Survey
* Your assessment is very important for improving the workof artificial intelligence, which forms the content of this project
* Your assessment is very important for improving the workof artificial intelligence, which forms the content of this project
Basic Concepts and Devices RT 31 Humidification Basic Concepts • Functions of the of the upper airway: assure that inspired gas is: – Warmed (convection). – Humidified via evaporation from the mucosa – Filtered • During exhalation: – Expired gas transfers heat back to the mucosa (convection) – Condensation occurs on the mucosal surfaces and water is reabsorbed by mucus (rehydration) Basic Concepts • As gas travels through the lungs it achieves BTPS: – Body temp ~ 37C – Barometric pressure – Saturation with water vapor (100% relative humidity @ 37C) Basic Concepts • The point at which this occurs is called the isothermic saturation boundary (ISB) – Usually occurs ~ 5 cm below the carina – If the upper airway is bypassed or VE is significantly higher than norm, • The ISB will be deeper into the lungs and HUMIDITY therapy may be indicated Basic Concepts • One of the most important, but least understood, aspects of pulmonary care is the role of humidity therapy. • Many care providers and most patients do not appreciate the role of hydration in liquefying secretions and facilitating the natural flow of mucus from the lower airways. Basic Concepts • Pulmonary patients need: – adequate humidification of their inspired gases – controlled fluid balance – otherwise patients can become dehydrated. • Dehydration can make secretions more viscous and inhibit the mucociliary escalator activity of the airways, making secretions difficult to dislodge. • If these secretions block functional gas flow through the distal airways infections, atelectasis and other respiratory problems can easily occur. Basic Physical Principles of Humidity • Humidity is essentially the water vapor in a gas. • This water vapor can be described in several ways, as: • 1. Absolute humidity - The actual content of water vapor in a gas measured in milligrams per liter. • 2. Potential humidity - The maximum amount of water vapor that a gas can hold at a given temperature. • 3. Relative humidity - The amount of water vapor in a gas as compared to the maximum amount possible, expressed as a percentage • 4. Body humidity - The absolute humidity in a volume of gas saturated at body temperature of 37 C; equivalent to 43.8 mg/L Formulas Used When Calculating Humidity • %RH=(absolute humidity/saturated capacity) x 100 – Refer to table 5-3 of Egan or 4-1 of Mosby. Calculations are based on temperature. See example on page 90 of Egan • %BH = (absolute humidity/43.8mg/L) x 100 – See example on page 97 of Egan • Absolute humidity: Refer to table 5-3 of Egan Primary Humidity Deficit • If the atmosphere's relative humidity is less than 100%, the air of the atmosphere has what is referred to as a humidity deficit. • If outside air at 20°C has 14 mg/l of water vapor, and needs to have 17.3 mg/l to be fully saturated, it is said to have a primary humidity deficit of 3.3 mg/l. – 17.3 mg/L (potential) – 14 mg/L (absolute) = 3.3 mg/L (primary deficit) • Remember that the potential is based temp • The primary humidity deficit occurs in the atmosphere and represents the difference between what humidity there is and what there could be. • Primary Humidity Deficit = Potential Water Vapor Content - Actual Water Vapor Content Secondary Humidity Deficit • This is the moisture deficit in the inspired air that the nose and upper airway need to compensate for. – The amount of water vapor the body needs to add to inspired air to achieve saturation at body temperature. • When air is breathed into the nasal cavity and heated to body temperature, its potential water vapor rises to 44 mg/l, which is the potential water vapor content of air at 37°C. • Therefore, unless the air of the atmosphere is at least 37°C and fully saturated, there exists a moisture deficit. • Secondary Humidity Deficit = 44 mg/l - Absolute Humidity. How does a patient develop a humidity deficit? • Breathing a gas with little or no humidity and very high minute volume evaporation of the respiratory mucosa occurs • Bypass of upper airway: intubation, tracheostomy. • Dehydration due to illness, exposure, etc..... • Please understand Figure 4-3 of Mosby, page 94. Water Losses • Insensible: skin and lungs • Sensible: urine, GI tract, sweat • Additive: vomiting, diarrhea, suction from intestines, severe burns, and fever • For each degree of temperature above 99F for over 24 hours, 1000m of fluid is required for replacement Water Vapor Correction • Water vapor acts in most ways like any other gas, it creates a partial pressure when it’s in a mixture of gases. • That partial pressure depends – The amount of water vapor present • Which in turn depends on the temperature. • Unlike other gases in the air, changes in the barometric pressure of the atmosphere under normal conditions do not have much impact on the partial pressure of water. Water Vapor Correction • As a result, it is best to calculate the partial pressures of the other gases in the air after the partial pressure of water vapor has been determined--especially when measuring the air within the lungs. • Inside the lungs, the partial pressure of water vapor is approximately 47 mm Hg. • This value is relatively constant because the air entering the lungs is normally saturated and at 37°C. • By subtracting the partial pressure of the water vapor from the total atmospheric pressure, you will find what is referred to as the dry gas pressure Importance of Humidity • It is needed to maintain normal bronchial hygiene • It promotes functions of the normal mucociliary escalator • It maintains the body's vital homeostasis • Without humidity: – the nearly 100 ml of mucus secreted daily would become quite thick and tenacious. – actual lung parenchyma would dry up, causing a loss of normal compliance which would restrict lung movement and reduce ventilation. Importance of Humidity If the upper airway were bypassed or dry gases were inhaled, a series of adverse reactions could occur, including: – Slowing of mucus movement – Inflammatory changes and possible necrosis of pulmonary epithelium – Retention of thick secretions and encrustation – Bacterial infiltration of mucosa (bronchitis) – Atelectasis – Pneumonia – Impairment of ciliary activity Importance of Humidity The general goals of humidity and aerosol therapy are to: 1. Promote bronchial hygiene 2. Loosen dried and/or thick secretions 3. Promote a effective coughs to clear secretions 4. Provide adequate humidity in the presence of an artificial airway 5. Deliver adequate humidity when administering dry gases therapies 6. Delivering prescribed medications Clinical Evaluation of the Need for Humidity and/or Aerosol Use • Patient's age and ability to move normal secretions • • • • Neuromuscular status Recent or planned surgeries