Today’s heart-rate monitors can do more than measure a pulse. They also can tally the
amount and rate of exercise, as well as the effort expended. They just need additional signal
conditioning. Stored data can be downloaded to a computer via USB or a wireless USB
dongle.
Bad eating habits, fast-paced daily routines, stress, and a lack of exercise all contribute
today’s rising rates of obesity. Consumers have a legitimate and growing concern about
getting healthier more than ever before, and there has been an increasing demand for
products that help track our physical activity, health, and wellness.
Related Articles



Heart-Rate AFE Monitors Remotely
Simple Heart Monitor AFE Multiplies Opportunities For Medical OEMs
Plethysmograph AFEs Follow Trend Toward Application Specialization
Heart-rate monitors, body-worn fitness trackers, and smart scales are just a few of the
options to measure and monitor personal fitness both during exercise and daily activity.
Tracking the amount of energy we burn in a day helps to change eating and exercise habits,
and it gives us a sense of how active we are and how much energy we are burning in our
daily activities. It provides a reference for the number of calories required to drop weight
while maintaining the same level of physical activity.
Activity monitors can track the number of steps and estimate the calories burned throughout
the day. These estimations would be more accurate if these devices could measure heart
rate. Existing products can measure heart rate, but users must touch them with their fingers
(in the case of sports watches) or wear them on the chest, which makes them uncomfortable
and unnatural for the person exercising. These existing solutions are based on reading
electrocardiograms (ECGs or EKGs).
An alternative to this method is the photoplethysmographic (PPG) technique used in existing
pulse-oximeters. These devices can continuously monitor a heart rate in any place where
there are blood vessels such as the wrist.
Reflective And Transmissive Pulse Oximetry
Pulse oximetry is a noninvasive photo-based technique that measures the light absorption
and refection properties of deoxygenated and oxygenated hemoglobin. The amount of light
absorbed in the hemoglobin is defined by the Lambert-Beer Law, which associates the
degree of light absorption with the wavelength of the beam light, the path length, and the
absorption coefficient of the blood constituents (Fig. 1).
1. Tracking the low and high concentration of light absorption provides noninvasive pulse-oximetry feedback.
By illuminating a tissue bed with LEDs and measuring the amount of light absorbed by the
tissue using a light-sensitive photodiode (PD), the concentration of oxygen in the arterial
blood, heart rate, and blood flow can be estimated. The LEDs typically shine visible red (650
nm) and infrared (IR) light (940 nm).
PPG signals can be measured by either transmittance or reflectance mode optical sensors.
Transmittance mode sensors (commonly used on finger clips) measure the amount of light
that passes through tissue by a photo-sensor located on the opposite side of a tissue bed
(such as the finger or earlobes). The reflectance operating mode measures the reflected light
by a photosensor positioned adjacent to the red and infrared LEDs. A pulse-oximeter that
employs a reflectance mode sensor can detect the PPG signals in any part of the human
body where there is a reasonable concentration of blood vessels (Fig. 2). It can be difficult to
measure the PPG signals obtained at the wrist, but other body sites can be used to obtain
the PPG signals.
2. PPG signals can be measured by either transmittance or reflectance mode optical sensors.
System Definition
Activity monitors measure both a person’s amount and rate of exercise by detecting the
subject’s posture (standstill, seated, sleeping, walking, or running) during periods of time.
These gadgets use this information to estimate the amount of energy consumed during the
process. In addition to measuring posture, the new generation of activity monitors will be
able to measure heart rate and therefore accurately determine the effort expended to
perform a specific activity like playing baseball, swimming, or walking.
By developing innovative ways to increase integration while reducing system power
consumption and providing smart connectivity, existing semiconductor components are
making the next generation of activity monitors more flexible, affordable, and accessible.
Driven by consumers’ desire to quickly know their heart rate, portable and battery-operated
fitness devices commonly have goals for extended battery life, high precision, and fast
response times.
