SECTION 7—TROUBLESHOOTING GUIDE SECTION 7—TROUBLESHOOTING GUIDE SYMPTOM: Alarm: Short beeps, long pause Concentrator not operating, power switch On. Beep......Beep...... Alarm: Continuous PROBABLE CAUSE: Main Power Loss: 1. Power cord not plugged in. 1. Insert plug into outlet. 2. No power at outlet. 2. Inspect house circuit breakers or fuses. If problem recurs, use a different outlet. 3. Tripped circuit breaker. 3. Call Service Provider. System Failure: 1. Unit overheating due to blocked air intake. 1a. Remove and clean cabinet filters. 2. Insufficient power at outlet. 2. DO NOT use extension cords. Move to another electrical outlet or circuit. 3. Internal repairs required. 3. Call Service Provider. Low Flow Alarm: 1. Kinked or blocked tubing, cannula or humidifier. 1. Inspect for kinks or blockages. Correct, clean or replace item. Once corrected, turn power Off for 60 seconds and then turn power back ON. 2. Flowmeter set at 0.5 L/min or less. 2. Reset flowmeter to prescribed flowrate. If prescribed flowrate is below 0.5 L/min, use the Pediatric Flowmeter IRCPF16 (Only with Perfecto2 5). Concentrator not operating, Power Switch On. Beeeeeeeep.... Alarm: Rapid Beep....Beep... Beep...Beep Perfecto2™Series SOLUTION: 26 1b. Move oxygen concentrator at least three inches away from walls, draperies or furniture. Part No 1143482 SECTION 7—TROUBLESHOOTING GUIDE SYMPTOM: YELLOW or RED light Illuminated. Alarm: Continuous On RED light only. * Only applies to IRC5PO2. PROBABLE CAUSE: SOLUTION: 1. Low oxygen purity.* 1. Clean or Replace filters. 2. Kinked or blocked tubing, cannula or humidifier.* 2. Inspect for kinks or blockages. Correct, clean or replace item. Once corrected, turn power Off for 60 seconds and then turn power back On. 3. Flowmeter set at 0.5 L/min or less.* 3. Check flowmeter is set to 1.0 L/min or more. 4. Unit overheating due to blocked air intake. 4a. Remove and clean cabinet filters. 4b. Move concentrator at least three inches from walls, draperies, and furniture. GREEN light with YELLOW Light Flashing. Part No 1143482 5. Insufficient power at outlet. 5. DO NOT use extension cords. Move to another electrical outlet or circuit. 6. Internal repairs required. 6. Call Service Provider. 1. Internal repairs required. 1. Call Service Provider. 27 Perfecto2™Series