Download Prepared By - Beckman Coulter

Survey
yes no Was this document useful for you?
   Thank you for your participation!

* Your assessment is very important for improving the workof artificial intelligence, which forms the content of this project

Document related concepts

Polysubstance dependence wikipedia , lookup

Stimulant wikipedia , lookup

Medication wikipedia , lookup

Prescription costs wikipedia , lookup

Pharmaceutical industry wikipedia , lookup

Plateau principle wikipedia , lookup

Bad Pharma wikipedia , lookup

Neuropsychopharmacology wikipedia , lookup

Pharmacognosy wikipedia , lookup

Urban legends about drugs wikipedia , lookup

Drug discovery wikipedia , lookup

Pharmacokinetics wikipedia , lookup

Drug interaction wikipedia , lookup

Transcript
Procedure:
Ecstasy
OSR9X229
This procedure is valid for the following chemistry analyzers:

AU400/AU400e

AU640/AU640e

AU480

AU680

AU600

AU2700

AU5400

AU5800
Prepared By
Date Adopted
Supersedes Procedure #
Review Date
Revision Date
Signature
# of
Distributed to
Copies
# of
Distributed to
© Beckman Coulter, Inc. April 2013
All printed copies are considered to be copies of the electronic original.
Copies
CLSI OSR9X229.03
Page 1 of 28
Procedure:
Ecstasy
OSR9X229
PRINCIPLE:
“Ecstasy” is the popular street name used to refer to
methylenedioxymethamphetamine (MDMA).2 Ecstasy and related drugs,
methylenedioxyamphetamine (MDA) and methylenedioxyethamphetamine
(MDEA), are amphetamine derivatives. The tablets, as sold illicitly in
Europe and North America, may also include amphetamine and
methamphetamine in the preparation.3,4
Ecstasy drugs are listed by the U.S. Drug Enforcement Administration as
Schedule l, which designates this drug as having no acceptable medical
application with great abuse potential. These compounds are central nervous
system stimulants that produce an initial feeling of euphoria and also a
feeling of increased well-being, self-esteem, with heightened mental and
physical capacity.5,6
MDMA is readily absorbed from the intestinal track. Peak plasma
concentrations occur approximately 2 hours after oral dose and are generally
low since MDMA passes readily into tissues. MDMA is metabolized in the
liver and the excretion of the drug occurs with over 95% of the drug cleared in
about 2 days7. Approximately 65% of the drug is excreted unchanged, with
10-15% of the dose excreted as methylenedioxyamphetamine (MDA) and a
similar small amount converted to amphetamine and methamphetamine.8
Methods historically used for detecting MDMA in biological fluids include
high-performance liquid chromatography, gas-liquid chromatography, and
enzyme immunoassay.9
INTENDED USE:
The Emit II Plus Ecstasy Assay is intended for use in the qualitative and
semi-quantitative analyses of methylenedioxymethamphetamine (MDMA)
and closely related drugs in human urine. Emit II assays are designed for
use with multiple Beckman Coulter AU analyzers.
© Beckman Coulter, Inc. April 2013
All printed copies are considered to be copies of the electronic original.
CLSI OSR9X229.03
Page 2 of 28
Procedure:
Ecstasy
OSR9X229
METHODOLOGY:
The Emit® II Plus Ecstasy Assay is a homogeneous enzyme immunoassay
used for the analysis of specific compounds in human urine. The assay is
based on competition between drug in the sample and drug labeled with the
recombinant glucose-6-phosphate dehydrogenase (rG6PDH) for antibody
binding sites. Enzyme activity decreases upon binding to the antibody, so the
drug concentration in the sample can be measured in terms of enzyme
activity. Active enzyme converts nicotinamide adenine dinucleotide (NAD) to
NADH in the presence of glucose-6-phosphate (G6P), resulting in an
absorbance change that is measured spectrophotometrically. Endogenous
serum G6PDH does not interfere because the coenzyme NAD functions only
with the bacterial (Leuconostoc mesenteroides) enzyme employed in the assay.
The Emit II Plus Ecstasy Assay provides only a preliminary analytical test
