REORDER NUMBER
ELD188
NEO96NYE
Medical
PALL Posidyne NEO Filter
Priming Instructions
INDICATIONS
1
2
3
The Pall Posidyne NEO Filter is an air eliminating filter with a
0.2 µm Posidyne membrane for up to 96 hours use. It is indicated
for the removal of inadvertent particulate debris, microbial
contaminants and their associated endotoxins and entrained air
which may be found in solutions for intravenous, epidural or
subcutaneous administration.
4
The filter has a needleless ‘Y’ access site which can be used for
delivering lipid emulsions or bolus injections *(for drugs that may
be incompatible with infusion fluid, medications or preparations
that may be contraindicated for administration through the filter,
or if rapid delivery is required).
F
E
4a
4b
NOTES
■ Always swab the needleless ‘Y’ access site prior to use
■ Maximum recommended working pressure is 1500 mmHg
(approx. 30 psi, 2 bar)
■ It is recommended that the filter is changed at 96 hours
■ When a pump is used in the primary line, position filter
downstream of pump.
D
■ This product is free of natural rubber latex.
4c
4d
C
R
PRIME POSITION
■ Prime extension/administration
set to its outlet
■ Close extension/administration
set clamp if appropriate
ATTACHING THE FILTER
B
A
■ Remove protective cap (A) from
filter inlet (B)
■ Connect extension/
administration set to inlet (B)
using a twisting motion (do not
over-tighten)
■ Hold filter with outlet (E)
uppermost and the plain side
visible
■ Do not remove the protective
cap (F) from filter outlet
R
PRIMING THE FILTER
■ Open extension/administration
set clamp and prime filter to
outlet (E)
■ Close extension/administration
set clamp
■ Verify that no air bubbles are
present on ribbed (patient) side
of housing and tubing
extension, include needleless
access Y-site (D)
■ Air bubbles can be removed by
slightly opening the extension/
administration set clamp to reestablish flow and gently
tapping the filter housing
■ Close extension/administration
set clamp and filter clamp (C)
CONNECTION
■ Remove protective cap (F) and
connect to patient
■ Regulate flow in usual manner.
R
USE OF Y-SITE FOR
LIPID DELIVERY
■ Swab needleless access Y-site
(D) with preferred disinfectant
■ Connect primed lipid filter to
needleless access Y-site (D)
with a twisting motion (do not
over-tighten)
■ Deliver lipids at flow rate
prescribed
R
Ordering Information
Description
Reorder No.
Packaging
USE OF Y-SITE FOR
BOLUS ADMINISTRATION
Pall Posidyne NEO
Filter with needleless
‘Y’ access site
NEO96NYE
50 units per case
(please see indications for
further information on this
application)*
Pall Lipipor NEO Filter
for lipid emulsion used
in parental nutrition
NLF1E
50 units per case
■ Close filter clamp (C)
■ Swab needleless access Y-site
(D) with preferred disinfectant.
Please contact your local Pall office for full details of other
Pall NEO filter variants.
■ Connect standard syringe
containing suitable flushing
solution with a clockwise twisting
motion. Do not over-tighten
Medical
■ Flush line and remove syringe
using an anticlockwise twisting
motion
+44 (0)23 9230 3452
+44 (0)23 9230 3324
[email protected]
■ Connect syringe for bolus
injection and administer
■ Flush line after bolus
administration
■ Open filter clamp (C) to
re-establish flow
Europa House, Havant Street
Portsmouth PO1 3PD, England
telephone
fax
E-mail
Visit us on the Web at www.pall.com/medical
This document is not for distribution in the USA and Canada
The information provided in this literature was reviewed for accuracy at the time
of publication. Product data may be subject to change without notice.
For current information consult your local Pall distributor or contact Pall directly.
Part numbers quoted above are protected by the Copyright of Pall Europe Ltd.
, Posidyne, Lipipor and Pall are trade marks of Pall Corporation.
Filtration. Separation. Solution. is a service mark of Pall Corporation.
Pall Medical a division of Pall Europe Ltd.
© 2003, Pall Europe Limited.
Printed in England. DBD/1M/12.2003