Download VII. Introduction to Charge Mode Accelerometers

Survey
yes no Was this document useful for you?
   Thank you for your participation!

* Your assessment is very important for improving the workof artificial intelligence, which forms the content of this project

Document related concepts

Scattering parameters wikipedia , lookup

Buck converter wikipedia , lookup

Dynamic range compression wikipedia , lookup

Sound reinforcement system wikipedia , lookup

Switched-mode power supply wikipedia , lookup

Resistive opto-isolator wikipedia , lookup

Metadyne wikipedia , lookup

Regenerative circuit wikipedia , lookup

Rectiverter wikipedia , lookup

Electrostatic loudspeaker wikipedia , lookup

Instrument amplifier wikipedia , lookup

Audio power wikipedia , lookup

Wien bridge oscillator wikipedia , lookup

Public address system wikipedia , lookup

Negative feedback wikipedia , lookup

Tube sound wikipedia , lookup

Amplifier wikipedia , lookup

Opto-isolator wikipedia , lookup

Transcript
Introduction to Charge Mode Accelerometers
Dytran charge mode accelerometers are designed to measure shock and vibration phenomena over a broad temperature range. These accelerometers, unlike
the Low Impedance Voltage Mode (LIVM) types, contain no built-in amplifiers.
Dytran’s charge mode accelerometers utilize high sensitivity piezoceramic crystals, of the lead zirconate titinate (PZT) family, to produce a relatively high
charge output in response to stress created by input vibration or shock acting
upon the seismic system.
standardization, ranging, filtering, integrating for velocity and displacement,
etc. and,
3. when adding or replacing accelerometers where existing charge amplifiers
must be used for economic or other reasons.
Two System Concepts
Because of the high impedance level of the charge mode signal generated by
the crystals, a special type of amplifier, called a charge amplifier, is used to
extract the very high impedance electrostatic charge signal from the crystals.
The charge amplifier has the ability to convert the charge signal to a low
impedance voltage mode signal without modifying it.
Charge mode accelerometers may be combined with a variety of electronic
components to create two basic measurement system classifications:
1. The conventional charge mode system, and
2. The Hybrid system.
PROTECTIVE CAP
SEISMIC MASS
PRELOAD SCREW
PIEZO CRYSTALS
ELECTRODE
STRAIN ISOLATION
BASE
10-32 ELECTRICAL
CONNECTOR
10-32 TAPPED HOLE
The conventional charge mode system utilizes a sophisticated laboratory
charge amplifier while the hybrid system features simple dedicated range
miniature in-line charge and voltage amplifiers operating in conjunction with
LIVM current source power units.
The Conventional Charge Mode System
The versatile laboratory charge amplifier is the main feature of the conventional charge mode system. This section will familiarize you with the theory, operating characteristics and features of the basic laboratory charge amplifier.
Figure 2: Elements of the conventional charge mode system
MOUNTING STUD
MODEL 6200
ELECTROSTATIC
CHARGE AMPLIFIER
CHARGE AMPLIFIER
4165
DC VOLTS
-10
0
DC
+10
LOW NOISE CABLE
Figure 1: Typical compression design, charge mode accelerometer
0
O/L
1
2
3
4
5
AC
AC VOLTS-RMS
METER
10
0
90
20
10
80
30
70
60
50
40
Figure 1 is a cross-section of a typical charge mode compression design
accelerometer, model 3100C6. The sensitivity is 100 pC/g (pC = pico coulomb
= 1 x 10-12 Coulomb) and the useful frequency range is up to 5 kHz. The
3100C6 operates at temperatures up to +500˚F.
SENSOR
SENSITIVITY
pC or mV/UNIT
.1 to 1.1
1.0 to 11.0
RESET
ZERO
100
500
IN
1K
200
2K
OUT
10
FILTER
100
5K
50
CHARGE MODE
ACCELEROMETER
10K 1K
20
20K
10
1
PWR
S
L
M
50K
CHG
RANGE
UNITS/VOLT
LIVM
TC
INSTRUMENTS, INC.
A heavy metal seismic mass is preloaded against the piezoceramic crystals with
an elastic preload screw. The mass converts the input acceleration into analogous stress on the crystals producing an output charge signal in direct proportion to instantaneous acceleration.
USA
Figure 2 illustrates a laboratory charge amplifier, model 4165 in use with a
model 3100C6 charge mode accelerometer. Series 6019A low-noise coaxial
cable is used to minimize triboelectric noise generated by cable motion. This is
a very versatile system whose signal conditioning options include:
When to Use Charge Mode Accelerometers
The question may be asked, “When should I consider using charge mode
accelerometers vs. LIVM types with built-in electronics?”
Charge mode accelerometers should be considered:
1. when making measurements at temperatures above +250˚F, the maximum
temperature for most LIVM instruments,
2. when the versatility of the laboratory charge amplifier is desired for system
1. Standardization of system sensitivity
2. Full Scale range selection
3. Discharge time constant choices
4. Filter options
5. 0-10 VDC out for sinusoidal input
6. Overload indication
7. Instant system zeroing (or reset)
8. External calibrate signal insertion
9. Front panel meter for observation of DC level of output signal
21592 Marilla Street, Chatsworth, California 91311 • Phone: 818.700.7818 • Fax: 818.700.7880
