex failure, significant orthostatic hypotension, cardiac arrhythmias, chronic atrial fibrillation, clinically significant cardiac valvular disease or hypertension secondary to a treatable cause, carotid artery atherosclerosis with >50% stenosis, previous implants or radiation in the carotid sinus region, currently implanted electrical medical devices, dialysis and pregnancy. Rheos® Baroreflex Hypertension Therapy™ System Update: 2011 2 Scheffers et al (2010) included 45 patients who underwent Rheos implantation at nine clinical centres between March 2004 and November 2007. The group consisted of 26 males and 19 females with a mean age of 54 ± 9 years, mean SBP of 179 ± 29 mm Hg, mean DBP of 105 ± 22 mm Hg, taking a median of five antihypertensive medications. The device was activated one month after implantation, and patients were followed up monthly for the first three months, and annually thereafter. Of the 45 patients who underwent Rheos implantation, three were excluded from the safety and efficacy analyses (as per protocol), four dropped out and one missed a visit, resulting in a dataset of 37 patients at three months. Twenty-six and 17 patients completed the 1-year and 2-year follow-up, respectively. An additional cohort of 10 eligible patients who declined to participate was followed up by regular care. Outcomes measured included office and ambulatory BP, ability to exercise during a six minute hallwalk test, renal function, carotid artery stenosis and serious adverse events associated with implantation (Scheffers et al 2010). Safety The study by Bisognano et al (2011) evaluated procedural safety, BAT safety, and device safety. Some complications occurred prior to randomisation; four patients did not exhibit an acute testing response and were not implanted with the device, and two patients had the device explanted prior to randomisation due to infection. A total of seven deaths, none of which were related to the procedure or device, were reported during the follow-up period. The causes of death included three intracerebral haemorrhages, two cardiopulmonary arrests, one ruptured aortic aneurysm, and one drug overdose. The authors suggested that the majority of these deaths were due to the normal sequelae of long-term hypertension. With regard to procedural safety, the event-free rate was 74.8%, which was less than the prespecified objective performance criterion of 82% (P=1.00). Procedural adverse events were not presented by treatment group, and included surgical complications (n=13, 4.8%), nerve injury with residual deficit (n=13, 4.8%), transient nerve injury (n=12, 4.4%), respiratory complications (n=7, 2.6%), and wound complications (n=7, 2.6%). All other procedural adverse events occurred at a rate of less than 2%, and the majority (76%) of procedural adverse events resolved completely. With regard to BAT safety, the event-free rate was 91.7% in Group A and 89.3% in Group B (P<0.001), and the only adverse event that occurred at a rate greater than 2% in either group was hypertensive emergency (Group A: n=9, 5%; Group B: n=7, 8.3%). With regard to device safety, the event-free rate was 87.2%, which exceeded the pre-specified objective performance criterion of 72% (P<0.001), and the only event that occurred at a rate greater than 2% was hypertension-related stroke, which occurred in a total of 6 patients (2.3%). Scheffers et al (2010) reported that a total of seven patients (16.7%) experienced procedurerelated serious adverse events, while one patient (2.4%) experienced a device-related serious adverse event. Procedural complications included explantation of the device following infection (n=3), perioperative stroke with minimal residual effects (n=1), tongue paresis attributed to intraoperative hypoglossal nerve injury (n=1) and moderate pulmonary oedema that resolved in six days (n=1). One death due to angioneurotic oedema occurred in the cohort, and while the specific cause could not be determined, the authors suspected an adverse drug reaction. The sole device-related complication occurred as a result of movement Rheos® Baroreflex Hypertension Therapy™ System Update: 2011 3 of the implantable pulse generator, which was resolved by further surgery to reposition the device. In terms of functional safety measures, the distance travelled during a six minute hall-walk test improved by 48 meters (P=0.01, n=14) and serum creatinine levels increased substantially (P=0.04, n=22), one year after implantation (Scheffers et al 2010). In addition, carotid artery stenosis was absent in all patients at the 1-year follow-up visit. During the three months after implantation, no cases of orthostatic hypotension or syncope were reported (n=32). Effectiveness The primary