* Your assessment is very important for improving the workof artificial intelligence, which forms the content of this project
Download CT1 - hullrad
Survey
Document related concepts
Transcript
CT Scanning Dr. Craig Moore Medical Physicist & Radiation Protection Adviser Radiation Physics Service CHH Oncology Brief History of CT Scanning First CT Scanner - 1972 • • • • • • Originally called CAT A = axial 80 x 80 resolution 4 min. per rotation 8 grey levels overnight reconstruction Here and Now • • • • 512 x 512 or 1024 x 1024 resolution Sub second rotation 4096 grey levels 100’s slices per rotation Components of a CT Scanner Principal components of CT scanner • X-ray tube, collimator, and detector array on a rotating gantry • Rotation axis is referred to as Z axis • Fan beam wide enough to cover patient crosssection • Narrower width in the z-axis • Behind the patient is a bank of detectors • Patient lies on a couch that is moved longitudinally through the gantry X-ray Tube • • • • • • • Tube parallel to patient movement – minimise anode heel effect X-rays are produced by firing electrons at a metal target – typically tungsten Capable of producing long exposure times at high mA – get very hot (require heat capacities up to 4MJ and active cooling mechanisms) Continuous scanning limited to around 90s Focal spot size typically 0.6 - 1mm High kVps (80 – 140) and beam heavily filtered (6-10 mm Al filters) to optimise spectrum Stops attenuation coefficients varying with depth via beam hardening Collimation and Filtration • • • Want a monoenergetic beam to avoid beam hardening artefacts – As beam passes through patient low energies are filtered – This results in the apparent reduction of attenuation and CT number of tissues – Computer reconstruction assumes monoenergetic beam – Not possible with X-ray tubes so they are heavily filtered – At least 6 mm aluminium or copper and typically high kVp – Some manufacturers use shaped filters such as bow-tie filters to even out the dose distribution (conform to the shape of an elliptical patient) Pre-patient collimator is mounted on the Xray tube – Beam is approx 50cm wide to cover cross section of patient – The size is variable in the z-axis – Multi-slice scanners between 0.5 and 40 mm thick beams Post patient collimation is not used with multi-slice scanner Bow Shaping Filters • Body is approx circularly shaped • Shape of beam that reaches detector is therefore non-uniform which can lead to image artefacts • Shaping filters are applied to the beam prior to make the dose distribution more uniform Detectors • Requirements: – – – – – – Small enough to allow good spatial resolution Up to 1000 detectors per scanner Typically 1.5 mm width but can be a small as 0.5mm High detection efficiency Fast response Wide dynamic range – massive variation in X-ray intensity – Stable and noise free – No afterglow – There needs to be separation between detectors to prevent light crossover • This reduces efficiency from 98% to 80% Detectors • Modern CT detectors are fabricated as detector modules • A side view shows individual detector elements coupled to photodiodes that are layered on electronics • Spaces in between filled with optical filler to reduce cross talk Detectors Generations of CT Scanner CT Imaging • Conventional radiography suffers from the collapsing of 3D structures onto a 2D image • However, CT scanning has extremely good low contrast resolution, enabling the detection of very small changes in tissue type – Almost true depiction of subject contrast • CT gives accurate diagnostic information about the distribution of structures inside the body • Generation of images in transaxial section – Perpendicular to the axis of rotation of the X-ray tube about the body – Perpendicular to the craniocaudal axis of patient Number of detectors and projections • Typically, for a 3rd generation scanner: – 650 – 900 detectors – 1000 to 2000 projections per rotation Collapse of 3D Data into 2D Plane Image contrast 2:1 • Planar imaging – 2D representation of 3D Distribution of Tissue – No depth information – Structures at different depths are superimposed • Loss of contrast Subject Contrast 4:1 X rays Typical CT Image CT Images • • • Commonly calculated on 512x512 matrix, but 256x256 and 1024x1024 are also used Each pixel is more accurately described as a voxel, because it has depth information The value stored in each voxel is referred to as the CT number which is related to the attenuation of a particular tissue: – CTn = 1000 x (µt - µw)/ µw • • • Sometimes referred to as Hounsfield Units Each CT number is assigned a certain shade of grey in the resulting image CT number represents x-ray attenuation coefficient of the corresponding voxel within the patient CT Numbers Tissue Range of CT Numbers Bone 500-3000 Muscle 40-60 Brain (grey matter) 35-45 Brain (white matter) 20-30 Fat -60 