Survey
* Your assessment is very important for improving the workof artificial intelligence, which forms the content of this project
* Your assessment is very important for improving the workof artificial intelligence, which forms the content of this project
NEHRU ARTS AND SCIENCE COLLEGE, COIMBATORE DEPARTMENT OF ELECTRONICS & COMMUNICATION SYSTEMS MEDICAL ELECTRONICS Part A Unit 1 1. The bio signal frequencies from various sections of the human body are in the 0 to few kHz. 2. The EEG signals are originated from the Glia cells 3. The Range of frequency and voltage related to EEG are 0.5 Hz to 100 Hz and 10 micro volts to 100 micro volts. 4. The bundle of muscle fibers in a muscle supplied by a single motor nerve fiber is called a motor unit 5. The bio electric generator of heart is situated at SA node. 6. The most abundant negative ions in our body are chlorides. 7. Among the following electrodes which have high Z? Micro electrodes 8. In the case of ERG what types of electrodes used to pickup signals?Retinal electrodes 9 .Loud speakers is also in the EMG recorder 10. Pressure transducer for measuring blood pressure is from strain gauge or capacitive Transducer Unit 2 1. The hydrogen ion concentration of the blood is most easily determined with a glass electrodes 2. To obtain good contact between the electrode and the skin the gap is filled with an electrode paste containing electrolytes 3. The active transducer in the measurement of pressure is piezoelectric transducer. 4. The amplifier which has no drift is chopper amplifier 5. To reduce common mode interference during recording of bio signals one can use differential amplifier 6. Resistively generated interference arises through incorrect grounding 7. For biomedical application the mostly used amplifier is differential amplifier 8. EMG deals with the study of muscular activity 9. Mingographs are connected with ECG 10. The heart sounds are recorded by phonocardiography. Unit 3 1. The heart dipole field is measured by lead system called orthogonal lead system 2. Arrhythmia can be diagnosed by ECG 3. The T wave is produced during repolarisation of the ventricles 4. Bicycle ergometer is related with ECG 5. For faithful reproduction of QRS complex of ECG signal, an amplifier band width should be 0.05-100 Hz 6. Artifacts are time varying half cell potentials 7. The resting potential of the inside of the neuron is about -70mV 8. EEG waves include frequency content ranging from less than 1 Hz to 50 Hz. 9. Recording of the peripheral nerves action potentials is called Electro neurography 10. The conduction velocity in a motor nerve is normally 50 m/s Unit 4 1. During myocardial infarction one can use pacemaker 2. Inflammation of the kidneys is called rephritis 3. In the case of defibrillator a double square pulse type is used to restart the heart rhythm after the open heart surgery. 4. The common source of energy for pacemaker is the mercury battery. 5. In the case of stable total AV block a pacemaker is chosen with constant frequency. 6. After the chest operation the patient feels difficult to breathe. Then the patient is connected to a Ventilator 7. Two low a blood pressure is known as hypotension 8. The normal pH of the blood is 7.4 9. The obstruction of blood flow is known as stasis 10. Blood flow can be measured using the electromagnetic principles because blood has a high electrical conductivity. Unit 5 1. To avoid electrode polarization and bio potential artifacts electromagnetic blood flowmeters are using a.c magnetic fields 2. Blood plasma is obtained by centrifuging blood that has been prevented from coagulating 3. Fluoroscopic observation of cardiac catheterization is made by X-ray imaging 4. In Biotelemetry FDM refers to frequency division multiplexing 5. The radiocapsules are biotelemetry transmitter 6. The term bionics means a science concerned with the application of data about the functioning of biological systems to the solution of engineering problems. 7. Let-go current of men increases with the log frequency 8. Microshocks may occur due to flow of few mA current across the cardiac muscles 9. Let-go current are higher for men than women 10.Ground fault oriented electric shocks can be avoided by using isolated power supplies. PART - B UNIT –I 1. Explain cell and its structure Cell Structure The cell is considered to be the smallest structure in biology that has all the properties of living things and an understanding of cells and the basics of cell structure and function is critical to making sense out of biology. All cells at their essence have at least three things in common: Cell Membrane : All cells have a phospholipid based cell membrane. The cell membrane is selectively permeable in that it allows some materials to pass into or out of the cell but not others. Cytoplasm : Cells are filled with a complex collection of of substances in a water based solution. This substance is called cytoplasm. Across all cells there are a number of common features to all cell cytoplasm. For example all cells have ribosome . Also, in all cells the first steps in cellular respiration take place in the cytoplasm. DNA : All cells contain DNA. In the simplest cells, the DNA is in one loop more loop like structures free in the cytoplasm. In some cells such as those making up our body the DNA is isolated from the cytoplasm in a special structure called a nucleus. 2. Write about the transport of ions through cell The cell is bound by an outer membranethat, in accord with the fluid mosaic model, is comprised of a phospholipid lipid bilayer with proteins— molecules that also act as receptor sites—interspersed within the phospholipid bilayerThere are three principal mechanisms of outer cellular membrane transport (i.e., means by which molecules can pass through the boundary cellular membrane). The transport mechanisms are passive, or gradient diffusion, facilitated diffusion, and active transport. Diffusion is a process in which the random motions of molecules or other particles result in a net movement from a region of high concentration to a region of lower concentration. Thus, if the concentration of the diffusing substance is very high at the source, and is diffusing in a direction where little or none is found, the diffusion rate will be maximized.. Facilitated diffusion is the diffusion of a substance not moving against a concentration gradient but which require the assistance of other molecules. These are not considered to be energetic reactions Transmembrane proteins establish pores through which ions and some small hydrophilic molecules are able to pass by diffusion. Active transport is movement of molecules across a cell membrane or membrane of a cell organelle, from a region of low concentration to a region of high concentration. Since these molecules are being moved against a concentration gradient, cellular energy is required for active transport. 3. Explain the purpose of electrode paste The dry oputer skin of the body is highly non-conductive and will not establish a good electrical contact with an el;ectrode. The skin should therefore be washed throughly and rubbed briskly to remove some of the outer cells. This area should then be coated with an electrically conductive paste called electrode paste that should be worked in by further rubbing. The electrode is hgen applied to the prepared site and held in place with a rubber strap. Thus the electrode paste decreases the impedance of the contact and it also reduces the artifacts resulting from movement of the electrode or patient. Generally the conductivity of the skin is directly proportional to moisture on the skin. Even after the application of electrode paste, the contact impedance decreases with he increase of frequency of the signal. Among alcohol, electrode paste, saline solution and multipoint electrode, multipoint electrode has less contact impedance about 4kiloohms and alcohol has greater contact impedance. 4. Describe Half-cell potential Half-cell potential corresponds to an equilibrium, in term of electrical current, between metal (iron of steels) dissolution and the associated reduction reaction, which leads to the formation of ions hydroxyls (OH- , from oxygen gas). It should be noted that a current is exchanged between dissolution zones and those of oxygen reduction. These zones can be numerous if metal has a large contact area with its surrounding medium. In concrete, dissolution zones are very often fixed and rather large An electric field (current line) connects these zones to those where oxygen is reduced. These current lines are normal on equipotential surfaces (for a given value of the potential), which can be intersected by the concrete surface. In other words, the potential value can be measured by using a reference electrode placed on the concrete surface The half-cell potential is a parameter used for assessing the reinforcement condition. 