* Your assessment is very important for improving the workof artificial intelligence, which forms the content of this project
Download X-Ray Intensive Medical Procedures Using a Standard
Positron emission tomography wikipedia , lookup
Neutron capture therapy of cancer wikipedia , lookup
Backscatter X-ray wikipedia , lookup
Medical imaging wikipedia , lookup
Radiation therapy wikipedia , lookup
Nuclear medicine wikipedia , lookup
Radiosurgery wikipedia , lookup
Industrial radiography wikipedia , lookup
Radiation burn wikipedia , lookup
Center for Radiological Research wikipedia , lookup
X-Ray Intensive Medical Procedures Using a Standard Fluoroscope and its Lowest Automated Radiation Settings Karikari IO, Brown CR, Anderson DG Introduction: Methods Results: The deleterious effects of ionizing radiation are well established, with its cumulative impact having known health risks both for healthcare providers1-56,7 and patients1,9. This is of particular concern with respect to fluoroscopically guided medical procedures, as these rely on constant exposure to radiation throughout the procedure. While fluoroscopically guided procedures enhance surgical accuracy and reduce surgical morbidity, they do so at the cost of increased radiation exposure. A prospective, randomized cadaveric study was undertaken to compare the utility and radiation exposure experienced during a fluoroscopically intensive medical procedure. Physicians with experience in performing kyphoplasties were asked to duplicate the procedure in cadaveric specimens, imitating precisely their routine when performing a kyphoplasty in their daily practice. Each surgeon repeated the procedure at one spinal level using LessRay® enhanced imaging and another using conventional fluoroscopy unaided by enhancement. These were performed at adjacent spinal levels, randomly starting with conventional or reduced radiation imaging. In the reduced radiation case, though, the fluoroscope was set to the 1 pulse and low dose settings, the lowest automated setting on a GE OEC 9900 fluoroscope. As well, the physicians were blinded to the fluoroscope screen and only allowed to use the LessRay® enhanced images as long as they remained viable. At any point, if the image clarity was not sufficient to perform the procedure safely, the physicians were allowed to increase the radiation or look at the fluoroscope screen. Seven spine surgeons performed fourteen (14) kyphoplasties throughout the thoracolumbar spines of 2 human cadavers. In no case did any of the physicians abandon the LessRay® enhanced images. The average duration of fluoroscopy for the entire procedure was 11.0s and 65.6s in the LessRay® enhanced and conventional fluoroscopy groups respectively (p<0.001). Despite statistically similar number of images with both methods prior to cement injection (24 Conventional vs. 21 LessRay®assisted, p=0.30), the pulsed/low dosed procedures achieved an overall 88.8% radiation reduction over conventional imaging (36.1mGy vs. 4mGy, p< 0.001). Altering the dose settings and pulse rate on a conventional fluoroscope are simple measures that can radically minimize this exposure to ionizing radiation. These techniques, advocated by the American College of Radiology10 and the FDA11, can reduce radiation usage more than 90% as compared to conventional fluoroscopy12. Unfortunately, the image quality is often compromised with these techniques, rendering themineffective.13Despite this fact, the FDA has clearly stated that“medical imaging examinations should use techniques that are adjusted to administer the lowest radiation dose that yields an image quality adequate for diagnosis or intervention (i.e., radiation doses should be "As Low as Reasonably Achievable")11 – or ALARA. Therefore physicians require technology that allows greater utility of low radiation imaging. LessRay® is a computer display system that caninterface with a standard fluoroscope to enhance image resolution, allowing for a greater ability to use the pulse and low dose settings on a fluoroscope. It therefore offers a physician the opportunity to use significantly reduced radiation imaging in any procedure that requires a fluoroscope, and potentially better abide by the principles of ALARA. In this study, we assess the ability to use the lowest automatic radiation settings on a standard fluoroscope throughout the entirety of a radiation intensive surgical procedure when used in conjunction with LessRay. Typically, the image quality of pulsed and low dose images are not adequate to safely perform a surgical intervention. Hence, we are attempting to identify if poor quality images could be