Blood Pressure Measurement:
Published on Patient Care Online (http://www.patientcareonline.com)
Blood Pressure Measurement:
February 01, 2003 | Atrial Fibrillation [1], Sleep Disorders [2], Hypertension [3]
By Christopher B. Mcfadden, MD [4] and Raymond R. Townsend, MD [5]
ABSTRACT: The most common errors in measuring blood pressure (BP) are using the
incorrect cuff size, not having the patient relax for 5 minutes before the measurement,
and deflating the cuff too quickly. Observer bias may compound technical errors. When
patients use the proper procedure, home BP measurements may be more reproducible
than office measurements. Brachial artery-based monitors are more accurate than fingeror wrist-based instruments. To ensure that patients measure their BP correctly, observe
their technique with their own monitors. Counsel patients to measure their BP at
predetermined times and to have their monitors validated periodically.
Blood pressure (BP) measurement is an essential component of most adult office visits-particularly
because 20% to 40% of these visits are related to hypertension.1 Although the American Heart
Association (AHA) and other organizations offer guidelines for accurate BP measurement,2-4 there is
evidence that these recommendations often are not followed.5,6
BP readings may be significantly affected by seemingly trivial distractions, such as talking, or by
incorrect positioning. The errors introduced by poor technique may appear minimal in individual
cases, but the cumulative effects may be substantial. As small an increase as 5 mm Hg in diastolic
BP (DBP) readings would significantly increase the number of persons classified as hypertensive.7
Mistakes in technique may be compounded by other errors, such as terminal digit preference (eg,
home BP readings are consistently rounded off to zero) and observer bias.8 Because of the frequency
of error and the potential impact of misdiagnosis-in addition to the inherent variability of
BP-attention to the details of proper technique is crucial.9
In this article, we review current guidelines for BP measurement, with an emphasis on common
pitfalls and how they may be avoided. We also address the implications of the growing use of home
and ambulatory BP monitoring.
AN EARLY INNOVATOR
In 1905, Nikolai Korotkoff described the onset and disappearance of sounds heard during
auscultation of the brachial artery while occlusive pressure is withdrawn. These subsequently
became known as Korotkoff sounds10 and have since served as estimates of systolic (onset) and
diastolic (disappearance) pressure (Table 1). Although Korotkoff sounds underestimate systolic BP
(SBP) by 3 mm Hg and overestimate DBP by 6 to 8 mm Hg, they are considered the most reliable
indicators of these pressures in outpatient medicine.11,12 Their prognostic value over the last century
as markers of actual and impending cardiovascular disease has been thoroughly documented.
CORRECT TECHNIQUE
The most accurate BP measurements are obtained when the techniques and conditions used in
clinical studies are closely emulated. The following steps have been compiled from published
guidelines2,4,9 and from our experience in a hypertension clinic. Following these steps ensures that
BP measurement closely duplicates the procedure used in most of the studies that have
demonstrated the association between clinical (particularly cardiovascular) end points and BP
elevation.
Before the reading. Advise patients to avoid caffeine and nicotine for 30 minutes beforehand,
because these agents can raise BP. A full bladder also stimulates BP elevation because of increased
sympathetic activity.13
Ensure that the patient has rest-ed comfortably in a quiet room without distractions for
approximately 5 minutes. To minimize pressure elevation from isometric contractions, seat the
patient with legs and back supported.2 In one study, lack of back support raised DBP 6 mm Hg; SBP
was unchanged.14
The appropriate instrument. Use a cuff whose inflatable portion surrounds at least 80% of the
circumference of the arm. False elevations in pressure result from the use of cuffs that are too small,
because such cuffs require an excessive amount of internal pressure to occlude the brachial
Page 1 of 5
Blood Pressure Measurement:
Published on Patient Care Online (http://www.patientcareonline.com)
artery.2,15 The lower edge of the cuff should be 2 to 3 cm above the antecubital fossa.
