Download DN302 - Ultraprecise Instrumentation Amplifier Makes Robust Thermocouple Interface

Survey
yes no Was this document useful for you?
   Thank you for your participation!

* Your assessment is very important for improving the workof artificial intelligence, which forms the content of this project

Document related concepts

Three-phase electric power wikipedia , lookup

Negative feedback wikipedia , lookup

History of electric power transmission wikipedia , lookup

Electrical ballast wikipedia , lookup

Thermal runaway wikipedia , lookup

Ground (electricity) wikipedia , lookup

Power inverter wikipedia , lookup

Variable-frequency drive wikipedia , lookup

Islanding wikipedia , lookup

Control system wikipedia , lookup

Signal-flow graph wikipedia , lookup

Resistor wikipedia , lookup

Electrical substation wikipedia , lookup

Immunity-aware programming wikipedia , lookup

Multimeter wikipedia , lookup

Ohm's law wikipedia , lookup

Current source wikipedia , lookup

Power electronics wikipedia , lookup

Power MOSFET wikipedia , lookup

Stray voltage wikipedia , lookup

Alternating current wikipedia , lookup

Rectifier wikipedia , lookup

Voltage optimisation wikipedia , lookup

Earthing system wikipedia , lookup

Test probe wikipedia , lookup

Voltage regulator wikipedia , lookup

Surge protector wikipedia , lookup

Two-port network wikipedia , lookup

Buck converter wikipedia , lookup

Mains electricity wikipedia , lookup

Resistive opto-isolator wikipedia , lookup

Metadyne wikipedia , lookup

Schmitt trigger wikipedia , lookup

Current mirror wikipedia , lookup

Network analysis (electrical circuits) wikipedia , lookup

Switched-mode power supply wikipedia , lookup

Opto-isolator wikipedia , lookup

Transcript
advertisement
Ultraprecise Instrumentation Amplifier Makes
Robust Thermocouple Interface – Design Note 302
Jon Munson
Introduction
The versatile and precise LTC®2053 instrumentation
amplifier provides an excellent platform for robust, low
power instrumentation products—as exemplified below
by the battery-powered thermocouple amplifier circuit. The
LTC2053 offers exceptionally low 10μV maximum input
offset along with 116dB typical CMRR and PSRR, a result
of a combination of switched capacitor and zero-drift op
amp technologies. It is optimized for low voltage supplies from 2.7V to 11V single ended or up to ±5.5V with
split supplies. The LTC2053 is ideal for battery-powered
instrument applications because of its low 850μA typical
current draw. The gain is easily programmed with two
resistors, as shown in Figure 1, just like a traditional noninverting op amp. The LTC2053 also features low 1/f noise
and rail-to-rail I/O to maximize dynamic range.
The Requirements of Thermocouple Amplification
A robust thermocouple amplification circuit must meet
several specific requirements. First, a commonly used
type K thermocouple develops 40.6μV/°C, and a standard
readout scale is 10mV/°C, so a precision amplifier with a
nominal gain of 246 is required. Also, thermocouple leads
are generally exposed to the electrical noise of an industrial
environment so the fully differential input capability of
an instrumentation amplifier helps eliminate errors due
to common mode noise pickup. Finally, fault protection
against accidental contact of the thermocouple to sources
of transients or high voltage is needed but the protection
cannot compromise accuracy.
The LTC2053 offers features that help meet all of these
requirements. It can withstand a 10mA of fault current
in any pin so 10kΩ protection resistors allow ±100V
hard faults or Level 4 ESD (8kV contact/15kV air-gap)
on the thermocouple junction without damage to the IC.
The LTC2053 uses a switched-capacitor input topology,
sampling at approximately 2.5kHz. With an internal input
sampling capacitance of ~1000pF, the RC transients of
the 10kΩ protection resistors settle within the ~180μs
01/03/302
sampling window so they do not contribute to offset errors as they might with a typical IA.
