REORDER NUMBER ELD188 NEO96NYE Medical PALL Posidyne NEO Filter Priming Instructions INDICATIONS 1 2 3 The Pall Posidyne NEO Filter is an air eliminating filter with a 0.2 µm Posidyne membrane for up to 96 hours use. It is indicated for the removal of inadvertent particulate debris, microbial contaminants and their associated endotoxins and entrained air which may be found in solutions for intravenous, epidural or subcutaneous administration. 4 The filter has a needleless ‘Y’ access site which can be used for delivering lipid emulsions or bolus injections *(for drugs that may be incompatible with infusion fluid, medications or preparations that may be contraindicated for administration through the filter, or if rapid delivery is required). F E 4a 4b NOTES ■ Always swab the needleless ‘Y’ access site prior to use ■ Maximum recommended working pressure is 1500 mmHg (approx. 30 psi, 2 bar) ■ It is recommended that the filter is changed at 96 hours ■ When a pump is used in the primary line, position filter downstream of pump. D ■ This product is free of natural rubber latex. 4c 4d C R PRIME POSITION ■ Prime extension/administration set to its outlet ■ Close extension/administration set clamp if appropriate ATTACHING THE FILTER B A ■ Remove protective cap (A) from filter inlet (B) ■ Connect extension/ administration set to inlet (B) using a twisting motion (do not over-tighten) ■ Hold filter with outlet (E) uppermost and the plain side visible ■ Do not remove the protective cap (F) from filter outlet R PRIMING THE FILTER ■ Open extension/administration set clamp and prime filter to outlet (E) ■ Close extension/administration set clamp ■ Verify that no air bubbles are present on ribbed (patient) side of housing and tubing extension, include needleless access Y-site (D) ■ Air bubbles can be removed by slightly opening the extension/ administration set clamp to reestablish flow and gently tapping the filter housing ■ Close extension/administration set clamp and filter clamp (C) CONNECTION ■ Remove protective cap (F) and connect to patient ■ Regulate flow in usual manner. R USE OF Y-SITE FOR LIPID DELIVERY ■ Swab needleless access Y-site (D) with preferred disinfectant ■ Connect primed lipid filter to needleless access Y-site (D) with a twisting motion (do not over-tighten) ■ Deliver lipids at flow rate prescribed R Ordering Information Description Reorder No. Packaging USE OF Y-SITE FOR BOLUS ADMINISTRATION Pall Posidyne NEO Filter with needleless ‘Y’ access site NEO96NYE 50 units per case (please see indications for further information on this application)* Pall Lipipor NEO Filter for lipid emulsion used in parental nutrition NLF1E 50 units per case ■ Close filter clamp (C) ■ Swab needleless access Y-site (D) with preferred disinfectant. Please contact your local Pall office for full details of other Pall NEO filter variants. ■ Connect standard syringe containing suitable flushing solution with a clockwise twisting motion. Do not over-tighten Medical ■ Flush line and remove syringe using an anticlockwise twisting motion +44 (0)23 9230 3452 +44 (0)23 9230 3324 [email protected] ■ Connect syringe for bolus injection and administer ■ Flush line after bolus administration ■ Open filter clamp (C) to re-establish flow Europa House, Havant Street Portsmouth PO1 3PD, England telephone fax E-mail Visit us on the Web at www.pall.com/medical This document is not for distribution in the USA and Canada The information provided in this literature was reviewed for accuracy at the time of publication. Product data may be subject to change without notice. For current information consult your local Pall distributor or contact Pall directly. Part numbers quoted above are protected by the Copyright of Pall Europe Ltd. , Posidyne, Lipipor and Pall are trade marks of Pall Corporation. Filtration. Separation. Solution. is a service mark of Pall Corporation. Pall Medical a division of Pall Europe Ltd. © 2003, Pall Europe Limited. Printed in England. DBD/1M/12.2003