5. Write about Needle Electrodes? Any terminal by which an electric current passes in or out of a conducting substance; for example, the anode or a cathode in an electrolytic cell. The anode is the positive electrode and the cathode is the negative electrode. The terminals that emit and collect the flow of electrons in thermionic valves (electron tubes) are also called electrodes: for example, cathodes, plates (anodes), and grids. Needle electrodes are used to record the peripheral nerves action potentials. The needle electrode resembles a medicine dropper or hypodermic needle. Ashort length of the fine insulated metal wire is bent at its one end and the bent portion is inserted through the iumen of the needle and is advanced into the muscle. The needle is with drawn and the bent wire is resting inside the muscle. When the reference electrode is placed on the skin, then the needle electrode is called monopolar. When we insert two insulated wires into the lumen of the needle, then the two wires constitute bipolar electrode such that one wire is active electrode and the other wire is reference electrode. UNIT II 1. Give an introduction about bio potential recorders The bio potential recorder is playing an important role in the biomedical instrumentation. If the recorder is not faithful or the fidelity is poor, then there is improper diagnosis. Thus the recorders are very useful to interpret the nature of the biological source during its working condition. During 1887, Waller recorded the first electrocardiogram using capillary electrometer, immersion type electrodes and moving light sensitive paper. Then Einthoven, the father of electrocardiography, introduced the string galvanometer during 1903. These are further improved by introducing servomotors, ink jet recorders to extend the frequency response and signal recovery circuits 2. What are the characteristics of recording system? Sensitivity of the recording system is the magnitude of input voltage required to produce a standard deflection in the recorded trace. The sensitivity is composed of gain of the amplification system and sensitivity of writer in the recorder. There are provisions to change the sensitivity. A recorder is said to be linear if the pen deflection is proportional o amplitude of the input signal. The range over which the system is linear is called the dynamic range. Beyond the dynamic range, the nonlinear distortion occurs. Non linearity can occur separately in the amplifier and writer. Frequency response of the recorder is the frequency characteristics of the system such that the variation of the sensitivity for different frequency components of the biosignal. The interval between the turn over frequencies is the bandwidth of the system. A low noise is maintained by special circuits and use of quality electronic components 3. Explain about writer and pen damping effects Sensitivity of the writer with respect to different frequencies The writer is usually a pen motor and is electromagnetic system of moving coil or moving iron type. The moving coils type has he pen attached to the coil, carrying current and so behaves like a magnet whose strength is proportional to the current. The coil rotates in the field of permanent magnet and the angle of rotation is proportional to the current in case of moving iron type of write. The sensitivity of the writer is dependent on frequency because of inertia, mechanical resonance and friction. Many writers are constructed so that the pen arm moves in an arc. This introduces both amplitude and timing errors which increase in nonlinear manner with the deflection and cause some distortion of he wave shape. 4. Explain the origin of electrocardiography Electrocardiography (ECG or EKG) is the recording of the electrical activity of the heart over time via skin electrodes. It is a noninvasive recording produced by an electrocardiographic device. The etymology of the word is derived from electro, because it is related to electrical activity, cardio, Greek for heart, graph, a Greek root meaning "to write". The electrocardiography deals with he study of electrical activity of the heart muscles. The potential originated in the individual fibers of heart muscle are added to produce the ECG waveform. The electro cardiogram is the recorded ECG waveform. It reflects the rhythmic electrical depolarization and repolarization of the myocardium associated with the contractions of the atria and ventricles. Any form of arrhythmia can be easily diagnosed using electrocardiogram ECG WAVEFORM 5. Explain about vectorcardiography In case of electrocardiography only the voltage generated by he electrical activity of the heart is recorded. But in vectorcardiography, the cardiac vector is displayed along with its magnitude and spatial orientation. Even though the cardiac vector ia a three dimensional, its projection on orthogonal planes converts into two dimensional vector. By means of orthogonal lead system, the spatial relations of he cardiac dipole vector are usually displayed on cathode-ray oscilloscope. Thisaccomplished by resolving the signal into three images, corresponding to frontal, sagittal and transverse planes. The vector cardiogram appears as loops in each plane. The iso-electric line in the standard lead system is represented by the endpoint of the vectors in the vector cardiogram . With the modern computer facilities, the vectorcardiography may be clinically used in an extensive manner, even though it is not used clinically now. UNIT – 3 1. Explain the origin of EEG Electroencephalography deals with the study of electrical activities of the brain. Initially it was thought that brain waves represent a summation of the action potentials of the neurons in the brain. It is believed that the electrical patterns obtained on the surface of the skull are the result of graded potentials on the dendrites of neurons in the cerebral cortex and other parts of the brain. Electrical charges are transferred between one nerve fiber and the other through a dendrite te of a post synaptic neuron during the release of acetylcholine. Action potentials of the brain Progressive transient disturbance of the resting potential along a nerve fiber is used to transmit information from one end to the other. This action potential is caused by very rapid change of membrane permeability to sodium ions followed by a recovery period. Evoked potentials Evoked potentials are the potentials developed in the brain as the response to external stimuli like light sound etc. The external stimuli are detected by the sense organs which cause changes in the electrical activity of the brain. Good recording techniques are to be used for evoked potential studies 2. Explain with block diagram EMG recording unit The surface electrodes or needle electrodes pick up potentials produced by certain muscle fibers. The surface electrodes are from Ag-AgCl are in disc shape. The surface of the skin is cleared and electrode paste is applied. The electrodes are kept in place by means of elastic bands. There are two types of conventional electrodes: bipolar and unipolar type electrodes. In case of bipolar electrode, the potential difference between two surface electrodes resting on the skin is measured. In case of unipolar electrode, the reference electrode is placed on the skin and needle electrode which acts as active electrode is inserted into the muscle The needle is inserted into the muscle. Further to record the action potentials from a nerve fiber, microelectrodes are used The amplitude of EMG signals depends upon the type and placement of electrodes used and the degree of muscle exertions. That is, the surface electrode picks up many overlapping spikes and produces an average voltage from various muscles and motor units. The needle electrode picks up the voltage from the single muscle fiber. Thus EMG is useful for studying neuro muscular function, neuron muscular condition, reflex response and extent of nerve lesion and diagnosing muscular disease. 