Use a mercury sphygmomanometer if one is available. Although an aneroid manometer may be as
accurate as the mercury device and is an acceptable alternative, it may be less reliable because of
the need for frequent calibration.8,16,17 Studies have shown 40% of aneroid sphygmomanometers to
be out of calibration by 4 mm Hg or more (with 30% off by 10 mm Hg or more), compared with 8% of
mercury devices.6,17,18As the use of mercury-based sphygmomanometers decreases, it is essential to
ensure that aneroid devices undergo regular calibration and maintenance.19 One program recently
demonstrated that this can be done successfully.20
Proper arm position. Measure BP in the dominant arm, with the upper arm angled so that the
stethoscope is at approximately the level of the heart. This will usually require placing something
firm under the patient's elbow to support and elevate the arm. You can also rest the patient's elbow
in your hand, although the same hand is needed to hold the stethoscope. A phlebotomy chair or
appropriately elevated table may also be used. Measure BP in the nondominant arm at least once
and take the pressure in the arm with the higher pressure in subsequent measurements.
Palpation and auscultation. Palpate the brachial artery before auscultation and inflate the cuff by
increments of 10 mm Hg at a time to a value 30 mm Hg over the pressure at which the radial pulse
disappears. The point of disappearance and subsequent reappearance of the pulse during deflation
approximates systolic pressure. This maneuver helps evaluate the presence of an auscultatory
gap-the occasional disappearance of Korotkoff sounds after phase I. If you inflate the cuff during
auscultation to the point of absent Korotkoff sounds while listening over the artery without previous
palpation, you may mistake phase II for phase I and significantly underestimate systolic pressure.
Place the bell of the stethoscope over the brachial artery superior and medial to the antecubital
fossa, where the brachial artery pulse is most intense to palpation. Some question exists as to
whether the bell should be used instead of the diaphragm,14 although the bell is thought to
reproduce more accurately phase IV and V Korotkoff sounds.2
Auscultate the brachial artery, with the arm in a supported position and the cuff at heart level. Inflate
the cuff to 30 mm Hg above the level of SBP previously estimated by palpation. Deflate the cuff at a
rate of 2 mm Hg/s. More rapid deflation decreases your ability to note changes in Korotkoff sounds,
and significantly slower deflation may increase venous congestion in the arm and raise the diastolic
value.
Noting the measurements. Record, to the nearest 2 mm Hg, the level at which the first of 2
consecutive sounds is heard. (Accuracy greater than to within 2 mm Hg cannot be obtained by either
a mercury sphygmomanometer or an aneroid device.) Odd-numbered readings on the patient's chart
point to use of automatic recording devices or to significant observer error.
Note the level at which sounds become muffled (phase IV) and disappear (phase V). If sounds
continue toward zero, note this as well.
AVOIDING PITFALLS
A 1990 study found the most common errors to be:
Use of the wrong-sized cuff.
Failure to have the patient relax for 5 minutes.
Too rapid deflation of the cuff.6
These findings have been confirmed in more recent studies.21 Tips on ensuring the accuracy of BP
measurement are listed in Table 2.
A recent review summarized the effects of various activities on BP.22 The authors note that:
Talking elevates SBP 17 mm Hg and DBP 13 mm Hg.
Acute exposure to cold elevates SBP 11 mm Hg and DBP 8 mm Hg.
Arm placement 10 cm below the level of the heart increases SBP and DBP 8 mm Hg; arm placement
10 cm above the level of the heart decreases SBP and DBP 8 mm Hg.
Lack of arm support increases SBP and DBP about 2 mm Hg.
An inappropriately small cuff low- ers SBP 8 mm Hg and raises DBP 8 mm Hg.
Another study found that the white coat effect (the nonspecific elevation in BP that many persons
experience when they enter a medical setting) can raise SBP by 11 to 28 mm Hg and DBP by 3 to 15
mm Hg.9
HOME AND AMBULATORY BP MEASUREMENT
Home and ambulatory BP monitors are increasingly used in the management of hypertension. These
instruments are particularly helpful in determining whether a patient has white coat hypertension
and in assessing the response to therapy. When taken correctly, home BP measurements may be
more reproducible than office measurements.23
Home and ambulatory devices record BP by auscultatory discrimination or oscillometric assessment.
Page 2 of 5
Blood Pressure Measurement:
Published on Patient Care Online (http://www.patientcareonline.com)
The latter technique recognizes mean arterial pressure based on the maximum recorded oscillation.
Systolic and diastolic blood pressure are subsequently determined by algorithm.10 Oscillometric
devices may not measure pressures accurately in patients with arrhythmias, particularly atrial
fibrillation.24
Home BP monitors. The benefit of these instruments is obviated if they are not reliably accurate
and if patients do not use them correctly. A number of home and ambulatory devices were tested
recently by researchers who used American and British assessment guidelines.25 The findings
support the well-recognized limitations of wrist- and finger-based readings. Most experts recommend
brachial artery-based monitors.2,26 Advise patients to purchase devices that meet international
standards, such as those listed on the Web site of the British Hypertension Society: www.
hyp.ac.uk/bhsinfo/bpmdigit.html.