efficacy endpoints reported by Bisognano et al (2011) included acute efficacy and sustained efficacy. Acute efficacy was defined as the proportion of subjects in Group A versus Group B that achieved at least a 10 mm Hg drop in SBP at month 6 compared with month 0, with a superiority margin of 20%. This study reported that 54% of patients in Group A and 46% in Group B were responders at six months, which was not significantly different (P=0.97). An additional efficacy analysis did show that the percentage of subjects achieving a SBP ≤140 mm Hg at six months was greater in Group A (42%) than Group B (24%) (P=0.005). Sustained efficacy was defined as the sustained response in SBP at month 12 in Group A patients who had a response at month 6, with an objective performance criterion of 65%. For this outcome, 88% of responders at six months maintained their response at 12 months (P<0.001). The mean change in SBP at six months, which was a secondary efficacy endpoint, was 16 ± 29 mm Hg in Group A and 9 ± 29 mm Hg in Group B (P=0.08). However at 12 months, at which point Group A had received BAT for 12 months and Group B six months, the mean decrease in SBP was 25 ± 32 mm Hg for Group A and 25 ± 31 mm Hg for Group B. In the study by Scheffers et al (2010), significant reductions in both office BP and ambulatory BP were observed at 3 months, 1 year, and 2 years post-implantation (Table 1). The authors reported that at each study visit, when the device was temporarily switched off in order to assess BP without carotid artery stimulation, BP immediately rose to baseline levels, before falling again once the device was turned on. Table 1: Blood pressure and heart rate measurements after implantation with the Rheos system (Scheffers et al 2010) Δ 3 months Δ 1 year Office Blood Pressure n = 37 n = 26 SBP (mm Hg) -21 ± 4 (P<0.001) -30 ± 6 (P<0.001) DBP (mm Hg) -12 ± 2 (P<0.001) -20 ± 4 (P<0.001) HR (beats/min) -8 ± 2 (P<0.001) -8 ± 2 (P=0.001) Ambulatory Blood Pressure n = 26 n = 15 SBP (mm Hg) -6 ± 3 (P=0.102) -13 ± 3 (P<0.001) DBP (mm Hg) -4 ± 2 (P=0.041) -8 ± 2 (P=0.001) HR (beats/min) -5 ± 2 (P=0.001) -6 ± 2 (P=0.012) All figures given are mean ± standard deviation. Baseline levels not reported. SBP: Systolic blood pressure; DBP: Diastolic blood pressure; HR: Heart rate. Rheos® Baroreflex Hypertension Therapy™ System Update: 2011 Δ 2 years n = 17 -33 ± 8 (P=0.001) -22 ± 6 (P=0.002) -11 ± 4 (P=0.008) n=8 -24 ± 8 (P=0.017) -13 ± 5 (P=0.049) -11 ± 34 (P=0.005) 4 COST IMPACT A recent study showed that the Rheos Baroreflex Stimulator may be cost effective, with an incremental cost-effectiveness ratio (ICER) between $50,000 and $100,000 per QALY (Young et al 2009). The characteristics of the cohort, such as age and SBP, were found to be important in determining the cost-effectiveness of the implantable carotid body stimulator for use in treating resistant hypertension. A Markov model was used in this study. Direct costs (2007 dollars), utilities, and event rates for future myocardial infarction, stroke, heart failure, and end-stage renal disease were included in the model. ETHICAL, CULTURAL OR RELIGIOUS CONSIDERATIONS No issues were identified from the retrieved material. OTHER ISSUES Numerous supplementary articles have been published reporting combined results from the European DEBuT-HT and US feasibility studies. The majority of these publications were excluded from this update, as the study published by Scheffers et al (2010) provides a more comprehensive representation of patients enrolled in the DEBuT-HT study. Also in some publications, data from the US Feasibility study has been pooled with the European study, resulting in patient overlap with Scheffers et al (2010). Despite this, several small case series provide additional data of interest. Sica et al (2009) reported on the effects of Rheos implantation in patients with chronic kidney disease (using a subset of data from the European DEBuT-HT trial). Results from this study demonstrated no change in the estimated glomerular filtration rate (eGFR), and a reduction in left ventricular mass index (LVMI) (data not reported). Based on these findings, the authors concluded that BAT may confer renoprotective effects, in addition to cardioprotective effects, in patients with resistant hypertension (Sica et al 2009). Karunaratne et al (2010) assessed whether the Rheos baroreflex stimulator had any deleterious interactions with pacemakers in four patients. This study demonstrated no oversensing at any pacemaker voltage setting, even when both devices were programmed to provoke interaction. This subset of patients had a mean follow up of 8.3 ± 4.6 months at the time of publication, during which no occurrence of observed or suspected oversensing