Download Advances in Environmental Biology Systems

Survey
yes no Was this document useful for you?
   Thank you for your participation!

* Your assessment is very important for improving the work of artificial intelligence, which forms the content of this project

Document related concepts

Medical imaging wikipedia , lookup

Positron emission tomography wikipedia , lookup

Transcript
Advances in Environmental Biology, 8(11) Special 2014, Pages: 1031-1042
AENSI Journals
Advances in Environmental Biology
ISSN-1995-0756
EISSN-1998-1066
Journal home page: http://www.aensiweb.com/AEB/
Different Energy Mapping Methods in CT-Based Attenuation Correction in PET/CT
Systems
Elham Soleimani, Ali Soleymani, Shahla Ahmadi, Majid Pouladian
Islamic Azad University, Science and Research Branch, Tehran, Iran.
ARTICLE INFO
Article history:
Received 15 June 2014
Received in revised form
8 July 2014
Accepted 4 September 2014
Available online 20 September 2014
Keywords:
Energy Mapping Methods, CT-Based
Attenuation, CT-Based Attenuation
ABSTRACT
This paper presents the result of assessing different attenuation correction methods for
PET data according to CT data (CTAC). These methods are intended for use with a
combined PET/CT scanner. We discuss five possible methods of energy mapping from
the CT energies to the required 511keV. CT images are obtained using a computerized
whole body phantom, 4DXCAT, which is simulation of physically and anatomically of
human body. Materials and methods: The aim of this study is to compare different
energy-mapping techniques: scaling, segmentation, hybrid, bilinear calibration curve
and dual energy approach through attenuation map generated from CT data through
XCAT phantom. The CT images acquired from XCAT phantom is applied to generate
µ-maps in 511keV.Then these generated µ-maps are compared to the image acquired
from XCAT phantom in 511keV as the gold standard image. For comparing methods
we use three ways: Assessing different ROIs, correlation coefficients and difference
images for comparing pixel by pixel. Results and Discussion: Nearly all energymapping methods shown similar results in soft tissues. A noticeable relative difference
is seen in lung tissues in Segmentation method which refers to the variability in
densities. Also a bias in bone in the same method, which is due to the extended borders
of the segments. In Scaling results for different tissues are acceptable beside bone as it
has a high photoelectric ratio. Hybrid and Bilinear are somehow good. Dual Energy
reports the best results.
© 2014 AENSI Publisher All rights reserved.
To Cite This Article: Elham Soleimani, Ali Soleymani, Shahla Ahmadi, Majid Pouladian, Different Energy Mapping Methods in CT-Based
Attenuation Correction in PET/CT Systems. Adv. Environ. Biol., 8(11), 1031-1042 2014
INTRODUCTION
Today, diagnostic imaging is consisting of various branches. The information of imaging is divided into
anatomical and function groups. In some systems as CT, MRI imaging based on anatomical techniques, the
tumor is detected when the tissue is physiologically changed but in some systems including PET,SPECT
imaging based on physiological techniques, metabolic changes are detected in the early stage. It is obvious that
there are many instances in which it would be desirable to integrate the information obtained from two
modalities of the same patient. The combination of PET-CT was introduced in the early 1990s by Townsend.
The first prototype of PET-CT scanner was made in 1998 [1-9].
In PET/CT systems, based on the data, CT is applied for attenuation correction of PET data. In CTAC, at
first the patient undergoes CT scan and then at the same position, PET scan is taken of the patient [10, 11].
In converting CT attenuation to be used as data in PET image, there are two problems. The first problem is
the high difference between the energy of CT, PET photons. As attenuation coefficient depends on energy and
the main issue in attenuation correction by CT is conversion of the coefficients of CT energy, the energy at120
kVp to PET energy, energy at 511keV and the second problem is the difference between single energy spectrum
in PET and extensive energy spectrum in CT.
In PET imaging, annihilation photons at 511kev are applied while CT diffuses the photons with extensive
energy spectrum 40 to140 kev. This photoelectric overcome in CT energies range and Compton overcome at
511keve create error in conversion of the images of CT energies to PET energy.
There are various methods for energy mapping to convert attenuation coefficients in CT energy to PET
energy. These methods are including scaling, segmentation, hybrid (scaling+ segmentation) [12], bilinear [9, 13]
and dual energy approach [14, 15].
Corresponding Author: Elham Soleimani, Islamic Azad University, Science and Research Branch, Tehran, Iran.