The use of incorrect technique to measure BP at home is a widely underappreciated problem.27 The
same careful procedures need to be followed in home measurement as in clinic measurement. These
include limiting activity for a set period before the reading, keeping the arm at heart level, and not
talking during the procedure. We ensure that patients are using the correct technique by having
them perform the measurement with their own device under our supervision. Counsel patients to
measure their BP at predetermined times, not just when they are feeling well or unwell. Remind
them that the accuracy of all devices should be independently validated periodically. At our clinic,
we have patients bring their monitors in annually for calibration against mercury devices, unless an
earlier calibration is warranted.
Ambulatory monitors. These devices are particularly useful in diagnosing suspected white coat
hypertension in patients who have no signs of target organ damage. Approximately 10% to 40% of
normotensive and nominally borderline hypertensive patients experience an appreciable white coat
effect, and many may be erroneously classified as hypertensive.9 However, there is concern about
ongoing target organ damage, particularly left ventricular hypertrophy, in untreated patients with
white coat hypertension.28,29
Ambulatory monitoring may be useful in other areas of hypertension. Differentiating "dippers" (those
whose BP falls during sleep) from "nondippers" may allow earlier identification of persons at
increased risk for target organ damage. A recent study showed a higher prevalence of left
ventricular hypertrophy and carotid intima-media thickening among nondippers in whom
hypertension had recently been diagnosed.30 Future research may determine whether
pharmacologic therapy based on dipping status is appropriate.
Home BP readings average 5/5 mm Hg lower than office readings. Although there is no definitive
consensus, most experts agree that the upper limits of normal for ambulatory daytime
measurements should range from 129 to 135 mm Hg for SPB and 79 to 85 mm Hg for DBP.26,31-33
Normal nocturnal pressures should be lower. This topic has been extensively reviewed elsewhere.22
New prognostic data show that ambulatory BP control predicts cardiovascular risk better than office
BP control in patients being treated for hypertension,34,35 although large-scale interventional trials
are lacking.
References:
REFERENCES:
1. Pardell H, Tresserras R, Armario P, Hernandez del Rey R. Pharmacoeconomic considerations in the
management of hypertension. Drugs. 2000;59:13-20.
2. Perloff D, Grim C, Flack J, et al. Human blood pressure determination by sphygmomanometry.
Circulation. 1993;88:2460-2470.
3. Pickering TG, James GD, Boddie C, et al. How common is white coat hypertension? JAMA. 1988;
259:225-228.
4. Feldman RD, Campbell N, Larochelle P, et al. 1999 Canadian recommendations for the
management of hypertension. CMAJ. 1999;161(suppl 12): S1-S17.
5. Grim CM, Grim CE. A curriculum for the training and certification of blood pressure measurement
for health care providers. Can J Cardiol. 1995;11 (suppl H):38H-42H.
6. McKay DW, Campbell NR, Parab LS, et al. Clinical assessment of blood pressure. J Hum Hypertens.
1990;4:639-645.
7. Campbell NR, McKay DW. Accurate blood pressure measurement: why does it matter? CMAJ.
1999; 161:277-278.
8. Beevers G, Lip GYH, O'Brien E. ABC of hypertension: blood pressure measurement. Part
II-conventional sphygmomanometry: technique of auscultatory blood pressure measurement. BMJ.
Page 3 of 5
Blood Pressure Measurement:
Published on Patient Care Online (http://www.patientcareonline.com)
2001;322: 1043-1047.
9. Reeves RA. Does this patient have hypertension? How to measure blood pressure. JAMA.
1995;273: 1211-1218.
10. Nichols WW, Michael F. McDonald's Blood Flow in Arteries. New York: Oxford University Press;
1998.
11. Stolt M, Sjonell G, Astrom H, Hansson L. Factors affecting the validity of the standard blood
pressure cuff. Clin Physiol. 1993;13:611-620.
12. Breit SN, O'Rourke MF. Comparison of direct and indirect arterial pressure measurements in
hospitalized patients. Aust N Z J Med. 1974;4:485-491.
