Download Multiple-Drug Toxicity Caused by the Coadministration of 4

Survey
yes no Was this document useful for you?
   Thank you for your participation!

* Your assessment is very important for improving the workof artificial intelligence, which forms the content of this project

Document related concepts

Psychopharmacology wikipedia , lookup

Polysubstance dependence wikipedia , lookup

Plateau principle wikipedia , lookup

Bad Pharma wikipedia , lookup

Pharmacognosy wikipedia , lookup

Medication wikipedia , lookup

Drug design wikipedia , lookup

Theralizumab wikipedia , lookup

Prescription drug prices in the United States wikipedia , lookup

Prescription costs wikipedia , lookup

Pharmaceutical industry wikipedia , lookup

Neuropharmacology wikipedia , lookup

Drug discovery wikipedia , lookup

Urban legends about drugs wikipedia , lookup

Pharmacogenomics wikipedia , lookup

Drug interaction wikipedia , lookup

Stimulant wikipedia , lookup

Pharmacokinetics wikipedia , lookup

Mephedrone wikipedia , lookup

Transcript
Journal of Analytical Toxicology, Vol. 34, April 2010
Case Report
Multiple-Drug Toxicity Caused by the Coadministration
of 4-Methylmethcathinone (Mephedrone) and Heroin*
Amber J. Dickson†, Shawn P. Vorce, Barry Levine, and Marilyn R. Past
Division of Forensic Toxicology, Armed Forces Medical Examiner System, Armed Forces Institute of Pathology,
Rockville, Maryland 20850-3125
Abstract
An accidental death caused by the combined use of a new designer
drug, 4-methylmethcathinone (mephedrone), and heroin is
reported. A 22-year-old Caucasian male was found unresponsive in
his living quarters and was transported to the hospital where he
died. During autopsy, needle marks were found along the
decedent’s lower legs and ankles. Investigators discovered the
decedent and his roommate had been using “Black Tar” heroin and
mephedrone. Routine toxicological analysis detected morphine in
the decedent’s blood at 0.06 mg/L. Additionally, 6-acetylmorphine,
morphine, codeine, and doxylamine were detected in his urine. A
designer drug screen, employing a basic liquid–liquid extraction
followed by pentafluropropionic anhydride derivatization, was
used to isolate mephedrone from both blood and urine specimens.
The derivatized extracts were analyzed by gas chromatography–
mass spectrometry (GC–MS) operating in full-scan mode.
Quantitative analysis of mephedrone was performed by
GC–MS operating in selective ion monitoring mode using
methamphetamine-d14 as an internal standard. Mephedrone was
confirmed in the decedent’s blood and urine at 0.50 and 198 mg/L,
respectively. The physiological and pharmacological effects of
mephedrone and any associated toxicity have not been reported.
However, because of its structural similarities with methcathinone
and the high concentration in the decedent’s blood, the overall
contribution of mephedrone to the death could not be minimized.
Therefore, the medical examiner reported the cause of death as
multiple-drug toxicity and the manner of death as accidental.
Introduction
According to the 2007 National Survey on Drug Use and
Health, approximately 153,000 Americans aged 12 or older
were current heroin users (2). In vivo, heroin is rapidly
hydrolyzed to a pharmacologically active metabolite, 6-acetylmorphine (6-AM), which is then hydrolyzed to morphine (1).
Heroin is excreted in the urine primarily as morphine and
morphine glucuronides. Other opium alkaloids, synthetic
artifacts, and cutting agents, such as quinine and diltiazem, can
be found in illicit heroin (1,3). A high risk of overdose is
associated with heroin because the purity of illicit heroin can
vary from 10% to more than 80% based on geographical location (1). Cyanosis, drowsiness, stupor, hypotension, and bradycardia are signs of heroin toxicity. Death from heroin toxicity
is caused by respiratory depression and failure (4). Any other
central nervous system depressant that is coadministered, such
as alcohol or benzodiazepines, dramatically increases the risk
of a fatal overdose (3,4).
4-Methylmethcathinone (mephedrone) is a synthetic stimulant and the 4-methyl aromatic analogue of methcathinone.
Mephedrone is commonly referred to on the streets as 4-MMC,
MMCAT, and Crab, reportedly because of its fishy smell (5,6).
Mephedrone can be found for sale on the internet as a research chemical or plant food and is also marketed as a “legal
