Download IC 15030870 rishabh_C12

Survey
yes no Was this document useful for you?
   Thank you for your participation!

* Your assessment is very important for improving the workof artificial intelligence, which forms the content of this project

Document related concepts

Oscilloscope history wikipedia , lookup

Audio power wikipedia , lookup

Flip-flop (electronics) wikipedia , lookup

Phase-locked loop wikipedia , lookup

Ohm's law wikipedia , lookup

Multimeter wikipedia , lookup

Josephson voltage standard wikipedia , lookup

Amplifier wikipedia , lookup

Analog-to-digital converter wikipedia , lookup

Radio transmitter design wikipedia , lookup

Power MOSFET wikipedia , lookup

Transistor–transistor logic wikipedia , lookup

Immunity-aware programming wikipedia , lookup

CMOS wikipedia , lookup

Integrating ADC wikipedia , lookup

Wilson current mirror wikipedia , lookup

Valve audio amplifier technical specification wikipedia , lookup

Resistive opto-isolator wikipedia , lookup

Surge protector wikipedia , lookup

Valve RF amplifier wikipedia , lookup

Operational amplifier wikipedia , lookup

Voltage regulator wikipedia , lookup

Schmitt trigger wikipedia , lookup

Power electronics wikipedia , lookup

Current mirror wikipedia , lookup

Opto-isolator wikipedia , lookup

Switched-mode power supply wikipedia , lookup

Rectiverter wikipedia , lookup

Transcript
Operating Instructions Transducer
for Phase angle difference Rish Ducer C12
9 10 11 12 13 14 15 16
RISHABH INSTRUMENTS PVT. LTD.
F-31, M.I.D.C., SATPUR,
NASIK-422 007, INDIA.
RISH Ducer C12
4
0.5
PHASE ANGLE DIFFERENCE TRANSDUCER
C11-31B1 D14E 90
SR.NO:07/01/12598
15-
16+
AUX: 230V, 50/60Hz,4 VA
3
UL1
INPUT
415 V
50 Hz
1A
11 N
2
N
4
DIN 43807
UL1 2
DIN 43807
1
CP = 1
9-
10+
OUTPUT :P: 4...20mA R max 750 W
I
-60...600 el
I
1
2
3
4 5
IC15030870
RISHABH
INSTRUMENTS
Measure, Control & Record with a Difference
6
7
8
RISHABH INSTRUMENTS PVT. LTD.
F-31, M.I.D.C., Satpur, Nashik 422 007, India.
Tel. : +91 253 2202160, 2202202,
Fax : +91 253 2351064
Email : [email protected]
www.rishabh.co.in
IC 150870 rishabh_c12 26 Aug 09
Safety precautions to be strictly observed are marked
with following symbols in the operating instructions :
6.5
19
9.Dimensional drawings
04.5
Operating Instructions
Transducer for Phase Angle Difference RishDucer C12
9 10 11 12 13 14 15 16
9 10 11 12 13 14 15 16
Measuring Range D
: ±1200 el
Nominal Frequency
: 50 or 60 Hz
Nominal Input Voltage : For both inputs
100/ 3, 110/ 3, 100, 110,
200, 230, 400 or 500 V
1
2
3
4
5
6
7
8
70
1
2
3
4
5
6
7
154
140
125
Measuring Input
130
125
4. Technical Data (Refer Fig.5)
8
126
12
70
125
Measuring Output
DC Voltage : 0....10 / 1....5 / -10....0....10 V.
Load capacity 20 mA
External resistance Rext (KW
) > UAN [V]
20mA
UAN = full output value
Contents
1.
2.
3.
4.
5.
6.
7.
8.
9.
10.
Read first and then .....................................2
Scope of supply ..........................................2
Brief description..........................................2
Technical data.............................................2
Mounting.....................................................3
Electrical connections.................................3
Commissioning and maintenance ..............4
Releasing the transducer............................4
Dimensional drawing ..................................5
Special Features.........................................5
DC Current : 0....1mA to 0....20mA / 4.....20 mA
-1...0...1 mA to -20...0...20 mA
Burden Voltage ±15 V
External Resistance Rext max. [KW
] = 15V
IAN [mA]
IAN = Full output value
1. Read first and then
The proper and safe operation of the
device assumes that the operating
instructions are read and the safety
warning given in the various sections
5. Mounting
6. Electrical Connections
are observed.
DC Voltage