E-mail: [email protected]
1032
Elham Soleimani et al, 2014
Advances in Environmental Biology, 8(11) Special 2014, Pages: 1031-1042
Although there are some studies performed regarding the comparison of energy mapping methods by
various groups [160 29], no comprehensive study is performed in this regard as all the methods are done with
the phantom physiologically and anatomically close to body anatomy. The performed studies are done based
on2 or 3 methods or physical phantoms are used in them with inevitable errors. The current study aimed to
study all the existing methods of energy mapping to obtain attenuation map at 511keV of CT images in
combinational PET/CT systems by XCAT human computer phantom [30,31] and the advantages and
disadvantages of these methods are reviewed.
1- Materials and methods:
This study was conducted in three stages including CT image simulation as basic image, PET image
simulation as reference image and the final stage applying energy mapping methods on CT image.
1-2 Simulation of CT image as basic image
To test the various methods of energy mapping on CT images, at first by XCAT-4D phantom, CT images
are built. To obtain CT images in XCAT phantom, at first the required energy as 74keV and effective energy at
120kVp in this paper were entered in the files and after creating phantom, the image is observed by Amide
software. It can be said that the obtained image of XCAT phantom is based on attenuation coefficient. Thus, to
use it in CT-based studies, at first the coefficients should be converted to CT numbers in accordance with
equation (1) and it is done in MATLAB software.
(1)
After the conversion of attenuation coefficients to CT numbers, the image is read by Amide software. CT
image of whole body XCAT phantom at 74keV equal to 120kVp is shown in Figure 1. CT image is acquired of
phantom 128×128, resolution 0.3215 and its output is as linear attenuation coefficients (cm-1) in Raw data
format.
Fig. 1: CT image of whole body XCAT phantom from sagittal and coronal view
2-2 Simulation of PET image as reference image:
The reference image is obtained by putting the value 511keV for energy in the required section. This image
is applied to compare the attenuation maps of various methods of mapping.
2-3 Applying mapping methods on CT image
After various energy mapping methods on CT image, the simulation is applied. By entering the image based
on CT numbers in MATLAB software, the changes of each method are applied on the image and then the image
is observed in Amide software.
2-3-1 Scaling methods
The scaling approach applies conversion coefficient equal to the ratio of attenuation coefficients of water at
CT energy to water attenuation coefficient at PET energy.
Conversion coefficient in scaling was 1.94:
Scaling factor= µwater (CT) / µwater (PET)
(2)
1033
Elham Soleimani et al, 2014
Advances in Environmental Biology, 8(11) Special 2014, Pages: 1031-1042
Scaling factor=1.94
The conversion coefficient was applied by MATLAB software on CT image and attenuation map is
generated from PET energy. The image of applying scaling methods on CT image is shown in Figure 2.
Fig. 2: The attenuation map generated from PET energy after applying scaling method on CT image.
In s scaling method, an effective energy is selected to display CT spectrum ranging 50 to 80 keV.
2-3-2 Segmentation method:
This method is consisting of two stages. In the first stage, the reconstructed CT image is segmented into
different tissue types. In the second stage, the linear attenuation coefficient value for each tissue type is then
replaced with appropriate attenuation coefficients at 511 keV.
Typical choices for tissue types are soft tissue, bone, and lung. This segmentation can be done based on CT
numbers as for 800<HU<0- lung tissue, for 0
for soft tissue and 300≤HU bone and for HU≤800
for air are considered.
Fig. 3: The attenuation map generated from PET energy after applying segmentation method on CT image.
2-3-3 Hybrid (segmentation-scaling) method:
The hybrid method is the combination of segmentation and scaling method. In this method for most
materials except bone, the ratio of the attenuation coefficient for all tissues in two energies is essentially
constant. Thus, at first a threshold is determined to remove the bones from CT image.
