Comparison of Simultaneously Recorded Central and Peripheral Arterial Pressure Pulses During Rest, Exercise and Tilted Position in Man By EDWIN J. KKOEKER, M.D. AND EARL H. WOOD, M.D., PH.D. Downloaded from http://circres.ahajournals.org/ by guest on June 16, 2017 Central (aortic or subclavian), brachial, radial and femoral pressure pulses were recorded simultaneously in 12 healthy subjects during conditions of rest, exercise and 70 degree head-up tilt. Peripheral systolic pressure at rest uniformly exceeded the central systolic pressure generated by the same heartbeat. The average radial pulse pressure was 146, 146 and 165 per cent of central pulse pressure during rest, exercise and tilt while radial mean pressures were 94, 93 and 9S per cent of central mean pressures respectively. Summation of the incident pulse wave with reflected waves from the periphery and i-esonance effects in the peripheral arterial systems may produce these changes in pressure and contour. ton and co-workers4 in 1936. Aortic and radial arterial pressure pulses were recorded simultaneously by Fuller and associates6 in 1952 during an operation for coarctation of the aorta. Using the technic of arterial catheterization, Schnabel and co-workers8 in 1952 reported measurements of the central pressure pulse in the intact human. The present investigation was initiated for the purpose of establishing normal pressurepulse data in man by the simultaneous recording of central and multiple peripheral arterial pressure pulses during conditions of rest, exercise and 70-degree head-up tilt. ALTHOUGH measurements of blood / % pressure in man by direct needle puncJL A . ture are now commonplace, actually, determinations of arterial pressure by this technic, are of relatively recent origin. The average and range of values encountered in man, especially in regard to central arterial pressure pulses, have not as yet been well documented. Direct arterial blood pressure was first recorded in man during limb amputations. Thus, Faivre 1 in 1856 repoi^ted the blood pressure in the femoral artery of man as being 120 mm. of mercury. Merke and Mviller,2 in 1925, surgically exposed a brachial artery in each of two critically ill patients and inserted T tubes for the recording of pressures. Brachial arterial pressures were obtained by needle puncture in 1931 by Wolf and von BonsdorfP in a series of normal subjects and of patients with cardiovascular disease. Hypodermic-type manometers were used to record simultaneous pressure pulses in the axillary, femoral and dorsalis pedis arteries by Hamil- METHODS The studies included 14 arterial catheterizations of 12 healthy male physicians whose average age was 32 years (range: 27 to 41), average height was 70 inches (range: 66 to 74) and average weight was 175 pounds (range: 150 to 205) (table 1). Subjects were supine during rest and exercise; however, during exercise the left foot was in the pedal of a bicycle ergometer, while the right leg, with an indwelling needle in the femoral artery in the right groin, remained at rest. The subject maintained exercise at a rate of approximately 45 r.p.m. with his left leg by watching a tachometer leading from the bicycle ergometer.7 Arterial pressure pulses at multiple sites were recorded at rest, after 3 to 6 minutes of exercise and after at least 10 minutes in the 70-degree head-up tilt position. Respiration and the electrocardiogram were registered simultaneously. Cardiac output was measured by the From the Section of Physiology, Mayo Clinic and Mnyo Foundation, Rochester, Minn. Abridgment of a portion of the thesis submitted by Dr. Kroeker to the Faculty of the Graduate School, University of Minnesota, in partial fulfillment of the requirements of a degreo of Master of Science in Medicine. Received for publication June 28, 1955. 