Technical Specifications for Fixed Diagnostic Radiology Equipment
Al-Mafraq and Jerasn-Ajloun Military Hospitals
NOTES

This tender shall be processed as a package; partial bidding will
be rejected.

Bidders are REQUIRED to fill out the product details,
compliance information, and brochure reference page number in
tabular format for all items below, in a clear, unambiguous
manner. This information shall be part and parcel of any
subsequent award contract
Technical Specifications for Fixed Diagnostic Radiology Equipment
Al-Mafraq and Jerasn-Ajloun Military Hospitals
Item 1. MRI SYSTEM (1.5T)
Qty.(2)
A magnetic resonance imaging system is required for whole body imaging
applications in a general hospital. The system must meet or transcend the minimum
requirements listed below.
A. Magnet & Gantry
1. Configuration: Closed
2. Strength: 1.5 T
3. Homogeneity (V-RMS) 40 cm DSV ppm: ≤0.3
4. Shimming: Active
5. Boil-off rate l/hr: ≤ 0.03
6. Cryogen refill frequency: ≥3 years
7. Bore diameter, cm: ≥ 60
8. Patient comfort: Audio, lighting, etc.
9. Acoustic noise reduction
B. Table
1. Two-speed longitudinal movement
2. Min. height, cm ≤ 70 cm
3. Manual table movement
4. Table control from console
5. Digital display of table position
6. Scan range ≥ 150 cm
7. Laser light indicator
8. Table weight limit: ≥ 150kg
C. Gradient System
1. Max. Amplitude, mT/m: ≥33
2. Standard slew rate, T/m/sec: ≥120
3. Cooling type: Water cooled system
D. RF System
1. Amplifier type: Solid state
2. Number of independant channels: ≥8
3. Calibration technique: Automatic
E. Coils: All available standard & optional types must be listed
F. Scanning Techniques
1. EPI pulse sequences: Please Specify
2. Neurologic imaging
Page 1 of 19
Technical Specifications for Fixed Diagnostic Radiology Equipment
Al-Mafraq and Jerasn-Ajloun Military Hospitals
3. Diffusion
4. ADC Maps
G. Image Processor
1. Processor, No.×speed: ≥ 2×2.6GHz, 64bit
2. Dedicated RAM: : ≥ 8GB
3. Maximum matrix: ≥1024×1024
4. Min. 2D slice thickness, mm: ≤0.3
5. Min. 3D slice thickness, mm: ≤0.1
6. Minimum FOV, cm: : ≤1
7. Maximum FOV, cm: ≥ 48
8. 3D reconstruction, MIP, MPR
9. Zooming and distance measurement
10. Pixel intensity value and ROI measurement
H. Control Console Workflow
1. Processor, No.×speed: ≥ 2×2.0GHz
2. Dedicated RAM: : ≥ 2GB
3. OS: 64bit System
4. Color LCD, ≥19” 1280×1024
5. HDD storage: ≥ 250000 image (2562)
6. MOD/CD/DVD Archiving
7. Full DICOM 3.0, RIS, HIS, PACS Support
8. Tele-radiology capability: Please Specify
9. Remote diagnostics
I. Supplementary Equipment
1. MRI compatible anesthesia machine with patient monitor (completely nonmagnetic)
2. MRI compatible Injector
3. UPS
4. MRI compatible wheel chair and trolley
5. Control room furniture( 8 chairs & one Table)
6. Chiller must be included
All available standard & optional features, packages, & accessories must be listed and
priced separately
Page 2 of 19
Technical Specifications for Fixed Diagnostic Radiology Equipment
Al-Mafraq and Jerasn-Ajloun Military Hospitals
Item 2. COMPUTED TOMOGRAPHY SYSTEM
Qty.(2)
A multi-slice spiral CT scanner is required for all types of CT examinations
including head, spine, chest, abdomen and pelvis in addition to cardiovascular
applications in a general hospital. The system must meet or transcend the minimum
requirements listed below.