was reported (Karunaratne et al 2010). Wustmann et al (2009) examined the effect of Rheos baroreflex stimulation on cardiovascular regulation in 21 patients. In addition to demonstrating a mean decrease in BP, this study also reported a fall in mean heart rate (P=0.001) and heart rate turbulence onset (P=0.004), and altered heart rate variability frequency domain parameters (P<0.001). The authors indicated that these cardiovascular changes were consistent with inhibition of sympathetic activity and increased parasympathetic activity, corresponding to the observed decrease in BP (Wustmann et al 2009). Including the European DEBuT-HT and US feasibility trials, there are currently six ongoing clinical trials evaluating the Rheos system (Table 2). Rheos® Baroreflex Hypertension Therapy™ System Update: 2011 5 Table 2: Ongoing and currently recruiting clinical trials using the Rheos® system ClinicalTrials.gov id n (estimated) Description NCT01077180 (ongoing) 16 NCT00710190 (ongoing) 45 NCT00710294 (ongoing) 50 NCT00442286 (ongoing) 300 NCT00718939 (ongoing) 60 NCT00957073 (recruiting) 540 United States Rheos® Feasibility Trial; case series to demonstrate the safety and efficacy of the Rheos® system. 13 month follow-up period. Device Based Therapy in Hypertension Trial (DEBuT-HT); case series to demonstrate the safety and efficacy of the Rheos® system. Continuous followup. European DEBuT-HT extension trial; collection of long-term safety and efficacy data to support regulatory approval and market release of the Rheos® baroreflex stimulator. Rheos® Pivotal Trial; prospective randomised controlled trial comparing clinical results between Rheos ‘ON’ and Rheos ‘OFF’ arms. 12 month follow-up period. Rheos® Diastolic Heart Failure Trial; prospective randomised double-blind controlled trial comparing Rheos® patients to non-implanted patients. 12 month follow-up period. Rheos® HOPE4HF Trial; prospective randomised controlled trial. Outcomes of primary interest are death and complications. Estimated date of completion March 2012 December 2010 December 2010 December 2009 January 2011 December 2014 2011 SUMMARY OF FINDINGS Although a significant number of articles have been published with regards to the safety and efficacy of the Rheos Baroreflex Stimulator, the majority of such articles are in the form of supplementary reports relating to several main trials. The study (Rheos Pivotal Trial) by Bisognano et al (2011), and the two year safety and efficacy data (DEBuT-HT study) presented by Scheffers et al (2010), represent the most comprehensive analysis of the Rheos system to date. At six months, Bisognano et al (2011) did not demonstrate a significant difference in the primary endpoint, number of acute responders, between the group that received stimulation and the group that did not have the device switched on; however, an additional analysis did show that more subjects in the stimulation group achieved SBP below 140 mm Hg. The device and stimulation therapy were found to meet pre-defined safety criteria, while for procedural safety the event-free rate was less than the pre-specified objective. In the study by Scheffers et al (2010), patients showed a reduction in office and ambulatory BP measurements following implantation, and these values continued to decrease two years post-implantation. Several serious complications were noted. 2011 HEALTHPACT ASSESSMENT The safety and effectiveness of the Rheos Baroreflex Stimulator remain under investigation. Several large clinical trials are ongoing. As these trials have the potential to expand Rheos® Baroreflex Hypertension Therapy™ System Update: 2011 6 knowledge relating to the safety and effectiveness of the Rheos system, it is recommended that this technology be monitored for 24 months. 2011 INCLUDED STUDIES Total number of studies Level II interventional evidence Level III-3 interventional evidence 2 1 1 2011 REFERENCES Bisognano JD, Bakris G, Nadim MK, Sanchez L, Kroon AA, Schafer J, de Leeuw PW, Sica DA. Baroreflex activation therapy lowers blood pressure in patients with resistant hypertension. Journal of the American College of Cardiology 2011; 58 (7): 765–773. Karunaratne H, Muluk S, Papademetriou V, Park M, Sample R, Irwin E. Implantation of a Carotid Baroreceptor Stimulator in Patients with Pacemakers and Hypertension. Pacing and Clinical Electrophysiology 2010;[Epub ahead of print] Scheffers IJM, Kroon AA, Schmidli J, Jordan J, Tordoir JJM, Mohaupt MG, Luft FC, Haller H, Menne J, Engeli S, Ceral J, Eckert S, Erglis A, Narkiewicz K, Philipp T, de Leeuw PW. Novel Baroreflex Activation Therapy in Resistant Hypertension – Results of a European Multi-Center Feasibility Study. Journal of the American College of Cardiology 2010; 