3. Explain with block diagram ERG recording unit The recording and interpreting the electrical activity of eye is called electroretinography. All sense organs are connected to the brain but the eye has a special relation ship as the retina is an extension of cerebral cortex. Potentials within the eye may be recorded relatively easily because of its exposed position. The corona is about 20 mv positive relative to the fundus of the eye. The fundus is back to the interior of the eyeball. If the illumination of the retina is changed, the potential changes slightly in a complex manner. The recording of these changes is called electroretinogram. Electrode placement : A bipolar recording technique is used. The exploring electrode is placed on a saline filled contact lens. The contact lens is tightly attached to an eye. During eye movement there is no slip of contact lens by using negative pressure attachment techniques. Specially made contact lenses are used to record the action potentials of eye during flash of light incident on the eye. Recording Techniques: When light falls on retina he absorption of photons by photo pigments localized in the outer segment of the retinas photoreceptor is taking place, This causes the breakdown of photo pigments which results in liberation of ions. This in turn results in the development of action potential that is transmitted down the optic nerve. This action potential is picked up. 4. Explain with block diagram EOG recording unit A record of corneal-retinal potentials associated with eye movements is called electrooculogram. This is more simpler to record than the electroretinogram. The electrodes are to be placed. One pair of disc like skin electrodes on either side of eye for recording of horizontal movement of the eyes and another pair of electrodes on the forehead and cheeks for recording of vertical movement of eyes. The output from these two pairs are given separately to the preamplifiers and then are recorded. During recording, the eye rhythmically rotated as the vision shifts between two fixed points Some disease which affect the steady potential of the eye can be studied using EOG: The effects of certain drugs on the eye movemebnt system can b e determned. The state of the semicircular canals is analysed by EOG Diagnosis of the neurological disorders may be possible The level of anesthesiacan be indicated by characteristic eye movements 5. What are the special applications of phonocardiogram Fetal phonocardiogram : A stethoscope microphone with a large chest piece is over that part of the maternal abdomen where auscultation reveals fetal heart sounds. Simultaneously with fetal sound tracing maternal ECG is recorded for comparison Esophageal phonocardiogram : In general, sounds and murmurs have lower frequencies than when recorded by conventional techniques. The heart sounds are with shorter duration. Tracheal phonocardiogram : Tracheal phonocardiogram have been recorded in patients by means of a tracheal canella. The technique of connecting the outer end of the cannula with a microphone by means of short piece of rubber tube. The heart sounds are with shorter duration and have W UNIT – 4 1. Give an introduction about pacemakers Pacemaker is an electrical pulse generator for starting and or maintaining the normal heartbeat. The output of he pacemaker is applied either externally to the chest or internally to he heart muscle. In the case of cardiac stand still, the use of pacemaker is temporary – just long enough to start a normal rhythm, In cases requiring long term pacing, the pacemaker is surgically implanted in the body and its electrodes are in direct contact with the heart. In cardiac diseases, where the ventricular rate is too own, it can be increased to normal rate by using pacemaker. The various arrhythmias that result in heart both and Adams stokes attacks represent a serious pathological condition. During that time, the patient becomes invalid because of the constant risk of sudden losing consciousness. By fixing artificial electronic pacemakers, he above defects in he heart can be eliminated. With conventional drug therapy, the failure within a year is 50%. Pacemaker’s therapy lowers to 15% and leads to a considerable improvement in the patient mental and physical well-being 2. What are the methods of stimulation There are two types of simulation or pacing : External stimulation and internal stimulation. External stimulation: It is employed to restart the normal rhythm of the heart in case of cardiac stand still. Stand still can occur during open-heart surgery or whenever here is a sudden physical shock or accident. Attaché paddle shaped electrodes are applied on he surface of the chest and the current in the range of 20 – 150 Ma are employed. Internal stimulation It is employed in cases during long term pacing because of permanent damage that prevents normal self-triggering of the heart. In some cases during restarting of the heart after open-heart surgery, spoon like electrodes are used. The e current range of 2-25 Ma is employed. The bipolar and unipolar electrodes are used. In the bipolar electrode there are stimulating electrode and contact electrode which serves as a return path for the current to the pacemakers. In the unipolar electrode, there is only stimulating electrode and the return path for current to the pacemaker is made through body fluids 3. What are the differences between external and internal pacemakers? External Pacemakers : The pacemaker is placed outside the body. It may be in the form of wrist watch or in the pocket, from that one wire will go into the heart through the vein The electrodes are called end cardiac electrodes and are applied to the heart by means of electrode c catheter with electrode tip situated in the apex of right ventricle. It does need necessitate he open chest surgery The battery can be easily replaced During placement, swelling and pain don’t arise due to minimal foreign body reaction They are used for temporary heart irregularities. Internal pacemakers : The pacemaker is miniaturized and is surgically implanted beneath the skin near the chest or abdomen with it output leads are connected direcly to the heart muscle The electrodes are called myocardial electrodes and are in contact with the outer wall of myocardium It requires an open chest minor surgery to place the circuit The battery can be replaced only after minor surgery During placement, swelling and pain arise due to foreign body reaction They are used for permanent damages 4. What are the different types of artificial heart valves Today cardiac surgery has advanced tremendously to enable repair of all congenital heart defects, replacement of damaged valves and bypassing coronary block. About 1 in 1000 babies is born with deformed heart. The imperfect functioning of heart is due to various defects like holes in the heart, atrial septal defects and ventricular septal defects. In several cases like pulmonary stenos is, the valves of the major vessels are narrow. When the major vessels of the heart are wrongly connected then there is mixing of pure and impure blood. This accounts for he blue co lour if the body There are two types of valves i) Prosthetic Valves : These are made from high grade plastics and metal. They need life long coagulating agent blood thinning agent. The average durability of these valves is about 8 to 10 years ii) Tissue valves These may be either homograph of hetero graft. Homograph valves are taken from human being. Heterograft valves aren taken fro animals. Here blood thinning agent is not necessary. 5. Give an introduction about Defibrillators A defibrillator is an electronic device that creates a sustained myocardial depolarization of a patients heart in order to stop ventricular fibrillation or atrial fibrillation. Ventricular fibrillation is a serious cardiac emergency resulting fro asynchronous contraction of heart muscle Defibrillation is the definitive treatment for the life-threatening cardiac arrhythmias, ventricular fibrillation and pulseless ventricular tachycardia Defibrillation consists of delivering a therapeutic dose of electrical energy to the affected heart with a device called a defibrillator. This depolarizes a critical mass of the heart muscle, terminates the arrhythmia, and allows normal sinus rhythm to be reestablished by the body's natural pacemaker in the sinoatrial node of the heart. Defibrillators can be external, transvenous, or implanted, depending on the type of device used or needed. Some external units, known as automated external defibrillators (AEDs), automate the diagnosis of treatable rhythms, meaning that lay responders or bystanders are able to use them successfully with little, or in some cases no training at all. UNIT –5 1.What is meant by biotelemetry system? Biotelemetry is an important technique for biomedical research and clinical medicine. Perhaps cardiovascular research and treatment have benefited the most from biotelemetry. Heart rate, blood flow, and blood pressure can be measured in ambulatory subjects and transmitted to a remote receiver-recorder. Telemetry also has been used to obtain data about local oxygen pressure on the surface of organs (and for studies of capillary exchange (that is, oxygen supply and discharge). Biomedical research with telemetry includes measuring cardiovascular performance during the weightlessness of space flight and portable monitoring of radioactive indicators as they are dispersed through the body by the blood vessels. Telemetry has been applied widely to animal research, for example, to record electroencephalograms, heart rates, heart muscle contractions, and respiration even from sleeping mammals and birds. Telemetry and video recording have been combined in research of the relationships between neural and cardiac activity and behavior. The use of telemetry methods for sending signals from a living organism over some distance to a receiver. Usually, biotelemetry is used for gathering data about the physiology, behavior, or location of the organism. Generally, the signals are carried by radio, light, or sound waves. Consequently, biotelemetry implies the absence of wires between the subject and receiver. 