Clarification No.3
Publication Ref: EuropeAid/138127/ID/SUP/TR
Subject: Supply of Equipment for Improvement of the Shared Industrial Infrastructure of the
Pistachio Processing Sector
Location – Siirt, Turkey
Further to the requests received from the tenderers, the following clarifications are provided:
ANNEX II + III : TECHNICAL SPECIFICATIONS + TECHNICAL OFFER
LOT 1
QUESTIONS
ANSWERS
For Article 1.1.4.4, the maximum operating
Q1 1.1 HPLC WITH DEDECTOR
1.1.4.4 Maximum operating pressure: at least 400 pressure of the quaternary pump shall not
take into consideration the psi value with
bar.
eluent generation.
The current specification is specific to the one
manufacturer (Agilent) although there are also other Please also note that 400 bar is a widely used
items which is originated in EU.
operating pressure for similar applications,
On the other hand, The desired device is not a UPLC therefore the related specification remains
system. In traditional HPLC applications are not rise unchanged.
to such high pressure. Also, in such high pressure the
normal HPLC colons make leak and the requested Please also refer to Question-5 under
colon is a normal in technical specification. Even 345 Clarification No.1
bar will be a very high pressure. So, to be
more competitive tender we kindly request you to
change this specification as below.
1.1.4.4 Maximum operating pressure: at least 5000
psi (345 bar).
Q2
1.1 HPLC WITH DEDECTOR
1.1.5 Autosampler Specifications
Question: Is it obligation for obtaining Autosampler
device at the same brand with HPLC device, since
most of the manufacturers supply autosampler with a
different brand?, can we offer different brand for
autosampler which meets with requirement with
HPLC device?
Autosampler can be provided from different
brand satisfying the minimum technical
requirements and compatible with the offered
HPLC device.
1/8
Q3
1.1.6.2.: Contains Peltier-controlled heat exchangers
for effective cooling and heating.
For rnycotoxin and aflatoxin analysis ; Column
temperature should be approximately 40 °C . So
heating is important for " Column Temperature
Control Module "
Cooling option is specific feature for some
companies.
The application of the lab shall not be limited
to one specific analysis. The technical
specification does not indicate any
temperature range for cooling. It requires
that the item shall have peltier unit operation,
not just heat-only units.
Specification remains unchanged.
We kindly ask you to let change this specification as
"Contains Peltier-controlled heat exchangers for
effective heating"
Q4
1.6.9.: "Interpupillary distance adjustment range (a-b):
a: min 30mm-max 90mm"
According to the above specification requested range
is given with min and max values. In order to prevent
any misuderstanding, could you please inform us we
can offer an "adjustment range" for requested
microscope between 30mm-90mm? (As an example:
40-80mm, 35-65mm etc..)
Q5
The lower limit of the interpupillary distance
shall be 30mm or higher, whereas the highest
distance shall be 90mm or lower. Any values
between these ranges (a-b) is acceptable.
With this regard, the provided examples in
the question also satisfies the minimum
technical requirements.
1.8 ELISA READER WITH TEST KIT
Specification remains unchanged.
1.8.2 Filter Capacity and number of filters: min 4
Please also see the Question-15 under
Clarification No.1
Wavelength range min. 400-700 nm or wider
range
The new systems are not required of filter and this is
better than requested. So, we kindly request you to
change this specification as below.
1.8.2 Filter Capacity and number of filters: min
4 or the device doesn’t need any filter
Wavelength range min. 400-700 nm or wider
range
Q6
1.8 ELISA READER WITH TEST KIT
1.8.11 “PC Operated. Supplied with related
connections and software” :
Question: Is it required to supply with PC or not?
Does Related software includes data analyzing?
The software shall be the standard bundled
software for the operation of the elisa reader.
The technical specification indicates that the
item shall be PC operated but a separate PC
shall not be supplied.
2/8
Q7
1.8 ELISA READER WITH TEST KIT
1.8.12 Lamp: Halogen Lamp
Halogen Lamp is an old technology and this lamp has
2.000 hours life. Xenon flash lamp is a new
technology and better than halogen lamp. Xenon flash
lamp has endless life. The new system use xenon flash
lamp. To be more competitive tender we kindly
request you to change this specification as below.
Please see the Corrigendum No.4 to the
Tender Dossier.
1.8.12 Lamp: Halogen or Xenon Flash Lamp
Q8
1.8 ELISA READER WITH TEST KIT
1.8.12 “Lamp: Halogen Lamp”:
Please see Answer-7.
Question: "Is halogen lamp a must? Our instrument
has a LED lamp which is much more advanced than
halogen lamp. Is this acceptable?"
Q9
A single box composed of 100 tests is
1.8.14 “Supplied with 100 units of Alfatoxin ELISA enough.