high” in head shops in the U.K. (7). Synthesized as a cathinone
analogue, the structural similarities between mephedrone and
methcathinone are displayed in Figure 1. Published data on
mephedrone’s chemical properties, pharmacology, and toxicity were not found. Most of the known information about its
effects can be discovered through internet drug user forums.
Heroin is a semisynthetic diacetyl derivative of morphine
first synthesized in 1874 and later used in 1898 as a pharmaceutical preparation (1). Currently, heroin is not approved or
available for medical purposes in the U.S. and is listed by the
Drug Enforcement Administration (DEA) as a Schedule I drug.
* Disclaimer: The opinion or assertions herein are those of the authors and do not necessarily
reflect the view of the Department of the Army, Navy, or the Department of Defense. This work
was supported in part by the American Registry of Pathology.
† Author to whom correspondence should be addressed. E-mail: [email protected].
162
Figure 1. Structure of 4-methylmethcathinone (mephedrone) and methcathinone.
Reproduction (photocopying) of editorial content of this journal is prohibited without publisher’s permission.
Journal of Analytical Toxicology, Vol. 34, April 2010
Based on user experiences, the stimulant and euphoriant effects of mephedrone are comparable to drugs such as cocaine,
ecstasy, and amphetamine (5). The most common routes of administration for mephedrone are oral and insufflation,
although there have been some reports of intravenous and
suppository administration (5). After administration of
mephedrone, the initial onset of stimulus and euphoria typically lasts 10–30 min, depending on dose and route of administration (5). According to users’ experiences, some individuals
may consume their entire mephedrone supply continuously in
order to maintain the initial effects of the drug (5). Side effects
of mephedrone reported by users include skin rashes, numbness, headache, and minor amnesia (5). More serious side
effects noted after large doses of mephedrone were skin discoloration, light-headedness, cramping, and aches (5). Because mephedrone is a research chemical, there have been no
published studies on its human or animal toxicity, addiction or
acute overdose potential, long-term effects, or allergic reactions. Structural similarities with methcathinone and user reported experiences would suggest that mephedrone shares
many of the same euphoric and stimulant effects.
Recently, there were two reported deaths in Sweden and
Denmark related to possible mephedrone use. One of the individuals was reported to have taken mephedrone, and the
other individual was found to be in possession of mephedrone
prior to death (5). Mephedrone was consequently banned in
both Sweden and Denmark, although the presence of
mephedrone in the two deaths was not confirmed by toxicology results (5). Currently, mephedrone is also banned in
Finland, Israel, and Norway, but it is still legal in other countries such as the U.K. and U.S. (5). According to multiple
sources from the DEA, mephedrone had not been encountered in the U.S. until June 2009 and is considered a rare drug
of abuse. However, in July 2009, a DEA Microgram Bulletin (8)
contained a mephedrone submission from Oregon, but the
presence of mephedrone was not confirmed because no authenticated standard was available. For this reason,
mephedrone is currently unscheduled in the U.S.; however, it
could possibly be controlled under the Federal Analogue Act of
1986 because of its chemical similarities with methcathinone.
This case report describes an accidental death from multipledrug toxicity of mephedrone and heroin as well as the
development of a screening and quantitative method for
mephedrone detection and quantitation in biological specimens.
Case History
The decedent, a 22-year-old Caucasian male, was found unresponsive in his living quarters in the U.S. After emergency
medical services arrived on the scene, they initiated CPR and
transported the man to a nearby hospital. At the hospital, all
life-saving measures were unsuccessful, and the man was pronounced dead. During the autopsy, multiple needle marks
were found along the medial aspect of the decedent’s lower legs
and ankles. There were no other signs of physical trauma on
the decedent’s body. After an investigative search of the dece-