In this study, the number 300CT is considered as a threshold because CT of bone tissues is above 300, then
due to the differences of physical attributes of bone and soft tissues for each part, a different conversion
coefficient is applied. For non-bone tissues, soft tissues, the conversion coefficient is the ratio of linear
attenuation coefficient of water at CT,PET energies, the same coefficient used in scaling and for the bones, the
conversion coefficient is the ratio of linear attenuation coefficient of cortical bone at PET, CT energies. The
linear attenuation coefficient of the bone at both energies is achieved in accordance with XCOM section [32]:
1034
Elham Soleimani et al, 2014
Advances in Environmental Biology, 8(11) Special 2014, Pages: 1031-1042
3) µbone(CT)= 0.4934 cm-1
4) µbone(PET)= 0.1715 cm-1
Thus, conversion coefficient for bone tissues is equal to:
5) )Bone scaling factor = µbone (CT) / µbone (PET)
Bone scaling factor=2.87
Now by having conversion coefficients for soft and bone tissues, in CT image of each of the tissues are
multiplied by the coefficient to obtain the linear conversion coefficient of each tissue at PET energy [26].
Fig. 4: Mapping image of hybrid method
2-3-4 bilinear method:
if the unknown matter X as the combination of water and A matter is imaged in CT imaging, , by equation
(6), and CT values of X,A matter, we can calculate linear attenuation coefficient of X at E energy.
µx(E)= px×µx(E)
In the above equation Px, µx(E) are density and mass attenuation coefficient of X, respectively. To convert
CT values to linear attenuation coefficients by equation (6), two linear equations are obtained.
As the body tissues are divided into two groups, in the first group the tissues their CT number is less than or
equal to zero (HUX≤0), they are the combinational of water and air (A=Air) and in the second group, the tissues
their CT is more than zero (HUX>0) and they can be assumed as the combination of water and bone. (A=
Cortical bone).
To calculate the required parameters in equation (6), the practical measurements are required. Table (1)
shows the equations of bilinear curves generated in 4 voltages of CT scanner.
Table 1: The equations of bilinear curves generated in 4 voltages of CT scanner
If HU≤0: LAC_511=96×10.6 HU+0.096
If HU>0: LAC_511=54×10.6 HU+0.096
If HU≤0: LAC_511=96×10.6 HU+0.096
If HU>0: LAC_511=63.54×10.6 HU+0.096
If HU≤0: LAC_511=96×10.6 HU+0.096
If HU>0: LAC_511=73.87×10.6 HU+0.096
If HU≤0: LAC_511=96×10.6 HU+0.096
If HU>0: LAC_511=81.44×10.6 HU+0.096
80 kVp
100 kVp
120 kVp
140 kVp
Based on calibration curve in various energies of figure (5), we can obtain for each CT image, linear
attenuation coefficient of one by one of its pixels.
1035
Elham Soleimani et al, 2014
Advances in Environmental Biology, 8(11) Special 2014, Pages: 1031-1042
Fig. 5: The calibration curve at different energies
In this study, calibration curves at 120kVp were applied by MATLAB software to CT image to acquire
linear attenuation coefficient of each pixel at 511 keV. Figure (6) illustrates bilinear map attenuation method.
Fig. 6: bilinear map attenuation method
2-3-5 Imaging method with dual energy:
In this method the CT image at two different energies is acquired. Attenuation coefficient is as the weight
sum of extracting photoelectric and Compton contributions. The different contributions (photoelectric and
Compton) can then be scaled separately in energy, then are converted separated in each energy and then are
added to acquire total attenuation.
Dual energy is an accurate method because both photoelectric and Compton components are converted
separately. Compton is decreased linearly with energy but photoelectric is combined as E-3 and finally with
each other and total is obtained at the required energy.
The attenuation map generated by dual energy method is shown in Figure 7.
Fig. 7: The attenuation map generated by dual energy method
1036
Elham Soleimani et al, 2014
Advances in Environmental Biology, 8(11) Special 2014, Pages: 1031-1042
2-4 The strategy of evaluation of various methods of energy mapping:
In this study, the comparison of the attenuation maps of various energy mapping methods and simulated
reference image is made by ROI plotting, designing correlation curves and differential image method to acquire
the error of each method to the reference.
2-4-1 The comparison via the regions of interest:
In ROI analysis comparison, at first some ROIs are defined on reference image and images generated of
various methods of energy mapping. Then, the linear attenuation coefficient (LAC) is determined for the tissue
and LAC values are compared in the interest regions in reference image and the images of 5 energy mapping
methods.
In order to locate the ROI (Regions of Interest) exactly in the similar coordinate, in the reference image of
Amide, ROIs are selected, then attenuation maps are opened in the same page. Thus, ROI coordinate selected in
reference image and the images of 5 methods are along with each other. An example of ROI in generated
attenuation maps is shown in Figure 8.