result. A more specific alternative chemical method must be used to obtain a
confirmed analytical result. Gas chromatography/mass spectrometry
(GC/MS) is the preferred confirmatory method1 but other chemical
confirmation methods are available. Clinical consideration and professional
judgment should be applied to any drug of abuse test result, particularly
when preliminary positive results are used.
SPECIMEN:
PATIENT / SAMPLE PREPARATION:
None required.
Additional instructions for preparation as designated by this laboratory:
© Beckman Coulter, Inc. April 2013
All printed copies are considered to be copies of the electronic original.
CLSI OSR9X229.03
Page 3 of 28
Procedure:
Ecstasy
OSR9X229
TYPE:
Urine samples are the recommended specimen type.
Additional type conditions as designated by this laboratory:
HANDLING CONDITIONS:
Urine specimens may be collected in plastic (i.e., polypropylene,
polycarbonate, polyethylene) or glass containers. Some plastics, other
than those listed, can absorb certain drugs.
Internal testing has shown that, if not analyzed immediately, specimens
may be stored unrefrigerated for up to 7 days. Specimens may be stored
refrigerated for 30 days before analysis. After 7 days unrefrigerated or 30
days refrigerated, samples should be stored frozen.10
Frozen specimens must be completely thawed and mixed thoroughly prior
to analysis.
Specimens with high turbidity should be centrifuged before analysis.
The recommended pH range for urine specimens is 3.0-11.0.
Adulteration of the urine specimen may cause erroneous results. If
adulteration is suspected, obtain another specimen.
Human urine specimens should be handled and treated as if they are
potentially infectious.
Additional handling conditions as designated by this laboratory:
© Beckman Coulter, Inc. April 2013
All printed copies are considered to be copies of the electronic original.
CLSI OSR9X229.03
Page 4 of 28
Procedure:
Ecstasy
OSR9X229
EQUIPMENT AND MATERIALS:
EQUIPMENT:
Beckman Coulter AU400/AU400e, AU480, AU600, AU640/AU640e,
AU680, AU2700, AU5400 and AU5800 analyzers.
MATERIALS:
Emit II Plus Ecstasy Assay
Antibody/Substrate Reagent 1 – Sheep polyclonal antibodies reactive to
methylenedioxymethamphetamine (MDMA), glucose-6-phosphate,
nicotinamide adenine dinucleotide, bovine serum albumin,
preservatives, and stabilizers
Enzyme Reagent 2 – methylenedioxyamphetamine (MDA) labeled with
bacterial rG6PDH, Tris buffer, bovine serum albumin, preservatives,
and stabilizers
Reagent storage location in this laboratory:
Test tubes 12 -16 mm in diameter or sample cups
(Cat No. AU1063).
Storage location of test tubes or sample cups in this laboratory:
© Beckman Coulter, Inc. April 2013
All printed copies are considered to be copies of the electronic original.
CLSI OSR9X229.03
Page 5 of 28
Procedure:
Ecstasy
OSR9X229
Emit II Plus Ecstasy Calibrators/Control products are packaged
individually and sold separately.
Emit II Plus 6AM/Ecstasy
Calibrators/Controls
OR
Emit II Plus
Ecstasy
Calibrators/Controls
9R529UL
OR
9X529
9R549UL
OR
9X549
9R569UL
OR
9X569
9R589UL
OR
9X589
Level
Level 1
(150ng/mL)
Level 2
(300ng/mL)
Level 3
(500ng/mL)
Level 4
(1000ng/mL)
Note: The Emit Calibrator/Controls contain stated concentrations of
(MDMA) methyenedioxymethamphetamine (ng/mL) for calibration of this
assay. Refer to package insert for concentration listings.
Storage location of the calibrator in this laboratory:
Preparation
The Emit II Plus Ecstasy Assay reagents are packaged in a ready to
use liquid form. No preparation is required.
© Beckman Coulter, Inc. April 2013
All printed copies are considered to be copies of the electronic original.
CLSI OSR9X229.03
Page 6 of 28
Procedure:
Ecstasy
OSR9X229
Note: Reagents 1 and 2 are sold as a matched set. They should not be
interchanged with components of kits with different lot numbers.
The Emit II Plus Ecstasy Calibrators/Controls are packaged in a
ready to use liquid form and may be used directly from the
refrigerator. Close the calibrator bottles when not in use. Caps must
