Download Snímek 1

Survey
yes no Was this document useful for you?
   Thank you for your participation!

* Your assessment is very important for improving the workof artificial intelligence, which forms the content of this project

Document related concepts

Radiosurgery wikipedia , lookup

Image-guided radiation therapy wikipedia , lookup

Medical imaging wikipedia , lookup

Nuclear medicine wikipedia , lookup

Positron emission tomography wikipedia , lookup

Technetium-99m wikipedia , lookup

Transcript
Nuclear Medicine
SPECT & Gamma Camera
Eugen Kvasnak, PhD.
Department of Medical Biophysics and Informatics
3rd Medical Faculty of Charles University
SPECT
Why SPECT?
Similar to X-ray Computed Tomography (CT) or Magnetic
Resonance Imaging (MRI), Single Photon Emission Computed
Tomography (SPECT) allows us to visualize functional information
about a patient's specific organ or body system.
SPECT
How does SPECT manage to give us functional information?
Internal radiation is administered by means of a pharmaceutical
which is labeled with a radioactive isotope. This so called
radiopharmaceutical, or tracer, is either injected, ingested, or
inhaled. The radioactive isotope decays, resulting in the emission of
gamma rays. These gamma rays give us a picture of what's
happening inside the patient's body.
SPECT
But how do these gamma rays allow us to see inside?
By using the most essential tool in Nuclear Medicine, the
gamma camera. The gamma camera can be used in planar
imaging to acquire 2-dimensional images, or in SPECT
imaging to acquire 3-dimensional images.
SPECT
What is SPECT?
SPECT is short for Single Photon Emission Computed Tomography.
As its name suggests (single photon emission), gamma ray
emissions are the source of information, rather than X-ray
transmissions as used in conventional Computed Tomography.
SPECT
How are these gamma rays collected?
The Gamma camera collects gamma rays that are emitted from
within the patient, enabling us to reconstruct a picture of where the
gamma rays originated. From this, we can determine how a
particular organ or system is functioning.
SPECT … history
• Although the first instance of SPECT was when Kuhl and Edwards
produced the first tomographs from emission data in 1963, the history of
SPECT detectors begins earlier.
In the 1940's crude spatial information about radioactive source
distributions within the brain were produced using a single detector
positioned at various locations around the head.
Ben Classen improved this method in the 1950's when he invented the
rectilinear scanner. This device produced planar images by mechanically
scanning a detector in a raster-like pattern over the area of interest. By
today's standards, this technique required very long imaging times
because of the sequential nature of the scanning.
A pin-hole in lead was used to project a gamma ray image of the source
distribution in 1953 by Hal Anger. The image was projected onto a
scintillating screen with photographic film behind it. This technique
required extremely long exposure times because of the huge inefficiencies
in the system (principally due to losses in the film). The inefficiencies in
the system resulted in extremely high radiation doses to patients.
SPECT … history
• In the late 1950's, Anger replaced the film and screen with a single
NaI crystal and PMT array. This formed the basis for the "Anger
Camera" which is now the standard clinical nuclear imaging device.
Modern Anger Cameras use a lead collimator perforated with many
parallel, converging or diverging holes instead of the original pinhole configuration.
Kuhl and Edwards were the first to present tomographic images
produced using the Anger Camera in 1963.
Everett, Fleming, Todd and Nightengale suggested the use of the
Compton effect for gamma-radiation imaging in 1977. This
technique is currently in use in astronomy. It's adaptation to SPECT
is non-trivial because of the vastly different source distributions and
geometry involved.
The investigation of the Compton Camera for SPECT began in
1983. Manbir Singh and David Doria proposed and experimented
with a basic design using solid state detectors, performed an
analysis of possible detector materials, and produced a small
prototype for testing.
Gamma Camera
Once a radiopharmaceutical has been administered, it is
necessary to detect the gamma ray emissions in order to attain
the functional information. The instrument used in Nuclear
Medicine for the detection of gamma rays is known as the
Gamma camera. The components making up the gamma
camera are the collimator, detector crystal, photomultiplier tube
array, position logic circuits, and the data analysis
computer. The purpose of each is briefly described below.
Gamma Camera
Gamma Camera1. Camera Collimator
The first object that an emitted gamma
photon encounters after exiting the body is
the collimator. The collimator is a pattern of
holes through gamma ray absorbing
material, usually lead or tungsten, that
allows the projection of the gamma ray
image onto the detector crystal. The
collimator achieves this by only allowing
those gamma rays traveling along certain
directions to reach the detector; this
ensures that the position on the detector
accurately depicts the originating location
of the gamma ray.
Gamma Camera
2. Scintillation Detector
In order to detect the gamma photon we use scintillation detectors. A Thalliumactivated Sodium Iodide [NaI(Tl)] detector crystal is generally used in Gamma
cameras. This is due to this crystal's optimal detection efficiency for the gamma ray
