Survey
* Your assessment is very important for improving the workof artificial intelligence, which forms the content of this project
* Your assessment is very important for improving the workof artificial intelligence, which forms the content of this project
Nephrol Dial Transplant (1998) 13: 2098–2103 Nephrology Dialysis Transplantation Technical Report Evaluation of an ultrasonic blood volume monitor Christian Johner1, Paul William Chamney2, Daniel Schneditz3 and Matthias Krämer1 1Fresenius Medical Care, Innovation and Technology, Bad Homburg, Germany, 2Lister Hospital, Stevenage, UK, and 3Beth Israel Medical Center, New York, USA Abstract Background. Hypotension complicates approximately 30% of all dialysis treatments. Although the genesis of hypotension is multifactorial, hypovolaemia is thought to play a major role as a direct result of decreased blood volume, particularly during ultrafiltration. The described blood volume monitor enables blood volume to be measured continuously by a non-invasive technique. Methods. The blood volume monitor is based on the principle that the total protein concentration, the sum of haemoglobin and plasma proteins in the vascular space, changes during ultrafiltration. Changes of total protein concentration are determined from the velocity of sound waves in blood, measured using a cuvette in the extracorporeal circuit designed for this purpose. The precision of the blood volume monitor was evaluated in 180 dialysis treatments in 49 patients. The relative blood volume obtained by the monitor was compared with a standard reference method involving calculation of relative blood volume from serial measurements of haemoglobin. Results. A very good agreement between the two methods was achieved (SD=1.70%, r>0.96). The results showed no sensitivity to changes in serum sodium concentration (range 130–145 mmol ). The ‘noise’ introduced in the blood volume signal was low (∏0.2%, sampling rate 10 s) allowing subtle blood volume changes to be detected with high resolution. In addition the device enabled the measurement of haematocrit (Hct) and haemoglobin (Hb) to be made since this is the largest blood component determining total protein concentration. A comparison with the centrifuge method revealed an accuracy of ±2.9 Hct-%, and a comparison with the photometer an accuracy of ±0.8 g Hb/dl. Conclusion. In summary the blood volume monitor allows precise and reliable measurement of relative blood volume. It provides the instrumentation essential for feedback control of relative blood volume during dialysis. Correspondence and offprint requests to: Dr Christian Johner, Fresenius Medical Care, Innovation & Technology, Daimlerstr. 15, D-61352 Bad Homburg, Germany. Key words: blood volume; haemodialysis; haemoglobin; instrumentation; ultrafiltration; ultrasound Introduction Managing optimal fluid status remains one of the most difficult problems despite advances in dialysis [1,2]. In the long term fluid overload is a contributing factor to cardiac stress. In particular the development of left ventricular hypertrophy (LVH ) is a severe risk factor in dialysis patients. More importantly, LVH is a reversible condition which can be achieved by careful monitoring of fluid status in addition to other contributing factors [3], since cardiac death remains the leading cause of mortality [4]. In the short term many dialysis treatments are complicated by the onset of hypotension [5]. There is considerable evidence that hypovolaemia in the vascular compartment plays an important role in the genesis of hypotension. Factors that interfere with blood volume homeostasis such as ultrafiltration, dry weight changes, posture changes, food ingestion, dialysate osmolarity, and antihypertensive drugs can have a profound effect on the changes in blood volume [6–8]. Several factors influence the degree of blood volume reduction during ultrafiltration, the most important is the degree of hydration of the extracellular volume. The ability to shift fluid from the extravascular to the vascular compartment is governed by physical factors such as capillary permeability and oncotic forces [9,10]. A number of sympathetic and hormonally mediated mechanisms cause vasoconstriction which promotes the rate of vascular refilling [11,12]. To provide more comprehensive information on fluid shifts in addition to the long- and short-term effects of fluid overload, objective measurements are necessary. Absolute blood volume, the volume in litres inside the vascular compartment, may be determined by measuring the concentration of an injected tracer or dye, which distributes homogeneously in the vascular compartment. This method is both invasive and intermittent and is not suitable for routine application in the monitoring and control of blood volume. During the © 1998 European Renal Association–European Dialysis and Transplant Association Evaluation of an