Trauma Disease conditions • The presence of any of these may impair the patient's ability to cough and move secretions. • Another problem may occur when patients develop very thick and abundant amounts of secretions which cannot be moved with normal muscle activity--making humidity or aerosol therapy necessary. Indications for delivery of humidified gases and aerosols • Primary indications for humidifying inspired gases include: • Administration of medical gases • Delivery of gas to the bypassed upper airway • Thick secretions in nonintubated patients Indications for delivery of humidified gases and aerosols • Additional indications for warming inspired gases: – Hypothermia – Reactive airway response to cold inspired gas Mucociliary Blanket • It’s natural escalator functions to clear airways via function of the ciliated mucosa. • This mechanism occurs from the larynx to the respiratory bronchioles. • Mucus is produced by goblet cells and submucosal glands. • Clara cells and tissue fluid transudation also contribute to airway secretions. • A wave-like motion of the cilia then move secretions upward toward the larynx where it is either swallowed or expectorated. Mucociliary Blanket Mucociliary Blanket Sources of Mucus • Secretion from goblet cells and bronchial (mucous) glands. • The goblet cells, which are distributed throughout the epithelium of the mucosa, synthesize and secrete mucus into the airway. Sources of Mucus • The mucous glands, which are in the submucosa, are the greater source of mucus. • Chronic irritation or disease can cause the number and size of goblet cells and mucous glands to increase, resulting in a larger and more viscous mucous blanket. Effects of Mucous Layer • Ciliary activity, which moves the mucus, can be adversely affected if the mucous layer is changed. – A higher ratio of gel to sol layer will affect the flow of mucus by increasing cilia workload. • a decrease in the watery sol layer • or an increase in the viscous gel layer – The cilia are capable of continuing to beat even if the workload increases, but only to a certain level. – If the cilia become tangled in the thick mucus or are unable to penetrate the dense layer, the transport of the mucous blanket would stop, causing secretions to become retained in the respiratory tract. Other factors that can impede ciliary activity and the flow of mucus include: • Tobacco smoke • Local environmental conditions • Pathology of the airway can impede clearance due to changes in the epithelium. Sign/Symptoms of Inadequate Airway Humidification • • • • Atelectasis Dry, nonproductive cough Increased airway resistance Increased in incidence of infection • Increased work of breathing • Substernal pain • Thick, dehydrated secretions Humidification Devices • The purpose of humidifiers is to deliver a gas with a maximum amount of water vapor content. • May be heated or unheated, and the factors affecting the efficiency of humidification devices include: – temperature – time of exposure between gas and water – surface area involved in the gas/water contact Humidification Devices • As temperature rises, the force exerted by the water molecules increases, enabling their escape into the gas, adding to the humidity. – So the higher the tempthe more humidity • Longer exposure of a gas to the water increases the opportunity for the water molecules to evaporate during the humidifier's operation. • The greater the area of contact between water and gas, the more opportunity for evaporation to occur. Humidification Devices • Space-efficient methods – Bubble diffusion – Aerosol – Wick technologies Humidification Devices • Bubble diffusion: – Stream of gas is directed underwater – The gas is broken up into small bubbles – As gas bubbles rise, evaporation increases the water vapor content within the bubble Humidification Devices • Aerosol: spraying water particles into gas – Aerosol (suspended water droplets) is generated in the gas stream – The greater the aerosol density (# of molecules), the greater the gas/water surface area available for evaporation Humidification Devices • Wick: – Use porous waterabsorbent materials to increase surface area – A wick draws water into its fine honeycombed structure by means of capillary action – The surfaces of the wick increase the area of contact between the water and gas Blow-By • This “pass-over” type humidifier directs a dry gas source over a water surface area, and flowing it to the patient. • Because exposure area and time of contact is limited and it is not heated, this unit is not very efficient. • These units are often used in incubators and in certain ventilators, although many times the use of a heated element is added to improve this humidification system • Wick type or membrane type Bubble Humidifier • • • • • • • • Low-flow gas system Provides flow lower than patient’s inspiratory needs. Oxygen or air is humidified at 30%-50% relative humidity. Gas is passed below the water’s surface in the form of bubbles. Increase exposure time result in good humidification Patient’s airway provide further humidification. Should be used for NC use at 4lpm or above, but the higher the flow rate the less the exposure time. Do not use with oximizers or venturi masks (these are common mistakes in the clinical setting) Jet Humidifier • Forms aerosol • Baffle system to break up particles into smaller sizes • Bernoulli’s principle • Low flow • Large-volume jets are used for bland aerosol Ultrasonic Nebulizers • Electrically powered • Uses piezoelectric crystal to generate aerosol – Transducer converts radio waves into high-frequency mechanical vibrations (sound) – These vibrations are transmitted to a liquid surface creating a geyser of aerosol droplets Mist tents and hood • Kids don’t like things on their face • So tents and hoods are used to deliver bland aerosols • Sometimes referred to as croup tents • High flow rates should be used to prevent CO2 build up • Must use some kind of cooling device to prevent heat retention: refrigeration devices or even ice Cascade Humidifier • High Flow: Provide vapor to entire gas flow • Used for mechanical ventilation, airway bypass (artificial airways) • 100% humidity • Usually heated to body temperature • Figure 4-7 of Mosby, page 99 Cascade Humidifier • Gas enters the cascade and travels to the bottom of the tower • Then it moves up through a sheet of plastic consisting of many tiny holes. • Tiny bubbles are produced and dissipate into water vapor, which are carried to the patient’s delivery circuit. • No back flow is allowed due to one-way valves • A heating element in the water reservoir heats the water to form warm gas. • Thermostat can be used to regulate and monitor temperature. Heat and Moisture Exchangers (HME) or Artificial Noses • Functions similarly to the upper airway • Captures s exhaled heat and moisture and using it to heat and humidify inhaled gas. • Do not add heat or moisture--Use the body’s own heat and moisture. • Book statement: should be used short term, flow less than 10 lpm, and in the absence of thick secretions • Practical purposed: Used all the time! Changed every 24 hours. Heat and Moisture Exchangers (HME) or Artificial Noses • Light weight • Less dead space • Reduce accumulation of condensation in tubing • Decrease risk of infection (maybe) • If moisture remains on filter for an extended period of time, airway resistance increases. • Must be removed during aerosol therapy • Dead space volume limits use for neonates and pediatrics. Aerosol Therapy Basic Concepts and Delivery Systems Aerosol Therapy • It is important to remember that an aerosol is not the same as humidity. • Humidity is water in a gas in molecular form, while an aerosol is liquid or solid particles suspended in a gas. • Examples of aerosol particles can be seen everywhere: as pollen, spores, dust, smoke, smog, fog, mists, and viruses. Aerosol Therapy • Aerosol therapy is designed to increase the water content delivered while delivering drugs to the pulmonary tree • Deposition location is of vital concern • Some factors that affect aerosol deposition are aerosol particle size and particle number. Aerosol Output • The actual weight or mass of aerosol that is produced by nebulization. • Usually measured as mg/L/min also called aerosol density • Aerosol output does not predict aerosol delivery to desired site of action. Particle Size The particle size of an aerosol depends on the device used to generate it and the substance being aerosolized. Particles of this nature, between 0.005 and 50 microns, are considered an aerosol. The smaller the particle, the greater the chance it will be deposited in the tracheobronchial tree. Particles between 2 and 5 microns are optimal in size for depositing in the bronchi, trachea and pharynx. Particle Size • Heterodisperse: – aerosol with a wide range of particle sizes (medical aerosols) • Monodisperse: – aerosol consisting of particles similar in size (laboratory, industry) Deposition • The aerosol particles are retained in the mucosa of the respiratory tract. They get stuck! • The site of deposition depends on size, shape, motion and physical characteristics of the AIRWAYS Mechanism resulting in Deposition: Inertial Impaction • Moving particles collide with airway surface. – Large particles (>5micros), upper and large airways • Physics: the larger the particle, the more likely it will remain moving in a straight line even when the direction of flow changes. • Physics: greater velocity and turbulence results in greater tendency for deposition Mechanism resulting in Deposition Table: Particle size and area of deposition. Particle Size in Microns 1 to 0.25 1 to 2 2 to 5 5 to 100 Area of Deposition Minimal settling Enter alveoli with 95% deposition Deposit proximal to alveoli Trapped in nose and mouth Mechanism resulting in Deposition: Sedimentation • Particles settle out of aerosol suspension due to gravity. • The bigger it is the faster it settles! • Medium particles: 1-5 microns, central airways • Directly proportional to time. • The longer you hold your breath the greater the sedimentation Mechanism resulting in Deposition: Diffusion • Actual diffusion particles via the alveolar-capillary membrane and to a lesser extent tissue-capillary membranes of respiratory tract • Lower airways: 2-5 microns • Alveoli: 1-3 microns • These values are from your book Deposition of Particles is also affected by: • Gravity – – Gravity affects large particles more than small particles, causing them to rain-out. • Viscosity - The viscosity of the carrier gas plays an important role in deposition. • For example, if a gas like helium, which has a low viscosity and molecular weight, is used as a carrier gas, gravity will have more of an effect upon the aerosol. • Helium is very light and hence can't carry these particles well, leading to rain-out and early deposition. Deposition of Particles is also affected by: • Kinetic activity - As aerosolized particles become smaller, they begin to exhibit the properties of a gas, including the phenomenon of "Brownian movement." • This random movement of these small (below lmm) particles causes them to collide with each other and the surfaces of the surrounding structures, causing their deposition. • As particle size drops below 0.1m, they become more stable with less deposition and are exhaled. Deposition of Particles is also affected by: • Particle inertia (repeated) Affects larger particles which are less likely to follow a course or pattern of flow that is not in a straight line. • As the tracheobronchial tree bifurcates, the course of gas flow is constantly changing, causing deposition of these large particles at the bifurcation. Deposition of Particles is also affected by: • Composition or nature of the aerosol particles - Some particles absorb water, become large and rain-out, while others evaporate, become smaller and are conducted further into the respiratory tree. • Hypertonic solutions absorb water from the respiratory tract, become larger and rain-out sooner. • Hypotonic solutions tends to lose water through evaporation and are carried deeper into the respiratory tract for deposition. • Isotonic solutions (0.9% NaCl) will remain fairly stable in size until they are deposited. Deposition of Particles is also affected by: • Heating and humidifying - As aerosols enter a warm humidified gas stream, the particle size of these aerosols will increase due to the cooling of the gas in transit to the patient. • This occurs because of the warm humidified gas cooling and depositing liquid (humidity) upon the aerosol particles through condensation. Deposition of Particles is also affected by: • Ventilatory pattern - RCPs easily control this by simple observation and instruction. • For maximum deposition, the patient must be instructed to: – Take a slow, deep breath. – Inhale through an open mouth (not through the nose). – At the end of inspiration, use an inspiratory pause, if possible, to provide maximum deposition. – Follow with a slow, complete exhalation through the mouth. Aerosol vs. Systemic • In many cases, aerosols are superior in terms of efficacy and safety to the same systemically administered drugs used to treat pulmonary disorders. • Aerosols deliver a high concentration of the drugs with a minimum of systemic side effects. • As a result, aerosol drug delivery has a high therapeutic index; especially since they can be delivered using small, large volume, and metered dose nebulizers. Methods of Aerosol Delivery • Aerosols are produced in respiratory therapy by utilizing devices known as nebulizers. • There are a variety of nebulizers in use today, but the most common is one in which the Bernoulli principle is used through a Venturi apparatus Bernoulli’s Principle and Nebulizers • When gas flows through a tube, it exerts a lateral wall pressure within that tube due to its velocity. • As the gas reaches a smaller diameter in the tube, the velocity is