Download Class3

Survey
yes no Was this document useful for you?
   Thank you for your participation!

* Your assessment is very important for improving the workof artificial intelligence, which forms the content of this project

Document related concepts

Magnetotellurics wikipedia , lookup

Transcript
System Characteristics
Recap
SYSTEM CHARACTERISTICS
Medical Measurement Constraints
The amplitude and frequency ranges for each parameter are the
major factors that affect the design of all instrument components
Proper measurand-sensor interface cannot be obtained
Medical variables are seldom deterministic
External energy must be minimized to avoid any damage
Nearly all biomedical measurements depend on some form of energy
being applied to the living tissue or to the sensor, for example:
X-ray, ultrasonic imaging and electromagnetic or Doppler
ultrasonic
Blood flow-meters depend on externally applied energy interacting
with living tissue
Safe level of applied energy is an important consideration
Equipment must be reliable
Additional Medical Measurement Constraints
Reliable
Simple to operate
Withstand physical abuse and exposure to
corrosive chemicals
Electrical safety (minimize electric shock
hazard)
Classification of Biomedical Instruments
Quantity being sensed:
Pressure, flow, temperature, potential, etc.
Advantage: easy comparison of different methods for measuring
any quantity
Principle of transduction:
Resistive, capacitive, inductive, ultrasonic or electrochemical
Advantages: a. different applications of each principle can be
used to strengthen understanding of each concept
b. newer applications readily apparent
Measurement technique for each physiological system:
Cardiovascular, pulmonary, nervous, endocrine
Advantage: isolates all important measurements for specialists
Disadvantage: considerable overlap of quantities sensed and the
principles of transduction used
Classification of Biomedical Instruments...
Clinical medicine specialties:
Pediatrics, Obstetrics, Cardiology, Radiology,
etc.
Advantage: valuable for medical personnel
interested in specialized instruments
Interfering and Modifying Inputs
Desired input: the measurand that the instrument is
designed to isolate and measure
Interfering inputs: quantities that inadvertently affect
the instrument as a consequence of the principles used
to acquire & process the desired inputs
Modifying inputs: undesired quantities that indirectly
affect the output by altering the performance of the
instrument itself
Modifying inputs can affect processing of either desired or
interfering inputs
Some undesirable quantities can act as both a modifying
input and an interfering input
Example

A simplified ECG recording system provides a good
example:
In this recording system:
The desired input is: Vecg –
electrocardiographic voltage
between 2 electrodes (RA & LA)
The interfering inputs are:
50 Hz or 60 Hz (power-line)
noise voltage induced in the
shaded loop by ac magnetic
fields
Also the difference between the
currents running through each
of the electrodes to the patient
and to the ground causes a
voltage on Zbody
Example...
In ECG, the example of a
modifying input is the
orientation of the patient
cables. If the plane of the
cable is parallel to the ac
magnetic field, magnetically
introduced interference is
zero. If the plane of the
cables is perpendicular to
the ac magnetic field,
magnetically introduced
interference is maximal.
Time–dependent changes in
electrode impedance
Electrode motion
Elimination of Interfering and Modifying Inputs
 To reduce or eliminate the effects of most
interfering and modifying inputs we have two
alternatives:
1. Alter the design of essential instrument
components (preferred, but hard to achieve)
2. Add new components to offset the undesired
inputs
Sensor characteristics
Static characteristics
The properties of the system after all transient effects
have settled to their final or steady state
Accuracy, Discrimination, Precision, Errors, Drift,
Sensitivity, Linearity, Hysteresis (backslash)
Dynamic characteristics
The properties of the system transient response to an
input
Zero order systems
First order systems
Second order systems
Accuracy & discrimination
Accuracy is the capacity of a measuring instrument to
give RESULTS close to the TRUE VALUE of the measured
quantity
Accuracy is related to the bias of a set of measurements
(IN)Accuracy is measured by the absolute and relative errors
More about errors in a later
Discrimination is the minimal change of the input necessary to
produce a detectable change at the output
Discrimination is also known as RESOLUTION When the increment is
from zero, it is called THRESHOLD
Precision
The capacity of a measuring instrument to give the same reading when
repetitively measuring the same quantity under the same prescribed
conditions
Precision implies agreement between successive readings, NOT closeness to the
true value
Precision is related to the variance of a set of measurements
Precision is a necessary but not sufficient condition for accuracy
Two terms closely related to precision
Repeatability
The precision of a set of measurements taken over a short time interval
Reproducibility
The precision of a set of measurements BUT
taken over a long time interval or
Performed by different operators or
with different instruments or
in different laboratories
Example
Shooting darts
Discrimination
The size of the hole produced by a dart
Which shooter is more accurate?
Which shooter is more precise?
Shooter A
Shooter B
Accuracy and Errors
Systematic errors
Result from a variety of factors
Interfering or modifying variables (i.e., temperature)
Drift (i.e., changes in chemical structure or mechanical stresses)
The measurement process changes the measurand (i.e., loading errors)
The transmission process changes the signal (i.e., attenuation)
Human observers (i.e., parallax errors)
Systematic errors can be corrected with COMPENSATION methods (i.e. feedback,
filtering)
Random errors
Also called NOISE: a signal that carries no information
True random errors (white noise) follow a Gaussian distribution
Sources of randomness:
Repeatability of the measurand itself (i.e., height of a rough surface)
Environmental noise (i.e., background noise picked by a microphone)
Transmission noise (i.e., 60Hz hum)
Signal to noise ratio (SNR) should be >>1
With knowledge of the signal characteristics it may be possible to interpret
a signal with a low SNR (i.e., understanding speech in a loud environment)
Example: systematic and random errors
More static characteristics
Input range
The maximum and minimum value of the physical variable that can be measured (i.e., 40F/100F in a thermometer)
Output range can be defined similarly
Sensitivity
The slope of the calibration curve y=f(x)
An ideal sensor will have a large and constant sensitivity
Sensitivity-related errors: saturation and “dead -bands”
Linearity
The closeness of the calibration curve to a specified straight line (i.e., theoretical behavior,
least-squares fit)
Monotonicity
A monotonic curve is one in which the dependent variable always increases or decreases as
the independent variable increases
Hysteresis
The difference between two output values that correspond to the same input depending on
the trajectory followed by the sensor (i.e., magnetization in ferromagnetic materials)
Backslash: hysteresis caused by looseness in a mechanical joint