ultrasonic blood volume monitor 2099 last decade there has been considerable interest in the development non-invasive (extracorporeal ) methods for the determination of relative blood volume (RBV ). The RBV is the ratio of the current blood volume to the initial blood volume at the beginning of dialysis. RBV(t)= c(0) ×100% c(t) [RBV ]=% (1) where c(t) is the time dependent concentration of the blood component, e.g. cells, haemoglobin, plasma protein, total protein. All techniques rely on the principle that these (‘solid’) blood components remain confined to the vascular space. As plasma water is removed by ultrafiltration the concentration of blood components increases. The key clinical application of blood volume monitoring is the routine feedback control of relative blood volume [13]. Preliminary algorithms for controlling the blood volume depending on predefined patient specific criteria have been applied [14] demonstrating a reduction in morbid events such as dizziness, nausea or symptomatic hypotension. A number of methods have been developed for blood volume monitoring including optical [15–17], electrical [18], and mechanical methods [19]. In the case of optical methods for example, it is necessary to take into consideration different absorption characteristics of light in oxygenated and deoxygenated haemoglobin in addition to different scattering properties of blood cells which are affected by changes in osmolarity and plasma protein concentration. The ultrasonic technique—initially introduced by Schneditz [20] has been adopted for the measurement of the relative blood volume for several reasons, among these is the availability of high precision transient time measurement technology and the low costs of disposable materials, and high achievable precision of RBV measurements. Subjects and methods Ultrasonic measurement of blood volume The ultrasonic measurement technique exploits the principle that sound speed in blood depends on total protein concentration, the sum of plasma proteins and haemoglobin [20]. Changes in sound speed can be related to changes in total protein concentration. The relative blood volume may be determined from protein concentration. A polycarbonate measuring cuvette is located in the arterial line of the extracorporeal circuit before the pump as shown in Figure 1. Blood passes through the cuvette from bottom to top, avoiding accumulation of air bubbles. An ultrasonic pulse is transmitted through the cuvette containing the blood. A silicon rubber insert ensures sound coupling to the measuring cuvette. The transit time of the pulse, which is dependent on the speed of sound, through the blood is measured by the BVM (Figure 1). A high-precision temperature measurement (precision <0.1°C ) is required to compensate for the dependence of sound velocity on blood temperature [21]. An empirical function found by Schneditz [22] is used to derive total protein concentration TPC from sound velocity and temperature. Using the principle of mass conservation (sim- Fig. 1. Schematic of the BVM sensor head. ilar to Eq. 1), the RBV can be determined from TPC as function of time. Haematocrit and haemoglobin can be calculated from TPC by simple linear equations assuming a mean plasma protein concentration of 72.5 g/l at the start of treatment. During the priming procedure when the circulating saline in the extracorporeal circuit has reached a stable temperature, the BVM performs automatic calibration. On detection of the presence of blood, measurement of relative blood volume commences as soon as a sufficient stable temperature and sound velocity is reached, typically after 1–3 min. Additional measurements In addition to the measurement of RBV, the BVM determines haematocrit and haemoglobin. In this study, the reference photometer method provided the haemoglobin measurements for comparison with the BVM. Haematocrit was determined via the microcentrifuge. Since some sensor systems for blood volume are known to show artefacts caused by osmolarity changes, both sodium and potassium were measured at periodic intervals during the treatment. Lipids (triglycerides and cholesterol ), plasma proteins, leukocytes, thrombocytes, urea, creatinine, anorganic phosphorus and glucose were measured at start and end of treatment in order to assess a possible influence of a varying blood composition on velocity of sound. Reference method To obtain discrete relative blood volume measurements, haemoglobin (Hb) and haematocrit (Hct) are frequently used. However, Hct is subject to variation caused by osmolality changes. Low plasma osmolality causes erythrocytes to swell and consequently Hct is increased, even if relative blood volume does not change. The photometric cyanmethaemoglobin method is regarded as the most accurate standard by which haemoglobin may be determined [23]: 10 ml of whole blood is withdrawn into a mini-cuvette manually with a pipette