A Battery-Powered Thermocouple Amplifier
Figure 2 shows the LTC2053 used in a battery-powered
thermocouple amplifier. The circuit is used as a plug-in
adapter for common digital multimeters and is completely
portable. This circuit employs the LT1025 thermocouple
compensator to improve accuracy over a wide range of
ambient conditions and is mounted close to the thermocouple connection points for optimal thermal tracking. It
precludes the need to temperature stabilize the thermocouple “cold junctions” and removes the accuracy penalty
of a static room temperature correction value.
The output of the LT1025 provides a 10mV/°C correction voltage for the ambient temperature difference from
0°C—normally about 250mV at room temperature. The
measured probe temperature is the sum of this compensation voltage and the amplified thermocouple voltage.
Simple connection of the output of the compensator to
the REF input of the LTC2053 is all that is needed to add
these two voltages. The only consideration with this configuration is that the correction voltage must be capable
, LTC and LT are registered trademarks of Linear Technology Corporation.
V+
8
V1
3
V2
2
+
7
LTC2053
0.1μF
RG
REF 6
5
–
EN
1
VO
R2
4
VREF
R1
V–
( )
VO = (V1 – V2) 1 + R2 + VREF
R1
Figure 1. Typical Connection of
LTC2053 Instrumentation Amplifier
www.BDTIC.com/Linear
of either sourcing or sinking the feedback resistor current
that flows. As the LT1025 only sources current, a precision
buffer can be used to drive the REF node (e.g. using an
LTC2050 zero-drift op amp). The limitation imposed by
using a single supply is that both the probe and amplifier
unit temperatures must be above 0°C for valid output. If
negative temperatures must be accommodated, a simple
charge-pump inverter, such as an LTC1046, can be used
to develop a minus supply rail. The excellent PSRR of the
LTC2053 precludes the need for regulated power supplies, and the additional design and space expense they
entail. Four AA alkaline cells supply the ICs in this circuit
with 3.5V to 5V, depending on state of charge, yielding a
minimum full scale output of 350°C. The total battery draw
is typically only 1.8mA. In a conventional line-powered
application, one can use a single LT1025 and buffer amplifier to correct several LTC2053 thermocouple amplifier
channels, provided all the thermocouple connections and
the LT1025 thermally track.
Filtering and Protection
Since the LTC2053 operates by sampling the input signal, the frequencies of interest are generally below a few
hundred Hz so it is useful to rolloff the amplifier response
by adding 0.1μF in the feedback circuit. The capacitors in
the thermocouple input network help absorb RF pickup
and suppress sampling artifacts from appearing on the
thermocouple leads. The resistors connected to the
thermocouple provide a high impedance bias of VS /2
to maximize common mode immunity without inducing
voltage drops in the leads. For short thermocouple lead
lengths, which minimize common-mode signals, the probe
junction may be grounded (note that with split supplies,
grounding would be optimal). The 5.1V Zener is used to
provide fault-induced supply overvoltage and reverse-battery protection in conjunction with the 560Ω ballast.
VS
10M
10M
1M
1M
0°C–350°C
TYPE K
THERMOCOUPLE
40.6μV/°C
3
10k
10k
–
ORANGE
0.001μF
0.1μF
8
+
YELLOW
+
7
RG
REF 6
5
0.1μF
LTC2053
2
0.001μF
THERMAL
COUPLING
–
EN
1
4
10mV/°C
51Ω
249k
1%
100Ω
SCALE FACTOR
TRIM
VS
VS
2
1.00k
1%
6
4
0.1μF
LTC2050
LTC1025
3
3
200k
4
VS
–
1
+
560Ω
6V
+
–
CMPZ5231B
5.1V
2
5
Figure 2. Complete Schematic of the Thermocouple Amplifier
Data Sheet Download
For literature on our Instrumentation Amplifiers,
call 1-800-4-LINEAR. For applications help,
call (408) 432-1900, Ext. 2156
http://www.linear.com/go/dnLTC2053
Linear Technology Corporation
dn302f LT/TP 0103 316.5K • PRINTED IN THE USA
FAX: (408) 434-0507 ● www.linear.com
© LINEAR TECHNOLOGY CORPORATION 2003
1630 McCarthy Blvd., Milpitas, CA 95035-7417
(408) 432-1900
●
www.BDTIC.com/Linear