A. Main Features
1. Type: Multislice, 64 Slices per second.
2. Application: General, Full Body
3. Shortest rotation time, sec, 360°: ≤ 0.5
B. Gantry
1. Gantry aperture, cm: 70
2. Gantry tilt, °: ±30
3. Scan localizer: Laser
4. Detector type: Solid state, Ceramic
5. Field of view (standard), cm: ~ 50
6. Slice width options, mm: ~ 0.5-10
7. Shortest rotation time, sec, 360°: ≤ 0.5
8. Laser light indicator
C. X-Ray Tube
1. Heat storage, MHU/ Heat dissipation, kHU/min: 8/1600 or a better
combination
2. Small focal spots, mm: ~0.6×1
3. Large focal spots, mm: ~1×1
4. Max. mA for smallest tube spot: 500
5. Max. scan time at maximum mA, sec: 10
D. X-Ray Generator
1. kW output: ≥60
2. kVp range: ~80-140
3. mA range: ~20-500
E. Performance
1. Spatial Resolution 0% MTF, lp/cm: ≥ ~20
2. Spatial Resolution 50% MTF, lp/cm: ≥ ~8
3. Low-contrast resolution, mm at % at 2 rads: ~4 mm at 0.3% at 2 rads
4. Noise, % at 2.5 rads: ~0.3% at 2.5 rads
5. Adaptive filteration for noise reduction
6. Max. Continuous scan time, sec: 100
Page 3 of 19
Technical Specifications for Fixed Diagnostic Radiology Equipment
Al-Mafraq and Jerasn-Ajloun Military Hospitals
7. Periodic User calibration: NONE, Automatic
8. Total time from fully off to scanning, min:≤ ~5
F. Patient Table
1. Vertical movement, cm: ~ 50-100
2. Longetudinal movement, cm: : ≥ 150
3. Scannable range, cm: ≥ 150
4. Max. load capacity, kg : ≥ 200
5. Table speed, mm/s: 1 – 100, or better
6. Positioning accuracy, mm: ≤ ±0.5
G. Radiation Dose
1. Pediatric-specific dose control
2. Prospective ECG gating
3. Low-dose cardiac Examination
4. AEC:
i. mA adjustment for patient size
ii. mA adjustment along the z-axis
H. Clinical Applications And Functionality
1. 3D Reconstruction, MIP, MPR
2. Cardiac Package
3. Neurology Package
4. Coronary artery calcification scoring
5. Lung function analysis
6. Pediatric Package
I. Image Reconstruction
1. Reconstruction matrix: 512×512
2. Reconstruction rate, (512 x 512), f/s: ~20
3. Simultaneous scanning and reconstruction
4. Simultaneous and routine analysis
5. Simultaneous scanning and/or hard copying
J. Main Console
1. Two color LCD monitor: ≥19” 1024×1024
2. HDD storage: ≥ 250000 image (5122)
3. MOD/CD/DVD Archiving
4. Full DICOM 3.0, RIS, HIS, PACS Support
5. Tele-radiology capability: Please Specify
6. Remote diagnostics
Page 4 of 19
Technical Specifications for Fixed Diagnostic Radiology Equipment
Al-Mafraq and Jerasn-Ajloun Military Hospitals
K. Main Computer
1. Processor, speed: ≥ 2.6GHz
2. Dedicated RAM: : ≥ 8GB
3. Word length: 64bit
4. Maximum matrix: ≥512×512
L. Independent Workstation
1. Processor, No.×speed: ≥ 2×3.0GHz
2. Dedicated RAM: : ≥ 2GB
3. HDD storage capacity: ≥ 140GB
4. MOD/CD/DVD Archiving
5. Full DICOM 3.0, RIS, HIS, PACS Support
M. Supplementary Equipment
1. Compatible Injector
2. UPS for computer system
3. Control room furniture ( 8 chairs & a table)
All available standard & optional features, packages, & accessories must be listed and
priced separately
Page 5 of 19
Technical Specifications for Fixed Diagnostic Radiology Equipment
Al-Mafraq and Jerasn-Ajloun Military Hospitals
Item 3. RADIOGRAPHIC/FLUOROSCOPIC SYSTEM
Qty. (2)
A Digital Radiographic/Fluoroscopic system is required for general hospital use.
The system must meet or transcend the minimum requirements listed below.