dent’s living quarters, the following items were discovered:
drug paraphernalia, syringes, suspected heroin, and other controlled substances including 2,5-dimethoxy-4-(n)propylthiophenethylamine (2C-T-7). The decedent’s roommate
admitted to possessing the following drugs: “Black Tar” heroin,
mephedrone, and 2C-T-7. He also confessed to providing the
heroin and mephedrone to the decedent. Based on the roommate’s statement, both of them administered mephedrone
doses by insufflation and intravenous and oral routes prior to
injecting heroin in the early morning hours before the decedent was found unresponsive. According to the medical examiner report, the cause of death was multiple drug toxicity and
the manner of death was accidental. The 22-year-old male
roommate submitted a urine sample that was sent to the Navy
Drug Screening Laboratory in Jacksonville, FL (NDSL JAX) for
routine drugs-of-abuse testing. An aliquot of his urine sample
was later submitted to the Division of Forensic Toxicology,
Armed Forces Medical Examiner System (AFMES) for special
testing of 2C-T-7 and mephedrone.
Experimental
Reagents and materials
All solvents were high-performance liquid chromatography
grade and purchased from Fisher Scientific (Pittsburgh, PA).
Monobasic and dibasic sodium phosphate, sodium acetate trihydrate, mepivicaine, and ethylmorphine were purchased from
Sigma Aldrich (St. Louis, MO). Mephedrone and mephedroned3 were purchased from Toronto Research Chemicals (North
York, ON, Canada). Pentafluropropionic anhydride (PFPA) was
purchased from Aldrich (Milwaukee, WI). Concentrated hydrochloric acid (HCl) was purchased from EMD Chemicals
(Gibbstown, NJ). Concentrated potassium hydroxide (KOH)
pellets were also purchased from Fisher Scientific. Methamphetamine-d14 and mescaline-d9 were purchased from Cerilliant (Round Rock, TX). Mixed-mode silica-based solid-phase
extraction columns (ZCDAU020) were purchased from United
Chemical Technologies (Bristol, PA).
Standard, calibrators, and control preparation
For the mephedrone extraction and quantitative procedure,
a stock standard of mephedrone was prepared in methanol
and stored at –15°C in an amber vial. Calibrator samples were
prepared in 2.0 mL of certified drug-free blood by spiking
mephedrone at concentrations of 0.05, 0.10, 0.20, 0.50, 1.00,
and 2.00 mg/L. A blood control was prepared from the same
stock standard source in certified drug-free blood at 0.50 mg/L.
Calibrator and control samples were prepared prior to each extraction. Mephedrone-d3 and methamphetamine-d14 spiking
solutions were prepared for use as the internal standards at a
concentration of 0.01 mg/mL. Negative and positive blood controls were included and extracted with each batch.
Instrumentation
Samples were screened and quantitated using an Agilent
Technologies (Palo Alto, CA) 6890 gas chromatograph (GC)
163
Journal of Analytical Toxicology, Vol. 34, April 2010
coupled to a 5975 mass spectrometer (MS). The separation
columns were from J&W (20 m × 0.18-mm i.d. × 0.18 µm,
Rancho Cordova, CA) with helium carrier gas maintained at a
constant flow of 1.0 mL/min. The MS source and quadrupole
temperatures were held at 230°C and 150°C, respectively. The
transfer line was set to 280°C. Full-scan EI MS data was collected over a mass range of m/z 42–550 and a detection
threshold of 150. Identification of targeted analytes was accomplished using selected ion monitoring (SIM). The SIM
ions collected were m/z 204, 160, 119 for mephedrone and
m/z 211, 163 for methamphetamine-d14; the quantitative ions
are underlined.
Basic drug screen extraction
Blood (1 mL) and urine (2 mL) samples were prepared by the
addition of 3 mL of 0.1 M phosphate buffer (pH 6.0). An internal standard solution containing 0.01 mg/mL mepivacaine
and ethylmorphine was added to each sample at a concentration of 0.50 mg/L. Blood samples were sonicated for 15 min
using an ultrasonic bath. All samples were centrifuged at 3000
rpm for 10 min prior to extraction. Extraction columns were
conditioned successively with 3 mL of methanol, 3 mL deionized water, and 2 mL of 0.1 M phosphate buffer (pH 6.0). Following sample application, the columns were washed with 2
mL of deionized water, 2 mL of 20% acetonitrile in deionized
water, and 2 mL of 0.1 M acetic acid. The columns were dried
under vacuum for 3 min. The final wash steps included 2 mL
of hexane followed by 3 mL of methanol. The columns were