623 CirciJallon RftrarcA, Volume III, jVoremfter 1MI 024 CENTRAL AND PERIPHERAL PRESSURE PULSES IN MAN Downloaded from http://circres.ahajournals.org/ by guest on June 16, 2017 direct Fick technic or by the dye-dilution technic, under all three conditions just before or after the recordings of aiterial pressures. Central arterial pressure pulses were obtained via a Petei-son-type aiterial catheter advanced up to the arch of the aorta or to the left subclavian artery through an IST-gage needle inserted into the femoral artery at the groin or into the left braehial artery in the antecubital fossa respectively. The catheter was 80 cm. in length, 1.0 mm. in external diameter and 0.5 mm. in internal diameter. The final position of the catheter tip as measured on the roentgenogram with correction for x-ray tube distortion was an average of 5.5 cm. (range: 1.5 to 9.3) from a point at the top of the aortic arch at the midsternal line. The catheter was inserted an average distance of 51.5 cm. (4S.5 to 54.5) in the five femoral aiterial catheterizations and 42.5 cm. (40 to 45) in the nine brachial aiterial catheterizations. The average distance from the third interspace at the sternum to the femoral-needle puncture site was 50 cm. (45 to 53), and from the brachialneedle puncture site 42.5 cm. (40 to 46) respectively in these subjects. The catheter-manometer system permitted immediately interchangeable recording with a capacitance* or strain-gage manometer. Both straingauge and capacitance manometer-catheter systems were optimally damped. The response of the straingauge manometer-catheter system to sine-wave pressure variations of increasing frequencies was within 10 per cent of static sensitivity out to 12 c.p.s., while the capacitance manometer-catheter system had a uniform response out to 60 cycles per second. It has been found that catheter-manometer systems with these characteristics are adequate to reproduce the practically important components of peripheral8 and central* pressure pulses in man. Peripheral arterial pressure pulses were recorded by means of a hypodermic strain-gauge manometer system described previously1"1" with a uniform response to sine-wave pressure variations up to 25 cycles per second. Dynamic response characteristics of all manometer systems to sine-wave and squarewave pressure variations were determined immediately following each procedure.10 Photokymographic recording was at paper speeds ninging from 75 to 150 mm. per second. The manometers were calibrated frequently during the course of a procedure, known pressures being projected into all the manometer systems simultaneously. The zero reference point for both the supine and the tilt position was taken as the midpoint of the anteroposterior diameter of the chest at the level of the third interspace at the sternum. Mean pres• A commercial version of the Lilly manometer manufactured by the Tochnitrol Engineering Company, Philadelphia, Pn. sures were obtained by planimetry of the photographic record. Only simultaneously recorded pulses were measured. Simultaneous recordings of central and radial pressure pulses were obtained in all 12 subjects. In addition simultaneous brachial pressure pulses were obtained in eight and simultaneous femoral pressure pulses in nine of these 12 subjects. RESULTS The values for the 12 subjects as regards age, height, surface area, metabolic rate, heart rate, cardiac indexes, and central and radial arterial pressures are shown in table I. Table 2 compares central arterial pressures recorded in different conditions in healthy men. The average increase in oxygen consumption produced by the exercise was 2.50 (200 to 320) per cent as compared to an increase of 69 (46 to 97) per cent in the cardiac output (table 1). Table 3 shows the relationship of peripheral arterial pressures to the pressures generated by the same heartbeat in the central arteries in healthy men. AVhen the men were resting, in the supine position, there was a uniform increase of systolic pressure toward the periphery. The diastolic, dicrotic and mean pressures showed a smaller but uniform decrease to the periphery. During exercise the pressure changes from the central to the peripheral arteries were similar to the changes at rest except for a greater decrement in dicrotic pressure to the brachial and radial arteries and an absence of the increase in systolic pressure at the femoral artery. During the 70-degree head-up tilt the average percentage increase of systolic pressure was larger and the decrement in mean pressure toward the periphery was smaller than in the resting state. Table 4 is a comparison of pulse pressures recorded from the aorta and from the femoral, brachial and radial arteries in healthy men. At rest, central pulse pressure was 45 mm. of mercury, but it increased to 57 mm. during exercise and decreased to 33 mm. at the 70 degree head-up tilt. At rest, brachial pulse pressure was 131 per cent, femoral ]39 per cent and radial 146 per cent of aortic pulse pressure. During exercise there was a striking 625 E. J. KROEKER AND E. H. WOOD TARI.K 1.