Survey
* Your assessment is very important for improving the work of artificial intelligence, which forms the content of this project
* Your assessment is very important for improving the work of artificial intelligence, which forms the content of this project
J. Micromech. Microeng. 10 (2000) 415–420. Printed in the UK PII: S0960-1317(00)11716-4 A pneumatically-actuated three-way microvalve fabricated with polydimethylsiloxane using the membrane transfer technique Kazuo Hosokawa† and Ryutaro Maeda Surface and Interface Technology Division, Mechanical Engineering Laboratory, AIST/MITI, 1-2 Namiki, Tsukuba, Ibaraki, 305-8564 Japan E-mail: [email protected] Received 8 February 2000, in final form 14 April 2000 Abstract. In this paper, a three-way microvalve system composed of three independent one-way valve units is presented. Each valve unit has a membrane, which is actuated by external negative air pressure. Intervals between the valve units are smaller than 780 µm, which opens up the possibility of realizing a high-density microvalve array. The small intervals were realized by providing the system with a layer of microchannels to conduct the air pressure to the valve units. In spite of the extra layer of microchannels, the device has been fabricated through a simple process by adopting polydimethylsiloxane (PDMS) as the material for the microchannel chips as well as the membrane. In particular, a newly developed technique for wafer level transfer of a PDMS membrane has been proven to be effective. Flow characteristics of the microvalve system for water are presented. The microvalve works in an on–off manner with hysteresis. No leakage has been observed in the closed state. In the open state, measured flow resistances (pressure drops) are within the range of 1.65–2.29 kPa (µl min−1 )−1 , and consistent with an electric circuit model. 1. Introduction Recently, microfluidic devices such as microvalves have been attracting more and more attention, especially for (bio)chemical applications including micro total analysis systems (µTAS) [1, 2]. For example, miniaturization of a flow injection analysis (FIA) system requires a set of microvalves to control a pulsed sample flow [3, 4]. In a typical design, a microvalve has a microfabricated channel interrupted by a valve seat. The fluidic route is opened/closed at the valve seat with a membrane, which is actuated via various methods [5]. A key issue in the valve design is material for the membrane. A large deflection comparable to the channel height—several tens of micrometers in many cases—is required for switching the liquid, which is essential in biochemical applications. It has recently been recognized that silicone elastomer is one of the most suitable materials for the microvalve membrane because of its low Young’s modulus and excellent sealing property [6–10]. On the other hand, a disadvantage of the silicone elastomer membrane is the limited variety of actuation methods. Only two methods—pneumatic [6–9] and thermopneumatic [10]—have been reported so far. These methods are problematic for fabricating a high-density array † Author to whom correspondence should be addressed. 0960-1317/00/030415+06$30.00 © 2000 IOP Publishing Ltd of microvalves. In the case of thermopneumatic actuation, insulation of heat between neighboring valves is difficult in a high-density array system. In the case of pneumatic actuation, microfabricated pneumatic channels are required to conduct the air pressure to the membranes. This means an extra layer of microchannels besides those for the fluid to be switched. In general, a complicated fabrication process is required for such a device containing multilayered microchannels. Pneumatically driven three-way [7] and four-way [8] microvalves have already been reported. They are composed of three and four independent one-way valve units, respectively. However, they are not equipped with pneumatic microchannels. Hence, the intervals between the valve units seem to be limited by the sizes of non-microfabricated pneumatic connectors. The intervals in these systems are larger than 2.5 mm. This paper describes a three-way microvalve system composed of three one-way valve units. Each valve unit has a silicone elastomer membrane, which is actuated by external negative air pressure. Intervals between the valve units are smaller than 780 µm. These small intervals have been achieved by providing the system with pneumatic microchannels. To simplify the fabrication process for this relatively complicated device containing multilayered microchannels, 415 K Hosokawa and R Maeda (A) (B) (C) Figure 1. (A) Schematic drawing of the design of the three-way microvalve system made with three one-way valve units. The fluidic chip has three fluidic ports (In1 