dried under vacuum for 10 min. The analytes were eluted with
3 mL of dichloromethane/isopropanol/ammonium hydroxide
(78:20:2), and eluates were evaporated under nitrogen at 25°C.
The eluates were reconstituted in 50 µL of acetonitrile and then
transferred to glass autosampler vials.
Three microliters of sample was injected using a pulsed split
program of 40 psi for 0.8 min at a 10:1 split and an inlet temperature of 250°C. An initial oven temperature of 70°C was held
for 1 min before ramping at 20°C/min to 300°C. The column
was held for 5.5 min at the final temperature of 300°C for a
total run time of 18 min.
Designer drug screen extraction
Three milliliters of sample was aliquoted, and 75 µL of the
internal standard solution containing 0.01 mg/mL methamphetamine-d14 and mescaline-d9 was added to each sample for
a final concentration of 0.25 mg/L. Four drops of concentrated
KOH and 5 mL of chlorobutane were added to all samples.
Next, the samples were capped and mixed for 20 min on a rotomixer. The samples were then centrifuged for 10 min at
3000 rpm. The top organic layer was transferred with a glass
pipette to clean 10-mL conical test tubes. Prior to drying down
under nitrogen at 40°C, 50 µL of 10% methanolic HCl was
added to each sample. The eluates were reconstituted in 50 µL
of ethyl acetate and 50 µL of PFPA. Each tube was capped,
vortex mixed, and incubated at 70°C for 15 min. The samples
were again evaporated to dryness under nitrogen at 40°C, reconstituted with 25 µL ethyl acetate, and then transferred to
glass autosampler vials.
Three microliters of sample were injected using a split pro-
164
gram of a 10:1 split and an inlet temperature of 225°C. An initial oven temperature of 100°C was held for 1.5 min before
ramping at 10°C/min to 235°C and then 35°C/min to 300°C.
The column was held for 1.5 min at the final temperature of
300°C for a total run time of 18.0 min.
Mephedrone extraction
Two milliliters of sample, calibrators, or control was pipetted
into clean, labeled 16 × 100-mm test tubes. Then 100 µL of the
internal standard solution containing 0.01 mg/mL methamphetamine-d14 was added to each sample for a final concentration of 0.50 mg/L. Four drops of concentrated KOH and 5
mL of chlorobutane were added to all samples. Next, the samples were capped and mixed for 20 min on a rotomixer. The
samples were then centrifuged for 10 min at 3000 rpm. The top
organic layer was transferred with a glass pipette to clean 10
mL conical test tubes. Prior to drying down under nitrogen at
25°C, 50 µL of 10% methanolic HCl was added to each sample.
The eluates were reconstituted in 50 µL of ethyl acetate and 50
µL of PFPA. Each tube was capped, vortex mixed, and incubated
at 70°C for 15 min. The samples were evaporated to dryness
under nitrogen at 55°C, reconstituted with 50 µL ethyl acetate, and then transferred to glass autosampler vials.
One microliter of sample was injected using a pulsed split
program of 40 psi for 1.0 min at a 30:1 split and an inlet temperature of 225°C. An initial oven temperature of 100°C was
held for 1 min before ramping at 20°C/min to 200°C and then
40°C/min to 300°C. The column was held for 0.5 min at the
final temperature of 300°C for a total run time of 9.0 min.
Method validation
Selectivity. The selectivity of the assay was determined by analyzing three sources each of negative blood and negative urine
for potential endogenous and extraneous interferences.
Accuracy and precision. The accuracy of the method was determined by comparing the measured mean quantitated results
with the theoretical spiked concentrations of mephedrone.
The within-day accuracy was measured using five aliquots at
four different concentrations in a single extraction. The between-day accuracy was measured at 4 different concentrations over 10 separate extractions. The measured values could
not deviate more than 20% from the theoretical concentrations
and were expressed as percent difference. The within-day precision was measured using five aliquots at four different concentrations in a single extraction. The between-day precision
was measured over 10 extractions at 4 different concentrations. The precision of the method was expressed as the coefficient of variation (CV) and could not exceed 15%.
Linearity. The minimal criterion for acceptability of a
method’s linearity is to have a line of regression with a correlation coefficient (r2) of 0.99 or greater. The linearity of the