improved by LessRay® to a level considered adequate to perform the procedure safely. The number of images taken prior to cement injection, seconds of fluoroscopy for the entire procedure, and the amount of radiation in mGy for the entire procedure were recorded for each surgeon, as was the need to convert to conventional imaging for better clarity, if necessary. The amount of radiation reduction was quantified as the percent reduction. The chi-square test was used to determine statistical significance. A p value <0.05 was considered statistically significant. The Technology LessRay® is a computer display system that interfaces directly with the fluoroscope. The captured images produced by the fluoroscope are transmitted in real time to the LessRay® computer where the images are enhanced and then displayed on the LessRay® monitor. LessRay® uses proprietary image processing algorithms to post-process low-dose / pulsed images, improving resolution and contrast to provide a clinically valuable image to the physician. This allows for greater utility of low dose imaging. As the radiation needed to obtain the image dictates the radiation exposure to the patient and physician, using low radiation images should strongly impact this exposure. When a pulsed and/or low dose image of this shot is taken, they appear on the C-Arm monitor as washed-out and grainy, but the general form of the anatomy is faintly apparent through the noise. LessRay® embellishes this grainy image to make it appear more like the original image taken (see Figure I). As long as the image is of the same anatomy in a similar orientation, LessRay® will, in most cases,improve the poor quality precursor from the C-Arm monitor to a useable quality image on the LessRay® monitor. This works for both static images and continuous fluoroscopy, which is key for a procedure like a kyphoplasty which requires both. Discussion: Figure 1: A lateral view of the spine, taken during the study at 1 pulse and low dose settings, prior to (upper) and after (lower) enhancement by LessRay®. Note the clarity of the anatomy only once LessRay® improves the image. Across nearly all surgical subspecialties, minimally invasive (MIS) procedures are becoming increasingly commonplace. While the benefits have been lauded in numerous studies13-16 they often require increased radiation exposure due to the extensive Note: In clinical practice, the amount of patient dose reduction achieved when a Pulse and/or Low Dose is improved with LessRay is dependent on the clinical task, patient size, anatomical location and clinical technique. The dose should only be lowered to the level to which the physician is able to achieve the adequate image quality needed for the particular clinical task. A consultation with a radiologist and a physicist may aid in determining the appropriate dose settings. imaging needed to make up for what is not directly visualized at the time of surgery6,1721 . Although image guidance techniques exist for some procedures, conventional fluoroscopy remains the mainstay for imaging in MIS procedures. Radiation exposure from fluoroscopy has known significant medical risks, including concerns for cataracts, CV disease, and most worrisome, cancer1,8,9. The cumulative impact of this radiation exposure can quickly result in a physician exceeding their yearly or lifetime limits set by the National Council on Radiation Protection 2 2 . As dozens of the most common medical procedures performed in the United States are fluoroscopically dependent, the concern for radiation over-exposure is most acute in these procedures 16,17,19,21,23,24. Kyphoplasty is a common surgical procedure thatis totally dependent on fluoroscopic guidance, making it an ideal radiation intensive procedure for investigating radiation reduction strategies. In their practice guideline for the performance of vertebral augmentation, the American College of Radiology (ACR) presses the physician to use imaging during these procedures in a manner “consistent with the ALARA radiation safety guidelines.”31 The options available for minimizing radiation exposure include pulse and low dose imaging. They can be employed with standard fluoroscopy to reduce radiation over 90%12. When acquiring an image with pulsed radiation, an 8 pulse image uses essentially 8 times the radiation of a 1 pulse image, and roughly 1/4 the radiation of standard imaging settings12. While this technique is lauded by the ACR as one of the most effective ways to decrease radiation10, it often cannot be fully utilized in medical procedures. Image clarity is negatively affected by decreasing the radiation settings13. Each