increased, which decreases lateral wall pressure. • This decrease in diameter within the tube is at a structure called a jet. • Just distal to the jet is a capillary tube that is immersed in a body of fluid. • The decreased pressure is transmitted to the capillary tube and fluid is drawn up it. • When the fluid reaches the jet, it is then atomized. • See Mosby Figure 4-25, pg 115 Bernoulli’s Principle and Nebulizers: The Baffle • The absolute humidity that will be delivered from these devices can be increased by the use of a heater. • A baffle is distal to this atomization process in the stream of gas/fluid flow. – Nebulization takes place here as the liquid is impelled against the baffle. – This baffle causes the larger particles to coalesce and collect in the reservoir. • The smaller particles will be delivered to the patient in aerosol form. Bernoulli’s Principle and Nebulizers: The Baffle • If the baffle is not used, the device is known as an atomizer. • When the baffle is used, it is then called a nebulizer. • In addition to the physically placed baffle, any 90° angle to gas flow can be considered a baffle. • Large bore corrugated tubing should be used with baffles. • This will enable the aerosol particles to be delivered to the patient. Aerosol delivery is accomplished in a variety of ways: nasal spray pump metered-dose inhaler (MDI) dry powder inhaler (DPI) jet nebulizer small volume nebulizer (SVN) large volume nebulizer small-particle aerosol generator (SPAG) mainstream nebulizers ultrasonic nebulizer (USN) intermittent positive pressure breathing (IPPB) devices Metered Dose Inhalers • Metered dose inhalers (MDIs) consist of a pressurized cartridge and a mouthpiece assembly. • The cartridge, which contains from 150-300 doses of medication, delivers a premeasured amount of the drug through the mouthpiece when the MDI is inverted and depressed. • See Mosby Figure 4-32, pg 119 Metered Dose Inhalers • The particle size of the drug released is controlled by two factors: – the vapor pressure of the propellant blend – the diameter of the actuator's opening. • Particle size is reduced as vapor pressure increases, and as diameter size of the nozzle opening decreases. • The majority of the active drug delivered by an MDI is contained in the larger particles, many of which are deposited in the pharynx and swallowed. Metered Dose Inhalers • Successful delivery of medications with an MDI depends on the patient's ability to coordinate the actuation of the MDI at the beginning of inspiration. • Proper instruction and observation of the patient are crucial to the success of MDI of therapy. • Patients need to be alert, cooperative, and capable of taking a coordinated, deep breath. Patients should be instructed to: Metered Dose Inhalers • Be sure to shake the MDI canister well before using. • Hold the MDI a few centimeters from the open mouth. • Holding the mouthpiece pointed downwards, actuate the MDI at the beginning of a slow, deep inspiration, with a 4-10 second breath hold. • Late actuation, or at the end of the inspiration, or stopping inhaling when the cold blast of propellant hits the back of the throat will cause the medication to have only a negligible effect. • Exhale through pursed-lips, breathing at a normal rate for a few moments before repeating the previous steps. • Patients should also be instructed to rinse their mouths after taking the medication. The advantages of MDI aerosol devices include: • They are compact and portable. • Drug delivery is efficient. • Treatment time is short Disadvantages – They require complex hand-breathing coordination. – Drug concentrations are pre-set. – Canister depletion is difficult to ascertain accurate – A small percentage of patients may experience adverse reactions to the propellants. – There is high oropharyngeal impaction and loss if a spacer or reservoir device is not used. – Aspiration of foreign objects from the mouthpiece can occur. – Pollutant CFCs, which are still being used in MDIs, are released into the environment until they can be replaced by non-CFC propellant material Reservoir Devices for MDI’s (Spacers) • These can be used to modify the aerosol discharged from an MDI. The purposes of these spacers or extensions include: • Allow additional time and space for more vaporization of the propellants and evaporation of initially large particles to smaller sizes. • Slow the high velocity of particles before they reach the oropharynx. • As holding chambers for the aerosol cloud released, reservoir devices separate the actuation of the canister from the inhalation, simplifying the coordination required for successful use. • See Mosby, Figure 4-33, pg 121 Dry powder inhalers (DPIs) • Consist of a unit dose formulation of a drug in a powder form, dispensed in a small MDI-sized apparatus for administration during inspiration. • Because these devices are breathactuated, using turbulent air flow from the inspiratory effort to power the creation of an aerosol of microfine particles of drug, they don't require the hand-breath coordination needed with MDIs. Dry powder inhalers (DPIs) • Cromolyn sodium and albuterol are the two primary drugs available in powder form. • Cromolyn sodium is dispensed in a device called the Spinhaler, which pokes holes in capsules containing the powdered drug. • The albuterol formulation is dispensed in a device called the Rotohaler, which cuts the capsule in half, dropping the powdered drug into a chamber for inhalation. • In both cases, a single-dose micronized powder preparation of the drug in a gelatin capsule is inserted into the device prior to inhalation. • See Mosby Fig 4-39,40 The advantages of using DPI devices for drug administration include: • They are small and portable. • Brief preparation and administration time. • Breath-actuation eliminates dependence on patient's hand-breath coordination, inspiratory hold, or head-tilt needed with MDI. • CFC propellants are not used. • There is not the cold effect from the freon used in MDIs, eliminating the likelihood of bronchoconstriction or inhibited inspiration. • Calculation of remaining doses is easy. The disadvantages encountered when relying on DPIs for drug administration include: • Limited number of drugs available for DPI delivery at this time. • Dose inhaled is not as obvious as it is with MDIs, causing patients to distrust that they've received a treatment. • Potential adverse reaction to lactose or glucose carrier substance. • Inspiratory flow rates of 60Lpm or higher are needed with the currently available cromolyn and albuterol formulations. • Capsules must be loaded into the devices prior to use. Small volume nebulizers (SVNs) • Gas powered (pneumatic) and are a common method of aerosol delivery to inpatients. • There are a variety of different SVNs available. Each has specific characteristics, especially in regard to output. • Bernoulli’s principle: make sure you understand this concept Small volume nebulizers (SVNs) • Two subcategories: mainstream and sidestream. • The mainstream nebulizer is one in which the main flow of gas passes directly through the area of nebulization. • The sidestream nebulizer is one in which the nebulized particles are injected into the main flow or stream of gas as with IPPB (Mosby 180-90) circuits. • Don’t spend too much time on IPPB, just know basic concepts and guidelines. • The main difference, based upon their construction, is that the larger particles tend to rain-out with a sidestream nebulizer. Advantages of SVN therapy: • Requires very little patient coordination or breath holding, making it ideal for very young patients. • It is also indicated for patients in acute distress, or in the presence of reduced inspiratory flows and volumes. • Use of SVNs allows modification of drug concentration, and facilitates the aeorsolization of almost any liquid drug. • Dose delivery occurs over sixty to ninety breaths, rather than in one or two inhalations. Therefore, a single ineffective breath won't ruin the efficacy of the treatment. Disadvantages of SVNs include: • The equipment required for use is expensive and cumbersome. • Treatment times are lengthy compared to other aerosol devices and routes of administration. • Contamination is possible with inadequate cleaning. • A wet, cold spray occurs with mask delivery. • There is a need for an external power source (electricity or compressed gas). Large-Volume Nebulizers • These units also have the capability for entraining room air to deliver a known oxygen concentration. • They can deliver varying concentrations of oxygen. When using these units, you should always match or exceed the patient's peak inspiratory flow rates. • This assures delivery of oxygen and nebulized particles. • These units produce particle sizes between two and ten microns and may be heated to improve output. Ultrasonic Nebulizers (USN) • Ultrasonic nebulizers work on the principle that high frequency sound waves can break up water into aerosol particles. • This form of nebulizer is powered by electricity and uses the piezoelectric principle (ability to change shape when a charge is applied). • This principle is described as the ability of a substance to change shape when a charge is applied to it. Ultrasonic Nebulizers (USN) • Contains a transducer that has piezoelectric qualities. • When an electrical charge is applied, it emits vibrations that are transmitted through a volume of water above the transducer to the water surface, where it produces an aerosol. • The frequency of these sound waves is between 1.35 and 1.65 megacycles, depending on the model and brand of the unit. Ultrasonic Nebulizers (USN) • Their frequency determines the particle size of the aerosol. • The transducers that transmit this frequency are of two types. • One type is the flat transducer, which creates straight, unfocused sound waves that can be used with various water levels. • The other type is a curved transducer, which needs a constant water level above it because its sound waves are focused at a point slightly above the water surface. • If the water level falls below this point, the unit loses its ability to nebulize. Ultrasonic Nebulizers (USN) • The particle size falls in the range of .5 to 3 microns. • The amplitude or strength of these sound waves determines the output of the nebulizer, which falls in the range of 0 to 3 ml/minute and 0 to 6 ml/minute. • Ultrasonic nebulizers also incorporate a fan unit to move the aerosol to the patient. This fan action also helps cool the unit. • The gas flow generated by this fan falls in the range of between 21 and 35 liters/minute. This flow of air also depends on the brand and model of the unit. Ultrasonic Nebulizers (USN) • The transducer of an ultrasonic nebulizer is often found in the coupling chamber, which is filled with water. • This water acts to cool the transducer and allows the transfer of sound waves needed for the nebulizer, which takes place in a nebulizer chamber. • The nebulizer chamber is found just above the coupling chamber. These two chambers are usually separated by a thin plastic diaphragm that also allows sound waves to pass. • When studying ultrasonic nebulizers, remember that output is controlled by amplitude, and particle size is controlled by frequency. The advantages of Ultrasonic Nebulization are: • High aerosol output • Smaller stabilized particle size • Deeper penetration into the tracheobronchial tree (alveolar level) • Useful in the treatment of thick secretions that are difficult to expectorate, and they can help to stimulate a cough. • The therapy can be delivered through a mouthpiece or face mask. Therapy can be given with sterile water, saline or a mixture of the two. Small-particle aerosol generator (SPAG) • This is a highly specialized jet-type aerosol generator designed to for administering ribavirin (Virazole), the antiviral recommended for treating high risk infants and children with respiratory syncytial virus infections. Advantages of Aerosol Therapy as a Whole: • Systemic side effects are fewer and less severe than with oral or parenteral therapy • Inhaled drug therapy is painless and relatively convenient. Aerosol doses are smaller than those for systemic treatments. • Onset of drug action is rapid. • Drug delivery is directly targeted to the respiratory system. Disadvantages as a Whole: • Special equipment is often needed for its administration. • Patients generally must be capable of taking deep, coordinated breaths. • There are a number of variables affecting the dose of aerosol drug delivered to the airways. • Difficulties in dose estimation and dose reproducibility. • Difficulty in coordinating hand action and breathing with metered dose inhalers. • Lack of physician, nurse, and therapist knowledge of device use and administration protocols. • Lack of technical information on aerosol producing devices. • Systemic absorption also occurs through oropharyngeal deposition. • The potential for tracheobronchial irritation, bronchospasm, contamination, and infection of the airway. The common hazards of aerosol therapy are: • Airway obstruction - Dehydrated secretions in the patient's airways may absorb water delivered via aerosol and swell up large enough to obstruct airways. – To avoid this, watch the patient very closely and let him progress with therapy at a reasonable rate. You may want to have suction apparatus on hand. • Bronchospasms - It is common for aerosol particles to cause this condition (especially among asthmatics) and it is more prevalent when administering a cold aerosol as compared to a heated one. – If a very large amount of coughing occurs, stop therapy and give the patient a rest. If this persists in farther therapy, stop treatment and notify the physician. The common hazards of aerosol therapy are: • Fluid overload - This can occur when administering continuous aerosol therapy. It can happen quite frequently when treating infants or patients in congestive heart