2.What are the considerations in designing a biotelemetry system The telemetering system should be selected to transmit he bioelectric signals with maximum fidelity and simplicity There would not be any constraint for living system due to these telemetry systems and there would not be any reaction The size and weight of the telemetry system should be small. In he case of long term units or implant units, the weight and size limit is in the order of 1% of th subject It should have more stability and reliability The power consumption should be very small to extend the source life time in case of implanted units The miniaturized radio telemetry system should be used to reduce noises 3.Explain about surgical diathermy Diathermy is the treatment process by which cutting coagulation etc of tissues are obtained. It is found hat when high frequency current in the range 1-3mhz is applied, heating of tissues takes place. The evoking steam bubbles in the tissues at the surgical tip continuously rupture the tissue and by that way the cutting action is obtained. Similarly during the passage of the high frequency current through the tissue, the tissue is heated locally. So that the tissue is melted instantaneously and sealing of capillary and other blood vessels is taking place. Thus the coagulation of the tissue takes place. The us of high frequency current is to avoid he intense muscle activity and the electrocution hazard, which occurs if low frequencies are used. 4.Explain about Microwave diathermy Here the frequency used is 2,450mhz corresponding to the wavelength of 12.25 cm Heating of the tissues is produces due o absorption of microwave energy. Better therapeutic results are obtained by using microwave diathermy than short wave diathermy. Here there is no pad shaped electrode. Instead the microwaves are transmitted into the portion of the body to be treated directly from the director of the unit. Normally magnetrons are used to produce microwaves A delay of about 3 or 4 minute is required for the warming of the magnetron. Arrangement is made such that a lamp lights up after 4 minutes indicating the magnetron is ready to deliver its output to the director. Proper cooling of the magnetron is provided. 5.What are the various electrosurgery techniques? Explain Fulguration : By passing sparks from a needle or ball electrode of small diameter to the tissue, the developed heat dries out the superficial tissue without affecting the deep seated tissues. This is called fulguration in which the electrode is held near the tissue without touching it and due to the passage of the electric arc, the destruction of the superficial tissue takes place Desiccation : The needle point electrodes are struck into the tissue and kept steady while passing electric current. This creates a high local increase in heat and drying of tissues takes place Electrotonmy : When the electrode is kept above the skin electrical arc is sent. The developed heat produces a wedge shaped narrow cutting of the tissue on the surface Coagulation: When the electrode is kept near the skin high frequency current is sent through the tissue in the form of burst and heating it locally so that it coagulates from inside Blending: When the electrode is kept above the skin the separated tissues or nerves can be welded or combined together by an electric arc PART-C UNIT -1 1. Explain in detail the chemical electrodes Hydrogen Electrode The standard hydrogen electrode (abbreviated SHE), also called normal hydrogen electrode (NHE), is a redox electrode which forms the basis of the thermodynamic scale of oxidation-reduction potentials. Its absolute electrode potential is estimated to be 4.44 ± 0.02 V at 25 °C, but to form a basis for comparison with all other electrode reactions, hydrogen's standard electrode potential (E0) is declared to be zero at all temperatures. Potentials of any other electrodes are compared with that of the standard hydrogen electrode at the same temperature. Hydrogen electrode is based on the redox half cell 2H+(aq) + 2e- → H2(g) This redox reaction occurs at platinized platinum electrode. The electrode is dipped in an acidic solution and pure hydrogen gas is bubbled through it. The concentration of both the reduced form and oxidised form is maintained at unity. That implies that the pressure of hydrogen gas is 1 bar and the concentration of hydrogen in the solution is 1 molar Reference electrode: The silver/silver chloride or Ag/AgCl reference electrode is many electrochemists' reference electrode of choice. It is easily and cheaply prepared. It is stable, and quite robust. It is sometimes referred to as "SSCE" (Silver/Silver Chloride Electrode) l The body of the electrode is made from 4 mm glass tube. Vycor porous glass is available in 4 mm diameter rod and serves as the ironically conducting electrical pathway between the inside of the reference electrode and the bulk of your cell. It has low electrical resistance (under 10 kohl for the common filling solutions) and a modest leak rate. The electrical resistance of the reference electrode 'frit' is an important factor in determining the stability . The leak rate may be important because of possible contamination of your solution by the reference electrode filling solution and vice versa. The Vycor frit (about 1/8" long) is attached to the glass tube by 'heat shrink' Teflon tubing. The heat-shrink tubing should be cut flush with the end of the Vycor frit to prevent trapping any air bubbles.. The cap is conveniently made out of scrap Teflon or plastic cap or protector made to fit 5/32" OD tubing. It should be snug, but easily removable for replenishing the filling solution. Saturated solutions of KCl or NaCl have the advantage that the concentration is reproducible even if the temperature changes (if solid salt is present) and are immune to the effects of water evaporation. However, the solid salts harden into an impenetrable block which may lead to a high impedance electrode. A "nearly saturated" solution (3.5M KCl or 3M NaCl) can change concentration due to evaporation. Glass pH electrode Most often used pH electrodes are glass electrodes. Typical model is made of glass tube ended with small glass bubble. Inside of the electrode is usually filled with buffered solution of chlorides in which silver wire covered with silver chloride is immersed. pH of internal solution varies - for example it can be 1.0 (0.1M HCl) or7.0 (different buffers used by different producers). Active part of the electrode is the glass bubble. While tube has strong and thick walls, bubble is made to be as thin as possible. Surface of the glass is protonated by both internal and external solution till equilibrium is achieved. Both sides of the glass are charged by the adsorbed protons, this charge is responsible for potential difference. This potential in turn is described by the Nernst equation and is directly proportional to the pH difference between solutions on both sides of the glass The majority of pH electrodes available commercially are combination electrodes that have both glass H+ ion sensitive electrode and additional reference electrode conveniently placed in one housing. For some specific applications separate pH electrodes and reference electrodes are still used - they allow higher precision needed sometimes for research purposes. 