56 (15): 1254–1258. Sica D, Rothstein M, Bisognano JD, Nadim MK, Lovett EG, Maufman D, de Leeuw P, for the DEBuT-HT and Rheos Feasibility Investigators. Baroreflex Activation Therapy preserves eGFR in hypertensive patients with Stage 2 and 3 chronic kidney disease: results from European and United States Trials of the Rheos® System. Journal of the American Society of Nephrology 2009; 20: 227A. Wustmann K, Kucera JP, Scheffers I, Mohaupt M, Kroon AA, de Leeuw PW, Schmidli J, Allemann Y, Delacrétaz E. Effects of chronic baroreceptor stimulation on the autonomic cardiovascular regulation in patients with drug-resistant arterial hypertension. Hypertension 2009; 54 (3): 530–536. Young KC, Teeters JC, Benesch CG, Bisognano JD, Illig KA. Cost-effectiveness of treating resistant hypertension with an implantable carotid body stimulator. Journal of Clinical Hypertension (Greenwich) 2009; 11 (10): 555-563. Rheos® Baroreflex Hypertension Therapy™ System Update: 2011 7 PRIORITISING SUMMARY (2008) REGISTER ID S000091 NAME OF TECHNOLOGY RHEOS HYPERTENSION THERAPY SYSTEMTM PURPOSE AND TARGET GROUP PATIENTS WITH DRUG RESISTANT HYPERTENSION STAGE OF DEVELOPMENT (IN AUSTRALIA) Yet to emerge Experimental Investigational Nearly established Established Established but changed indication or modification of technique Should be taken out of use AUSTRALIAN THERAPEUTIC GOODS ADMINISTRATION APPROVAL Yes No Not applicable INTERNATIONAL UTILISATION COUNTRY Czech Republic Germany Latvia Netherlands United States Poland Switzerland Trials Underway or Completed LEVEL OF USE Limited Use Widely Diffused IMPACT SUMMARY CVRx Inc. has developed the Rheos Hypertension Therapy System to treat people with drug resistant hypertension. At the time of writing, this technology is not available in Australia and is limited by Federal law in the United States to investigational use only. BACKGROUND Hypertension, also referred to as high blood pressure, is defined as systolic blood pressure (SBP) more than 140 mm Hg and/or diastolic blood pressure (DBP) more than 90 mm Hg (Illig et al 2006). It is a major risk factor for heart disease, stroke, peripheral vascular disease and kidney disease failure (Berlowitz et al 1998; Spranger et al 2006). Rheos® Baroreflex Hypertension Therapy™ System Update: 2008 1 Treatment of hypertension depends on its severity and whether the patient is in a high risk group. For those stratified as being at high or very high risk (five year risk of cardiovascular event 15% or greater), drug treatment is required in addition to lifestyle modifications (Vial 2004). There are many classes of medications used to treat hypertension including low-dose thiazide diuretics, beta blockers, calcium channel blockers, ACE inhibitors and angiotensin-II receptor antagonists (Vial 2004). The choice is influenced by the presence of contraindications for particular classes or the presence of co-indications for drugs of a particular class (Vial 2004). Most patients with hypertension requiring drug therapy need combination therapy with medications from two or more classes to achieve target blood pressure levels (Vial 2004). In Australia in 2005-06, agents acting on the rennin-angiotensin system were by far the most popular medicines prescribed or supplied for hypertension (53.4 per 100 problems), followed by calcium-channel blockers (18.9 per 100 problems), betablocking agents (11.9 per 100 problems) and diuretics (7.4 per 100 problems), while antihypertensives were seldom prescribed (1.7 per 100 problems) (AIHW 2007). Despite the vast number of oral medications designed to help control hypertension, not all patients are able to achieve desired blood pressure targets. There are numerous reasons for blood pressure being inadequately controlled including issues of compliance to prescribed medication, cost and incorrectly chosen medication combinations (Sica and Lohmeier 2006). However, a significant number of patients fail to achieve adequate blood pressure control despite adherence to maximal doses of several pharmacological agents (Fillipone and Bisognano 2007). These patients are said to have resistant hypertension, defined as failure to achieve a blood pressure of < 140/90 mmHg despite taking three anti-hypertensive medications, including a diuretic for at least one month (Fillipone and Bisognano 2007; Johnson 2007). In such patients new treatment options are sought to reduce hypertension induced cardiovascular morbidity. The Rheos Hypertension (HT) Therapy System (Rheos, CVRx. Inc., Maple Grove, MN, USA) is a new device based treatment for drug resistant hypertension which works by electrically activating the baroreceptors in the carotid arteries. Activation of the baroreceptors initiates the baroreflex - a