A. Main Features
1. Type: Digital
2. Application: General Purpose
3. Configuration: Overcouch tube
4. Remote Operation from control area
B. Imaging Tower
1. Longitudinal movement, cm: ≥150
2. Lateral movement, cm: ≥20
3. Vertical movement, cm: ≥30
4. Maximum SID, cm: ~150
5. Minimum SID, cm: ~115
6. Ambidextrous Controls
7. Fluoroscopy control: Footswitch
8. Removable Grid
9. Integrated table/tower movements controls
C. Detector
1. Technology: Image Intensifier+CCD
2. Detector Size, cm: ~40
3. Field Sizes: Please Specify
D. Fluoroscopy
1. kVp pulsed fluoroscopy
2. Pulse rate: ~0-30
3. User selectable filtration
4. Additional filtration, mm Cu: ~0.1, 0.2, 0.3
5. Maximum acquisition matrix size: 1k
6. Acquisition bit depth:≥10
7. Grid Pulsed fluoroscopy is an dvantage
E. Radiography, Spot Filming
1. Acquisition matrix size: 1k
2. Maximum spatial resolution, lp/mm: ~5
3. Frames per second up to 6 or more
4. Acquisition bit depth:≥10
5. Conventional radiography: size, cm: 43×43
6. Retractable grid
Page 6 of 19
Technical Specifications for Fixed Diagnostic Radiology Equipment
Al-Mafraq and Jerasn-Ajloun Military Hospitals
F. Imaging Capabilities
1. Last image hold
2. Virtual collimation
3. Edge enhancement and noise reduction
4. Zoom and pan functions
5. Manual pixel shift
G. Patient Table
1. Maximum load kg: ≥200
2. Lateral movement, cm: ≥~20
3. Longitudinal movement, cm: ≥~100
4. Table tilt, °: ±90
5. Maximum patient weight for tilt, kg: ≥180
6. Tilt indicator
7. Motorised Movements
H. Generator
1. High-Frequency type generator
2. Generator power, kW: ≥60
3. Radiographic kVp range: ~50-150
4. Radiographic mA range: ~0-1000
5. Fluoroscopic kVp range: ~40-110
6. Fluoroscopic mA range: ≥ ~0-20
I. X-Ray Tube
1. Tube power, kW: ≥60
2. Focal spot sizes, mm: ~0.6, 1.2 or better
3. Tube heat capacity, kHU: ≥ 700 or equivalent
4. Controls on X-ray tube: please specify
J. Image Display
1. Two overhead color LCD monitors: ≥18”
2. Control room color LCD monitor: ≥18”
K. Connectivity
1. Full DICOM 3.0, RIS, HIS, PACS Support
2. Remote diagnostics
L. Accessories
1. Handgrips
2. Removable footrest
3. Shoulder rests
All available standard & optional features, packages, & accessories must be listed and
priced separately
Page 7 of 19
Technical Specifications for Fixed Diagnostic Radiology Equipment
Al-Mafraq and Jerasn-Ajloun Military Hospitals
Item 4. RADIOGRAPHIC SYSTEM, GENERAL PURPOSE
Qty.(4)
A General purpose radiography system is required for hospital use with a
compatible, high-throughput CR system. The system must meet or transcend the
minimum requirements listed below.