2. Explain in detail about Resting and Action Membrane Potentials Resting Potential There is normally a charge difference across the plasma membrane of a neuron. The outside of the membrane has a positive charge. The inside has a negative charge. Resting potential results from differences between sodium and potassium positively charged ions and negatively charged ions in the cytoplasm. Sodium ions are more concentrated outside the membrane, while potassium ions are more concentrated inside the membrane. This imbalance is maintained by the active transport of ions across the membrane known as the sodium potassium pump. This difference or potential is measured in mill volts. The resting potential is usually about -70mv. Action Potential When a neuron receives signals, an abrupt, temporary reversal in the polarity is generated (an action potential). The inside becomes more positive. Any membrane that can produce action potentials is said to show membrane excitability. Voltage change causes voltage-gated channels in the membrane to open. As a result of ion flow through these channels, the inside of a neuron briefly becomes more positive than outside. Two properties of the neuron membrane permit a resting potential: The lipid bilayer bars the free passage of potassium ions and sodium ions. Ions can flow from one side to the other through channels in transport proteins. There are more potassium ions inside and more sodium ions outside the resting neuron membrane. Potassium ions have a tendency to leak out by facilitated diffusion through channel proteins. Most of the sodium channels are "gated" and remain closed most of the time, keeping the concentration outside high. However, small amounts of sodium do leak in and must be pumped out (and potassium pumped in) by the sodium-potassium pump. When a stimulus reaches a certain minimum, a threshold-gated channels open and sodium rushes in. In an accelerating way, more and more gates open (At threshold, the opening of more gates no longer depends on the stimulus but is self-propagating All action potentials are the same size. If stimulation is below threshold level, no action potential occurs. If it is above threshold level, the cell is always depolarized to the same level. Polarization, depolarization,hyperpolarization: Polarization literally means having poles. Because a cell is negative inside (as a result of being dominated by the electrochemical equilibrium potential of potassium) it is said to be polarized. Increasing that polarization (i.e. making is more negative inside) is a hyper polarization and decreasing it (i.e. making it less negative inside) is a depolarization. Repolarization : When depolarization in one region is ended, the sodium gates close and potassium gates open. The sodium-potassium membrane pumps also become operational to fully restore the resting potential. Movement of potassium out of the cell repolarizes the cell. The action potential is self-propagating and moves away from the stimulation site to adjacent regions of the membrane undiminished. A brief (refractory) period follows at each depolarization site. Sodium gates shut and potassium gates open. During this period the membrane is insensitive to stimulation. 3. Explain about nerve tissues and organs Nervous tissue is the main component of the nervous system-the brain, spinal cord, and nerves-which regulates and controls body functions. It is composed of neurons which transmit impulses, and the neuroglia which assist propagation of the nerve impulse as well as provide nutrients to the neuron. Every time you get pinched, part of your nerve tissue is damaged. All nervous tissue of a living organism makes up its nervous system which includes the brain, spinal cord, and nerves throughout the organism. Nervous tissue is made of nerve cells that come in many varieties, all of which are distinctly characteristic by the axon or long stem like part of the cell that sends action potential signals to the next cell. All living cells have the ability to react to stimuli. Nervous tissue is specialized to react to stimuli and to conduct impulses to various organs in the body which bring about a response to the stimulus. Nerve tissue (as in the brain, spinal cord and peripheral nerves that branch throughout the body) are all made up of specialized nerve cells called neurons. Neurons are easily stimulated and transmit impulses very rapidly. A nerve is made up of many nerve cell fibers (neurons) bound together by connective tissue. A sheath of dense connective tissue, the epineurium surrounds the nerve. This sheath penetrates the nerve to form the perineurium which surrounds bundles of nerve fibers. Blood vessels of various sizes can be seen in the epineurium.. The cell body is enclosed by a cell (plasma) membrane and has a central nucleus. Granules called Nissl bodies are found in the cytoplasm of the cell body. Within the cell body, extremely fine neurofibrils extend from the dendrites into the axon. The axon is surrounded by the myelin sheath, which forms a whitish, noncellular, fatty layer around the axon. Outside the myelin sheath is a cellular layer called the neurilemma or sheath of Schwann cells. The myelin sheath together with the neurilemma is also known as the medullary sheath. This medullary sheath is interrupted at intervals by the nodes of Ranvier. structure of a nerve cell. The nerve impulse enters the nerve cell by the dendrites and is then carried away from the cell body by the axon Tissues : Cells group together in the body to form tissues - a collection of similar cells that group together to perform a specialized function. There are 4 primary tissue types in the human body: epithelial tissue, connective tissue, muscle tissue and nerve tissue. 1. Epithelial Tissue - The cells of epithelial tissue pack tightly together and form continuous sheets that serve as linings in different parts of the body. Epithelial tissue serve as membranes lining organs and helping to keep the body's organs separate, in place and protected. Some examples of epithelial tissue are the outer layer of the skin, the inside of the mouth and stomach, and the tissue surrounding the body's organs. 2. Connective Tissue - There are many types of connective tissue in the body. Generally speaking, connective tissue adds support and structure to the body. Most types of connective tissue contain fibrous strands of the protein collagen that add strength to connective tissue. Some examples of connective tissue include the inner layers of skin, tendons, ligaments, cartilage, bone and fat tissue. In addition to these more recognizable forms of connective tissue, blood is also considered a form of connective tissue. 3. Muscle Tissue - Muscle tissue is a specialized tissue that can contract. Muscle tissue contains the specialized proteins actin and myosin that slide past one another and allow movement. Examples of muscle tissue are contained in the muscles throughout your body. 4. Nerve Tissue - Nerve tissue contains two types of cells: neurons and glial cells. Nerve tissue has the ability to generate and conduct electrical signals in the body. These electrical messages are managed by nerve tissue in the brain and transmitted down the spinal cord to the body. Organs : Organs are the next level of organization in the body. An organ is a structure that contains at least two different types of tissue functioning together for a common purpose. There are many different organs in the body: the liver, kidneys, heart, even your skin is an organ. In fact, the skin is the largest organ in the human body and provides us with an excellent example for explanation purposes. The skin is composed of three layers: the epidermis, dermis and subcutaneous layer. The epidermis is the outermost layer of skin. It consists of epithelial tissue in which the cells are tightly packed together providing a barrier between the inside of the body and the outside world. Below the epidermis lies a layer of connective tissue called the dermis. In addition to providing support for the skin, the dermis has many other purposes. The dermis contains blood vessels that nourish skin cells. It contains nerve tissue that provides feeling in the skin. And it contains muscle tissue that is responsible for giving you 'goosebumps' when you get cold or frightened. The subcutaneous layer is beneath the dermis and consists mainly of a type of connective tissue called adipose tissue. Adipose tissue is more commonly known as fat and it helps cushion the skin and provide protection from cold temperatures. UNIT –2 1. Explain in detail the ECG Lead configurations In electrocardiography, the word, "lead" (rhymes with 'speed') refers to the signal that goes between two electrodes These electrodes are attached to the patient's body, usually with very sticky circles of thick tape-like material ECG leads record the electrical signals of the heart from a particular combination of recording electrodes which are placed at specific points on the patient's body. Placement of electrodes Ten electrodes are used for a 12-lead ECG. They are labeled and placed on the patient's body as follows vs. bipolar leads There are two types of leads—unipolar and bipolar. Bipolar leads have one positive and one negative pole. In a 12-lead ECG, the limb leads (I, II and III) are bipolar leads. Unipolar leads have only one true pole (the positive pole). The negative pole is a "composite" pole made up of signals from lots of other electrodes. In a 12-lead ECG, all leads besides the limb leads are unipolar (aVR, aVL, aVF, V1, V2, V3, V4, V5, and V6). Limb leads In both the 5- and 12-lead configuration, leads I, II and III are called limb leads. The electrodes that form these signals are located on the limbs—one on each arm and one on the left leg. The limb leads form the points of what is known as Einthoven's triangle. Lead I is the signal between the (negative) aVR electrode (on the right arm) and the (positive) aVL electrode (on the left arm). Lead II is the signal between the (negative) aVR electrode (on the right arm) and the (positive) aVF electrode (on the left leg). Lead III is the signal between the (negative) aVL electrode (on the left arm) and the (positive) aVF electrode (on the left leg). Augmented limb leads Leads aVR, aVL, and aVF are 'augmented limb leads'. They are derived from the same three electrodes as leads I, II, and III. However, they view the heart from different angles because the negative electrode for these leads is a modification of 'Wilson's central terminal', which is derived by adding leads I, II, and III together and plugging them into the negative terminal of the ECG machine. This zeroes out the negative electrode and allows the positive electrode to become the "exploring electrode" or a unipolar lead. This is possible because Einthoven's Law states that I + (-II) + III = 0. The equation can also be written I + III = II. It is written this way (instead of I - II + III = 0) because Einthoven reversed the polarity of lead II in Einthoven's triangle, possibly because he liked to view upright QRS complexes. Wilson's central terminal paved the way for the development of the augmented limb leads aVR, aVL, aVF and the precordial leads V1, V2, V3, V4, V5, and V6. Lead aVR or "augmented vector right" has the positive electrode (white) on the right arm. The negative electrode is a combination of the left arm (black) electrode and the left leg (red) electrode, which "augments" the signal strength of the positive electrode on the right arm. Lead aVL or "augmented vector left" has the positive (black) electrode on the left arm. The negative electrode is a combination of the right arm (white) electrode and the left leg (red) electrode, which "augments" the signal strength of the positive electrode on the left arm. Lead aVF or "augmented vector foot" has the positive (red) electrode on the left leg. The negative electrode is a combination of the right arm (white) electrode and the left arm (black) electrode, which "augments" the signal of the positive electrode on the left leg. The augmented limb leads aVR, aVL, and aVF are amplified in this way because the signal is too small to be useful when the negative electrode is Wilson's central terminal. Together with leads I, II, and III, augmented limb leads aVR, aVL, and aVF form the basis of the hexaxial reference system which is used to calculate the heart's electrical axis in the frontal plane. In modern digital equipment, the augmented leads are derived from the limb leads by simple calculation: aVR = -(I + II)/2 aVL = I - II/2 aVF = II - I/2 Precordial leads The electrodes for the precordial leads (V1, V2, V3, V4, V5, and V6,) are placed directly on the chest. Because of their close proximity to the heart, they do not require augmentation. Wilson's central terminal is used for the negative electrode, and these leads are considered to be unipolar (recall that Wilson's central terminal is the average of the four limb leads.).The precordial leads view the heart's electrical activity in the so-called horizontal plane. The heart's electrical axis in the horizontal plane is referred to as the Z axis. Leads V1, V2, and V3 are referred to as the right precordial leads and V4, V5, and V6 are referred to as the left precordial leads. The QRS complex should be negative in lead V1 and positive in lead V6. The QRS complex should show a gradual transition from negative to positive between leads V2 and V4. The equiphasic lead is referred to as the transition lead. When the transition occurs earlier than lead V3, it is referred to as an early transition. When it occurs later than lead V3, it is referred to as a late transition. There should also be a gradual increase in the amplitude of the R wave between leads V1 and V4. This is known as R wave progression. Poor R wave progression is a nonspecific finding. It can be caused by conduction abnormalities, myocardial infarction, cardiomyopathy, and other pathological conditions. Ground An additional electrode (usually green) is present in modern four-lead and twelve-lead ECGs. This is the ground lead and is placed on the right leg by convention, although in theory it can be placed anywhere on the body. With a threelead ECG, when one dipole is viewed, the remaining lead becomes the ground lead by default. 2. Explain in dertail the ECG waveform analysis A typical ECG tracing of a normal heartbeat (or cardiac cycle) consists of a P wave, a QRS complex and a T wave.] A small U wave is normally visible in 50 to 75% of ECGs. The baseline voltage of the electrocardiogram is known as the isoelectric line. Typically the isoelectric line is measured as the portion of the tracing following the T wave and preceding the next P wave. The four deflections were originally named ABCDE but renamed PQRST after correction for artifacts introduced by early amplifiers. P wave During normal atrial depolarization, the main electrical vector is directed from the SA node towards the AV node, and spreads from the right atrium to the left atrium. This turns into the P wave on the ECG, which is upright in II, III, and aVF (since the general electrical activity is going toward the positive electrode in those leads), and inverted in aVR (since it is going away from the positive electrode for that lead). A P wave must be upright in leads II and aVF and inverted in lead aVR to designate a cardiac rhythm as Sinus Rhythm. The relationship between P waves and QRS complexes helps distinguish various cardiac arrhythmias. The shape and duration of the P waves may indicate atrial enlargement. Absence of the P wave may indicate atrial fibrillation QRS complex The QRS complex is a structure on the ECG that corresponds to the depolarization of the ventricles. Because the ventricles contain more muscle mass than the atria, the QRS complex is larger than the P wave. In addition, because the His/Purkinje system coordinates the depolarization of the ventricles, the QRS complex tends to look "spiked" rather than rounded due to the increase in conduction velocity. A normal QRS complex is 0.08 to 0.12 sec (80 to 120 ms) in duration represented by three small squares or less, but any abnormality of conduction takes longer, and causes widened QRS complexes. Not every QRS complex contains a Q wave, an R wave, and an S wave. By convention, any combination of these waves can be referred to as a QRS complex. However, correct interpretation of difficult ECGs requires exact labeling of the various waves. Some authors use lowercase and capital letters, depending on the relative size of each wave. For example, an Rs complex would be positively deflected, while a rS complex would be negatively deflected. If both complexes were labeled RS, it would be impossible to appreciate this distinction without viewing the actual ECG. The duration, amplitude, and morphology of the QRS complex is useful in diagnosing cardiac arrhythmias conduction abnormalities, ventricular hypertrophy myocardial infarction electrolyte derangements, and other disease states. Q waves can be normal (physiological) or pathological. Pathological Q waves refer to Q waves that have a height of 25% or more than that of the partner R wave and/or have a width of greater than 0.04 seconds. Normal Q waves, when present, represent depolarization of the interventricular septum. For this reason, they are referred to as septal Q waves, and can be appreciated in the lateral leads I, aVL, V5 and V6. Q waves greater than 1/4 the height of the R wave, greater than 0.04 sec (40 ms) in duration, or in the right precordial leads are considered to be abnormal, and may represent myocardial infarction "Buried" inside the QRS wave is the atrial repolarization wave, which resembles an inverse P wave. It is far smaller in magnitude than the QRS and is therefore obscured by it. ST segment The ST segment connects the QRS complex and the T wave and has a duration of 0.08 to 0.12 sec (80 to 120 ms). It starts at the J point (junction between the QRS complex and ST segment) and ends at the beginning of the T wave. However, since it is usually difficult to determine exactly where the ST segment ends and the T wave begins, the relationship between the RT segment and T wave should be examined together. The typical ST segment duration is usually around 0.08 sec (80 ms). It should be essentially level with the PR and TP segment. QT interval The QT interval is measured from the beginning of the QRS complex to the end of the T wave. Normal values for the QT interval are between 0.30 and 0.44 secondsThe QT interval as well as the corrected QT interval are important in the diagnosis of long QT syndrome and short QT syndrome Long QT intervals may also be induced by antiarrythmic agents that block potassium channels in the cardiac myocyTE The QT interval varies based on the heart rate, and various correction factors have been developed to correct the QT interval for the heart rate. The QT interval represents on an ECG the total time needed for the ventricles to depolarize and repolarize. The most commonly used method for correcting the QT interval for rate is the one formulated by Bazett and published in 1920Bazett's formula is, where QTc is the QT interval corrected for rate, and RR is the interval from the onset of one QRS complex to the onset of the next QRS complex, measured in seconds. However, this formula tends to be inaccurate, and over-corrects at high heart rates and undercorrects at low heart rates. 