failure, renal failure or patients who are very old and immobile. • In the infant, because of the smaller body size and possible underdeveloped fluid control mechanism, a quantity of water that an adult can easily handle will cause fluid overload. • In a patient with congestive heart failure, any addition of fluid to the vascular system will put an increased strain on the heart. • In a patient with renal failure who is probably already in fluid overload, it is easily seen that you will not want to increase the fluid volume. • In older patients, the fluid control mechanisms may be impaired due to age. Physician orders for aerosol therapy should contain identification of: • • • • • • • Type of aerosol Source gas (FI02) Fluid composition (NaCl, water, etc.) Delivery modality Duration of therapy Frequency of therapy Temperature of the aerosol Charting should include: • time of administration • • • • • • • • • duration of therapy type or composition of the aerosol (NaCl) pulse respiratory rate and pattern breath sounds characteristics of sputum if sputum was or was not produced the ease of breathing benefits observed and any other relevant observations. The reasons for administering aerosol therapies include: • • • • For bronchial hygiene Hydrate dried secretions Promote cough Restore mucous blanket • Humidify inspired gas • Deliver prescribed medications • Induce sputum lab culture Bronchial Hygiene Bronchial Hygiene • Techniques designed to help mobilize and remove secretions and improve gas exchange • PDPV, CPT, modified breathing/coughing techniques, and new devices • Broad application is ineffective and expensive • If combined with exercise, and used when indicated, it can be a improve lung function – Component of comprehensive respiratory care Bronchial Hygiene • Insufficient evidence to support or refute its use with COPD, CB, or bronchiectasis • Successful outcomes require: knowledge of normal/abnormal physiology patient evaluation and selection clear definition of therapeutic goals rigorous application of appropriate methods on-going assessment follow-up evaluation Normal Clearance requires: • a patent airway • functional mucocilliary escalator (larynx to respiratory bronchioles) • effective cough most important protective reflex Four components to an effective cough: • • • • Irritation Inspiration Compression Expulsion Four components to an effective cough: • Irritation Abnormal stimulation provokes sensory fibers to send impulses to he brain’s medullary cough center Stimulus is either inflammatory, mechanical, chemical or thermal Four components to an effective cough: • Inspiration: Cough center generates a reflex stimulation of the respiratory muscles to initiate a deep inspiration Four components to an effective cough: • Compression Reflex nerve impulses cause glottic closure and a forceful contraction of the expiratory muscles This causes rapid rise in pleural and alveolar pressure Four components to an effective cough: • Expulsion – Glottis opens – Large pressure gradient is present – Causes a violent, high-velocity, expulsive flow combined with dynamic airway compression creates a shearing force that displaces mucus for the walls into the airstream Abnormal Clearance is caused by an alteration in • Airway patency • Mucociliary function • Effectiveness of cough reflex Abnormal Clearance • Airway patency full airway obstruction mucus plugging can result in atelectasis with the possiblitiy of deoxygenation due to shunting Inadequate humidification can result in this partial obstruction (reduced airflow) increase work of breathing airtrapping overdistention VQ mismatch Abnormal Clearance • Mucociliary function high FiO2 can impair directly or due to tracheobronchitis Abnormal Clearance • Effectiveness of cough reflex Abnormal Clearance • Therapeutic interventions • Abnormal clearance in the presence of a pathogenic organism may result in infection • Infectious process inflammatory response and release of chemical mediators damage to airway epithelium and increase mucus production cyclical activity Phase Disruption • Irritation Anesthesia CNS depression Narcotics Phase Disruption • Inspiration: • • • • Pain Neuromuscular dysfxn Pulmonary restriction Abdominal restriction Phase Disruption • Compression Laryngeal nerve damage Artifical airway No mucocillary escalator Erosion of trachea Prevent closure of glottis Abdmonial muscle weakness Abdominal surgery Phase Disruption • Expulsion Airway compression Airway obstruction Abdominal muscle weakness Inadeaqute lung recoil Diseases • Internal obstruction or external compression FBO Mucus hypersecretion Inflammatory changes Bronchospasm Asthma CB Pneumonia pneumonitis • Tumor • Kyphoscoliosis Diseases • Alteration in mucocilliary escalator CF (viscous secretions) Ciliary diskinetic syndromes (cilia don’t work right) Bronchiectasis (occurs w/ CF & Ciliary diskinetic syndromes) Permenent airway damage Dilated airway Constant obstruction Diseases • Reflex Neuromuscular disorders Muscular dystrophy Amytrophic muscular sclerosis MS Polymyelitis • Cerebral palsy Goals • Mobilize and remove retained secretion • Improve gas exchange • Reduce WOB Indications • Acute Acutely ill with copious secretions Acute respiratory failure with clinical signs of retained secretions Lobar atelectasis V/Q abnormalities due to unilateral lung infiltrates or consolidation Probably not helpful for: Pneumonia without significant sputum production COPD Uncomplicated asthma Indications • Chronic: > 25-30 ml/day to be effective ( 1 fluid oz or shot glass full) CF Bronchiectasis Ciliary dyskinetic syndromes Chronic bronchitis Indications • Prevention Body position Pt Mobilization PDPV combined with exercise to maintain normal function in CF – Possible NM disorders Determining the need: • Bedside assessment Ineffective cough Absent or increased sputum production Labored breathing pattern Decreased breath sounds Crackles or rhonchi Tachypnea, tachycardia Fever General physical fitness Posture, muscle tone Determining the need: • Chart H/O secretion retention or dz process indicating such Upper abdominal or thoracic surgery Age H/O COPD Obesity Nature of procedure Type of anesthesia Duration of procedure Intubation or trach CXR: atelectasis or infiltrates PFT ABG Bronchial Hygiene Methods: all can be used alone or in combination with another • PD & P includes – postural drainage and turning – Percussion – Vibration • Coughing and repulsion techniques – – – – PAP adjunts PEP CPAP Expiratory PAP (EPAP) • High-frequency compression/oscillation methods • Mobilization/exercise Postural Drainage Therapy: • Involves the use of gravity and mechanical energy to Aid in mobilizing secretions Improve V/Q balance Normalize FRC • Includes Turning Drainage Percussion vibrations Turning • Kinetic Therapy or continuous lateral rotational therapy • Done by – Patient – Caregiver – Rotational bed • RotoRest Delta Bed rotates continuously side to side (124 degree angle