Dosimetry for mammography
DS /CM/JlH
IAEA
International Atomic Energy Agency
Mammography dosimetry
• Measurements using phantoms
• Measurements on patients
Kampala
2
Mean glandular dose (MGD or Dg) - UK
• 5.3 cm standard breast.
• DG = g53 c53 s Ki
•
g53 c53 determined by
computer
• s depends on targets and anodes
• Ki IAK
b
2a
Kampala
3
Mammography equipment
• 45 mm thick PMMA phantom
for use in breast dosimetry
• Use of a multi-slab phantom
allows the possibility of
measurements at thickness
other than 45 mm
INTERNATIONAL ELECTROTECHNICAL
COMMISSION, Medical Electrical Equipment Dosimeters with Ionization Chambers and/or
Semi-conductor Detectors as used in X-Ray
Diagnostic Imaging, IEC-61674
Kampala
4
Measurements on Patients
• Measuring the mean glandular dose to the standard breast
has for many years been the main means of assessing the
doses from mammographic systems.
• Although this is a valuable and simple approach to dose
measurement it provides limited information on the doses
received by individual women.
• Consequently the periodic measurement of doses for
samples of women undergoing mammographic
examinations is recommended in the UK by the NHSBSP.
Kampala
5
Selection of patients
•
•
Kampala
For each patient and patient exposure
record the following data: selected
target/filter combination, tube voltage, tube
loading (post exposure mAs) and indicated
breast thickness. Data should be
recorded separately for left and right
breasts and the radiographic projection
used for each exposure should be noted.
Collect data for the required series of
patients.
6
Calculations for each patient case
DG =gc s Ki
Kampala
7
Results – Perth Royal Infirmary
Kampala
8
NHS BSP Software
Kampala
9
Scottish National Picture
Kampala
10