Download MEG381.LABS.LVDT

Survey
yes no Was this document useful for you?
   Thank you for your participation!

* Your assessment is very important for improving the workof artificial intelligence, which forms the content of this project

Document related concepts

Heterodyne wikipedia , lookup

Ground loop (electricity) wikipedia , lookup

Power inverter wikipedia , lookup

Stepper motor wikipedia , lookup

Current source wikipedia , lookup

Pulse-width modulation wikipedia , lookup

Islanding wikipedia , lookup

Stray voltage wikipedia , lookup

Rectifier wikipedia , lookup

Buck converter wikipedia , lookup

History of electric power transmission wikipedia , lookup

Schmitt trigger wikipedia , lookup

Voltage optimisation wikipedia , lookup

Resistive opto-isolator wikipedia , lookup

Mains electricity wikipedia , lookup

Power electronics wikipedia , lookup

Electric machine wikipedia , lookup

Three-phase electric power wikipedia , lookup

Voltage regulator wikipedia , lookup

Alternating current wikipedia , lookup

Switched-mode power supply wikipedia , lookup

Galvanometer wikipedia , lookup

Inductor wikipedia , lookup

Loading coil wikipedia , lookup

Ignition system wikipedia , lookup

Rectiverter wikipedia , lookup

Opto-isolator wikipedia , lookup

Coilgun wikipedia , lookup

Tesla coil wikipedia , lookup

Transformer wikipedia , lookup

Resonant inductive coupling wikipedia , lookup

Transcript
Department of Mechanical Engineering
MEG 381 Instrumentation
University of Bahrain
Linear Variable Differential Transformer
Introduction:
One of the most useful of the many variable inductance transducers is the differential
transformer or more specifically the so-called Linear Variable Differential Transformer or
LVDT. This displacement transducer provides an output voltage proportional to the
displacement of a ferrous core passing through the windings of the transformer. It is a mutual
inductance device making use of a primary and two secondary windings arranged generally
as shown in the figures below.
The primary winding is energized from an external AC source. For some designs such as
those used in the lab, the external source is a DC source which powers an AC generator
(oscillator) contained within the transducer body. The two secondary or pickup coils are
connected together differentially in phase opposition. When the core is perfectly centered the
output of each secondary coil is equal in magnitude but opposite (0, 180 degrees) in phase
and so they cancel exactly to yield a zero output. As the core is moved one way or the other,
the magnitudes change and the phase indicates in which direction the motion is taking place
(with a +/- 180 degree phase change )
Dr. Mostafa S. Habib
MEG 381 Instrumentation- LVDT
1
Objectives:
Determination the end deflection and stiffness of the cantilever.
Theory of Operation
LVDT is much like any other transformer in that it consists of a primary coil, secondary
coils, and a magnetic core. An alternating current, known as the carrier signal, is produced in
the primary coil. The changing current in the primary coil produces a varying magnetic field
around the core. This magnetic field induces an alternating (AC) voltage in the secondary
coils that are in proximity to the core. As with any transformer, the voltage of the induced
signal in the secondary coil is linearly related to the number of coils. The basic transformer
relation is:
Where:
Vout is the voltage at the output,
Vin is the voltage at the input,
Nout is the number of windings of the output coil, and
Nin is the number of windings of the input coil.
As the core is displaced, the number of coils in the secondary coil exposed to the coil changes
linearly. Therefore the amplitude of the induced signal varies linearly with displacement. The
LVDT indicates direction of displacement by having the two secondary coils whose outputs
are balanced against one another. The secondary coils in an LVDT are connected in the
opposite sense (one clockwise, the other counter clockwise). Thus when the same varying
magnetic field is applied to both secondary coils, their output voltages have the same
amplitude but differ in sign. The outputs from the two secondary coils are summed together,
usually by simply connecting the secondary coils together at a common center point. At an
equilibrium position (generally zero displacement) a zero output signal is produced. The
induced AC signal is then demodulated so that a DC voltage that is sensitive to the amplitude
and phase of the AC signal is produced.
Procedure:-
1-Use scale and a voltmeter as given in the lab and we determine LVDT core location (x) and
its corresponding output voltage (v).
2- Tabulate the measured (x) and (v).
3- Repeat for many core location covering the core full range.
4- plot the relationship of (x) versus (v)
5- Obtain the null position of the core, hence; plot the (x) versus v.
6- identify the linear and nonlinear regions
7- Remove the nonlinear points from your data.
8- Obtain the best fit (linear regression) for the linear region.
9- Determine the slop (=sensitivity in mV of the measured v per Volt of the input per mm of
(x)) of the LVDT.
10- Draw conclusions.
Dr. Mostafa S. Habib
MEG 381 Instrumentation- LVDT
2