drop in pulse also causes a drop in image quality. Therefore, lower pulse rate settings often do not produce adequate image quality needed to safely perform a procedure13. In this study, all of the LessRay® procedures were performed exclusively using one pulse and low dose imaging (i.e. the lowest setting on a conventional C-arm) without the need to supplement these low quality images with any standard dose images to ensure safety or efficacy. The effectiveness of the LessRay® technology was demonstrated by the observation that no additional imaging was needed when LessRay® was utilized. Moreover there was no compromise in surgical accuracy and no procedure was aborted due to poor image quality. The technique of lowering the ionizing radiation emitted by the fluoroscope by altering the pulse rate is not unique to this study. Decreasing the pulse rate is strongly advocated by the ACR and the Society for Pediatric Radiology in their Image Gently and Image Wisely campaigns10. It has been employed in numerous studies, both preclinical15,26 and clinical13,27-30 to address radiation exposure. Although some neonatal and pediatric studies have shown the ability to lower the pulse rate down to less than 6pps (pulses per second)26, no study focusing on adults or adult anatomy has gone lower than 6pps13,26,28,2930Further, no study has gone to 1pps, and they typically do not include the addition of low dose imaging. The reason for this is probably best articulated in a recent study of minimally invasive spine fusions which sought to decrease radiation using 8pps in combination with low dose imaging wherever image clarity allowed that decrement in radiation. Unfortunately, "for fine detail,” when required to perform the procedure accurately, because the "detail of the pedicle is lost with the low-dose pulsed fluoroscopy,” the authors required steering away from that technique13. It should be noted that identifying accurately the borders of the pedicle is required in exactly the same fashion in performing the surgical intervention used in this paper, yet image enhancement allowed our physicians to continue to perform the procedure throughout at a lower radiation setting than their 8pps and not abandon it when anatomic detail was required. Conclusions: By digitally improving low radiation images, this study demonstrates that LessRay® allows entire procedures to safely be performed using the lowest automated radiation settings on a fluoroscope. The procedure can progress as the physician desires without altering the steps or the number of images. Setting the C-arm to one pulse and a low dose decreases radiation emitted by an order of magnitude. LessRay® makes these images useful. Together, they are a powerful combinationresulting in a safer environment for everyone in the operating room. 1. 2. 3. 4. 5. 6. 7. Hoffman DA, Lonstein JE, Morin MM, Visscher W, Harris BS, 3rd, Boice JD, Jr. Breast cancer in women with scoliosis exposed to multiple diagnostic x rays. Journal of the National Cancer Institute.Sep 6 1989;81(17):1307-1312. Adams MJ, Hardenbergh PH, Constine LS, Lipshultz SE. Radiation-associated cardiovascular disease. Critical reviews in oncology/hematology. 2003;45(1):55-75. Freedman DM, Sigurdson A, Rao RS, et al. Risk of melanoma among radiologic technologists in the United States. International journal of cancer. Journal international du cancer. Feb 10 2003;103(4):556-562. Mohan AK, Hauptmann M, Freedman DM, et al. Cancer and other causes of mortality among radiologic technologists in the United States. International journal of cancer. Journal international du cancer. Jan 10 2003;103(2):259-267. Yoshinaga S, Mabuchi K, Sigurdson AJ, Doody MM, Ron E. Cancer Risks among Radiologists and Radiologic Technologists: Review of Epidemiologic Studies. Radiology. 2004;233(2):313-321. Chou LB CS, Harris AH, Tung J, Butler LM. Increased breast cancer prevalence among female orthopedic surgeons. Journal Womens Health. 2112;21(6):683-689. Fisher BHaR. Influence of a prolonged period of lowdosage xrays on the optic and ultrastructural 8. 9. 10. 11. 12. 13. 14. 15. 16. 17. 18. 19. 20. 21. 22. 23. 24. 25. 26. 27. 28. 29. 30. 31. appearances of cataract of the human lens. Br J Opthalmol. 1979;63(7):457-464. Brenner DJ HE. Computed Tomography--An Increasing Source of Radiation Exposure. The New England Journal of Medicine. Nov 29 2007;357(22):2277-2284. Balter S HJ, Miller DL, Wagner LK, Zelefsky MJ. Fluoroscopically guided interventional procedures: a review of radiation effects on patients' skin and hair. Radiology. February 2010;254(2):326-341. Hernanz-Schulman M GM, Bercha IH, Strauss KJ. "Pause and Pulse: Ten Steps That Help Manage Radiation Dose During Pediatric Fluoroscopy". American Journal of Radiology. 