always be replaced on the original containers.
Precautions:
1. The Emit II Plus Ecstasy Assay and Calibrator/Controls are for in
vitro diagnostic use.
2. Reagent 1 contains non-sterile sheep antibodies. Reagents 1 and 2
contain non-sterile bovine serum albumin.
3. No known test method can offer complete assurance that products
derived from human sources or inactivated microorganisms will not
transmit infection. Reagents, calibrators, and human specimens
should be handled using prevailing good laboratory practices to
avoid skin contact or ingestion.
4. Do not use reagents or calibrators after expiration date.
5. Components contain sodium azide, which may react with lead and
copper plumbing to form highly explosive metal azides. If waste is
discarded down the drain, flush it with a large volume of water to
prevent azide buildup. Dispose of properly in accordance with local
regulations.
6. This Emit II Plus Ecstasy Assay is qualified for use only with the
Emit Calibrators listed in the Calibrator section
Storage Requirements:
© Beckman Coulter, Inc. April 2013
All printed copies are considered to be copies of the electronic original.
CLSI OSR9X229.03
Page 7 of 28
Procedure:
Ecstasy
OSR9X229
Any reagents not loaded in the reagent refrigerator on the analyzer or
any calibrators not in use should be stored at 2-8°C (36-46ºF), upright,
and with caps tightly closed. Do not freeze reagents or calibrators.
Avoid exposure to temperatures above 32°C for prolonged periods of
time.
Unopened reagents are stable until the expiration date printed on the
label if stored as directed. Refer to Assay Methodology Sheets for
additional on-board, open vial stability information.
Opened and unopened calibrators are stable until the expiration date
printed on the vial label when stored as directed.
Improper storage of reagents or calibrators can affect assay
performance. Stability depends on handling reagents and calibrators
as directed.
Additional storage requirements as designated by this laboratory:
Indications of Deterioration:
Discoloration (especially yellowing) of the reagents or calibrators,
visible signs of microbial growth, turbidity, or precipitation in reagent
or calibrators may indicate degradation and warrant discontinuance of
use.
© Beckman Coulter, Inc. April 2013
All printed copies are considered to be copies of the electronic original.
CLSI OSR9X229.03
Page 8 of 28
Procedure:
Ecstasy
OSR9X229
PERFORMANCE PARAMETERS:
The following performance characteristics represent total system
performance and should not be interpreted to refer only to reagents. Studies
were performed on the Beckman Coulter AU analyzer series. Results may
vary due to analyzer-to-analyzer differences. Positive results were confirmed
by GC/MS.
PRECISION
Within run precision was performed and calculated according to Clinical
and Laboratory Standards Institute (CLSI EP5-A) by running 2 replicates
of each cutoff calibrator and positive and negative controls twice a day for
20 days (N=80). Total precision was calculated from this data. Results for
these studies are summarized in the following table.
MDMA 300 ng/mL Cutoff
Within-Run Precision
Total Precision
Cutoff
Calibrator
Control
75%
Control
125%
Cutoff
Calibrator
Control
75%
Control
125%
Mean
mAU/min
466
435
500
466
435
500
SD
4.2
5.6
5.6
4.9
6.7
7.5
%CV
0.9
1.3
1.1
1.1
1.5
1.5
MDMA 500 ng/mL Cutoff
Within-Run Precision
Cutoff
Control
Control
Total Precision
Cutoff
© Beckman Coulter, Inc. April 2013
All printed copies are considered to be copies of the electronic original.
Control
Control
CLSI OSR9X229.03
Page 9 of 28
Procedure:
Ecstasy
OSR9X229
Calibrator
75%
125%
Calibrator
75%
125%
Mean
mAU/min
585
539
612
585
539
612
SD
2.7
3.1
2.9
3.4
3.3
3.2
%CV
0.8
1.0
0.8
1.0
1.0
0.9
COMPARISON
Clinical urine specimens were tested using the Emit II Ecstasy Assay on
an Beckman Coulter AU analyzer and on a reference analyzer. Specimens
positive by either method were confirmed by GC/MS analysis. The results
are summarized below showing the number of positive and negative