13. Fagius J, Karhuvaara S. Sympathetic activity and blood pressure increases with bladder
distension in humans. Hypertension. 1989;14:511-517.
14. Cushman WC, Cooper KM, Horne RA, Meydrech EF. Effect of back support and stethoscope head
on seated blood pressure determinations. Am J Hypertens. 1990;3:240-241.
15. O'Brien E. A century of confusion: which bladder for accurate blood pressure measurement? J
Hum Hypertension. 1996;10:565-572.
16. Sloan PJ, Zezulka A, Davies P, et al. Standardized methods for comparison of
sphygmomanometers. J Hypertens. 1984;2:547-551.
17. Bailey RH, Knaus VL, Bauer JH. Aneroid sphygmomanometers. An assessment of accuracy at a
university hospital and clinics. Arch Intern Med. 1991;151:1409-1412.
18. Burke MJ, Towers HM, O'Malley K, et al. Sphygmomanometers in hospital and family practice:
problems and recommendations. Br Med J. 1982;285:469-471.
19. Sheps SG, Jones DW. Is it time to eliminate the mercury sphygmomanometer? Presented at: 17th
Annual Scientific Meeting and Exposition of the American Society of Hypertension; May 14-18, 2002;
New York.
20. Canzanello VJ, Jensen PL, Schwartz GL. Are aneroid sphygmomanometers accurate in hospital
and clinic settings? Arch Intern Med. 2001;161: 729-731.
21. Villegas I, Arias IC, Botero A, Escobar A. Evaluation of the technique used by health-care workers
for taking blood pressure. Hypertension. 1995;26: 1204-1206.
22. McAlister FA, Straus SE. Evidence-based treatment of hypertension. Measurement of blood
pressure: an evidence-based review. BMJ. 2001;322: 908-911.
23. Jyothinagaram SG, Rae L, Campbell A, Padfield PL. Stability of home blood pressure over time. J
Hum Hypertens. 1990;4:269-271.
24. O'Brien E, Beevers G, Lip GY. ABC of hypertension: blood pressure measurement. Part IVautomated sphygmomanometry: self blood pressure measurement. BMJ. 2001;322:1167-1170.
25. O'Brien E, Waeber B, Parati G, et al. Blood pressure measuring devices: recommendations of the
European Society of Hypertension. BMJ. 2001; 322:531-536.
26. Asmar R, Zanchetti A. Guidelines for the use of self-blood pressure monitoring: a summary
report of the First International Consensus Conference. Group Evaluation and Measure of the French
Society of Hypertension. J Hypertens. 2000;18:493-508.
27. Townsend RR. Common questions and answers in the management of hypertension: home blood
pressures kits. J Clin Hypertens. 2000;2:288-289.
28. Owens PE, Lyons SP, Rodriguez SA, O'Brien ET. Is elevation of clinic blood pressure in patients
with white coat hypertension who have normal ambulatory blood pressure associated with target
organ changes? J Hum Hypertens. 1998;12:743-748.
29. Moser M. White-coat hypertension-to treat or not to treat: a clinical dilemma. Arch Intern Med.
2001;161:2655-2656.
30. Cuspidi C, Macca G, Sampieri L, et al. Target organ damage and non-dipping pattern defined by
two sessions of ambulatory blood pressure monitoring in recently diagnosed essential hypertensive
patients. J Hypertens. 2001;19:1539-1545.
31. Schettini C, Bianchi M, Nieto F, et al. Ambulatory blood pressure: normality and comparison with
other measurements. Hypertension. 1999;34:818-825.
32. O'Brien E, Staessen JA. What is "hypertension"? Lancet. 1999;353:1541-1543.
33. Graves JW, Sheps SG. Out of office self measurement of blood pressure. J Clin Hypertens.
1999;1: 120-124.
34. Verdecchia P, Reboldi G, Porcellati C, et al. Risk of cardiovascular disease in relation to achieved
office and ambulatory blood pressure control in treated hypertensive subjects. J Am Coll Cardiol.
2002;39: 878-885.
35. Ohkubo T, Imai Y, Tsuji I, et al. Prediction of mortality by ambulatory blood pressure monitoring
versus screening blood pressure measurements: a pilot study in Ohasama. J Hypertens.
Page 4 of 5
Blood Pressure Measurement:
Published on Patient Care Online (http://www.patientcareonline.com)
1997;15:357-364.
Source URL: http://www.patientcareonline.com/articles/blood-pressure-measurement