A. Main Features
1. Application: General Purpose
2. Tube assembly: Ceiling Suspended
3. Configuration: Table and upright Bucky
4. Type: CR compatible system
B. Table
1. Type: 4-way float-top
2. vertical movement, cm : ≥30 Motorized
3. Longitudinal movement, cm: ≥~100
4. Lateral movement, cm: ≥~25
5. X-ray density: ≤1mm Al
6. Table weight capacity, kg: ≥200
7. Table support: Fixed Pedestal Base
8. Locking system: Electromagnetic
C. Table Bucky System
1. Type: Motorised/Manual with auto sensing
2. Size: All CR Cassette sizes
3. AEC: 3-field
4. Grid Ratios: 10:1 or higher
D. Upright Bucky System
1. AEC: 3-field
2. vertical movement, cm : ≥~100 Motorized
3. Size: Standard Chest CR Cassette size
4. Grid Pulsed fluoroscopy is an dvantage
E. Generator
1. High-Frequency type generator
2. Generator power, kW: ≥60
3. kVp range: ~40-150
4. mAs range: ~0.5-800
5. Automatic exposure programs
Page 8 of 19
Technical Specifications for Fixed Diagnostic Radiology Equipment
Al-Mafraq and Jerasn-Ajloun Military Hospitals
F. X-Ray Tube
1. Tube power, kW: ≥60
2. Focal spot sizes, mm: ~0.6, 1.2 or better
3. Tube heat capacity, kHU: ≥ 700 or equivalent
4. Controls on X-ray tube: please specify
5. Manual collimation
G. Ceiling Suspension
1. Travel range: L, T, and V: please specify
2. Locking system: Electromagnetic
3. Tube rotation: please specify
4. Power assist (option)
H. CR System
1. Application: all radiographic exams
2. Resolution: ≥ 10 pixel/mm
3. 12-bit image plate reader or better
4. Cassette buffer capacity: ≥ 4 (input)
5. Throughput per hour: : ≥ ~100 (all sizes)
6. Patient ID: Manual entry, and DICOM modality worklist
7. Image Storage: ≥ ~2500 images
8. Processor: Core 2 duo 2.6 GHz or better
9. RAM: ≥ 4GB
10. High-resolution LCD monitor: ≥ 17”
11. Automated image processing
12. Image quality assurance
13. Zoom and roam function
14. Contrast enhancement
15. Full DICOM 3.0, RIS, HIS, PACS Support
16. Archiving: CD/DVD
All available standard & optional features, packages, & accessories must be listed and
priced separately
Page 9 of 19
Technical Specifications for Fixed Diagnostic Radiology Equipment
Al-Mafraq and Jerasn-Ajloun Military Hospitals
Item 5. MAMMOGRAPHY SYSTEM (SCREEN-FILM)
Qty.(2)
A Radiographic Mammography unit with a compatible, high-resolution CR system
is required to provide excellent image quality for screening, and diagnostic
mammography exams. The system must meet or transcend the minimum
requirements listed below.
A. Generator
1. Type: High Frequency
2. Generator Power, kW: : ≥~3.2
3. kV range:Lower ~20, Upper: ≥~35, 1kV steps
4. mAs range:Lower ~4, Upper: ≥ 500
5. mA range: Max. ≥ 100
6. Automatic Exposure Control
B. X-Ray Tube
1. Anode type: Rotating
2. Heat capacity, kHU: ≥ 300
3. Heat dissipation rate, kHU/min: ≥ 60
4. Target/filter combinations: Mo/Mo, Mo/Rh
5. Focal spot size, mm: ~ 0.1 and 0.3
C. Positioning Assembly
1. Auto Collimation, 18 x 24 cm & 24 x 30 cm
2. Movement Locks: Electromagnetic
3. Assembly movement: Motorized
4. Rotation range, °: -135 to +180 or wider
5. Vertical Travel, cm: ~70-140
6. SID, cm: ~65
7. Display: at least pressure and thickness
8. Removable patient face shield
D. Compression System
1. Motorized with foot switch and manual mode
2. Force, N: Up to 200, variable
3. Compression Modes: Please Specify
4. Safety Features: Please Specify
Page 10 of 19
Technical Specifications for Fixed Diagnostic Radiology Equipment
Al-Mafraq and Jerasn-Ajloun Military Hospitals
E. Screen-Film system
1. Grid Ratio: 5:1
2. Bucky for both film sizes: 18 x 24 cm & 24 x 30 cm
3. Magnification Device
4. Film ID System
F. CR System
1. Application: Mammography
2. Resolution: ≥ 10 pixel/mm
3. 12-bit image plate reader or better
4. Cassette buffer capacity: ≥ 2 (input)
5. Patient ID: Manual entry, and DICOM modality worklist
6. Image Storage: ≥ ~2500 images
7. Processor: Core 2 duo 2.6 GHz or better
8. RAM: ≥ 4GB
9. High-resolution LCD monitor: ≥ 17”
10. Automated image processing for mammography exams
11. Image quality assurance
12. Zoom and roam function
13. Contrast enhancement
14. Full DICOM 3.0, RIS, HIS, PACS Support
15. Archiving: CD/DVD
All available standard & optional features, packages, & accessories must be listed and
priced separately
Page 11 of 19
Technical Specifications for Fixed Diagnostic Radiology Equipment
Al-Mafraq and Jerasn-Ajloun Military Hospitals
Item 6. DIAGNOSTIC ULTRASOUND SYSTEM
Qty.(6)
A general-purpose, mid- to high-range diagnostic ultrasound system is required for
general hospital use at the radiology department. The system must meet or
transcend the minimum requirements listed below.