method was determined with a 7-point calibration curve evaluated over 10 extractions. Each calibration batch contained a
negative control and four positive controls spiked at four different concentrations. The calculated value of each calibration point and control could not exceed 20% of the theoretical
spike. The upper limit of linearity (ULOL) was established by
measuring two aliquots at six different concentrations with a
Journal of Analytical Toxicology, Vol. 34, April 2010
seven-point calibration curve. The ULOL was defined as the
highest analyte concentration where the measured value was
within ± 20% of the theoretical concentration and the qualifier
ratios were within ± 20% of the average ion ratios based on the
calibrators.
Limits of detection (LOD) and quantitation (LOQ). The
LODs and the LOQs were established by measuring two
aliquots at five different concentrations with a seven-point calibration curve. The LOD was defined as the lowest analyte
concentration where all qualifier ion ratios were within ± 20%
of the average ion ratios established from the calibrators. The
LOQ was defined as the lowest analyte concentration where the
measured value was within ± 20% of the theoretical concentration and all qualifier ion ratios were within ± 20% of the average ion ratios based on the calibrators.
Recovery. The recovery was determined by comparing an unextracted mephedrone sample to an extracted mephedrone
sample in blood in which the internal standard was added
post-extraction. Recovery was determined at four different
concentrations throughout the linear range. The recovery
values were reported as the percent of drug recovered during
the extraction procedure.
Specificity. A comprehensive panel of commonly encountered alkaline compounds and a designer drug panel were analyzed to determine if any chromatographic or mass spectral
interference would occur with this method.
Chromatographic criteria. The retention time (tR) of the
analytes was required to be within ± 2% of the tR of the calibrator. Qualifier ion ratios of the analytes needed to be within
± 20% of the average ion ratios established from the calibrators. A chromatographic peak should be Gaussian in shape
with tailing factors less than 2.0. The peaks were required to
have resolution from any adjacent peaks so that the valley between peaks is < 10% of the analyte’s peak height.
Results
± 20% of the theoretical concentration). Additionally, the qualifier ion ratios of the LOD and LOQ values were within ± 20%
of the average ion ratios determined from the calibrators.
Recovery. The mephedrone recovery was calculated to be
greater than 80%.
Specificity. There were no interferences observed.
Case samples
A standard postmortem toxicology panel was performed on
the decedent’s blood and urine, which included headspace GC
volatile quantitation, an immunoassay urine drugs-of-abuse
panel, and a full-scan GC–MS basic drug screen. No volatiles
were detected from the GC headspace volatile quantitation including ethanol at cutoff of 20 mg/dL. The urine screened positive by immunoassay for 6-acetylmorphine (6-AM) and was
confirmed by GC–MS SIM at 0.53 mg/L. No 6-AM was detected
in the blood by GC–MS at an LOQ of 0.005 mg/L. Morphine was
detected in the urine by immunoassay and confirmed in blood
and urine by GC–MS SIM at 0.06 mg/L and 27.13 mg/L, respectively. Codeine was detected by GC–MS SIM in the urine at
1.86 mg/L; no codeine was detected in the blood by GC–MS at
an LOQ of 0.05 mg/L. The urine opiate confirmation procedure
involved an acid hydrolysis step; therefore, the previously mentioned urine morphine and codeine values represent total drug
concentrations.
A full-scan GC–MS basic drug screen in the urine detected
mephedrone, doxylamine, morphine, and 6-AM (Figure 2). A
GC–MS SIM basic drug quantitation confirmed the presence of
doxylamine in the urine at 1.99 mg/L; no doxylamine was detected in the blood by GC–MS at an LOQ of 0.05 mg/L. A fullscan GC–MS designer drug screen performed on the decedent’s blood and urine specimens confirmed the presence of
mephedrone in both specimens. However, the GC–MS designer
drug screen did not detect 2C-T-7 in the blood or urine at an
LOD of 0.025 mg/L. A quantitative GC–MS SIM method was developed for the mephedrone confirmation using methamphetamine-d14 as the internal standard with an LOD of 0.010
