Atlas AF1000 Friabilator
The Atlas AF1000 Friabilator is designed to meet the
specifications in the USP/EP and other pharmacopoeias. The unit is
supplied with two drums as standard – these can be mounted
individually or together on to the front of the unit. This gives better
visibility than side mounted drums.
The unit is easy to program using the alphanumeric key pad. The test may be
set to a specified number of rotations or a fixed time. The speed at which
the drum rotates is fixed at 25rpm
When the test begins the drum automatically moves into the load
position and the weight of the tablets is entered.
At the end of the test, the unit will automatically tip the tablets into
the receiving tray. The tablets should then be carefully brushed to
remove any powder then reweighed. This value is entered and the the
percentage weight loss calculated. This value along with the other test
parameters may be printed on a suitable printer.
The unit is supplied with a foldable leg that enables a test to be run at 10°
as per the USP if required.
Full IQ/OQ documentation is available.
Key Features
• Alpha-Numeric splash proof keyboard for entry of sample name, sample Number/ ID
number for data authentication.
• Selectable report printout complies to GLP requirements.
• Built-in Real Time Clock (RTC) for date, time display & authenticity of printout.
• Meets specification of USP, BP, EP.
• Two modes of operation - count or time mode.
• Automatic discharge of the sample into individual sample tray after completion of each test.
• 10° tilting of drums as per USP recommendation.
• Easy front loading system.
• Supports optional abrasion drum.
• Calculation mode for friability -percentage weight-loss.
• Balance interface for online sample weight transfer.
Specifications
Voltage (needs to be specified)
Program modes
Rotation speed
Count accuracy
Standard Drum
Optional drum
Keypad
Output
Tilt angle
Dimension (approx)
Shipping weight (approx)
115V/60Hz or 230V/50Hz
By rotations (up to 9999) or time (99:59.59)
25 +/-1 RPM (fixed)
+/- 1 rotation on setting
Dual chamber
Abrasion
Alphanumeric splash waterproof soft keys
Printer- Dot matrix 80 column or Deskjet Parallel Port
PC compatible RS232 for data download (software not provided)
Legs provide tilt of 10o
310 x 310 x 490mm (W x D x H)
10kg
SMI-Labhut Ltd. Unit D8 Churcham Business Park Churcham Gloucestershire GL2 8AX UK