A. Application:
1. Abdominal
2. Musculoskeletal
3. Neonatal Pediatric
4. Small Parts
5. Urological
6. Vascular
7. Transcranial
8. OB/GYN
B. Configuration:
1. Mobile cart configuration
2. Four full swivel castors with brakes
3. Full control keyboard
4. High-resolution LCD: ≥ 17”
5. Storage for transducer cables
6. Auxiliary storage compartments/drawers
7. Minimum of 3 transducer connections
8. User-adjustable presets
9. Adjustable control panel: up/down & rotate
10. Adjustable monitor position: up/down, rotate, swivel, & tilt
C. General Features:
1. Grayscale levels: 256
2. Power up time: ≤ 2 minutes
3. Self-diagnostic routine during power up
4. Touch screen capability
5. Acquisition frame rate: ≥ 500 fps
6. Simultaneous processing bandwidth capability: 1-16MHz
7. No. of digitally processed channels: ≥ 9000
8. Input dynamic range: : ≥ 180dB
9. Cine loop: ≥ 1000 frames
10. Automatic real-time Doppler trace capability including calculation and display
of user-selected measurements
11. Measurements on replay
12. Real-time Pan/Zoom
Page 12 of 19
Technical Specifications for Fixed Diagnostic Radiology Equipment
Al-Mafraq and Jerasn-Ajloun Military Hospitals
13. Frozen image Pan/Zoom
14. Automatic optimization of base line and scale
15. Local HDD patient search through name and/or ID
16. Full DICOM 3.0, RIS, HIS, PACS Support
D. Image Storage:
1. HDD: ≥ 160 GB
2. Removable Storage: CD/DVD, USB
E. Multifrequency Transducers:
1. Linear Array:
~5-12 MHz with biopsy guides
2. Curved Array: ~1-5 MHz with biopsy guides
3. Phased array: Microconvex for pediatric applications
F. Imaging Modes:
1. 2D
2. B-mode
3. Tissue harmonic imaging
4. M-mode and colour M-mode
5. Dual imaging
6. Doppler imaging
7. Colour Doppler imaging
8. Colour Power Angio Imaging
9. Pulsed Wave (PW) Doppler
10. Simultaneous PW Doppler and 2D (Dual Mode)
11. Simultaneous 2D, colour Doppler, and PW Doppler (Triple Mode)
12. Contrast imaging
13. Compounding imaging
14. Speckle reduction
All available standard & optional features, packages, & accessories must be listed
and priced separately
Page 13 of 19
Technical Specifications for Fixed Diagnostic Radiology Equipment
Al-Mafraq and Jerasn-Ajloun Military Hospitals
Item 7. MINI-PACS
Qty.(2)
An integrated, modular, and scalable mini PACS (hardware and software) with
essential RIS functions is required to support all modalities listed above (items 1-6)
Pricing should be based on a minimal working configuration with all options listed
and priced separately. The system must meet or transcend the minimum
requirements listed below.