mg/L and a ULOL of 5.0 mg/L. Mephedrone was confirmed in
Table I. Validation Data for Mephedrone in Blood
Method validation
The mephedrone validation results including accuracy, precision, linearity, LOD, LOQ, and recovery are summarized in
Table I.
Selectivity. There were no interferences observed in the negative blood and negative urine matrices analyzed.
Accuracy and precision. Precision and accuracy met all acceptance criteria. The within-day CVs (n = 20) were less than
6.3% for mephedrone, and between-day CVs (n = 42) were less
than 9.9% for mephedrone. Accuracy was determined for each
aliquot analyzed (n = 62), and no samples exceeded 15.8% of
their theoretical value for mephedrone.
Linearity. Linearity was established from 0.025 to 5.000
mg/L for mephedrone, and the r2 value was 0.9976.
LOD and LOQ. The LOD was determined to be 0.010 mg/L.
The LOQ was established at 0.025 mg/L and defined as the
lowest calibrator. The LOQ values met the criteria for acceptability in regards to precision (CV < 15%) and accuracy (within
Mephedrone
Linearity (n = 10)
0.025–5.000 mg/L
r2
0.9976
LOD
0.010 mg/L
LOQ
0.025 mg/L
Recovery (n = 4)
> 80 %
Precision
Within-day (n = 20)
Between-day (n = 42)
CV < 6.3%
CV < 9.9%
Accuracy*
Within-day (n = 20)
Between-day (n = 42)
2.7–15.8%
0.2–12.4%
* Expressed as % difference from the theoretical value.
165
Journal of Analytical Toxicology, Vol. 34, April 2010
both the decedent’s blood and urine at 0.50 mg/L and 198
mg/L, respectively.
A standard drug-of-abuse toxicology panel was performed on
the roommate’s urine specimen at the NDSL JAX. His urine
screened positive by immunoassay for 6-AM and was confirmed
by GC–MS SIM at 0.007 mg/L. The urine screened positive by
immunoassay for opiates, and both total codeine and morphine were confirmed by GC–MS SIM at 0.30 and 7.67 mg/L,
respectively. Special testing for 2C-T-7 and mephedrone was
performed at the Division of Forensic Toxicology, AFMES. A
GC–MS designer drug screen detected the presence of
mephedrone, but 2C-T-7 was not detected at an LOD of 0.025
mg/L. Mephedrone was confirmed and quantitated in the urine
by GC–MS SIM at 8.10 mg/L.
Discussion
base peak ion (m/z 119) used for the mephedrone confirmation.
Instead, methamphetamine-d14 was used as the internal standard for the mephedrone confirmation because of its similar
chemical structure.
The detection and confirmation of mephedrone in this case
is consistent with the roommate’s statement that he had provided the mephedrone to the decedent just prior to his death.
The roommate also stated the decedent had administered the
mephedrone by multiple routes of administration (insufflation, oral, and intravenous). Because there have been no other
reported mephedrone concentrations in blood or other biological specimens, it is difficult to correlate the quantitative
values or describe the effects that the decedent experienced
with his mephedrone blood concentration. Additionally, the effects and toxicity of mephedrone would be difficult to predict
because of the coadministration of heroin.
Prior to the decedent’s death, the roommate stated that he
had also provided heroin to the decedent and that the decedent
had administered the heroin intravenously. Heroin has a short
half-life of about 2–6 min in blood because it is rapidly deacetylated to 6-AM by blood esterases (9). The half-life of 6-AM,
which is 6–25 min in blood, is longer than that of heroin (9).
6-AM is further hydrolyzed to morphine, which has a half-life
of 2–3 h (9). The presence of 6-AM in the urine demonstrates
that the decedent administered heroin prior to his death. 6-AM
was not detected in the blood because it had already metabolized to morphine. Morphine was detected in the blood at 0.06
In this case, our laboratory’s standard postmortem toxicology panel would not have been able to detect the presence
of the two designer drugs, 2C-T-7 and mephedrone, unless a
specific test request was made. A full-scan GC–MS basic drug
screen of the urine detected an unidentified peak at 6.405 min
with a base peak of m/z 58. A mass spectrum match could not
be made for this peak when using searchable reference libraries. Because of the unidentified peak in the base screen and
the possible use of 2C-T-7 and
mephedrone, the toxicology testing was
expanded to include a GC–MS designer
drug screen. With the ability to detect 32