Fig. 8: The attenuation map generated from CT image of phantom with some of the regions of interest in
coronal and sagittal views
2-4-2 The comparison of correlation curve methods:
By the descriptive statistical methods, we can express the features of a set of data. Correlation coefficient is
a mathematical index being applied about the two or multi-variate distributions. Dispersion diagram is applied
to represent the correlation between two variables. To review the error of each energy mapping by the data
generated of the attenuation coefficients in various tissues, correlation curve is plotted. The diversion value of
y=x line acquires the correlation and error range.
2-4-3 The comparison of differential image method:
In this method, the existing feature of amide software is applied and the reference images generated of
XCAT phantom of the image of each method are subtracted.
3-Results:
3-1 The results of the comparison via the regions of interest:
The linear attenuation coefficients generated of various energy mapping methods are compared with the
reference image.
The percent of relative difference between LAC in ROI of attenuation maps with LAC value in ROI of
reference image as a criterion to compare attenuation maps were calculated. The less the value of error, it means
that the required method is suitable for that tissue. Table 3 illustrates the error percent of each of energy
mapping methods.
Table 2: The linear attenuation coefficients acquired of various energy mapping methods compared to reference state
Dual energy
Scaling
bilinear
Hybrid
Segmentation
Standard
0.027
0.027
0.025
0.025
0.028
0.026
0.087
0.097
0.087
0.087
0.087
0.092
0.090
0.100
0.090
0.090
0.091
0.096
0.091
0.097
0.087
0.087
0.091
0.097
0.091
0.101
0.090
0.091
0.091
0.096
ROI
Lung
Soft tissue
Brain
Muscle
Kidney
1037
Elham Soleimani et al, 2014
Advances in Environmental Biology, 8(11) Special 2014, Pages: 1031-1042
0.092
0.092
0.117
0.136
0.102
0.102
0.127
0.166
0.092
0.092
0.110
0.125
0.092
0.091
0.115
0.110
Table 3: The error percent of each of energy mapping methods
Dual energy
Scaling
bilinear
Hybrid
3.8
3.8
3.8
3.8
5.4
5.4
5.4
5.4
4.2
4.1
6.2
6.2
8.9
0
10.3
10.3
5.2
5.2
6.2
5.2
5.1
5.1
5.1
5.1
5.1
5.1
5.1
8.9
6.3
15.4
0
4.5
4.6
27.6
3.8
15.3
0.091
0.091
0.101
0.172
Segmentation
7.6
5.4
5.2
8.9
5.2
8.9
8.9
8.1
32.3
0.097
0.097
0.110
0.130
Standard
0.026
0.092
0.096
0.097
0.096
0.097
0.097
0.110
0.130
Heart
Liver
Spines bone
Ribs bone
ROI
Lung
Soft tissue (Water)
Brain
Muscle
Kidney
Heart
Liver
Spines bone
Ribs bone
As is shown in error percent table:
In lung bilinear and hybrid methods had low error, thus for energy mapping, the images of this region can
be suitable. The segmentation method reported considerable error. This is because in some tissues from one
region to another one, the density is changed continually. This value in lung tissue reaches 30%. Thus, any error
in tissues segmentation leads into the error in attenuation coefficient value. Thus, the segmentation method for
lung tissue can not be a good method for lung tissue.
In soft tissue, bilinear, hybrid, segmentation methods had equal error percent, thus the three methods are
applied for energy mapping.
In the kidney, the hybrid and segmentation methods were suitable.
In the heart, hybrid and bilinear methods had low error percent.
In the liver, bilinear had low error.
For the bones, bilinear method had considerable low error. In bone region, the maximum error was
dedicated to scaling because bone has high calcium content and its photoelectric deficit is high.
3-2 The results of the comparison by correlation curves methods:
Figure (9) shows the correlation curves of attenuation coefficients in different tissues, the equation of this
curve and correlation coefficient for scaling, segmentation, hybrid and bilinear and dual energy methods.
a
b
1038
Elham Soleimani et al, 2014
Advances in Environmental Biology, 8(11) Special 2014, Pages: 1031-1042
c
d
e
Fig. 9: Correlation curves of attenuation coefficients in different tissues, the equation of this curve and
correlation coefficient a) scaling, b) segmentation, c) hybrid, d) bilinear, E) dual energy.
3-3 The results of the comparison of differential image method:
In figure 10, the reference image of XCAT phantom, the image of each mapping method and differential
image were shown and by the results we can find about the existing error of each method.