and 2 and Out), and the pneumatic chip has three control ports (C1–3). The main working region A is detailed in (B), where the dashed lines indicate microchannels between the pneumatic chip and the membrane. Sectional views (section B–B ) of a one-way valve unit are shown in (C). we adopted polydimethylsiloxane (PDMS)—a kind of silicone elastomer—as material for the device. Recently, the replica molding technique of PDMS has been widely used to simplify the fabrication process of microfluidic devices [9, 11–19]. Commonly used procedures are: (1) a PDMS chip with grooves on its surface is molded against a microfabricated negative master, and (2) the grooves are sealed with a flat substrate to make microchannels. The simple fabrication process is enabled by two features of PDMS. One is the submicrometer replica fidelity [11], which is sufficient for most applications. The other is the selfadhesion property. To seal the grooves, a PDMS surface can be reversibly bonded to various surfaces—glass, silicon, PDMS itself, and so on—without the need for an elaborate bonding technique. If necessary, they can also be bonded irreversibly by treating the surfaces with an oxygen plasma before they are brought into contact [13, 14]. The device described here is composed of two PDMS microchannel chips and a PDMS membrane. They have been assembled using both the bonding techniques, reversible and irreversible. The latter has been utilized to transfer the PDMS membrane to one of the PDMS chips from a silicon wafer, on which the membrane was originally spin-coated. Besides the simple fabrication process, the use of PDMS has additional advantages, as follows. This material is tough, inexpensive, transparent (suitable for optical detection), and biocompatible (even implantable into human bodies). However, it is swelled by many organic solvents and hence unsuited for use with such solvents. The device consists of two PDMS chips—referred to as ‘fluidic chip’ and ‘pneumatic chip’—and a PDMS membrane sandwiched between the two chips. Although the device has no specific flow direction, three access ports in the fluidic chip are referred to as ‘inlet ports’ (In1 and 2) and ‘outlet port’ (Out) for convenience. On the other hand, the pneumatic chip has three control ports (C1–3). On the fluidic and the pneumatic chips, there are microfabricated grooves with depths of 25 µm and 70 µm, respectively. The grooves are sealed with both sides of the membrane to make two layers of microchannels. The fluidic chip is reversibly bonded to the membrane, whereas the pneumatic chip is irreversibly bonded to it. The functional microstructures are magnified in figure 1(B). Three 100 µm wide fluidic channels from the inlet and the outlet ports meet at the center of the chip. All the channels have one-way valve units, which are equivalent in essence. Sectional views of a valve unit are illustrated in figure 1(C). A 50 µm gap interrupting the fluidic channel works as a valve seat. The valve unit is normally closed by sealing the valve seat with the membrane. External negative air pressure, supplied via a control port and a 200 µm wide pneumatic channel, deflects an active area (450 µm by 200 µm) of the membrane to open the valve unit. For example, the fluidic route from In1 to Out opens by applying negative air pressure to C1 and C3 at the same time. The dead volume of each valve unit is smaller than the cavity on the pneumatic chip: 450 µm × 200 µm × 70 µm = 6.3 nl. 2.2. Fabrication 2. Design and fabrication 2.1. Design Figure 1 shows the design of the three-way microvalve system made with three independent one-way valve units. 416 The device fabrication process is outlined in figure 2. First, the pneumatic chip was fabricated by a molding technique as follows. To obtain a negative pattern for the pneumatic channels with a height of 70 µm, an ultrathick photoresist (SU-8; MicroChem, USA) was spin-coated on a silicon wafer A pneumatically-actuated PDMS three-way microvalve A Closed 0 kPa 0 kPa Closed Closed 0 kPa B Open -60 kPa -60 kPa Open Closed Figure 2. Schematic diagram of the fabrication process of the three-way microvalve. 