www.dytran.com • For permission to reprint this content, please contact [email protected]
Charge Amplifier, Basic Theory
Adding Versatility to The Charge Amplifier
A charge amplifier is a special high gain, high input impedance inverting voltage amplifier with capacitive feedback. The amplifier is usually an operational
amplifier (op-amp) with near infinite voltage gain.
Standardization of system sensitivity, say to exactly 10.00 or 100.00 mV/g is
accomplished by adjusting the amount of feedback with a potentiometer as
shown. Standardization is often necessary because accelerometers are rarely
ever made to exact sensitivities. The use of a precision multi-turn potentiometer
with turns counting dial allows the standardization of the system sensitivity by
dialing in the accelerometer sensitivity.
S1
S1
Cf
Rf
q
Rf
-
in
-
q in
-A
CA
Referring to Figure 3b, changing the range of the charge amplifier is accomplished by switching various values of feedback capacitor into the feedback
path. This is accomplished with a rotary switch which has maybe 10 values of
precision capacitor arrayed around it.
Cf
-v out
-A
CA
+
-v out
10
+
Standardization
Potentiometer
0
(a)
(b)
(a)
(a)
Again referring to Figure 3b, feedback resistor Rf gives DC stability to the circuit
and establishes the discharge time constant (TC) of the amplifier thereby setting the low frequency response of the amplifier. A momentary reset switch (S1)
discharges the residual charge in the feedback capacitor returning the system
output to zero.
Figure 3: The charge amplifier
Referring to Figure 3, the input charge qin is applied to the summing junction
(inverting input) of the charge amplifier and is distributed to the input capacitance of the amplifier CA and the feedback capacitor Cf. We may write the equation:
qin = qA + qf
The Hybrid System
A Dytran hybrid system combines charge mode accelerometers with miniature
in-line fixed sensitivity charge amplifiers. These charge amplifiers are powered
by standard 2-wire LIVM power units.
Eq 1
Using the electrostatic equation q = Cv and substituting in equation 1:
qin = vA CA + vf Cf
The circuit shown in Figure 3b represents only the first stage of a rudimentary
charge amplifier. One or more stages of filtering, integration and other features
can be added all in one compact package.
Eq 2
LIVM POWER UNIT
CHARGE MODE ACCELEROMETER
Using equation 2 and making the appropriate substitutions and solving for the
output voltage of the amplifier in terms of input charge, amplifier loop gain,
and input and feedback capacitance we have:
4114B
CURRENT SOURCE
SHORT
NORMAL
OPEN
LOW NOISE COAXIAL CABLE
0
12
24
SENSOR BIAS VDC
SERIES 4705A
IN-LINE CHARGE
AMPLIFIER
2
3
1
-qin
Vout =
-qin
1
=
x
CA / A + Cf (A + 1) Cf (1 + 1 / A) 1 + CA / Cf (A + 1)
Eq 3
where A is the open loop gain of the op-amp.
Now, letting gain A approach infinity, we have:
-qin
Vout =
Eq 4
Cf
This result (Eq 4) shows clearly that the transfer function (gain) of a charge
amplifier is a function only of the value of the feedback capacitor Cf. Notice
that input capacitance CA has no effect on the sensitivity of the charge amplifier. This means that cable capacitance, for example, has no effect on the sensitivity, a significant find when switching cable lengths and types.
USA
INSTRUMENTS, INC.
INPUT
INSTRUMENTS, INC.
USA
SIG/PWR
MOD 4705AXX
4
CHANNEL
MONITOR
S/N XXXX
ON
PWR
GENERAL PURPOSE COAXIAL
CABLE
Figure 4: A hybrid system with in-line charge amplifier
The hybrid system, (refer to Figure 4) is ideal for field use because of its small
size and rugged construction of the miniature charge amplifiers, for example,
Models 4751B and 4505A. These amplifiers are powered by conventional LIVM
constant current power units and transmit the output signal over the same two
wires as do conventional LIVM systems. The power unit separates the signal
information from the DC bias of the amplifier and couples it to the readout
instrument. As with conventional charge mode systems, low noise coaxial cable
is used to couple the accelerometer to the charge amplifier to minimize triboelectric noise.
21592 Marilla Street, Chatsworth, California 91311 • Phone: 818.700.7818 • Fax: 818.700.7880
www.dytran.com • For permission to reprint this content, please contact [email protected]
When to Use The Hybrid System
The hybrid system should be considered when:
1. The accelerometer is used to measure events at a temperature which is above
that recommended for LIVM instruments, i.e., instruments that have built-in
amplifiers and the environment is not favorable to laboratory charge amplifiers.
2. System cost is an important factor. Per channel cost of the hybrid system is a
fraction of that for the conventional charge amplifier system.
3. LIVM current source power units are in-hand and must be utilized.
4. Ruggedness and small size of the measurement system is imperative.
5. A dedicated system without the versatility of a laboratory charge is sufficient
for the measurement task.
21592 Marilla Street, Chatsworth, California 91311 • Phone: 818.700.7818 • Fax: 818.700.7880
www.dytran.com • For permission to reprint this content, please contact [email protected]