results identified and the percent agreement between analyzers.
Assay
Cutoff
Positive
Negative
% Agreement
Ecstasy
300 ng/mL
56
52
99
Ecstasy
500 ng/mL
51
48
100
ANALYTICAL RECOVERY
Negative human urine specimens were spiked with known concentrations
of methylenedioxymethamphetamine (MDMA). Specimens spiked with
drug concentrations lower than the cutoff concentration were analyzed
qualitatively and correctly identified as negative 100% of the time.
Specimens spiked with drug concentrations greater than the cutoff were
© Beckman Coulter, Inc. April 2013
All printed copies are considered to be copies of the electronic original.
CLSI OSR9X229.03
Page 10 of 28
Procedure:
Ecstasy
OSR9X229
correctly identified as positive 100% of the time. Results from semiquantitative analysis of the specimens are listed in the following table.
300 ng/mL Cutoff
Concentration (ng/mL)
EIA Type I
Mean (ng/mL)
Poligonal
Mean (ng/mL)
100
107
78
250
257
260
375
393
402
450
448
455
Concentration (ng/mL)
EIA Type I
Mean (ng/mL)
Poligonal
Mean (ng/mL)
250
259
267
375
404
413
450
485
468
550
613
644
750
791
871
500 ng/mL Cutoff
© Beckman Coulter, Inc. April 2013
All printed copies are considered to be copies of the electronic original.
CLSI OSR9X229.03
Page 11 of 28
Procedure:
Ecstasy
OSR9X229
CALIBRATION:
QUALITATIVE ANALYSIS
Perform a one-point calibration (AB) using a water blank (blue rack) and
the appropriate EMIT Ecstasy Calibrator / Control for the desired cutoff;
Level 2 = 300 ng/mL, or Level 3 = 500 ng/mL. Refer to Analyzer Specific
Protocols for calibration set-point options and analyzer settings.
Three options are available for Qualitative Calibration:
Option 1: On the “Specific Test Parameters” menu, the “General
Tab”, program the Correlation B factor as 0.0. On the same screen,
under the “Range Tab” program the Value/Flag Level H as 999999.
Under Calibration Specific Parameters menu set the Calibration type
to MB. Blank the test using the blue rack. The cutoff calibrator (300
or 500) is run in a white rack. Each sample response is compared to
the cutoff calibrator response to determine if the sample is positive or
negative. Positive samples will not be flagged. Comparison of sample
responses is a manual process.
Option 2: On the “Specific Test Parameters” menu, the “General
Tab”, program the Correlation B factor as 0.0. On the same screen,
under the “Range Tab” program the Value/Flag Level H as 100. Under
Calibration Specific Parameters menu set the Calibration type to AB
with Formula as Y=ax+b. The Conc. for the calibrator should be
entered as 100. Blank the test using the blue rack. Calibrate by
placing the designated cutoff calibrator (300 or 500) in the assigned
position in the calibration rack (yellow rack). Positive samples will be
flagged (P) and will printout as greater than or equal to 100.
Option 3: On the “Specific Test Parameters” menu, the “General
Tab”, program the Correlation B factor as -100. On the same screen,
under the “Range Tab” program the Value/Flag Level H as 0.0. Under
Calibration Specific Parameters menu set the Calibration type to AB
© Beckman Coulter, Inc. April 2013
All printed copies are considered to be copies of the electronic original.
CLSI OSR9X229.03
Page 12 of 28
Procedure:
Ecstasy
OSR9X229
with Formula as Y=ax+b. The Conc. for the calibrator should be
entered as 100. Blank the test using the blue rack. Calibrate by
placing the designated cutoff calibrator (300 or 500) in the assigned
position in the calibration rack (yellow rack). Positive samples will be
flagged (P) and will printout as greater than or equal to zero.
SEMI-QUANTITATIVE ANALYSIS
Perform a multi-point calibration (4AB) using a water blank (blue rack)
and the Emit® Calibrator / Controls: Level 1, Level 2, Level 3, and Level 4
for the 300 or 500 Cutoff Assays. Calibration parameters are set to
perform calibration and set up the calibration curve. Refer to analyzer
User’s Guide or Analyzer Specific Protocol sheets for analyzer settings.