Test Kits along with pipettes with standard 96
The pipettes can be multichannel compatible
wells/kit with ≤0.05ppb sensitivity and high
with standard 96 wells application. There is
recovery (>80%) and suitable for aflatoxin analysis
no specific requirement or obligation with
in pistachio and peanuts.” :
number of channels for a multichannel
1.8 ELISA READER WITH TEST KIT
Question: Is it required in ‘units’ or ‘tests’ since one
unit has a high cost, but 100 tests are included in only
one box (as kit).
pipette system.
On the other hand, what is the channel number for
pipettes?
Q10 1.8 ELISA READER WITH TEST KIT
1.8.13 Rapid Aflatoxin Test Kit (B1-B2-G1-G2)
1.8.14 Supplied with 100 units of Alfatoxin ELISA
Test Kits along with pipettes with standard 96
wells/kit with ≤0.05ppb sensitivity and high recovery
(>80%) and suitable for aflatoxin analysis in pistachio
and peanuts.
Question 1:
Is Origin obligatory for those Test Kits in this item
also?
Items 1.8.13 and 1.8.14 are required for
initial operational requirements of the unit
and the ELISA system. Tenderers are
required to supply such consumables
separately (which can be of different brand)
for this item and since these articles are not
indicated as separate items, the “Technical
Documents to be provided during the
implementation period” as mentioned in
Article 2 of the General Requirements does
not apply for items 1.8.13 and 1.8.14, but
only for the ELISA Reader machine itself.
Please also see Question-16 under
Clarification No.1
3/8
Q11 1.13 AUTOCLAVE
Item 1.13.8 “Automatic control for heating,
sterilization, water filling.” :
This item restricts the potential tenderers to prepare
satisfied proposal for this tender. To provide more
attendance, to provide best competition and to
eliminate cancellation possibility of this tender, we
kindly request the contracting authority to change this
item in the technical specifications to be read as
follows:
“Automatic or manuel control for heating,
sterilization, water filling.”
Q12 1.15 FREEZER
1.15.2 Freezing capacity range: -10 ºC to -25 ºC or
wider range
The freezing capacity of laboratory freezer generally
starting from -20 ºC . Providing a lower temperature is
more useful. We kindly request you to change this
specification as below
Please see the Corrigendum No.4 to the
Tender Dossier.
The requested “-20 ºC to -45 ºC or wider
range” already meets the minimum
requirement of “-10 ºC to -25 ºC or wider
range”, therefore the specification remains
unchanged.
1.15.2 Freezing capacity range: -20 ºC to -45 ºC or
wider range
Q13 1.15.7. Shelves: PVC laminated or aluminum shelves.
Shall have minimum 4 baskets or isolated
compartments.
Please see the Corrigendum No.4 to the
Tender Dossier.
In general, for laboratory usage stainless steel material
is one of the best and most efficient solution for both
interior and exterior of the coolers. For item 1.15
stainless steel interior has been already requested but
shelves are requested to be PVC laminated, Then,
aluminum material was also added to TS. We strongly
recommend the contracting authority to also add
"stainless steel shelves" to above mentioned
specification in terms of durability and efficiency. It
will also help to enable competition.
4/8
Q14 1.16.1.: Container Capacity: min 15 liters
Could you please clarify that the mentioned container
will be used as a feeder for the grinder or will be used
to collect the grinded materials after the grinding
process. If it will be used after the grinding process,
could you please inform us whether we can offer the
container seperately?
The technical specification does not indicate
the container type, since solutions may be
different according to the manufacturers. For
some systems the feeding container can be a
separate unit, whereas some items use
integrated containers.
It is requested that the grinder shall be
capable of operating with min 15 liter
capacity (pistachio and similar material
qualification) either with a separate container
and motor unit, or an integrated container and
motor.
With this regard a separate container can also
be offered.
Q15
1.16 GRINDER
1.16.1 “Container Capacity: min 15 liters” :
1.16.4 “Blade Speed: min 6500 rpm” :
It is declared from many manufacturers that container
capacity and blade speed are inversely proportional
since one decreases by the other increases. It is said
that the grinder having min 6500 rpm can not be with
a min 15 liter container capacity. So, is it min 1,5
liters and min 6500 rpm actually?
The required minimum blade speed does not
indicate a criterion to be achieved with a
fully loaded container capacity. The blade
speed refers to the performance of the
standard motor/blade speed.
Specification remains unchanged.
On the other hand, to provide more attendance, we
kindly request the contracting authority to change the
item below also to be read as follows :
"Blade Speed: Motor Power: 0,6 – 0,9 kW 750 / 1500
rpm"
Q16 1.17 LAB BLENDER:
In the market, there is only one or two brands
producing laboratory type blenders. Unfortunately,
these blenders are not manufactured in an eligible
country. Therefore, we kindly request contracting
authority to give derogation for this item or remove
this item from item list.