drugs, including 2C-T-7, our GC–MS designer drug screen was further expanded
to include a reference standard of
mephedrone. An unextracted mephedrone reference standard was analyzed,
both underivatized and derivatized with
PFPA, using the GC–MS designer drug
screen. The mass spectrum of mephedrone and mephedrone-PFPA (Figure 3)
were added to a custom in-house reference library. Once identifiable by GC–MS,
both extraction and instrumentation
methods were developed in order to detect and confirm the presence of
mephedrone.
Additionally, a deuterated mephedroned3 reference standard was purchased to
be used as the internal standard. Unfortunately, the mephedrone-d3 MS fragmentation pattern was very similar to
mephedrone-d0, sharing the same base
peak ion (m/z 119). Furthermore, they
could not be separated chromatographically. For this reason, the use of
mephedrone-d3 as the internal standard
Figure 2. Gas chromatogram and full-scan mass spectrum of mephedrone extracted from the decedent’s
was not possible because it was conurine using a basic drug screen.
tributing to the abundance of the qualifier
166
Journal of Analytical Toxicology, Vol. 34, April 2010
mg/L and was present in the urine at 27.13 mg/L. Because 6AM was detected in the urine, it is most likely that the morphine found in the blood and urine was from the metabolism
of 6-AM. In cases involving victims of heroin overdoses, their
morphine blood concentrations are indicative of recent drug
use (9). From prior chronic use of heroin, overdose victims’
urine, bile, and tissue morphine concentrations can represent
both past and recent drug exposure (9). There are two reasons
that could explain the presence of codeine in the urine. First,
the decedent could have also taken codeine in addition to the
other drugs. Another explanation is that 6-acetylcodeine or
codeine may have been present as an impurity in the heroin
(9). Illicit heroin is made from raw opium that primarily contains morphine and codeine. During the synthesis, morphine
is acetylated to produce heroin, and any codeine present is
also acetylated to form 6-actetylcodeine (10–12). The 6-acetylcodeine may have been metabolized to codeine, or the codeine
was actually present unchanged in the heroin and therefore detected in the urine.
In this case, there was a significant difference between
the drug concentrations detected in the decedent’s urine
versus his blood. The difference in the urine and blood drug
concentrations led to notably high urine/blood ratios observed. Based on the medical examiner’s report, the decedent
and his roommate administered mephedrone and heroin in
the “early morning” hours on May 21, 2009. An unknown
amount of mephedrone doses were administered by various
routes of administration (insufflation, intravenous, and oral).
This was followed by an intravenously administered dose of
an unknown amount of heroin. The decedent was pro-
nounced dead later that same day at 4:47 p.m. Based on this
information, there was a significant time lapse between the
multiple drug administrations and the decedent’s time of
death. Therefore, it is difficult to draw any significant conclusions about the urine/blood ratios because of the unknown amount of mephedrone and heroin doses in addition
to the multiple routes of administration.
The decedent’s roommate stated that he had self-administered the mephedrone by multiple routes of administration, including insufflation, oral, and intravenous, prior to administering heroin. It is difficult to predict the effects of mephedrone
and heroin in this case because it is hard to correlate drug concentrations in urine with effects as well as the fact that both
drugs were recreationally administered. The fact that one individual died while the other individual survived after self-administration of mephedrone and heroin illustrates the complexity of predicting adverse outcomes after abuse of multiple
drugs.
Because mephedrone has a structural similarity to amine
compounds, it was of interest to know if mephedrone would
cross-react with either amphetamine or ecstasy (MDMA) urine
drug screening kits. The following urine drug screening assay
kits were used: Roche Abuscreen Online® Amphetamine (Indianapolis, IN), Siemens Emit® II Plus Amphetamine (Newark,
DE), and Microgenics DRI® Ecstasy (Fremont, CA). Negative
urine specimens were spiked at the following mephedrone