Fig. (1-10): (a) The reference image of energy 511, b. The image of scaling, C. The differential image of scaling
and reference image
1039
Elham Soleimani et al, 2014
Advances in Environmental Biology, 8(11) Special 2014, Pages: 1031-1042
Fig. (2-10): (a) The reference image of energy 511, b. The image of mapping, C. The differential image of
hybrid and reference image
Fig. (3-10): (a) reference (b) bilinear method mapping (c) the differential image
Fig. (4-10): (a) Reference image, (b) mapping image of dual energy (c) Differential image
1040
Elham Soleimani et al, 2014
Advances in Environmental Biology, 8(11) Special 2014, Pages: 1031-1042
4- Discussion and conclusion:
In this study, it was attempted to do the comparison of various energy mapping methods accurately in
which the errors of the patient and physical phantoms don’t exist. To do so, at first by 4DXCAT software
phantom, the reference image (the image at 511 keV) was made and then various energy mapping were applied
on CT image to be compared with reference image. Based on the results of 3 methods being used to evaluate the
image mapping methods, after the comparison of the attenuation maps generated from reference image, it was
defined that in bone region, bilinear and dual energy mapping methods and in muscle tissue, scaling had the best
results.
In other regions as soft tissue, lung and heart, the errors were similar.
The following points about each of mapping methods are considerable.
4-1 scaling mapping method:
The reported results showed that scaling method was suitable for the regions in which Compton has the
dominant effect and in bone tissue, the error percent of scaling method is considerable and this is because the
bone due to calcium content has high photoelectric deficit.
4-2 Segmentation mapping method:
In segmentation method whole-body is segmented into water, air and bone and as in each tissue, the
attenuation tissue is not constant and in most of the tissues the density changes are high in various parts.
In this method, the error for soft tissue, brain and kidney was 5.45, 5.2%, 5.2%, respectively less than other
tissues.
The error in spines bone, heart and kidney was 8.1%, 8.9%, 8.9%, respectively showing that considerable
changes of tissue density are from one region to another.
Thus, in these tissues, considering a value for LAC for a tissue causes error. The more whole body is
segmented to more tissues; the attenuation map is more exact.
This method presents a good attenuation coefficient for the regions with dominant Compton and for the
regions with the dominance of photoelectric, there is no good approximation and the error is high for the tissues
with high atomic number.
4-3 Hybrid mapping method:
At low energies, the ratio of LAC in CT and PET is equal for water, muscle and air, but it is different
significantly for the bone. This shows that attenuation coefficient of air, water and muscle is associated with
electron density due to Compton interaction. But the attenuation coefficient of bone in CT energy (low energies)
is associated with photoelectric due to the existence of calcium. After the measurement of error by this method,
it was defined that for lung, bone and heat tissues, the error is 3.8%, 4.5%, 5.1%. The error is because
bone at CT energy is associated with photoelectric but
The error of this method is because of the ratio of bone
of
of water, air and muscle is associated with Compton.
in different dual energy both in low energies range.
4-4 Bilinear method:
The minimum error in this method is dedicated to bone tissue. Then, lung with 3.8%, heart and liver and
soft tissue 5.1%, 5.1%, 5.4%, respectively show the error percent. The maximum error observed in this method
was dedicated to muscle tissue with 10.3 % difference.
In the previous studies, to apply this method, to reduce the noise, high currents were used to increase the
dose of the patient. However, in this paper as in XCAT, the noise was ignored and ideal case was considered in
its design, there is no need to high current and the same result can be achieved in imaging at low current.
4-5 Dual energy method :
In this method, to obtain at 511 keV, the Compton and photoelectric effects are obtained separately and
then are combined with each other. Thus, the physical details of the tissues are considered and the acceptable
results are achieved.
The drawbacks of this method are delivery of extra dose to the patient. On the other hand, as two images are
subtracted, the noise is increased. The image mostly depends upon the effective energy. Thus, the error in
generating the effective CT energy creates error in the results. Dual energy method is theoretically accurate but
its SNR is lower compared to other mapping methods.
In this method, the lowest error is dedicated to lung and brain tissue. In bone tissue, there was a
considerable difference compared to other methods as the error of bone tissue in this method was 4.6% , 6.3%
and it is considerable compared to the error of other methods (15.4% segmentation and 15.3% hybrid).