— Vikd Statistics awl Simultaneous Central and Radial Arterial Pressures in IS Healthy White Men at Resl in Supine Position* Age, Years Cardiac Index, Heart Metabolic Rate at L./min./M Height, Surface Area, Rate, Rest, Inches Win. Rest Exercise Tilt 40 69 69 66 71 71 68 2.12 1.95 1.95 1.90 2.02 2.10 2.11 2.15 1.86 1.93 1.7S 2.06 2.07 1.81 Avcrune 32 70 1.9S 40 31 2S 28 2<J 30 Downloaded from http://circres.ahajournals.org/ by guest on June 16, 2017 29 28 27 31 41 70 70 68^ 71 72 74 +2 -7 -3 -2 0 +2 +5 +1 +32 -3 -2 -8 + 10 +2 +3 Standard deviation Arterial Pressure (mm. Hg) During Supine Rest Systolic 66 4.1 3.1 3.7 3.3 5.6 3.0 6.5 5.7 — 4.S 8.6 5.9 2.7 — 2.6 — 2.9 — 114 US 114 115 12S 121 133 113 137 120 131 119 136 146 64 3.8 6.3 2.S S O.S 1.1 0.2 60 58 59 64 63 72 3.2 60 5S 54 66 62 63 88 5.5 2.4 A Q 7 ^ q i — — 3.3 3.S 5.3 7.3 2.7 q o a c q 0 — Right radial artery Central artery (A) (A) (A) (A) (S) (S) (A) (S) (S) (S) (S) (S) (S) (S) Diastolic 71 73 73 76 7S 7S S4 Diastolic Mean 116 139 136 12S 15S 144 65 72 6S 72 74 75 77 64 75 75 S2 76 78 S3 S3 91 88 91 97 96 100 SI 100 96 106 94 104 107 Mean Systolic 92 93 91 94 101 99 SS 00 113 113 149 US 146 144 142 12S 147 153 126 SI 102 140 75 96 10 6 7 12 5 7 72 80 S4 S5 86 104 92 105 104 110 103 * Arranged in order of increasing central arterial diastolic pressure. S •= Pressure recorded from left subclavian artery by catheter inserted via bracliial artery. A = Pressure recorded from aorta by catheter inserted via femoral artery. TABLE 2.—Central* Arterial Pressures Recorded During Different Coiulitions in Healthy Men Condition Rest Kxercisc Tilt Subjects 12 12 S Constant Average Range Average Range Average Range Heart Rate, Beats per Minute Pressure, ram. Hg Systolic Diastolic Dicrotic Mean 64 126 SI 10S 102 54-SS 99 71-125 113-140 71-90 96-124 91-113 144 S7 109 112 116-187 69-101 87-125 92-133 9S S9-105 S9 116 S3 102 69-113 104-122 75-S9 90-110 * Aorta or subclavian artery near aorta. decrease in the degree of amplification of the pulse pressure in the femoral artery, femoral pulse pressure being only 114 per cent of aortic pulse pressure, while brachial and radial pulse pressures relative to aortic pulse pressure showed little change from the resting condition. During the 70-degree head-up tilt the absolute aortic pulse pressure was reduced but the amplification of peripheral pulse pressure in relation to aortic pulse pressure was increased at all the sites studied to levels of 14G per cent in the brachial, 159 per cent in the femoral and 165 per cent in the radial artery. These differences in pressure levels between peripheral and central pressure pulses are associated with the characteristic differences in contour of these pulses. Figure 1 shows a series of simultaneously recorded pulses in a 31 year old healthy subject. There is a gradual increase in the time interval between the peak of the R wave in the electrocardiogram and the onset of the pulse wave and a gradual increase in pulse pressure, especially the systolic peaks, toward the periphery. The anaerotic shoulder is present in the aortic pulse but barely visible in the femoral pulse. There 626 CENTRAL AND PERIPHERAL PRESSURE PULSES IX MAN TABI^E 3.—Peripheral Arterial Pressures* as Percentages of Central Arterial Pressure Recorded Simultaneously in Healthy Men % of Central Pressure Pressure Pulse Constant Systolic Diastolic Dicrotic Mean 98 92-101 94 SS-9S 96 93-99 At rest Brachial Radial Femoral 8 12 9 Averuge Range Average Range Average Range 109 96 92 101-llS 90-100 89-96 SS 83-95 94 SS-99 112 93 102-123 SS-99 110 94 104-119 90-97 During exercise Downloaded from http://circres.ahajournals.org/ by guest on June 16, 2017 Brachial Radial Femoral S 12 S Average Range Average Range Average Range 111 97 89 97 101-122 90-101 93 86-99 88-93 83 73-SS 92-101 113 9S-12S 93 S9-100 101 95 97 97 95-107 S7-100 S3-104 92-101 9S 96-102 95 92-9S 98 96-101 93 8S-9S 89 84-95 95 93-97 99 96-103 9S 92-101 100 9S-102 During 70 degree lead-up tilt Brachial Radial Femoral S S 3 Average Range Average Range Average Range 111 105-1IS 115 103-121 123 112-139 * The average and range of distances from the needle puncture site to the midsternum at the third interspace were 50 (45-53), 42 (40—15) and 68 (60-72) cm. for the femoral, brachial and radial arterial puncture sites respectively. TABLE 4.