Please see the Corrigendum No.4 to the
Tender Dossier.
5/8
Q17 1.17 LAB BLENDER
Specification remains unchanged.
1.17.4 “Max attainable RPM: min 10000 rpm –
variable speed option.”:
To provide more attendance, to provide best
competition and to eliminate cancellation possibility
of this tender, we kindly request the contracting
authority to change this item in the technical
specifications to be read as follows:
"Max attainable RPM: Motor Power: 300-500 W –
9000 rpm"
Q18 1,19.14.: "Display features: HEPA and UV life span,
temperature, air flow and speed."
According to our market research, instead of life span
for UV and HEPA, there are counters showing the
elapsed time. This might also be more efficient
solution for HEPA and UV. Therefore, we kindly
request contracting authority to change this
specification as;
The life span means that the cabinet shall
provide the user proper information
concerning the need for a replacement of the
related filters or monitoring the filter status.
This can be achieved with different
mechanisms or displays.
Specification remains unchanged.
"Display features: "HEPA and UV counter;
temperature, air flow and speed" or "HEPA and UV
life span or counter; temperature, air flow and speed"
Q19 Although there are items satisfy not only rule of origin
but also technical specifications, it is imposible to
submit the Certificate of origin due to the they are
cheap and number of items is low. We kindly request
you not to request the below items as seperate items.
For Items 1.21 Ph-Meter
Specification remains unchanged.
Q20 1.21.2.: Electrode: shall have temperature and pH
electrodes.
Some manufacturers in the market are able to
combine the measuring of PH and the temperature in a
single electrode. This might be more practical than
requested. Therefore, we kindly request the
contracting authority to modify this specification as;
"Electrode: shall have temperature and p1-1
electrodes or PH electrode with temperature sensor."
Q21 1.21.6.: "Temperature sensitivity: maximum 0.1 °C"
In order to prevent misunderstanding and confusion,
could you please clarify the mentioned "sensivity"
means/indicates "resolution value"?
Technical specification is indicated as
“electrode”, therefore a combined solution
mentioned in the question is also acceptable.
It does not indicate specifically single or
multi electrode solutions.
Specification remains unchanged.
Some manufacturers use different
terminology. Sensitivity also means
resolution.
6/8
ANNEX II + III : TECHNICAL SPECIFICATIONS + TECHNICAL OFFER
LOT 2
QUESTIONS
ANSWERS
Please see the Corrigendum No.4 to the
Q22
2.2 PROJECTOR WITH SCREEN
2.2.7 Standards: Projector shall be compliant with Tender Dossier.
EMC Directive 2004/108/EC, Low Voltage
Directive 2006/95/EC, ENERGY STAR®
No projector does not have Energy Star Standard.
Please find devices with this certification in this
link https://www.energystar.gov/products. To be
competitive tender, we kindly request you to change
this specification as below.
2.2.7 Standards: Projector shall be compliant
with EMC Directive 2004/108/EC, Low Voltage
Directive 2006/95/EC. The device should have
energy saving feature.
Q23
2.2 PROJECTOR WITH SCREEN
2.2.7 Standards: Standards: Projector shall be
compliant with EMC Directive 2004/108/EC, Low
Voltage Directive 2006/95/EC, ENERGY STAR®
Please see Answer-22.
After our wide market research we have reached to
the conclusion that there is no projection device
having ENERGY STAR® certification. Therefore
we kindly request you to remove ENERGY
STAR® from this article.
ANNEX II + III : TECHNICAL SPECIFICATIONS + TECHNICAL OFFER
LOT 3
QUESTIONS
ANSWERS
Item 3,1 Office Desk: On lline 3,1,10 metal
Please see the Corrigendum No.1 to the
Q24
Tender Dossier. The related specification was
components are requested as cromium plated.
already changed. Please also see Question 41
Could you please kindly clarify if electrostatic
under Clarification No.1.
powder paint after anti-corrosion surface treatment
is also acceptible.
Q25
Item 3,3 Meeting Table: On lline 3,3,4 metal
components are requested as cromium plated.
Could you please kindly clarify if electrostatic
powder paint after anti-corrosion surface treatment
is also acceptible.
Please see the Corrigendum No.1 to the
Tender Dossier. The related specification was
already changed. Please also see Question 42
under Clarification No.1.
7/8
Q26
Item 3,12 Double Visitor Seat: On line 3,12,4
seating width is requested as 55-65 cm. Industry
standard is 45-50 cm. Could you please kindly
revise seating width dimension.
Please see the Corrigendum No.1 to the
Tender Dossier. The related specification was
already changed. Please also see Question 43
under Clarification No.1.