4.6 The comparison of the study on XCAT phantom and physical and software phantom
1041
Elham Soleimani et al, 2014
Advances in Environmental Biology, 8(11) Special 2014, Pages: 1031-1042
By the comparison of the results of applying the research method on XCAT phantom and the comparison
with other results of software and physical phantoms as K2HPO4, we can say that regarding the studies
performed on K2HPO4 phantom, the bone was equal to soft tissue and as in soft tissue, the dominant effect is
Compton, the bone attenuation coefficient was E-1 and it meant that of bone was higher than the real value.
On the other hand, in the study performed on K2HPO4 phantom, there was a problem that all kinds of bones
such as a bone tissue (cortical bone) are considered. It means that bone density for different bone tissues is
considered equal but XCAT phantom is designed as in terms of the tissues properties as atomic number and
density are close to the real values. Thus, XCAT phantom showed photoelectric and Compton effects as with
direct association with the atomic number of the tissue with accurate results.
6- Future recommendation:
The current study aimed to acquire attenuation maps at 511 keV by various energy mapping methods and
the comparison of these methods with each other. As XCAT phantom is designed as error factors as noise are
eliminated, the study can be performed with the conditions in which there is noise to be close to the real
conditions. In addition, in the images from the phantom, activity effects are ignored. In the rest of this study, we
can apply attenuation map obtained for reconstruction of PET images. The comparative evaluation in this study
is in the stage of the comparison of attenuation maps and the review of the effect of attenuation map on PET
data, after the attenuation correction process can be done in the following of this study.
Also, applying the hidden anatomical data in CT images can help the performance of dispersion correction
methods.
REFERENCES
[1] Ay, M.R., 2005. Monte Carlo and Experimental Assessment of CT-Based Attenuation Correction in PET,
in Department of Radiology. University of Geneva: Geneva. p: 156.
[2] Hasegawa, B-H-, B-K. B- Stebler, Rutt, 1991. et al., A prototype high-purity Germanium detector system
with fast photon-counting circuitry for medical Imaging- Med- Phys-, 18(5): 900-909.
[3] Lang, T-F-, B-H- Hasegawa, S-C. Liew, 1992. et al., Description of a prototype Emission-transmission
computed tomography imaging system- J Nucl Med, 33(10): 1881-1887.
[4] Hasegawa, B-H-, T-F- Lang, J-K- Brown, et al., 1993. Object-specific attenuation correction of SPECT
with correlated dual-energy x-ray CT- IEEE Trans- Nucl- Sci- 40: 1242-1252.
[5] Beyer, T-, D-W- Townsend, T. Brun, 1369-1379. et al., A combined PET/CT scanner for clinical oncologyJ- Nucl- Med-, 2000-41[6] Beyer, T-, D-W- Townsend and T.M. Blodgett, 2002. Dual-modality PET/CT tomography for clinical
oncology- Q J Nucl Med, 46(1): 24-34.
[7] Townsend, D.W., J.P.J. Carney, J.T. Yap, 2004. et al., PET/CT today and tomorrow Nucl Med, 45(90010):
4S-14.
[8] Townsend, D.W. and S.R. Cherry, 1968. Combining anatomy and function: the path to true image fusion.
Eur Radiol, 2001. 11(10): 1974
[9] Townsend, D.W., 2008. Multimodality imaging of structure and function. Phys. Med. Biol., (53): 1-39.
[10] Dekemp, R.A. and C. Nahmias, 1994. Attenuation correction in PET using single photon transmission
measurment. Med. Phys., 21(6): 771-778.
[11] Zaidi, H., 2006. Recent developments and future trends in nuclear medicine instrumentation. Med. Phys.,
16: 5-17.
[12] Kinahan, P.E., B.H. Hasegawa and T. Beyer, 2003. X-ray-based attenuation correction for positron
emission tomography/computed tomography scanners. Semin Nucl Med, 33(3): 166-179.
[13] Bai, C., L. Shao, A. Da Silva, et al., 2003. A generalized model for the conversion from CT numbers to
linear attenuation coefficients. IEEE Trans. Nucl. Sci., 50(5): 1510-1515.
[14] Kinahan, P.E., A.M. Alessio and J.A. Fessler, 2006. Dual energy CT attenuation correction methods for
quantitative assessment of response to cancer therapy with PET/CT imaging. Technol Cancer Res Treat,
5(4): 319-327.