—Relationship of Pulse Pressures Recorded Simultaneously From Different Arteries in Healthy Men Condition Rest Exercise Tilt Constant Subjects Average Range Subjects Average Range Subjects Average Range Aortic Pulse Pressure, mm. Hg 12 45 33-57 12 57 46-86 12 33 25-40 is a secondary wave following the primary peak but preceding the dicrotic notch in the brachial-radial system and there is an absence of this wave in the femoral-dorsalis pedis system. The incisura, sharp and short in the (Peripheral Pulse Pressure/Aortic) X 100 Femoral Brachial Radial 9 139 118-170 8 131 104-156 S 127 103-147 S 146 119-16S 12 146 119-192 12 146 106-191 S 165 122-1SS s 114 95-136 3 159 142-177 aortic pulse, is lost during transmission of the pulse wave peripherally. The dicrotic notch, which is drawn out and deep in the brachial-radial system, is practically nonexistent in the femoral and so drawn out and E. J. KROEKER AND E. H. WOOD 027 RIGHT : RADIAL d LEFT F LEFT BRACHIAL ~- AORTIC LJ £ Downloaded from http://circres.ahajournals.org/ by guest on June 16, 2017 RIGHT FEMORAL - RIGHT DORSALIS L_ PEDIS ECG FIG. 1. Comparison of central and various peripheral arterial pulses (as labeled) from a normal male. Discussion in toxt. deep in the dorsalis pedis pulse that it approaches end diastolic pressures. Figure 2 shows the transformation of the subclavian type of pulse to the brachial-radial type at graded distances from the subclavian artery near the aorta. The records have been mounted so that the R waves of the electrocardiogram are superimposed. On the left are pulses recorded by the strain-gauge manometer system and on the right those recorded by the higher-frequency capacitance manometer. Note the similarity of the pulses recorded by these two widely different manometer systems. There is an increase in the interval from the R wave of the electrocardiogram to the onset of the pulse toward the periphery; the systolic peaks, however, tend to be stationary for the first 20 cm. There is a gradual increase in amplitude, with higher systolic peaks. The anacrotic shoulder is present and similar in both systems but disappears after the 10 cm. withdrawal. In this subject preoscillations were recorded in the central pulse only by the faster capacitance manometer system and were absent after the 10 cm. withdrawal. In the subclavian artery the ineisura is fol- lowed by a second negative deflection. On withdrawal these two merge and the dicrotic notch becomes progressively deeper. Central pulse preoscillations have been classically described in the horse by Chaveau and Marey12 and in the dog by Frank." The first preoscillation occurring in the latter part of diastole consists of a broad positive pressure wave and was attributed by the foregoing authors to atrial systole. The second preoscillation is a short positive pressure wave occurring immediately preceding the onset of the aortic pulse. Between the two waves there is a negative deflection with a sharp peak, somewhat resembling the ineisura. The genesis of these waves is still in dispute. However, Laszt and Miiller," in simultaneous recording of left ventricular and aortic pressure pulses in dogs, found the lowest dip of the negative deflection to coincide with the rapid increase in left ventricular pressure during isometric contraction. Preoscillations were seen in eight of the 12 healthy subjects investigated in the present study and were more clearly demarcated in recordings by the faster manometer systems 628 CENTRAL AXU PERIPHERAL PRESSURE PULSES IN MAN STRAIN GAUGE MANOMETER I CAPACITANCE MANOMETER POSITION: PROXIMAL SUBCLAVIAN ARTERY Downloaded from http://circres.ahajournals.org/ by guest on June 16, 2017 RADIAL ARTERY AT WRIST Fio. 2. Curves showing transformation of central to peripheral pulses at graded distances from subclavian artery near aorta. Discussion in text. than in recordings by the slower systems. The first preosoillation recorded in animal studies could not, however, be discerned in the