CALIBRATION STABILITY
Studies have shown the calibration stability to be at least 14 days.
Recalibrate as indicated by control results or with a new lot of reagent.
Calibration stability may vary from laboratory to laboratory depending on
the following: handling of reagents, maintenance of analyzer, adherence to
operating procedures, establishment of control limits, and verification of
calibration.
Note: When using a new set of reagents with the same lot number,
recalibration may not be required. Validate the system by assaying
controls.
QUALITY CONTROL:
During operation of the Beckman Coulter AU analyzer at least two levels of
control material should be tested a minimum of once a day. Controls should
be performed after calibration, with each new set or lot of reagent, and after
specific maintenance or troubleshooting steps described in the appropriate
© Beckman Coulter, Inc. April 2013
All printed copies are considered to be copies of the electronic original.
CLSI OSR9X229.03
Page 13 of 28
Procedure:
Ecstasy
OSR9X229
User’s Guide. Quality control testing should be performed in accordance with
regulatory requirements and individual laboratory’s standard procedures. If
more frequent verification of test results is required by the operating
procedures within your laboratory, those requirements should be met.
QUALITATIVE ANALYSIS:
Validate the calibration by assaying controls. Ensure that the result from
the negative control is negative (or lower) relative to the
Calibrator/Control set-point. Ensure that the result from the positive is
positive (or higher) relative to the Calibrator/Control set-point. Once the
calibration is validated, run urine specimens.
SEMI-QUANTITATIVE ANALYSIS:
Validate the calibration by assaying controls. Ensure that control results
fall within acceptable limits as defined by the testing facility. Once the
calibration is validated, run urine specimens.
QUALITATIVE AND SEMI-QUANTITATIVE ANALYSIS:
1. Follow government regulations or accreditation requirements for
quality control frequency. At least once each day of use, analyze two
levels of Quality Control (QC) material with known MDMA
concentrations. Follow your laboratory internal QC procedures if the
results obtained are outside acceptable limits.
2. Refer to the instrument operator’s manual for appropriate instrument
checks.
PARAMETERS:
A complete list of test parameters and operating procedures can be found in
the appropriate User’s Guide and at www.beckmancoulter.com. The
Analyzer Specific Protocol Sheets may also be used.
© Beckman Coulter, Inc. April 2013
All printed copies are considered to be copies of the electronic original.
CLSI OSR9X229.03
Page 14 of 28
Procedure:
Ecstasy
OSR9X229
CALCULATIONS:
None required.
REPORTING RESULTS:
REFERENCE RANGES:
No reference ranges are defined for drugs of abuse testing.
Expected reference ranges in this laboratory:
PROCEDURES FOR ABNORMAL RESULTS
The laboratory must define procedures to be used in reporting high
concentration results to the patient’s physician.
Abnormal results are flagged by the listed analyzers according to the
normal values entered by the user into the instrument parameters.
REPORTING FORMAT:
Semi-quantitative results are automatically printed in ng/mL at 37C.
Interpretation of Results
Qualitative Analysis -- When the Emit II Plus Ecstasy Assay is used
as a qualitative assay; the amount of drugs and metabolites detected
by the assay in any given sample cannot be estimated. The assay
results distinguish positive from negative samples only. The specific
© Beckman Coulter, Inc. April 2013
All printed copies are considered to be copies of the electronic original.
CLSI OSR9X229.03
Page 15 of 28
Procedure:
Ecstasy
OSR9X229
Emit® Ecstasy Calibrator/Control designated as the cutoff level is used
as a reference for distinguishing “positive” from “negative” specimens.
Positive Results: A specimen that gives a result equal to or higher
than the Calibrator/Control set point is interpreted as positive: The