0 kPa and processed according to the manufacturer’s instructions (figure 2(A)). After development, the wafer was baked at 150 ◦ C for 5 min in an oven to reinforce the adhesion, and then it was gradually cooled to room temperature over 1–2 h. For mold release, the wafer was coated with a fluorocarbon layer polymerized by a CHF3 plasma in a reactive ion etching (RIE) machine (System VII SLR 730/740; Plasma-Therm, USA) for 2 min under the following conditions: CHF3 gas flow of 50 sccm, pressure of 160 mTorr, and electric power of 200 W. A prepolymer solution of PDMS (Sylgard 184; Dow Corning, USA, base:curing agent = 10:1) was poured onto the wafer with a frame for holding the solution (figure 2(B)). It was cured in the oven at 65 ◦ C for 1 h, followed by the second cure at 100 ◦ C for 1 h. The cured PDMS chip was peeled from the wafer, and three access holes of 1.5 mm diameter were punched in the chip using a metal pipe (figure 2(C)). Second, the PDMS membrane was formed on another silicon wafer and transferred onto the pneumatic chip using a new technique, as follows. In advance, the silicon wafer was coated with a CHF3 -plasma-polymerized fluorocarbon layer using the process described above. The prepolymer solution of PDMS was spin-coated on the wafer at 3000 rpm for 30 s and cured in the oven at 100 ◦ C for 1 h. As a result, a 25 µm thick PDMS membrane was obtained. For the irreversible bonding of the pneumatic chip and the membrane, both surfaces were treated with an oxygen plasma in the RIE machine for 1 min under the following conditions: oxygen gas flow of 100 sccm, pressure of 300 mTorr, and electric power of 200 W. Immediately after removal from the plasma chamber, the two surfaces were brought into contact, and baked in the oven at 100 ◦ C for 2 h (figure 2(D)). Since the two parts were irreversibly bonded, they could be peeled off together from the silicon wafer without crumpling the membrane (figure 2(E)). Finally, all the parts were assembled. The fluidic chip was fabricated in the same way as the pneumatic chip, C Open -60 kPa -60 kPa Open Open -60 kPa 500 µm Figure 3. Optical micrographs of the three-way microvalve. (A) All the fluidic routes are closed. (B) The fluidic route from In1 to Out is opened by applying a negative pressure of −60 kPa to the control ports C1 and C3. (C) All the fluidic routes are opened by applying a negative pressure of −60 kPa to all the control ports. except that the thickness of the SU-8 photoresist was 25 µm. The fluidic chip was reversibly bonded to the composite of the pneumatic chip and the membrane only by bringing the two surfaces into contact. Since PDMS is transparent, alignment is fairly easy using a video microscope (VH-6300; KEYENCE, Japan) and a homemade tool based on an x–y–z stage. Six glass pipes were inserted into the access holes, and glued with PDMS (figure 2(F)). Figure 3 shows optical micrographs of the fabricated microvalve system in different states. 3. Flow characteristics 3.1. Experimental set-up and methods Flow characteristics of the three-way microvalve system for water were evaluated using the set-up depicted in figure 4. 417 K Hosokawa and R Maeda Figure 4. Experimental set-up for evaluation of flow characteristics of the three-way microvalve for water: (1) vacuum pump, (2 and 3) vacuum regulators, (4–7) manual three-way valves, (8) trap for water, (9) the device, (10 and 11) silicone tubes filled with pure water. Negative pressure was used not only for controlling the valve membranes but also for pumping the water, because the reversible bonding between the fluidic chip and the membrane is not perfect against positive internal pressure higher than 10 kPa. The negative pressure was supplied by a vacuum pump (DA-5D; ULVAC Sinku Kiko, Japan), and regulated for water pumping and valve control independently using two vacuum regulators (VR200-G; Koganei, Japan). Four manual three-way valves were used for switching the pressures in the ports from vacuum to atmospheric pressure and vice versa. Pressure in each port is denoted by pport name . The water pumped from the outlet port was trapped in a bottle to prevent the regulator from sucking in the water. The inlet ports were connected to 1 m long, 1 mm internal diameter silicone tubes filled with pure water. The ends of the tubes were exposed to the atmosphere. Neglecting the height difference of several centimeters between the device and the tubes lying on a table, we regard the pressures pIn1 and pIn2 as atmospheric. We measured the moving velocities of the water menisci in the tubes to calculate the volumetric flow rates denoted by q1 and q2 for In1 and In2, respectively. Figure 5. Flow characteristics of the three-way microvalve for water. (A) Opening/closing behavior of the fluidic route In1–Out. The other route, In2–Out, was kept closed. (B) Flow rate against pumping pressure for the route In1–Out in the open state. The other route, In2–Out, was kept closed. (C) Flow rates against pumping pressure for both of the fluidic routes in the open states. 