records obtained. The negative V-shaped dip was most easily recognized and was usually followed by a short positive wave, which would be analogous to the second preoscillation obtained in animal studies. Preoscillations were less readily recognized during exercise than at rest but, when seen, the V-shaped dip was closer to the central pulse and the ascending limb of the pressure pulse seemed to arise from the onset of the second preoscillation. The preoscillations were most often and most clearly seen in the 70 degree head-up tilt position and then were a greater interval in advance of the ascending limb of the aortic pulse. Preoscillations were seldom seen beyond 10 cm. from the aortic arch, midsternal line. Table 5 shows the relationship of the onset of the central pressure pulse to the peak of the R wave of the electrocardiogram and to the anacrotic shoulder on the ascending limb of the pressure pulse in different conditions in healthy men. The interval from the peak of the R wave to the onset of pulse was shortest during exercise (64 msec), longest during the 70 degree head-up tilt (108 msec), the resting value occupying an intermediate position of 88 milliseconds. The anacrotic wave was present in the central pulse in all subjects at rest and exercise. In two subjects the anacrotic wave became the point of maximal pressure after the 70 degree head-up tilt and by definition could no longer be called an anacrotic wave. The TABLE 5.—Relationship of the Onset of the Central Pressure Pulse to the Peak of the R Wave of the Electrocardiogram and to the Anacrotic Wave* Condition Rest Exercise Tilt Constant Interval from Peak of R Wave to Onset of Central Pulse, Milliseconds Average Range Average Range Average Range SS 73-103 64 52-91 10S 97-1IS Interval Heart from Onset of Pulse to Rote, Annr.ro tic Beat* per Wave,Milli- Minute seconds 73 51-96 60 4S-S0 5S 46-87 64 54-SS 02 71-100 S9 69-113 * Data from 10 arterial catheterizations of eight healthy men. E. J. KROEKER AND E. H. WOOD TABLE G29 8.—Trayismission Times of Pressure Pulses Recorded Simultaneously From Different Arteries in Healthy Men Transmission Time* Constant Condition | lit II [J 160 12 77 8.8 183 S 75 It Rest Exercise Tilt Subjects Average Range Subjects Average Range Subjects Average Range Brachial Radial I 11 Femora ill I 8 Brachial to radial II ill 59 7.2 170 9 75 6.3 8 57 7.4 172 3 63 6.5 11 III S 20.6 12.5 144-186 65-105 6 .6-13 8 126-160 47-83 5 4-S.5 151-lSS 64-100 4.7-10.4 12-28 0 3-22 5 9 12 8 s 7.7 129 140 75 8.9 119 55 61 7.8 23 11.0 102-169 52-95 7 .2-13 8 100-143 37-69 6 2-11.6 91-142 41-71 5.4-12.4 16-28 9 6-16 9 168-195 62-95 147 7 S 14.1 .3-10 5 147-179 46-75 5 3-9.4 157-178 55-72 5.7-7.3 15-23 10 0-16 9 165 8.8 IS Downloaded from http://circres.ahajournals.org/ by guest on June 16, 2017 * Tho time interval from the peak of the R wave to the onset of the central pulse is shown in table 5. I I — msec, from peak of R wave of electrocardiogram to onset of pulse wave. J II = msec, from onset of central pulse to onset of peripheral pulse. § I I I = calculated pulse-wave velocity from central to peripheral recording site, meters per second. TABLE 7.—Buildup Times of Pressure Pulses Recorded Simultaneously From Different Arteries of Healthy Men Interval from On&et of Pulse to Maximal Pressure, msec. Condition Constant Radial Brachial 166 12 102 109 118-214 92-112 94-131 13S 117-160 12 96 S 102 9 114 82-120 74-131 92-139 8 93 S 94 3 91 85-100 S6-100 77-106 Central Rest Exercise Tilt Subjects Average Range Subjects Average Range Subjects Average Range 12 • anacrotic time, the time interval from the onset of the pulse wave to the point of maximal convexity of the anacrotic wave, showed little change during the physiologic conditions studied. Table 6 illustrates the transmission times of pressure pulses recorded simultaneously from different arteries in healthy men. The measurement from the peak of the R wave to the onset of the peripheral pulse is an absolute reference point for all pulses. It includes the possible latent period between the beginning of electric and the beginning of mechanical left ventricular systole, the period of isometric cardiac contraction and the transmission time of the impulse from the semilunar valves to the catheter tip and the peripheral arterial needle. 