series of responses to physiologic stimuli that determine the balance of sympathetic and parasympathetic activity in the heart and peripheral vasculature, and thereby participate in both short and long term control of blood pressure (Braunwald et al 2001). The Rheos Hypertension (HT) Therapy System is made up of three components - the Rheos Implantable Pulse Generator, Rheos Carotid Sinus Leads and the Rheos Programmer System. Electrodes implanted on the exterior surface of the carotid sinus wall are connected to the battery powered, programmable, impulse generator. The carotid sinus leads conduct the activation energy from the impulse generator to stimulate the baroreceptor fibres in the vessel walls of both carotid sinuses. These impulses are transmitted to the brainstem, where the increased nerve traffic originating from the baroreceptor afferents is interpreted as an elevated blood pressure and results in central nervous systems modulation of sympathetic and vagal outflows which reduce blood pressure and heart rate (Tordoir et al 2007). The programmable system is linked telemetrically to the impulse generator and allows the physician to non-invasively adjust the stimulation parameters delivered to the carotid sinus leads (Tordoir et al 2007). Rheos® Baroreflex Hypertension Therapy™ System Update: 2008 2 The degree to which cardiovascular risk is reduced with treatment of resistant hypertension is unknown. However, Calhoun et al (2008) suggest the benefits of successful treatment, are likely to be substantial as suggested by hypertension outcome studies in general and by the early Veterans Administration cooperative studies, which demonstrated a 96 % reduction in cardiovascular events over 18 months with use of triple antihypertensive regimens compared with placebo in patients with severe hypertension (diastolic blood pressure 115 to 129 mm Hg) (Veterans Administration cooperative study group on antihypertensive agents, 1967). CLINICAL NEED AND BURDEN OF DISEASE The number of Australians with hypertension in 2004-05 was 2,100,700 (AIHW 2008a). It is the most common problem managed in general practice, at 6.5 % of all problems in 2005-06 (AIHW 2007). Nearly 8 % of the burden of disease in Australia in 2003 could be attributed to hypertension (AIHW 2008b). A recent population based survey of 11,247 Australian adults found that the overall prevalence of hypertension was 29 % (Johnson 2007). However, only 47 % of these people were receiving antihypertensive medications, and less than 20 % had adequate blood pressure control (BP < 140/90 mm Hg) (Johnson 2007). The prevalence of resistant hypertension ranges from 3 to 5 % in primary health care to 21 % in groups of patients referred to tertiary institutions (Johnson 2007). DIFFUSION At the time of writing the CVRx Rheos Hypertension System was an investigational device only, undergoing clinical evaluation in Europe and the United States to demonstrate its safety and effectiveness. The data collected from this study by the FDA will be used to evaluate whether this device should be made available to patients with drug resistant hypertension in the US. COMPARATORS Lifestyle modifications (reduction in salt and alcohol intake and weight loss) combined with drug therapy are the current methods used to control resistant hypertension. SAFETY AND EFFECTIVENESS ISSUES Four case series studies (level IV intervention evidence) were selected for inclusion in this prioritising summary. Three of these are 1 page summary articles from supplement issues of journals (Bisognano et al 2008; De Leeuw et al 2008; Scheffers et al 2008). Bisognano et al (2008) recorded ambulatory blood pressure in 16 patients with resistant hypertension during 1 year of Baroreflex Hypertension TherapyTM. Ambulatory pressure was measured using an automated cuff programmed to inflate every 30 minutes during the day and every 60 minutes during the night. Only tests of at least 21 hours duration and with 70% of more available readings were used in their analysis. No mention was made of the patient’s details in the paper including whether they were taking any hypertension medication in addition to undergoing the Baroreflex Hypertension TherapyTM. Scheffers et al (2008) report the first two year data (blood pressure and heart rate) on 16 patients with resistant hypertension (8f /8m, age: 52 ± 9 yrs, BMI: 31 ± 6 lg/m2) enrolled in the DEBuT-HT trial (Device Based Therapy in Hypertension). The DEBuT-HT Clinical Trial is a multi-center European clinical evaluation of the Rheos Hypertension System in patients Rheos® Baroreflex Hypertension Therapy™ System Update: 2008 3 with drug resistant high blood pressure.The patients were a cohort from a total of 45 patients implanted