3.2. Results and discussion In figure 5(A), the flow rate q1 is plotted against the control pressures pC1 and pC3 , which were varied together from 0 to −70 kPa. The pressures pC2 and pOut were kept constant as shown in the insert. This curve illustrates the opening/closing behavior of the fluidic route In1–Out when the other route In2–Out is closed. The curve has almost no linear range in practice. In other words, the set of the two valve units, controlled by C1 and C3, works as an on–off valve. Since all three valve units are virtually equivalent, other combinations (C2–C3 and C1–C2) should have on–off characteristics similar to figure 5(A). The hysteresis between the opening and the closing pressures is considered to be caused by sticking between the membrane and the valve seats. No leakage could be detected in the closed state. Although the current device does not stand positive internal pressure higher than 10 kPa, it is worthwhile to estimate the behavior of the valve operated with a high 418 positive pressure. As shown in figure 5(A), the required control pressure to open the valve is −30 to −40 kPa, which means a 10–20 kPa pressure difference across the membrane, because pOut was kept at −20 kPa. Therefore, in the case of pIn1 > 20 kPa and pOut = 0, the valve would become a ‘normally-open’ valve which could be closed by a positive control pressure. Another conceivable operation mode is the ‘passive mode’ with variable inlet pressure and constant control pressure. When the inlet pressure is lower than a threshold value (10–20 kPa higher than the control pressure), the valve would be closed. By increasing the inlet pressure, the valve would be abruptly opened at the threshold value. To render the microvalve resistant to a high positive pressure, all the parts should be irreversibly bonded except for the active areas of the membrane. Such selective irreversible bonding would be possible by using a protective layer, such as polyimide or Parylene, patterned onto the active areas. A pneumatically-actuated PDMS three-way microvalve 4. Conclusions Figure 6. (A) Electric circuit model of the three-way microvalve system. (B) Designed lengths of the fluidic channels. In figure 5(B), the flow rate q1 is plotted against the pumping pressure pOut , which was varied from 0 to −30 kPa. The control pressures were kept constant as pC1 = pC3 = −60 kPa and pC2 = 0 kPa, which rendered the fluidic route In1–Out open and the route In2–Out closed. A photograph of this state is presented in figure 3(B). As shown in figure 5(B), the flow rate is proportional to the pumping pressure. Therefore, the flow resistance (pressure drop) can be defined as the factor of proportionality. By fitting the curve in figure 5(B), the flow resistance is calculated as RA = 1.65 kPa (µl min−1 )−1 . Figure 5(C) shows the results of an experiment similar to the above. By applying −60 kPa pressure to all the control ports, all the fluidic routes were kept open, as shown in figure 3(C). As expected, the two flow rates q1 and q2 in figure 5(C) are mutually balanced and proportional to the pumping pressure. By fitting the two curves, the flow resistances are calculated as RB = 2.24 kPa (µl min−1 )−1 and RC = 2.29 kPa (µl min−1 )−1 for the fluidic routes In1–Out and In2–Out, respectively. The above flow resistances are consistent with the electric circuit model shown in figure 6(A), where the symbol R denotes the flow resistance of each channel. Neglecting the small difference between RB and RC calculated above, we assume that the system is symmetric with respect to the two inlet ports. In this case, the two resistance values should be represented by the mean value RB = (RB +RC )/2 = 2.27 kPa (µl min−1 )−1 . Considering the combined resistances, the experimental results are expressed as: R0 + R1 = RA R0 + (R1 /2) = RB /2. Rearranging the above equations, we obtain: R0 = RB − RA = 0.62 kPa (µl min −1 )−1 R1 = 2RA − RB = 1.03 kPa (µl min −1 )−1 . The difference between R0 and R1 probably originates from the difference in the channel length illustrated in figure 6(B), because, in general, the flow resistance of a microchannel with a uniform cross section is proportional to its length. This relationship is formulated as Hele–Shaw flow or Poiseuille flow in the low Reynolds number regime [20]. As for the above experiments, the ratio of the flow resistances, R0 /R1 = 0.60, roughly agrees with the ratio of the channel lengths, L0 /L1 = 0.66. This agreement suggests that the flow resistances in the open states are dominated by the straight regions of the microchannels, not by the valve units. A pneumatically-actuated three-way microvalve system composed of three one-way valve units was constructed. Intervals between the valve units are smaller than 780 µm, which opens up the possibility of realizing a high-density microvalve array. The device containing multilayered microchannels was fabricated through a relatively simple process, which was enabled by the use of PDMS. In particular, a new technique for wafer level transfer of a PDMS membrane has been proven to be effective. Since PDMS can be irreversibly bonded