2. Explain wih block diagrm ECG Recording unit The important parts of ECG recorder are 1. Patient cable and Defibrillator protection circuit : Fibrillation circuit. It consists of buffer amplifiers and over voltage protection circuit. The leads are connected with the buffer amplifier such that one buffer amplifier for each patient lead. This reduces means the input impedanc is in creased and the effects arising from the variations in the electrode impedance. Further over voltage protection circuit is necessary to avoid any damage to the bioamplifiers in the recorder. The over voltage protection circuits consists of a network of resistors and neon lamps which fire when a pulse from a defibrillator is present 2. Lead selector switch: After the defibrillator protection circuit, there is lead selector, which is used to feed the input voltage from the appropriate electrode to the preamplifier. .3. Calibrator: A push button allows the insertion of standardization voltage of 1mv to the preamplifier. This enables the technician to observe the output on display unit and adjust the scale so that a known own deflection corresponds to a I mv input signal. From the leads selector switch the ECG signal goes to bioamplifier. 3. Bio-Amplifier : The bio-amplifier consists of preamplifier and power amplifier. The sensitivity or the gain of the amplifier can be varied. Followed by the preamplifier there is power amplifier, which is used to drive the recorder. Power amplifiers are required with high power gain. To avoid the cross over distortion in a push pull amplifier, an ideal no inverting amplifier is inserted on the input. The offset control is provided by resistance and is used to position the output stylus pen. 4. Auxiliary amplifier : Since the electrode impedance are not equal, a differential amplifier doesn’t completely reject the common mode signals. The common mode signals can be reduced to a minimum level by adding an auxiliary namplifier between the driven right leg lead and the ECG unit. 5. Isolated power supply : The isolated power supply is used to give power to the bioamplifier and by means that, the electrical safety for the patient is increased. 6. Output Unit : The output unit is cathode ray oscilloscope or a paper chart recorder. In the case of paper chart recorder, the power or pen amplifier supplies required power to drive pen motor. A position control on the pen amplifier is used to position the pen at the center ion recording paper. Power switch : The power switch of the recorder has three positions. ON,OFF and RUN UNIT - III 1. Explain in detail the placement of electrodes inn EEG In EEG electrodes are placed in standard locations on the skull in an arrangement called 10-20 system, a placement scheme devised by the international federation of societies of EEG i) Draw a line on the skull from the nasion, the root of the nose, to the inion, ossification center on the occipital lope ii) Draw a similar luine from the left preauricular point to the right preauricular point iii) Mark the intersection of these two lines which is the mid point of the distance between nasion and anion. iv) Mark points atb10,20,20,20,20, and 10%of the total nasion-inion distance v) Mark points at 10,20,20,20,20 and 10/% of the total distance between the preauricular points Before placing the electrodes, the scalp s cleaned, tightly abraded and electrode paste is applied between electrode and skin. Both bipolar and un bipolar electrod systems are used to facilitate location of foci, that is cortical areas from which abnormal waves spread. The phase relationship of the waves indicates the position of the focus and in sometimes it enables the velocity at which the waves spread to be calculated. Aian bipolar technique the difference in potential between two adjacent electrodes is measured. In the monopolar technique the potential of each electrode is measured with respect to he reference electrode attached to the ear lobe 2. Explain in the the recording set up of EEG unit The patient cable contains of 21 electrodes and is connected to the eight channel selector. The electrodes are attached to the channel selector in groups og eight called montage of electrodes. The right ear electrode acts as referenc electrode for the righ brain electrodes and the left ear electrode acts as reference electrode for the left brain electrodes. The 50hz interference is reduced by employing differential amplifiers and preamplifiers with more than 80db CMRR and ny use of 50 hz notch filters. The effect of notch filter in signal distortion is not so much because important EEG signals have frequencies below 30hz. . The 50hz interference is reduced by employing differential amplifiers and preamplifiers with more than 80db CMRR and ny use of 50 hz notch filters. The effect of notch filter in signal distortion is not so much because important EEG signals have frequencies below 30hz. Because the source of brain wave has high internal impedance, th input impedance of the preamplifier should be more to prevent reducion of signal amplitude. Further by cascading the gain of the amplifier is increased. Ahe output vbvoltage fro the preamplifier may either be applied directly to the eight channel display through filter bank or it may be stored in data on a tape recorder or in computer memory for further processing. 3. Explain in detail Phonocardiography The graphic record of heart sounds is calle phonogram. Because the sound is from the heart it is called phonocardiogram. The instrument used to measure the heart sounds is called phonocardiogram. This instrument uses a phonocather, a device similar to a convential catheter, with a microphone at the tip. The basic aim of phonocardiogram is to pick up different heart sounds, filter out the heart sounds and to display them or record them. Heart bsounds sare acustoic events of heart can be divided into two categories heart sounds amd murmurs. Heart sounds have a transient character and are of short dration. Heart murmurs have a noisy characteristic and last fo longer time Heart sounds : Heart sound are clasified into four types i) Valve closure sounds ii) Ventricullar filling sounds iii) Valve openiong sounds iv) Extra cardiac sounds Valve closure sounds: These sounds occur at he beginning of the systolem and the beginning of the diastole. The first heart sound is due to clousre of mitral and trcuspid valves. The second heart sound is due to clousre of arotivc pulmonary valves Ventricullar filling sounds: These sounds occur either at the period of rapid filling of the ventricles or during terminal phas of ventricular filling. These sounds are normally inaudiable Valve openiong sounds: These sounds occur at the time of opening of the artiventricvular valves and semilunaar valved Extra cardiac sounds: These sounds occur in mid or late systole or early systole and are believed to causeby thickened pericardium Heart sounds and murmurs are characterised by hre physical ,properties they are 1. Frequency 2. Amplitude 3. Quality UNIT –4 1. Explain in detail about square wave and double square wave defibrillators Square wave defibrillators : Here the capacitor is discharged through the subject on a series silicon controlled rectifier. When sufficient energy has been delivered to the subject a shunt SCR short circuits the capacitor and terminates the pulse. The output can be controlled by varying he voltage on the capacitor or duration of discharge. Here the defibrillation is obtained at less peak current and so there is no side effect. Digital circuits can also produce a square pulse used for defibrillation Double Square wave defibrillators: Double square pulse defibillator is used normally after the open heart surgery. Conventional d.c and a,.c defibrillators are producing myocardial injury with a diminished ventricular function for a period of approximately 30 minbutes. I f he chest is open only lower energy electric shock should be given. When the first pulse is delivered some of he fibrillating cells will be excitable and will be depolarised. In order to obtain oal defibrillation the second pulse operates on this later gyoup of cells. Using double square pulse defibrillator, efficie and quick recovery of heart norml manner without any sid effect. 