Additional requirements may drive the need for more memory to allow for data logging and
wired or wireless interfaces for data exchange or access to the sensor data. Visual feedback
for displaying the heart rate, calories burned, or more complex step-by-step utilization
instructions may be required as well. Adding all these features in a stylish and small form
factor without increasing system power consumption is a significant challenge.
A PPG-based heart-rate monitor comprises (Fig. 3):
3. The elements of a PPG-based heart-rate monitor include sensors, the AFE, the ultra-low-power microcontroller,
and wireless modules.
• PPG sensors
• An analog front end (AFE) for detecting the reflective PPG signal
• An ultra-low-power microcontroller (MCU) for calculating the heart rate
• A wireless module based on Bluetooth Low Energy (BLE) for exchanging information with
smart phones and tablets
• A motion sensor for monitoring the user’s activity
• A photosensor and LED board
• Ferroelectric RAM (FRAM) for data logging
• A lithium-polymer rechargeable battery
• A battery charger
• A battery fuel gauge
• PPG sensors and an AFE
The key components (Fig. 4) required for acquiring and signal-conditioning the PPG signals
are the LED, photodetector, and AFE.
4. The key components required for acquiring and signal-conditioning the PPG signals include the LED,
photodetector, and AFE.
This example circuit for driving the LEDs is a classic H-bridge driver (Fig. 5). It multiplexes
the ON times between the IR and red LED by applying pulse-width modulation (PWM)
signals to the base of the bipolar junction transistors (BJTs). LEDs consume the most power
(currents of 4 to 20 mA).
5. The LED driver provides light for the photodetector in a basic PPG heart-rate monitor design.
Most schemes use 25% nominal duty cycle for each LED, and the duty cycle can be scaled
down to 1% to 5% to save power. How low the duty cycle can be scaled depends on the
strength of the signal received at the PD, the speed of the acquisition system, and the input
referred noise of the photodiode. A good photodiode with low input referred noise helps save
power and extends battery life.
The photodiode requires a signal conditioning circuit (Fig. 6). It is based on the classic
resistor-feedback transimpedance amplifier (TIA) and capacitor-feedback switched
integrators suitable for pulse-oximetry applications
6. The photodiode requires a signal conditioning and amplifier circuit.
Some commercially available AFEs integrate both the LED driver circuitry and the
photodiode signal conditioning circuitry in a single package (Fig. 7). This new generation of
AFEs can drive the LED currents in using an H-bridge configuration capable of driving up to
150 mA/leg, with short-circuit protection. They can also increase the dynamic range greater
than 105 dB and create a current reference independent of the IR and red LEDs.
7. Commercially available AFEs like TI’s AFE4400 integrate the LED driver circuitry and the photodiode signal
conditioning circuitry in a single package.
The photodiode circuitry embedded into these devices can amplify currents below 1 µA with
13 bits of resolution. It is ultra-low-power (<4 mW) and has a programmable TIA. The AFE
consumes less than 3 mA of current when active.
Microcontroller
In this design example, the microcontroller is used to calculate the heart rate, merge the
motion sensor data, and process the AFE information. The microcontroller should have
specific features including the ability to maintain the context at all times. It should also have a
limited power budget because it will be continuously running and nobody wants to drain the
batteries. The average power consumption of this device should be limited to less than 10
mW. System developers should consider the following features when evaluating MCUs for
this application:
• Low current consumption when the CPU is active: The current consumption in active mode
should be under 100 µA/MHz.
• Efficient and intelligent low-power modes: The MCU should have multiple low-power
schemes that can disable specific internal modules, memory sections, and clock sources to
save energy.
• Smart peripherals: These devices should be able to operate and process information while
the MCU is in any of the low-power modes.
• Fast wakeup times: The MCU should be able to recover from low-power modes in less than
5 µs with the goal of staying in low-power modes for longer periods of time. It should also be
able to enter into low-power modes in less than 1 µs.
• Low current consumption when CPU is in standby mode: Current consumption in standby
mode should be less than 2 µA.