A. Main Features:
1. Scalable archive software
2. Diagnostic DICOM viewer, router, and DICOM to CD/DVD
3. Image manipulation
4. Integrate patient registration and DICOM worklist
5. Remote image viewing capability
6. Thumbnail gallery to fast preview and select images
7. Import and export images of DICOM/JPEG/BMP/TIFF format
8. Lossy and lossless compression
B. Hardware Features:
1. Low-latency, high-end hardware
2. Storage Hardware: Raid (NAS/SAN)
3. Max. Storage Capacity: ≥ 5 years worth of storage guaranteed by supplier
4. Long-term storage: DVD
5. Std. peripheral devices: Please specify
6. Backup features: please specify
7. Network Jukebox Storage system: optional
8. UPS must be included
C. Network Features:
1. Operating System(s): Windows, Unix, Linux or SUN systems
2. Multi-criterion patient query /search
3. HL7 & HIS interfacing capability
9. Max. No. Of terminals: Please Specify
4. Security: 128-bit SSL
5. Database management: Please specify
D. Diagnostic Workstation(s):
1. At least two stations 3MP and 5MP resolution
2. Independent login
3. Admin-controlled worklist
4. Auto notification of prior exams
5. Image manipulation : please specify
Page 14 of 19
Technical Specifications for Fixed Diagnostic Radiology Equipment
Al-Mafraq and Jerasn-Ajloun Military Hospitals
6. User-definable hanging protocols
7. Key image select
E. Web Image Access:
1. Number of licenced users: Please specify
2. Max. Number of concurrent users: Please specify
3. Patient search criteria: please specify
4. Automatic image compression
5. Image manipulation: : please specify
6. Thumbnail image selection
F. Image Sharing:
1. Printing support
2. Self-executing CD-ROM production
3. Remote viewing from KHMC
G. System Admin:
1. GUI-based tools
2. Several access clearance levels
3. Patient manage
4. Hardware manage
5. Back-up settings
6. Auto alert of system failure
H. DICOM 3.0 Services:
1. Query/retrieve SCP
2. Query/retrieve SCU
3. Worklist management
4. Other: please specify
I. RIS Functions:
1. Scheduling
2. Results reporting
3. Case retrieval
4. Word processing
All available standard & optional features, packages, & accessories must be listed
and priced separately
Page 15 of 19
Technical Specifications for Fixed Diagnostic Radiology Equipment
Al-Mafraq and Jerasn-Ajloun Military Hospitals
SPECIAL TERMS
1. All equipment must be of the latest models of the range specified by the
technical specifications (authenticated certificate by the manufacturer must be
submitted), old models will be rejected.
2. Submission of a formally endorsed certificate of FDA approval/ clearance for
selling equipment of USA origin to USA healthcare facilities (installation
references must be included in the offered documents), CE mark (MDD)
certification for European Equipment that meet the country of origin
requirements clearly indicating the certifying body number for class ΙΙ
equipment. JIS (Japanese Industrial standards) approval for Japanese
equipment, CSA (Canadian Standards Association) approval for Canadian
equipment, offers not including such certificates will be rejected.
3. Country of origin:
3.1. USA, Canada, Japan, UK, Sweden, Finland, Denmark, Switzerland,
Belgium, Germany, France, Netherlands.
3.2. Products manufactured by well-known companies based in countries
mentioned in item 3.1 that have manufacturing plants in different parts of
the world and hold free-sale certificates in the original country where the
mother company is based will be taken into consideration.
3.3. For countries not mentioned in 3.1, ONE of the following conditions
must be met:
a. A formally endorsed certificate from the Institute of Biomedical
Technology/ Royal Medical Services, indicating that the system itself or
an older model of the same type has been used by the Royal Medical
Services for a minimum period of five consecutive years with an
acceptable service record, and without any written claims.
b. The offered model must have successfully passed through the process of
medical equipment assessment program in accordance with the
regulations of the assessment committee in the Royal Medical Services
(acceptance certificate must be submitted).
Page 16 of 19
Technical Specifications for Fixed Diagnostic Radiology Equipment
Al-Mafraq and Jerasn-Ajloun Military Hospitals
4. Offers must include an original copy of free-sale certificates in the original
country where the mother company is based endorsed by the ministry of health
of that country.
5. A list of Reputable institutions worldwide where the same offered model is
operating especially in the USA or Germany. Reference sites will be verified
and evaluated at the discretion of the purchasing committee.
6. Bidders must submit their reservations / queries regarding tender
specifications and / or special terms within the first third of the tender closing
period.
7. Tender awards will be based upon the best value for money and best
performance expectations (reputation, after sale support etc.).
8. Offers must include a two-year full manufacturer warranty, including parts and
labour.