12 121 94-157 8 109 59-164 s Femoral 9 By and large, these results with subjects supine and at rest are similar to the values recorded in the literature and obtained by sphygmographic means.16'16 The pulse-wave velocity to the radial artery averaged 8.8 meters per second, to the brachial artery 7.2 meters per second and to the femoral artery 6.3 meters per second. Pulse-wave velocity from the brachial artery at the antecubita) fossa to the radial artery at the wrist averaged 12.5 meters per second. There was little change in the transmission time to the radial and brachial arteries during exercise and tilt. However, the average pulsewave velocity to the femoral artery increased by 1.5 meters per second during exercise, and the pulse-wave velocity from the brachial to 630 CENTRAL AND PERIPHERAL PRESSURE PULSES IN MAN TABLE 8.—Systolic Times of Pressure Pulses Recorded Simultaneously From Different Arteries of Healthy Men Interval from Onset of Pulse Wave to Dicrotic Notch, msec. Condition Rest Exercise Tilt Constant Subjects Average Range Subjects Average Range Subjects Average Range Central Radial Brachial Femoral 12 311 296-321 12 271 228-298 8 231 210-244 12 330 310-344 12 296 244-326 8 242 211-272 8 320 300-339 8 299 27S-317 S 237 213-267 9 336 306-365 9 291 227-325 3 234 214-263 Downloaded from http://circres.ahajournals.org/ by guest on June 16, 2017 the radial artery increased by an average of 1.6 meters per second during the tilt. Table 7 gives the buildup times of pressure pulses recorded simultaneously in healthy men. The average buildup time (onset of pulse to maximal pressure) of the central pulse was longest at rest (166 msec), shorter during exercise (121 msec.) and shortest during the tilt (109 msec). The buildup times of the peripheral pulses were uniformly shorter than the buildup times of the central pulses generated by the same heart beats. At rest, the average buildup time of radial pulses was 102 msec, of brachial pulses 109 msec, and of femoral pulses 138 msec. During exercise and the 70-degree head-up tilt, shortening of the buildup time peripherally persisted but to a lesser degree. Table 8 shows systolic times of pressure pulses recorded simultaneously from different arteries of healthy men during conditions of rest, exercise and 70 degree head-up tilt. All times were measured from the onset of the pulse wave to the lowest dip in pressure of the incisura or dicrotic notch. In the femoral pulses, in which the dicrotic notch is poorly defined, the descending limb of the primary peak was extended to meet the extension of a Line from the ascending slope of the undulation following the primary peak. Central systolic time was longest at rest (311 msec), shorter during exercise (271 msec.) and shortest during the 70 degree tilt (231 msec). The shortening of the systolic time during the tilt did not seem to be related merely to the heart rate, for the heart rate Heart Rate, Beats/min. 12 64 54-SS 12 99 71-125 S 89 69-113 was more rapid during exercise than in the tilt position. At rest, the peripheral systolic time was uniformly prolonged beyond that of the central pulse, brachial-artery values being closest to those of the central pulse and femoral-artery values deviating the most. COMMENT In 1890 Hiirthle17 noted a higher systolic and wider pulse pressure in the crural than in the carotid artery of the dog. Frank11 considered the increase in pulse pressure and the absolute increase in systolic pressure of the peripheral over the central pulse as its most prominent features and this led him to doubt the hypothesis of simple transmission of the pulse wave as in an endless elastic tube. These amplitude changes have been noted in man by Hamilton and co-workers4 and by Schnabel and co-workeis.8 These investigators found that, although there was an increase of systolic pressure peripherally, the diastolic and mean pressures remained relatively constant. Wood and co-workers13 demonstrated that in addition to the increase in systolic pressure peripherally there was a small but significant drop in diastolic and mean pressures. These latter findings have been confirmed by our present studies. The hypotheses of the transformation of the central to the peripheral pulse have been summarized by Hamilton and Dow.19 Our findings are best explained