specimen contains methylenedioxymethamphetamine (MDMA)and
closely related drugs.
Negative Results: A specimen that gives a result lower than the
Calibrator/Control set point is interpreted as negative: Either the
specimen does not contain MDMA and closely related drugs, or that
MDMA and closely related drugs are present in concentrations
below the cutoff level for this assay.
Semi-Quantitative Analysis -- When used semi-quantitatively, the
Emit® II Plus Ecstasy Assay yields approximate, cumulative
concentrations of the drug and metabolites detected by the assay. The
semi-quantitation of positive results enables laboratories to determine
an appropriate dilution of the specimen for confirmation by GC/MS.
Semi-quantitation also permits laboratories to establish quality control
procedures and assess control performance.
Immunoassays that produce a single result in the presence of multiple
detectable components cannot fully quantitate the concentration of
individual components. Interpretation of results must take into
account that urine concentrations can vary extensively with fluid
intake and other biological variables. A more specific alternative
chemical method must be used to obtain a confirmed analytical result.
Additional reporting information as designated by this laboratory:
© Beckman Coulter, Inc. April 2013
All printed copies are considered to be copies of the electronic original.
CLSI OSR9X229.03
Page 16 of 28
Procedure:
Ecstasy
OSR9X229
LIMITATIONS:
1. The Emit II Plus Ecstasy Assay is designed for use only with human
urine.
2. A positive result from the assay indicates the presence of Ecstasy and
closely related drugs, but does not indicate or measure intoxication.
3. Boric acid is not recommended as a preservative for urine.
4. Other substances and/or factors not listed (e.g., technical or procedural
errors) may interfere with the test and cause false results.
5. Interpretation of results must take into account that urine concentrations
of MDMA can vary extensively with fluid intake and other biological
variables.
6. Immunoassays that produce a single result in the presence of a drug and
its metabolites cannot fully quantitate the concentration of individual
components.
SENSITIVITY:
The minimum detection limit of the Emit® II Plus Ecstasy Assay is less
than 75 ng/mL. This level represents the lowest concentration of
methylenedioxymethamphetamine (MDMA) that can be distinguished
from 0 ng/mL with a confidence level of 95%.
SPECIFICITY
The Emit® II Plus Ecstasy Assay detects methlenedioxymethamphetamine
(MDMA) and closely related drugs in human urine.
© Beckman Coulter, Inc. April 2013
All printed copies are considered to be copies of the electronic original.
CLSI OSR9X229.03
Page 17 of 28
Procedure:
Ecstasy
OSR9X229
The following table lists the compounds this assay detects and the levels
at which the compounds have been found to give a response
approximately equivalent to that of the cutoff calibrators (Emit® Ecstasy
Calibrator/Control Level 2 (300 ng/ml) or Level 3 (500 ng/mL). Each
concentration represents the reactivity level for the stated compound
when it is added to a negative urine specimen. If a specimen contains
more than one compound detected by the assay, at lower concentrations
than those listed in the table, these may combine to produce a rate
approximately equivalent to or greater than that of the cutoff calibrator.
Compound
Concentration (g/mL)
300 ng/mL Cutoff
Concentration (g/mL)
500 ng/mL Cutoff
MDA
0.33
0.61
MDEA
0.29
0.50
MBDB
0.20
0.43
BDB
0.22
0.78
PMA
13
22
Compound
Concentration (g/mL)
300 ng/mL Cutoff
Concentration (g/mL)
500 ng/mL Cutoff
PMMA
3.1
9.0
HMMA
1400
2100
D-Amphetamine
160
430
D-Methamphetamine
37
130
D,L-Methamphetamine
18
53
D,L-Amphetamine
93
230
L-Amphetamine
220
310
L-Methamphetamine
30
87
© Beckman Coulter, Inc. April 2013
All printed copies are considered to be copies of the electronic original.
CLSI OSR9X229.03
Page 18 of 28
Procedure:
Ecstasy
OSR9X229
4-Chloramphetamine
9
12
D,L 4-Methylamphetamine
13
―
Benzphetamine