and then mixed with a reagent of hexacyanoferrate and potassium cyanide. This 2100 C. Johner et al. procedure lyses the red cells, eliminating light scattering effects. Haemoglobin was determined using the Dr Lange Miniphotometer LP2, which has a numerical resolution of 0.1 g Hb/dl. This device contains an accurate light source at a wavelength of 560 nm and calibration is achieved automatically. However, the overall precision of these measurements is ultimately limited by the accuracy of manual handling. For increased precision each blood sample was measured three times and handled in accordance with NCCLS (National Committee for Clinical Laboratory Standards [23]). The presence of large amounts of lipoproteins may elevate the apparent actual haemoglobin value, because of turbidity. For this reason patients with these conditions (triglycerides >500 mg/dl ) were excluded. Haematocrit was measured from whole blood taken from arterial lines. The blood was centrifuged for at least 5 min at 10 000 r.p.m. using heparin-coated microcentrifuge capillaries. No further anticoagulant was used, and the haematocrit was not corrected for trapped water. Usually a standard deviation of ±0.5 Hct-% was achieved with double readings. Patients Forty-nine patients (29 male, 20 female) with end-stage renal failure and requiring intermittent haemodialysis three times a week were selected for the study. Pregnant patients, patients with hyperlipidaemia (triglycerides >500 mg/dl ), or patients undergoing haemofiltration or haemodiafiltration were excluded. The treatments lasted between 2D and 4 h, the ultrafiltration rate, varied between 390 and 1700 ml/h. Each patient was monitored once per week for three dialysis sessions. Informed consent was obtained for participation in the study, in addition to approval of the ethical committee. The study was undertaken in five centres. This multi-centre co-operation enabled all data to be acquired within 4 months. Fig. 2. Correlation of relative blood volume: the data measured by the BVM are plotted against those calculated by Hb measurements (reference method): RBV =0.97×RBV +2.4%. BVM photometer monitor (BVM ) and the RBV determined from the photometer. The correlation was highly significant (r>0.96, n=882). The data correlate via RBV = BVM 0.97×RBV +2.4%. The mean error between photometer the two methods was 0.07%, standard deviation (SD)= 1.70%, demonstrating the high accuracy and reproducibility of the BVM ( Figure 3). No dependencies of the error on the different centres or ultrafiltration rates could be observed. Protocol During each treatment a total of six blood samples were withdrawn from the extracorporeal system. Samples were taken at the start and at the end of ultrafiltration, with additional samples at 30, 60 and 180 min after start of ultrafiltration. A final sample was taken 10 min after ultrafiltration to monitor blood volume rebound. Haemoglobin was measured from each sample as described above. In order to minimize intra-observer errors concerned with pipetting, the haemoglobin concentration was determined from the mean of three measurements of the same blood sample. Haematocrit was determined from a mean of two measurements by microcentrifuge. Statistics Relative blood volume, haematocrit, and haemoglobin were compared to the corresponding reference method by means of linear regression analysis. The correlation coefficient r, the mean deviation, and standard deviation (SD) were derived. Results Relative blood volume measurements Figure 2 shows the correlation between the relative blood volume (RBV ) measured by the blood volume Fig. 3. Frequency distribution of the deviation in RBV measurements between the BVM and reference method. The frequency distribution shows the percentages of measurements revealing a deviation as indicated on the horizontal axis. Relative blood volume measurements carried out with the BVM deviate only slightly from the reference method (Hb – measurement with photometer): deviation= 0.07%±1.70%. Evaluation of an ultrasonic blood volume monitor 2101 Fig. 4. Intradialytic relative blood volume by BVM and by haemoglobin measurement against time. The insert shows the low noise on the RBV signal (<0.2%). An example of a typical dialysis treatment is displayed in Figure 4. The inlay of this figure shows that the noise on the RBV signal is very low (<0.2%, sampling rate 10 s). Haemoglobin and haematocrit measurements Figure 5 indicates the strong correlation between the photometer Hb measurements and the BVM, respectively (r=0.92). It is apparent from Figure 6 there are no systematic deviations between these two methods. The mean deviation was 0.1 g/dl, SD=0.8 g/dl. The correlation between the Hct value derived indirectly by the BVM and the microcentrifuge Hct measurements was significant (r=0.88). The mean deviation in Hct was −0.5 Hct-%, SD=2.9%. Measurement artefacts The data from