[15] Guy, M.J., I.A. Castellano-Smith, 1998. Flower, et al., DETECT-dual energy transmission estimation CTfor improved attenuation correction in SPECT and PET. IEEE Trans Nucl Sci, 45(3): 1261-1267.
[16] Beyer, T., P.E. Kinaham, D.W. Townsend, 1995. et al. The use of X-ray CT for attenuation correction of
PET data. in Proc. IEEE Nuclear Science Symposium and Medical Imaging Conference. Rome, Italy.
[17] LaCroix, K.J., B.M.W. Tsui, B.H. Hasegawa, 1994. et al., Investigation of the use of X-ray CT images for
attenuation compensation in SPECT. IEEE Trans Nucl Sci, 41(6): 2793-2799.
[18] Tang, H., C. Schreck, B. Hasegawa, et al., ECT attenuation maps from X-ray CT images. J Nucl Med,
[19] Watson, C.C., V. Rappoport, Faul, D., 2006. et al., A method for calibrating the CT based attenuation
correction of PET in human tissue. IEEE Trans Nucl Sci, 53(1): 102-107.
1042
Elham Soleimani et al, 2014
Advances in Environmental Biology, 8(11) Special 2014, Pages: 1031-1042
[20] Hasegawa, B.H., K. Iwata, K.H. Wong, 2002. et al., Dual-modality imaging of function and physiology.
Acad Radiol, 9(11): 1305-1321.
[21] Kinahan, P.E., D.W. Townsend, T. Beyer, 1998. et al., Attenuation correction for a combined 3D PET/CT
scanner. Med Phys, 25(10): 2046-2053.
[22] Kamel, E., T.F. Hany, C. Burger, 2002. et al., CT vs 68Ge attenuation correction in a combined PET/CT
system: evaluation of the effect of lowering the CT tube current. Eur J Nucl Med Mol Imaging, 29(3): 346350.
[23] Bai, C., L. Shao, 2002. Da Silva, et al. A generalized model for the conversion from CT numbers to linear
attenuation coefficients. in Proc. IEEE Nuclear Science Symposium and Medical Imaging Conference.
2002. Nov. 13-16, Norfolk, VA.
[24] Rappoport, V., J.P.J. Carney and D.W., 2004. Townsend. CT tube-voltage dependent attenuation correction
scheme for PET/CT scanners. in IEEE Nuclear Science Symposium and Medical Imaging Conference. Oct.
19-22, Rome, Italy.
[25] Ay, M.R. and H. Zaidi, 2006. Computed Tomography-based attenuation correction in neurological positron
emission tomography: evaluation of the effect of xray tube voltage on quantitative analysis. Nucl Med
Commun, 27(4): 339-346.
[26] Shirmohammad, M., M.R. Ay, S. Sarkar, H. Ghadiri, A. Rahmim, 2008. Comparative assessment of
different energy mapping methods for generation of 511keV attenuation map from CT images in PET/CT
systems, a phantom study. Proceeding of the 4 th European Conference of the international federation for
medical and biological engineering, Nov 23-27;Antwerp, Belgium.
[27] B. Teimourian, M.R., M. Ay, Shamsaei Zafarghandi, H. Ghadiri, 2009. A novel dual energy CT-based
attenuation correction method in PET/CT systems:A phantom study. Iran J Nucl Med., 17: 2.
[28] W.P. Segars, B.M.W., 2009. Tsui, MCAT to XCAT:The evolution of 4-D computerized phantoms in
medical research. Proceeding of the IEEE, Vol. 97, No.12, December
[29] Ay, M.R., M. Shirmohammad, S. Sarkar, A. Rahmim, H. Zaidi, 2010. Comparative assessment of different
energy mapping approaches in CT-Baesd attenuation correction for PET,MoI Imaging Biol,
[30] Segars, W.P., M. Mahesh, T.J. Beck, E.C. Ferry and B.M.W. Tsui, 2008. Realistic CT simulation using the
XCAT phantom, American Association of Physics in Medicine.
[31] Segars, W.P. and B.M.W. Tsui, 2009. MCAT to XCAT:The evaluation of 4-D computerized phantoms for
imaging research, proceeding of the IEEE,Vol.97,No.12.
[32] Rameshwar Prasad, Mohammad R. Ay., 2009. Osman Ratib, and Habib Zaidi, CT-based attenuation
correction on the FLEX Triumph™ preclinical PET/CT scanner, IEEE Nuclear Science Symposium
Conference Record, M09-308.