energies of radionuclide emission common to Nuclear Medicine. A detector crystal
may be circular or rectangular. It is typically 3/8" thick and has dimensions of 30-50
cm.
A gamma ray photon interacts with the detector by means of the Photoelectric Effect
or Compton Scattering with the iodide ions of the crystal. This interaction causes the
release of electrons which in turn interact with the crystal lattice to produce light, in a
process known as scintillation.
Gamma Camera
3. Photomultiplier Tubes
Only a very small amount of light is given off from the scintillation
detector. Therefore, photomultiplier tubes are attached to the back of the
crystal. At the face of a photomultipler tube (PMT) is a photocathode which, when
stimulated by light photons, ejects electrons. The PMT is an instrument that
detects and amplifies the electrons that are produced by the photocathode. For
every 7 to 10 photons incident on the photocathode, only one electron is
generated. This electron from the cathode is focused on a dynode which absorbs
this electron and re-emits many more electrons (usually 6 to 10). These new
electrons are focused on the next dynode and the process is repeated over and
over in an array of dynodes. At the base of the photomultiplier tube is an anode
which attracts the final large cluster of electrons and converts them into an
electrical pulse.
A Photomultiplier Tube Array
Each gamma camera has
several photomultiplier tubes
arranged in a geometrical
array. The typical camera has
37 to 91 PMT's.
Gamma Camera
4. Position Circuitry
The position logic circuits immediately
follow the photomultiplier tube array and
they receive the electrical impulses from
the tubes in the summing matrix circuit
(SMC). This allows the position circuits to
determine where each scintillation event
occurred in the detector crystal.
Gamma Camera
5. Data Analysis Computer
Finally, in order to deal with the incoming projection data and to process
it into a readable image of the 3D spatial distribution of activity within the
patient, a processing computer is used. The computer may use various
different methods to reconstruct an image, such as filtered back
projection or iterative reconstruction, both of which are further described
in this tutorial.
Acquisition Protocols
Various different acquisitions can be performed
with a SPECT camera….
1.
2.
3.
4.
Planar Imaging
Planar Dynamic Imaging
SPECT Imaging
Gated SPECT Imaging
• 1. Planar Imaging
•
The simplest acquisition protocol is the
planar image. With planar imaging, the
detector array is stationary over the
patient, and acquires data only from this
one angle. The image created with this
type of acquisition is similar to an X-ray
radiograph. Bone scans are done
primarily in this fashion.
• 2. Planar Dynamic Imaging
•
Since the camera remains at a fixed position in a
planar study, it is possible to observe the motion of a
radiotracer through the body by acquiring a series of
planar images of the patient over time. Each image is a
result of summing data over a short time interval,
typically 1-10 seconds. If many projections are taken
over a long time, then an animation of the tracer
movement can be viewed and data analysis can be
performed. The most common dynamic planar scan is to
measure glomerular filtration rate in the kidneys.
• 3. SPECT Imaging
•
If one rotates the camera around the
patient, the camera will acquire views of
the tracer distribution at a variety of
angles. After all these angles have been
observed, it is possible to reconstruct a
three dimensional view of the radiotracer
distribution within the body.
• 4. Gated SPECT Imaging
• As the heart is a moving object, by performing a
regular SPECT of the heart, the end image
obtained will represent the average position of
the heart over the time the scan was taken. It is
possible to view the heart at various stages of its
contraction cycle however, by subdividing each
SPECT projection view into a series of subviews, each depicting the heart at a different
stage of it's cycle. In order to do this, the
SPECT camera must be connected to an ECG
machine which is measuring the heart beat.
• Reconstruction
• The most common algorithm used in the tomographic
reconstruction of clinical data is the filtered
backprojection method. Other methods also exist,
please refer to the section on iterative reconstruction
methods.
• 1. Projection of Original Data
2. Transformation of Data into Fourier Domain
3. Filtering of Data
4. Transformation of Data Back into Spatial Domain
5. Backprojection
1. Data Projection
As a SPECT camera rotates around a patient, it creates a series of
planar images called projections. At each stop, only photons
moving perpendicular to the camera face pass through the
collimator. As many of these photons originate from various depths
in the patient, the result is an overlapping of all tracer emitting
organs along the specific path, much in the same manner that an Xray radiograph is a superposition of all anatomical structures from
three dimensions into two dimensions. A SPECT study consists of
many planar images acquired at various angles.
A set of projections taken of a patient's bone scan.
After all the projections are acquired, they are subdivided
by taking all the projections for a single, thin slice of the
patient at a time. All the projections for each slice are then
ordered into an image called a sinogram as shown below.
It represents the projection of the tracer distribution in the
body into a single slice on the camera at every angle of
the acquisition.
• The aim of the reconstruction