9. One set of operation manuals, and one set of service manuals including
schematics and a spare-part list (CD/DVD is acceptable) must be delivered with
each unit.
10.Pricing must include services of sale, shipment (Customs clearance shall be the
responsibility of the JAF), transportation and delivery to site, installation, and
bringing into service of the equipment.
11.Power requirements: single phase 220V, 50Hz, or 3-phase 380V.
12.Offers must include clear technical brochures and data sheets.
13.Offers must include compliance sheets in tabular format listing our minimal
technical system requirements (as per tender documents) on one side, and a
clear and unequivocal response to each and every entry on the other side, with
reference (page and line numbers) to technical brochure(s). Failure to comply,
gives the purchasing committee the right to reject the offer in whole, or in
part. Bidders will be held responsible for all information in compliance sheets.
Erroneous or misleading entries will be recorded for future reference and may
eventually render bidders in question blacklisted.
14.Any accessories and consumable items necessary to operate the offered
system(s) must be clearly identified and priced. Prices must be fixed for a
minimal period of 5 years. A start-up kit of accessories and consumable items
must be provided with the system(s).
Page 17 of 19
Technical Specifications for Fixed Diagnostic Radiology Equipment
Al-Mafraq and Jerasn-Ajloun Military Hospitals
15.Spare parts availability must be guaranteed for a minimum of 10 years starting
from date of installation and commissioning.
16.Offers must include a comprehensive, priced list of spare parts valid for a
minimal period of 5 years. Any item not listed will be considered as free of
charge. Another copy of the list must be submitted at the Institute of
Biomedical Technology (IBMT) before an installation-and-commissioning
certificate can be issued.
17.For PC based systems, a complete restoration CD/DVD (Operating system and
application software) must be supplied. Only computers made by hp/Compaq,
IBM/Lenovo, Fujitsu, Dell, SUN systems or by the same manufacturer of the
offered system(s) are acceptable.
18.This tender shall be processed as a package; partial bidding will be rejected.
19.The package must include a teleradiology solution for diagnostic, highresolution image viewing from KHMC department of radiology. Bidders are
encouraged to meet with senior radiologists at KHMC to acquaint themselves
with the required set up.
20.All items must be DICOM 3.0 and HL7 compliant.
21.Pre-installation:
21.1. For all items, pre-installation shall be the responsibility of the supplier,
and shall include all floor ducting and finishing, false ceiling, room
lighting, AC, and any electromechanical work and/or civil work
involving site modifications necessary to accommodate the new system
in coordination with the civil and/or electromechanical and/or IT
infrastructure contractor(s)
and in accordance with the system
manufacturers' specifications and international hospital standards.
21.2. For Item 1 (Magnetic Resonance Imaging System), A high-quality and
efficiency chiller sufficient to satisfy all the cooling requirements must be
supplied and installed with the system.
21.3. For each and every one of the X-Ray Imaging and computed
tomography systems, it is the suppliers’ responsibility to provide and
install the following:
a. An efficient, dedicated A/C system
Page 18 of 19
Technical Specifications for Fixed Diagnostic Radiology Equipment
Al-Mafraq and Jerasn-Ajloun Military Hospitals
b. Lead glass viewing window
c. Beam supports for x-ray tube weight bearing ceiling suspensions (where
applicable)
d. Partitions (where applicable)
e. Lead rubber under-door seals
f. A set of three lead aprons of different sizes complete with heavy-duty
wall-mounted racks for safe storage.
g. UPS where applicable.
21.4. Bidders are invited to conduct a field survey of the site before submitting
their offers.
22.Training:
Offers must include certified training programs (including user training
for at least one user (minimum period 10 working days) and service
training for at least one biomedical engineer/technician (minimum priod
10 working days) at a reputable training center sanctioned by the system
manufacturer for each item. Tickets, expenses, full accommodation and
board must be included. The program must conform to the following
standards:
 Training must be conducted on a system of identical make, model,
and configuration to that purchased by DRMS.
 Certificates must be endorsed and officially sealed by the system
manufacturer, legally empowering trainees to engage in user and
service activities according to operation and service manual(s).
Page 19 of 19