with Baroreflex Hypertension Therapy SystemTM. At the time Scheffer et al (2008) were writing their paper the mean duration of the patient’s treatment with the device was 33 ± 5 months. The number of antihypertensive medications taken by the patients remained constant throughout the trial starting at 4.8 at baseline (when the device was activated) and 4.8, 4.7 and 4.6 at 3 months, 1 year and 2 years respectively. De Leeuw et al (2008) investigated changes in blood pressure, heart rate, left ventricular structure and number of anti-hypertension medications in 16 patients with resistant hypertension (12 Europe/4 US, 7 M/9 F, Age 50.4±11.5 yrs, BMI 33.1±7.8 kg/m2) during 12 months of Baroreflex Hypertension TherapyTM. The patients, who were implanted at 4 different centers, had stage II hypertension (systolic BP ≥ 160 mmHg) and were taking ≥ 3 anti hypertension drugs (≥ diuretic) at the time of implantation of the device. The patients continued to take the drugs whist undergoing the Baroreflex Hypertension TherapyTM. Tordoir et al (2007) report the short term outcome of 16 patients implanted with the Baroreflex Hypertension Therapy SystemTM as part of a multi-center feasibility trial for the treatment of resistant hypertension. The patients (8m/8f, age: 52.3±8.6 yrs, BMI: 31.0±6.2 kg/m2) all had severe hypertension despite a multi drug therapy with a mean of >5 concomitant antihypertensive drugs. Tests of hemodynamic responses (heart rate and blood pressure) to acute device activation were conducted. The settings for the tests included: continuous bilateral stimulation with a frequency of 100Hz and an impulse width of 480 µs. Stimulus amplitude (voltage) was increased in steps of 1V from 1-6V, each for 5 minutes. Patient responses to the device were monitored for 3 months of therapeutic electrical activation. It is not stated in the paper as to whether the patients remained on the antihypertensive drugs during the evaluation of the device. Safety Safety outcomes were not reported in the study by Bisognano et al (2008), it is therefore unclear if any adverse events occurred. Scheffers et al (2008) reported that no unexpected system or procedure related serious adverse events occurred during the observation period in their study (534 patient months). The only mention of safety outcomes by De Leeuw et al (2008) was that no unanticipated adverse events occurred. In the study by Tordoir et al (2007) the implantation of the baroreflex activating system was reported to have been well tolerated with no unexpected serious procedure or device related adverse events or perioperative deaths occurring. In the perioperative period (the time during which the procedure was being implanted and a 30 day period following the procedure) there were 38 procedure related adverse events reported in 17 patients. Of these events three were classified as serious adverse events: 1. Infection, which led to the complete surgical removal of the device in one patient. The implantable pulse generator was infected with a spreading infection along the leads to the neck incision at both sides. High fever and serious pain with redness necessitated surgical intervention. 2. Procedure related hypoglossal nerve injury with symptoms of hoarseness and eating disturbances which improved during follow up. In this patient the electrodes had to be placed high on the carotid bifurcation which resulted in an accidentally injury to the nerve and 3. One case of intraoperative bradycardia (to 20 Rheos® Baroreflex Hypertension Therapy™ System Update: 2008 4 beats/min), which recovered spontaneously without any sustained effect. Details of all adverse events are summarised in Table 1. Table 1: Perioperative device and/or procedure related adverse events (Tordoir et al 2007) Adverse event Device related Procedure related N events Infection Hypoglossal nerve injury Introperative bradycardia Pain Wound complication Extravascular tissue stimulation Anaesthesia complications Injury to local tissue Other Total 0 0 1 1 1 1 N patients with event 1 1 2 1 2 2 1 0 6 3 7 3 5 3 0 1 1 1 0 2 2 2 1 1 1 22 38 23 41 16 17 3 Effectiveness In the study of 16 patients by Bisognano et al (2008) significant reductions from baseline in 24 hour mean blood pressure, daytime systolic blood pressure, nighttime systolic blood pressure, pulse pressure, blood pressure load (percent of time systolic blood pressure exceeded 140 mmHg), and trough:peak ratio (ratio of mean systolic pressure for the first 2 hours after wake-up to the last two hours before going to bed) were observed following 12 months Baroreflex Hypertension TherapyTM (Table 2). Table 2: Changes in ambulatory measures following 12 months BHT (N=16) (Bisognano et al 2008) Measure Baseline 