2011(197):475-481. Administration UFaD. Initiative to Reduce Unnecessary Radiation Exposure from Medical Imaging. 2010; http://www.fda.gov/RadiationEmittingProducts/RadiationSafety/RadiationDoseRed uction/ucm199994.htm. Healthcare G. Hitting the Mark With Dose. 2008; http://www.gehealthcare.com/usen/xr/surgery/docs/hit _mark_dose_power.pdf. Clark JC JG, Marciano FF, Tumialán LM. Minimally invasive transforaminal lumbar interbody fusions and fluoroscopy: a low-dose protocol to minimize ionizing radiation. Neurosurg Focus. 2013;35(2). Harrington JF FP. Open versus minimally invasive microdisectomy: comparison of operative times, length of hospital stay, narcotic use, and complications. Minima Invasive Neurosurg. 2008;2008(51):30-35. Jae Hun Cho M, Jae Yun Kim, MD, Joo Eun Kang, MD, Pyong Eun Park, MD, Jae Hun Kim, MD, Jeong Ae Lim, MD, Hae Kyoung Kim, MD, and Nam Sik Woo MD. A Study to Compare the Radiation Absorbed Dose of the C-Arm Fluoroscopic Modes. The Korean Journal of Pain. 2011;24(4):199-204. Theocharopoulos N PK, Damilakis J, Papadokostakis G, Hadjipavlou A, Gourtsoyiannis N. Occupational exposure from common fluoroscopic projections used in orthopaedic surgery. . J Bone Joint Surg Am003;85(A (9)):1698-1703. Efstathopoulos EP, Pantos I, Andreou M, et al. Occupational radiation doses to the extremities and the eyes in interventional radiology and cardiology procedures. The British journal of radiology. Jan 2011;84(997):70-77. Ron E JP. Late health effects of ionizing radiation: bridging the experimental and epidemiologic divide. Radiati Res. Dec 2010;174(6):789-792. Sinnott B RE, Schneider AB. Exposing the thyroid to radiation: a review of its current extent, risks, and implications. Endocr Rev. Oct 2010;31(5):756-773. Martha S. Linet TLS, Donald L. Miller, Ruth Kleinerman, Choonsik Lee, Preetha Rajaraman, Amy Berrington de Gonzalez. Cancer Risks Associated with External Radiation From Diagnostic Imaging Procedures. CA Cancer J Clin. Feb 3 2012;62(2):75100. G. S. Occupational radiation exposure to the surgeon. Journal American Academy of Orthopedic Surgery. 2005;13(1):69-76. (U.S.) NCoRPaM. Review of the current state of radiation protection philosophy : recommendations of othe National Council on Radiation Protection and Measurements. . 1975;Publication No. 43. Gledo I, Pranjic N, Drljevic K, Prasko S, Drljevic I, Brzezinski P. Female breast cancer in relation to exposure to medical iatrogenic diagnostic radiation during life. Contemp Oncol (Pozn). 2012;16(6):551556. Brown KR, Rzucidlo E. Acute and chronic radiation injury. Journal of vascular surgery. Jan 2011;53(1 Suppl):15S-21S. Long SS, Morrison WB, Parker L. Vertebroplasty and kyphoplasty in the United States: provider distribution and guidance method, 2001-2010. AJR. American journal of roentgenology. Dec 2012;199(6):13581364. Ward VL BC, Strauss KJ, Lebowitz RL, Venkatakrishnan V, Stehr M, McLellan DL, Peters CA, Zurakowski D, Dunning PS, Taylor GA. Radiation exposure reduction during voiding cystourethrography in a pediatric porcine model of vesicoureteral reflux. Radiology. 2006;238(1):96-106. Rogers DP EF, Lozhkin K, Lowe MD, Lambiase PD, Chow AW. Improving safety in the electrophysiology laboratory using a simple radiation dose reduction strategy: a study of 1007 radiofrequency ablation procedures. Heart. 2011;97(5):366-370. Georges JL, Livarek B, Gibault-Genty G, et al. Reduction of radiation delivered to patients undergoing invasive coronary procedures. Effect of a programme for dose reduction based on radiationprotection training. Archives of cardiovascular diseases. Dec 2009;102(12):821-827. Greene DJ, Tenggadjaja CF, Bowman RJ, Agarwal G, Ebrahimi KY, Baldwin DD. Comparison of a reduced radiation fluoroscopy protocol to conventional fluoroscopy during uncomplicated ureteroscopy. Urology. Aug 2011;78(2):286-290. M. H-S. Science to practice: can fluoroscopic radiation dose be substantially reduced? Radiology. Jan 2006;238(1):1-2. http://www.acr.org/~/media/ACR/Documents/PGTS/ guidelines/Vertebral_Augmentation.pdf Note: In clinical practice, the amount of patient dose reduction achieved when a Pulse and/or Low Dose is improved with LessRay is dependent on the clinical task, patient size, anatomical location and clinical technique. The dose should only be lowered to the level to which the physician is able to achieve the adequate image quality needed for the particular clinical task. A consultation with a radiologist and a physicist may aid in determining the appropriate dose settings.