the BVM was analysed for sensitivity to different blood compositions. No deviations from the reference method could be correlated with concentrations of the blood components mentioned above, as electrolytes, solutes, lipids, glucose or proteins (r<0.3). Nevertheless the possibility of errors cannot be excluded in case of extreme pathological blood compositions since no such results were observed during this study. Fig. 5. The haemoglobin measurements by the BVM are plotted vs the photometer measurements. The results correlate well (r=0.92). Discussion Symptomatic hypotension can already occur at 5–10% blood volume reduction [24]. Many patients suffering from hypovolaemic hypotension may be resuscitated 2102 Fig. 6. Haemoglobin measurements carried out with the BVM deviate only slightly from the reference method (photometer): deviation= 0.1 g/dl±0.8 g/dl. by administration of as little as 200 ml of saline and postural changes, indicating the sensitivity of the patient to small changes in blood volume. In an average patient with a blood volume of approximately 5000 ml, such a saline infusion represents only a 4% change in blood volume. In order to resolve this change, a blood volume sensor should provide an accuracy of 2%. The results of this evaluation have shown that the blood volume monitor (BVM ) has a standard deviation of 1.7%, which satisfies the measurement criteria required for routine blood volume monitoring. Furthermore, it is necessary to consider that the resolution of the photometer is 0.1 g/dl in the measurement of haemoglobin. This resolution already corresponds to a 1% uncertainty in the determination of relative blood volume (RBV ) by the photometer. The measurement error of the BVM adds only another 0.7% to the observed standard deviation. A clean RBV signal with minimal noise is necessary in order to assure a rapid detection of sudden blood volume changes. Figure 4 demonstrates the quality of the signal which has a noise level of <0.2%. This is a precondition for using the BVM for automatic control of blood volume. The RBV measurement is very precise since it is only the relative change in the concentration of blood constituents which is important. Any technique applied for the monitoring of relative blood volume assumes that the absolute mass of a particular marker (such as haematocrit or protein) stays constant throughout the dialysis treatment. Therefore, irrespective of the absolute value or type of a particular marker at the start of dialysis, all markers must change in the same proportion as ultrafiltrate is removed from the blood during ultrafiltration. In the case of the BVM, the initial value of the aggregate of the marker, determines the speed of sound at the start of dialysis. Changes in the concentration of markers during ultrafiltration C. Johner et al. changes the speed of sound which may be related accurately to a change in relative blood volume. In order to control the dose of erythropoietin in an individual patient it is necessary to monitor haematological parameters on a regular basis introducing additional treatment costs. The BVM measures haematocrit (Hct) with an accuracy of ±2.9% and haemoglobin (Hb) with an accuracy of ±0.8 g Hb/dl. Whereas the error in relative blood volume, which is derived form total protein concentration, is small, the relative error in Hct and Hb measurement is somewhat higher, because the BVM cannot distinguish between intracellular proteins and plasma proteins. Since plasma protein concentration varies between 65 and 80 g/l and the BVM assumes a mean plasma protein concentration (p.p.c.) of 72.5 g/l, the variation in p.p.c. leads to a slightly increased variation in Hct and Hb. In patients whose plasma protein concentration is largely deviating from the normal range of 65–80 g/l the error in Hct or Hb measurement by the BVM could become unacceptably high. Such patients are rare and none was found in the study population. For such patients a comparison of the BVM results to laboratory analyses should be performed to assess whether the BVM shows a sufficient precision. The accuracy of the BVM for measurement of Hct or Hb does not match the precision of a laboratory analysis, but the BVM measurement is sufficient for reliable routine use in the dialysis setting. Some of the error introduced between the laboratory analyser and the BVM measurement of Hct will be due to the fact that the BVM is not disturbed by changes in osmolality in the observed range (265–325 mOsm/kg). A decrease in plasma osmolality, for example due to a decrease of sodium or plasma concentration, leads to swelling of the erythrocytes which would not be detected by the BVM. Therefore the BVM measures a osmolaritycorrected Hct. Conclusion The blood volume monitor provides a continuous, accurate and non invasive measurement of relative blood volume ( RBV ). The measurement is not significantly affected by