by the hypothesis that the increase in amplitude and change of contour in passage from the central to the peripheral pulse is due to the summation of the incident E. J. KROEKER AND E. H. WOOD Downloaded from http://circres.ahajournals.org/ by guest on June 16, 2017 wave with the reflected wave from the periphery. Resonance effects in the peripheral arterial systems probably also play a role, especially under conditions of cardiovascular stress such as the Valsalva maneuver.20 The increase in central pulse pressure amplification to the radial artery in the tilt above that seen at rest and exercise (table 4) may at least partially be attributed to compensatory vasoconstriction in this condition. The peripheral resistance would be increased and there would lie an increase of positive wave reflection. The smaller drop in mean pressure in this condition as compared to rest and exercise would lie compatible with reduced flow secondary to the vasoconstriction. The decrease in amplification of central pulse pressure at the femoral artery during exercise was marked as compared to the rest and tilt conditions (table 4). As the exercise consisted of pedalling with the opposite leg, a decrease in peripheral resistance would be expected secondary to vasodilatation in the vascular bed supplied by the femoral arteries; a decrease in positive wave reflection would result and therefore a decrease in amplification of pulse pressure. I t is probable that the character of the systolic ejection from the left ventricle also affects the degree of peripheral amplification of the pulse pressure. In severe aortic stenosis, for instance, the increase in systolic pressure peripherally is small or an actual decrease occurs, while in contrast the amplification of the pressure pulse peripherally is exaggerated in patients with severe aortic insufficiency (unpublished). SUMMABY AND CONCLUSIONS Central (aortic or subclavian), brachial, radial, and femoral pressure pulses were recorded simultaneously by means of intraarterial needles and catheters in 14 studies on 12 healthy subjects during conditions of rest, exercise and 70 degree head-up tilt. The average central arterial pressure was 126 (113-146) mm. Hg systolic and 81 (71 to 90) diastolic. Preoscillations and anacrotic oscillations were evident. As this pulse wave was transmitted peripherally, striking changes in contour occurred, along with a small pro- 631 gressive decrease in diastolic and mean pressure and a larger progressive increase in systolic pressure. Peripheral systolic pressure at rest uniformly exceeded the central systolic pressure generated by the same heartbeat; brachial was 109 per cent, radial 112 per cent and femoral 110 per cent of central systolic pressure. The average radial pulse pressure was 146, 146 and 165 per cent of central pulse pressure during rest, exercise and tilt respectively, while radial mean pressures were 94, 93 and 98 per cent of central mean pressures respectively. Femoral pulse amplitude was similar to that of the radial except during exercise of the opposite leg when femoral pulse presure was only 114 per cent of central pulse pressure. The velocity of the pulse wave was increased toward the periphery, the buildup time was shortened while the systolic time was prolonged. The contour of the pulse wave was uniformly strikingly different in the braehialradial and femoral-dorsalis pedis system. A double systolic wave developed in the pressure pulse during transmission peripherally from the subclavian artery. The primary systolic wave increased in amplitude, the secondary systolic wave usually decreased and the dicrotic dip was exaggerated during transmission down the arm. In contrast, the pressure pulse at the femoral artery consisted of a single systolic wave followed by a poorly defined or absent dicrotic dip. These time relations and changes in contour, in conjunction with corresponding pressure alterations, best describe the changes that the pulse wave undergoes in transit to the periphery. REFERENCES 1 1 quoted by HAUCK, E. G.: Blutdruckstudien am Menschen. Ztschr. f. Biol. 98: 3S7, 1937-193S. FAIVRE, MERKE, F. AND MCLLER, A.: Experimentelles zur Hydromechanik und Hiimodynamik. Ztschr. f. d. ges. exper. Med. 46: 332, 1925. ' W O L F , H. J. AND VON BONSDORFF, B.: Blutige Messung des absoluten Sphygmogmmms beim Menschen. Ztschr. f. d. ges. exper. Med. 79: 569, 1931. 