• Low-power memory: The MCU should have non-volatile RAM technology with fast-write
and fast-read memory technology with ultra-low-power consumption when data received
from the AFE and the motion sensors is read/written. FRAM technology fulfills all of these
requirements. As a reference for this example, the ultra-low-power MSP430 microcontroller
from Texas Instruments fulfills these requirements while providing best-in class performance
per microwatt.
Motion Sensors
Sensors are a fundamental part of the human machine interface (HMI). They help the system
identify the context and environmental conditions. Motion sensors such as accelerometers,
gyroscopes, and magnetometers help identify whether a person is seated, walking, or
running. They are key elements to identify the orientation of the arm, wrist, or other specific
part of the body where the activity monitor is located. They also help to track the travel
distances and provide a more accurate position of the system by increasing the resolution of
the GPS with dead-reckoning algorithms.
The key requirements for selecting a sensor are power consumption (should be less than 0.6
mW), resolution (higher than 12 bits/axis), and dynamic range and size (less than 2 mm2).
Figure 8 shows one of the nine-axis combo sensors used in the design example. The active
current consumption of the motion sensor is 3.5 mA.
8. Motion sensor circuits are fundamental for the design of a human machine interface (HMI) for a heart-rate
monitor.
Communication Link
The system described in this article has both wireless and wired communication links. The
wireless communication link is based on BLE and is based on the BR-LE4.0-S2A, an FCCcertified (Federal Communications Commission) system-in-PCB (printed-circuit board)
module available online that only requires a few external components.
This module works with AT-based commands and is easy to use since it includes a network
processor that handles all the transactions required by the Bluetooth 4.0 stack. The wired
communication is based on USB 2.0. The microcontroller’s built-in module requires only a
few external components. USB is also used for charging the lithium-polymer battery.
Battery Charger And Fuel Gauge
The battery charger operates from either a USB port or ac adapter and supports charge
currents up to 1.5 A. The input voltage range with input overvoltage protection supports
unregulated adapters. The USB input current limit accuracy and startup sequence allow the
battery charger to meet the USB-IF inrush current specification. Additionally, the input
dynamic power management prevents the charger from crashing incorrectly configured USB
sources.
The battery fuel gauge circuits an easy-to-configure microcontroller peripheral that provides
system-side fuel gauging for single-cell lithium-ion batteries. The device requires minimal
user configuration and system microcontroller firmware development. The battery fuel gauge
uses the impedance track algorithm for fuel gauging and provides information such as
remaining battery capacity (mAh), state-of-charge (%), and battery voltage (mV). Battery fuel
gauging requires connections only to PACK+ (P+) and PACK– (P–) for a removable battery
pack or embedded battery circuit.
Power Consumption
The system can run up to six hours with BLE radio enabled transmitting 300 bytes/s
continuously and up to 10 hours continuously measuring the heart rate with the BLE radio
disabled. Table 1 shows the current consumption and power consumption of the overall
system when the BLE radio is disabled. Enabling the BLE radio does increase power
consumption significantly as expected (Table 2).
Conclusion
The PPG-based heart rate monitor described in this article is a low-power, low-cost, and
flexible system that provides a high-performance alternative for designers creating the next
generation of activity monitors. The AFE produces a high-quality PPG signal that helps to
minimize the complexity of the algorithms running on the MCU.
By running lean and efficient algorithms, the MCU is in low-power modes for longer periods
of time and therefore reduces the average current consumption. This helps to maintain the
system average current consumption below to 12 mA when the BLE radio is disabled and
less than 32 mA when the radio is enabled and transmitting the heart rate once a second.
Daniel Torres is an applications engineering manager at Texas Instruments (TI) and has
more than 10 years of experience in electronic systems. He has published several
application notes and papers describing multiple applications for microcontrollers. His
professional interests are MEMS systems and sensors, ultra-low-power systems, motor
control, power electronics, analog systems, and digital signal processing.