changes in the blood composition, especially changes in electrolytes and small solutes. The low noise on the RBV signal allows a precise and rapid detection of blood volume changes. This facilitates more objective blood volume control on an individual basis. The BVM additionally offers the determination of haemoglobin and haematocrit continuously with an accuracy that is acceptable for routine haematological monitoring. Acknowledgements. The authors wish to thank Sue Cooke, Prof. Dr Roger Greenwood, Richard Humber (Lister Hospital Stevenage, UK ); Thomas Fernsebner, Dr Martin Gottsmann ( Kliniken Traunstein, Trostberg, Germany); Günther Grimm, Dr Alfred Krause, Klaus Metzner (Fresenius Medical Care International, Innovation and Technology); Dr H.-J. Heinicke, Lutz Herrmann, Silke Symanc ( Krankenhaus Dresden Friedrichstadt, Germany); Evaluation of an ultrasonic blood volume monitor Prof. Dr Nathan W. Levin, Katja Martin (Beth Israel Medical Center, New York, USA); Prof. Dr Helmut Mann, Dr Siegfried Stiller, Patrick Wurth( KfH Aachen, Germany); and Christiane Rode, Prof. Dr Volker Wizemann (Georg-Haas Dialysezentrum Gieben, Germany). 2103 13. 14. 15. References 1. Wizemann V, Schilling M. Dilemma of assessing volume state— the use and limitations of a clinical score. Nephrol Dial Transplant 1995; 10: 2114–2117 2. Ifudu O. The concept of ‘dry weight’ in maintenance hemodialysis: flaws in clinical application. Int J Artif Organs 1996; 19(7): 384–386 3. Kooman JP, Leunissen KML. Cardiovascular aspects in renal disease. Curr Opin Nephrol Hypertens 1993; 2: 791–797 4. USRDS Annual Report, 1996 5. Henderson LW. Symptomatic hypotension during dialysis. Kidney Int 1980; 17: 571–576 6. Kouw PM, Olthof CG, Gruteke P, de Vries PMJM, Meijer JH, Oe PL. Influence of high and low sodium dialysis on blood volume preservation. Nephrol Dial Transplant 1991; 6: 876–880 7. Bogaard HJ, de Vries JPPM, de Vries PMJM. Assessment of refill and hypovolaemia by continuous surveillance of blood volume and extracellular fluid volume. Nephrol Dial Transplant 1994; 9: 1283–1287 8. Leunissen KML., Dry weight in dialysis patients. Dial J 1996 9. Landis EM, Pappenheimer JR. Exchange of substances through the capillary walls. In: Hamilton WF, Dow P ed. Handbook of Physiology. Washington DC, 1963; 961–1034 10. Schneditz D, Roob JM, Oswald M, Pogglitsch H, Moser M, Kenner T. Nature and rate of vascular refilling during hemodialysis and ultrafiltration, Kidney Int 1992; 42: 1425–1433 11. Mann H, Konigs F, Heintz B, Gladziwa U, Kirsten R, Stiller S. Vasoactive hormones during hemodialysis with intermittent ultrafiltration. ASAIO Trans 1993; 36(3): M367–369 12. Enzmann G, Bianco F, Paolini F, Panzetta G. Autonomic 16. 17. 18. 19. 20. 21. 22. 23. 24. nervous function and blood volume monitoring during hemodialysis. Int J Artif Organs 1995; 18(9): 504–508 Santoro A, Mancini E. Clinical significance of intradialytic blood volume monitoring. Int J Artif Organs 1997; 20(1): 1–6 Stiller S, Wirtz D, Waterbar F, Gladziwa U, Dakshinamurty K, Mann H. Less symptomatic hypotension using blood volume controlled ultrafiltration. ASAIO Trans 1991; 37: M139–141 Aldridge C, Wilkinson JS, Fleming SJ, Greenwood RN, Cattell WR. Continous measurement of blood hydration during ultrafiltration using optical methods. Med Biol Eng 1987; 25: 317–323 Paolini F, Mancini E, Bosetto A, Santoro A. Haemoscan: a dialysis machine-integrated blood volume monitor. Int J Artif Organs 1995; 18(9): 487–494 Steuer RR, Harris DH, Weiss RL. Evaluation of a noninvasive hematocrit monitor: a new technology. Am Clin Lab 1991; 10(6): 20–22 Maeda K, Shinzato T, Yoshida F et al. Newly developed circulating blood volume-monitoring system and its clinical application for measuring changes in blood volume during hemofiltration. Artif Organs 1986; 10(6): 452–459 Greenwood RN, Aldridge C, Cattell WR. Serial blood water estimations and in-line blood viscometry: the continuous measurements of blood volume during dialysis procedures. Clin Sci 1984; 66: 575–583 Schneditz D, Moser M, Smolle-Juttner FM, Dorp E, Pogglitsch H, Kenner T. Methods in clinical hemorheology: the continous measurement of arterial blood density and blood sound speed in man. Biorheology 1989; 27: 895–902 Schneditz D, Pogglitsch H, Horina J, Binswanger U. A blood protein monitor for the continuous measurement of blood volume changes during hemodialysis. Kidney Int 1990; 38: 342–346 Schneditz D, Heimel H, Stabinger H. Sound speed, density and total protein concentration of blood. J Clin Chem Clin Biochem 1989; 27: 803–806 National Committee for Clinical Laboratory Standards (NCCLS). Reference Procedure for the Quantitative Determination of Hemoglobin in Blood Vol 4, No. 3, 1984 Maeda K, Morita H, Shinzato T et al. Role of hypovolemia in dialysis-induced hypotension. Artif Organs 1988; 12(2): 116–121 Received for publication: 18.8.97 Accepted in revised form: 30.3.98