concentrations: 100, 200, and 250 mg/L. The mephedrone
samples were analyzed on two different urine drug screening
instruments: a Hitachi 912 and a Hitachi Modular P (Tokyo,
Japan). Both the amphetamine and ecstasy kits were calibrated
at 0.50 mg/L with d-amphetamine and
MDMA, respectively. All three of the previously mentioned mephedrone concentrations screened negative when using
both amphetamine kits and the ecstasy
kit. Therefore, a case that is screened for
amphetamines and MDMA using the
three kits would not screen positive if it
contained mephedrone at a concentration of 250 mg/L or less in urine.
Conclusions
Figure 3. Gas chromatogram and full-scan mass spectrum of the pentafluoropropionic anhydride (PFPA)
derivative of mephedrone extracted from the decedent’s blood.
Typical non-fatal investigative cases in
our laboratory are tested for volatile analysis and abused drug testing but are not
tested with a full-scan GC–MS basic drug
screen or full-scan GC–MS designer drug
screen. Commonly used immunoassay
screening kits proved to have no cross-reactivity with mephedrone. In addition,
none of the three reference mass spectral
libraries used in our laboratory included
mephedrone. Therefore, if an investigative case was submitted in which the
individual was known to be using mephe-
167
Journal of Analytical Toxicology, Vol. 34, April 2010
drone without a specific request for mephedrone testing, a
negative test result may be reported. This case illustrates the
importance of investigators and medical examiners relaying
pertinent information about their cases to the forensic toxicology laboratory. Otherwise, new drugs not before encountered may be left unidentified, and medical examiners may report a death determination without complete information.
The medical examiner in this case reported that the cause of
death was attributed to multiple-drug toxicity associated with
mephedrone and heroin use. This is the first known reported
death in the U.S. attributed to the use of mephedrone as well
as the combined use of mephedrone and heroin. Despite
mephedrone being sold as a research chemical and plant feeder,
individuals are abusing this “legal high” drug because of its reported stimulant and euphoric effects. More research needs be
conducted on mephedrone, including duration of action, longterm effects, addiction potential, and toxicity.
Acknowledgment
This work was funded in part by the American Registry of
Pathology (Washington, D.C. 20306-6000).
2. Substance Abuse and Mental Health Services Administration. Results from the 2007 National Survey on Drug Use and Health: National Findings. Office of Applied Sciences, Department of Health
and Human Services, Series H-34, DHHS Publication No. SMA
08-4343. Rockville, MD, 2008.
3. S. Darke and D. Zador. Fatal heroin “overdose”: a review. Addiction 91: 1765–1772 (1996).
4. Goodman and Gilman’s The Pharmacological Basis of Therapeutics, 11th ed., L.L. Brunton, J.S. Lazo, and K.L. Parker, Eds. The
McGraw-Hill Companies, New York, NY, 2006.
5. Erowid. The Vaults of Erowid. 4-Methylmethcathinone section.
www.erowid.org, July 2009.
6. R.P. Archer. Fluoromethcathinone, a new substance of abuse.
Forensic Sci. Int. 185: 10–20 (2009).
7. S. Baker. ‘Legal highs’ crackdown is doomed to failure, say experts. The Independent. www.independent.co.uk, 25 May 2009.
8. U.S. Drug Enforcement Administration. 4-Methylmethcathinone
in Oregon. Microgram Bull. 42(7): 62 (2009).
9. R.C. Baselt. Disposition of Toxic Drugs and Chemicals in Man, 7th
ed. Biomedical Publications, Foster City, CA, 2004.
10. A.W. Jones, A. Holmgren, and F.C. Kugelberg. Driving under the
influence of opiates: concentration relationships between morphine, codeine, 6-acetylmorphine, and ethyl morphine in blood.
J. Anal. Toxicol. 32: 265–272 (2008).
11. S. Paterson and E. Cordero. Comparison of the various opiate alkaloid contaminants and their metabolites found in illicit heroin
with 6-monoacetylmorphine as indicators of heroin ingestion.
J. Anal. Toxicol. 30: 267–273 (2006).
12. R. Brenneisen, F. Hasler, and D. Würsch. Acetylcodeine as a urinary marker to differentiate the use of street heroin and pharmaceutical heroin. J. Anal. Toxicol. 26: 561–566 (2002).
References
1. Principles of Forensic Toxicology, 2nd ed., B. Levine, Ed. AACC
Press, Washington, D.C., 2006.
168
Manuscript received October 2, 2009;
revision received November 12, 2009.