The device should only be handled by appropriately
trained personnel who are familiar with it and
authorised to work in electrical installations.
: 24, 115, 120, 230 V or 240 V ± 20%
42 to70 Hz
Power consumption approx. 4VA
: 24....90 V or 90....240 V
-15 /+33%
Power Consumption approx. 4W
Accuracy
Reference value
Basic accuracy
2. Scope of supply (Fig. 1)
: Output span
: Class 0.5
9 10 11 12 13 14 15 16
RISHABH INSTRSUMENT PVT.DLT .
Electrical Safety
F-31, M.I.D.C., SATPUR,
NASIK-422 007, INDIA.
RISH Ducer C12
4
15-
Installation Category : CAT III, Max. voltage to
ground 500 V
Pollution degree
: 2
Test Voltage
: 4 KV between electrically
isolated circuits and versus
housing
16+
AUX: 230V, 50/60Hz,4 VA
11 N
INPUT
DIN 43807
UL1
415 V
50 Hz
1A
DIN 43807
UL1 2
3
N
2
4
CP = 1
9-
10+
OUTPUT :P: 4...20mA R max 750 W
I
-60...600 el
I
1
2
3
4 5
Measuring range D
0
0
1 between ±60 el. and ± 175 el.,
0
but for angle < ±120 el.,
additional error 0.5%
Nominal Frequency fN
2 between 16 and 400 Hz
apart from the standard ranges 50 or 60 Hz
Limitation at fN > 100 Hz :
addtional error 0.2%
Limitations at 16 £
fN < 50 Hz :
0
posssible only with measuring ranges >± 60 el.
additional error 0.3 %
response time t90 < 2 s
Power supply derived from measuring input not
possible
6
7
8
Fig.1
3. Brief Description :
Environmental conditions
The RishDucer C12 measuring transducer is used to
measure phase angle difference between two
synchronised single phase balanced network having
sine wave form. The output signal is proportional load
independent DC voltage or current.
The transducer is available in AC or DC auxiliary
supply and also in self power version.
Climatic range
Operating Temperature
Storage Temperature
Relative humidity
of annual mean
2
: Climate class 3Z acc. to
VDI/VDE 3540
: -250C to +550C
: -40 to +700C
: £
75%
7
9
Bipolar-symmetrical load-independent
DC current
Ranges between -1...0...1 and -20...0...20 mA, other
than the standard ranges -1...0...1/-2.5...0...2.5/
-5...0...5/-10...0...10 & -20...0...20mA
£
0.5% p.p. instead of £
2% p.p.
Limitations :
possible only with nominal frequency ³
50
response time < 1 s
(not suitable for synchronising operations)
Power supply
without separate power supply connection
11
Power supply from voltage input signal
(³
24 V to 500, fN 50 or 60 Hz)
Limitations :
Reference condition :
Input voltage UN ± 15%
with UN ³
170 V
Impulse withstand voltage acc. to IEC 255-4, CI. II :
1 kV,1.2/50 µs, 0.5 Ws or
overload capacity of the voltage input max.
680 V~, 2 s
The additional power taken from the input voltage
signal is approx. 4 VA
Output signal A (measuring output A)
Unipolar load-independent DC voltage*
4
Ranges between 0...1 and 0....15 V, other than
the standard range 0.....10 V
6
Live-zero
Ranges between 1..5 and 4...20 mA, other than
the standard range 4...20 mA
10
Nominal Input voltage UN
3 between 10 and 600 V, other than the standard
values 100/ 3, 110/ 3, 100, 110, 200, 230,
400 or 500 V
Limitation :
with UN > 500 V overload capacity 2000 V, 2s
5
8
Residual ripple in output current
0.5
PHASE ANGLE DIFFERENCE TRANSDUCER
C11-31B1 D14E 90
SR.NO:07/01/12598
1
10. Special Features
Power Supply
AC Voltage
Fig.8 Fig.7 RISH Ducer C12 in housi ng E16 with the screw
hole brackets pulled out for wall mount ing.
Fig.7 RISH Ducer C12 in housing E16 clipped onto a top hat
rail (35 x 15 mm or 35 x 7.5 mm, acc. to EN 50 022).
Live-zero*
Ranges between 0.2...1 and 3...15 V, other than
the standard range 1...5 V
*Limitation with UAN < 4 V
Additional error :
Dependence on external resistance
D
Rext max. = 0.2%, reference conditions:
External resistance 2 x Rext min. ±
20%
12
Bipolar-symmetrical load-independent
DC Voltage*
Ranges between -1..0...1 and -15...0...15 V,
other than the standard range -10...0...10 V
with AC Voltage
any voltage between 24 and 500 V, ± 20 %,