4 HAMILTON, W. F., WOODBURY, R. A. AND HARPER, H. T., JR.: Physiologic relationships between 632 CENTRAL AND PERIPHERAL PRESSURE PULSES IN MAN intrathoracic, intraspinal tind arterial pressures. J.A.M.A. 107: S53, 1936. 'FULLER, J., TAYLOR, B. E., CLAGETT, 0. T. AND WOOD, E. H.: Blood pressure in the aorta during resection and repair of coarctation of the aorta. J. Lab. & Clin. Med. 39: 10, 1952. 6 SCHNABEL, T. G., JR., FITZPATRICK, H. F., PKTERSON, L. H., RASHKIND, W. J., TALLEY, ])., Ill, AND RAPHAEL, R. L.: A technic of Downloaded from http://circres.ahajournals.org/ by guest on June 16, 2017 vascular catheterization with small plastic catheters. Circulation 5: 257, 1952. 7 WOOD, E. H.: Special technics of value in the cardiac catheterization laboratorj'. Proc. Staff Meet., Mayo Clin. 28: 58, 1953. 8 WOOD, E. H.: Study of minimal dynamic response characteristics of manometer systems required for adequate recording of peripheral artery pressure pulses in man (abstract). Am. J. Physiol. 163: 762, 1950. 11 CHAVEAU, AND MAREY, E. J., L., AND MOLLER, A.: Uber 16 WEZLER, K. AND BSGER, A.: Die Dynamik des quoted by LASZT, den Dnickverlauf im linkeii Ventrikel und Vorhof und in der Aorta ascendens. Helvet. physiol. et pharmacol. acta. 9: 55, 1951. 11 FRANK, O.: Der Puls in den Arterien. Ztschr. f. Biol. 46: 441, 1905. 14 LASZT, L., AND MULLER, A.: Beitrag zur Herzdynamik. Ztschr. f. Kreislaufforsch. 42: 2S7, 1953. 15 WIGGERS, C. J.: Circulatory Dynamics: Physiologic Studies. New York, Grune & Stratton, 1952. WOOD, E. H., LKUSBN, I. R. AND WARNER, H. R.: arteriellen Systems: Der arterielle Blutdruck und seine Komponenten. Ergebn. Physiol. 41: 292, 1939. 17 HURTHLE, K.: Ueber den Ursprungsort der sekundaren Wellen der Pulscurve. Arch. f. d. ges. Physiol. 47: 17, 1890. Comparison of aortic pressure pulses recorded by strain-gauge and capacitance manometercatheter systems in man (abstract). Am. J. Physiol. 171: 7S0, 1952. 10 WOOD, E. H.: Special instrumentation problems encountered in physiological research concerning the heart and circulation in man. Science 112: 707, 1950. 11 WOOD, E. H., AND SUTTERER, W.: Improved resistance wire strain-gauge manometers adaptable for biologic measurements. J. Lab. & Clin. Med. 46: 153, 1955. Intraluminal pressures recorded simultaneously from different arteries in man (abstract). Am. J. Physiol. 167: S3S, 1951. 19 HAMILTON, W. F., AND DOW, P.: An experimental study of the standing waves in the pulse propagated through the aorta. Am. J. Physiol. 125: 48, 1939. 10 KROEKER, E. J., AND WOOD, E. H.: Comparison of central and peripheral arterial pressure pulses in man. Abstracts of Communications XLX. International Physiological Congress, Montreal, Aug. 31-Sept. 4, 1953, p. 535. 11 18 WOOD, E. H., FULLER, J., AND CLAGETT, 0. T.: Comparison of Simultaneously Recorded Central and Peripheral Arterial Pressure Pulses During Rest, Exercise and Tilted Position in Man EDWIN J. KROEKER and EARL H. WOOD Downloaded from http://circres.ahajournals.org/ by guest on June 16, 2017 Circ Res. 1955;3:623-632 doi: 10.1161/01.RES.3.6.623 Circulation Research is published by the American Heart Association, 7272 Greenville Avenue, Dallas, TX 75231 Copyright © 1955 American Heart Association, Inc. All rights reserved. Print ISSN: 0009-7330. Online ISSN: 1524-4571 The online version of this article, along with updated information and services, is located on the World Wide Web at: http://circres.ahajournals.org/content/3/6/623 Permissions: Requests for permissions to reproduce figures, tables, or portions of articles originally published in Circulation Research can be obtained via RightsLink, a service of the Copyright Clearance Center, not the Editorial Office. Once the online version of the published article for which permission is being requested is located, click Request Permissions in the middle column of the Web page under Services. Further information about this process is available in the Permissions and Rights Question and Answer document. Reprints: Information about reprints can be found online at: http://www.lww.com/reprints Subscriptions: Information about subscribing to Circulation Research is online at: http://circres.ahajournals.org//subscriptions/