to various materials including glass and silicon [13], this technique is considered to be widely applicable to multilayer microchannel systems. There are at least three issues for future study. First, resistance to positive internal pressure should be improved, as discussed in section 3.2. Second, the robustness for practical applications, such as lifetime and particle tolerance, should be tested. Finally, the geometrical design should be optimized. The miniaturization is currently limited by alignment inaccuracy—tens of micrometers—in the assembly process. To reduce it to several micrometers, a more sophisticated technique using a mask aligner is under development. As for the thickness of the membrane, it has been reported that a 2 µm thick PDMS membrane can be spin-coated and peeled off [21]. Therefore, from the fabrication point of view, we consider that the microvalve system can be further miniaturized by a factor of ten. However, from the application point of view, there are some trade-off relationships. For example, miniaturization would cause an increase in the flow resistance, and an extremely thin membrane would be problematic for robustness. These relationships should be quantitatively characterized to enable optimized valve design according to its application. References [1] van den Berg A and Bergveld P (eds) 1995 Micro Total Analysis Systems (Dordrecht: Kluwer) [2] Harrison D J and van den Berg A (eds) 1998 Micro Total Analysis Systems 98 (Dordrecht: Kluwer) [3] Manz A, Verpoorte E, Raymond D E, Effenhauser C S, Burggraf N and Widmer H M 1994 µ-TAS: miniaturized total chemical analysis systems Proc. Micro Total Analysis Systems (Enschede, The Netherlands) pp 5–27 [4] Sjölander S and Urbaniczky C 1991 Integrated fluid handling system for biomolecular interaction analysis Anal. Chem. 63 2238–45 [5] Shoji S and Esashi M 1994 Microflow devices and systems J. Micromech. Microeng. 4 157–71 [6] Vieider C, Ohman O and Elderstig H 1995 A pneumatically actuated micro valve with a silicone rubber membrane for integration with fluid-handling systems Proc. IEEE Int. Conf. Solid-State Sensors Actuators (Stockholm, Sweden) vol 2 pp 284–6 [7] Ohori T, Shoji S, Miura K and Yotumoto A 1997 Three-way microvalve for blood flow control in medical micro total analysis systems (µTAS) Proc. IEEE Micro Electro Mechanical Systems (Nagoya, Japan) pp 333–7 [8] Bousse L, Dijkstra E and Guenat O 1996 High-density arrays of valves and interconnects for liquid switching Proc. Solid-State Sensor and Actuator Workshop (Hilton Head, USA) pp 272–5 [9] Hosokawa K and Maeda R 2000 A normally closed PDMS (polydimethylsiloxane) microvalve Trans. IEE Japan E 120 177–8 419 K Hosokawa and R Maeda [10] Yang X, Grosjean C, Tai Y C and Ho C M 1999 A MEMS thermopneumatic silicone rubber membrane valve Sensors Actuators A 64 101–8 [11] Delamarche E, Bernard A, Schmid H, Michel B and Biebuyck H 1997 Patterned delivery of immunoglobulins to surfaces using microfluidic networks Science 276 779–81 [12] Effenhauser C S, Bruin G J M, Paulus A and Ehrat M 1997 Integrated capillary electrophoresis on flexible silicone microdevices: analysis of DNA restriction fragments and detection of single DNA molecules on microchips Anal. Chem. 69 3451–7 [13] Duffy D C, McDonald J C, Schueller O J A and Whitesides G M 1998 Rapid prototyping of microfluidic systems in poly(dimethylsiloxane) Anal. Chem. 70 4974–84 [14] Duffy D C, Schueller O J A, Brittain S T and Whitesides G M 1999 Rapid prototyping of microfluidic switches in poly(dimethyl siloxane) and their actuation by electro-osmotic flow J. Micromech. Microeng. 9 211–7 [15] Kenis P J A, Ismagilov R F and Whitesides G M 1999 Microfabrication inside capillaries using multiphase laminar flow patterning Science 285 83–5 420 [16] Armani D, Liu C and Aluru N 1999 Re-configurable fluid circuits by PDMS elastomer micromachining Proc. IEEE Micro Electro Mechanical Systems (Orlando, FL) pp 222-7 [17] Hosokawa K, Fujii T and Endo I 1999 Droplet-based nano/picoliter mixer using hydrophobic microcapillary vent Proc. IEEE Micro Electro Mechanical Systems (Orlando, FL) pp 388–93 [18] Hong J W, Hosokawa K, Fujii T, Seki M and Endo I 1999 An inexpensive PDMS (polydimethylsiloxane) microchip for capillary gel electrophoresis Proc. IEEE Int. Conf. Solid-State Sensors and Actuators (Sendai, Japan) pp 760–3 [19] Hosokawa K, Fujii T and Endo I 1999 Handling of picoliter liquid samples in a poly(dimethylsiloxane)-based microfluidic device Anal. Chem. 71 4781–5 [20] Brody J P, Yager P, Goldstein R E and Austin R H 1996 Biotechnology at low Reynolds numbers Biophys. J. 71 3430–41 [21] Jackman R J, Duffy D C, Cherniavskaya O and Whitesides G M 1999 Using elastomeric membranes as dry resist and for dry lift-off Langmuir 15 2973–84