process is to retrieve the
radiotracer spacial distribution
from the projection data as it is
illustrated below. This surface
rendered image was
reconstructed using a fully 3D
OSEM algorithm.
2. Fourier Transform of Data
If the projection sinogram data were reconstructed at this point, artifacts would appear
in the reconstructed images due to the nature of the subsequent backprojection
operation. Additionally, due to the random nature of radioactivity, there is an inherent
noise in the data that tends to make the reconstructed images rough. In order to
account for both of these effects, it is necessary to filter the data. When we filter
data, we can filter it directly in the projection space, which means that we convolute
the data by some sort of smoothing kernel.
Convolution is a computationally intensive task however and so it is useful to avoid
using it when possible. It turns out that the process of convolution in the spatial
domain is equivalent to a multiplication in the frequency domain. This means that
any filtering done by the convolution operation in the normal spatial domain can be
performed by a simple multiplication when transformed into the frequency domain. To
see what is meant by the spatial domain and the frequency domain consider the
following illustrative example.
Consider a picket fence surrounding Old Lady Fourier's yard. Since Old Lady Fourier
has lived here for a long time and has never looked after her picket fence, it is rather
decrepit. At one time, the pickets were all evenly spaced apart and there were exactly
33 pickets over the 10 meter width of her yard when expressed in the spatial
domain. We can express this in the frequency domain however by saying that the
picket frequency is 3.3 pickets per m, or 3.3 m-1.
When the fence was new, we can plot graphically in the spatial domain, the number of
pickets vs the length in the yard. The same plot is shown as plotted in the frequency
domain. In the spatial domain, there is one picket spaced every 0.33 m along the
fence over the entire 10 m length. When transformed, we see that there is a large
peak at the frequency 3.33 m-1, which corresponds to all the pickets being spaced
equally apart.
As some of the pickets disappear, there is a change in these plots to the ones shown
below. Some of the pickets are missing from the spatial plot, and we see that in the
frequency space, there is a second peak emerging at 1 m-1 as now not all the pickets
are 0.33 meters apart. These pickets are now 1.0 meters apart and so the frequency
has decreased to 1.0 m-1 for these pickets.
This change in the way the same data is displayed is called a transform. In SPECT
imaging we make a similar transform of the projection data into the frequency space
whereby we can more efficiently filter the data. The transform that we make use of is
called the one dimensional Fourier Transform, so named after Old Lady Fourier.
3. Data Filtering
Once the data has been transformed to the frequency domain, it is then
filtered in order to smooth out the statistical noise. There are many
different filters available to filter the data and they all have slightly
different characteristics. For instance, some will smooth very heavily
so that there are not any sharp edges, and hence will degrade the final
image resolution. Other filters will maintain a high resolution while only
smoothing slightly. Some typical filters used are the Hanning filter,
Butterworth filter, low pass cosine filter, Weiner filter, etc. Regardless
of the filter used, the end result is to display a final image that is
relatively free from noise and is pleasing to the eye. The next figure
depicts three objects reconstructed without a filter true (left), without a
filter noisy (middle) and with a Hanning filter (right).
4. Inverse Transform of the Data
As the newly smoothed data is now in the frequency domain, we
must transform it back into the spatial domain in order to get out the
x,y,z information regarding spatial distribution. This is done in the
same type of manner as the original transformation is done, except
we use what is called the one dimensional inverse Fourier
Transform. Data at this point is similar to the original (left) sinogram
except it is smoothed as seen below (right).
5. Backprojection
The main reconstruction step involves a process known as
backprojection. As the original data was collected by only allowing
photons emitted perpendicular to the camera face to enter the
camera, backprojection smears the camera bin data from the filtered
sinogram back along the same lines from where the photon was
emitted from. Regions where backprojection lines from different
angles intersect represent areas which contain a higher
concentration of radiopharmaceutical.
The backprojection process.
SPECT Applications
… some examples of the many studies that can be performed with a
SPECT camera …
1. Heart Imaging
2. Brain Imaging
3. Kidney/Renal Imaging
4. Bone Scans
1. Heart Imaging
The following figure is a
myocardial MIBI scan taken
under stress conditions. Regions
of the heart that are not being
perfused will display as cooler
regions.
2. Brain Imaging
This figure is is a transverse
SPECT image of the
brain. Note the hot spots
present in the right posterior
region.
3. Kidney/Renal Imaging
The following is a renal planar scan using MAG3 tracer
(a glucose analog).
4. Bone Scans
Bone scans are typically
performed in order to assess
bone growth and to look for
bone tumours. The tumors are
the dark areas seen in the
picture below.
Thank you for your attention