36
88
Buproprion
2000
4400
Chloroquine
6000
6000
L-Ephedrine
230
2000
Fenfluramine
5
10
Mephentermine
180
380
Methoxyphenamine
6900
>70000
Nor-pseudoephedrine
330
780
Phenmetrazine
3400
7400
Phentermine
700
1700
Phenylpropanolamine (PPA)
700
2200
Propanolol
1000
3200
Pseudoephedrine
220
530
Compound
Concentration (g/mL)
300 ng/mL Cutoff
Concentration (g/mL)
500 ng/mL Cutoff
Quinacrine
5000
5000
Tranylcypromine
420
630
Tyramine
1000
1600
Next is a list of the compounds that have produced a positive result.
Specificity testing was performed at the 300 ng/mL and 500 ng/mL cutoffs,
© Beckman Coulter, Inc. April 2013
All printed copies are considered to be copies of the electronic original.
CLSI OSR9X229.03
Page 19 of 28
Procedure:
Ecstasy
OSR9X229
which represents the greatest potential for cross-reactivity. Each
concentration represents the reactivity level for the stated compound
when it is added to a negative urine specimen.
Compound
Concentration (g/mL)
300 ng/mL Cutoff
Concentration (g/mL)
500 ng/mL Cutoff
m-Chlorophenylpiperazine
(M-CPP)
(Trazodone and Nefazodone
metabolite)
41
150
Dobutamine
49
240
Haloperidol
16
85
Isoxsuprine
47
165
Labelatol
35
80
Mebeverine
0.13
0.19
Methylone
24
74
Nylidrin
24
70
Trazodone
7
24
The following table lists the compounds that produce a negative result by
the Emit® II Plus Ecstasy Assay. These compounds were tested at the
300, and 500 ng/mL cutoff level, which represents the greatest potential
for cross-reactivity.
© Beckman Coulter, Inc. April 2013
All printed copies are considered to be copies of the electronic original.
CLSI OSR9X229.03
Page 20 of 28
Procedure:
Ecstasy
OSR9X229
Compound
Concentration
(g/mL)
300 ng/mL
Cutoff
Concentration
(g/mL)
500 ng/mL
Cutoff
Acetaminophen
1000
1000
-Acetyl-N,N-dinormethadol (dinor LAAM)
25
25
L--Acetylmethadol (LAAM)
25
25
N-Acetylprocainamide (NAPA)
400
400
Acetylsalicylic Acid
1000
1000
Albuterol
1000
1000
p-Aminobenzoic Acid (PABA)
1000
1000
Amitriptyline
100
500
Amoxicillin
1000
1000
Atenolol
1000
1000
Benzoylecgonine
1000
1000
Buprenorphine
1000
1000
Caffeine
1000
1000
Carbamazepine
250
250
Carisoprodol
1000
1000
Chlorpheniramine
500
500
Compound
Concentration
(g/mL)
300 ng/mL
Cutoff
Concentration
(g/mL)
500 ng/mL
Cutoff
Chlorpromazine
500
500
© Beckman Coulter, Inc. April 2013
All printed copies are considered to be copies of the electronic original.
CLSI OSR9X229.03
Page 21 of 28
Procedure:
Ecstasy
OSR9X229
Cimetidine
1000
1000
Clomipramine
2.5
2.5
Clonidine
1000
1000
Codeine
500
500
L-Cotinine
100
100
Cyclobenzaprine
125
125
Desipramine
800
800
Dextromethorphan
1000
1000
Dextrorphan
280
280
Diphenhydramine
1000
1000
Donepezil
59
170
Doxepin
250
250
Doxylamine
1000
1000
L-Epinephrine
1000
1000
2-Ethylidene-1,5-dimethyl-3,3diphenylpyrrolidine
(EDDP)
1000
1000
Fenoprofen
1000
1000
Fluoxetine
125
500
Furosemide
1000
1000
Glutethimide
500
500
Ibuprofen
1000
1000
Compound
Concentration
(g/mL)
300 ng/mL
Cutoff
Concentration
(g/mL)
500 ng/mL
Cutoff
© Beckman Coulter, Inc. April 2013
All printed copies are considered to be copies of the electronic original.
CLSI OSR9X229.03
Page 22 of 28
Procedure:
Ecstasy
OSR9X229
Imipramine
750
750
Ketamine
100
100
Ketoprofen
1000
1000
Ketorolac Tromethamine
350
350
Lidocaine
1000
1000
LSD
0.15
0.15
Meperidine HCl
1000
1000
Mescaline
1500
1500
Metaclopramide
1000
1000
Methadone
1000
1000
Methaqualone
1500
1500
D,L-Methyldopa
1000
1000
L-Methyldopa
1000
1000
Monoethylglycinexylidide (MEGX)
1000
1000
Morphine
1000
1000
Nalmefene
20
20
Naloxone
500
500
Naproxen
1000
1000
Nicotinic Acid
500
500
Nitroglycerin
1000
1000
Noracetylmethadol (nor LAAM)
25
25
11-nor-delta9-THC-9-COOH
100
100
© Beckman Coulter, Inc. April 2013
All printed copies are considered to be copies of the electronic original.
CLSI OSR9X229.03
Page 23 of 28
Procedure:
Ecstasy
OSR9X229
Compound
Concentration
(g/mL)
300 ng/mL
Cutoff
Concentration
(g/mL)
500 ng/mL
Cutoff
Nortryptyline
1000
1000
Ofloxacin
100
100
Oxazepam
300
300
Paroxetine
5
5
Phencyclidine
1000
1000
Phenelzine