12-months BHT 24-hour mean (mm Hg) 171 ± 22 157 ± 24 Daytime SBP (mm Hg) 174 ± 22 159 ± 24 Nighttime SBP (mm Hg) 160 ± 26 148 ± 28 Pulse Pressure (mm Hg) 70 ±14 65 ± 13 BP load > 140 (%) 90 ± 16 71 ± 33 SD 24-hr SBP (mm Hg) 20 ± 4 18 ± 5 Trough:Peak 0.93 ± 0.11 1.04 ± 0.15 *p < 0.05 It is not stated in the paper as to what the error bars represent ∆ 12 months -14* -14* -11* -5* -19* -2 0.11* Scheffers et al (2008) observed significant blood pressure and heart rate reductions compared to baseline after 3 months, 1 year and 2 years of Baroreflex Hypertension TherapyTM (Table 3). A drop in systolic blood pressure of at least 20 mmHg was achieved in 12 of 16 (75 %) patients at 2 years and 5 of 16 (31%) achieved a systolic blood pressure of less than 140 mmgHg at 2 years. Rheos® Baroreflex Hypertension Therapy™ System Update: 2008 5 Table 3: Office based mean ± SD baseline and mean ± SE deltas (Scheffers et al 2008) N=16 Systolic BP (mm Hg) Diastolic BP (mm Hg Heart rate (bpm) *P<0.001 **P= 0.004 Baseline 191 ± 32 116 ± 22 81 ± 11 ∆ 3 months -34 ± 7* -20 ± 4* -14 ± 3* ∆ 1 year -38 ± 7* -27 ± 5* -12 ± 3* ∆ 2 year -35 ± 8* -24 ± 6* -12 ± 4** De Leeuw et al (2008) reported significant reductions in office cuff systolic blood pressure, office cuff diastolic blood pressure, septal wall thickness, left ventricular posterior wall thickness, left ventricular mass index and relative wall thickness from baseline after 3 months and 12 months of Baroreflex Hypertension TherapyTM. Left ventricular mass index decreased in 15 out of 16 patients after a year of Baroreflex Hypertension TherapyTM. Tordoir et al (2007) reported significant mean maximal reductions (p ≤ 0.0001) in systolic and diastolic blood pressure and heart rate in system tests conducted 1-3 days postoperatively. The tests, which were repeated monthly, showed consistent acute dose dependent reductions in systolic and diastolic blood pressure and heart rate in that the degree of change was directly related to the amplitude (voltage) of stimulation (Table 4). Reductions in blood pressure and heart rate were also observed within amplitudes between the 1 month (baseline values) and 4 month (3 months of baroreflex activation therapy) measurements (Table 4). A repeated measures ANOVA of blood pressure and heart rate readings during testing across voltage increments (0 volts to 6 volts) and visits (1 and 4 months) demonstrated significant differences across doses (p < 0.0001 for each) and by visit (p = 0.003 for systolic blood pressure, p = 0.0001 for diastolic blood pressure and p < 0.0001 for heart rate. Table 4: Blood pressure and heart rate during 1 and 4 month system tests, n = 16 (Tordoir et al 2007) Parameter and Baroreflex Activation IPG Stimulation Amplitude (V) Time Point 0 1 2 3 4 SBP (mm Hg) 1-month 184 ± 28 180 ± 29 176 ± 30 167 ± 32 160 ± 36 4-month 165 ± 21 164 ± 21 163 ± 27 160 ± 23 155 ± 27 DBP (mm Hg) 1-month 105 ± 17 104 ± 18 105 ± 17 98 ± 21 94 ± 22 4-month 95 ± 17 94 ± 16 93 ± 17 92 ± 15 89 ± 17 HR (BPM) 1-month 80 ± 14 78 ± 14 77 ± 15 74 ± 14 72 ± 14 4-month 68 ± 11 67 ± 10 68 ± 11 65 ± 11 65 ± 11 SBP; systolic blood pressure DBP; diastolic blood pressure HR; heart rate BPM; beats per minute It is not stated in the paper as to what the error bars represent 5 6 153 ± 31 147 ± 30 144 ± 36 143 ± 30 92 ± 21 84 ± 17 84 ± 24 81 ± 17 71 ± 15 64 ± 12 68 ± 14 62 ± 12 COST IMPACT At the time of writing a price for the Rheos Hypertension Therapy System had not been established as the product was still under clinical investigation. ETHICAL, CULTURAL OR RELIGIOUS CONSIDERATIONS No issues were identified from the retrieved material. Rheos® Baroreflex Hypertension Therapy™ System Update: 2008 6 OTHER ISSUES Each of the four case series had a person from the company manufacturing the Rheos Hypertension Therapy System (CVRx Inc.) as a co-author on the paper. In the paper by Tordoir et al (2007) the authors from CVRx, RK and RC, declare potential conflicts of interest as employees and shareholders of the sponsor CVRx Inc. and the authors EI and TP are consultants for the sponsor. SUMMARY OF FINDINGS The four papers reviewed all reported that Baroreflex Hypertension TherapyTM resulted in sustained reductions in blood pressure over the period the patients were monitored. The longest time investigated was two years (Scheffers et al 2008). Another outcome reported by De Leeuw et al 2008 was that left ventricular hypertrophy, a factor that increases risk of myocardial ischemia, regressed with Baroreflex Hypertension TherapyTM. Serious complications were reported in one of the studies which conducted a safety analysis (Tordoir et al 2007). It is important to note that these are only case series studies (level IV intervention evidence) that have presented preliminary results from feasibility clinical trials. Long term randomised controlled trials are required to further investigate the effectiveness and safety of Baroreflex Hypertension TherapyTM. Several long term clinical trials sponsored by CVRx Inc. are due to finish in 2009/2010 (http://clinicaltrials.gov; search term baroreflex). HEALTHPACT ACTION Given the promising results to date and that the Rheos Hypertension Therapy System is still under clinical investigation it is recommended that this system is monitored for 24 months. NUMBER OF STUDIES INCLUDED Total number of studies 4 Level IV intervention evidence REFERENCES Australian Institute of Health and Welfare. 