over 3-4 minutes) • Reposition can be accomplished by using automated inflation and deflation of air-filled mattress compartments Turning • Primary Purpose – Promote lung expansion – Improve oxygenation – Prevent retention of secretions • Other benefits – Reductions of venostasis – Prevention of skin ulcers Turning • Absolute contraindications – Unstable spinal chord injuries – Traction of arm abductors • Relative contraindications – – – – – – – Severe diarrhea Marked agitation Rise in intracranial pressure (ICP) Large drops in blood pressure (>10%) Worsening dyspnea Hypoxia Cardiac dysrhythmias Turning • Hazards – Ventilator disconnection – Accidental extubation – Aspiration of ventilator condensate – Disconnection of vascular lines or urinary catheters Turning • Proning – Used in pts with Acute Lung Injury (ALI) – Improves oxygenation without negative effects on hemodynamics • May allow for lower FiO2 and lower pressure • Not shown to improve survival though Turning • Possible reasons for improved oxygenation – Transpulmonary pressure in this position probably exceeds the airway opening pressure in the lung regions where atelectasis, shunt, and V/Q mismatch are most severe – May shift blood away from shunt regions via gravity, which induces recruitment of previously atelectatic but healthy areas – Reduces further injury from PPV Postural Drainage • Use of gravity to help move secretions from distal lung segments – May be coughed up – Or suctioned out • Affected lung segmental bronchus to be drained in a vertical position relative to gravitational pull • Positions are usually held for 3-15 minutes – Depends on tolerance and condition Postural Drainage • Most effective if – Sputum production is >25-30 ml/day – Head-down positions exceed 25 degree below horizontal – Pt is adequetly hydrated • Airwaymay need bland aerosol • SystemicIV NS – Performed every 4-6 hours • Or as appropriate given pt response Postural Drainage • Technique – Identify appropriate lobe or segment – Determine position and need for position modification given your assessment • • • • Unstable hemodynamics HTN Cerebrovascular disorders Orthopnea – Schedule treatment at least 1.5-2 hours after meals to prevent aspiration – Assess need for pain meds Postural Drainage • Assess pt surroundings – – – – Monitors IV or other lines NG O2 • Explain procedure to the patient – Secretions don’t always come up immediately. – May take several txs to be successful • Assess vitals and pulse-ox pre, during and post • Assess breath sounds pre and post • Encourage appropriate coughing techniques pre, during, and post Postural Drainage • Other assessments – Subjective response – Breathing pattern, symmetrical movement, etc. – Mental function – Skin color – SpO2 – ICP Postural Drainage • Recommended interventions upon complications – Hypoxia • Give higher FiO2 during procedure • If hypoxia occurs during tx, give 100% FiO2stop therapyreturn to original position – Increased ICP • Stop therapyreturn to original position Postural Drainage • Recommended interventions upon complications – Acute hypotension during tx • Stop therapyreturn to original position – Pulmonary Hemorrhage • Stop therapyreturn to original positioncall Doc immediatelyO2maintain airway – Pain or injury • Stop therapyreturn to original position carefully Postural Drainage • Recommended interventions upon complications – Vomiting/Aspiration • Stop clear airway/suctionO2maintain airway return to original positioncall Doc – Bronchospasm • Stop return to original position O2call Docbronchodilators as ordered – Dysrhythmias • Stop return to original position O2call Doc Postural Drainage • Outcome assessment: criteria indicating positive response • Should be assessed every 24 hrs for critical and every 3 days for others or upon change in status – Worsening breath sounds is not necessarily bad • Example: diminished to rhonchisecretions have loosened Postural Drainage • Outcomes – – – – – – – – Pt’s subjective response to treatment Vitals and ECG Breathing pattern, rate, chest expansion, etc. Sputum production Breath sounds Chest X-ray SaO2, SpO2, ABGs Ventilator variables Postural Drainage • Charting – – – – – Date and time Position(s) Time in position(s) Patient tolerance Subject/objective indicators of tx effectiveness • Sputum color, viscosity, volume – Pre, during, post assessment – Signature Right Lung (3 Lobes) Right Upper Lobe Right Middle Lobe Right Lower Lobe Left Lung Left Upper Lobe Left Lower Lobe Bronchi-Carina Right Upper Lobe Bronchi Right Middle Lobe Bronchi Right Lower Lobe Bronchi Left Upper Lobe Bronchi Left Lower Lobe Bronchi Right Upper Lobe Segmental Anatomy : Apical UPPER LOBES Apical Segment/1 • Bed or drainage table flat. • Patient leans back on pillow at 30 degree angle. • (Clap over area between clavicle and top of scapula on each side.) Right Upper Lobe Segmental Anatomy : Posterior UPPER LOBES Posterior Segment/3 • Bed or drainage table flat. • Patient leans over folded pillow at 30 degrees angle. • (Clap over upper back on each side of chest.) Right Upper Lobe Segmental Anatomy : Anterior UPPER LOBES Anterior Segment/2 • Bed or drainage table flat. • Patient lies flat on back with pillow under knees. • (Clap between clavicle and nipple on each side of chest.) Right Middle Lobe Segmental Anatomy : Medial Right Middle Lobe Segmental Anatomy : Lateral RIGHT MIDDLE LOBE: Lateral Segment-4 Medial Segment-5 • Foot of table or bed elevated 14 inches or about 15 degrees. • Patient lies head down on left side and rotates 1/4 turn backward. Pillow may be placed behind patient from shoulder to hip. • Knees should be flexed. (Clap over right nipple area.) Right Lower Lobe Segmental Anatomy : Superior Right Lower Lobe Segmental Anatomy : Posterior Basilar Right Lower Lobe Segmental Anatomy : Medial Basilar Right Lower Lobe Segmental Anatomy : Anterior Basilar Right Lower Lobe Segmental Anatomy : Lateral Basilar Left Upper Lobe Segmental Anatomy : Anterior Left Upper Lobe Segmental Anatomy : Apicoposterior Left Upper Lobe Segmental Anatomy : Anterior Left Upper Lobe Segmental Anatomy : Superior Lingular Left Upper Lobe Segmental Anatomy : Inferior Lingular LEFT UPPER LOBE Lingular Segment-Superior-4 Inferior-5 • Foot of table or bed elevated 14 inches or about 15 degrees. • Patient lies head down on right side and rotates 1/4 turn backward. • Pillow may be placed behind patient from shoulder to hip. Knees should be flexed. • (Clap over left nipple area.) Left Lower Lobe Segmental Anatomy : Anterior MedialBasilar LOWER LOBES: Anterior Basal Segment/8 • Foot of table or bed elevated 18 inches or 30 degrees. • Patient lies on side, head down, pillow under knees. • (Clap over lower ribs just beneath axilla.) Left Lower Lobe Segmental Anatomy : Superior LOWER LOBES: Superior Segment/6 • Bed or table flat. Patient lies on abdomen with pillows under hips • (Clap over middle of back below tip of scapula on either side of spine.) Left Lower Lobe Segmental