100
100
1-Phenylcyclohexylamine (PCA)
50
50
Phenytoin
1000
1000
Phthalic Acid
1000
1000
1-Piperidinocyclohexane Carbonitrile
50
50
Procainamide
1000
1000
Promethazine
1000
1000
Propoxyphene
1000
1000
Ranitidine
1000
1000
Sertraline
125
125
Scopolamine
500
500
Secobarbital
1000
1000
Thioridazine
100
100
Tolmetin Sodium
2000
2000
Tramadol
1000
1000
© Beckman Coulter, Inc. April 2013
All printed copies are considered to be copies of the electronic original.
CLSI OSR9X229.03
Page 24 of 28
Procedure:
Ecstasy
OSR9X229
Compound
Concentration
(g/mL)
300 ng/mL
Cutoff
Concentration
(g/mL)
500 ng/mL
Cutoff
Trifluoperazine
1000
1000
Trimethobenzamide
500
500
Trimethoprim
1000
1000
Verapamil
1000
1000
Zidovudine (AZT)
2000
2000
Zolpidem
100
100
Diethylpropion HCL
1000
1000
D,L-Isoproterenol
1000
1000
Mephedrone
400
>600
Metaproterenol
10
10
3,4 Methylendioxypyrovalerone (MDPV)
440
>800
Methylphenidate (Ritalin®)
1000
1000
Phendimetrazine
400
400
Phenethylamine
20
20
Phenylephrine
20
20
Propylhexedrine
125
125
3-OH-Tyramine (dopamine)
300
300
Sympathomimetic Amines
© Beckman Coulter, Inc. April 2013
All printed copies are considered to be copies of the electronic original.
CLSI OSR9X229.03
Page 25 of 28
Procedure:
Ecstasy
OSR9X229
Laboratory specific procedure notes:
© Beckman Coulter, Inc. April 2013
All printed copies are considered to be copies of the electronic original.
CLSI OSR9X229.03
Page 26 of 28
Procedure:
Ecstasy
OSR9X229
NON-INTERFERING SUBSTANCES:
Each of the following compounds when added to urine containing
methylenedioxymethamphetamine (MMDA) at ± 25% concentration of the
cutoff do not yield a false response relative to the 300 and 500 ng/mL
Compound
Concentration
Acetone
1.0 g/dL
Ascorbic Acid
1.5 g/dL
Bilirubin
2.0 mg/dL
Creatinine
0.5 g/dL
Ethanol
1.0 g/dL
Gamma Globulin
0.5 g/dL
Glucose
2.0 g/dL
Hemoglobin
115 mg/dL
Human Serum Albumin
0.5 g/dL
Oxalic Acid
0.1 g/dL
Riboflavin
7.5 mg/dL
Sodium Chloride
6.0 g/dL
Urea
6.0 g/dL
SENSITIVITY
The sensitivity level of the EMIT II Plus Ecstasy Assay is less the 75
ng/mL. This level represents the lowest concentration of
© Beckman Coulter, Inc. April 2013
All printed copies are considered to be copies of the electronic original.
CLSI OSR9X229.03
Page 27 of 28
Procedure:
Ecstasy
OSR9X229
methylenedioxymethamphetamine (MMDA) that can be distinguished from 0
ng/mL with a confidence level of 95%.
REFERENCES:
1. Hawks RL, Chiang CN, eds. Urine Testing for Drugs of Abuse. NIDA
research monograph 73, National Institute on Drug Abuse (NIDA),
Rockville, MD. 1986.
2. MDMA, ONDCP Drug Policy Information Clearinghouse, Fact Sheet,
February 2004. NCJ-2013867 www.whitehousedrugpolicy.gov.
3. Cole J, et al. The Content of Ecstasy Tablets: Implications for the Study of
their Long-term Effects. Addiction 2002, 97, 1531-1536.
4. Valentine G. MDMA and Ecstasy. Psychiatric Times. February 2002. 21:2.
5. Monks T, et al. The Role of Metabolism in 3,4, (+)-Methylenedioxyamphetamine and 3,4, (+)-Methylenedioxymethamphetamine
(Ecstasy) Toxicity. Therapeutic Drug Monitoring. April 2004. 26:2.
6. Freese T, et al. The Effects and Consequences of Selected Club Drugs.
Journal of Substance Abuse Treatment, 2002. (23). 151-156.
7. Kalant H. The Pharmacology and Toxicology of “Ecstasy” (MDMA) and
Related Drugs. CMAJ 2001; 165(7):917-28.
8. Baselt RC. Disposition of Toxic Drugs and Chemicals in Man, Chemical
Toxicology Institute, 6th Ed. 2002, 687-688.
9. Zhao H, et al. Profiles of Urine Samples Taken from Ecstasy Users at
Rave Parties: Analysis by Immunoassays, HPLC, and GC-MS. Journal of
Analytical Toxicology, 2001; 25:258-269.
10. Mandatory Guidelines for Federal Workplace Drug Testing Programs.
Federal Register, Volume 69, No. 71, Section 2.4(i) p. 19661.
11. United Kingdom Laboratory Guidelines for Legally Defensible Workplace
Drug Testing: Urine Drug Testing. Version 1.0, March 2001.
© Beckman Coulter, Inc. April 2013
All printed copies are considered to be copies of the electronic original.
CLSI OSR9X229.03
Page 28 of 28