2007. Medicines for cardiovascular health: are they used appropriately? Senes S & Penm E. Cardiovascular Disease Series Number 27. Cat. No. 36. Canberra: AIHW. Australian Institute of Health and Welfare. High Blood Pressure in Australia. 2008a http://www.aihw.gov.au/cvd/high_blood_pressure.cfm (Accessed September 2008). Australian Institute of Health and Welfare. 2008. Australia’s health 2008. Cat. No. AUS 99. Canberra: AIHW. Berlowitz DR, Ash AS, Hickey EC, Friedman RH, Glickman M, Kader B, Moskowitz MA. Inadequate management of blood pressure in a hypertensive population. N Engl J Med 1998; 339(27): 1957-1963. Bisognano J, de Leeuw P, Irwin E, Rossing M. Long term baroreflex hypertension therapy significantly lowers ambulatory blood pressures in patients with resistant hypertension. The Journal of Clinical Hypertension 2008; 10(supp A, no.5): A117, p. 271. Rheos® Baroreflex Hypertension Therapy™ System Update: 2008 7 Braunwald E, Zipes D, Lippy P. Heart Disease, A textbook of Cardiovascular Medicine. WB Saunders Company Philadelphia, USA 2001. Calhoun DA, Jones D, Textor S, Goff DC, Murphy TP, Toto RD, White A, Cushman WC, White W, Sica D, Ferdinand K, Giles TD, Falkner B & Carey RM. Resistant hypertension: diagnosis, evaluation, and treatment: a scientific statement from the American Heart Association Professional Education Committee of the Council for High Blood Pressure Research. Circulation 2008; 117(25): e510-e526. De Leeuw P, Gangahar D, Bach D, Lovett E. Left ventricular reverse remodelling with chronic treatment of resistant hypertension using an implantable device: results from European and United States trials of the Rheos Baroreflex Hypertension Therapy Journal of Hypertension 2008; 26(Suppl 1):S471. Filippone JD, Bisognano JD. Baroreflex stimulation in the treatment of hypertension. Current Opinion in Nephrology and Hypertension 2007; 16: 403-408. Illig KA, Levy M, Sanchez L, Trachiotis GD, Shanley C, Irwin E, Pertile T, Kieval R, Cody R. An implantable carotid sinus stimulator for drug-resistant hypertension: surgical technique and short-term outcome from the multicenter phase II Rheos feasibility trial. J Vasc Surg 2006; 44(6): 1213-1218. Johnson D. Resistant hypertension. Australian Doctor 2007; April: 27-34. Scheffers I, Schmidli J, Kroon AA, Tordoir J, Mohaupt M, Allemann Y, Jordan J, Engeli S, Liebeskind U, Luft FC, Eckert S, Hansky B, Baal, T, de Leeuw PW. Sustained blood pressure reduction by baroreflex hypertension therapy with a chronically implanted system: 2-year data from the Rheos debut-HT study in patients with resistant hypertension. Journal of Hypertension 2008; 26(Suppl 1): S19. Sica DA, Lohmeier TE. Baroreflex activation for the treatment of hypertension: principles and practice. Expert Rev. Med. Devices 2006; 3(5): 595-601. Spranger CB, Ries AJ, Berge CA, Radford NB, Victor RG. Identifying gaps between guidelines and clinical practice in the evaluation and treatment of patients with hypertension. Am J Med 2004; 117(1): 14-18. Tordoir JHM, Scheffers I, Schmidli J, Savolainen H, Liebeskind U, Hansky B, Herold U, Irwin E, Kroon AA, de Leeuw P, Peters TK, Kieval R and Cody R. An implantable carotid sinus baroreflex activating system: surgical technique and short-term outcome from a multicenter feasibility trial for the treatment of resistant hypertension. European Journal of Vascular and Endovascular Surgery 2007; 33: 414-421. Rheos® Baroreflex Hypertension Therapy™ System Update: 2008 8 Veterans Administration Cooperative Study Group on Antihypertensive Agents. Effect of treatment on morbidity in hypertension: results in patients with diastolic blood pressure averaging 115-129 mm Hg. JAMA 1967; 202: 1038-1034. Vial J. Hypertension – which therapy first? Australian Doctor 2004; October: 25-32. SOURCES OF FURTHER INFORMATION http://www.cvrx.com/medpros/index.php?id=15 http://www.cvrx.com/patients/index.php?id=33 SEARCH CRITERIA TO BE USED Rheos, baroreflex, carotid sinus AND resistant hypertensio Rheos® Baroreflex Hypertension Therapy™ System Update: 2008 9