to -150 Lung -300 to -800 Image Display • CT image represented by a range of CT numbers from -1000 to + 3000 (ie 4000 levels of grey) • Human eye dose not have the capacity to distinguish so many grey levels • If 4000 shades of grey displayed altogether there would be very little difference between different tissues Window Width and Level • The appearance of the image on the screen can be changed by altering the window width and level • Window width refers to the range of CT numbers selected for display • This range of CT numbers is centred at a particular level called the window level – e.g. if imaging bone window level should be ~1000 • Can spread a small range of CT numbers over a large range of grayscale values Window Level –593 Window Level –12 Window Width 500 Window Width 400 Good contrast in lungs Good soft tissue contrast Only see CT numbers +/- 250 around -593 Only see CT numbers +/200 around -12 How do we get the images? • Tube and detector rotate smoothly around the patient • X-rays are produced continuously and the detectors sample the X-ray beam approx 1000 times during one rotation • Typically 2 to 4 revolutions per second • • • In reality not always parallel to detectors Each voxel is traversed by one or more x-ray beams for every measurement (1000 per rotation) Number of measurements taken in single axial section depends on – number of detectors – Number of samples per rotation • • • Assume 800 detectors measured at 0.5° intervals per 360 ° rotation This is 576,000 measurements More than needed as we only need 260,000 measurements (512 x 512) How do we get the picture? • Back Projection – Reverse the process of measurement of projection data to reconstruct image – Each projection if smeared back across the reconstructed image Back Projection – the basics • • • • • • • • • Consider cylindrical uniform body with a hole down the centre A beam passing through this body from one direction will have a transmitted profile in its central region This single measurement cannot determine the position of the hole other than identifying that it is in the line of the pencil beam passing through the centre of the body Pixel values along this line are decreased by the amount of attenuation measured These values are projected back along the field of view A second projection at 90° provides a second band of grey This is then projected back across field of view Progressive projections are shown in the final figure – a star like pattern We now have an image that looks similar to what we are scanning Back Projection 3 6 3 3 3 6 6 3 3 3 6 3 • Back Project each planar image onto three dimensional image matrix Back Projection 3 6 3 1 2 1 1 2 1 1 2 1 • Back Project each planar image onto three dimensional image matrix Back Projection 3 6 3 3 21 2 2 3 1 2 6 31 2 4 31 3 21 23 1 2 • Back Project each planar image onto three dimensional image matrix Back Projection 3 6 3 3 4 6 4 3 6 6 8 6 6 3 4 6 4 3 3 6 3 • Back Project each planar image onto three dimensional image matrix Back Projection 3 6 3 3 4 6 4 3 6 6 8 6 6 3 4 6 4 3 3 6 3 • Back Project each planar image onto three dimensional image matrix Back Projection • More views – better reconstruction • 1/r blurring, even with infinite number of views Filtered Back Projection • Back projection produces blurred transaxial images • Projection data needs to be filtered before BP • Different filters can be applied for different diagnostic procedures – Smoother filters for viewing soft tissue – Sharp filters for high resolution imaging • After filtration, back projection same as before – Data from neighbouring beams are used – Some data is subtracted – Some data is added • Filters are convolved with the blurred image data in Fourier Space Filtered Back Projection Filtered Back Projection • Usually done in frequency (Fourier) space • Filter planar views prior to back projection • Correction of 1/r blurring requires ‘Ramp’ Filter – Gives increasing weight to higher spatial frequencies – Amplifies Noise SPECT FIlters signal 1 Ramp 0.8 Shepp-Logan 0.6 Modified Shepp_Logan 0.4 Hanning 0.2 Hamming 0 0 0.2 0.4 0.6 0.8 -0.2 Frequency (fraction of Nyquist) 1 Butterworth Problems with Filtered Back Projection • Back projection is mathematically correct, but real life images require Filtered Back Projection • Back Projection can introduce noise and streaking artefacts – Not good with attenuation correction • Filtered Back Projection can reduce noise and artefacts, but may degrade resolution Iterative Reconstruction • FBP was primary method for reconstructing CT images for many decades • Modern computing power now enables CT scanners to use iterative reconstruction • This process begins with an initial estimate of what the object may look like • This guess is used to compute forward projected data sets, which are compared with the measured projection data sets • The