42 to 70 Hz.
Power consumption approx. 4 VA
apart from the standard voltages 24, 115, 120, 230
and 240 V
Climatic
Climate class 3Z acc. to VDI/VDE 3540, but
13
0
temperature continuously -25 to +55 C.
Relative humidity £
90% annual mean
(application class HVR acc. to DIN 40 040)
Unipolar load-independent DC Current
Ranges between 0...1 and 0...20 mA, other than
the standard ranges 0...1/0...5/0...10 & 0...20 mA
5
IC 150870 rishabh_c12 26 Aug 09
5. Mounting
RISHABH INSTRUMENTS PVT. LTD.
F-31, M.I.D.C., SATPUR,
NASIK-422 007, INDIA.
Manufacturer
RISH Ducer C12
4
0.5
PHASE ANGLE DIFFERENCE TRANSDUCER
C12-32A1 Z09E A0
SR. NO.: 07/01/12598
15-
Type
Works No.
16+
Note "Environmental conditions" in section
"4.Technical Data" while deciding the place
of installation!
Power Supply
AUX: 230V, 50/60HZ, 4 W
3
UL1
2
INPUT
415 V, 50 Hz
11
INPUT
415 V, 50 Hz
N
2
N
4
Measuring range
Measured quantity
IL1
6
Nominal Frequency
5.1 Top-hat rail mounting
Output signal
External resistance
Simply clip the device onto the Top-hat rail
(EN 50 022) (See Fig. 2)
Input
DIN 43807
UL1
DIN 43807
1
3
10+
9OUTPUT: 4...20mA
I
I
Fixing the housing to a wall
or mounting plate using two
4 mm diameter screws.
Drill holes as shown in the
drilling pattern (Fig.4)
The RISHDucer C 12 can be mounted either on a top-hat
rail or directly onto a wall or mounting plate.
Output
R max 750 W
Fig.4. Drilling Pattern
6. Electrical connections
Make connection as per printed label on transducer
(Fig.5,example of a nameplate)
0
-60...60 el.
Fig.5. Declaration of type label
Electrical Connections
Meaning of symbols on device
Front
Warning of danger (Caution, see documentation)
Double insulation Class II device
4
Impending danger by high input voltage
or high mains voltage.
Be aware of danger of open Current
transformer secondary. Make sure that
the measuring input cables are not
live(potential-free) when making the
connections !
Fig.2 Mounting onto
top-hat rail 35 x 15
or 35 x 7.5 mm.
(_) (+ )
_ +
Test voltage 4 KV
9 10 11 12 13 14 15 16
Note : It is strongly recommended to employ a circuit
breaker in building installation to make provision for
disconnection of the power supply to device and to
provide protection in case of short circuit. The circuit
breaker should be close to device, easily reachable
and appropriately identified.
5.2 Wall mounting
The screw hole brackets (1) can be released and
pulled out by pressing on the latch (4). They can be
pushed in after pressing the latch (5).
7. Commissioning and maintenance
Switch on the power supply and the measuring input.
During the operating, you can disconnect the output
and connect a test epuipment e.g. For a functional
test.
No maintenance is required.
Note that,.....
... the data required to carry out the
prescribed measurement must
correspond to those marked on the
nameplate of the RISHDucer C12
(
measuring input,
measuring
output and
power supply. See Fig.5)
1 2 3 4 5 6 7 8
8. Releasing the transducer
Release the transducer from a top-hat rail as
shown in Fig. 6
UG
(U1)
1
2
3
4
5
6
7
8
... the total loop resistance connected to
the output (receiver plus leads) does not
exceed the maximum permissible value
Rext! See "Measuring output" in section
"4.Technical data" for maximum values
of Rext!
US
(U2)
... the measurement output cables should
be twisted pairs and run as far as possible
away from heavy current cables !
U G = Measuring input, generator voltage
Fig.3. Rear at device
U S = Measuring input, bus voltage
= Measuring output
Fig. 6
= Power supply
4
(1) Screw hole brackets
(2) Latch for pulling the screw
hole brackets out
In all other respects, observe all local
regulations when selecting the type of electrical
cable and installing them!
(3) Latch for pushing the screw
hole bracket in
(4) Rubber buffer
3
IC 150870 rishabh_c12 26 Aug 09