2, Explain the working of heartlung machine During open heart surgery for installation of valla prosthesis or correction of a congenital mal information, the heart cannot maintain th circulation. It is then necessary o provide extra-corporeal circulation with a special machine call heart lung machine. The Heart lung machine replaces the functions of heart and lungs there by providing the rest of the body with a continuous supply of oxygenated blood while the heart is stopped In the lungs the blood is oxygenated to about 95 to 98% saturation and that it is going to hr left atrium of heart which acts as a receiver chamber From there the blood flows into the left ventricle through the mistrals valve i) In an intact heart venous or unoxygenated blood is returned to the right side of the heart ii) From the right side of the heart, the blood is pumped into lungs through pulmonary artery iii) In the lungs the blood is oxygenated to about 95 to 98% saturation and that it is going to hr left atrium of heart which acts as a receiver chamber iv) From there the blood flows into the left ventricle through the mistrals valve v) The ventricle ejects the blood into the aorta with peak pressures ranging from 100 to 150 mm Hg vi) The heart pumps about 5 liters of blood per lire. At any time the veins contain 75 o 80 % of blood volume and arteries contain 20 to 25 percents of blood volume 3. Explain with diagram Verntricular synchronous pacemaker and Ventricular inhibited pacemaker Verntricular synchronous pacemaker Patients with only short periods of AV block or bundle block can be supplied with a ventricular synchronous pacemaker. This type of pacemaker does not compete with the normal heart activity. A single transverse electrode placed in the right ventricle both senses the R wave and delivers the stimulation, thus no separate sensing electrode is required. A Rwave from an arterial generated ventricular contraction triggers the ventricular synchronized pacemaker, which provides an impulse falling in the lowest part of normal QRS complex. This ensures that the pacemaker does not interfere with the sinus rhythm. If a trial generated ventricular contractions are absent, the pacemaker provides impulse at a basic frequency 0f 70impulse/minute. Thus it provides impulses only when the atrial generated ventricular contractions are absent, thus conveying energy Using he sensing electrode, the heart rate is detected and is given to the timing circuit in the pacemaker. If the detected heart rate is below a certain minimum level, the fixed rate pacemaker is turned on to pace he heart. The lead used to detect R wave is now used to stimulate the heart. If a natural contraction occurs, the asynchronous pacemaker timing circuit is reset so that it will time its next pulse to detect heart beat. Otherwise the asynchronous pacemaker produces pulses at its preset rate Ventricular inhibited pacemaker The R wave inhibited pacemaker also allows the heart to pace at its normal rhyhm when it is able to do so. If the R wave is missing for a preset period of time, the pacer will supply the stimulus. Therefore if the heart rate falls below a predetermined value, the pacemaker will turn on and provide the heart a stimulus. The sensing electrode picks up the R wave The refractory circuit provides a period of time following a output pulse or sensed R wave during which the amplifier in the sensing circuit will no respond to outside signals. The sensing circuit detects the R wave and resets the oscillator. The reversion circuit allows the amplifier to detect R wave in low level signal to noise ratio. In the absence of R wave it allows the oscillator in the timing circuit to deliver pulses at the preset rate. The pulse width rate determines duration of the pulse delivered to the heart. Then the output of the pulse width circuit fed into the rate limiting circuit, which limits the pacing rate to a maximum of 120 pulses. The output circuit provides a proper pulse to stimulate the heart Thus the timing circuit, pulse width circuit, rate limiting circuit and output circuit are used to produce the desired pacemaker pulses to pace the heart. The energy compensation circuit produces an increase in he pulse duration as the battery voltage decreases to maintain constant stimulation energy to the heart UNIT –5 1. Explain short-wave diathermy and ultrasonic diathermy with diagran The heating of the tissues is carried out at a high frequency of 27.12mhz and a wave length of 11 meters. When we use currents having high frequencies the motor or sensory nerves are not stimulated and there is no contraction of muscles. The RF energy creates the tissues and promotes the healing of injured tissues and inflammations. There are sev real provisions to regulate the intensity of current passed through the patient circuits. The pads are placed so that the portion of body to be treated is sandwiched between them When RF current is applied to the pads the dielectric loss of capacitor produces heat in the intervening tissues. This technique is called condenser or capacitor method. There is also an inductive method in which a flexible cable is coiled around the arm or knee or any other portion of the patient body where plate electrodes are inconvenient to use Utrasonic diathermy Ultrasonic therapy is used where short wave treatment is failed and in cases where localization of heat effect is desires. The ultrasonic diatheraphic units are helpful to cure he diseases of peripheral nervous systems like neuritis, skeletal and muscular systems. The heating effect is produced because of absorption of ultrasonic energy by tissues. The effect of ultrasonic on the tissues is high speed vibration of micro massage. Micro massage is use in the treatment of soft tissue lesions. I t is better than manual massage because of depth of massage and can be obtained without any pain to the paient The transducer is made of bariumtitanite The ultrasonic waves can be applied in continuous or pulsed mode. The treatment timer is an electrically operated contact which can be set from 1 to 15 minutes 3. Explain wih block diagram Multiple channel telemetry system For most biomedical instruments it is desirable to have simultaneous recording of several signals for correlation study. Each signal require a telemetry channel. There are two types i) Frequency division multiplex ii) Time division multiplex Frequency division multiplex System : Each signal is frequency modulated on a sub carrier frequency. Then these modulated sub carrier frequencies are combined to modulate the main R.F.carrier. A t the receiver side the modulated sub carriers will be separated by proper band pass filters after the first discrimination The individual signals are recovered from these modulated sub carriers by the second set of discriminator The frequency of sub carriers has to be carefully selected to avoid interferences. The low p[ass filters are used to extract signals without any noise. Time division multiplex system: Since most biomedical signals have low frequency bandwidth requirements we can use time division multiplex system by the timesharing scheme. The transmission channel is connected to each signal channel input for a short time to sample and transmit the signal. Then the transmitter is switched to the next signal channel in a definite sequence. When all the channels have been scanned once a cycle is com placed the next will start. The operation is repeated again. Th the receiver end the process is reversed. The sequenciality is arranged, signal pulse are distributed to individual channels by a synchronized switching circuit. 4. What are the problems in biotelemetry? Give its uses Problems in biotelemetry: For long term monitoring implant telemetry is more useful one. The whole electronic circuit is fully packed as a capsule and then implanted For implant telemetry, the size and weight limitations are much more serous and the reliability requirements is more critical Body reaction size, weight, surface condition and shape of the implant system will have effects on the body reaction. Mean while medical grade silastic, Teflon, glass and some metals which are used as enclosures cause little foreign body reaction on tissue The coating materials of electronic circuits to protect them from body fluid are silicon ruby, epoxy, plastics, paraffin, glass and metal Power supply : Two special power supplies are used for long term implant telemetry units along with the mercury and lithium cells i) Environmental Power supply : Radio induction has been applied to transmit mill watt of power to the implanted telemetry units for months ii) Microwatt Power supply circuits using piezoelectric crystals laced on any blood vessel or aortas Uses of Biotelemetry : Biotelemetry helps us record biosignals over long periods and while the patient is engaged in his normal activities The medical attendant or computer can easily diagnosis the nature of disease by seeing the telemetries signals without attending the patient room Patient is in his room without any mechanical or physical disturbance during recording by mean of biotelemetry For future reference the biotelemetry is essential one For recording animals, particularly for research ths biotelemetry is greatly used For monitoring the persons