Anatomy : Lateral Basilar LOWER LOBES: Lateral Basal Segment/9 • Foot of table or bed elevated 18 inches or 30 degrees. • Patient lies on abdomen, then rotates 1/4 turn upward. • Upper leg can be flexed over a pillow for support. (Clap over uppermost portion of lower ribs.) Left Lower Lobe Segmental Anatomy : Posterior Basilar LOWER LOBES: Posterior Basil Segment/10 • Foot of table or bed elevated 18 inches or 30 degrees. • Patient lies on abdomen, head down, with pillow under hips. Upper leg can be flexed over a pillow for support. • (Clap over lower ribs close to spine on each side of chest.) Assignment for Upcoming Labs • Memorize the segments of the lungs • Memorize the appropriate positions for each segment! Percussion and Vibration • Application of mechanical energy to the chest wall – Hands – Pneumatic devices • Percussionbreak secretions loose for TB tree • Vibration aids in moving secretions toward the central airways Percussion and Vibration • Unclear as to how much force or frequency should be used to be effective • Effectiveness is controversial • Used in conjunction with postural drainage • Percussion over the lobe or segment being drained Percussion • This should be done with the hands in the cupped position, with the thumb and fingers closed to trap air. Percussion • Hold your arms with the elbows partially flexed and wrists loose • Rhythmically strike the chest wall in a waving motion using both hands alternately in sequence. • Percuss back and forth in a circular pattern over the specific segment for 3-5 minutes Vibration technique • Place hands on either side of the chest • After the pt takes a deep breath, exert slightto-moderate pressure ont eh chest wall • Initiate a rapid vibratory motion of the hands throughout expiration Mechanical Percussion and Vibration • Devices have both frequency and force control – 20-30 cycles/second – 20-30 Hz – Noise, excess force, mechanical failure and electrical shock are all potential hazards Coughing • Directed cough (DC) to clear or mobilize secretions is a component of bronchial hygiene – Directed Cough is a deliberate maneuver that is taught, supervised, and monitored. • Forced expiratory technique (FET, or huff cough) and manually assisted cough are examples of directed cough. Coughing • Seeks to mimic the attributes of an effective spontaneous cough (or series of coughs) • To help to provide voluntary control over reflex • To compensate for physical limitations – – – – increasing glottic control inspiratory and expiratory muscle strength coordination airway stability • Patient should assume position best for exhalation and allows for easy thoracic compression • •Surgical (Thoracic/Abdominal): Splinting to limit pain and anxiety "CASCADE TECHNIQUE" • Breathe in slowly and deeply through the nose. • Breathe out slowly and completely through pursed lips. • Breathe in slowly and deeply once again, then hold breath briefly. • Cough several times until lungs feel empty. The cough should produce a sharp sound. • Avoid taking sharp, quick breaths between coughs. Forced Expiratory Technique "HUFF TECHNIQUE" • Forced expirations of middle to low lung volume without closure of the glottis • Breathe in slowly and deeply through nose. • Breathe out slowly and completely through pursed lips. • Breathe in slowly and deeply once again, then hold breath briefly. • Instead of coughing, let the air out in several short bursts while saying "huff." • A "huff" sound is produced rather than a sharp sound. "ASSISTED COUGH TECHNIQUE” • Pt Breathes in slowly and deeply through the nose and then out slowly and completely through pursed lips. • Pt Breathes in slowly and deeply once again, then holds breath briefly. • Assisting person places hands on pt’s sides at the lower rib cage or on stomach above belly button. • Pt coughs while the person assisting applies gently pressure. • Stop applying pressure when the patient is finished breathing out, but don’t remove hands PAP • Positive airway pressure (PAP) adjuncts are used to mobilize secretions and treat atelectasis and include – continuous positive airway pressure (CPAP) – positive expiratory pressure (PEP) – expiratory positive airway pressure (EPAP). • Cough or other airway clearance techniques are essential components of PAP therapy when the therapy is intended to mobilize secretions PAP: CPAP • The patient breathes from a pressurized circuit against a threshold resistor (water-column, weighted, or spring loaded) that maintains consistent preset airway pressures from 5 to 20 cm H2O during both inspiration and expiration – (By strict definition, CPAP is any level of aboveatmospheric pressure.) • CPAP requires a gas flow to the airway during inspiration that is sufficient to maintain the desired positive airway pressure. PAP: CPAP • Types of threshold resistors: all of these valves operate on the principle that the level of PAP generated within the circuit depends on the amount of resistance that must be overcome to allow gas to exit the exhalation valve. • They provide predictable, quantifiable, and constant force during expiration that is independent of the flow achieved by the patient during exhalation PAP: CPAP • Underwater seal resistor: – expiratory port of the circuit is submerged under a column of water, the level of CPAP is determined by the height of the column • Weighted-ball resistor: – consists of a steel ball placed over a calibrated orifice, which is attached directly above the expiratory port of the circuit PAP: CPAP • Spring-loaded: – rely on a spring to hold a disc or diaphragm down over the expiratory port of the circuit. • Magnetic valve resistors – contain a bar magnet that attracts a ferromagnetic disc seated on the expiratory port of the circuit the amount of pressure required to separate the disc from the magnets is determined be the distance between them. PAP: PEP • The patient exhales against a fixed-orifice resistor, generating pressures during expiration that usually range from 10 to 20 cm H2O • PEP does not require a pressurized external gas source. • The amount of PEP varies with the size of the orifice and the level of expiratory flow produced by the patient. The smaller the orifice the greater the pressure. PAP: PEP • Thus the patient must be encourage to generated a flow high enough to maintain expiratory pressure at 10-20 mm H2O • Ideal I:E of 1:3 or 1:4 • The patient should perform 10-20 breaths through the device and then perform 2-3 huff breath coughs • This should be repeated 5-10 times during a 15-20 minute session PAP: EPAP • The patient exhales against a threshold resistor, generating preset pressures of 10 to 20 cm H2O (similar to CPAP expiration) • EPAP does not require a pressurized external gas source. • EPAP utilizing threshold resistors does not produce the same mechanical or physiologic effects that PEP does when a fixed orifice resistor is used. • Further study is necessary to determine how these differences affect clinical outcome.