difference between the computed and measured projections creates an ‘error matrix’ • The error matrix is used to correct each iteration until the CT image looks like the thing that was scanned Iterative Reconstruction Helical and Multi-Slice Scanning Helical Scanning • Have discussed simplest form of CT scanning – Can produce transaxial slices with the patient being moved along the z-axis between each rotations – ‘Step and shoot’ – This is now very rare • Helical scanning is now the standard – Slip rings – Continuous table feed through gantry Helical Scanning • Slip ring technology – X-ray tube has to be supplied with constant power – Detectors have to pass signals to computer – Not possible if gantry was wired – cables would become entangled and overstretched – Slip ring is a metal ring mounted on the gantry – Good connection while the gantry is free to rotate First we need the ‘scout view’ • Scout views are needed prior to the scan • Performed to allow the planning of the CT sequence • Scout views are produced lines by line at a fixed projection angle – Typically AP Helical Scanning • Patient moves continuously through the gantry as the X-ray tube and detectors rotate • Continuous acquisition of data in a single exposure • Can be visualised as a ribbon wrapped around the body • This technology minimises slice misregistration Contiguous scan Helical Scanning • The position at which sections can be reconstructed can be anywhere within scanned volume other than at the ends • For example: – 300 mm long volume scanned – 10 mm slice width – Pitch = 2 • Only 15 rotations required • From the measured data, 30 contiguous slices, each 10mm thick can be reconstructed (for a single slice scanner) • For thinner slices we need multi-slice scanners Advantages of Helical Scanning • Speed – – – – – – No need to pause between scans for table movement Pitches greater than 1 allowed (reduction in dose) Longer scan lengths within breath hold Reduced patient movement artefacts Increased throughput Reduced use of contrast medium Disadvantages of Helical Scanning • Broadening of Slice profile – Effective slice thickness increases – poorer z axis resolution – Higher noise • Helical artefacts not seen in axial scanning • Possibility of very high dose if pitch < 1 • Lot of tube heating and loading Multi Slice CT Remember!! Multiple detectors in single row Multislice CT • In its original form, CT scanning was twodimensional – 2D slices through the body • True 3D imaging requires isotropy – Voxel size must be equal in all directions • Under these circumstances, data generated in a 3D matrix can be reconstructed in any plane Multislice CT • Voxel size in axial plane is dependent on matrix size and field of view • Typically 1mm • Single slice helical scanners have the capability of collimating the beam width (in the patient direction) to 1mm, but this is restricted by scan time • Not possible in practice so we need to have multislice technology Multislice CT • • • Key to 3D scanning is the multislice scanner These scanners use solid state detectors with multiple rows of detectors Typical configuration for an 8 slice scanner – 12 curved detector rows – Each row has approx 800 detectors – Each row has minimum possible gap between them – Central rows have approx half the length of 2 outer rows – Length of central rows 0.5 – 1mm – Rows can be used separately or in combination Multislice CT • Four possible combinations here are possible: – – – – (a) 8 x 1mm slices (b) 8 x 2mm slices (c) 4 x 4mm slices (d) 2 x 8mm slices • Eight slice scanner here is capable of producing scans at four slice widths, 1,2,4, or 8mm • In each case, slice width is determined by the detector size and by collimation • Scanners up to 256 slices now available Beam width is varied Physical acquisition Computer reconstruction Single slice 3 rotations Multi slice One rotation Single slice Multi slice Dose and Multi-Slice Scanners • Considerations similar to those of single slice scanners • Dose utilisation on z axis usually poorer than with single slice scanners – X ray beam width is generally broader than the total imaged width – Geometric efficiency down to 50% for very small slice thicknesses (sub mm) Geometric Efficiency Dose and Helical CT • All helical scanning requires extra irradiation at the end of each run to obtain sufficient interpolation data to reconstruct the required volume • On multi-slice scanners this extra length can be quite long Helical & Multi-Slice in the UK • Helical & Multi-Slice scanning represent significant steps forward in CT – Better scanning of previous scans – Expansion of workload • Nearly all scanners sold in UK are multi-slice • Technology is still advancing – 32/40/64/256 slice scanners now available – More slices in the future? Next Lecture: • Dosimetry • Operator controls and affects on dose and image quality • artefacts